Beruflich Dokumente
Kultur Dokumente
TM
Operating Instructions HFM 436 Lambda
Contents
Chapter II Installation
Chapter V Appendix
03/07 J:\HFM436\Contents.doc
HFM 436 Chapter I
General Information
03/07 J:\HFM436\Chapter1.doc
HFM 436 General Information
Information
NETZSCH-Gerätebau GmbH
Wittelsbacherstraße 42
D - 95100 SELB
Telephon: 09287/881- 0
Telefax: 09287/881- 144
E-Mail: at@ngb.netzsch.com
Internet: http://www.netzsch-thermal-analysis.com
03/07 J:\HFM436\Chapter1.doc 1
HFM 436 General Information
In this manual, the symbols described below are used to simplify orientation.
NOTE
ATTENTION!
The tools listed after this symbol are required for the
installation or modification of your instrument.
03/07 J:\HFM436\Chapter1.doc 2
HFM 436 General Information
Unauthorized Access The operator must make sure that only trained personnel
work on the instrument.
Improper Operation Any type of operation which reduces the safety of the user
and the operability of the instrument should be avoided.
03/07 J:\HFM436\Chapter1.doc 3
HFM 436 General Information
Maintenance The operator of the system must ensure that the instrument is
Obligation operated in perfect condition at all times.
Proper Set-up of the The operator must guarantee the openness and cleanliness of
Work Stations the work stations at the instrument through appropriate
instruction and inspections.
03/07 J:\HFM436\Chapter1.doc 4
HFM 436 General Information
03/07 J:\HFM436\Chapter1.doc 5
HFM 436 Chapter II
Installation
03/07 J:\HFM436\Chapter2.doc
HFM 436 Installation
Contents
INSTALLATION............................................................................................ 1
03/07 J:\HFM436\Chapter2.doc
HFM 436 Installation
Installation
03/07 J:\HFM436\Chapter2.doc 1
HFM 436 Installation
Location Requirements
03/07 J:\HFM436\Chapter2.doc 2
HFM 436 Installation
Installation schematics
520
measuring part HFM 436 computer printer table
230 V, 10 A weight 10.5 Kg weight 8 Kg
monitor
weight 3.5 Kg
wide range power pack
100-240 V, 6 A
measures in “mm”
Option
or provided by the customer
floor
HFM 436/3/1E
470 500
400 625 depth 450 depth 500
depth 500 depth 700
490
515
520
measures in “mm”
Option
or provided by the customer
floor
03/07 J:\HFM436\Chapter2.doc 3
HFM 436 Installation
HFM 436/6/1
~1000
depth ~850
470 500
depth 450 depth 500
400
depth 500 ~850
490
520
computer printer table
cooler measuring part HFM 436 weight 10.5 Kg weight 8 Kg
230 V, 10 A
weight ~186 Kg monitor
weight 3.5 Kg
wide range power pack
100-240 V, 6 A
measures in “mm”
Option
or provided by the customer
floor
03/07 J:\HFM436\Chapter2.doc 4
HFM 436 Installation
PC
SERIAL
PARALLEL SERIAL
RESET
power supply
03/07 J:\HFM436\Chapter2.doc 5
HFM 436 Installation
HFM 436/3/1E
PC
SERIAL
PARALLEL SERIAL
RS232
RESET
OUTLET
INLET
power supply
03/07 J:\HFM436\Chapter2.doc 6
HFM 436 Installation
HFM 436/6/1
PC
SERIAL
RESET
OUTLET
INLET
power supply
03/07 J:\HFM436\Chapter2.doc 7
HFM 436 Installation
HFM 436/3/1E
cooler
power supply
HFM 436/3/1E rear
OUTLET
INLET
PARALLEL SERIAL
RS232
RESET
power supply
03/07 J:\HFM436\Chapter2.doc 8
HFM 436 Installation
HFM 436/6/1
Connect the cooler via the RS232 interface.
cooler
power supply
HFM 436/6/1 rear
OUTLET
INLET
SERIAL PARALLEL
RS232
RESET
power supply
03/07 J:\HFM436\Chapter2.doc 9
HFM 436 Chapter III
System Components
03/07 J:\HFM436\Chapter3.doc
HFM 436 System Components
Contents
MEASURING UNIT....................................................................................... 1
03/07 J:\HFM436\Chapter3.doc
HFM 436 System Components
Measuring Unit
height
515 mm
depth
700 mm
width
625 mm
height
850 mm
depth
850 mm
width
1000 mm
03/07 J:\HFM436\Chapter3.doc 1
HFM 436 System Components
Front panel
A
POW ER
R ESET
R UN
1 2 3 ES C * ENTE R
R OUGH
4 5 6 Y SPAC E N
B 7
.
