Sie sind auf Seite 1von 15

Documentation

User Information

User Name
Address

Tel
Fax
E-Mail

Notes
Device parametrized
Name, Date, Signature

Device parametrization checked


Name, Date, Signature

Device parametrization released


Name, Date, Signature

14-01-2020 9:42 AM Page 2


Prowirl 200

Prowirl 200

Prowirl 200
Access status tooling Service

Operation / Operation
Language English Locking status ---
Access status display Maintenance

Operation / Display / Display


Format display 4 values Display interval 5s
Contrast display 40 %

Operation / Totalizer handling / Totalizer


Control Totalizer 1 Totalize Preset value 2 0.00000 kg
Preset value 1 0.00000 t Control Totalizer 3 Totalize
Control Totalizer 2 Totalize Preset value 3 0.00000 Nm³

Operation / Totalizer handling / Operation


Reset all totalizers Cancel

Setup / Setup
Device tag Prowirl

Setup / Commissioning / HART info


HART short tag PROWIRL HART address 0
Device tag Prowirl

Setup / Commissioning / System units


Volume flow unit m³/h Mass flow unit t/h

Setup / Commissioning / Medium property


Select medium Steam Fixed temperature 20.0000 °C

Setup / Commissioning / HART output


Assign PV Mass flow Assign TV Totalizer 1
Assign SV Temperature Assign QV Totalizer 2

Setup / Commissioning / Current output 1


Assign current output 1 Mass flow Current span 4...20 mA NAMUR

14-01-2020 9:42 AM Page 3


Prowirl 200

Setup / Commissioning / Current output 1


4 mA value 0.000 t/h Failure mode Max.
20 mA value 9.476 t/h

Setup / Commissioning / Totalizer 1


Assign process variable Mass flow Unit totalizer 1 t
Failure mode Stop

Setup / Medium selection / Medium selection


Select medium Steam
Steam calculation mode Saturated steam (T-
compensated)

Setup / System units / System units


Volume flow unit m³/h Energy flow unit kW
Volume unit m³ Energy unit kWh
Mass flow unit t/h Velocity unit m/s
Mass unit kg Density unit kg/m³
Corrected volume flow unit Nm³/h Specific volume unit m³/kg
Corrected volume unit Nm³ Dynamic viscosity unit cP
Pressure unit bar Length unit mm
Temperature unit °C

Setup / Current output 1 / Current output


Assign current output 1 Mass flow 20 mA value 9.476 t/h
Current span 4...20 mA NAMUR Damping output 1 2.0 s
4 mA value 0.000 t/h Failure mode Max.

Setup / Pulse/frequency/switch output 1 / Pulse/frequency/switch output


Operating mode Pulse Pulse width 100.00 ms
Assign pulse output 1 Mass flow Failure mode No pulses
Value per pulse 0.26322 kg Invert output signal No

Setup / Low flow cut off / Low flow cut off


Sensitivity 5.0 Turn down 100 %

Setup / Advanced setup / Advanced setup


Enter access code 0

Setup / Advanced setup / External compensation / External compensation


External value Off 2nd temperature delta heat 20.0000000 °C
Delta heat calculation Off Fixed process pressure 1.0133 bar
Fixed temperature 20.0000 °C

Setup / Advanced setup / Sensor adjustment / Sensor adjustment


Inlet configuration Off Inlet run 825.0001 mm

14-01-2020 9:42 AM Page 4


Prowirl 200

Setup / Advanced setup / Sensor adjustment / Sensor adjustment


Mating pipe diameter 0.0000 mm Installation factor 1.0000

Setup / Advanced setup / Totalizer 1 / Totalizer 1


Assign process variable Mass flow Failure mode Stop
Unit totalizer 1 t

Setup / Advanced setup / Totalizer 2 / Totalizer 2


Assign process variable Mass flow Failure mode Stop
Unit totalizer 2 kg

Setup / Advanced setup / Totalizer 3 / Totalizer 3


Assign process variable Corrected volume flow Failure mode Stop
Unit totalizer 3 Nm³

Setup / Advanced setup / Display / Display 1


Format display 4 values Value 3 display Calculated saturated steam
Value 1 display Mass flow pressure
0% bargraph value 1 0.0000 t/h 0% bargraph value 3 0.0000 bar
100% bargraph value 1 270.0000 t/h 100% bargraph value 3 120.0000 bar
Decimal places 1 x.xx Decimal places 3 x.xx
Value 2 display Totalizer 1 Value 4 display None
Decimal places 2 x.xx Language English

Setup / Advanced setup / Display / Display 2


Display interval 5s Header Device tag
Display damping 0.0 s Separator .

