Beruflich Dokumente
Kultur Dokumente
System Manual
Software Version 2.1
cobas® 4800 System
Systeni Manual
Document information
Edition notice This manual is for users of the cobas 4800 System.
Every effort has been made to ensure that all the information contained in this
manual is correct at the time of printing. However, Roche reserves the right to make
any changes necessary without notice as part of ongoing product development.
Any customer modification to the cobas x 480 instrument or cobas z 480 analyzer
will render the warranty or service agreement null and void.
Software updates are done by Roche Service representatives.
The screenshots in this publication have been added exclusively for the purpose of
ilustration. Configurable and variable data such as parameters, resuits, path names
etc. visible therein must not be used for laboratory purposes.
Intended use The system is intended to be used as a diagnostic or screening system providing
samplepreparation, amplification and detection of specific targets from human
samples.
It is important that the operator reads this manual and the test-specific Operator’s
Manual thoroughly before using the system.
The system is to be used by laboratory professionals trained in laboratory techniques
and by instruction on the use of the system.
Feedback Every effort has been made to ensure that this manual fulfills the intended purpose as
mentioned above. All feedback on any aspect of this manual is welcome and will be
considered during updates. Please contact your Roche representative, should you
have any such feedback.
Instrument approvals The instrument and the analyzer meet the protection requirements laid down in IVD
Directive 98/79/EG.
Furthermore, the analyzer is manufactured and tested according to the following
international standards:
• lEG 61010-1
• lEG 61010-2-101
• UL61O1O-1
• CAN/CSA C22.2 No. 61010-1
This manual meets the European Standard EN ISO 18113-3.
Compliance is demonstrated by the following marks:
Roche Diagnostics
2 System Manual Version 1.0
.
cobas® 4800 System
Systeni Manual
C®(JS
Contact addresses
Roche Diagnostics
System Manual Version 1.0
.
3
cobas® 4800 System
System Manual
Roche Diagnostics
4 System Manual Version 1.0
cobas® 4800 System
Sys(em Iftinunl
Table of contents
3 Hardware
Instrument 39
Analyzer 65
Control unit 69
4 Software
Basic software elements 73
Database 81
Configuration
5 Configuration
Configuration 87
User management 92
Exporting support data 97
Result counters 98
Maintenance
6 Maintenance
Safety information 103
Instrument maintenance 104
Analvzer maintenance 115
Roche Diagnostics
System Manual Version 1.0
. 5
cobas® 4800 System
System I’4anual
Roche Diagnostics
6 System Manual Version 1.0
cobas® 4800 System
Systeni Manual
Preface
The system integrates fully automated total nucleic acid isolation directly from
primary and secondary tubes, automated PCR setup, and real time PCR.
This manual covers the complete system comprised of the cobas’ 4800 software, the
instrument used for sample preparation, and the analyzer used for amplification and
detection using real time PCR.
• Keep this manual in a safe place to ensure that t is not damaged and remains available
for use.
• This manual should be easily accessible at all times.
To help you find information quickly, there is a table of contents at the beginning of
the manual and each chapter. In addition, a complete index can be found at the end
of the manual.
In addition to this manual, the following documents are also provided to assist in
finding desired information quickly:
Test-specfic Operator Manuals For each test, there is a specific Operator’s Manual which describes how to prepare
and perform a run, and how to handle results.
Test-spec ficpackage insert For each test, there is a package insert which provides additional information. For
example, instructions on storage and handling of reagents, samples and controls.
Visual cues are used to help locate and interpret information in this manual quickly.
This section explains formatting conventions used in this manual.
( Start of procedure
End of procedure
List item
(j) Cross-reference
C’ Cail-up (software reference)
Table 2 Symbols used
Roche Diagnostics
System Manual Version 1.0 7
cobas® 4800 System
Systern Manual
‘.‘ Electrical and electronic equipment marked with this symbol are
/Y\ covered by the European Directive 2002/961EC on waste electrical
and electronic equipment (WEEE).
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Table 2 Symbols used
Abbreviation Definition
AD Amplification and detection
ANSI American National Standards Institute
cc cubic centimeter
cLLD Capacitive Liquid Level Detection
CSA Canadian Standards Association
dBA decibel weighted against the A-frequency response curve. This curve
approximates the audible range of the human ear.
DIL Diluent
DWP Deepweil Plate
EC European community
e.g. exempli gratia — for example
EMC ElectroMagnetic Compatibility
EN European standard
ie. id est — that is to say
IEC International Electrical Commission
iSWAP internal Swivel Arm Plate handler
IVD In Vitro Diagnostic
KVA kilovolt-Ampere
LED Light Emitting Diode
LIS Laboratory information system
MWP Microweli plate
n/a not applicable
PCR Polymerase Chain Reaction
QC Quality Control
SD Standard deviation
UL Underwriters Laboratories mc.
UPS Uninterruptible Power Supply
WEEE Waste Electrical and Electronic Equipment
XML Extensible Markup Language
Table 3 Abbreviations
Roche Diagnostics
8 System Manual Version 1.0
.
cobas® 4800 System
Systern Manual
Roche Diagnostics
System Manual Version 1.0
. 9
cobas® 4800 System
Systeni Manual
Roche Diagnosti
10 System Manual Version 1.0
System description
In this chapter, you will find information on the safe operation of the system.
Roche Diagnostics
System Manual Verson 1.0
.
13
1 General safety information cobas® 4800 System
Table of contents System Manual
Disposal 27
Disposal of the instrument 27
Disposal of the analyzer 27
Disposal of control unit components 27
Roche fliagnostics
14 System Manual Version 1.0
.
cobas® 4800 System 1 General safety information
Systern Manual Safety c1assfications
Safety classifications
Safety information
The safety alert symbol by itself without a signal word is used to promote awareness to
hazards which are generic or to direct the reader to safety information provided
elsewhere in the document.
These symbols and signal words are used for specific hazards:
WARNING
1 Indicates a hazardous situation which, if not avoided, could result in death or serious
Injury.
WARNING
CAUTION
Indicates a hazardous situation which, if not avoided, could result in minor or moderate
CAUTION injury
NOTICE NOTICE
lndicates a hazardous situation which, if not avoided, could result in damage to
equipment.
For more information about product safety labels, see Safety labels on the systern (p. 25)
Important information which is not safety relevant is indicated by the following
symbol:
Tip
Indicates additional information on correct use of the system or useful tips.
Roche Diagnostics
System Manual Version 1.0
. 15
1 General safety information cobas® 4800 System
Safetyprecautions System Manual
Safety precautions
Safety information
Particular attention must be paid to the following safety precautions. 1f these safety
precautions are ignored, the operator may suffer serious or fatal injury. Each
precaution is important.
Operator qualification
Personal protective equipnient • Be sure to wear appropriate protective equipment, including, but not limited to,
safety glasses with side shields, fluid resistant lab coat, and approved disposable
gloves.
• Wear a face shield ifthere is a chance of splash or splatter.
Accuracy/precision of measured • Do not use reagents that have exceeded their expiry date, otherwise inaccurate
results data may be obtained.
• For diagnostic purposes always assess the resuits in conjunction with the patient’s
medical history, clinical examination, and other findings.
• Each laboratory must verify that reagent performance meets the published
specifications.
Correct use • Use the instrument only for preparing liquid samples with the provided reagents.
Operating conditions • Operation outside of the specified ranges, e.g. with excessive environmental
temperature and/or humidity, may lead to incorrect resuits or malfunction of the
instrument or analyzer.
• Use the instrument and analyzer as indicated in the technical specifications.
