There are multiple blending methods, ranging from the simplest (and widely
used) splash blending, to ratio or proportional blending (usually done at a ter-
QinetiQ North America
minal rack). The FluidScan device can be used to monitor blend verification
www.SpectroInc.com at different tanks, or rack sections. Page 1
FSC-08/ 4 September 2009
QinetiQ North America
Application Note
Capability to Identify Biodiesel Content in Diesel Fuel
Controlled blends of biodiesel and ultra low sulfur diesel fuel were generated in the laboratory for
the requirements of the test. Mixes were prepared with volume percent biodiesel from 0.1 to 99.8%
(samples were produced individually, not serially). The pure diesel and biodiesel were also measured.
A calibration was developed using the absorption band for esters at 1744 cm-1. The IR spectra for a
number of blends, as measured by the FluidScan, can be seen in Figure 2.
B0 B0.25 B0.5
1.95 B1 B2.5 B5
B7.5 B8.75 B10
1.75 B11.25 B12.5 B15
B17.5 B20 B25
1.55 B30 B40 B50
B60 B70 B70
1.35 B80 B90 B95
B97 B99 B99.8
Absorbance
1.15
0.95
0.75
0.55
0.35
0.15
-0.05
1000 1100 1200 1300 1400 1500 1600 1700 1800 1900
-1
Frequency (cm )
2
R = 0.99
100
60
40
20
0
0 20 40 60 80 100 120
Laboratory Blending (% volume biodiesel)
Separate sets of biodiesel/diesel fuel mixes were generated and measured to test the robustness of the
calibration on “unknown” samples. Two samples of B100 were obtained from an international supplier
and were produced from atypical feedstocks. These were blended with two alternate, aged diesel fuels
samples to generate a matrix of test samples. The samples were measured on two FluidScan units, to
give a full picture of the reproducibility of the measurement.
Figures 5a and 5b show the excellent correlation obtained from the FluidScan measurements on this
set of “real world” or “unknown” test samples. In Figure 5a, it can be seen that the FluidScan reading
correlate well with the known laboratory blending numbers over the entire range of blends from B0
to B100. The correlation factor is over 99%. Figure 5b shows the same data, highlighting the lower
range of laboratory blends below 15% biodiesel.
% Volume Biodiesel in Diesel Fuel- Test Samples % Volume Biodiesel in Diesel Fuel- Test Samples
14
100
R2 = 0.99 R2 = 0.98
12
FluidScan (% volume biodiesel)
FluidScan (% volume biodiesel)
80
10
60 8
6
40
20
2
0 0
0 20 40 60 80 100 0 2 4 6 8 10 12 14
Laboratory Blending (% volume biodiesel) Laboratory Blending (% volume biodiesel)
Figures 5a and 5b, FluidScan Biodiesel in Diesel Fuel Correlations to Laboratory Blends
from B0 to B100 (5a) and B0 to B15 (5b)
160 Ayer Road • Littleton, MA 01460 USA • Tel: (978) 486-0123 • Fax: (978) 486-0030
Page 3
QinetiQ North America
• E-mail: sales@spectroinc.com • World Wide Web: www.spectroinc.com
Figure 6, FluidScan Biodiesel in Diesel Fuel Correlations to Laboratory Blend
Innovation, Quality and Support
Conclusion
The tests and data as described in this application note demonstrate that the FluidScan can provide
quality analytical results for biodiesel content in diesel fuel comparable to those provided by laboratory
instruments.
With the FluidScan, these results can be obtained in the field and the measurement are quickly and
clearly displayed to the user. The measurements can be stored, viewed and trended on the device or
exported to a central database. Database software, called FluidScan Manager, is also provided for a
personal computer to store, trend, set alarms, and report all collected data, Figure 6.
The handheld nature of the FluidScan, ease of use, and rapid results make fast, accurate on-site read-
ings of biodiesel content in diesel fuel possible.