Beruflich Dokumente
Kultur Dokumente
P/N: 01.54.113701-10
Statement
This manual will help you understand the operation and maintenance of the product better. It
is reminded that the product shall be used strictly complying with this manual. User’s
operation failing to comply with this manual may result in malfunction or accident for which
Edan Instruments, Inc. (hereinafter called EDAN) can not be held liable.
EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any
materials contained in this manual shall not be photocopied, reproduced or translated into
other languages.
Materials protected by the copyright law, including but not limited to confidential information
such as technical information and patent information are contained in this manual, the user
shall not disclose such information to any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or implicitly, any
right or license to use any of the intellectual properties of EDAN.
EDAN holds the rights to modify, update, and ultimately explain this manual.
The electrical installation of the relevant room complies with national standards, and
Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other
information to help qualified technician to maintain and repair some parts, which EDAN may
I
define as user serviceable.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment,
produce inaccurate data, or invalidate a procedure.
II
Table of Contents
Chapter 1 Warranty and Service............................................................................................ 1
Chapter 2 Safety Guidance...................................................................................................... 4
2.1 Introduction ................................................................................................................... 4
2.2 General Information ...................................................................................................... 4
2.3 Safety Precautions ......................................................................................................... 5
Chapter 3 Installation .............................................................................................................. 7
3.1 Environment Requirements........................................................................................... 7
3.2 Electrical Requirements ................................................................................................ 7
3.3 Safety Requirements ..................................................................................................... 7
3.4 Installing the Monitor.................................................................................................... 8
3.5 Connecting to AC Power............................................................................................... 9
Chapter 4 Functional Checks................................................................................................ 10
4.1 Switching on Check .................................................................................................... 10
4.2 LCD Screen Check...................................................................................................... 10
4.3 Keys Check ................................................................................................................. 10
4.4 Touch Screen Check.................................................................................................... 10
4.5 Printing Check............................................................................................................. 11
4.6 Alarms Check.............................................................................................................. 11
Chapter 5 System Configuration .......................................................................................... 12
5.1 Opening User Maintain Menu..................................................................................... 12
5.2 Entering Demo Mode .................................................................................................. 12
5.3 Selecting Lead Style.................................................................................................... 12
5.4 Changing the Bed No. ................................................................................................. 12
5.5 Changing Network Bed No. ........................................................................................ 13
Chapter 6 Maintenance ......................................................................................................... 14
6.1 Maintenance Inspection............................................................................................... 14
6.2 Maintenance of the Monitor........................................................................................ 14
Chapter 7 Principle Introduction ......................................................................................... 16
7.1 Module Principle ......................................................................................................... 16
7.1.1 ECG Principle ................................................................................................... 16
7.1.2 NIBP Principle .................................................................................................. 19
7.1.3 SpO2 Principle ................................................................................................... 20
III
7.1.4 TEMP Principle................................................................................................. 21
7.1.5 IBP Principle ..................................................................................................... 22
7.1.6 C.O. Principle.................................................................................................... 22
7.1.7 CO2 Principle.................................................................................................... 22
7.1.8 AG Principle...................................................................................................... 23
7.2 System Principle Block Diagram ................................................................................ 23
7.2.1 Main Control Board .......................................................................................... 24
7.2.2 ECG Module ..................................................................................................... 25
7.2.3 ECG (E6/E8) Module........................................................................................ 25
7.2.4 NIBP (V6) Module............................................................................................ 26
7.2.5 SpO2 (A8) Module ............................................................................................ 27
7.2.6 Print Controlling Board..................................................................................... 27
7.2.7 Power Module ................................................................................................... 28
7.2.8 Key Board ......................................................................................................... 28
7.2.9 M80 Interface Board ......................................................................................... 29
7.2.10 M80 Interface Board ....................................................................................... 30
7.2.11 LVDS LCD ..................................................................................................... 31
7.3 Interfaces ..................................................................................................................... 32
7.3.1 DB9 Interface .................................................................................................... 32
7.3.2 RJ45 Interface ................................................................................................... 32
7.3.3 USB Interface.................................................................................................... 32
7.3.4 DB15 Interface .................................................................................................. 33
7.3.5 BNC Interface ................................................................................................... 33
7.3.6 SD Card Interface.............................................................................................. 33
7.3.7 Nurse Call Interface .......................................................................................... 33
Chapter 8 Troubleshooting ................................................................................................... 34
8.1 Monitor Booting Failures ............................................................................................ 34
8.2 Display Failures........................................................................................................... 35
8.3 Touch Screen Failures................................................................................................. 35
8.4 Operation Failures ....................................................................................................... 35
8.5 Recorder Failures ........................................................................................................ 35
8.6 Network Failures ......................................................................................................... 36
8.7 Alarm Failures............................................................................................................. 36
