Beruflich Dokumente
Kultur Dokumente
CHROMATOGRAPH
TRAINING COURSE
Applies to both
Daniel Danalyzer Gas Chromatograph
Rosemount Analytical Gas Chromatograph
JULY 2006
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
GAS CHROMATOGRAPH
TRAINING MANUAL
NOTICE
DANIEL MEASUREMENT AND CONTROL, INC. AND ROSEMOUNT ANALYTICAL, INC. (COLLECTIVELY,
“SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR
OMISSIONS FROM THIS MANUAL. SELLER MAKES NO WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING THE IMPLIED W ARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE W ITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL SELLER BE LIABLE FOR ANY
SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION,
LOSS OF PROFITS, ETC.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND
MAY BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES.
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND
W HILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE
CONSTRUED AS W ARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE
PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. W E RESERVE THE
RIGHT TO M ODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
SELLER DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY
PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND M AINTENANCE OF ANY SELLER
PRODUCT REMAINS SOLELY W ITH THE PURCHASER AND END-USER. DANIEL AND THE DANIEL LOGO
ARE REGISTERED TRADEMARKS OF DANIEL INDUSTRIES, INC. THE ROSEMOUNT AND ROSEMOUNT
ANALYTICAL LOGO THE ARE REGISTERED TRADEMARKS OF ROSEMOUNT ANALYTICAL, INC. THE
EMERSON LOGO IS A TRADEMARK AND SERVICE MARK OF EMERSON ELECTRIC CO.
COPYRIGHT © 2006
BY DANIEL MEASUREMENT AND CONTROL, INC.,
HOUSTON, TEXAS,
U.S.A.
PREFACE i
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
WARRANTY
1. LIMITED W ARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise
expressly provided herein, Daniel Measurement and Control, Inc. and Rosemount Analytical, Inc., (collectively “Seller”)
warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods
manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use
and care until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date
of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first.
Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty
extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a
reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer discovers any
warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option,
promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of
manufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the
defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal
wear and usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper installation,
modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited
warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any costs or charges incurred by Buyer
or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs
of dismantling, reinstallation and freight and the time and expenses of Seller's personnel for site travel and diagnosis
under this warranty clause shall be borne by Buyer unless accepted in writing by Seller. Goods repaired and parts
replaced during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90)
days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a
writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement,
THERE ARE NO REPRESENTATIONS OR W ARRANTIES OF ANY KIND, EXPRESSED OR IMPLIED, AS TO
MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER W ITH RESPECT
TO ANY OF THE GOODS OR SERVICES. It is understood that corrosion or erosion of materials is not covered
by our guarantee.
2. LIM ITATION OF REM EDY AND LIABILITY: SELLER SHALL NOT BE LIABLE FOR DAMAGES
CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF
W ARRANTY HEREUNDER SHALL BE LIMITED TO REPAIR, CORRECTION, REPLACEMENT OR REFUND
OF PURCHASE PRICE UNDER THE LIMITED W ARRANTY CLAUSE IN SECTION 1 HEREIN. IN NO EVENT,
REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (W HETHER BASED IN CONTRACT,
INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERW ISE), SHALL SELLER'S
LIABILITY TO B UYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC
GOODS MANUFACTURED OR SERVICES PROVIDED BY SELLER GIVING RISE TO THE CLAIM OR CAUSE
OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS
CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAM AGES. THE
TERM "CONSEQUENTIAL DAMAGES" SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF
ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE AND COST OF CAPITAL.
ii PREFACE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
DATE:________________
Please print your name as you would like it to appear on your certificate. Also please give a brief
description of what you expect from this course.
