CUSTOMER : M/S FONTUS WATER PVT. LTD.

D-4, FREEDOM FIGHTER ENCLAVE, NEB SARAI, IGNOU ROAD NEW DELHI
Calibration Certificate of : DIGITAL WEIGH BALANCE Certificate No. – HTCS/03170
Make : INTERFACE Service Request No. : HTCS/03170
Range : 0-300 Kg Service Request Date : 25.04.2012
L.C. : 50 gm Date of Calibration : 26.04.2012
I.D. No. : FWPL/WB-02 Suggested due Date : 26.04.2013
Sl. No. : 9110064 Page : 1 of 1
: :
Visual Inspection OK Calibration Method WI-28
:
Environmental Condition of Lab.- : Uncertainty of Measurement : ±5.8 g
Temperature : (20+/-1) Deg. C (At 95% Confidence Level,
Humidity (50+/-10)% RH k=1.96)

DETAILS OF STANDARD EQUIPMENTS USED FOR CALIBRATION


Name of the Instrument : STANDARD WEIGHTS : :
Make : WEIGH INDIA : :
Certificate No. : WMCL/M/2008/25 : :
Sl. No./ID No. : HTCS/SW/01 : :
: 18/11/2012 : :
Calibration Due Date
: WEIGHTRONICS : :
Traceable By

Calibration Results
Sl. Standard Observed Value Sl. Standard Observed Value
No. Value (UUC) No. Value (UUC)
01 01.00 Kg 01.00 Kg 08 120.00 Kg 120.10 Kg
02 02.00 Kg 02.01 Kg 09 150.00 Kg 150.20 kg
03 05.00 Kg 05.02 Kg 10 180.00 Kg 180.50 Kg
04 10.00 Kg 10.02 Kg
05 20.00 Kg 20.05 Kg
06 50.00 Kg 50.05 Kg
07 100.00 Kg 100.10 Kg

Remarks:-

Notes :-
1. This calibration results reported in this certificate are valid at the time of & under the stated conditions of measurements.
2. This report should not be reproduced except in full without our prior permission in writing.
3. This certificate relates only to the item calibrated.
4. Format No.-F01(5.10/01).

Calibrated By Approved By