Scanvaegt International A/S

Legal Set-up and


Calibration Manual

Scanvaegt International A/S


P.O. Pedersens Vej 18
DK-8200 Aarhus N
Denmark

8500 Indicator Series Legal Set-up and Calibration Manual SN 912475 Rev. C
Legal Set-up and Calibration Manual
Table of contents
Introduction .................................................................... 3
Non-ratiometric calibration .......................................... 3
Legal set-up and calibration ......................................... 4
How to start legal set-up and calibration ............ 4
Keyboard functions during legal set-up and
calibration ............................................................... 5
Legal set-up menu points....................................... 6
Gravity compensation .......................................... 12
How to set the scale for gravity
compensation ................................................ 13
Correction factor table ................................................ 16
Calibration menu points ...................................... 20
Calibration error codes ....................................... 23
How to finish legal set-up and calibration ......... 25
Cold power-up .............................................. 25
Monitor menu ............................................................... 26
How to start the monitor menu........................... 26
Monitor menu points ........................................... 28
How to finish the monitor menu ......................... 28
Table of legal set-up and calibration menu points .... 29

912475 2002-01-16 Page 1 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

© 2002 Scanvaegt International A/S

912475 2002-01-16 Page 2 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Introduction
The intention of developing a legal set-up and calibration procedure
for the 8500 indicator series is to ease the working process and to
minimize the number of errors that may occur during legal set-up
and calibration.

Therefore, the operator will be guided through the part of the


procedure that concerns the legal set-up. At any legal set-up menu
point it is possible to select only allowed values placed in a rolling
memory stack.

The best calibration results will be obtained by warming up the


scale for at least 60 min. prior to calibration!

60 min.

This manual covers different indicators in the 8500 indicator series,


with different keyboard layouts. Therefore, it is necessary to show
the different keys in this manual. The keys are separated by a “/”.
The key in front of the “/” is active on an 8564 MK-I indicator, and
the key after the “/” is active on the rest of the indicator series.

Non-ratiometric calibration
The 8500 series of indicators initially has two calibration sequences:

• Non-ratiometric calibration (NRC).


• Legal set-up and calibration.

The NRC is an internal calibration of the A/D converter on the


weighing board and, when performed, the A/D converter will
compensate for any offset errors not related to ratiometric
measurements.
The first time ever the indicator is powered up,
followed by will appear in the weight display, and
the indicator will be forced into the NRC mode. The NRC is to be
performed as described:
1. In the NRC mode the indicator shows flashing in

912475 2002-01-16 Page 3 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
the weight display.

2. Press the / (registration key). The indicator now


performs the NRC, and will be forced into legal set-up and
calibration when finished by displaying in the
weight display.

3. Proceed with the legal set-up and calibration.

The non-ratiometric calibration needs only to be done when


changing the load cell, load unit, cable length or weighing
board! This is done by typing in the password for non-
ratiometric calibration. See schematic below, and follow above
description.

Password entry

In the user menu and the password entry, there will be an


automatic time-out after 30 sec., if the keys are untouched.

8564 MK I: Pressing and holding the keys and during power-


on will always start the non-ratiometric calibration!

8564 MK II, 8526, 8544 : Pressing and holding the keys and
during power-on will always start the non-ratiometric calibration!
Legal set-up and calibration
How to start legal set-up and calibration

912475 2002-01-16 Page 4 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

• Press the / (registration key) for more than 1.5 sec.

• Select menu by use of (zero-set key) or


(tare key), and press the / (registration key).

• The indicator is now in password entry where you will be able to


key in the password by use of the numerical keyboard.

• Key in the password for the legal set-up and calibration menu.
The indicator is now in the legal set-up and calibration menu.

User menu

Password entry

In the user menu and the password


entry, there will be an automatic time-out
after 30 sec., if the keys are untouched.

Keyboard functions during legal set-up and


calibration

When the indicator is in the legal set-up and calibration menu, the
different keys contain two functions:

• A primary function
• A secondary function.
The primary function is accessed by a short click on the key. A
click and hold on the key for more than 1.5 sec. access the
secondary function.

912475 2002-01-16 Page 5 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

If an invalid key is activated, a long beep


signal is given, and the activation will be
ignored.
When a valid and operational key is
activated, a short beep signal is given.

8564-1 8564-2 8526 8544 Primary key Secondary key


function function

Increment menu Fast increment


point menu point

Decrement menu Fast decrement


point menu point

Decrement menu Fast decrement


point value menu point value

Not Increment menu Fast increment


available point value menu point value

Exit calibration

Please note that changes made to menu point values in the legal
set-up and calibration menu only are stored after pressing
(zero-set key) or (tare key).