• Make sure that the instrument’s and analyzer’s ventilation openings remain
unobstructed at all times.
Roche Diagnostics
16 System Manual Version 1.0
.
cobas® 4800 System 1 General safety information
Systern Manual Safetyprecautions
Approvedparts Use of non-approved parts or devices may result in malfunction of the instrument or
analyzer and may render the warranty nuli and void. Only use parts and devices
approved by Roche Diagnostics.
Third-party software Installation of any third-party software that is not approved by Roche Diagnostics
may result in incorrect behavior of the system or the software. Do not instali any
non-approved software.
Power interruption A power failure or momentary drop in voltage may lead to data loss. Operation with
an uninterruptible power supply (UPS) is strongly recommended. Check the UPS
regularly to make sure it functions properly.
1f there is a power failure, the analyzer is disconnected from the control unit. The
displayed maintenance status of the analyzer may not be correct. To get the correct
maintenance status, refresh the analyzer under Overview> System > cobas z480 tab.
Electrornagneticfields Devices that emit electromagnetic waves may cause the instrument or analyzer to
malfunction. Do not operate the following devices in the same room where the
instrument or analyzer are installed:
• Mobile phone
• Transceiver
• Cordless phone
• Other electrical devices that generate electromagnetic fields
Data backup There is an automatic backup process that stores the data on the D drive. It is the
customers responsibility to perform regular backups of all measurement results.
Relocation and transportation Do not attempt to relocate or transport the instrument or analyzer. Leave relocation
and transportation to personnel trained or authorized by Roche.
For additional details, see Disposal (p. 27)
Roche Diagnostics
System Manual Version 1.0
. 17
1 General safety information cobas® 4800 System
Safety surnmay Systeni Aianual
Safety summary
This safety summary contains the most important and general warning and caution
messages. Additionally, you will find specific safety information at the beginning of
Part Ivfaintenance (p. 101) and in the test-specific Operator’s Manuals.
For more information about the safe use of the system, refer to the test-specific package
insert.
Warning messages
Electrical safety
Optical safety
Loss of sight due to stanng into laser beam
The barcode reader on the autoloader of the instrument contains a class II laser diode.
WARNING Do not stare into the laser transmitter beam as eyesight may be severely damaged.
Roche Diagnostics
18 System Manual Version 1.0
cobas® 4800 System 1 General safety information
System j’vlanual Safrty summary
Biohazardous materials
Waste
Roche Diagnostics
System Manual Version 1.0 19
1 General safety information cobas® 4800 System
Safety surnniary Systeni Manual
Caution messages
Mechanical safety
Roche Diagnostics
20 System Manual Version 1.0
cobas® 4800 System 1 General safety information
System Aianual Safety suniniary
Roche Diagnostics
System Manual Version 1.0 21
1 General safety information cobas® 4800 System
Safety sunimary Systeni Ivianual
Incorrect resuits due to two barcodes on the same tube or mixing up specimen
barcodes
CAUTION Duplicate barcodes are not accepted on the system within the same run.
Take adequate measures to avoid placing incorrect barcode on specimen.
Make sure that there is only one barcode on each specimen tube.
Carryover
Hot surfaces
Roche Diagnostics
22 System Manual Version 1.0
.
cobas® 4800 System 1 General safety information
Systeni Manual Safety sunirnary
Data security
Unauthorized access and data loss due to malicious software and hacker attacks
Portable storage media can be infected with and transmit computer malware, which may
CAUTION be used to gain unauthorized access to data or cause unwanted changes to software.
The system is not protected against malicious software and hacker attacks.
The customers are responsible for IT security of their IT infrastructure and for protecting it
against malicious software and hacker attacks. Failure to do so may result in data loss or
render the system unusable.
Roche recommends the following precautions:
Allow connection to authorized external devices only
Ensure that all external devices are protected by appropriate security software.
Ensure that access to all external devices is protected by appropriate security
equipment. cobas IT firewall must be used when the system is integrated into a
network.
Do not copy or install any software on the software control unit unless it is part of the
system software or you are nstructed to do so by a Roche Service representative.
It additional software is required, contact your Roche Service representative to ensure
validation of the software in question.
Do not use the USB ports to connect other storage devices unless you are instructed to
do so by official user documentation or a Roche Service representative.
Exercise utmost care when using external storage devices such as USB flash drives,
CDs, or DVDs. Do not use them on public or home computers while connecting to the
system.
Keep all external storage devices in a secure place and ensure that they can be
accessed by authorized persons only
Do not enter any confidential patient-relevant information into the work order file.
There is the risk of unauthorized access to patient data.
Notices
Moving parts
Hoche Diagnostics
System Manual Version 1.0
.
23
1 General safety information cobas® 4800 System
Safety suniniary System Manual
Spillage
Roche Diagnostics
24 System Manual Version 1.0
.
cobas® 4800 System 1 General safety information
Systeni Manual Safety labels on the system
Warning labels draw your attention to areas of potential hazard. The labels and their
definitions are listed below according to their location.
The safety labels comply with the following standards: ANSI Z535.6, EN 15223-1,
IEC 61010-1, or ISO 7000.
1f the labels are damaged, they must be replaced by Roche Service personnel. For
replacement labels, contact your local Roche representative.
ID Label Meaning
A Pipetting Arm
Do not move pipetting arm by hand.
B Biohazard warning
4r),,
1 Instrument deck may contain biohazardous or chemically
contaminated materials.
C Power connection
Connect only to earth-grounded outlet.
Connection to PC
Use only the appropriate shielded cables.
USB connection
Having a total cable distance of more than 5 m, signals can be
interfered.
Table 1-1 Description of safety labels on the instrument
Roche Diagnostics
System Manual Version 1.0
. 25
1 General safety information cobas® 4800 System
Safety labels on the system Systeni .lv.tanual
ID Label Meaning
13 Barcode reader laser beam
Do not stare into beam of class 2 laser.
E Moving parts
Moving arm inside transparent cover. Aborts the run if cover is
opened.
F Biohazard warning
,(ç).. Waste may contain biohazardous or chemically contaminated
materials.
Table 1-1 Description of safety labels on the instrument
,\ —
Biohazard warning
Take adequate precautions when
•
•
On the Xenon lamp unit
On the margin of the microweil
plate loader
working with potentially infectious
material.
Table 1-2 Safety labels on analyzer
In addition to safety labels, there are safrty notes in the corresponding parts of this
manual and all other relevant manuals.
These safety notes give more detailed information about potentially hazardous
situations that may arise during daily operation or when carrying out maintenance
procedures.
During operation, be sure to observe both the safety labels and the safety notes in all
relevant manuals.
Roche Diagnostics
26 System Manual Version 1.0
.
cobas® 4800 System 1 General safety information
Systern Manual Disposal
Disposal
Roche Diagnostics
System Manual Version 1.0 27
1 General safety information cobas® 4800 System
Disposal Systeni Mamal
Roche Diagnostics.
28 System Manual Version 1.0
cobas® 4800 System 2 Overview
Table of contents Systeni Alanual
Overview
In this chapter you get a basic overview about the whole system inciuding the
instrument, the analyzer, and the software.
Roche Diagnostics
System Maniial Version 1.0 29
2 Overview cobas® 4800 System
Table of contents Systern Manual
Roche Diagnostics
ao System Manual Version 1.0
cobas® 4800 System 2 Overview
Systern Manual Systern overview
System overview
LIS
Orders Results
(test-specific)
4
cobas 4800 software
—-
.