NAME: ____________________________________________
COMPANY: ________________________________________
POSITION: _________________________________________
COMMENTS:____________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
PREFACE iii
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
iv PREFACE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
TABLE OF CONTENTS
COURSE OUTLINE
CHROMATOGRAPHIC THEORY
MAINTENANCE PRACTICES
MAINTENANCE CHECKLIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
TROUBLESHOOTING CHECKLIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
FUNCTION CODE DICTIONARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
DEW POINT OF VAPOR SAMPLES (Vapor Pressure of Components) . . . . . . . . . . . . . . . . . 43
VAPOR PRESSURE OF COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
THERMAL CONDUCTIVITY EXAMPLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
TABLE OF CONTENTS v
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
SAMPLE SYSTEM
APPENDIX
COURSE EVALUATION
vi TABLE OF CONTENTS
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
I. INTRODUCTION
A. Interviews
1. Evaluation of Students' level of experience, technical background,
and specific expectations for training
B. Chromatographic Equipment
1. Introduction to major parts, components, peripherals and functions
performed
A. Detector
C. Chromatographic Principles
1. Carrier Gas(es)
2. Temperature Control
(Day 2- 8 hours)
III. DOCUMENTATION
A. Drawings
B. Specifications
C. Parameter Sheets
D. Chromatograms
E. Reference Manual
COURSE OUTLINE 1
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
A. Operational Modes
1. Keyboard Procedures
2. Prompts
V. CALCULATIONS
VI. ALARMS
A. Functions of LED(s)
B. Alarm Conditions
2 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
(Days 3 and 4)
VIII. MAINTENANCE
A. Routine Checks
B. Pre-Amp Balance
F. Valve Overhaul
(Day 5 - 4 hours)
IX. TROUBLESHOOTING
COURSE OUTLINE 3
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
4 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
Chromatography was first employed in 1905 to separate mixtures of gases and vapors. The
following year, Tswett1 obtained discrete colored bands of plant pigments on a chromatographic
column. He coined the term "chromatography" (color writing) which in the light of present methods
is a misnomer.
Based on the work of the Englishmen Martin and Synge for which they were later awarded the Nobel
Prize, James and Martin introduced gas-liquid chromatography in 1952.
The sensitivity, speed, accuracy, and simplicity of this method for the separation, identification, and
measurement of volatile compounds has resulted in a phenomenal growth.
Gas chromatographs were available commercially about 1955. In this relatively short time, this
technique has become by far the most widely used analysis method in the world.
BASIC PRINCIPLES
Gas chromatography employs elements for injecting a sample, separating the components of the
sample and for identifying and measuring the separated components. It is used to analyze gases,
liquids, and solids in their vapor phase.
1. Tswett, M., Ber. deut. botan. Ges. 24, 316, 384 (1906)
CHROMATOGRAPHIC THEORY 5
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
The analysis method is a batch type procedure. Discrete samples are injected separately. For best
results, the sample should be injected as quickly as possible.
Figure 1
6 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
• Carrier gas system - for transporting the component through the column at a constant
flow rate.
• Sampling apparatus - for measuring and injecting the sample into the carrier gas ahead
of the column.
Although the fundamental principles are the same, the physical differences between laboratory and
process chromatographs are extensive. We shall discuss the common fundamental principles first
and the aspects peculiar to the process analyzer later.
We shall now discuss briefly the elements of the chromatograph in the order of their importance.
CHROMATOGRAPHIC THEORY 7
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
CHROMATOGRAPHIC COLUMN
The heart of the chromatograph is the column. It is here that the components in the sample are
separated so they may be identified and measured. The column is a length of tubing (1/16" - 1/4"
diameter up to 30 feet long) packed with particles.
Gas chromatography is a technique for separating volatile substances by flowing a gas stream over
a stationary phase. The column packing forms the stationary phase.
If the stationary phase is the particle, we speak of gas solid chromatography (GSC) and the
separation of the sample components is achieved by selective adsorption of the components on the
surface of the particles.
If the stationary phase is a liquid coating on the particle, we speak of gas liquid chromatography
(GLC) and the separation of the sample components is the result of the partitioning of each
component between its vapor phase and a non-volatile solvent (stationary phase) which is coated on
the inert particles (solid support).
The solvent (stationary phase) retains the sample components, according to their distribution
(Partition) coefficients, until they form separate bands in the carrier gas. These component bands
leave the column in the gas stream and are recorded as a function of time by the detector and
recording system. Figure 2 illustrates the formation of a chromatogram, the record showing three
peaks at the right of the diagram.
8 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
The following paragraphs will attempt to elaborate on the mechanism by which the separation of the
gaseous components in a sample is achieved.
Figure 2
CHROMATOGRAPHIC THEORY 9
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
First perhaps, a crude analogy will be helpful. Refer to Figure 3. Imagine the sketch represents the
following situation.
Figure 3
A groups of animals, elephants, deer, and foxes, are peacefully at home in a large area at the left
bounded by sheer rock walls, when a large grass fire whipped by winds blowing from left to right
bears down upon them and stampedes them toward the densely forested, but more restricted area at
the center of the diagram. Here, as indicated, the trees are closely spaced and the ground is muddy.