Legal set-up menu points

As earlier mentioned legal set-up and calibration starts by entering


the password. Upon doing this, the indicator will respond with the
number of the first legal set-up menu point.

The meaning and function of the legal set-up menu points and the
calibration points will be explained in the following.
Changing the legal set-up menu points and/or performing a new
calibration will result in a change of the check number for the legal
set-up and calibration menu. On verified scales this will result in a
new verification of the scale.

912475 2002-01-16 Page 6 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

Menu point Description Options Default Note


Initial zero setting 0 Semi automatic 1 Setting the
Select initial zero 1 Automatic option to 0 will
setting condition. result in that the
initial zero
setting is
performed by
pressing the

(set-zero
key).

Menu point Description Options Default Note


Warm power-on 0 No 1
Select warm power-on, 1 Yes
when power-off is
shorter than 30 sec.

Menu point Description Options Default Note


Number of decimals 00 / 0 / 0.0 / 0.00 / 0.00
Select actual number of 0.000
decimals on display.
This is a point
which, if
altered, requires
new weight
calibration at
calibration
points

and

912475 2002-01-16 Page 7 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Menu point Description Options Default Note
Actual scale interval 1/2/5 1
Select actual scale for 0 to 0.000
interval d, as typed on
the sticker on the front 10 / 20 / 50 This is a point
layer of the indicator for 00 which, if
cabinet. altered, requires
It is shown in display new weight cal-
divisions depending on ibration at
legal set-up menu point calibration
points

and

Ex 1:
If value of is , the values for

are , or

.
Ex 2:

If value of is , the values for

are , or

Menu point Description Options Default Note


Max. capacity 250-6000 3000 In non-verified
Select max. capacity, as [divisions] [divisions] scales the option
typed on the sticker on is 250-7500
the front layer of the step 250 divisions.
indicator cabinet. [divisions]

This is a point
which, if
altered, requires
new weight cal-
ibration at
calibration
points

and

Max. capacity is calculated in the following way:

Max.capacity = Display division * No. of divisions

Value of = Value of * No. of divisions

912475 2002-01-16 Page 8 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Menu point Description Options Default Note
Is the required lb 0 No 1 Yes This entry has
capacity less than the 1 Yes only relevance
chosen kg capacity? for dual
ex1: graduation
60 kg - 132 lb scales with
60 kg * 2.2 = 132 lb d = 2 g/0.005 lb,
132 lb132 lb ? Yes d = 20 g/0.05 lb,
(1) d = 200 g/0.5 lb
ex2: d = 2 kg/5 lb.
40 kg - 100 lb
40 kg * 2.2 = 88 lb
100 lb88 lb ? No
(0)
ex3:
30 kg - 60 lb
30 kg * 2.2 = 66 lb
60 lb66 lb ? Yes (1)

Menu point Description Options Default Note


Overload limit 0 [divisions] 8
Select overload limit 8 [divisions] [divisions]
before display blanks.

Menu point Description Options Default Note


Digital filter 1/2/3/4 1 Increasing this
Select type of digital value will result
filter. in a slower, but
more stable
scale. This can
be of advantage
in certain
installations.

Menu point Description Options Default Note


No-motion window 0.1 [divisions] 1.0 On non-verified
Select no-motion 0.2 [divisions] [division] scales the
window size. 0.5 [divisions] options are:
1.0 [divisions] 0.1/0.2/0.5/1.0/
2.0/5.0
[divisions]

Menu point Description Options Default Note


No-motion time 0.1 - 1.0 [sec.] 0.5 [sec.]
Select no-motion time. step 0.1 [sec.]

Menu point Description Options Default Note

912475 2002-01-16 Page 9 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Zero track window 0.5 0.5 On non-verified
Select zero track [divisions] [divisions scales the
window size. ] options are:
0.5/1.0/2.0/5.0/1
0.0 [divisions]

Menu point Description Options Default Note


Zero track speed 0.1 / 0.2 / 0.5 0.5 [sec.] On non-verified
Select zero track speed. [sec.] scales the
options are:
0.1/0.2/0.5/1.0
[sec.]

Menu point Description Options Default Note


Minimum legal 20 20 On non-verified
weighing limit [divisions] [divisions scales the
] options are:
1/2/5/10/20
[divisions]

Menu point Description Options Default Note


Indicator type 8526 / 8564-1
Select indicator type. 8544 / 8564-1 /
8564-2

Menu point Description Options Default Note


Indicator alibi 0-125 0
identifier step 1
Select indicator alibi
identifier.