8.8 Technique Alarms ....................................................................................................... 36
IV
8.9 Power Board Failures .................................................................................................. 36
8.10 Data Storage .............................................................................................................. 37
8.11 ECG Monitoring Failures.......................................................................................... 37
8.12 SpO2 Monitoring Failures ......................................................................................... 38
8.13 NIBP Monitoring Failures......................................................................................... 38
8.14 TEMP Monitoring Failures ....................................................................................... 38
8.15 CO2 Monitoring Failures........................................................................................... 38
8.16 AG Monitoring Failures ............................................................................................ 39
Chapter 9 Modules’ Malfunction Verification .................................................................... 40
9.1 Verifying Malfunction of the Main Control Board..................................................... 40
9.2 Verifying Malfunction of the ECG Module................................................................ 42
9.3 Verifying Malfunction of the NIBP Module............................................................... 43
9.4 Verifying Malfunction of the SpO2 Module ............................................................... 45
9.5 Verifying Malfunction of the IBP Module (4 Channel IBP) ...................................... 46
9.6 Verifying Malfunction of the CO2 Insulation Power Board ....................................... 47
9.7 Verifying Malfunction of the Power Module.............................................................. 48
Chapter 10 Disassembling the Monitor................................................................................ 50
10.1 Tools Required .......................................................................................................... 50
10.2 Replacing Fuses......................................................................................................... 50
10.3 Disassembling the Main Unit.................................................................................... 51
10.4 Disassembling the Upper Assembly ......................................................................... 52
10.4.1 Replacing the Touch Screen Controlling Board ............................................. 52
10.4.2 Replacing the Key Board ................................................................................ 53
10.4.3 Replacing the Power Inverter Board ............................................................... 53
10.4.4 Replacing the LCD.......................................................................................... 55
10.4.5 Replacing the Touch Screen............................................................................ 57
10.5 Disassembling Module Unit...................................................................................... 57
10.5.1 Replacing the Main Control Board ................................................................. 57
10.5.2 Replacing the NIBP Module ........................................................................... 58
10.5.3 Replacing the SpO2 Module............................................................................ 60
10.5.4 Replacing the ECG Module ............................................................................ 61
10.5.5 Replacing the IBP Module .............................................................................. 62
10.5.6 Replacing the CO2 Insulation Power Board.................................................... 64
10.6 Disassembling Main Unit.......................................................................................... 65
V
10.6.1 Replacing the Recorder ................................................................................... 65
10.6.2 Replacing the Speaker..................................................................................... 65
10.6.3 Replacing the Power Module .......................................................................... 66
10.6.4 Replacing the Fan............................................................................................ 67
10.6.5 Replacing the Power Interface Board.............................................................. 68
Appendix 1 Renewal Parts .................................................................................................... 71
VI
M80 Patient Monitor Service Manual Warranty and Service
Standard Service
EDAN provides a one-year-warranty for the warranted products (accessories are included). The
warranty period begins on the date the products are shipped to customers. If a customer promptly
notifies EDAN of customer’s warranty claim hereunder, EDAN will either repair, adjust or
replace (with new or exchange replacement parts) EDAN’s products. EDAN warrants that any
service it provides to customers will be performed by trained individuals in a workmanlike
manner.
Limitation of Warranty
Direct, indirect or final damage and delay caused by the following situations for which EDAN is
not responsible may void the warranty:
Service Procedure
Fill in the SCF with detailed information including: Model Name, Serial Number (SN) and
Problem Phenomena.
EDAN should not have any obligation to take over the case without this information. The
form can be downloaded at: http://www.edan.com.cn or obtained from EDAN’s Service
Department.
Once the service department receives the fully filled SCF, EDAN’s engineer will offer a
solution in three working days. EDAN will follow out the case based on the two conditions
below:
-1-
M80 Patient Monitor Service Manual Warranty and Service
Within Warranty:
i) After receiving the Return Material Authorization (RMA) form from EDAN service
department, the customer sends EDAN the defective parts and informs about the shipment
tracking number. Then we will dispatch new part(s) to your confirmed address with
confirmed shipping invoice.
ii) The customer signs the Declaration Form and sends it back by email or fax. This form is
legally certificated to make sure the customer or end-user will return the defective parts to
EDAN on time. We will, at this option, dispatch the replacement one(s) with confirmed
shipping invoice.
NOTES:
(1) Both Return Material Authorization Form and Declaration Form are offered by
EDAN service department once the SCF is confirmed by service engineer.
(2) The customer is responsible for freight & insurance charges when the equipment
is shipped to EDAN for service, including custom charges. EDAN is responsible
for the freight, insurance & custom charges from EDAN to the customer.
Out of Warranty:
After receiving the RMA form from the service department, the customer sends defective
parts to EDAN in advance. We will analyze the problems and discuss with the customer about
either repairing or replacing the part(s). Once the maintenance fee is invoiced and paid, we
will make sure to dispatch good part(s) to the confirmed address.
NOTE: The customer is responsible for any freight & insurance charge for the
returned product.
Before the shipment of the materials, the customer must obtain an RMA form from our
service department, in which the RMA number, description of returning parts and shipping
instructions are included. The RMA number should be indicated on the outside of the
shipping container.
-2-
M80 Patient Monitor Service Manual Warranty and Service
NOTE:
EDAN should not have any obligation to the end-user or customer who returns the
goods without the notification by EDAN’s service department. The sender takes full
responsibility for the accounted fee.
Please disassemble the parts with anti-static facility, do not touch the parts with naked hand.
Please describe the returned parts as ‘sample of *****’ and put the total value on the invoice,
and note on the invoice as ‘sample, no commercial value’.
Contact Information
If you have any question about maintenance, technical specifications or malfunctions of devices,
do not hesitate to contact us.
E-mail: support@edan.com.cn
-3-
M80 Patient Monitor Service Manual Safety Guidance
2.1 Introduction
This service manual is a reference for periodic preventive maintenance and corrective service
procedures for the M80 patient monitor.
WARNING
M80 Patient Monitor (hereinafter called monitor) is designed to comply with the international
safety requirements IEC/EN 60601-1 for medical electrical equipment. It is class I
equipment.
The protective categories against electric shock of the patient connections are:
This symbol indicates that the electric shock defending grade of this instrument is Type BF.
This symbol indicates that the electric shock defending grade of this instrument is Type CF.
-4-
M80 Patient Monitor Service Manual Safety Guidance
To avoid the possibility of injury, observe the following precautions during the operation of the
instrument.
WARNING
1 The monitor must be serviced by authorized and qualified personnel only. EDAN do
not accept responsibility for safety compliance, reliability and performance if
modifications or repairs are carried out by unauthorized personnel. Identical
replacement parts must be used.