The large and heavy elephants encounter considerable difficulty in penetrating and passing through
the restricted area. The smaller and lighter deer encounter less difficulty and the small, light and
agile foxes even less trouble. So as all three are kept under pressure by the fire to keep moving to
the right, the foxes move through the restricted area fastest, the deer next, and finally the elephants
bring up the rear. The conditions as stated have served to separate the animals into three distinct
groups and as they emerge into the free world at the right, we can easily identify each group as well
as count them. If we should make a chart or graph of the animals as they emerge as a function of
time, we would have the equivalent of a chromatogram.
10 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
In Figure 4, we can easily relate some of the basic elements of a chromatogram to the drama just
played by the animals.
Figure 4
CHROMATOGRAPHIC THEORY 11
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
Getting a little closer to reality, the actual mechanism leading to the separation may be described as
follows:
A few micro liters of the sample to be analyzed is injected into a flowing stream of inert carrier gas
that is being swept through the column under carefully regulated conditions. During the transit of
the sample through the packed column, its components become separated according to their
distribution (partition) coefficients.
Figure 5
A closed container is partially filled with a solvent A. When solute B in the form of a gas is
introduced in a small amount above the liquid, an equilibrium will be quickly reached between the
fraction of B remaining in the vapor state and the fraction that goes into solution in liquid A.
At a particular temperature the ratio of the amount of solute per unit volume of liquid phase to the
amount of solute per unit volume of gaseous phase is a constant, or:
C liq
K(t) = S)))) distribution (partition) coefficient
C gas
12 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
If some of the gas should be removed from the space above the liquid, some of the gas dissolved in
the liquid will come out of the solution in order to maintain K constant.
Imagine now that Figure 6 is a magnified particle of support material in the column coated with the
liquid (stationary) phase and that this particle in the column is near the entrance of the column.
Figure 6
CHROMATOGRAPHIC THEORY 13
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
As the sample being swept through first encounters this particle, a fraction of each component in it
will be dissolved in the stationary phase according to its distribution coefficient. As the carrier
continues to sweep by the particle, the least soluble components will leave the stationary phase first
in their attempt to maintain the value of their distribution coefficient. Therefore, as they sweep
through the column, the components with the lower distribution coefficients will emerge first and
those with the highest coefficients last.
The value of K is high when most of a substance is retained in the liquid phase. This means the
substance moves slowly down the column because only a small fraction will be in the carrier gas at
any given time. Thus, separation between two components is possible, if their distribution
coefficients are dissimilar.
The time that elapses between the start of the analysis and the formation of the peak maxima
corresponding to each component in the sample is called "retention time" for the respective peak
(Figure 7).
Figure 7
14 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
Under fixed conditions, retention times are highly reproducible and hence, are used to identify the
sample components.
• Volume flow rate of the carrier gas - an increase in flow rate will decrease retention
times.
• Length of the column - an increase in length will increase retention times and resolution
(Figure 9).
Factors that affect the ability of column to efficiently separate the components in a mixture:
• solid support
• type and amount of liquid (stationary) phase
• method of packing
• length and temperature of the column
• size and composition of the sample
• mode of injection of the sample
• type and rate of flow of the carrier
CHROMATOGRAPHIC THEORY 15
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
Figure 8
16 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
Figure 9
CHROMATOGRAPHIC THEORY 17
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
Solid Support
As stated previously, the solid support should expose a very large area of the stationary phase to the
sample in order that gas-liquid equilibrium may be achieved rapidly.
A good packing material should have a specific area of 2 or 5 meters2/gram (very porous) and
uniform particle size.
Each 100 parts by weight of the support material should have from 5 to 30 parts of the stationary
phase absorbed upon it.
• A porous structure with uniform pore diameter in the range of 10 microns or less (.1mm).
Diatomaceous earth, teflon, crushed firebrick, and celite are often used as support materials.
18 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
Stationary Phase
Good solvent efficiency results in greater separation of peaks (greater ratios of retention times) or
better resolution, (Figure 10).
Figure 10
CHROMATOGRAPHIC THEORY 19
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
Resolution refers to the true separation of two consecutive peaks. It relates to the narrowness of
peaks and the separation between maxima (Figure 11).
Figure 11
As a general rule, the liquid (stationary) phase should be as similar in chemical class to the
components of a mixture to be separated as possible. There should be no possibility of reactivity
between the chemical groups of the stationary phase and those in the components of the analytical
mixture.