Menu point Description Options Default Note


Alibi system 0 No alibi 0
Select alibi system. 1 Alibi printer
2 Alibi memory
3 Alibi harddisk

Menu point Description Options Default Note


Software option 0 Normal 0
Select software option. 1 Swinging
Load Scale
2 Positive Hopper
Scale
3 Negative Hopper
Scale

Menu point Description Options Default Note


Graduation 0 Single graduation 0
Select graduation 1 Dual graduation
Single (kg)
Dual (kg/lb)
Menu point Description Options Default Note

912475 2002-01-16 Page 10 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Start-up graduation 0 kg 0
Select start-up 1 lb
graduation on dual
graduation scale.

Menu point Description Options Default Note


Initial delay 0.0 - 10.0 [sec.] 0.0 [sec.]
Select initial delay.
Software option 1, 2 step 0.1 [sec.]
and 3 only.

Menu point Description Options Default Note


Cycle time 0.3 - 10.0 [sec.] 0.3 [sec.]
Select cycle time.
Software option 1, 2 step 0.1 [sec.]
and 3 only.

Menu point Description Options Default Note


Termination window 5-125 5
Select termination [divisions] [divisions]
window.
Software option 1, 2 step 1
and 3 only. [divisions]

Menu point Description Options Default Note


Gravity correction 0 – 45.00 12.56 Refer to section
factor. Step 0.01 56°, 0 m “Correction
Denmark. factor table”.

Menu point Description Options Default Note


Minimum load
calibration starts
Calibration points. registration of
minimum load. Calibration
Refer to section point
“Calibration menu
points” for more changes to
information about is performed
menu point
calibration. before
. calibration point

(refer to
“Calibration
menu points”).

Menu point Description Options Default Note


Fine calibration -3.0 - 3.0 0.0
Select fine calibration at [divisions] [divisions]
maximum capacity.
step 0.5 Normally, it is
[divisions] not necessary to
make a fine
calibration.
Fine calibration
will be reset at
every new
weight
calibration.
Menu point Description Options Default Note

912475 2002-01-16 Page 11 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Where to linearize 100 to max. capacity 100
Select where to [divisions]
linearize. Step
100 [divisions] Normally, it is
not necessary to
linearize.
Linearization
will be reset at
every new
weight
calibration.

Menu point Description Options Default Note


How much to linearize -3.0 - 3.0 0.0
Select how much to [divisions] [divisions]
linearize.
step 0.1 Normally, it is
[divisions] not necessary to
linearize.
Linearization
will be reset at
every new
weight
calibration.

Gravity compensation
When moving around the world, gravity changes due to change in
latitude and altitude. The gravity is highest at the poles, and lowest at
equator.

This also means that the gravity affects the weighing results from a
scale if it is calibrated at one position, and then moved to another
position.

Example: A 3 kg / 1 g scale calibrated in Aarhus, Denmark, and


then moved to Mellac, France.

912475 2002-01-16 Page 12 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

Aarhus, Denmark: Latitude = 56°, Altitude = 0 meters


Mellac, France: Latitude = 48°, Altitude = 0 meters

Moving the scale from Aarhus, Denmark, to Mellac, France, will


actually affect the result with approx. 2.2 g. The scale will then
readout 2.2 g less at maximum capacity than expected, as you move
the scale closer to equator, where the gravity is lower.

Please refer to section “How to set the scale for gravity


compensation” for a description on how to set the scale for gravity
compensation.

How to set the scale for gravity compensation

The following procedure describes what to do to set the scale for


gravity compensation.

1. First the latitude and the altitude for the calibration position
need to be provided. Then the correction factor table to find the
calibration correction factor. (Please refer to section “Correction
factor table”).
2. Set the calibration point to the found correction
factor.

Use the keys (zero-set key) or (tare key) to select


calibration point .

Use the keys / or / to change the value. If the


keys are pressed and released (< 1.5 sec.), the value will change
only with steps of 0.01.
If the key is pressed and hold (> 1.5 sec.), the value will change
with steps of 0.10.

912475 2002-01-16 Page 13 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
3. Calibrate the scale. (Please refer to section “Calibration menu
points”).

4. After the calibration, finish the legal set-up and calibration menu
(refer to section “How to finish legal set-up and calibration”),
and perform the necessary weighing tests (e.g. linearization).

5. Provide the latitude and altitude of the destination position, and


find the destination correction factor in the correction factor
table. (Please refer to section “Correction factor table”).