2 The service personnel should be familiar with the operation of this monitor. Refer to
Patient Monitor User Manual for details.
3 Perform periodic safety testing to ensure proper patient safety. This should include
leakage current measurement and insulation testing. The recommended testing
interval is once per year.
4 Disconnect power cord before changing the fuses. Replace them with those with the
same specifications only.
5 SHOCK HAZARD - Do not attempt to connect or disconnect a power cord with wet
hands. Make certain that your hands are clean and dry before touching a power cord.
6 SHOCK HAZARD - Do not remove the top panel cover during operation or while
power is connected. Only authorized service personnel could remove the unit cover.
8 Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data
processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the system standard IEC/EN
60601-1-1. Anybody who connects additional equipment to the signal input connector
or signal output connector to configure a medical system must ensure that the
system complies with the requirements of the valid version of the system standard
IEC/EN 60601-1-1. If in doubt, consult our technical service department or your local
distributor.
-5-
M80 Patient Monitor Service Manual Safety Guidance
11 Don’t solder the leading wire and the battery terminal directly.
CAUTION
2 The device is designed for continuous operation and is “ordinary” (i.e. not drip or
splash-proof).
3 Do not operate the unit if it is damp or wet because of condensation or spills. Avoid
using the equipment immediately after moving it from a cold environment to a warm,
humid location.
4 While the battery is charged, used or stored, keep it away from objects or materials
with static electric charges.
-6-
M80 Patient Monitor Service Manual Installation
Chapter 3 Installation
WARNING
Working
Temperature: 5 ºC ~ 40 ºC ( 41ºF ~ 104ºF)
Relative Humidity: 25% ~ 80% (non-condensing)
Atmospheric Pressure: 860hPa ~ 1060hPa
Storage
Temperature: -20 ºC ~ 55 ºC
Relative Humidity: 25% ~ 93% (non-condensing)
Atmospheric Pressure: 700hPa ~ 1060hPa
Do not install the monitor in a flammable atmosphere where concentrations of flammable
anesthetics or other materials may occur.
Keep the environment clean. Avoid vibration. Keep it far from corrosive medicine, dust area,
high-temperature and humid environment.
Do not touch signal input or output connector and the patient simultaneously.
Equipment and devices that connect to the monitor should form an equipotential body to
ensure effective grounding.
Do not switch on the monitor until all cables have been properly connected and verified.
-7-
M80 Patient Monitor Service Manual Installation
CAUTION
1 Shut the display completely flat before mounting the monitor to the wall.
2 Make sure the wall mounting board is firmly fixed to the wall. If there is any doubt, do
not hang the monitor to this board.
3 Make sure the monitor is safely hung on the posts of the board before releasing your
hands from the monitor.
-8-
M80 Patient Monitor Service Manual Installation
-9-
M80 Patient Monitor Service Manual Functional Checks
You are not required to open the device case for functional checks.
WARNING
Only qualified service personnel should perform a full functional check procedure.
Check if the power indicator lights up, if a start-up tone is heard, and if the screen lights up in a
few seconds.
If any failure is detected, refer to section 8.1 Monitor Booting Failures for details.
If any failure is detected, refer to section 8.2 Display Failures for details.
If any failure is detected, refer to section 8.4 Operation Failures for details.
If you touch the center of a key but the monitor does not execute the right operation, calibrate the
-10-
M80 Patient Monitor Service Manual Functional Checks
touch screen. Refer to section Calibrating Touch Screen in Patient Monitor User Manual for
details.
If any operation failure is detected, refer to section 8.3 Touch Screen Failures for details.
If any failure is detected, refer to section 8.5 Recorder Failures for details.
NOTE:
Please make sure the paper loading and setting are correct before start printing.
If any failure is detected, refer to section 8.7 Alarm Failures for defective details.
-11-
M80 Patient Monitor Service Manual System Configuration
WARNING
-12-
M80 Patient Monitor Service Manual System Configuration
3 Select a device No. from 1 to 254 and press the knob to confirm it.
CAUTION
Make sure the device numbers of the monitors in the same system do not overlap.
-13-
M80 Patient Monitor Service Manual Maintenance
Chapter 6 Maintenance
6.1 Maintenance Inspection
(1) Visual Inspection
Users should perform the following inspection before using the monitor:
Check the monitor and accessories to see if there is any visible damage that may affect
patient’s safety or monitor performance.
Check all the outer cables, power socket and power cables.
If any damage is detected, do not use the monitor until it is repaired by EDAN service engineer or
professional service personnel of the dealer.
The overall check of the monitor, including the safety check and functional check, should be
performed by qualified personnel every 6 to 12 months, and each time after service.
The equipment should undergo periodic safety test to ensure proper patient isolation from live
parts. This should include leakage current measurement and insulation testing. The recommended
testing interval is once a year or as specified in the institution’s test and inspection protocol.
WARNING
If the hospital or institution using the monitor can’t practice a satisfying maintenance
plan, abnormal device invalidation may occur and human health may be endangered.
-14-
M80 Patient Monitor Service Manual Maintenance
The gathering of dew on the screen may occur with abrupt temperature or humidity changes. A
stable environment is recommended.
-15-
M80 Patient Monitor Service Manual Principle Introduction
The electrochemistry activity of “excitable cells” in cardiac muscle makes the muscle excite and
the heart contract mechanically. The activation of the heart generates close action current that
flows in human body to every part, which causes potential difference on the surface of every part.
The electrocardiogram records the real time potential difference changes on the body surface.
ECG Lead
This electric activity of heart is first discovered and recorded by WALLER in 1889. In 1903, a
German EINTHOVEN improved the recording equipment and brought in some concepts, which
are still widely in use today, including marking the ECG wave with letters PQRSTU and the
concept of leads.