20 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
One of the major advantages of GC over the distillation is the ability to use solvents selected to
provide the best separation of the components in the sample. Thus components with same boiling
point which are impossible to separate by distillation may be separated on a GC.
There are several interaction forces which can aid in the GC separation. These are atomic and
molecular forces of various types. Without attempting to describe or explain them individually, it
may be said that their combined effects are expressed by the partition coefficient K.
• Good absolute for sample components - if solubility is low, components elute rapidly, and
separation is poor.
• Non-volatile - vapor pressure .01 to .1mm at operating temperature for reasonable column
life.
• Thermally stable.
CHROMATOGRAPHIC THEORY 21
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
Column Temperature
The partition coefficient K decreases with increasing temperature. The fraction of the solute in the
gas phase will increase with rising temperature and hence, the elution time will decrease, which also
decreases the width of the peaks. This results in decreased separation since it is the liquid phase
which performs the separation. So, to achieve better separation, lower temperatures should be used,
but this will result in longer analysis times (Figure 12).
Figure 12
22 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
Carrier Gas
A high pressure gas cylinder serves as the source of carrier gas. A pressure regulator is used to
assure a uniform pressure to the column inlet and thereby a constant rate of gas flow. At a given
temperature, this constant rate of gas flow will elute components at a characteristic time (the
retention time).
Commonly used gases are hydrogen, helium and nitrogen. The carrier gas should be:
Column efficiency depends upon choosing the proper linear gas velocity.
The optimum flow rate can be easily determined experimentally by making a simple van Deemter
plot of HETP vs. linear gas velocity. HETP means Height Equivalent of a Theoretical Plate. Plates
as referred to in the term are somewhat analogous to those in a distillation column.
The number of theoretical plates (N) in a column are determined from a chromatogram as indicated
below (Figure 13).
Figure 13
CHROMATOGRAPHIC THEORY 23
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
The Height Equivalent of a Theoretical Plate is then simply the length of the column divided by N.
HETP = L/N
The most efficient flow rate is the one giving the minimum value of HETP. In other words, it is the
flow rate permitting the use of the shortest or most efficient column, for L = HETP x N. Where
HETP is minimum, so also will be L (Figure 14).
Figure 14
24 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
From the foregoing, it is apparent that a great many things affect the efficiency of a chromatographic
column. Its ability to produce sharp, well separated peaks in the chromatogram in the minimum
amount of time is a function of:
Detectors
After a chromatographic column has successfully separated the components in a sample, the problem
of detecting the components in the carrier gas for purposes of identification and quantitative
measurement remains.
To solve this problem, several different types of detectors have been devised. The detectors most
commonly used can be separated into two categories — concentration and mass flow rate.
The concentration detectors sense a bulk property, e.g. thermal conductivity, of both the carrier gas
and component. Thus, there must be a great difference in the value of this property between the
carrier gas and the sample components. An example of this type is the thermal conductivity detector.
Helium as a carrier gas has high thermal conductivity. Sample components have low thermal
conductivity. The change in thermal conductivity of the gas within a detector indicates the presence
of a component.
The mass flow rate detectors sense a specific property of the sample components and not of the
carrier gas. For example, in the flame ionization detector, the hydrogen flame ionizes the sample
components but not the carrier gas. Thus, the response is proportional only to the flow rate of
component molecules as they are ionized in the flame.
The chromatogram produced by a detector consists of a series of peaks, each of which corresponds
to a different component. The area under each peak is proportional to the total mass of that
component. Hence, the chromatographer may calculate weight percent compositions from area ratios
on the chromatogram. With thermal conductivity detectors, the peak area is inversely proportional
to carrier gas flow rate, so the flow rate must be kept constant for accurate quantitative analysis.
CHROMATOGRAPHIC THEORY 25
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
For detectors responding to mass flow rate, such as flame ionization detectors, the peak area is
independent of carrier gas flow rate.
Generally speaking, the usefulness of a detector regardless of the type, may be evaluated in terms
of the following characteristics:
• Selectivity
• Sensitivity
• Response
• Noise
• Linear Range
Secondly, to these things, it should be simple, rugged and relatively insensitive to changes in flow
rate and temperature.
The Flame Ionization and Thermal Conductivity detectors will now be discussed in more detail as
examples of the two types.