6. Start the legal set-up and calibration menu (refer to section


“How to start legal set-up and calibration”) and set the
calibration point to the found destination
correction factor.

Use the keys (zero-set key) or (tare key) to select


calibration point .

Use the keys / or / to change the value. If the


keys are pressed and released (< 1.5 sec.), the value will change
only with steps of 0.01.
If the key is pressed and hold (> 1.5 sec.), the value will change
with steps of 0.10.

7. Finish the legal set-up and calibration menu. (Please refer to


section “How to finish legal set-up and calibration”).

On the following pages you will find an example showing the


procedure for a scale that is calibrated in Aarhus, Denmark, and
moved to Mellac, France.

912475 2002-01-16 Page 14 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Example (from Aarhus, Denmark, to Mellac, France):

In this example the scale is calibrated in Aarhus, Denmark, (latitude


56 and altitude 0 meter) and then sent to the destination Mellac,
France, (latitude 48° and 0 meters).

1. From the correction factor table the correction factor for Aarhus,
Denmark, is found to be 12.56.
1000 m

250 m
0 meter
90°
75°
60°

45°

30°
Aarhus, Denmark latitude 56° and Altitude
0 meters ~ correction factor = 12.56
15° (The place where the scale was calibrated).

2. The calibration correction factor for Aarhus, Denmark, was found


to be 12.56. This value has to be stored in point
in the legal set-up and calibration menu.

Use the keys (zero-set key) or (tare key) to select


calibration point .

Press the key / or / until the display shows


(default value).

3. Calibrate the scale. (Please refer to section “Calibration menu


points”).
4. Finish the legal set-up and calibration menu, and perform

912475 2002-01-16 Page 15 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
necessary weighing tests.

5. From the correction factor table the correction factor for Mellac,
France, is found to be 17.93.
90° 1000 m
75°
60°
250 m
0 meter
45°

30°
Mellac, France latitude 48° and Altitude 0
meters ~ correction factor = 17.93
15° (The destination for the scale).

6. Enter the legal set-up and calibration menu. The destination


correction factor was found to be 17.93. This value has to be
stored in point in the legal set-up and
calibration menu.

Use the keys (zero-set key) or (tare key) to select


calibration point .

Press the key / or / until the display shows


.
Press either the keys (zero-set key) or (tare key) to
leave the menu point .

7. Finish the legal set-up and calibration menu. (Refer to section


“How to finish legal set-up and calibration”).

Correction factor table

912475 2002-01-16 Page 16 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

Altitude[m]

Latitude
[°] 0 200 400 600 800 1000 1200 1400 1600 1800 2000
0 39,79 40,26 40,73 41,20 41,68 42,15 42,62 43,09 43,56 44,04 44,51
1 39,78 40,25 40,72 41,19 41,66 42,14 42,61 43,08 43,55 44,02 44,50
2 39,74 40,21 40,68 41,16 41,63 42,10 42,57 43,04 43,52 43,99 44,46
3 39,68 40,15 40,62 41,10 41,57 42,04 42,51 42,98 43,46 43,93 44,40
4 39,60 40,07 40,54 41,01 41,48 41,96 42,43 42,90 43,37 43,84 44,32
5 39,49 39,96 40,43 40,90 41,38 41,85 42,32 42,79 43,26 43,74 44,21
6 39,36 39,83 40,30 40,77 41,24 41,72 42,19 42,66 43,13 43,60 44,08
7 39,20 39,67 40,15 40,62 41,09 41,56 42,03 42,50 42,98 43,45 43,92
8 39,02 39,50 39,97 40,44 40,91 41,38 41,85 42,33 42,80 43,27 43,74
9 38,82 39,29 39,77 40,24 40,71 41,18 41,65 42,13 42,60 43,07 43,54
10 38,60 39,07 39,54 40,01 40,48 40,96 41,43 41,90 42,37 42,84 43,32
11 38,35 38,82 39,29 39,77 40,24 40,71 41,18 41,65 42,13 42,60 43,07
12 38,08 38,55 39,02 39,50 39,97 40,44 40,91 41,38 41,86 42,33 42,80
13 37,79 38,26 38,73 39,21 39,68 40,15 40,62 41,09 41,56 42,04 42,51
14 37,48 37,95 38,42 38,89 39,36 39,84 40,31 40,78 41,25 41,72 42,20
15 37,14 37,61 38,09 38,56 39,03 39,50 39,97 40,45 40,92 41,39 41,86
16 36,79 37,26 37,73 38,20 38,67 39,15 39,62 40,09 40,56 41,03 41,51
17 36,41 36,88 37,36 37,83 38,30 38,77 39,24 39,71 40,19 40,66 41,13
18 36,02 36,49 36,96 37,43 37,90 38,38 38,85 39,32 39,79 40,26 40,73
19 35,60 36,07 36,54 37,02 37,49 37,96 38,43 38,90 39,38 39,85 40,32
20 35,17 35,64 36,11 36,58 37,05 37,53 38,00 38,47 38,94 39,41 39,89
21 34,71 35,19 35,66 36,13 36,60 37,07 37,55 38,02 38,49 38,96 39,43
22 34,24 34,72 35,19 35,66 36,13 36,60 37,07 37,55 38,02 38,49 38,96
23 33,76 34,23 34,70 35,17 35,64 36,11 36,59 37,06 37,53 38,00 38,47
24 33,25 33,72 34,19 34,67 35,14 35,61 36,08 36,55 37,03 37,50 37,97
25 32,73 33,20 33,67 34,15 34,62 35,09 35,56 36,03 36,5 36,98 37,45
26 32,19 32,67 33,14 33,61 34,08 34,55 35,02 35,5 35,97 36,44 36,91
27 31,64 32,11 32,59 33,06 33,53 34,00 34,47 34,95 35,42 35,89 36,36
28 31,08 31,55 32,02 32,49 32,96 33,44 33,91 34,38 34,85 35,32 35,8
29 30,50 30,97 31,44 31,91 32,38 32,86 33,33 33,80 34,27 34,74 35,22
30 29,91 30,38 30,85 31,32 31,79 32,26 32,74 33,21 33,68 34,15 34,62

