Beruflich Dokumente
Kultur Dokumente
Abstract— In this work, electrical impedance spectroscopy to extract quantitative parameters of its impedance where
(EIS) has been used to characterize the electrical properties of different methods exist. Previous studies have demonstrated
the electrode-electrolyte interface and the electrode-skin that the measurement of the DC skin-electrode impedance
interface. The impedance was measured from face to face does not provide sufficient information [6]. Because of the
Ag/AgCl electrode pairs, as well as the impedance of Ag/AgCl capacitive component presented by the epidermal layer, AC
electrodes placed on the human body surface, in the frequency measurement is needed to obtain a precise estimate of the
band from 0.05Hz to 1MHz. The obtained frequency responses skin-electrode interface [6].
of each interface were analyzed to determine their electrical This paper reports the results of impedance
models, with parameterization of the described impedance
measurements for both interfaces by EIS. Then, an analysis
models using an optimization program written in Matlab
of the frequency responses (magnitude and phase) permit to
source software. The thermal noises were also calculated for
both interfaces. define the electrical models. The obtained results provide
valuable information for the design of the input stage of
Keywords— Impedance measurement; Electrode modeling; biopotential amplifier in order to avoid signal distortion and
Skin-Electrode interface; Impedance spectroscopy; Electrode to compute the electrode noise as well as the electrode-
noise. amplifier noise.
400
Impedance (Ohms)
350
450 measurement
300 simulation C1=160µF
400 simulation C1=250µF
250 simulation C1=300µF
350
Impedance (Ohms)
200
300
150
250
100 200
-1 0 1 2 3 4 5 6
10 10 10 10 10 10 10 10
150
Frequency (Hz)
100
Fig. 1. Impedance response of the electrode-electrolyte interface
50
-1 0 1 2 3 4 5 6
10 10 10 10 10 10 10 10
Frequency (Hz)
5
-5
-10 5
-15 0
-5
-20
-10
Phase (Deg)
-25
-15
-30
10
-1 0
10 10
1
10
2 3
10
4
10
5
10
6
10 -20
Fig. 5. Comparison between the measured data (phase) and the simulation
data of circuit with different values of C1
Fig. 3. Equivalent circuit model of the electrode-electrolyte impedance
50
TABLE I. FACE TO FACE ELCTRODE’S IMPEDANCE PARAMETERS
0
Parameters R1 R2 C1 C2 -10
Face to face
314Ω 118Ω 250µF 60pF -20
electrode
Phase (Deg)
One electrode 157Ω 59 Ω 125µF 30pF -30
-40
-60
The impedance is measured from electrodes placed on -1 0 1 2 3 4 5 6
10 10 10 10 10 10 10 10
the body surface of two subjects (man and woman) where Frequency (Hz)
measurements data are depicted in Fig. 6. From this figure it Fig. 8. Phase response of the electrode-skin interface
can be seen that the impedance of the electrode-skin contact
is higher for woman than for man in low frequency. The By analyzing the magnitude and phase responses of the
conductivity is also various in different parts of the body and electrode-skin contact, we defined the equivalent circuit
is highest in the forehead [12]. However, because of the model of its impedance presented in Fig. 9. The circuit
complexity of the phenomena defining the value of the model comprises the double layer capacitance C1 in parallel
interface’s impedance, this latter is extremely variable from with the charge transfer resistance R1 and a series resistance
one experiment to another. R2 as already described [11][13]. The model’s parameters
In order to model this impedance, we consider only the were estimated following the previous procedure and Table
data measured from woman performed from very low 2 shows the impedance parameters.
frequencies compared to these of man. Both magnitude and
phase responses are given hereafter in Fig.7 and Fig.8.
At very low frequencies the impedance is dominated by
the series combination of R1 and R2 whereas it tends to R2 at
high frequencies. Between these two limits, the impedance is Fig. 9. Equivalent circuit model of the electrode-skin impedance
influenced by the parallel capacitance so it decreases with
increasing frequencies. The phase angle decrease from 0°, TABLE II. ELCTRODE/SKIN’S IMPEDANCE PARAMETERS
reach -60° then it increases toward 0°.
Parameters R1 R2 C1
6
1x10
5
9x10
Measurement data_Woman
Measurement data_Man
value 929KΩ 647Ω 20nF
5
8x10
Impedance (Ohms)
5
7x10
5
The simulation results of the defined model for both
6x10
5
magnitude and phase were compared to the measured data.