8
0
9
+/-
FINE
POINT
B Front Door
03/07 J:\HFM436\Chapter3.doc 2
HFM 436 System Components
UP/DOWN Switch
Is used to raise (lower) the upper plate. When a sample is to be placed into the sample
chamber, the UP/DOWN switch should be pressed to raise the upper plate. After the
sample is correctly positioned, press the UP/DOWN switch again to lower the plate so
that it properly clamps the sample.
Over-ride Switch
Is used when it is necessary to realign the drive mechanism for the upper plate. Under
certain circumstances, the drive mechanism for the upper plate could become misaligned
and the upper plate will not be driven to the full upper (lower) position. If this happens,
press both the Over-ride button and the UP/DOWN switches to drive the plate to the
uppermost position so that proper operation of the system can be restored.
03/07 J:\HFM436\Chapter3.doc 3
HFM 436 System Components
POWER
RESET
RUN
1 2 3 ESC * ENTER
ROUGH
4 5 6 Y SPACE N
FINE
7 8 9
POINT
. 0 +/-
Figure 3: HFM Unit Controller
LEDs
Keypad
ENTER
Key is used to accept information and go to the next step. If you are at
the end of the last step in a sequence, the controller will present the
previous step.
ESC
Key is used to return to the previous step. Any change that was made to
the present display will be erased.
SPACE
Key is used for erase character.
Y N
Keys are used for when employ a specific feature (yes or no decision).
Keys for moving the cursor to another line within the list.
*
Key is used for enter a point.
03/07 J:\HFM436\Chapter3.doc 4
HFM 436 System Components
See chapter IV for starting a measurement via the HFM unit controller
(Q-Test).
Front door
The sample compartment is accessed by opening the door on the front of the unit.
front
door
03/07 J:\HFM436\Chapter3.doc 5
HFM 436 System Components
Up/Down
switch
front
door
03/07 J:\HFM436\Chapter3.doc 6
HFM 436 System Components
Back panel
A
PARALLEL SERIAL
C
RESET
03/07 J:\HFM436\Chapter3.doc 7
HFM 436 System Components
HFM 436/3/1E
A
PARALLEL SERIAL
B
RS232
C
RESET
OUTLET
I
F
INLET
D E
03/07 J:\HFM436\Chapter3.doc 8
HFM 436 System Components
HFM 436/6/1
A
B RS232 SERIAL PARALLEL
C
RESET
OUTLET
D
F
INLET
03/07 J:\HFM436\Chapter3.doc 9
HFM 436 System Components
Cooler
A cooler is required to operate the instrument HFM 436/3/1E and 436/6/1. This handles
thermostatic control of instrument components relevant to the measurement and
cooling of the thermoelectric system. The cooler is connected to the (inlet-outlet)
connections on the back panel of the measuring unit.
Technical data for the thermostat can be found in the accompanying operating
instructions.
03/07 J:\HFM436\Chapter3.doc 10
HFM 436 Chapter IV
03/07 J:\HFM436\Chapter4.doc
HFM 436 Operating the Instrument
Contents
03/07 J:\HFM436\Chapter4.doc
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc
HFM 436 Operating the Instrument
Operating Principle
LVDT-system
Hoisting device
Upper heat sink
Cold Plate
Peltier system
Lower heat sink
LVDT-system
Hoisting device
Upper heat sink
Electronics and data
acquisition system
Peltier system
Hot plate
Heat flux transducers
Direction of
Test sample
Heat Flow
Cold Plate
Peltier system
Lower heat sink
External chiller
for HFM 436/X/1E
versions
03/07 J:\HFM436\Chapter4.doc 1
HFM 436 Operating the Instrument
The Sample is placed between two heated plates, set at different temperatures. The
heat flow (q) through the sample is measured by a (calibrated) heat flux transducer.
After reaching a thermal equilibrium, the test is done. Only the sample center (100x100
mm for the HFM 436/3/x versions) is used for the analysis. The heat flux transducer
output is calibrated with a standard.
The Fourier heat flow equation ( Equation 1) gives the relationship between these
parameters when the test section reaches thermal equilibrium.