912475 2002-01-16 Page 17 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

Latitude Altitude[m]
[°] 0 200 400 600 800 1000 1200 1400 1600 1800 2000
31 29,30 29,77 30,24 30,72 31,19 31,66 32,13 32,60 33,08 33,55 34,02
32 28,69 29,16 29,63 30,10 30,57 31,04 31,52 31,99 32,46 32,93 33,40
33 28,06 28,53 29,00 29,47 29,95 30,42 30,89 31,36 31,83 32,31 32,78
34 27,42 27,89 28,37 28,84 29,31 29,78 30,25 30,73 31,20 31,67 32,14
35 26,78 27,25 27,72 28,19 28,67 29,14 29,61 30,08 30,55 31,02 31,50
36 26,12 26,60 27,07 27,54 28,01 28,48 28,96 29,43 29,90 30,37 30,84
37 25,46 25,94 26,41 26,88 27,35 27,82 28,29 28,77 29,24 29,71 30,18
38 24,80 25,27 25,74 26,21 26,68 27,15 27,63 28,10 28,57 29,04 29,51
39 24,12 24,59 25,07 25,54 26,01 26,48 26,95 27,42 27,90 28,37 28,84
40 23,44 23,91 24,39 24,86 25,33 25,80 26,27 26,75 27,22 27,69 28,16
41 22,76 23,23 23,70 24,17 24,65 25,12 25,59 26,06 26,53 27,01 27,48
42 22,07 22,54 23,02 23,49 23,96 24,43 24,90 25,38 25,85 26,32 26,79
43 21,38 21,86 22,33 22,80 23,27 23,74 24,21 24,69 25,16 25,63 26,10
44 20,69 21,16 21,64 22,11 22,58 23,05 23,52 24,00 24,47 24,94 25,41
45 20,00 20,47 20,94 21,42 21,89 22,36 22,83 23,30 23,78 24,25 24,72
46 19,31 19,78 20,25 20,72 21,20 21,67 22,14 22,61 23,08 23,56 24,03
47 18,62 19,09 19,56 20,03 20,50 20,98 21,45 21,92 22,39 22,86 23,34
48 17,93 18,40 18,87 19,34 19,81 20,29 20,76 21,23 21,70 22,17 22,65
49 17,24 17,71 18,18 18,65 19,13 19,60 20,07 20,54 21,01 21,49 21,96
50 16,56 17,03 17,50 17,97 18,44 18,91 19,39 19,86 20,33 20,80 21,27
51 15,88 16,35 16,82 17,29 17,76 18,23 18,71 19,18 19,65 20,12 20,59
52 15,20 15,67 16,14 16,62 17,09 17,56 18,03 18,5 18,98 19,45 19,92
53 14,53 15,00 15,47 15,95 16,42 16,89 17,36 17,83 18,31 18,78 19,25
54 13,87 14,34 14,81 15,28 15,76 16,23 16,70 17,17 17,64 18,11 18,59
55 13,21 13,68 14,16 14,63 15,10 15,57 16,04 16,52 16,99 17,46 17,93
56 12,56 13,04 13,51 13,98 14,45 14,92 15,4 15,87 16,34 16,81 17,28
57 11,93 12,40 12,87 13,34 13,81 14,29 14,76 15,23 15,70 16,17 16,65
58 11,30 11,77 12,24 12,71 13,19 13,66 14,13 14,60 15,07 15,55 16,02
59 10,68 11,15 11,62 12,10 12,57 13,04 13,51 13,98 14,46 14,93 15,40
60 10,07 10,55 11,02 11,49 11,96 12,43 12,90 13,38 13,85 14,32 14,79