Setup / Advanced setup / Heartbeat setup / Heartbeat base settings / Heartbeat base settings
Plant operator -------------------------------- Location --------------------------------

Setup / Advanced setup / Configuration backup display / Configuration backup display


Operating time 2d16h55m02s Backup state None
Last backup Comparison result No backup available
Configuration management Cancel

Setup / Advanced setup / Administration / Administration


Define access code 0 Device reset Cancel

Diagnostics / Diagnostics
Actual diagnostics -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Operating time from restart 0d00h08m34s
Previous diagnostics -------- Operating time 2d16h55m02s

Diagnostics / Diagnostic list / Diagnostic list


Diagnostics 1 -------- Timestamp 0d00h00m00s

14-01-2020 9:42 AM Page 5


Prowirl 200

Diagnostics / Diagnostic list / Diagnostic list


Diagnostics 2 -------- Diagnostics 4 --------
Timestamp 0d00h00m00s Timestamp 0d00h00m00s
Diagnostics 3 -------- Diagnostics 5 --------
Timestamp 0d00h00m00s Timestamp 0d00h00m00s

Diagnostics / Event logbook / Event logbook


Filter options All Clear event list Cancel

Diagnostics / Device information / Device information 1


Device tag Prowirl Device name Prowirl 200
Serial number P70B6E19000 Order code 7F2C80-4DC6/0
Firmware version 01.03.01 ENP version 2.02.00

Diagnostics / Device information / Device information 1 / Extended order code


Extended order code 1 7F2C80-AABCJDACAAAD2 Extended order code 3
Extended order code 2 SKA1+AA

Diagnostics / Device information / Device information 2


Device revision 0x4 Device type 0x1138
Device ID 0x4a5bf4 Manufacturer ID 0x11

Diagnostics / Measured values / Process variables / Process variables


Volume flow 0.0000 m³/h Heat flow difference 0.0000 kW
Corrected volume flow 0.0000 Nm³/h Reynolds number 0.0000
Mass flow 0.0000 t/h Density 0.7866 kg/m³
Flow velocity 0.0000 m/s Specific volume 1.2714 m³/kg
Temperature 108.1155 °C Pressure 1.0133 bar
Calculated saturated steam pressure 1.3592 bar Compressibility factor 0.0000
Energy flow 0.0000 kW Degrees of superheat 0.0000 K

Diagnostics / Measured values / Totalizer / Totalizer


Totalizer value 1 0.00000 t Totalizer overflow 2 0
Totalizer overflow 1 0 Totalizer value 3 3399895.00000 Nm³
Totalizer value 2 58863.68000 kg Totalizer overflow 3 0

Diagnostics / Measured values / Output values / Output 1


Output current 1 4.00 mA Terminal voltage 1 24 V
Measured current 1 4.00 mA

Diagnostics / Measured values / Output values / Output 2


Pulse output 0.000 Hz

Diagnostics / Data logging / Data logging


Assign channel 1 Off Assign channel 3 Off
Assign channel 2 Off Assign channel 4 Off

14-01-2020 9:42 AM Page 6


Prowirl 200

Diagnostics / Data logging / Data logging


Logging interval 1.0 s Clear logging data Cancel

Diagnostics / Heartbeat / Performing verification / Performing verification


Year 10 Verification mode Internal verification
Month January Start verification Cancel
Day 1d Progress 100 %
Hour 12 h Status Not done
Minute 0 min Overall result Not done

Diagnostics / Heartbeat / Verification results / Verification results


Overall result Not done

Diagnostics / Simulation / Process value simulation


Assign simulation process variable Off

Diagnostics / Simulation / Output simulation


Current output 1 simulation Off Pulse output simulation Off

Diagnostics / Simulation / Diagnostic event simulation


Device alarm simulation Off Diagnostic event simulation Off
Diagnostic event simulation 65533

Expert / Expert
Locking status --- Access status tooling Service
Access status display Maintenance Enter access code 0