Result reports
_ —
S
LK Control unit
Microweli plate
Samples with prepared samples
Reagents
Consumables
Tests that only use the PCR Only workflow, have the following components:
Roche Diagnostics
System Manual Version 1.0 31
2 Overview cobas® 4800 System
Systenz overview System ivlanual
LIS
t Resuits
•) .z=
Result reports
t
Control ooit
IVlicroweli plate
with manually 0
prepared specimens
Figure 2-2 System overview for tests which use only the PCR Only workflow.
Instrument
The instrument prepares the samples. The instrument is loaded with samples,
consumables and reagents. After sample preparation, the microweli plate with the
PCR-ready samples is unloaded, sealed, and transferred to the analyzer for
amplification and detection using real time PCR.
The procedure for sample preparation is as follows;
1. Specimen-specific processing
2. Generic binding, washing, and elution
3. Working master mix preparation and PCR setup
Roche Diagnostics
32 System Manual . Version 1.0
cobas® 4800 System 2 Overview
Systeni Manual Systeni overi’iew
Analyzer
The analyzer utilizes fluorescence signal to detect nucleic acids amplified by using
real time PCR methodologv.
Sealing, loading, and unloading the microweli plate are the only manual
interventions.
Roche Diagnostics
System Manual Version 1.0
.
33
2 Overview cohas® 4800 System
Systent overview Systern Mannal
The system software runs on a dedicated control unit. The software guides the user
through the entire process from sample preparation to amplification and detection
and result interpretation.
A hand-held barcode reader is connected to the control unit. It is used to scan reagent
and reagent reservoir barcodes during loading of reagents as well as sample barcodes
to set up the ivork order file.
For details about the control unit and the hand-held barcode reader, see Control unit
(p. 69)
115
The system can be connected to a Laboratory information system (LIS). Orders are
downloaded automatically from the LIS after samples are loaded onto the
instrument. After reviewing of the final results, results have to be manually uploaded
to the LIS using the export function in the Results work area.
The System overview tab displays the status of the LIS connection.
Roche Diagnostics
34 System Manual Version 1.0
.
cobas® 4800 System 2 Overview
System Manual Systeni overvielv
Adby St4os
1C
TrSt4,
-*, — os
Aveb(y
[J [J .
L IS
Trnsemn5
• Using result upload to LIS, results of the selected runs can be uploaded to the LIS. The
administrator can define whether only accepted resuits or all results can be uploaded
to the [IS.
• PCR Only runs are not fully supported by the [IS i.e. accepted results can be uploaded
to LIS but work orders cannot be received from LIS. Use the sample editor to create the
work orders for a PCR Only run.
• The LIS connection (if the LIS is present or not) can be configured by the lab manager
or administrator.
• The LIS time-out (maximal time between an LIS query and answer) can be configured
by the lab manager or administrator.
For information about the LIS host protocols, refer to the cobas 4800 System Host
Interface Manual.
Consumables
Roche Diagnostics
System Manual Version 1.0
. 35
2 Overview cobas® 4800 System
Systern overview Systern Man,ial
Roche Diagnostics
36 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Table of contents Systevn Manual
Hardware
Roche Diagnostics
System Manual Version 1.0
. 37
3 Hardware cobas® 4800 System
Table of contents Systens Manual
Roche Diagnostics
38 System Manual Version to
cobas® 4800 System 3 Hardware
Systeni Manual Instrument
Instrument
A Pipetting arm with pipetting head, D Stationary carrier with heater/shaker unit,
8 pipetting channels, and ISWAP magnet plate, and holders for microweli
B Instrument deck plate and deepweli plate.
Roche Diagnostics
System Manual Version 1.0
. 39
3 Hardware cobas® 4800 System
Instrument Systeni Manual
Covers
The front cover protects you from moving parts inside the instrument. The front
cover can be opened for user intervention during maintenance. Always close the
front cover when you have finished maintenance work. During a run, the front cover
must be closed.
Roche Diagnostics
40 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systeni Manual Instrument
LEDs
The loading status on the instrument deck is indicated by an LED bar above the
autoload tray. The status LEDs inform you where to place the carriers on the
autoload tray and which tracks on the instrument deck are occupied by a carrier.
A Track position
B [ED bar above the autoload indicating the loading status
Instrument deck
The work area of the instrument is called instrument deck. The instrument deck
hoids:
• removable carriers for samples, reagents, plates, and consumables.
• a stationary carrier used for sample processing. The stationary carrier is not
removable and hoids heater and shaker unit, magnet plate, and the plate holders
for the deepweil plate and the microweil plate.
Roche Diagnostics
System Manual Version 1.0
. 41
3 Hardware cobas® 4800 System
Instrument System Manual
Instrument deck layout The instrument deck is divided into 54 equal tracks. The tracks are numbered so that
you can readily identify the location where a carrier must be loaded. The track
positions are stamped on the surface of the autoload tray. A corresponding LED on
the LED bar above the autoload tray indicates the loading status of each track.
Track assignment is test-specific. The instrument deck layout below shows you an
example of the areas for each carrier type.
A Plate carrier for deepweli plate and microwell plate E Tip carrier (right)
B Tip carrier (left) F Stationary carrier used for sample processing
C Sample loading area G Reagent area for reagent reservoir carriers (200 mL and 50 mL)
D Depending on the test, sample loading area or tip carrier (middle) and reagent carrier
Track positions The following table summarizes the track positions on the instrument deck.
(test-specijic)
Track position Used for carrier
1-6 Platecarrier
7-10 Not used
11-16 Left tip carrier
17-34 Sample loading area (varies depending on the test)
29-34 Middie tip carrier
35-40 Right tip carrier
4 1-47 Stationary carrier for sample preparation
48-49 200 mL reagent reservoir carrier
50 50 mL reagent reservoir carrier
51-53 Reagent carrier
54 Not used
Table 3-2 Track positions on instrument deck of the instrument
Sample loading area The sample loading area can hold any combination of sample carriers.
For additional details about sample carriers, see Sample carriers (p. 50)
Roche Diagnostics
42 System Manual Version 1.0
cobas® 4800 System 3 Hardware
Systeni Manual Instrument
Autoload unit
The autoload unit consists of the autoload tray the platform where the carriers are
-
placed and the autoloader. The autoloader detects waiting carriers with their tubes
-
Inserting carriers on autoload tray The autoload tray contains slide bio cks that engage tracks in the carriers to guide the
carriers when they are loaded. When inserting a carrier onto the autoload tray, be
sure that the tracks on the carrier properly engage the slide blocks. Insert the carriers
into their designated position on the autoboad tray until they touch the stop hooks on
the far side of the tray.
Insert the carriers into the tracks between the front and rear slide blocks of the
Autoboad tray until they touch the stop hooks on the far side of the tray.
Roche Diagnostics
System Manual Version 1.0
. 43
3 Hardware cobas® 4800 System
Instrument Systeni Manual
Safety guards The autoload tray has safety guards attached on each side. Extend the safety guards to
avoid bumping into the carriers on the autoload tray.
S --
•
è --‘-
A Satetyguard
Autoloader with barcode reader During loading, the barcode reader on the autoloader scans the barcode of the carrier
and those of the supplies on the carrier (samples, reagents, and disposables).
The following barcode types are supported:
• Codabar (without check sum)
• Code 39 (without check sum)
• Code 128, subset B and C (with check sum)
• GTIN (inciuding JAN) (with check sum). This barcode type is not enabled by
default. 1f you require this barcode type, contact Roche Service.
• Interleaved 2 of 5 (without check sum). This barcode type is not enabled by
default. 1f you require this barcode type, contact Roche Service.