Lead means the waveform of potential difference between two or more parts of the body surface
changing with heart beat circle. The earliest I, II, III lead defined by EINTHOVEN is called
Bipolar Limb Lead. Later GOLDBURGER defined Augmented Unipolar Limb Lead aVH, aVL
and aVF, and WILSON defined Nonpolar Precordial Lead V1, V2, V3, V4, V5 and V6. They are
all standard ECG leads which are still used in clinics at present.
The heart is solid, a lead waveform indicates the electric activity of heart on a projection surface;
the 12 leads indicate the heart electric activity on 12 different projection surfaces. Consequently
physicians can diagnose the pathological changes on different parts of the heart.
-16-
M80 Patient Monitor Service Manual Principle Introduction
At present, the standard electrocardiograph in clinics places the limb electrodes on wrists and
ankles. But for the monitor, the electrodes are placed on the patient’s chest and abdomen. In spite
of the different positions, they are equivalent and have the same definition. Therefore, the
monitoring leads correspond to electrocardiograph leads; they have the same polarity and
waveform.
The monitor monitors 3 or 6 leads, displays the waveforms of one or two leads simultaneously,
and picks up parameters through waveform analyzing. Powerful monitors can monitor 12 leads
and analyze their waveforms, and then picks up ST band and arrhythmia events.
But ECG monitoring can not completely take the place of standard electrocardiograph. For the
monitored ECG waveform at present can not provide slighter structure. This is mainly because of
their different purpose, the monitor observes the patient’s real-time heart rhythm for a long period;
the electrocardiograph gets the result of certain condition in a short time. As a result, these two
types of equipment have different amplifier pass band widths in measuring circuit; the
electrocardiograph’s band is 0.05-80Hz; that of the monitor is 1-25Hz.
The ECG signal is a faint electrical signal and interfered easily by outside environment. Even the
manufacturers have taken this into consideration and taken some measures to enhance the
machines’ anti-jamming ability, some of the interference are still unconquerable.
-17-
M80 Patient Monitor Service Manual Principle Introduction
Power-line interference comes from the alternating current (50Hz-60Hz) that inputs to the
monitor. This is a common interference. Because of its constant frequency it is easy for the
equipment to avoid it with a band-pass filter.
The contraction of the muscles under the electrodes generates muscle electricity, which will
interfere with ECG signals. Its same spectrum and bandwidth as ECG signal makes it impossible
to be avoided with filter.
The patient’s breathing causes low frequency excursion of ECG signal. But they are lower than
the low frequency extreme of the amplifier; the monitor can handle this interference.
The patient’s movement causes the change of ECG signals. The degree of influence depends on
the movement’s amplitude and frequency. If it’s within the ECG amplifying bandwidth, it will be
hard for the monitor to conquer.
Any interference on the path of the body to ECG amplifier will cause strong noise, which will
make the ECG waveform indistinct. The noise might be caused by the bad contact between
electrodes and skin, cable disconnection or bad grounding of the equipment, for example, the
electrodes are not attached to the skin firmly, electric paste of electrodes is dry, the skin is not
well prepared, or the cable is not well connected or broken. To avoid these, the user has to be
very careful when checking and using the equipment. Besides, the equipment has to be grounded
reliably, which is not only good for anti-inference, but also for the safety of both the patient and
the operator.
If a high frequency electric knife is used during an operation, the ECG waveform will be badly
influenced. As the electrical energy on the body is much stronger than the ECG signals in
amplitude; its rich frequency composition makes the ECG amplifier reach saturated condition and
not be able to monitor the normal ECG waveform. For the time being no monitor can avoid this
type of interference. As a result, the anti-interference requirement for electric knife of the monitor
-18-
M80 Patient Monitor Service Manual Principle Introduction
is: the monitor resumes order after the high frequency electric knife is removed for 5 seconds.
Blood pressure is the force of blood pushing against the walls of the vases; this force changes
with every systole and diastole. The pressure to the arteries and veins are different, so is the
pressure of different parts of the body. Normally we concern the systolic pressure and diastolic
pressure of the artery in the upper arm at the height of the heart as the blood pressure, they are
called systolic pressure and diastolic pressure.
-19-
M80 Patient Monitor Service Manual Principle Introduction
The precondition of oscillometric method is to find out regular arterial pressure pulsation. Or it
might result in unreliablility of the measurement result, increasing of measuring time or even
measuring failure. For example, if the patient’s movement or outer distraction interferes with the
pressure change of air in the cuff during measurement, the monitor will not be able to detect the
regular arterial wave, which might cause measuring failure. This the same thing as that happens
in measuring blood pressure with Korotkoff Sound Method, if the patient’s arm moves, the doctor
can not hear pulse sound clearly or distinguish blood pressure.
Although oscillometric method is an objective and effective method of measuring NIBP, it has its
weaknesses and limitations in clinical use.
2. Severe arrhythmia. The result might be interfered if the patient’s heart beat is irregular
because of serve arrhythmia.
3. Severe shock. Because of the shock patient’s reduced blood flowing to the periphery, the
arterial pulse reduces or vanishes. The measuring result is even more unreliable.
4. Too low or too high blood pressure or heart rate. They will cause the inaccuracy of the
measuring result.
As a result, the oscillometric method is not omnipotent. In order to assure the measuring result’s
validity, user should estimate it according to the real condition.
Besides, make sure to have an ideal condition when measuring NIBP with oscillometric method;
mind the cuff’s size, position and degree of tightness.