Expert / System / Display / Display 1


Language English Decimal places 2 x.xx
Format display 4 values Value 3 display Calculated saturated steam
Value 1 display Mass flow pressure
0% bargraph value 1 0.0000 t/h 0% bargraph value 3 0.0000 bar
100% bargraph value 1 270.0000 t/h 100% bargraph value 3 120.0000 bar
Decimal places 1 x.xx Decimal places 3 x.xx
Value 2 display Totalizer 1 Value 4 display None

Expert / System / Display / Display 2


Display interval 5s Separator .
Display damping 0.0 s Contrast display 40 %
Header Device tag Access status display Maintenance

Expert / System / Configuration backup display / Configuration backup display


Operating time 2d16h55m02s Backup state None
Last backup Comparison result No backup available
Configuration management Cancel

14-01-2020 9:42 AM Page 7


Prowirl 200

Expert / System / Diagnostic handling / Diagnostic handling


Alarm delay 0s

Expert / System / Diagnostic handling / Diagnostic behavior / Diagnostic behavior


Assign behavior of diagnostic no. 022 Alarm Assign behavior of diagnostic no. 834 Warning
Assign behavior of diagnostic no. 122 Warning Assign behavior of diagnostic no. 835 Warning
Assign behavior of diagnostic no. 350 Alarm Assign behavior of diagnostic no. 841 Warning
Assign behavior of diagnostic no. 371 Warning Assign behavior of diagnostic no. 844 Warning
Assign behavior of diagnostic no. 441 Warning Assign behavior of diagnostic no. 870 Warning
Assign behavior of diagnostic no. 442 Warning Assign behavior of diagnostic no. 871 Off
Assign behavior of diagnostic no. 443 Warning Assign behavior of diagnostic no. 873 Off
Assign behavior of diagnostic no. 801 Alarm Assign behavior of diagnostic no. 874 Off
Assign behavior of diagnostic no. 828 Warning Assign behavior of diagnostic no. 945 Warning
Assign behavior of diagnostic no. 829 Warning Assign behavior of diagnostic no. 947 Alarm
Assign behavior of diagnostic no. 832 Warning Assign behavior of diagnostic no. 972 Off
Assign behavior of diagnostic no. 833 Warning

Expert / System / Diagnostic handling / Diagnostic limits / Diagnostic limits


Reynolds number limit 5000.0000 Degrees of superheat limit 5.0000 K

Expert / System / Administration / Administration


Define access code 0
Device reset Cancel
Activate SW option 63916652
Software option overview Extended HistoROM, Mass
flow, Natural gas, Air +
industrial gas , Heartbeat
Verification

Expert / Sensor / Measured values / Process variables / Process variables


Volume flow 0.0000 m³/h Density 0.7866 kg/m³
Corrected volume flow 0.0000 Nm³/h Specific volume 1.2714 m³/kg
Mass flow 0.0000 t/h Pressure 1.0133 bar
Flow velocity 0.0000 m/s Saturation temperature 107.8661000 °C
Temperature 108.1155 °C Degrees of superheat 0.0000 K
Calculated saturated steam pressure 1.3592 bar Compressibility factor 0.0000
Energy flow 0.0000 kW Vortex frequency 0.0000 Hz
Heat flow difference 0.0000 kW Enthalpy 2687.8120 kJ/kg
Reynolds number 0.0000

Expert / Sensor / Measured values / System values / System values


Vortex amplitude 0.0005 DSC sensor temperature PT1 107.8707 °C
Gap capacity 56.7504 pF DSC sensor temperature PT2 105.4769 °C
Gap capacity difference 1.8059 pF Sensor leakage current 0.6117 %

14-01-2020 9:42 AM Page 8


Prowirl 200

Expert / Sensor / Measured values / Totalizer / Totalizer


Totalizer value 1 0.00000 t Totalizer overflow 2 0
Totalizer overflow 1 0 Totalizer value 3 3399895.00000 Nm³
Totalizer value 2 58863.68000 kg Totalizer overflow 3 0

Expert / Sensor / Measured values / Output values / Output 1


Output current 1 4.00 mA Terminal voltage 1 24 V
Measured current 1 4.00 mA

Expert / Sensor / Measured values / Output values / Output 2


Pulse output 0.000 Hz

Expert / Sensor / System units / System units


Volume flow unit m³/h Energy unit kWh
Volume unit m³ Velocity unit m/s
Mass flow unit t/h Density unit kg/m³
Mass unit kg Specific volume unit m³/kg
Corrected volume flow unit Nm³/h Dynamic viscosity unit cP
Corrected volume unit Nm³ Specific enthalpy unit kJ/kg
Pressure unit bar Length unit mm
Temperature unit °C Date/time format dd.mm.yy hh:mm
Energy flow unit kW