For details about barcode specifications, see Barcode reader (p. 63)
Auto matically unloading sample The instrument can automatically unload sample carriers i.e. samples, when pipetting
carriers is finished. This allows you to further process or store the samples without having to
wait for the run to finish. You can also define that the sample carriers be left in the
instrument and manually unloaded when the run is finished.
Roche tiiagnostics
44 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systeni Manual Instrument
Pipetting arm
The instrument is equipped with a pipetting arm containing eight pipetting channels
and the iSWAP device. During operation, the pipetting arm moves left-and-right (x
direction), picks up the required number of disposable pipetting tips, aspirates,
dispenses, and mixes the liquids, and then ejects the used tips into the tip solid waste
bag.
Each air-displacement pipetting channel can move forward-and-back (y-direction)
and up-and-down (z-direction) during the pipetting operation. A disposable
pipetting tip is picked up by each pipetting channel at the beginning of a pipetting
cycle. The pipetting tip is ejected into the tip solid waste bag at the completion of
pipetting for that cycle.
Each pipetting channel contains a pipette stop disc which aligns the disposable
pipetting tip when it is picked up by the pipetting channel at the beginning of a
pipetting cycle. Aspiration and dispensing is achieved by air displacement. The
pipette stop disc does not contact the liquid that is being aspirated into and dispensed
from the disposable pipetting tip. The pipette stop discs must be cleaned on a weekly
basis.
For additional details, see Weekly niaintenance (p. 110)
Roche Diagnostics
System Manual Version 1.0 45
3 Hardware cobas® 4800 System
Instrument System Manual
iSWAP
The landscape iSWAP is a robotic arm that picks up deepweli plates and microweli
plates and moves them to the desired location on the instrument deck during a run.
The ISWAP is mounted on the pipetting arm.
A iSWAP
Teaciting needies
The instrument uses eight teaching needies to automatically check (and adjust if
necessary) z-positioning and over- and under-pressure of each pipetting channel at
the appropriate time. The teaching needies are stored near the waste station, next to
the tip waste. The teaching needies are cleaned during the preventive maintenance
visit.
A Teaching needle
Roche Diagnostics
46 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systern Manual Instrument
Waste station
Safety information
Make sure that you have read and understood the chapter General safety information
(p. 13). The following safety messages in particular are relevant:
Warning messages:
Infection by biohazardous waste (p. 19)
Contanination of the environment by liquid waste and solid waste (p. 19)
Tip waste Solid vaste bags hold used pipetting tips after they have been ejected by the pipetting
arm.
The instrument can be installed on either a bench top or on a trolley that
accommodates a larger solid waste bag.
• 1f the instrument is installed on a bench top, a small solid waste bag is used. This
solid waste bag has a capacity for tips of at most one full run.
E9 Exchange the small solid waste bag each time before starting a new run to avoid overfilling
of the tip waste.
For details about replacing the small solid waste bag, see Daily niaintenance (p. 105) and
Veekly niaintenance (p. 110)
A Initialization/waste block
B Support frame
C Small solid waste bag with a capacity for tips of at most one full run
Figure 3—10 Small solid waste bag
Roche Diagnostics
System Manual Version 1.0
. 47
3 Hardware cobas® 4800 System
Instrument Systeni Manual
• The waste chute and waste box cannot be reused and must be replaced when they are
full.
• 1f required, replace the waste chute in the same way as the small tip solid waste bag.
• 1f required, fold up a new waste box as indicated in Figure 3-12 Folding up a new waste
box (p. 49)
For details about replacing the large solid waste bag, see Daily maintenance (p. 105) and
Weekly maintenance (p. 110)
Waste box The waste chute and waste box cannot be reused and must be replaced when they are
full. Fold up a new waste box in the following way:
Roche Diagnostics
48 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systern Manual Instrument
Liquid waste The liquid waste from the instrument is collected in an external waste container. The
level of the liquid waste is monitored by the system and a warning is issued when it is
not possible to do a full run. In this case, it is not possible to start the run.
The liquid vaste is emptied during daily or weekly maintenance. The liquid waste
container is reusable.
For details about emptying the liquid waste container, see Daily maintenance (p. 105) and
tVeekly maintenance (p. 110)
/
/
A
A Liquid waste cap with level detection sensor and waste tube
B Liquid waste container
Roche Diagnostics
System Manual Version 1.0 49
3 Hardware cobas® 4800 System
Instrument System Manual
A Insert the liquid waste cap as indicated B Close the liquid waste cap
Figure 3—14 Inserting and closing the liquid waste cap of the liquid waste container
Carriers
Sample carriers
Different sample carriers allow the use of samples in primary and secondary tubes.
Depending on the test, up to 94 patient specimens can be loaded onto the instrument
per run.
Roche Diagnostics
50 System Manual Version 1.0
cobas® 4800 System 3 Hardware
Systeni Manual Instrument
Each run has test specific controls e.g. positive/negative control. Controls are loaded
together with the reagents onto the reagent carrier.
Barcode dirnensions c B c
.
EL1
A
U
A Label length: max. 80 mm
B Code Ienght: max. 74 mm
C Quiet zone: min. 3 mm
D Label width: min. 12 mm
E Code width: min. 12 mm
F Distance from code to label edge: max. 1 mm
Positioning barcode labels on The barcode must fit within a range of between 20 mm and 100 mm from the bottom
sample tubes of the tube.
100 mm
2OmmL
Roche Diagnostics
System Manual Version 1.0
. 51
3 Hardware cobas® 4800 System
Instrument Systenz Manual
24-position saniple carrier The 24-position sample carrier hoids up to 24 primary or secondary tubes.
(test-specflc)
An aliquot can be manually transferred from the primary tube to the secondary if
necessary.
The 24-position sample carrier uses one track on the instrument deck.
EJ3 For details about the types of tubes you can use on the sample carrier, refer to the test
specific package insert.
Infection by samples and associated materials
1-track carriers (such as a sample carrier) can fall over and cause injury or contamination.
CAUTION Handle 1 -track carriers with particular care, position them on the autoload tray or place
several carriers together to minimize this risk.
•1v
t— B
,- S.
A Carrier barcode
8 Primary or secondary tubes
Figure 3-18 24-position sample carrier
32-position sample carrier The 32-position sample carrier is optionally available and hoids up to 32 primary or
(test-spec fic) secondary tubes.
An aliquot can be manually transferred from the primary tube to the secondary tube
if necessary.
The 32-position sample carrier uses one track on the instrument deck.
For details about the types of tubes you can use on the sample carrier, refer to the test
specific package insert.
Collection medium carrier The collection medium carrier is optionally available and hoids up to 12 Collection
(test-specfic) medium primary tubes.
This carrier uses two tracks on the instrument deck.
Open the collection medium containers before inserting the carrier on the autoload tray.
Roche Diagnostics
52 System Manual Version 1.0
cobas® 4800 System 3 Hardware
Systern Man ual Instrument
Reagent carriers
Reagent reservoir carriers The reagent reservoir carriers hold the reagent reservoirs. The reagent reservoir
carriers are available in two sizes:
• 200 mL reagent reservoir carrier for 200 mL reservoirs
• 50 mL reagent reservoir carrier for 50 mL reservoirs
The required reagents are poured into the reagent reservoirs which are then placed
onto their dedicated positions on the reagent reservoir carriers. The reagent reservoir
barcodes must face to the right of the carrier.
For details about reagents and the placement on reagent reservoir carriers, refer to the
test-specific Operator’s Manual.
Aø
Roche Diagnostics
System Manual Version 1.0
. 53
3 Hardware cobas® 4800 System
Instrument Systeni Manual
A 50 mL reagent reservoir
B Carrier barcode
Reagent carrier The reagent carrier hoids the test-specific reagents for sample processing and PCR
setup (controls, master mix, metal co-factor(s), and so on).