• ∆T
Q = λA Equation 1
∆x
One or two heat flow transducers measure the heat flow though the sample
(see Figure 1). The signal of a heat flow transducer (in Volts (V)) is proportional to the
heat flow through the transducer. In the HFM 436 Heat Flow Meter instrument, the area
of the heat flow transducer represents the area through which the heat flows and is the
same for all samples; therefore:
•
Q = N ⋅V Equation 2
where N is the calibration factor that relates the voltage signal of the heat flow
transducer to the heat flux through the sample. Solving for λ we derive the thermal
conductivity:
V∆x
λ=k=N Equation 3
∆T
03/07 J:\HFM436\Chapter4.doc 2
HFM 436 Operating the Instrument
The heat flow meter method is a standardized test technique. The application
(insulating materials) is strongly connected with standards:
• ISO 8301:
Standard test technique for measurements of insulating materials using the
heat flow meter method.
• ASTM C518:
American standard for measurements of insulating materials using the heat
flow meter method.
• DIN EN 12667/12939:
European standard for measurements of insulating materials using the heat
flow meter method or the guarded hot plate technique.
• DIN EN 13163:
European standard for characterization of foam insulations for building
applications using the heat flow meter method or the guarded hot plate
technique.
03/07 J:\HFM436\Chapter4.doc 3
HFM 436 Operating the Instrument
The heat flow meter can be used to test solid or loose-fill samples.
Solid sample
03/07 J:\HFM436\Chapter4.doc 4
HFM 436 Operating the Instrument
The heat flow meter can be used to determine the thermal conductivity of a loose-fill
material (such as a powder or flakes) by using a frame to contain the material (see for
dimensions of the frame). A frame can be obtained from NETZSCH or can be made
locally. The height of the frame is dependent on the particle size of the material as
indicated in Table 1.
Less than 20 mm 25 mm
More than 20 mm 50 mm
To construct a frame:
1. If the testing is to be performed below 70°C, the frame can be constructed from
Styrofoam. For temperatures above 70°C use a fibrous ceramic material.
2. Cut a piece of strong paper to fit across the base of the frame.
3. Attach the piece of paper to the frame using a high temperature adhesive.
frame
03/07 J:\HFM436\Chapter4.doc 5
HFM 436 Operating the Instrument
4. The loose-fill sample can be filled into the frame. The loose-fill sample should not
exceed the upper edge of the frame.
loose-fill material
frame
Compressible Samples
It is possible to test compressible material with the HFM 436. It is very important that the
operator visually verify that there is good thermal contact between the sample and the
upper plate. With poor thermal contact the interface resistances increase or decrease
due to convection in the gaps dramatically which cause the measured Thermal
Resistance value to be off.
03/07 J:\HFM436\Chapter4.doc 6
HFM 436 Operating the Instrument
1. Switch on the power to the heat flow meter, computer (if present), monitor (if
present), and printer.
Signing On IMS4831
2. Press the RESET button on the rear of the instrument to provide power to the
heaters.
PARALLEL SERIAL
RESET
RESET
03/07 J:\HFM436\Chapter4.doc 7
HFM 436 Operating the Instrument
• The red RESET LED on the front panel of the HFM Unit Controller will be
turned off.
POWER
RESET RESET
RUN
1 2 3 ESC * ENTER
ROUGH
4 5 6 Y SPACE N
FINE
7 8 9
POINT
. 0 +/-
03/07 J:\HFM436\Chapter4.doc 8
HFM 436 Operating the Instrument
front
door
sample
03/07 J:\HFM436\Chapter4.doc 9
HFM 436 Operating the Instrument
front
door
03/07 J:\HFM436\Chapter4.doc 10
HFM 436 Operating the Instrument
Q-Lab software
mean temperature
delta T between of two plates (K)
plates (K)
03/07 J:\HFM436\Chapter4.doc 11
HFM 436 Operating the Instrument
• Select Calibration as
Type of Measurement.
03/07 J:\HFM436\Chapter4.doc 12
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc 13
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc 14
HFM 436 Operating the Instrument
• Select tab
Equilibrium.
• Equilibrium
Settings:
This table contains
values used by the
software to determine
when test conditions
have reached
equilibrium. You
specify how often data
points are recorded,
how long the
averaging period
(block) should be, and
how tight a tolerance
the points in the block
must satisfy.
03/07 J:\HFM436\Chapter4.doc 15
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc 16
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc 17
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc 18
HFM 436 Operating the Instrument
• Select tab
Equilibrium.