Expert / Sensor / System units / User-specific units / User-specific units


User volume text User vol. User density factor 1.0000
User volume factor 1.0000 User specific-enthalpy text User enth.
User mass text User mass User specific-enthalpy offset 0.0000
User mass offset 0.0000 User specific-enthalpy factor 1.0000000
User mass factor 1.0000 User energy text User en.
User corrected volume text UserCrVol. User energy offset 0.0000
User corrected volume offset 0.0000 User energy factor 1.0000000
User corrected volume factor 1.0000 User pressure text User pres.
User density text User dens. User pressure offset 0.0000
User density offset 0.0000 User pressure factor 1.0000

Expert / Sensor / Process parameters / Process parameters


Flow override Off Flow damping 1.00 s

Expert / Sensor / Process parameters / Low flow cut off / Low flow cut off
Sensitivity 5.0 Assign process variable Volume flow
Calculated min. flow coefficient 11.0000 On value low flow cutoff 0.0000 m³/h
Turn down 100 % Off value low flow cutoff 50 %
Calculated min. vortex frequency gas 8.5000 Hz

14-01-2020 9:42 AM Page 9


Prowirl 200

Expert / Sensor / Measurement mode / Measurement mode


Select medium Steam
Steam calculation mode Saturated steam (T-
compensated)

Expert / Sensor / External compensation / External compensation


External value Off 2nd temperature delta heat 20.0000000 °C
Delta heat calculation Off Fixed process pressure 1.0133 bar
Fixed temperature 20.0000 °C

Expert / Sensor / Sensor adjustment / Installation settings


Inlet configuration Off Mating pipe diameter 0.0000 mm
Inlet run 825.0001 mm Installation factor 1.0000

Expert / Sensor / Sensor adjustment / Additional settings


Temperature calculation mode PT1 + PT2 Max. vortex frequency gas 1400.0000 Hz
Min. flow coefficient 11.0000 Sensor adjustment Cancel
Min. vortex frequency gas 8.5000 Hz DSP mode ---

Expert / Sensor / Calibration / Calibration


Calibration factor 3027.7810000 1/m³ Sensor diameter 73.7000 mm
Meter body properties DN80F-PN40 Calibration date/time 11.07.19 18:46
Tube inner diameter 82.5000 mm

Expert / Sensor / Calibration / Recalibration / Recalibration


Calibration factor 1.000 1/m³ Hour 12 h
Year 10 Minute 0 min
Month January Confirm settings No
Day 1d

Expert / Sensor / Calibration / Logbook / Logbook


Calibration counter 1 Timestamp 11.07.19 18:46
Select calibration log entry 1 Calibration factor 3027.781 1/m³

Expert / Sensor / Testpoints / Testpoints


Signal quality 0.0000 Temperature reference 1 266.6624 °C
Resonance frequency NaN Hz Temperature reference 2 130.5168 °C
Reference frequency drift 0.0250 % Pre-amplifier temperature 40.0 °C
Overall reference capacity 84.4028 pF

Expert / Output / Current output 1 / Current output 1-1


Assign current output 1 Mass flow 20 mA value 9.476 t/h
Current span 4...20 mA NAMUR Damping output 1 2.0 s
4 mA value 0.000 t/h Response time 3.0000 s

14-01-2020 9:42 AM Page 10


Prowirl 200

Expert / Output / Current output 1 / Current output 1-2


Failure mode Max. Measured current 1 4.00 mA
Output current 1 4.00 mA Terminal voltage 1 24 V
Start-up mode Min.

