Sie sind auf Seite 1von 10

LUNA-1K

CARM 1K
CCD CAMERA TEST MANUAL

Ver 1

COMED MEDICAL SYSTEMS Co.,LTD


LUNA-1K Test Manual

INDEX

1. Camera Component-----------------------------------------------------------------------------P. 3
2. Camera Control Board Test-------------------------------------------------------------------P. 4
3. Camera Power Board Test --------------------------------------------------------------------P. 6
4. Test form---------------------------------------------------------------------------------------------P. 7

2 COMED Medical Systems Co., Ltd


LUNA-1K Test Manual

1. Camera Component

Picture 1. Camera Component

① Camera Control Board


② Camera POWER Board
③ Lens

Picture 2. Camera Control Board Picture 3. Camera Power Board

3 COMED Medical Systems Co., Ltd


LUNA-1K Test Manual

2. Camera Control Board Test

Picture 4. Camera Control Board Top

Picture 5. Camera Control Board Bottom

4 COMED Medical Systems Co., Ltd


LUNA-1K Test Manual

 camera control board connector and function

• J1: IRIS connector

• J2: analog power input connector

• J3: digital power input connector

• J4: digital power input connector

• J7: digital power input connector

• J8: digital power input connector

5 COMED Medical Systems Co., Ltd


LUNA-1K Test Manual

 check digital POWER

Check digital power follow sequential order by DVM(Multimeter) after connect J2 (on camera PWR board) and J2 (on

camera control board) (reference picture 4)

• TP1/ TP7(GND): 0 VDC

• TP2(+1.2VDC): +1.185 VDC ~ +1.197 VDC

• TP3(+3.3VDC): +3.292 VDC ~ +3.332 VDC

• TP4(+2.5VDC): +2.495 VDC ~ +2.518 VDC

• TP5(+1.25VDC): +1.241 VDC ~ +1.259 VDC

• TP6(+1.8VDC): +1.791 VDC ~ +1.808 VDC

 check analog POWER

Check analog POWER follow sequential order by DVM(Multimeter) after connect J3 (on Camera PWR Board) and J8

(on Camera Control Board) (reference picture4)

• TP1/ TP7(GND): 0 VDC

• TP15(+15VDC): +15 VDC

• TP9(+12VDC): +12 VDC

• TP10(-8.5VDC): -8.5 VDC

• TP13(-4VDC): -4 VDC

• TP12(-5VDC): -5 VDC

• TP11(over +4.9VDC ): over +4.9 VDC

• TP14(-2VDC): -2 VDC

6 COMED Medical Systems Co., Ltd


LUNA-1K Test Manual

3. Camera Power Board Test

그림 6. Camera Power Board

 camera power board connector and function

• J1: 24V power input connector

• J2: analog power input connector

• J3: digital power input connector

7 COMED Medical Systems Co., Ltd


LUNA-1K Test Manual

 check, input power

Check, J1 input power by DVM(multimeter) (reference picture 5)

• TP1(+24VDC): +23.9 ~ +24.1 VDC

• TP2(IN_GND): 0 VDC

 check analog POWER

Check, J2 output power by DVM(multimeter) (reference picture 5)

• TP3(-15VDC): -15.29 VDC ~ -14.99 VDC/-10.61 VDC ~ -10.43 VDC(connect with camera)

• TP4(+15VDC): +14.93 VDC ~ +15.24 VDC/+14.71 VDC ~ +14.96 VDC(connect with camera)

• TP5(+5VDC): +4.968 VDC ~ +5.014 VDC/+4.967 VDC~ +4.982 VDC(connect with camera)

• TP6(OUT_GND): 0 VDC

 check, digital POWER

Check, J3 output power by DVM(multimeter) (reference picture)

• TP7(+3.3VDC): +3.311 VDC ~ +3.344 VDC

• TP8(+2.5VDC): +2.507 VDC ~ +2.528 VDC

• TP9(+1.2VDC): +1.198 VDC ~ +1.204 VDC

• TP10(+1.25VDC): +1.257 VDC ~ +1.272 VDC

• TP11(+1.8VDC): +1.799 VDC ~ +1.813 VDC

• TP6(OUT_GND): 0 VDC

8 COMED Medical Systems Co., Ltd


LUNA-1K Test Manual

4. LUNA-1K Camera QC Program Manual

 QC Program installation manual

• extract the QC.zip file.

• install the vcredist_x86.exe file in the QC folder, if not installed visual studio 2008 in the PC.

 QC Program

• execute the VideoViewer.exe file in the QC folder.

• connect the camera and PC.

• if you push the Grab Start button, image will be appear on the VideoViewer windows.

• In case of nothing Lens on the Camera, blind the sensor by hand to check image is getting darkness on the

VideoViewer windows.

• If image is strange, sensor or camera problems

• In case of Lens on the Camera, check image on the VideoViewer windows.

• If image is strange, sensor or camera or Lens problems.

9 COMED Medical Systems Co., Ltd


LUNA-1K Test Manual

 VideoViewer

7 8

1 9 10
2

11

 Composition of VideoViewer windows


1. Grab Start: import the image from the Camera, when you push the Grab Start button after connected
Device.
2. Grab Stop: stop image will be appears on the windows, when you push the Grab Stop button.
3. Save Images: save the current Grab image in the appointed folder.
A. No 9: a number of saved images.
B. No 10: progress bar – according to the number of saved images. (Max 100 image)
C. File nave: “MM.DD HH_MM_SS_SSS”(Month. Date. hr. min. sec. msec).
4. RunQC: execute the QC program.
5. fps: execute the Quality_Checker_Base.exe program by appointed address at VV.ini file.
6. : Bit per Pixel, Width, Height.
7. : Dialog window for saved image folder.
8. : saved image folder address.
A. You can use the button 7 and insert the address directly.
9. : a number of saved images.
10. : progress bar – according to the number of saved images. (Max 100 image)
11. : display the Grab image on windows.

 INI file (VV.ini)


 Program setting information.
 [DIR]
 CUR_DIRECTORY: current executed program Directory
 INI_PATH: address of VV.ini file.
 QC_EXE: address of Quality_Checker_Base.exe file.
 IMG_SAVE_DIRECTORY: address of image save folder.
 [IMAGE_FORMAT]
 BITPERPIXEL: Bit Per Pixel of image
 WIDTH: width resolution of image.
 HEIGHT : height resolution of image.

10 COMED Medical Systems Co., Ltd