• Equilibrium
Settings:
This table contains
values used by the
software to determine
when test conditions
have reached
equilibrium. You
specify how often data
points are recorded,
how long the
averaging period
(block) should be, and
how tight a tolerance
the points in the block
must satisfy.
03/07 J:\HFM436\Chapter4.doc 19
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc 20
HFM 436 Operating the Instrument
• If required, the
calibration file can be
saved.
03/07 J:\HFM436\Chapter4.doc 21
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc 22
HFM 436 Operating the Instrument
03/07 J:\HFM436\Chapter4.doc 23
HFM 436 Operating the Instrument
The HFM 436 controller is used to carry out measurements without the software. It
provides a series of displays that can be used to enter operating parameters, calibrate
the system, initiate data collection, view the actual value of a number of parameters
during operation of the unit and shut the system down. The program is designed to be
very flexible and allows data entry and viewing of instrument settings in a logical,
structured manner.
Calibration Run *
Test Run
Define Moving
View
Print Test Results
Calibrate LVDT
Save System Data
Shut Down
Q-LAB
Reset
Each entry on the main menu leads to a new display, which is used to set parameters,
initiate data collection or view data. To access a display, move the asterisk (*) to the
appropriate line and press Enter.
03/07 J:\HFM436\Chapter4.doc 24
HFM 436 Operating the Instrument
Main Menu
Calibration Run * 1. Select Calibration Run in the Main Menu and
Test Run
press Enter.
Define Moving
View
Print Test Results
Calibration LVDT
Save System Data
Shutdown
QLAB
Reset
MoveUp 5.000
2. The window shows preselected values which must
not be changed. Press Enter to skip the menu.
Move Down 2.600
Tolerance 0.010
CurrentDX 2.600
MoveUp
3. Select Continue Run and press Enter.
Move Down
Continue Run *
Material
4. Define Material (e.g. 1450C), I.D., Density,
Operator and Serial No.
I.D.
Thickness The thickness is read by the HFMs transducer and
displayed here.
Density
QMet Length QMet Length and QMet Width are preselected
QMet Width
and cannot be changed.
Operator
Serial No.
03/07 J:\HFM436\Chapter4.doc 25
HFM 436 Operating the Instrument
# MeanT°C DeltaT En
5. Enter the mean temperature (the averaged
temperature calculated from T Upper and T Lower)
1 10.00 20.00 Y
and temperature difference (delta) between hot
2 20.00 20.00 Y and cold plates and set the check mark Y or N (yes
3 30.00 20.00 Y or no) for the respective data point.
4 40.00 20.00 Y
HFM HFM HFM HFM
5 50.00 20.00 Y
436/0 436/3/1 436/3/1E 436/6/1
6 60.00 20.00 Y Programmable 1 10 10 10
7 70.00 20.00 Y Data Points
8 80.00 20.00 Y Max. 30 K 30 K 30 K 30 K
9 80.00 20.00 Y DeltaT
Start Test *
6. Select Start Test to start the calibration
measurement.
Stop Test
03/07 J:\HFM436\Chapter4.doc 26
HFM 436 Operating the Instrument
Main Menu
Calibration Run 1. Select Test Run in the Main Menu and press
Test Run *
Enter.
Define Moving
View
Print Test Results
Calibration LVDT
Save System Data
Shutdown
QLAB
Reset
MoveUp 5.000
2. The windows shows preselected values which
must not be changed. Press Enter to skip the
Move Down 2.600
menu.
Tolerance 0.010
CurrentDX 2.600
MoveUp
3. Select Continue Run and press Enter.
Move Down
Continue Run *
Material
4. Define Material, I.D., Density, Operator and
Serial No.
I.D.
Thickness The thickness is read by the HFMs transducer
and
Density
displayed here.
QMet Length
QMet Width
QMet Length and QMet Width are preselected
and cannot be changed.
Operator
Serial No.
03/07 J:\HFM436\Chapter4.doc 27
HFM 436 Operating the Instrument
# MeanT°C DeltaT En
5. Enter the mean temperature (the averaged
temperature calculated from T Upper and T
1 10.00 20.00 Y
Lower) and temperature difference (delta)
2 20.00 20.00 Y between hot and cold plates and set the check
3 30.00 20.00 Y mark Y or N (yes or no) for the respective data
4 40.00 20.00 Y point.