Ionization detectors operate on the principle that the electrical conductivity of a gas is directly
proportional to the concentration of charged particles within the gas. In the FID, effluent gas from
the column is mixed with hydrogen and burned in air or oxygen. Ions formed in the flame enter the
electrode gap which causes current to flow across the gap and through a series resistor. The resulting
voltage drop can then be amplified and recorded. The background current due to the carrier is
normally canceled out so that the effects of the sample components only are recorded.
The flame detector responds to all but a relatively short list of compounds (inorganics) which include
water, carbon disulphide, and also air. The lack of response to air and water make the FID especially
suitable for the analysis of air pollutants or aqueous samples.
26 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
FID performance depends on the proper choice of auxiliary gas (hydrogen and oxygen) flow rates.
Sensitivity and stability are affected and such detectors should be calibrated to determine the flow
rate that gives the best response.
Figure 15
CHROMATOGRAPHIC THEORY 27
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
Since the Gas Chromatograph (GC) uses a TC detector, the following paragraphs will elaborate on
this method of measurement.
The TC detector is widely used primarily because it responds to all types of compounds and can
sense trace concentrations as well as very high concentrations.
The TC detector operates on the principle that a hot body will lose heat at a rate proportional to the
composition of the surrounding gas. The rate of heat loss is then used as a measure of gas
composition.
The detector usually takes the form of a Wheatstone Bridge where the four arms of the bridge are
made of very fine wires or filaments of platinum or a tungsten alloy. The bridge current from a well-
regulated power supply is high enough to heat the bridge filaments to several hundred degrees
fahrenheit. The bridge wires are mounted on suitable supports and inserted in cavities in a small cell
block which are interconnected by passages through which the sample and the reference gases are
passed. Such an arrangement minimizes the effect of temperature and flow disturbances (Figure
16).
Figure 16
28 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
Most systems use four wire bridges where two legs are exposed to the sample and two to the
reference material. In some cases, two wire bridges are used and then only one heat sensitive
element responds to the sample and one to the reference.
When sample molecules mixed with the carrier (reference) gas pass by one pair of the bridge
elements, the rate of heat loss from the elements will be reduced and their resistance will increase,
thereby unbalancing the bridge. This bridge unbalance is usually impressed across an attenuator
network where either all, or some known fraction of it is applied to the input of a recorder and the
peak recorded.
The principle of operation is that the ability to conduct heat from a filament is a function of the
molecular weight of the gas surrounding it.
Figure 17 illustrates an arrangement for producing a chromatogram with a TC detector and an auto
ranging amplifier.
CHROMATOGRAPHIC THEORY 29
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
Figure 17
30 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
The four temperature elements of the thermal conductivity bridge are R1, R2, R3, R4. These four
elements are constructed to be as nearly identical as possible and they are mounted in the cavities
of the cell block to provide as nearly identical thermal environments as possible. The reasons for
matching elements and environments are to make the unit as insensitive to operating and
environmental changes as possible. A balanced Wheatstone Bridge tends to be self-regulating.
Elements R1 and R3 are mounted in the cell cavity through which pure carrier gas will pass and
elements R2 and R4 will be exposed to the effluent gas from the column. Electrical adjustments M
and N are provided for bridge balance and bridge current adjustment respectively.
Chromatogram baseline reference is obtained by passing pure carrier gas through both passage ways
in the cell simultaneously.
Signal Attenuation
Most detectors are shunted with an attenuator so that signal input to the recorder will result in a
readable peak even though there may be a very large differences in the absolute value of the bridge
output signals for different sample components. For example, the sample being analyzed may
consist of three principle components as follows: Component A 90%, Component B 9.9%, and
Component C .1%. In order that Component C produce a good measurable chromatogram peak
height, say of at least 10% of full scale on the recorder, the attenuator switch S is set to apply the full
bridge unbalance voltage to the recorder input terminals.
Now, peak B passes through the detector and in order to prevent this peak height from exceeding the
full scale span of the recorder, the attenuator must be reset so that only a known fraction of the
bridge output for this peak is applied to the recorder input. As Component A comes through, in all
probability, the attenuator will have to be reset again to accommodate the very large bridge output
it will produce.
The attenuator switch may be manually operated based on known sample composition or from
observation of peak heights as they are being recorded, or its manipulation may be programmed into
an automatic system that will initiate many other operations such as column switching, purging,
backflushing the system, etc., which are a part of overall system operation.
CHROMATOGRAPHIC THEORY 31
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
It may be noted here that this is a very demanding application for conventional analog recorders.
This is true for four principle reasons.
• To the recorder, the attenuator is a signal source with widely varying internal resistance.
