Beruflich Dokumente
Kultur Dokumente
zero band
slit on
slit
box
Figure 1. Schematic of the modular spectroscope. Each component is freely adjustable. The setup shows a mercury lamp as the calibration
light source.
a calibration light source since its emission spectrum is known The zero band consists of light from the source that passes
and spans a wide range of visible wavelengths. In Figure 1, through the grating without diffraction. Resolution is in-
the slit after the light source consists of narrowly spaced ra- creased when the grating is moved farther from the projec-
zor blades taped onto a cardboard flat. A second razor blade tion since the bands become farther apart. Resolution is also
slit is on the side of a cardboard box, with the opposite side increased when the slit width in the box is decreased since
open, that has been prepared by the instructor. The box serves the bands do not change position but decrease in size, which
to prevent extraneous light from reaching the dispersed pro- increases the precision of determining the center of each band.
jection on the wall. A holographic grating1 suspended by a These effects dramatically illustrate the advantages of using
clamp and lab stand is placed after the box. long monochromators in conjunction with narrow slit widths
The LED light source and power supply are illustrated for maximum spectral resolution.
in Figure 2. The diodes and circuit board were purchased Once the operation of the spectroscope is understood
from Radio Shack. The diodes have clear rather than colored and resolution is optimized, students calibrate the apparatus
plastic and each emits a different wavelength. A regulated by marking with a pencil the centers of each band on the
power supply that outputs 4 V is sufficient. Slightly lower or projection. Calibration of the system involves measuring the
higher voltages may be necessary if a different resistor or vari- center-to-center distance between every band and the zero
able transformer2 is incorporated into the setup shown in Fig- band position. A plot of distance-from-the-zero-band versus
ure 2. Variable transformers are commonly found in wavelength may then be used to determine wavelengths of
chemistry laboratories since they are used for controlling the other sources shown through the spectroscope. When deter-
power supplied to heating mantles. The transformer is used mining the distance from the zero band for a new light source,
to make fine adjustments to the voltage applied across the the zero band of the new light source and that of the marked
LED. A voltmeter with 0.01 mV resolution is connected the zero band position from the mercury lamp must overlap.
across the LED to monitor the voltage. In order to make this alignment and maintain calibration,
LEDs are easy to short circuit. The manufacturer’s speci- the optimal spectroscope component to adjust is the light
fications stated on the packaging of a diode specify its cur- source. The wavelengths of emission of the diodes are deter-
rent and voltage maxima. To safeguard each LED, the output mined using the plot described above.
of the power supply is kept constant and the voltage applied To determine Planck’s constant, a minimum voltage
across the LED is controlled with the transformer. The in- measurement for each diode is necessary. Minimum-voltage
structor provides either the maximum transformer setting or measurements are determined by slowly increasing from zero
the maximum voltage allowed for each diode. The 1.1 kΩ the voltage applied across each diode and stopping when the
resistor in Figure 2 limits the current. first faint glimmer of light is observed. This is a very fine
measurement and the use of a voltage transformer greatly sim-
Procedure plifies this measurement.
LED variable
transformer
transformer
output
voltmeter
circuit 1.1 kΩ resistor
board
to regulated
power supply transformer
input
Figure 2. Schematic of the LED light source and power supply. Figure 3. The emission spectrum of a mercury lamp after pass-
The power supply for the LED consists of a regulated power sup- ing through the spectroscope. The center band, which underwent
ply connected to the input (i.e., plug) of a variable transformer. no diffraction, is referred to as the zero band.
The output of the transformer is connected across the LED and
monitored with a voltmeter.
W
Supplemental Material
hc 1 M
V = + (5)
qe λ qe Instructions for the students and a sample of a student’s
data are available in this issue of JCE Online.
where V is the minimum voltage necessary to observe emis-
sion of wavelength λ. From eq 5 it can be seen that a plot of Notes
1 1. The grating, part 3054510, can be purchased from
V versus (6)
λ Edmund Scientific (1-800-728-6999).
2. The transformer, part 09-521-130, can be purchased from
will produce a slope of Fisher Scientific (1-800-766-7000).
hc Literature Cited
(7)
qe
1. Wakabayashi, F.; Hamada, K.; Sone, K. J. Chem. Educ. 1998,
from which Planck’s constant may be determined. 75, 1569.
2. Wickun, W. G. J. Chem. Educ. 1998, 75, 1477.
Conclusion 3. Cortel, A.; Fernandez, L. J. Chem. Educ. 1986, 63, 348.
4. Edwards, R. K.; Brandt, W. W.; Companion, A. L. J. Chem.
This exercise involves manipulation of the components Educ. 1962, 39, 147.
of a spectroscope to gain insight into its operation, calibra- 5. Fowles, G. F. In Introduction to Modern Optics; Dover: New
tion, and resolution. A determination of Planck’s constant is York, 1989; p 288.