5 50.00 20.00 Y
HFM HFM HFM HFM
6 60.00 20.00 Y 436/0 436/3/1 436/3/1E 436/6/1
7 70.00 20.00 Y Programmable 1 10 10 10
8 80.00 20.00 Y Data Points
Start Test *
6. Select Start Test to start the sample measurement.
Stop Test
03/07 J:\HFM436\Chapter4.doc 28
HFM 436 Operating the Instrument
Define Moving
The Define Moving menu allows you to set the distance between the upper and lower
plates. This function is generally used when testing compressible samples. The
recommended procedure is as follows:
3. Enter your calculated sample thickness on the Move Up or Move Down line.
Be sure when running your test you select the correct thickness, entered in
Move Up or Move Down.
4. You should visually check to make sure there is good contact between the
upper plate and the test sample. It might be necessary to subtract 2% or 3%
from the average sample thickness to ensure good contact between the
sample and the top plate.
5. Once the test is finished and the apparent Thermal Resistance is measured it
is recommended this value be adjusted by the % compression of the sample.
This will give you an accurate measure of the apparent Thermal Resistance of
the sample when not compressed. The relationships are as follows:
R = dT / K
R = Thermal Resistance
dT = Measured thickness of the sample by the thickness transducer.
K = Thermal Conductivity
Since Thermal Conductivity does not change strongly with a small change in
thickness or volume of the sample it is acceptable to use the measured
apparent Thermal Conductivity value from the same test where the apparent
Thermal Resistance was measured. Therefore:
R = dM / K
R = Thermal Resistance
dM = Measured thickness of the sample by the operator
K = Thermal Conductivity
03/07 J:\HFM436\Chapter4.doc 29
HFM 436 Operating the Instrument
In Figure 6 a value of 5.00 cm has been entered on the Move Up line and a value of 2.50
cm has been entered on the Move Down line.
Move Up 5.00
Move Down 2.50
Tolerance 0.01
No. Iter 3
Move Up/Down allows defined movement or placement of the upper plate during
sample change. Move Up defines a position the upper plate moves to before a sample is
removed or put in. Move Down defines the position after the sample is put in the
system. This feature is ideal to measure compressible samples at well defined thicknesses.
The allowable Tolerance and the number of adjustments (No. Iter) can also be entered in
this menu. The factory setting for the allowable tolerance is 0.01 cm and the number of
adjustments is set to 3. After the top plate has been adjusted once the % error is
checked between the desired position (entered by the operator on the Move Up or
Move Down line) and the measured position of the thickness transducer. If the % error
is greater than 0.01 cm then the HFM 436 will make two more adjustments to achieve
the desired position. It is not recommended to change the allowable tolerance beyond
0.01cm; however, the number of adjustments the top plate is allowed to make can be
set to the operator’s choice.
Once these values are set they will appear in the first menu of the Calibration Run or
Test Run menus. The second menu under Calibration Run or Test Run will allow you to
move the top plate to the desired position.
03/07 J:\HFM436\Chapter4.doc 30
HFM 436 Operating the Instrument
View Menu
The View menu (Figure 7) can be accessed before a run or during a run. It is used to
present a number of displays, which are used to view and/or set parameters.
a) The HFM 436/1/0 system will not present the Cal Set Point, Test Set Point and Idle
Set Point entries.
b) The HFM 436/1/0 systems will present the Set Point entry after the Test Params
View entry.
03/07 J:\HFM436\Chapter4.doc 31
HFM 436 Operating the Instrument
Cal Set Points presents a listing of the set points, which are to be used to calibrate
the system (Figure 8).
o
# Mean T C DeltaT En
1 24.00 20.00 Y
2 0.00 0.00 N
3 0.00 0.00 N
4 0.00 0.00 N
5 0.00 0.00 N
6 0.00 0.00 N
7 0.00 0.00 N
8 0.00 0.00 N
9 0.00 0.00 N
10 0.00 0.00 N
Any value can be edited and enabled (disabled) as desired. At least one line must be
enabled to calibrate the system.
The Cal Set Points entry is not presented for HFM 436/1/0 system. In this situation,
the Set Point entry (which is presented after the Test Parameters View entry),
provides for entry of a single mean/Delta T pair.
Cal Params View presents the parameters used for calibration of the system
(Figure 9).
Material 123_
I.D. 456
Thickness 2.54
Density 130
QMet Length 10.16
QMet Width 10.16
Operator 789
Serial No. 776
03/07 J:\HFM436\Chapter4.doc 32
HFM 436 Operating the Instrument
Reference Type
Reference Type presents the options for the calibration reference (Figure 10).