Expert / Output / Current output 1 / Current output 1-3


Trim Off Status 128

Expert / Output / Pulse/frequency/switch output / Pulse/frequency/switch output


Operating mode Pulse

Expert / Output / Pulse/frequency/switch output / Settings


Assign pulse output 1 Mass flow Failure mode No pulses
Value per pulse 0.26322 kg Pulse output 0.000 Hz
Pulse width 100.00 ms

Expert / Output / Pulse/frequency/switch output / Additional settings


Invert output signal No

Expert / Communication / Diagnostic configuration / Diagnostic configuration


Event category 022 Failure (F) Event category 122 Maintenance required (M)

Expert / Communication / Diagnostic configuration / Diagnostic configuration


Event category 350 Out of specification (S) Event category 371 Maintenance required (M)

Expert / Communication / Diagnostic configuration / Diagnostic configuration


Event category 441 Out of specification (S) Event category 443 Out of specification (S)
Event category 442 Out of specification (S)

Expert / Communication / Diagnostic configuration / Diagnostic configuration


Event category 801 Failure (F) Event category 844 Out of specification (S)
Event category 828 Out of specification (S) Event category 870 Out of specification (S)
Event category 829 Out of specification (S) Event category 871 Out of specification (S)
Event category 832 Out of specification (S) Event category 873 Out of specification (S)
Event category 833 Out of specification (S) Event category 874 Out of specification (S)
Event category 834 Out of specification (S) Event category 945 Out of specification (S)
Event category 835 Out of specification (S) Event category 947 Out of specification (S)
Event category 841 Out of specification (S) Event category 972 Out of specification (S)

Expert / Communication / HART input / Configuration / Configuration


Capture mode Off

Expert / Communication / HART input / Input / Input


Value 0.0000 Status Bad

14-01-2020 9:42 AM Page 11


Prowirl 200

Expert / Communication / HART output / Configuration / Configuration


HART short tag PROWIRL HART address 0
Device tag Prowirl No. of preambles 5

Expert / Communication / HART output / Burst configuration / Burst configuration 1 / Burst configuration
1 / Burst message 1
Burst mode 1 Off Burst trigger mode Continuous
Burst command 1 Command 2 Min. update period 1000 ms

Expert / Communication / HART output / Burst configuration / Burst configuration 2 / Burst configuration
1 / Burst message 2
Burst mode 2 Off Burst trigger mode Continuous
Burst command 2 Command 2 Min. update period 1000 ms

Expert / Communication / HART output / Burst configuration / Burst configuration 3 / Burst configuration
1 / Burst message 3
Burst mode 3 Off Burst trigger mode Continuous
Burst command 3 Command 2 Min. update period 1000 ms

Expert / Communication / HART output / Information / HART info


Device revision 0x4 HART descriptor PROWIRL
Device ID 0x4a5bf4 HART message PROWIRL
Device type 0x1138 Hardware revision 1
Manufacturer ID 0x11 Software revision 4
HART revision 7 HART date code 2009-07-20

Expert / Communication / HART output / Output / HART output


Assign PV Mass flow Assign TV Totalizer 1
Primary variable (PV) 0.000 t/h Tertiary variable (TV) 0.000 t
Assign SV Temperature Assign QV Totalizer 2
Secondary variable (SV) 107.713 °C Quaternary variable (QV) 58863.680 kg

Expert / Application / Application


Reset all totalizers Cancel

Expert / Application / Totalizer 1 / Totalizer 1


Assign process variable Mass flow Preset value 1 0.00000 t
Unit totalizer 1 t Failure mode Stop
Control Totalizer 1 Totalize

Expert / Application / Totalizer 2 / Totalizer 2


Assign process variable Mass flow Preset value 2 0.00000 kg
Unit totalizer 2 kg Failure mode Stop
Control Totalizer 2 Totalize

14-01-2020 9:42 AM Page 12


Prowirl 200

Expert / Application / Totalizer 3 / Totalizer 3


Assign process variable Corrected volume flow Preset value 3 0.00000 Nm³
Unit totalizer 3 Nm³ Failure mode Stop
Control Totalizer 3 Totalize

Expert / Application / Inventory counter / Inventory counter


Totalizer value 9768906.0000 l Unit l
Totalizer overflow 0

Expert / Diagnostics / Diagnostics


Actual diagnostics -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Operating time from restart 0d00h08m34s
Previous diagnostics -------- Operating time 2d16h55m02s

Expert / Diagnostics / Diagnostic list / Diagnostic list


Diagnostics 1 -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Diagnostics 4 --------
Diagnostics 2 -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Diagnostics 5 --------
Diagnostics 3 -------- Timestamp 0d00h00m00s