The required reagents are decapped and then placed onto their dedicated positions
on the reagent carrier. The reagent barcodes must face to the right of the carrier.
For details about reagent placement on reagent carriers, refer to the test-specific
Operator’s Manual.
The reagent carrier is the same carrier as the 24-position sample carrier.
Roche Diagnostics
54 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systent Manual Instrument
Plate carrier
The plate carrier hoids the deepweli plate and the microweli plate. The deepweil plate
is used to process the samples. The microweli plate is used to hold the final prepared
samples that are ready for PCR.
Place the deepweil plate on position 1 and the microweli plate on position 3 of the
plate carrier. The other positions on the carrier are not used. The deepweli plate
barcode must face to the right of the carrier. The deepweil plate barcode is read
during loading of the plate carrier. The microweil barcode is read after loading being
brought in front of the barcode reader on the autoloader by the iSWAP.
At the end of the run, the plates are placed back on the plate carrier. Unload the plate
carrier, seal the microweil plate and dispose of the deepweil plate according to the
local regulations unless further off-the-system use is intended.
The total number of pipetting tips per run varies and depends on several criteria (test type,
sample media, run size, and so on)
The instrument tracks tip usage from run to run. The instrument checks if enough pipetting
tips have been loaded to perform the run. 1f there are not enough, a message is displayed.
To avoid this message, fully bad all the required tip rack carriers for each run.
Partially used racks can be used in next run.
To perform a run, all the required tip rack carriers must be loaded into the instrument.
Roche Diagnostics
System Manual Version 1.0 55
3 Hardware cobas® 4800 System
Instrument Systeni Manual
A Tip raak with disposable pipetting tips. Each tip rack holds 96 tips.
Stationary carrier
The stationary carrier is the area where the samples are processed. It is mounted on
the instrument deck and is not removable. The stationary carrier hoids four units:
• heater/shaker unit
• magnet plate
• deepweli plate holder
• microwell plate holder
Heater/shaker unit The heater/shaker unit provides consistent and reliable heating and shaking during
sample processing.
Roche Diagnostics
56 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systern Manual Instrument
The heater/shaker unit is powered from an external controller box located outside of
the instrument. The heater/shaker unit must be switched on and offon the external
controller box.
Do not forget to switch on the heater/shaker unit on its external controller box.
Always switch off the heater/shaker unit before unplugging It.
Magnetpiate The magnet plate immobilizes magnetic glass particles during sample processing.
Be aware, that the magnets are not affixed to the plate so any metal can remove the
magnets from its holder.
Holders for deepweil plate and The holders for the deepweli plate and the microweli plate are where the washing and
microweliplate PCR setup occurs during sample processing.
Roche Diagnostics
System Manual Version 1.0
.
57,
3 Hardware cobas® 4800 System
Instrument System Manual
Consumables
NOTICE Instrument damage and incorrect resuits due to use of unsuitable consumables
Use of non-Roche consumables may damage the instrument or lead to incorrect resuits.
Use only Roche consumables designed for use on the system.
Do not reuse consumables. All consumables are designed for one time use only.
Inspect consumables before use. Do not use damaged consumables.
Do not use consumables that have exceeded their expiry date. Replace expired
consumables with unexpired consumables before sample processing.
Pipetting tips
Disposable pipetting tips with filters are used for pipetting of all liquids during
sample processing. The pipetting tips are available in tip racks of 96 tips each. Tip
racks are placed into tip rack carriers.
For details about tip rack carriers, see Tip rack carriers (p. 55)
Reagent reservoirs
The reagent reservoirs hold the reagents used for sample preparation. The reagent
reservoirs are barcoded and are filled manually by the operator (scan-scan-pour
place principle) for each run.
Scan-scan-pour-placeprinciple To minimize handling errors the reagent reservoirs are filled and placed using the
scan-scan-pour-place principle:
1. Scan the barcode of the required reagent using the hand-held barcode reader.
2. Scan the barcode of an unused reagent reservoir using the hand-held barcode
reader.
3. Pour the reagent in the scanned reagent reservoir.
4. Place the filled reagent reservoir onto the required position of the reagent
reservoir carrier as indicated in the wizard.
Roche Diagnostics
58 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systern ?vlanual Jos trument
Reagent reservoir sizes The reagent reservoirs are available in two sizes:
• The reagent reservoirs are designed for one time use. The software tracks the use of
the reservoirs and rejects previously used reagent reservoirs.
• Do not fiN reagent reservoirs above the maximal filI height. A sign within the reagent
reservoir indicates the maximal fili height.
• Always pour the entire reagent vial in the scanned reagent reservoir to avoid
underfilling.
• Handle filled reservoirs with particular care to avoid spiashes and tipping over.
• It is advisable to pour the reagent into the reservoir in a lengthwise movement to
minimize the risk of spiashing and resulting reagent loss.
• Do not pour reagents into reservoirs that are already placed onto a reagent rack.
Always follow the scan-scan-pour-place principle.
• Do not reuse reagents from previous runs.
Deepweli plate
The deepweil plate is used for sample processing. The deepweil plate hoids up to
96 samples.
The deepweil plate is placed in the plate carrier on position 1 prior to a run.
For details about the plate carrier, see Plate carrier (p. 55)
• The deepweli plate is barcoded and designed for one time use. The software tracks the
use of the plate and rejects previously used deepweli plates.
• Open the primary packaging of the deepweli plate before use. Open the packaging
only in a clean environment to prevent contamination of the plate.
• Never touch the weNs of a deepweli plate to avoid contamination.
• Do not seal the deepwell plate before loading the plate into the instrument.
Roche Diagnostics
System Manual Version 1.0
.
59
3 Hardware cobas® 4800 System
Instrument Systeni !ulanual
Microweli plate
The microweli plate (AD-plate 0.3 mL) hoids the processed samples mixed with
working master mix reagent and is used for amplification and detection on the
analyzer.
The microweli plate is placed in the plate carrier on position 3 prior to a run.
For details about the plate carrier, see Plate carrier (p. 55)
After final sample processing on the instrument the microweil plate is placed back on
the plate carrier by the iSWAP. Unload the plate carrier and seal the microweli plate
using the sealing film.
iJ For details about sealing of the microweli plate, refer to the test-specific Operator’s
Manual.
The microweil plate is barcoded and designed for one time use. The software tracks
the use of the plate and rejects previously used microweli plates.
• Open the primary packaging of the microweli plate before use. Open the packaging
only in a clean environment to prevent contamination of the piste.
• Never touch the weils of a microweil piste to avoid contamination.
• Do not seal the microweli piste before loading the piste into the instrument.
Roche Diagnostics
60 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
System A’fanual Instrument
N
B
r
The waste chute and waste box cannot be reused and must be replaced when they are full
Secondary tubes
You can use specific tubes when a secondary tube is needed on the instrument for a
24-position or 32-position sample carrier.
For details about the types of secondary tubes you can use, refer to the test-specific
package insert.
The use of any tube other than those described, may lead to incorrect results or
processing failures.
Every effort has been made to ensure that all the information contained in these
specifications is correct at the time of printing. However, Roche reserves the right to make
any changes necessary without notice as part of ongoing product development.
Roche Diagnostics
System Manual Version 1.0
. 61
3 Hardware cobas® 4800 System
Instrument System Manual
Roche Diagnosties
62 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systenz Manual Instrument
Roche Diagnostics
System Manual Version 1.0
. 63
3 Hardware cobas® 4800 System
Instrument Systeni Manual
Roche Diagnostics
64 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
Systern Manual Analyzer
Analyzer
The analyzer is a rapid thermal block cycler with integrated real-time detection
capabilities.