1450c *
1450b
1451
K VS T
The reference type display is used to indicate the reference, which is used for
calibration of the system. The calibration curves for NIST 1450C, 1450B and 1451 are
incorporated into the program; if another reference is used to calibrate the system,
select K VS T.
If K VS T is selected, the display provides for manual entry of data for your reference
sample (Figure 11).
o
# KValue Temp C
1 0.0000 0.00
2 0.0000 0.00
3 0.0000 0.00
4 0.0000 0.00
5 0.0000 0.00
6 0.0000 0.00
7 0.0000 0.00
8 0.0000 0.00
9 0.0000 0.00
10 0.0000 0.00
03/07 J:\HFM436\Chapter4.doc 33
HFM 436 Operating the Instrument
Test Set Points presents a listing of the set points, which are to be used to
determine the thermal conductivity of the sample (Figure 12).
o
# MeanT C DeltaT En
1 24.00 20.00 Y
2 0.00 0.00 N
3 0.00 0.00 N
4 0.00 0.00 N
5 0.00 0.00 N
6 0.00 0.00 N
7 0.00 0.00 N
8 0.00 0.00 N
9 0.00 0.00 N
10 0.00 0.00 N
Any value can be edited and enabled (disabled) as desired. At least one line must be
enabled to calibrate the system.
The Test Set Points entry is not presented for HFM 436/1/0 system. In this situation,
the Set Point entry (which is presented after the Test Parameters View entry),
provides for entry of a single mean/Delta T pair.
Test Params presents a display similar to that for Cal Params (Figure 9), which lists
the Test Parameters. The parameters cannot be edited at this juncture.
Idle Set Point presents the Idle Set Point conditions, which refer to the conditions
that the instrument should be set to when the system is at rest (i.e. waiting for the
next test) (Figure 13).
o
MeanT C DeltaT
Idle 23.000 20.00
Users of systems which include fixed temperature (HFM 436/1/0) operation cannot
set the Idle set Point, and this command is not presented.
03/07 J:\HFM436\Chapter4.doc 34
HFM 436 Operating the Instrument
Status View
Status View presents the operational status of the system (Figure 14).
Equilib Idle
SetPoint # 0
Elapsed 00:00
The Status View display indicates the present operation of the system (e.g. Idle,
Rough or Fine), the present set point and the period of time that has elapsed since
the start of the operation of the unit. The display will update automatically.
Meters View
Meters View provides a table that includes the present value for a number of
instrumental parameters (Figure 15).
K-User 0.03449
dXUser 2.53
K 0.03422
deltX 2.54
deltT 28.00
meanT 45.00
DelTC 28.20
MeaTC 45.01
TUpPl 59.00
TLwPl 31.00
QUpr 1200
QLwr 1225
VDelt 1.23
VMean 23.00
TCldJ 23.00
The Meters View display will update automatically. The parameters are not editable.
A detailed description of the various entries is presented in Chapter V (Parameter
Definitions for HFM Controller).
03/07 J:\HFM436\Chapter4.doc 35
HFM 436 Operating the Instrument
Heat Flow Meters is used to select the Heat Flow meter(s) that should be
employed during the measurement (Figure 16).
Raw Readings View presents the value for a variety of devices in the format in
which they are actually measured by the instrument (Figure 17). As an example, the
temperature values are expressed in terms of µV, rather than in oC or oF. To view the
data in the more customary format, refer to Meters View.
TDeCo -57
TMeCo -190
TUpPl -206
TLwPl -185
QUpr 0
QLwr 3187
LVDT 993420
Offst -35154
TCldJ 34224
03/07 J:\HFM436\Chapter4.doc 36
HFM 436 Operating the Instrument
Equip Params Set presents three parameters, which relate to the specific instrument
to which the controller is connected to (Figure 18).
ATTENTION!
The values for this display can be edited. It should be noted, however, that they are
set at the factory during the manufacture of the system and should not be edited
during normal operation.
Gain(x) 0.9402
Filter 0
OffSet(µV) 425
T adj.(µV) 0
PID Params Set provides information about the Proportional Integral Derivative
control (Figure 19). A detailed description of the various entries in the display is
presented in Chapter V (Parameter Definitions for HFM Controller).
ATTENTION!
The values for this display can be edited. It should be noted, however, that they are
set at the factory during the manufacture of the system and should not be edited
during normal operation.