Noninvasive arterial pulse oxygen saturation (SpO2) measuring is a technique developed in early
1980s. Based on the different absorption character of haemoglobin (Hb) and oxyhemoglobin
(HbO2) to light, it applies red glow (660nm) and infrared ray (940nm) to penetrate the tissue
respectively, and then is received by photosensor and transferred into electric signals. Meanwhile,
the absorption character of other elements in the tissue, such as skin, skeleton, muscle and vein
-20-
M80 Patient Monitor Service Manual Principle Introduction
blood is constant, only the absorption character of Hb and HbO2 in arteries changes periodically
with pulse. This feature also makes it possible to acquire the SpO2 via processing the received
signals.
This method only measures the SpO2 in arterial blood, and the necessary condition is there is
pulsed arterial blood flow. As a result, in clinic only the parts with arterial blood flow and thin
tissue are used to place the sensor, such as fingers, toes and earlaps.
1 Movement. The measurement bases on the regular arterial pulse absorbing light signals. The
movement of the measured part will interfere with the extraction of the regular pulse signals,
which makes it impossible to process a measurement.
2 Weak perfusion. The patient’s badly blocked twig circle will result in the reduced arterial
blood flow in the measured part, which makes it impossible to process a measurement, such
as severe blood loss, and low temperature at the measured part.
3 Optics interference. If there is strong light irradiating on the sensor, the photosensor’s work
might depart from the normal range, which will cause the inaccuracy of the measurement. So
avoid strong light when measuring.
The temperature sensor applies thermal resistor with negative temperature coefficient, whose
resistance changes with temperature.
Attentions
1 The temperature sensor can be placed at any place of the body. But due to the temperature
difference between different parts of a human being, the measurement result indicates only
the temperature of the human part where the sensor is placed, which might be different from
-21-
M80 Patient Monitor Service Manual Principle Introduction
2 When measuring the temperature, it will take a few minutes for the sensor to bring into
thermal balance with the human body, so the temperature at the beginning is not the real
temperature of the measured part.
3 When measuring the temperature, make sure the sensor is firmly attached to the human body.
If the sensor is loose or there is space between the sensor and skin because of patient’s
movement, the result might be lower than the real temperature.
IBP is measured by means of a catheter inserted directly into the circulatory system. A pressure
transducer connected to the catheter converts the mechanical force exerted by the blood into an
electrical signal, which is displayed graphically as pressure versus time on a monitor screen or
numerically on digital display. The monitor measures direct blood pressure of one selected blood
vessel through two channels or four channels, and displays waveforms and pressure of measured
direct blood pressure (SYS, DIA and MAP).
-22-
M80 Patient Monitor Service Manual Principle Introduction
of patient sample. The relation between partial pressure and percentage of CO2 concentration is
given below:
There are Mainstream and Sidestream modules according to different connecting way of infrared
sensor. Sidestream module consists of circuit board, built-in sidestream infrared sensor, air pump
and control unit. When using Sidestream, the user should also use exterior water trap, dry tube
and sampling tube; Mainstream consists of circuit board and external mainstream infrared sensor.
IR sensor requires preheating; in Sidestream mode, the pumping rate can be set as 100, 150 or
200 ml/min according to patient situation, and the user can set up compensation in anesthetic
monitoring, such as water vapor, oxygen, temperature, Des, etc. when not performing CO2, it is
recommended to close Sidestream air pump, Mainstream module sensor and IR source, for
extending the lifespan of module and reducing power consumption. In Mainstream mode, IR
preheating time is relative long, and the module has no air way like that of Sidestream.
7.1.8 AG Principle
Concentration of GAS (Anesthetic gas) is measured for its characteristic of absorbing infrared ray.
All the anesthetic gases measured in “AG module” have this characteristic, and each gas has its
own absorption characteristic. First the gas to be measured is driven into a sample cell. Then
optic infrared filter selects the infrared ray with special wavelength to penetrate this gas. For a
given volume, the higher gas concentration is, the more infrared rays are absorbed. It means the
higher the concentration of the absorbed infrared is, the fewer infrared rays penetrated the gas.
First measure the quantity of infrared rays that have penetrated the gas and then calculate the gas
concentration via specialized formula. For multiple gases measuring, it need install various
infrared filters in “AG module”.
-23-
M80 Patient Monitor Service Manual Principle Introduction
ARM926EJ-S
-24-
M80 Patient Monitor Service Manual Principle Introduction
GPIO 10
(1) ECG
3 Leads: 1 channel
Waveform
5 Leads: 2 channels (standard), 7 channels (max)
-25-
M80 Patient Monitor Service Manual Principle Introduction
Accuracy + 1%
Resolution 1 bpm
(2) TEMP
Channel Qty. 1
Resolution 0.1 ºC
25 ~ 45 ºC: ±0.1 ºC
Accuracy
0 ~ 25 ºC, 45 ~ 50 ºC: ±0.2 ºC
Auto Circle Mode Interval 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 240, 480 minute(s)
-26-
M80 Patient Monitor Service Manual Principle Introduction
The pulse extent of optical signal changes during monitoring. SpO2 parameter, pulse rate signal
and pleth wave will be acquired after calculation. These data will be transmitted to the PC by
special communication protocol. The specifications are listed as follows:
SpO2
Resolution 1%
Pulse
Resolution 1 bpm
Monitoring
circuit
CPU system
-27-
M80 Patient Monitor Service Manual Principle Introduction
The M80 printer module includes: microprocessor MC9S12A64 circuit, power switch and control,
communication interface, motor control part, recorder paper detecting circuit, time control for
printing power and heating, paper lacking temperature control and thermo sensitive head
protecting circuit.
Interface definition
PS900K power module mainly consists of PS900K power board, PS900K power control board,
and PS900K DC output board.