This may affect measurement sensitivity unless the recorder input impedance is made quite
high.
• Narrow recorder spans of one mv or less are needed to record the peaks of trace components.
This demands high measurement sensitivity.
• High recorder pen speeds, one second for full scale travel or faster are required to record the
tops of the high peaks without attenuation. This requires high servo power.
• Good input filtering is needed to minimize the effects of transverse stray voltages which may
originate in the bridge power supply, for example, and which otherwise might affect the
performance of the recorder.
For a more complete discussion of the recorder in the chromatograph, the reader is referred to: The
Chromatography Slave - The Recorder, by Rob B. Bonsall, Lab and Test Markets, Honeywell Inc.,
Fort Washington, Pennsylvania.
SAMPLE INJECTION
The size and mode of injection of the sample is very important in the design of a good
chromatograph.
Since retention time identifies the components in the sample, the time of sample injection should be
established accurately.
The rate of sample injection should be rapid so that the total sample may be considered to have been
introduced into the system at one time and not spread out over an extended period. The width of the
narrowest peak in a chromatogram can never be less than the time required to introduce the sample.
The size of the sample should be small. This permits equilibrium to be reached rapidly between the
sample and the stationary phase in the column. In laboratory chromatographs liquid sample injection
is frequently done with a suitably prepared microliter syringe. With it a few microliters (1 to 5) are
quickly injected into the carrier gas system at the input end of the column. Not only must the
injection time be short, but the sample size must be known and reproducible so that repeat runs and
check runs may be reliably compared.
32 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
For gas sample injection, very precise and elaborate sample injection valves, as well as gas tight
syringes, are used for introducing samples.
Reproducibility of sample size is of the order of .5% with valves, whereas about 2% reproducibility
is typical with syringes.
In gas chromatography, data processing and presentation involves the methods used in handling the
detector output to provide identification and amount of each of the sample components. These are,
respectively, qualitative analysis and quantitative analysis. The various techniques utilized to
accomplish qualitative and quantitative analysis will now be discussed.
Qualitative Analysis
Gas chromatography has a fantastic ability to separate components. One of the problems associated
with the technique is to positively identify all of the separated peaks emerging from the column.
Ability to identify peaks has not progressed as fast as ability to produce them.
The literature describes several methods for identifying sample components from their
chromatograms and most deal with retention data in various ways. Identification is based on
comparison of the retention time of the unknown component with that obtained from a known
compound analyzed under identical conditions. With constant carrier gas flow rate and constant
column temperature, retention time is peculiar to the compound for a given column. With well
designed equipment, retention data are quite reproducible. Confirmation of identifications where
retention data is difficult to interpret may be made using infrared, mass spectrographic, magnetic
resonance or other methods.
Quantitative Analysis
Component concentration is determined from peak heights or area or from other measurements
directly related to peak area.
CHROMATOGRAPHIC THEORY 33
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
High accuracies can be obtained providing the proper techniques are employed. Relating to accuracy
is the care with which the analyst recognizes and accounts for the many sources of error present in
the analyzing system. Some of these error sources are:
1. Does the sampling device and your sampling technique actually take the sample you wish
to analyze?
2. Does it actually put the sample into the chromatograph and without any changes in its
composition?
4. Does the detector respond properly and consistently? Is it stable and sensitive?
5. Does the recorder have the capabilities to faithfully record the output of the detector in the
face of the conditions as listed previously?
6. Is the method selected for measuring and/or calculating peak area valid?
If these questions have been adequately considered and their effects taken into consideration, then
the relative areas of the peaks are a good measure of the composition of the sample.
Some of the ways (with advantages and disadvantages) of relating peak shape to sample
concentration are:
1. Peak height; peak heights may be affected by sample size as well as baseline drift.
2. Planimetry; peak area measured with a hand planimeter. Time consuming and subject to
individual variations.
3. Height x width at half height. This assumes symmetrical peaks of triangular shape and
reasonable width. Technique is rapid and simple (Figure 18).
.
Figure 18
34 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
4. Triangulation; height measured from the baseline to the intersection of two tangents (Figure
19).
Figure 19
CHROMATOGRAPHIC THEORY 35
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
5. Cutting and weighing paper; peaks are actually cut out of paper and weighed. Accurate, but
very time consuming and chromatogram is destroyed.
6. Integration
a. Ball and disc integrator driven directly from the recorder with "counts" being
recorded as the pen in recording a peak deviates from the baseline.