Coarse Smooth
RateDlt 0.00 0.00
ResetDt 3.80 0.10
PropDlt 27.00 0.00
PBandDt 5.00 5.00
LBandDt 0.10 0.02
RateMea 0.00 0.00
ResetMn 15.00 20.00
PropMea 300.00 150.00
PBandMn 1.00 1.00
LBandMn 0.20 0.05
03/07 J:\HFM436\Chapter4.doc 37
HFM 436 Operating the Instrument
Power Set
Power Set is used to indicate the minimum and maximum potential, which should
be applied to the controller channels (Figure 20).
ATTENTION!
Although the operator can use this menu to change power settings, they should
never be changed from the factory settings without consulting a qualified NETZSCH
Instruments service engineer!
Equilibrium Params
The parameters can be edited at this juncture. The values that are selected for these
parameters have an impact on the period of time that it takes for the equilibration
and the accuracy of the measurement. If you select a short update rate, small Blk sizes
and high % Error values; the system will perform the measurement more quickly,
however the accuracy of the measurement will be reduced.
03/07 J:\HFM436\Chapter4.doc 38
HFM 436 Operating the Instrument
Temperature Offsets
Temp. Offsets is used to set temperature offset options and parameters. There is
an inherent temperature difference between the specified control temperatures
and the temperature at the plates because the system controls the mean and Delta
T independently from the plate temperature. The HFM 436 heat flow meter controls
the system so that each control thermocouple is controlled at its set point plus its
offset.
This feature is provided so that the user can obtain the desired temperature at the
plates by adjusting the control temperature offsets. The operator can adjust these
temperature offsets manually; however, this is not recommended unless you are
very familiar with the performance of the instrument. The “Automatic Mode”
should be used to let the system adjust them automatically (Figure 22).
The Auto Plate Adjust entry is used to indicate if the automatic feature should be
enabled (Y) or disabled (N).
The Secs to Settle field is used to specify the minimum amount of time between
offset adjustments. The system will make adjustments only when rough equilibrium
has been achieved and this minimum duration has elapsed. Normally this value is
set to 300 (5 minutes). The parameter has no affect if automatic mode is disabled.
The Delta and Mean Offsets fields are used to set the temperature offset (zero
should be used if no offset is desired). If automatic mode is enabled, the system will
adjust these values as a run progresses. If automatic mode is disabled, the offsets
are used for control but are not adjusted by the software.
03/07 J:\HFM436\Chapter4.doc 39
HFM 436 Operating the Instrument
Units Set
Units Set is used to indicate the data format to be used in the system (Figure 23).
S.I. *
British
Cal Results Set displays the calibration data (Figure 24) and this display can be
used to manually enter calibration data. This feature is especially useful if the
calibration data is determined using two (or more) samples.
o
# NValue Temp C
1 0.03 40
2 0.032 50
3 0.035 60
4 0.038 70
5 0.041 80
6 0.044 90
7 0.047 100
8 0.050 110
9 0.054 120
10 0.057 130
ATTENTION!
If you are entering the calibration data manually, use care to ensure that the data is
correctly entered. An error in entering calibration data may have a serious impact
on the accuracy of the results reported by the instrument.
03/07 J:\HFM436\Chapter4.doc 40
HFM 436 Operating the Instrument
No Printer *
Data Points
Just Results
The Printer Model Set (Figure 26) is used to indicate the printer that is used to
output the results directly from the measurement port.
HP *
Epson
The setting is not relevant if No Printer was selected on the Printer Options display
(Figure 25).
Messages View
The display can be scrolled to view additional messages. The last 20 messages are
saved in memory. The most recent message is presented on the top of the display.
When a message is presented, all previous messages are scrolled down one position.
A detailed discussion of all messages that are provided by the system is presented in
Appendix 3.
03/07 J:\HFM436\Chapter4.doc 41
HFM 436 Operating the Instrument
Date/Time Set
Date/Time Set is employed to set the present time and date (Figure 28).
Year 1996
Month 7
Day 19
Hour 10
Min 44
Sec 33
Dump to Printer
The Dump to Printer selection on the View menu presents a series of diagnostic
information on the printer. The information provided by this command is
potentially useful in determining the cause of a difficulty, but is not of utility to the
operator under normal circumstances.
Set Config
Version View
Version V1.20
MachineType 1
Min Mean -15
Max Mean 215
Max Delta 30.00
The values for this display are set at the factory during the manufacture of the
system and cannot be edited during normal operation.