Multiplex real time PCR assays allow the simultaneous detection of signals from
rnultiple dyes. This leaves It open how many filters are used by different IVD assays.
As a benefit of this, multiple resuits can be reported from a single run.
NOTICE Analyzer damage and incorrect resuits due to unsealed microweil plate
Before loading the microweli plate into the analyzer, it has to be sealed properly with the
self-adhesive sealing film. Sealing the plate is crucial to eliminate evaporation at high
temperatures. Unsealed microwell plates may damage the analyzer or lead to incorrect
results.
Always seal the microweil plate before loading the plate into the analyzer.
For details about sealing of the microwell plate, refer to the test-specific Operator’s
Manual.
Thermal block cycler unit • Thermal block cycler with block cycler cover, ventilation, microwell plate loader,
and microweil plate barcode reader.
Roche Diagnostics
System Manual Version 1.0
.
65
3 Hardware cobas® 4800 System
Analyzer Systern Manual
BÇ
Roche Diagnostics
66 System Manual Versian 1.0
.
cobas® 4800 System 3 Hardware
Systern Manual Analyzer
LEDs
The front of the analyzer provides two status LEDs which inform the operator of the
hardware status. The bad button for opening and closing the microweil plate loader
is bocated next to the two LEDs.
1!.)
Every effort has been made to ensure that all the information contained in these
specifications is correct at the time of printing. However, Roche reserves the right to make
any changes necessary without notice as part of ongoing product development.
Roche Diagnostics
System Manual Version 1.0
.
67
3 Hardware cobas® 4800 System
Analyzer Systern Manual
Roche Diagnostics
68 System Manual Version 1.0
.
cobas® 4800 System 3 Hardware
System Manual Control unit
Control unit
A dedicated control unit runs the software, controls the instrument, and the analyzer.
Incorrect resuits and malfunction due to use of unauthorized software
Use only manufacturers original installation CD-ROM or DVD sets for the operating
CAUTION system, and the original software.
Any manipulation of system data files or other information determining or affecting
system functions can result in erroneous results or equipment failure.
Only use the control unit that is supplied with the system.
1
A Display
B Display gooseneck
C Control unit
Figure 3-35 Control unit with display gooseneck
A hand-held barcode reader is connected to the control unit. It is used to scan reagent
and specimen barcodes during creation of the vork order file for microweli plate
preparation.
Usage The barcode label is read by holding the head of the barcode reader in close proximity
to the barcode label and then pressing the button on the barcode reader. The barcode
ID is then automatically read and displayed on the screen.
Be sure to hold the head of the hand-held barcode reader over the correct barcode label.
It is recommended to visually verify the displayed scanned characters with the human
readable characters on the barcode label.
Roche Diagnostics
70 System Manual Version 1.0
cobas® 4800 System 4 Software
Table of contents Systern Manual
Software
Roche Diagnostics
System Manual Version 1.0
. 71
4 Software cobas® 4800 System
Table of contents System Manual
Roche Diagnostics
72 System Manual Version 1.0
cobas® 4800 System 4 Software
System Manual Basic software elenzents
The screen of the software is divided into the following dedicated areas making the
software easy to understarid and intuitive to use:
Status area
• Tab navigation bar
• Main work area
• Global action bar
Alarm area
The screen representations shown in this chapter and throughout this manual are for
illustrative purposes only. The screens do not necessarily show valid data.
A[
Ft.l..,• R*sats Messagos U1lI,t,es
B
—
0061010
LIS
-
,a0412406
6
LJ LJ
00
a rw l69t9e —
N0.vaIlaOe
100 69e
.,
/
E
A The status area displays the currently logged D The global action bar contains buttons used
in user as well as and date and time. for general software functions. These
buttons are always available.
B The tab navigation bar displays the tabs. E The alarm area displays the most recent
Choose a tab to open t. In the first row of the alarms and warnings that are not yet
navigation bar, the tabs of the main work confirmed by the operator. Select an alarm in
areas are displayed. In the second row, the the list and choose the Alarm button on the
subtabs belonging to the currently selected left to get more details about the selected
work area are displayed. alarm.
C This is the main work area. It displays the
content of the currently selected tab.
Roche Diagnostics
System Manual Version 1.0
. 73
4 Software cobas® 4800 System
Basic software elements Systenz Manual
Colors
Color Meaning
Normal condition. The status is 0K.
S Green
Warning condition. Status is not 0K, but immediate intervention is
I Yellow not required.
. Red
Alarm condition. Immediate intervention is required. Operation
may have stopped.
The item is selected or active.
S Blue
Table 4-1 Color concept
Tabs
Tabs are used to group information and software functions into units that can be
displayed on one screen. The second row of tabs represent subgroups of the tab
selected in the first row. The tab in the third row is a subtab of the tab selected in the
first row second row.
Navigating within the software needs only one or two clicks.
L Choose a main tab in the first tab row. Here you have access to the five main work
areas Overview, Workpiace, Resuits, Messages, and Utilities.
2. Choose a subtab in the second row. Here you have access to all tabs of the selected
work area.
A Main tabs. Overview is selected. The first row of tabs give access to the main work areas.
B Subtabs. System is selected. The second row of tabs show all tabs of the selected work area.
C Subtab belonging to lnstruments tab. It is only visible t the Instruments tab is selected.
Roche Diagnostics
74 System Manual Version 1.0
.
cobas® 4800 System 4 Software
Systern Manual Basic software elerneuts
Buttons
Text buttons Choose a button to start the associated function. Some text buttons have a triangular
marking in a corner. This triangle telis you what exactly happens on screen if you
choose the button.
Button Behavior
r 1 Performs a specified action in the current window
CanceI j
A triangle in the top right corner of a button telis you that a new
L J dialog box will be displayed when you choose the button.
1 A triangle in the bottom left corner of a button Leus you that the
r
L J current window or dialog box will be closed when you choose the
button.
Table 4—2 Triangular markings in text buttons
Global action buttons On the right side of the screen, the global action buttons are available. They are
always visible and perform the following functions:
[ ij
.
ï1
Exit
‘€1 Export
Roche Diagnostics
System Manual Version 1.0
. 75
4 Software cobas® 4800 System
Basic software elements Systern Alanual
IEI
Log oS
Lists
The software provides powerful sorting and filtering tools for information displayed
in tables. These allow a lengthy list to be condensed in ways that provide fast access to
items of interest. It is also possible to hide certain columns, display additional
columns, and change their left-to-right order.
This section just gives a general overview. Some functions are not available in some of
the work areas, or less options are available, eg. in a context menu.
Organizing Iists
The sorting and filtering tools are accessed by right- or left-clicking the respective
column headings.
1
aliî 201fl 11 2149 Admm D112 AOr 201 s DOO,Ç 1 1 (J) Ad
•12.Açr.2010 11 352B Adoin 112Apr-201, O(’)
(N880)
5
tz-gc,.lo yJ-xx
G0O
12-40,201080 80)2
12-40, 2010 8080x
12-40,20108080:80
12-440-21080:80)2
400,
12-44,201000:00.80
12-4,201280:80.80
1240201380:80.00
12.44,201080:80:80
80:80:80
IZD40-2010 80:80:00
12-40-201080:80:80
10-40-201001-0080
12-440-201080:0080
C..’iit.
A Click the column header to sort the table in ascending or descending order. The sort order is
indicated by an arrow.
B Right-click the column header to display context menu for sorting and grouping information
items.
C Click the filter symbol (funnel icon) to display a list with appropriate filter options for this
column.