912475 2002-01-16 Page 18 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

Latitude Altitude[m]
[°] 0 200 400 600 800 1000 1200 1400 1600 1800 2000
61 9,48 9,95 10,42 10,89 11,37 11,84 12,31 12,78 13,25 13,73 14,20
62 8,90 9,37 9,84 10,31 10,78 11,26 11,73 12,20 12,67 13,14 13,62
63 8,33 8,80 9,27 9,74 10,22 10,69 11,16 11,63 12,10 12,58 13,05
64 7,77 8,25 8,72 9,19 9,66 10,13 10,61 11,08 11,55 12,02 12,49
65 7,23 7,71 8,18 8,65 9,12 9,59 10,07 10,54 11,01 11,48 11,95
66 6,71 7,18 7,65 8,13 8,60 9,07 9,54 10,01 10,49 10,96 11,43
67 6,20 6,67 7,15 7,62 8,09 8,56 9,03 9,51 9,98 10,45 10,92
68 5,71 6,18 6,66 7,13 7,60 8,07 8,54 9,02 9,49 9,96 10,43
69 5,24 5,71 6,18 6,65 7,13 7,60 8,07 8,54 9,01 9,49 9,96
70 4,78 5,25 5,73 6,20 6,67 7,14 7,61 8,09 8,56 9,03 9,50
71 4,35 4,82 5,29 5,76 6,23 6,71 7,18 7,65 8,12 8,59 9,06
72 3,93 4,40 4,87 5,34 5,82 6,29 6,76 7,23 7,70 8,17 8,65
73 3,53 4,00 4,47 4,94 5,42 5,89 6,36 6,83 7,30 7,78 8,25
74 3,15 3,62 4,09 4,57 5,04 5,51 5,98 6,45 6,93 7,40 7,87
75 2,79 3,26 3,74 4,21 4,68 5,15 5,62 6,10 6,57 7,04 7,51
76 2,46 2,93 3,40 3,87 4,34 4,82 5,29 5,76 6,23 6,70 7,18
77 2,14 2,61 3,08 3,56 4,03 4,50 4,97 5,44 5,92 6,39 6,86
78 1,85 2,32 2,79 3,26 3,73 4,21 4,68 5,15 5,62 6,09 6,57
79 1,58 2,05 2,52 2,99 3,46 3,93 4,41 4,88 5,35 5,82 6,29
80 1,33 1,80 2,27 2,74 3,21 3,69 4,16 4,63 5,10 5,57 6,05
81 1,10 1,57 2,04 2,52 2,99 3,46 3,93 4,40 4,88 5,35 5,82
82 0,90 1,37 1,84 2,31 2,78 3,26 3,73 4,20 4,67 5,14 5,62
83 0,72 1,19 1,66 2,13 2,60 3,08 3,55 4,02 4,49 4,96 5,44
84 0,56 1,03 1,50 1,98 2,45 2,92 3,39 3,86 4,34 4,81 5,28
85 0,43 0,90 1,37 1,84 2,32 2,79 3,26 3,73 4,20 4,67 5,15
86 0,32 0,79 1,26 1,73 2,21 2,68 3,15 3,62 4,09 4,57 5,04
87 0,23 0,71 1,18 1,65 2,12 2,59 3,07 3,54 4,01 4,48 4,95
88 0,17 0,65 1,12 1,59 2,06 2,53 3,01 3,48 3,95 4,42 4,89
89 0,14 0,61 1,08 1,55 2,02 2,50 2,97 3,44 3,91 4,38 4,86
90 0,13 0,60 1,07 1,54 2,01 2,48 2,96 3,43 3,90 4,37 4,84

912475 2002-01-16 Page 19 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Calibration menu points

The calibration of the 8500 indicator series consists of two parts.