Area Peak A
Area Percentage = S)))))))))))) x 100
Area all Peaks
36 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
SUMMARY
Because of somewhat complicated effects that changes in the operating variables have on qualitative
and quantitative aspects of the data, an outline of these effects is given as follows. Please note that
these are the normally expected effects for the respective changes. Under certain conditions this may
not be valid, but those conditions would be considered abnormal. Also note that if the entire
chromatogram was properly processed by a computer, many of these changes would have no effect
on achieving accurate qualitative and quantitative analysis. However, since the primary aim of this
introductory primer is process gas chromatography, the effect on "peak height" by changing variables
is emphasized.
A. At a constant retention time, the peak height of a component will change with a change in
any of the following:
4. Detector temperature.
CHROMATOGRAPHIC THEORY 37
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
B. Retention time of a component will change with a change in any of the following:
3. A large increase in sample size will result in a noticeable increase in retention time.
4. A decrease in column liquid phase (caused by operating column over the temperature
limit) will result in decreased retention time.
C. At constant carrier flow rate, peak height increases as retention time decreases.
D. In TC detectors, peak height (sensitivity) increases as carrier flow rate decreases. This is
caused by decrease in the specific heat effect and increase in gas diffusion of component
molecules to filament area.
38 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
MAINTENANCE CHECKLIST
Date Performed ___ /___ / ___
Sales Order No. ____________
BI-MONTHLY AS FOUND AS LEFT NOMINAL
HELIUM CARRIER
CYLINDER
Cylinder Pressure Reading (High) ____ psig ____ psig
Cylinder Pressure Outlet Reading ____ psig ____ psig 110-115 psig
SAMPLE SYSTEM
Sample Line Pressure(s) (1)___ psig ____ psig 5-15 psig
(2)___ psig ____ psig 5-15 psig
(3)___ psig ____ psig 5-15 psig
* Adjusting carrier pressure panel regulator will result in retention time changes. Do not adjust
without consulting Customer Service.
MAINTENANCE PRACTICES 39
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
ANALYZER
SAMPLE SYSTEM
GC GRI PAZ (12-bit AD) Value (12-bit AD) PAZ (16-bit AD) Value (16-bit AD)
1 0.0 to 0.0 600 to 800 0 to 4095 4800 to 6400 -32767 to 32767
2 0.8 to 1.1 600 to 800 0 to 4095 4800 to 6400 -32767 to 32767
3 0.8 to 1.1 600 to 800 0 to 4095 4800 to 6400 -32767 to 32767
4 0.8 to 1.1 1150 to 1500 0 to 4095 9200 to 12000 -32767 to 32767
40 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
CHROMATOGRAM
(SPECTRUM “ON”)
Check baseline
Check component values on report
Number of peaks
Retention times
Date and file
TEMPERATURE
(see Section 5.5.7)
Detector Temperature °C or °C or
Thermocouple Wire #1 (Type J) ____ mV ____ mV 78-83°C
Heater Block Temperature °C or °C or
Thermocouple Wire #2 (Type J) ____ mV ____ mV 78-83°C
MAINTENANCE PRACTICES 41
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
42 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
For multi component vapor mixtures, calculation of dew point can be performed according to the
following formula:
(P) (Y)
X = ————
VP
Step 2. Find the vapor pressure of each component at the above temperature.
Step 4. Calculate X for each component. If the sum of the X's for all the components equal
to 1.00 or more, there will be some dropout of liquid. If the sum of the X's for all
the components equal to 0.95 or less, the sample will remain vapor.
MAINTENANCE PRACTICES 43
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
NC 4 7.4 12 18 27 39 54 74 4.6
44 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
MAINTENANCE PRACTICES 45
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
46 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
SAMPLE SYSTEM 47
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
48 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
SAMPLE SYSTEM 49
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
50 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
SAMPLE SYSTEM 51
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
52 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
TUBING
1/8 x 0.030 wall 0.65 3.9 sec
1/8 x 0.020 wall 1.1 6.6 sec
1/4 x 0.030 wall 5.5 33.0 sec
3/8 x 0.035 wall 14 1.4 min
½ x 0.035 wall 29 2.9 min
5/8 x 0.049 wall 43 4.3 min
3/4 x 0.049 wall 66 6.6 min
PIPE
1/8 Sch 40 11 1.1 min
1/4 Sch 40 21 2.1 min
3/8 Sch 40 38 3.8 min
½ Sch 40 60 6.0 min
3/4 Sch 40 105 10.5 min
1 Sch 40 170 17.0 min
SAMPLE SYSTEM 53
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
54 TRAINING COURSE
JU L 2006 GAS CHROMATOGRAPH TRAINING MANUAL
56 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
58 TRAINING COURSE
GAS CHROMATOGRAPH TRAINING MANUAL JU L 2006
This appendix provides a sample standard Component Data Table as well as a table of the
ISO-related components.