5x10
5
4x10
The phase plots (Fig.10) show that the simulation plot
5
3x10
matches the measurements in a small frequency range only;
5
2x10
this indicates that the three element model is not suitable for
5
1x10
describing the electrode-skin impedance [6]. Kaczmarek and
0 Webster presented a more accurate model which describes
10
0 1
10 10
2 3
10
4
10 10
5 6
10 the skin-electrode interface as a double time constant system
Frequency (Hz)
with probably time-varying parameters [14]. In previous
work [6] [14], the magnitude and phase responses returned
Fig. 6. Impedance of the electrode-skin contact in Man and Woman satisfactory results when the double time constant model was
1x10
6
used.
5 10
9x10
5 0 Measure
8x10
Simulation
Impedance (Ohms)
5 -10
7x10
5 -20
6x10
5 -30
5x10
Pase (Deg)
5 -40
4x10
5 -50
3x10
5 -60
2x10
5 -70
1x10
-80
0
-1 0 1 2 3 4 5 6
10 10 10 10 10 10 10 10 -90
Frequency (Hz) -1 0 1 2 3 4 5 6
10 10 10 10 10 10 10 10
Frequency (Hz)
Fig. 7. Impedance response of the electrode-skin interface
Fig. 10. Comparison between the measured and the simulation data
(phase response)
51
C. Noise Analysis decreases with frequency, both for electrodes placed face to
face and for electrodes placed on the skin. Measurements on
In the recording of biomedical signals, a noise electrodes placed face to face showed that the impedance of
component is introduced by all subsystems in the metal-gel interface presents low impedance value. These
measurement chain. There are two noise sources that are latter imply that the high impedance of electrodes placed on
inherent to the measurement procedure: the electrode noise the skin mainly originates from the gel-skin interface and the
and the amplifier noise. Skin-electrode interface noise was upper skin layer. The electrode voltage noise power spectral
simulated with low noise bioamplifier developed by us [15]. density also appears to originate for the main part, from the
Fig.11 shows the bioamplifier input voltage noise with its gel-skin interface. Electrode-amplifier noise is characterized
inputs shorted, then with impedance presenting the by 1/f spectrum in low frequency range, and a white noise in
electrode-skin interface connected to its inputs. It can be higher frequencies.
clearly seen that the noise in both case has a 1/f character
In future works, the electrode-skin interface must be
with a maximum of about 5µV/√Hz at low frequencies.
modeled using the double time constant system. On the other
However, amplifier voltage noise falls and tends to
hand, for further investigation into the origin of the noise of
16nV/√Hz at higher frequencies along the desired frequency
surface electrodes, the noise of the interface must be
range (of EEG and ECG), while electrode-amplifier noise
measured then compared to the equivalent thermal noise
tends to about 205nV/√Hz. Thus, we notice that the
from the impedance electrode.
significant noise component is introduced by the electrode-
skin interface impedance.
REFERENCES
Furthermore, we computed the thermal noise
corresponding to the real part of the impedance both for [1] S.R. Wiese, P. Anheier, R.D. Connemara, A.T. Mollner, T.F. Neils,
J.A. Kahn, J.G. Webster, “Electrocardiographic motion artifact versus
electrode placed face to face and for electrodes placed on the electrode impedance,” IEEE Trans. Biom. Eng., Vol.52,No.1, pp.136-
skin, as shown in Fig.12. We observe that thermal noise of 139, 2005.
pre-gelled electrodes placed face to face appears to be lower [2] A.L. Toazza, F. Mendes de Azevedo, J.M. Neto, “Microcontrolled
than the noise generated from the electrode–skin interface system for measuring skin/electrode impedance in biomedical
impedance. recordings,” IEEE Int. Conf. Dev, Circ. and Sys., pp.278-281, 1998.
[3] M. Fernandez, M.A. Garcia, M. Vergas, J.J. Ramos Castro, “Noise in
10
instrumentation,” Wiley Encyc. Biom. Engi., John Wiley & Sons,
Amplifier with E/S impedance
Amplifier with inputs shorted 2006.
8 [4] M. Fernandez, R.Pallas-Areny, “Ag-AgCl electrode noise in high
resolution ECG measurements,” Biom. Instr. Tech., pp. 125-130, 34
)
2,2
VN(nV/Hz0.5)
52