-28-
M80 Patient Monitor Service Manual Principle Introduction
Alarm
light
Main board
Interface definition
(1) Interface connecting to PS900K power control board
-29-
M80 Patient Monitor Service Manual Principle Introduction
Interface definition:
Interface connecting to main control board
-30-
M80 Patient Monitor Service Manual Principle Introduction
Temperature Range:
-31-
M80 Patient Monitor Service Manual Principle Introduction
7.3 Interfaces
DB 9 interface (RS232)
RJ45 interface (Network)
USB interface
DB15 interface (VGA visual signal interface)
BNC interface (synchronous defibrillation/analog output interface)
SD card interface (data storage interface)
Nurse call interface
It is used to connect the monitor to a computer for updating or monitoring information collection,
or to a monitoring system such as MFM-CMS.
It is used to connect external devices that support USB protocol to the monitor.
-32-
M80 Patient Monitor Service Manual Principle Introduction
It is used to output the VGA visual signals of monitor to the external display device that supports
VGA signals.
-33-
M80 Patient Monitor Service Manual Troubleshooting
Chapter 8 Troubleshooting
EDAN supports replacement of PCBs and major subassemblies for this product. Verify the
malfunction of the PCBAs/Modules prior to disassembling with the method described in chapter
9 Modules’ Malfunction Verification. When replacement is needed, follow the procedures
described in chapter 10 Dissembling the Monitor.
-34-
M80 Patient Monitor Service Manual Troubleshooting
After switching on, LCD has LCD cable defective Repair or replace the LCD cable.
no display or is black when
working. Main control board
Replace the main control board.
defective.
No paper advances. No paper in the drawer. Load paper and close the drawer.
Paper bail is not pressed
Press down the paper bail.
down.
-35-
M80 Patient Monitor Service Manual Troubleshooting
-36-
M80 Patient Monitor Service Manual Troubleshooting
-37-
M80 Patient Monitor Service Manual Troubleshooting
No NIBP numeric value. Loose cuff or patient activity. Tie up the cuff or quiet the patient.
TEMP numeric value Bad connection of TEMP Connect and fix the TEMP transducer
incorrect. transducer. well.
-38-
M80 Patient Monitor Service Manual Troubleshooting
-39-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
NOTE: For all the interfaces, the first pin is the one with a square solder pad.
1) Open the main unit using the procedures described in section 10.3.
3) Measure the voltage to earth of the pins listed below, using a multimeter:
Item Pin Reference result
○
1 TP1 VDD_5V (5.0V) ±5%
○
2 TP3 +3V3(3.3V)±5%
○
3 TP2 EXT_12V (12.0V) ±5%
○
4 TP5 VDDMCU_1V2 (1.2V) ±5%
○
5 TP6 VDDFPGA_2V5 (2.5V) ±5%
○
6 TP7 VDDFPGA_1V2 (1.2V) ±5%
-40-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
J7
5 TP6
○
1 TP1
○
6 TP7
○
3 TP3
○
4 TP5
○
2 TP2
○
J6
J3 J4 J2 J5
-41-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
4) Compare the measurement results with the reference results in the list. If any one of the
results exceeds the reference range, the main control board defection is confirmed.
Replacement of the main control board is recommended.
1) Make sure the malfunction of ECG measurement is not caused by other defects, such as bad
connection or defective electrodes, main control board, etc.
2) Open the main unit using the procedures described in section 10.3.
3) Disconnect the cable between main control board and ECG module.
5) Measure the voltage to earth of the following pins of the main control board, using a
multimeter:
○
1 1st pin of J5 VDD_5V±5%
○
2 3rd pin of J5 UART_TXD1 (serial port signal with 115200bps baud rate)
○
3 4th pin of J 5 UART_TXD1 (serial port signal with 115200bps baud rate)
6) Measure the 3rd pin of J5 by using an oscillograph, a 0V ~ 5V pulse string signal should be
detected.
7) If all the results are in the reference range, connect the ECG module to the main control
-42-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
board. Measure the voltage to earth of the following pins on the ECG board.
J4
8) If any one of the results exceeds the reference range, the ECG module defection is confirmed.
Replacement of the module is recommended.
1) Make sure the malfunction of NIBP measurement is not caused by other defects, such as bad
connection or defective cuff, extension tube, main control board, etc.
2) Open the main unit using the procedures described in section 10.3.
3) Disconnect the cable between main control board and NIBP module.
5) Measure the voltage to earth of the following pins on the main control board, using a
-43-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
multimeter:
○
1 2nd pin of J7 (+5±5%) V
○
2 3rd pin of J7 (+12±5%) V
○
3 7th pin of J7 UART_RXD0 (Serial port signal with 9600bps baud rate)
○
4 8th pin of J7 UART_RXD0 (Serial port signal with 9600bps baud rate)
6) Use an oscillograph to measure the 8th pin of J7. A 0V ~ 5V pulse string signal should be
detected.
7) If all the results are in the reference range, connect the NIBP module to the main control
board. Measure the voltage to earth of the following pins on the NIBP board.
J4
8) If any one of the results exceeds the reference range, the NIBP module defection is
confirmed. Replacement of the module is recommended.
-44-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
1) Make sure the malfunction of SpO2 measurement is not caused by other defects, such as bad
connection or defective transducer, main control board, etc.
2) Open the main unit using the procedures described in section 10.3.
3) Disconnect the cable between main control board and SpO2 module.
5) Measure the voltage to earth of the following pins of the main control board, using a
multimeter:
○
1 1st pin of J3 4.75V ~ 5.25V
UART_TXD3 (Serial
○
2 3rd pin of J3 port signal with
4800bps baud rate)
UART_TXD3 (Serial
○
3 4th pin of J3 port signal with
4800bps baud rate)
6) Use an oscillograph to measure the 3rd pin of J3. A 0V ~ 5V pulse string signal should be
detected.