Roche Diagnostics
76 System Manual Version 1.0
.
cobas® 4800 System 4 Software
Systeni Pvlanual Basic software elements
Sort by column
Click a column header to sort the table by the entries of this column. Successive clicks
switch between ascending and descending sort order, the order is indicated by the
arrow that appears in the header.
Not every column is sortable. 1f a column is sortable, en arrow indicating the sorting
direction will appear in the column header after clicking the column header.
A SortAscending
Sc’rt Descending
f’lear Scrtinq
A Select the sort order, or dear any sorting. D Remove the selected column from the table.
B Group listed items by the values of this E Add or remove a column from the table
column (dick + or next to individual
-
The Column Chooser dialog box contains all available information items that are not
included in the table. Which items are available depends on the table from which you
started the Column Chooser.
Roche Diagnostics
System Manual Version 1.0 77
4 Software cobas® 4800 System
Basic software elernents Systern Manual
To add a column (information item) to the table, select the item in the Column
Chooser and drag it to the appropriate place in the table header.
To remove a column from the table, select its header and drag it into the Colunm
Chooser.
Some columns (e.g Sample ID, Flags) are mandatory and cannot be removed.
A Select Custom to display the Custom AutoFilter dialog box for reflning the filter criterion.
B Select the item that should serve as a filter criterion.
C Click the filter symbol (tunnel icon) to display the filter context menu.
When a filter criterion is applied, its details are displayed in a bar at the bottom of the
list (A).
1 jaJtrr:.-a
IItError
1 Warning 1204.2010 1010.
120420101154
IA Error 120420101417
Iterror 1204,20101154
Ikerror 120420101417
x 5
To return to the full view at any time (no filtering), choose the close button in the
filter details bar. Prevlously used filter settings remain accessible from the filter
history.
For more information, see Managingfilters and advancedfiltering (p. 79).
Custom AutoFilter
The Custom AutoFilter option is available only in those columns that have a filter
symbol (funnel icon) in their column header.
With the help of the Custom AutoFilter you can apply one or two filter conditions,
logically linked by AND or OR.
Click the filter icon in a column heading and select (Custom) to display the Custom
AutoFilter dialog box.
Roche Diagnostics
78 System Manual Version 1.0
-
cobas® 4800 System 4 Software
Systern Manual Basic software elernents
AB C DE
\ 1 II
[SeJentyj <Fatar
A Click to close the filter details bar and D Click to display previously used filters [filter
remove all filters. listory].
B Click to switch current filter on and off. E Filter history list: Click a filter entry to apply
C Current filter condition this prevlously used filter.
For information about filtering in the Resuits work area, refer to the test-specific
Operator’s Manual.
Tool tips
Tool tips provide brief functional descriptions when the mouse is paused over a tool
or application button. They mainly occur in the sample editor and give information
to the user when there is a mistake in editing.
Test Information
Creation 2701.2011 15:53:50
Microweli Plate ID
Oarcode is mandator_L__
B
A Box where an error occurred fl Tool tip informing about the nature of the
error
Roche Diagnostics
System Manual Version 1.0
. 79
4 Software cobas® 4800 System
Basic software elemeats Systeni Manual
Wizard
I
Figure 4—11 Steps in a run
Lond
End
It is not possible to go back to a previous step in a run. Follow the guidelines outlined in
this manual to avoid losing reagents, samples, or disposables.
24 LtARa,, 4447.444
— c
Roche Diagnostics
80 System Manual Version 1.0
.
cobas® 4800 System 4 Software
System Manual Database
Database
No run can be started when the database has reached the maximal size. Use the purge
and archive function to clean up the database. To optimize performance, It is
recommended to purge and archive results older than 6 months.
For information about the purge and archive function, see Archiving (p. 90).
The purge and archive function only covers run data and results. Messages are kept in the
software until they are deleted manually. Therefore, it is recommended to manually delete
confirmed messages from time to time to free up additional space in the database.
The actual size of the database is displayed in the System overview tab.
A
jr
Li
0
‘—4,4—,—
Audit trail
Roche Diagnostics
System Manual Version 1.0 81
4 Software cobas® 4800 System
Database Systeni Manual
041h1 1 14 01 40
3
1rI..9 420tfl 34.1 dI 011204004 13130403
3 3313147131
fl2 fl01..C34.k .432320030004
11.23,2013122012 t04o,r, 44C2C018501034 3,33 0.940 344 33 $00
lldç,.2013 122449 043330334. o4.3200I8000JI 0,4. 04937
B- 25
1
3
lI.Oo,.?fl1
4
4 2 3001,4033% 44.C2031203100 0,,, 0333.. 333 04, $10
ki4,o.4.,,
—r—
W,9{,9.3&d203f2, 0
B—...,4 — c
--
You can export the audit trail as a CSV file by selecting one or more entries and
choosing the Export button.
Roche Diagnostics
82 System Manual Version 1.0.
Configuration
5 Configuration 85
cobas® 4800 System 5 Configuration
Table of contents Systern J’vianual
Configuration
In this chapter the configuration of the system and the export of support data is
described.
Roche fliagnostics
System Manual Version 1.0 85
5 Configuration cobas® 4800 System
Table of contents Systeni Manual
Roche Diagnostics
86 System Manual Version 1.0
.
cobas® 4800 System 5 Configuration
Systern Manual Configuration
Configuration
The accessible configuration settings are dependent on the user group of the logged in
operator.
• Users assigned to the Supervisor group can access all configuration settings
described below.
• Users assigned to the Operator group can only export support data (data for Roche
Service) and change their password.
System settings
soaoo.orw. 010
Ir 1115 44* 044orllpsOd r
,00o0 ar
bLlS401orertr*n
40, sporrso..saldalr
11150000* 40 s4Wior500orr.004050 ‘s
-
*olwlrflal015qo0rT,rWolO 500 MMrortor lig .04 pr,tlonr 000110
€0
El
Roche Diagnostics
System Manual Version 1.0
. 87
S Configuration cobas® 4800 System
Configuration Systent Manual
Reporting
- L1... —:Z:
SIwOflotyIflIU,Ifl, Det.ayt*Odfl,NIflflk9by 0
Roche Diagnostics
88 System Manual Version 1.0
.
cobas® 4800 System 5 Configuration
Systern Manual Configuration
Tests
Use the Tests group to disable or enable tests on the system. Disabled tests cannot be
run on the system. You can also assign a test to a user group and sort the order of how
the test are displayed.
Ove4viW •l R,its
Epot1 tippot d{ Rt S
3ttI*O
1 - -
Sort order for workflow Number (0 to Define a number to sort the order of how
‘Test A Full” 20) the test are displayed on the Select test
Sort order for workflow Number (0 to dialog box. For example, a number higher
“Test B Full” 20) than 0 will more the test downwards on the
Select test dialog box.
Workflow “Test A Full” is All From the drop-down list, choose a user
allowed for group None group.
Supervisor
Operator
Table 5-3 Tests parameters
Roche Diagnostics
System Manual Version 1.0
. 89
5 Configuration cobas® 4800 System
Configuration System Manual
Archiving
,rrn41
Lh.,D
Overview Resuits Messages 10.103
Usurs Export support dsta Rosult 000310 a
Roche Diagnostics
90 System Manual Version 1.0
.
cobas® 4800 System 5 Configuration
System Manual CoJ/iguration
Maximum age of audit trail Number (10 to Number of days after which the audit
records 10000) trail records will be deleted.
Configured by the administrator.
Purge and archive p j’ Select the check box to enable the
purge and archive function.
Deselect the check box to disable the
purge & archive function.