First the calibration of the minimum load, and then the calibration of
the maximum load.

Minimum load calibration:


Select minimum load calibration point . At this
point only the keys (zero-set key) and (tare key) will be
accepted.
• Pressing (tare key) decrements the menu point to
.
• Pressing (zero-set key) will start the minimum load
calibration .

1. Respond to this by emptying the goods plate.

2. Press the (zero-set key) to start registration of minimum


load.
3. The display will now show flashing bars .

4. At this point two things may happen:

• If the registration can be completed without any errors, the


indicator will respond by showing the calibration point
(maximum load calibration).
• An error code may occur. If this happens, please refer to the
section “Calibration error codes”, page 23.
Maximum load calibration:

912475 2002-01-16 Page 20 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

1. Pressing the (zero-set key) will result in displaying of the


required calibration load which normally equals the selected
maximum capacity.

2. Now place the required calibration load on the goods plate.

3. Press the (zero-set key) to start registration of maximum


load.

4. The display will now show flashing bars .

5. Like at calibration point (minimum load


calibration), two things may happen:

• If the registration is completed successfully, the indicator will


respond by showing the first calibration menu point, which is
calibration point .
• An error code may occur. If this happens, please refer to the
section “Calibration error codes”, page 23.

912475 2002-01-16 Page 21 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

It is also possible to perform the maximum load calibration, using a


calibration load weighing less than maximum capacity. This
procedure is certainly not recommendable, but may be used at
very large capacities, or in an emergency situation. In any case
it is important to choose as large a calibration load as possible.

Any calibration load can be selected within certain limits:


• Calibration load cannot exceed the maximum capacity

• Calibration load cannot be below 100 divisions

.
When the display is showing the required calibration load

, it will be possible to activate the keys / or


/ and thereby decrease or increase the calibration load by 1
display division.

Keeping the keys / or / activated will result in the


following:

• The requested calibration load decreases/increases by steps of 10


display divisions.
• The decreasing/increasing will run at high speed until the button
is released.
When calibration has been finished, a weighing can be
performed. After having performed this weighing test, it will
be known to the operator if any fine calibration or
linearization is needed (normally not necessary). But if this is
a fact, the indicator will have to be put back into the
calibration mode. Please refer to "How to start legal set-up
and calibration", page 4.
By stepping down with the key (tare key) from
, it is possible to reach the calibration point
numbers , and

912475 2002-01-16 Page 22 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Calibration error codes

This section provides a listing of the error codes that can occur
during calibration.

This error code may be given during calibration at calibration points


and .

Meaning: During calibration the signal from the load


cell is too unstable.

Example: This error code is given if the goods plate


vibrates during calibration.

How to continue: Press the / (registration key).

The indicator responds by stepping back to the


current calibration point, either
or .

It will then be necessary to perform this


calibration point once more.

This error code may be given during calibration at calibration point


.

Meaning: The dynamic range is not big enough.

Examples: This error code will be given if the goods plate

912475 2002-01-16 Page 23 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
is not unloaded, before minimum load
calibration, or if the pre-load is too big.

How to continue: Press the / (registration key).

The indicator steps back to calibration point


. It will then be necessary to
perform this calibration point once more.

This error code may be given during calibration at calibration point


.
Meaning: The load on the goods plate or the pre-load is
too heavy, or a hardware error has been
discovered.

How to continue: Press the / (registration key).

The indicator responds by stepping back to the


calibration point .

This error code may be given during calibration at calibration point


or .

Meaning: The load on the goods plate or the pre-load is


too light, or a hardware error has been
discovered.
How to continue: Press the / (registration key).

912475 2002-01-16 Page 24 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual

The indicator responds by stepping back to the


previous calibration point.

This error code may be given during calibration at calibration point


.

Meaning: The registered maximum load, incl. 5% of the


dynamic range, exceeds the permitted
maximum value.

How to continue: Press the / (registration key).

The indicator responds by stepping back to


calibration point .
Step back to calibration point
by pressing (tare key). Find out what
caused the error. Either the calibration load or
the pre-load is too heavy.

It is now necessary to perform a complete


calibration at calibration points
and once
more.

How to finish legal set-up and calibration

Press the / (registration key) for more than 1.5 sec.

When exiting the legal set-up and calibration menu, a check number
for the legal set-up and calibration will be created, due to change in
the legal set-up and/or a new calibration. The indicator will then
perform a cold power-up.
Cold power-up

912475 2002-01-16 Page 25 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
During cold power-up the check number, the indicator type and the
software version are shown in the weight display in sequences of 1.5
sec. After that the display and all annunciators will illuminate before
initial zero setting.