7) If all the results are in the reference range, connect the SpO2 module to the main control
board. Measure the voltage to earth of the following pins on the SpO2 board.
-45-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
J4
8) If any one of the results exceeds the reference range, the SpO2 module defection is
confirmed. Replacement of the module is recommended.
1) Make sure the malfunction of IBP measurement is not caused by other defects, such as bad
connection or defective electrode, main control board, etc.
2) Open the main unit using the procedures described in section 10.3.
3) Disconnect the cable between main control board and IBP module.
5) Measure the voltage to earth of the following pins on the main control board, using a
multimeter: (Refer to section 9.3 for J4’s location.)
○
1 1nd pin of J4 (+5±5%) V
○
2 3rd pin of J4 UART_TXD2 (Serial port signal with 38400bps baud rate)
○
3 4th pin of J4 UART_TXD2 (Serial port signal with 38400bps baud rate)
○
4 5th pin of J4 (+5±5%) V
○
5 7th pin of J4 UART_TXD4 (Serial port signal with 38400bps baud rate)
○
6 8th pin of J4 UART_TXD4 (Serial port signal with 38400bps baud rate)
6) Use an oscillograph to measure the 3rd or 7th pin of J4. A 0V ~ 5V pulse string signal should
-46-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
be detected.
7) If all the results are in the reference range, connect the IBP module to the main control board.
Measure the voltage to earth of the following pins on the IBP board.
○
1 1st pin of J4 4.75V ~ 5.25V
○
2 3rd pin of J4 Serial port signal with 38400bps baud rate
○
3 4th pin of J4 Serial port signal with 38400bps baud rate
J4
8) If any one of the results exceeds the reference range, the IBP module defection is confirmed.
Replacement of the module is recommended.
1) Make sure the malfunction of CO2/GAS measurement is not caused by other defects, such as
bad connection or defective electrode, main control board, etc.
2) Open the main unit using the procedures described in section 10.3.
3) Disconnect the cable between main control board and CO2 insulation power board.
5) Measure the voltage to earth of the following pins on the main control board, using a
multimeter:
○
1 1st pin of J2 (+12±5%) V
○
2 2nd pin of J2 (+5±5%) V
-47-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
○
3 5th pin of J2 UART_TXD3 (Serial port signal with 9600bps baud rate)
○
4 6th pin of J2 UART_TXD3 (Serial port signal with 9600bps baud rate)
6) Use an oscillograph to measure the 3rd pin of J3 (refer to section 9.1 for J3’s location). A 0V ~
5V pulse string signal should be detected.
7) If all the results are in the reference range, connect the CO2 insulation power board to the
main control board. Measure the voltage to earth of the following pins on the CO2 insulation
power board. (refer to section 9.1 for J1’s location)
○
1 1st pin of J1 (+12±5%) V
○
2 5th pin of J1 Serial port signal with 9600bps baud rate
○
3 6th pin of J1 Serial port signal with 9600bps baud rate
J4
8) If any one of the results exceeds the reference range, the CO2 insulation power board
defection is confirmed. Replacement of CO2 insulation power board is recommended.
1) Open the main unit using the procedures described in section 10.3.
3) Measure the voltage to earth of the following pins on the main control board, using a
multimeter: (Refer to section 9.1 for J6’s location.)
-48-
M80 Patient Monitor Service Manual Modules’ Malfunction Verification
○
1 1st pin of J6 (+5±5%)V
○
2 4th pin of J6 (+12±5%)V
4) If any one of the results exceeds the reference range, the power module defection is
confirmed. Replacement of the module is recommended.
-49-
M80 Patient Monitor Service Manual Disassembling the Monitor
WARNING
2 Switch off the monitor and remove the AC power cable before disassembling the
monitor.
3 After any repair of the monitor, perform safety tests before using the monitor again.
1 – A cross-head screwdriver
2 – A flat-head screwdriver
3 – A M3 nut driver
4 – A pair of pliers
2) Carefully place the monitor upside down on a flat surface covered with cloth or other
protecting pad.
3) Pull out the fuse box from the groove, and you will see the fuse.
-50-
M80 Patient Monitor Service Manual Disassembling the Monitor
4) Remove the old fuse. Take out a new fuse supplied by EDAN or of the same specification:
Size: Ф5mm*20mm; Model: T1.6AL 250V.
5) Place the new fuse in the fuse socket, and push the box into position.
2) With a cross-head screw driver, remove the 4 screws securing the upper assembly to the
lower assembly.
3) Turn over the monitor again. Remove the 4 screws securing the lower assembly.
4) Place the monitor upright now. Gently separate the two assemblies, and disconnect the LCD
connecting wire and keyboard connecting wire. See the parts marked in red circles.
5) Carefully part the two assemblies and place the upper assembly on the side of the lower
assembly.
-51-
M80 Patient Monitor Service Manual Disassembling the Monitor
Replace the board in a reversal procedure. Refer to the following illustration when connecting
wires. Ensure they are all correctly connected.
-52-
M80 Patient Monitor Service Manual Disassembling the Monitor
To key board
To touch screen
Replace the board in a reversal procedure. Refer to the following illustration when connecting
wires. Ensure they are all correctly connected.
2
1
No. Connect to
4 1. Main Control Board
2. Inverter
3. Control Knob
4. Touch screen controlling board
5. Alarm indicator
10.4.3 Replacing the Power Inverter Board
To remove the power inverter board,
-53-
M80 Patient Monitor Service Manual Disassembling the Monitor
Replace the board in a reversal procedure. Refer to the following illustration when connecting
wires. Ensure they are all correctly connected.
-54-
M80 Patient Monitor Service Manual Disassembling the Monitor
2
No. Connect to
1. LCD Backlight Wire
2. Key Board
3. LCD Backlight Wire
2) Disconnect all the wires from the power inverter board, alarm indicator board and touch
screen controlling board (shown in circles).