Expert / Diagnostics / Event logbook / Event logbook


Filter options All Clear event list Cancel

Expert / Diagnostics / Device information / Device information


Device tag Prowirl Order code 7F2C80-4DC6/0
Serial number P70B6E19000 Configuration counter 133
Firmware version 01.03.01 ENP version 2.02.00
Device name Prowirl 200

Expert / Diagnostics / Device information / Device information / Extended order code


Extended order code 1 7F2C80-AABCJDACAAAD2 Extended order code 3
Extended order code 2 SKA1+AA

Expert / Diagnostics / Sensor information / Sensor information


DSC sensor serial number @f1902676116 Meter body ID 71333107
DSC ID 71345567 Pressure cell serial number
Meter body serial number 4210005 Pressure cell ID

Expert / Diagnostics / Pre-amplifier module / Pre-amplifier module


Hardware revision Serial number
Module name

Expert / Diagnostics / Mainboard module / Mainboard module


Software revision 10301 Hardware revision V1.01.00
Build no. software 2 Module name V308-HART

14-01-2020 9:42 AM Page 13


Prowirl 200

Expert / Diagnostics / Mainboard module / Mainboard module


Serial number 0156654881 Bootloader revision 701

Expert / Diagnostics / I/O module / I/O module


Software revision 20005 Module name IO211_212_SB
Build no. software 2124 Serial number 0157044438
Hardware revision V2.00.01 Bootloader revision 641

Expert / Diagnostics / Display module / Display module


Software revision 10204 Module name SD02
Build no. software 1 Serial number 0155646160
Hardware revision V2.00.00 Bootloader revision 630

Expert / Diagnostics / Data logging / Data logging


Assign channel 1 Off Assign channel 4 Off
Assign channel 2 Off Logging interval 1.0 s
Assign channel 3 Off Clear logging data Cancel

Expert / Diagnostics / Min/max values / Min/max values


Reset min/max values Cancel

Expert / Diagnostics / Min/max values / Terminal voltage / Terminal voltage


Minimum value 24 V Average value 24 V
Maximum value 29 V

Expert / Diagnostics / Min/max values / IO module temperature / IO module temperature


Minimum value 24.0 °C Average value 29.1 °C
Maximum value 33.0 °C

Expert / Diagnostics / Min/max values / Pre-amplifier temperature / Pre-amplifier temperature


Minimum value 24.2 °C Maximum value 54.6 °C

Expert / Diagnostics / Min/max values / Medium temperature / Medium temperature


Minimum value 20.0 °C Maximum value 194.4 °C

Expert / Diagnostics / Min/max values / Flow velocity / Flow velocity


Maximum value 2633.5 m/s

Expert / Diagnostics / Min/max values / Gap capacity / Gap capacity


Minimum value 56.4280 pF Maximum value 57.2661 pF

Expert / Diagnostics / Min/max values / Gap capacity difference / Gap capacity difference
Minimum value 1.5244 pF Maximum value 2.4995 pF

Expert / Diagnostics / Min/max values / DSC sensor temperature PT1 / DSC sensor temperature PT1
Minimum value 24.0 °C Maximum value 194.4 °C

14-01-2020 9:42 AM Page 14


Prowirl 200

Expert / Diagnostics / Min/max values / DSC sensor temperature PT2 / DSC sensor temperature PT2
Minimum value 23.9 °C Maximum value 192.7 °C

Expert / Diagnostics / Min/max values / Vortex frequency / Vortex frequency


Maximum value 1184.5000 Hz

Expert / Diagnostics / Heartbeat / Heartbeat base settings / Heartbeat base settings


Plant operator -------------------------------- Location --------------------------------

Expert / Diagnostics / Heartbeat / Performing verification / Performing verification


Year 10 Verification mode Internal verification
Month January Start verification Cancel
Day 1d Progress 100 %
Hour 12 h Status Not done
Minute 0 min Overall result Not done

Expert / Diagnostics / Heartbeat / Verification results / Verification results


Overall result Not done

Expert / Diagnostics / Simulation / Process value simulation


Assign simulation process variable Off

Expert / Diagnostics / Simulation / Output simulation


Current output 1 simulation Off Pulse output simulation Off

Expert / Diagnostics / Simulation / Diagnostic event simulation


Device alarm simulation Off Diagnostic event simulation Off
Diagnostic event simulation 65533

14-01-2020 9:42 AM Page 15