1. Random check number for the legal set-up


and calibration.
2. Indicator type number: 1.
t64 means SCI 8564, t26 means ScanWI
8526, t44 means ScanWIC 8544 option
(00 means Option 0, 01 means option 1,
etc.).
3. Current software version. 2.
4. Display and all announciators are
illuminated.
5. Ready for initial zero setting.

3.
Option 0: Normal weighing
Option 1: Swinging load weighing
Option 2: Positive Hopper weighing
Option 3: Negative Hopper weighing

4.

5.

Monitor menu
The monitor menu is for service purposes only, and is primarily used
to check the weighing performance of the indicator.
How to start the monitor menu

912475 2002-01-16 Page 26 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
• Press the / (registration key) for more than 1.5 sec.
• Select menu by use of the (zero-set key) or
the (tare key), and press the / (registration key).

• The indicator is now in password entry where you will be able to


key in the password by use of the numerical keyboard.

• Key in the password for the monitor menu. The indicator is now
in the monitor menu.

In the user menu and the password entry, there will be an


automatic time-out after 30 sec., if the keys are untouched.

912475 2002-01-16 Page 27 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Monitor menu points

The monitor mode has 15 different monitors. The monitor numbers


are indicated with the binary system of notation by means of the
"zero", "gross", "net" and “preset tare” annunciators.

Furthermore, the monitor number is shown in the secondary guide


display (SGD). See below for an overview of the different monitor
numbers.

Monitor 1: Counts from A/D-converter 1.

Monitor 2: Counts at zero calibration.

Monitor 3: Counts at max. capacity calibration.

Monitor 4: Weight result in subdivisions. 10


subdivisions = 1 division.

Monitor 5: Weight result in divisions.

Monitor 6: Weight result.

Monitor 7-15: Are not presently used.

How to finish the monitor menu

When pressing the / (registration key) for more than 1.5


sec., the indicator returns to weighing mode.

912475 2002-01-16 Page 28 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Table of legal set-up and calibration menu
points
Additional values for program version n (non-verified): ().
[s] = Seconds , [d] = Division
The high marked values are default.

Point Description Value


010 Semi automatic / automatic initial 0/1
zero setting
015 Cold / warm power-up 0/1
025 Number of decimals 0.000 / 0.00 / 0.0 / 0 / 00
030 Count - by division size 1 / 2 / 5 (10 / 20 for number of deci-
mals = 00)
035 Max. capacity 250 to 6000 (7500) [d], shown in dis-
(Count - by division size * number play division. Default: 3000 divisions
of divisions)
040 lb capacity < kg capacity (yes/no) 0/1
060 Overload limit 0 / 8 [d]
065 Filter type 1/2/3/4
080 No-motion window 0.1 / 0.2 / 0.5 / 1 (/ 2 / 5) [d]
085 No-motion time 0.1 / 0.2 / 0.3 / 0.4 / 0.5 / 0.6 / 0.7 /
0.8 / 0.9 / 1.0 [s]
095 Zero track window 0.5 (/ 1.0 / 2.0 / 5.0 / 10.0) [d]
100 Zero track speed 0.1 / 0.2 / 0.5 (/ 1.0) [s]
105 Legal minimum (Min) 20 (1 / 2 / 5 / 10 ) [d]
130 Indicator type 8526 / 8544 / 8564-1 / 8564-2
131 Indicator identifier (part of the alibi 0 – 125
number)
132 Alibi system (none / alibi printer / 0/1/2/3
alibi memory / alibi harddisk)
135 Software option 0/1/2/3
136 Single / dual graduation. 0/1
140 Start-up graduation with dual gradu- 0/1
ation, kg/lb
190 Initial delay before averaging 0.0 - 10.0 [s]
191 Averaging time (cycle time) 0.3 - 10.0 [s]
192 Termination window when 5 – 125 [d]
averaging
194 Gravity correction factor 0.00 – 45.00 [12.56]

912475 2002-01-16 Page 29 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series
Legal Set-up and Calibration Manual
Point Description Value
Lo. Calibration, min. load 0 [d]
Calib.
Hi. Calibration, max. capacity 100 to max. capacity [d]
Calib.
220 Fine calibration at max. capacity -3.0 to 3.0, 0.0
224 Where to linearize 100 to (max. capacity - 100) [d]
225 How much to linearize -3.0 to 3.0, 0.0

912475 2002-01-16 Page 30 of 30 Legal Set-up and Calib. Manual 8500 Indicator Series