-55-
M80 Patient Monitor Service Manual Disassembling the Monitor
4) Place the upper assembly upright and remove the display assembly stander, as shown below:
5) Remove the four screws around the display to remove the LCD.
screws
Replace the LCD in a reversal procedure. Refer to the following illustration when connecting
wires. Ensure they are all correctly connected.
-56-
M80 Patient Monitor Service Manual Disassembling the Monitor
2) Follow the steps of taking out LCD and remove the touch screen after removing the display
assembly stander, as shown below.
-57-
M80 Patient Monitor Service Manual Disassembling the Monitor
Replace the main control board in a reversal procedure. Refer to the following illustration when
connecting wires. Ensure they are all correctly connected.
10
No. Connect to
1. SpO2 Module
2. IBP module
9
3. CO2 module
4. ECG module
5. Key board
1 6. Printer signal
7. Interface board signal
8
(small)
8. NIBP module
2
9. Interface board signal
(big)
7
10. LCD
3
5
10.5.2 Replacing the NIBP Module
To remove the NIBP module,
2) Remove the three screws marked in circles and disconnect all the wires.
-58-
M80 Patient Monitor Service Manual Disassembling the Monitor
3) Take out the module unit. Remove the four screws securing the NIBP module, and remove
the plastic tube connecting to NIBP socket.
To main
board
Replace the module in a reversal procedure. Ensure the wires are correctly connected.
-59-
M80 Patient Monitor Service Manual Disassembling the Monitor
2) Disconnect all the wires on the SpO2 module, as shown in the following red circles.
3) Remove the four screws securing the module, as shown in the following figure.
To probe board
To main board
Replace the SpO2 module in a reversal procedure. Refer to the following illustration when
connecting wires. Ensure they are all correctly connected.
No. Connect to
i SpO2 Socket
(i) ii Bottom Interface
Board
(ii)
-60-
M80 Patient Monitor Service Manual Disassembling the Monitor
2) Remove the module unit. You’ll see the ECG module as shown in the following figure.
3) Disconnect all the wires on the ECG module. Remove the four screws securing the module.
To probe board
To probe board
To main board
To replace the ECG module,
-61-
M80 Patient Monitor Service Manual Disassembling the Monitor
Replace the ECG module in a reversal procedure. Refer to the following illustration when
connecting wires. Ensure they are all correctly connected.
(i)
No. Connect to
( ii )
( iii )
10.5.5 Replacing the IBP Module
Replacing 2 Channel IBP Module
2) Remove the module unit. You’ll see the 2 channel IBP module as shown in the following
figure.
Replace the 2 channel IBP module in a reversal procedure. Refer to the following illustration
when connecting wires. Ensure they are all correctly connected.
-62-
M80 Patient Monitor Service Manual Disassembling the Monitor
2) Remove the module unit. You’ll see the 4 channel IBP module as shown in the following
figure.
3) Remove the three screws marked in red circles and take off the above board card, see the
following figure.
-63-
M80 Patient Monitor Service Manual Disassembling the Monitor
Replace the 4 channel IBP module in a reversal procedure. Refer to the above illustration when
connecting wires. Ensure they are all correctly connected.
2) Remove the module unit. You’ll see the CO2 module as shown in the following figure.
Replace the CO2 insulation power board in a reversal procedure. Refer to the following
illustration when connecting wires. Ensure they are all correctly connected.
-64-
M80 Patient Monitor Service Manual Disassembling the Monitor
No. Connected to
2) Open the paper casing, and remove the two screws. See the following figure.
3) Gently pull out the recorder, and disconnect the wires connecting the recorder and the
interface board.
No. Connect to
1 Power wire
2 Signal wire
3 Setscrews
1 2
3
2) If a recorder is installed, remove the recorder first. See the following figure.
3) Remove the eight screws securing the main unit on the front, and remove the four screws on
-65-
M80 Patient Monitor Service Manual Disassembling the Monitor
the bottom.
4) Take out the main unit and you’ll see the speaker on the back of main unit. Remove the two
screws to remove the speaker.
2) Take out the main unit in the procedure of replacing the speaker.
3) Disconnect all the wires on the module. Remove the four screws securing the module.
Replace the power module in a reversal procedure. Refer to the following illustration when
connecting wires. Ensure they are all correctly connected.
-66-
M80 Patient Monitor Service Manual Disassembling the Monitor
1 2
No. Connect to
1. Battery interface
4
board
2. Main board
3. AC socket
4. Interface board
2 Take out the main unit in the procedure of replacing the speaker.
-67-
M80 Patient Monitor Service Manual Disassembling the Monitor
Replace the fan in a reversal procedure. Ensure the wires’ connector is correctly connected to
connector on the power cable bundle.
2 Take out the main unit in the procedure of replacing the speaker.
3 Remove all the wires on M80 interface board, and remove the four screws securing the power
interface board.
-68-
M80 Patient Monitor Service Manual Disassembling the Monitor
4 Remove all the wires on M80 interface board, and remove the three screws securing the
interface board.
5 Remove all the wires on M80 interface board, and remove the four screws securing the
battery casing to take out the battery casing.
-69-
M80 Patient Monitor Service Manual Disassembling the Monitor
Replace the power interface board in a reversal procedure. Ensure the wires are correctly
connected.
-70-
M80 Patient Monitor Service Manual Renewal Parts
WARNING
Speaker 01.14.038010
Fan 01.58.47066
LCD 01.16.078211
Inverter 02.08.113741
V6 Module 02.01.100996
E6 Module 12.01.108958
E8 Module 02.01.109095
Printer 12.01.19084
-71-