All values depend on a base pressure of 14.73 PSIA and a base temperature of 60 oF
(15.56 oC).
D aniel
C om ponent M ol R eid R el D ens R el D ens L b/gal GPM G ross D ry N et D ry AGA8 Sim
N am e Wt V apor G as L iquid F actor BTU BTU N o. 2251
I.D . N o.
C arbon M onoxide 28.01 0.0 0.9671 0.8010 6.6800 0.0 321.2 321.2 9 15
C arbon D ioxide 44.01 0.0 1.5196 0.82275 6.8596 0.16946 0.0 0.0 3 17
2,3-D imethbutan 86.18 7.4040 2.9753 0.6664 5.5560 0.4096 4756.0 4403.1 15 *
2,2-D imethpenta 100.21 3.4920 3.4596 0.6782 5.6540 0.4682 5494.6 5091.4 16 *
2,4-D imethpenta 100.21 3.2920 3.4596 0.6773 5.6470 0.4686 5499.4 5096.0 16 *
3,3-D imethpenta 100.2 2.7730 3.4596 0.6976 5.8160 0.4550 5501.5 5098.2 16 *
D aniel
C om ponent M ol R eid R el D ens R el D ens L b/gal GPM G ross D ry N et D ry AGA8 Sim
N am e Wt V apor G as L iquid F actor BTU BTU N o. 2251
I.D . N o.
Ethyl alchohol 46.07 2.3000 1.5906 0.7940 6.6200 0.1839 1602.8 1451.5 0 *
Ethyl benzene 106.17 0.3710 3.6655 0.8718 7.2680 0.3858 5234.3 4982.0 17 *
Ethylene O xide 44.05 0.0 1.4900 0.0 0.0 0.0 1459.4 1410.2 0 36
3-Ethyl pentane 100.21 2.0120 3.4596 0.7028 5.8590 0.4517 5513.4 5110.1 16 *
i-Propyl benzene 120.19 0.1880 4.1498 0.8663 7.2230 0.4396 5976.6 5674.0 18 *
i-O ctane 114.23 1.7080 3.9439 0.6962 5.8040 0.5199 6246.1 5792.2 17 *
M ethyl alcohol 32.04 4.6300 1.1063 0.7960 6.6400 0.1275 868.7 767.9 0 *
2-M ethylhexane 100.21 2.2710 3.4596 0.6830 5.6940 0.4647 5507.3 5104.0 16 *
3-M ethylhexane 100.21 2.1300 3.4596 0.6917 5.7670 0.4589 5511.3 5107.8 16 *
m-X ylene 106.17 0.3260 3.6655 0.8687 7.2430 0.3871 5219.9 4967.8 17 *
n-D ecane 142.285 0.05889 4.9127 0.73404 6.12 0.61405 7760.8 7206.63 19 *
n-N onane 128.258 0.1796 4.4284 0.72185 6.0183 0.56288 7012.386 6508.02 18 38
n-O ctane 114.231 0.5366 3.9441 0.70696 5.8942 0.51188 6263.26 5809.41 17 20
D aniel
C om ponent M ol R eid R el D ens R el D ens L b/gal GPM G ross D ry N et D ry AGA8 Sim
N am e Wt V apor G as L iquid F actor BTU BTU N o. 2251
I.D . N o.
Sulfur D ioxide 64.06 88.000 2.2117 1.3970 11.650 0.1453 0.0 0.0 3 43
A. Chromatographic Theory Q Q Q
B. Chemistry Q Q Q
C. Programming Q Q Q
D. Sample System Q Q Q
E. Sampling Q Q Q
F. Blueprint Reading Q Q Q
G. Total System Operation Q Q Q
A. Subject Knowledge Q Q Q
B. Clarity of Presentation Q Q Q
C. Course Organization Q Q Q
D. Courtesy Q Q Q
Other Comments:
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
7. Course Material: Good Fair Poor
A. Manual Q Q Q
B. Drawings Q Q Q
Comments:
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________