03/07 J:\HFM436\Chapter4.doc 42
HFM 436 Operating the Instrument
Print Test Results leads to the printing of the results from the determination of the
thermal conductivity of the sample. When this option is selected, the operator is asked
to enter information about the last test run. The results from the completed test is then
printed in a standard format. The system stores only one set of test results in active
memory. When a new test is started or the system is rebooted, the results from the last
test are removed. Therefore, if the standard report is required, this report procedure
should be executed, by the operator, immediately after the completion of a test. A
typical printout is shown in Figure 30.
03/07 J:\HFM436\Chapter4.doc 43
HFM 436 Operating the Instrument
Calibrate LVDT
Version V1.20
MachineType 1
Min Mean -15
Max Mean 215
Max Delta 30.00
03/07 J:\HFM436\Chapter4.doc 44
HFM 436 Operating the Instrument
When Enter is pressed, the display will present instructions regarding the calibration
(Figure 32).
1) Install Spacer A
2) Wait 8 seconds
3) Press Enter
4) Install Spacer B
5) Wait 8 seconds
6) Press Enter
7) Done
After you have installed spacer A and the time indicated above has elapsed, press Enter.
Repeat the process for spacer B. After the system has been calibrated with spacer B, the
display will present Figure 32 again to indicate that the calibration is complete.
When the calibration is complete, press Esc to return to the main menu. If you press
Enter again, the calibration will be repeated.
Save System Data is used to save the operating conditions (equilibrium parameters,
temperature offsets, set points, instrument constants, etc.) and calibration data in non-
volatile memory. When the system is powered up, the stored operating conditions and
calibration data will be employed. When using Q-Test exclusively, you must Save
System Data after every calibration or this data will be lost when the
instrument is turned off.
03/07 J:\HFM436\Chapter4.doc 45
HFM 436 Operating the Instrument
Shutdown
Shutdown is used to access the display that leads to shut down of the instrument
(Figure 33).
Don’t Shut Down returns the display to the main menu (Figure 5). All parameters and
data are maintained.
This display indicates that the system has been properly shut down. All parameters and
data will be lost unless they were saved via Save System Data.
The reset button on the back of the instrument should be pressed to reinitialize the unit.
Q-LAB
Q-LAB provides a display to select whether the unit should be operated under remote
control (e.g. via a PC running Q-Lab) as shown in Figure 35.
Disable Q-LAB *
Enable Q-LAB
When you enable Remote Control, the last LED will be illuminated. Additional LEDs will
be illuminated during the handshake procedure. When the system is under Remote
Control (i.e. the handshaking procedure is complete), all LEDs will be illuminated.
Reset
Reset is used to reinitialize the controller. This command has no effect on the
instrument.
03/07 J:\HFM436\Chapter4.doc 46
HFM 436 Operating the Instrument
If the thermal conductivity of a rigid sample such as urethane foam or a loose fill
material in a frame is being measured, the thickness reported by the transducer should
be used for the test. For compressible samples, caution should be taken not to over
compress the sample.
03/07 J:\HFM436\Chapter4.doc 47
HFM 436 Chapter V
Appendix
03/07 J:\HFM436\Chapter5.doc
HFM 436 Appendix
Contents
03/07 J:\HFM436\Chapter5.doc
HFM 436 Appendix
Technical Data
Cooling System Forced Air Forced Air External Cooler External Cooler
Programmable 1 10 10 10
Data Points
Thermal 0.1 to 8.0 0.1 to 8.0 m2K/W 0.1 to 8.0 m2K/W 0.1 to 8.0
Resistance m2K/W m2K/W
Range
Accuracy ± 1 to 3 % ± 1 to 3 % ± 1 to 3 % ± 1 to 3 %
Weight 64 Kg 64 Kg 64 Kg 186 Kg
03/07 J:\HFM436\Chapter5.doc 1
HFM 436 Appendix
03/07 J:\HFM436\Chapter5.doc 2
HFM 436 Appendix
The Messages View selection of the View menu presents a list of information about the
operation which may be useful if a problem is observed in the system, along with the
action that should be taken by the operator.
Lawson Reset The DAC board was reset during normal operation.
There maybe a problem with the DAC board or the
connection between the board and the controller.
Menu: Out of Memory The memory for the controller has been exceeded.
Offline or Paper Out The printer was not connected to the instrument or it
ran out of paper when a print command was presented.
Factory Defaults Set Original (Factory Set) default factors have been
implemented.
Enable Test Point The operator attempted to exit the Run Set point display
without enabling any Mean T/Delta T line.
Enable CAL Point The operator attempted to exit the Calibration Set point
display without enabling any Mean T/Delta T line.
03/07 J:\HFM436\Chapter5.doc 3