Purge and archive path Directory path Location where the resuits are
archived. The archive folder can be
located on a server.
Type the full path to the archive folder
and use Windows naming
conventions. (e.g. D:\Archive)
Allernatively, choose and browse
for the path.
Maximum age of results Number (10 to 500) Number of days after which the resuits
will be archived and purged from the
database. All results that are older than
the specified days will be archived and
purged.
To optimize performance, it is
recommended to purge and archive
results older than 6 months
(approximately 180 days).
Roche Diagnostics
System Manual Version 1.0 91
5 Configuration cobas® 4800 System
User management System lt’fanual
User management
Utilities> Users
There are three groups of functions available for managing users:
• Changing passwords
• Managing user accounts
• Setting up rules for passwords and user accounts
• Only users that are assigned as Supervisor can manage users and can set up rules for
passwords and user accounts.
• Users assigned as Operator can change their password only.
• Keep user access information on a secure place and do not make t public to prevent
unauthorized access to the system.
Password management
Changing a password
Any user can change their password. Lab managers can change the passwords of all
users. The password must follow the password rules that are defined in the software.
For more information, see Password rules (p. 96).
Roche Diagnostics
92 System Manual Version 1.0
.
cobas® 4800 System 5 Cunfiguration
System ivlanual User management
Res W.esses - -
UWkce* UW
Epot ,ppoh d,CU5
- UW -
t VT1
KI=’
UW -
--
- -j
Ii
UWWW LJ
H)
1 -
•fl..aaflU,
User riglits A user assigned to the Operator group has the following user rights:
• Starting and aborting runs
• Viewing result details and accepting, exporting, and printing results
• Performing maintenance
• Confirming messages
A user assigned to the Supervisor group has the same user rights as the Operator
group plus:
• Editing configuration settings
• Managing user accounts and editing account rules
• Deleting messages
Roche Diagnostics
93
System Manual Version 1.0
5 Configuration cobas® 4800 System
User management Systeni Manual
Sza —,
t R___.,._
rr
ri
— ,,,,,..,,M.wop,,a S
r)
!._._.
Figure 5—6 Defining a new user account
Roche Diagnostics
94 System Manual Version 1.0
.
cobas® 4800 System 5 Configuration
System Manual User management
0
Deleting user accounts
0
Unlocking user accounts
By default, user accounts become locked after five unsuccessful logon attempts.
Locked users are indicated in the Users list.
0
Rules
Roche Diagnostics
System Manual Version 1.0 95
5 Configuration cobas® 4800 System
User management System Manual
Account rules
Accoont rules
Password rules
Password rutes
mh,rn,uçnnundeeofdets r
mlrsmum number of nondphanurrrenc characters (eg $5. %)
Roche Diagnostics
96 System Manuai Version 1.0
.
cobas® 4800 System 5 Configuration
Systern Manual Exporting support data
1f necessary, support data for individual runs or a batch of runs can be exported to an
external storage device. The support data will help Roche Service to support you
optimally.
114 5H
—
[E]
Roche Diagnostics
System Manual . Version 1.0 97
5 Configuration cobas® 4800 System
Result counters System Manual
Result counters
View information about the number of tests that were completed, failed, and invalid.
— Utilities > Result counters
U44U
!
Overyw Rs4!ts Messae I5NU,. —
Ur Epo4ppoid4t4 ;ÇWT
441.448 ‘t 44 28 tÇ4
To filter the result counter, activate the date filter, define a date range, and choose the
Update counters button.
You can use the Export button to export the resuits counters as an XML file. Only the
displayed resuits are exported.
You can use the Print button to print out the result counters.
Roche Diagnostics
98 System Manual Version 1.0
.
Maintenance
6 Maintenance 101
cohas® 4800 System 6 Maintenance
Toble of contents Systern Monual
Maintenance
In this chapter .
Roche Diagnostics
System Manual Version 1.0
. 101
6 Maintenance cobas® 4800 System
Table of contents Systern Manual
Roche Diagnostics
102 System Manual Version 1.0
.
cobas® 4800 System 6 Maintenance
Systeni Manual Safety information
Safety information
Before performing any maintenance, read the safety messages carefully. 1f you ignore
safety messages, you may suffer serious or fatal injury.
Roche Diagnostics
System Manual Version 1.0 103
6 Maintenance cobas® 4800 System
Instrument maintenance System Manual
Instrument maintenance
Maintenance for the instrument is initiated and managed directly from the software.
The System tab shows the actual maintenance status of the instrument. Maintenance
is started from the cobas x 480 tab.
Periodic maintenance must be performed in order to ensure safe and reliable
operation of the instrument. A maintenance procedure is completed once the
procedure has been fully executed and the resuits are within the specifications.
Aborting maintenance procedures A maintenance procedure must always be finalized. Do not abort a maintenance
procedure. Aborting a maintenance procedure will lead to a failed status, and
maintenance will need to be started again. After aborting a maintenance procedure,
switch off the instrument and switch it on again after 10 seconds.
Weekly maintenance Recommended at the end of the week before shutting down the instrument.
1f an error is encountered during a maintenance procedure, try to resolve the
problem and re-start the maintenance procedure. 1f you cannot resolve the error
yourself, cail Roche Service.
Preventive inaintenance As part of the preventive service program two scheduled preventive rnaintenance
service visits per year will be required. Preventive maintenance is performed by
Roche Service. The duration of each preventative maintenance procedure will be
approximately 4 hours.
Roche Diagnostics
104 System Manual Version 1.0
.
cobas® 4800 System 6 Maintenance
Systeni Ivlanual Instrument maintenance
The system control unit is normally left on, except for a daily restart (e.g. as part or
before starting the daily maintenance).
For startup procedures, if the control unit is powered off, refer to the test-specific
Operator’s Manual.
Safety
Make sure that you have read and understood the chapter General safety information
(p. 13) and Safety information (p. 103)
The surface of the instrument should be cleaned on a weekly basis with a lint-free
cloth moistened with deionized water. 1f a spul of reagents occurs, the instrument
should be cleaned with a 70% ethanol solution.
Removable internal parts of the instrument, such as all carriers, should be weekly
rinsed with deionized water followed by 70% ethanol.The area around the instrument
should be periodically checked to ensure that air flow around the instrument is
unrestricted and that books, papers, or other supplies do not interfere with air flow.If
a specimen or other biohazardous material spills on the instrument or any of the
instrument carriers, the area or carriers should be cleaned with a decontaminating
and nucleic acid destroying agent, followed by deionized water.
• Do not use other organic solvents (eg., petroleum, benzene, or other solvents) because
they can damage plastic materials.
• Do not use an alcohol solution with a concentration greater than 70°/s because this
may damage transparent viewing windows.
• Only use completely dry carriers on the instrument.
Daily maintenance
Roche Diagnostics
System Manual Version 1.0 105
6 Maintenance cobas® 4800 System
Instrument maintenance System Manual
DaiIymatnIenore cIekdeckcteanog
ILZ...i:z_:.JI 1
The operator now has access to the instrument deck to check if cleaning is
required.
5 Open the front cover and inspect that all the parts of the instrument or carriers
are clean.
• 1f the instrument deck is clean continue with step 6.
• 1f the instrument deck requires cleaning, stop daily maintenance. Perform
weekly maintenance. To stop daily maintenance, choose the Cancel button in
the message box.
6 Choose the 0K button.
The next message box is displayed.
j Cancel
Roche Diagnostics
106 System Manual Version 1.0
.
cobas® 4800 System 6 Maintenance
Systern Manual Instrument maintenance
7
A Initialization/waste block C Tip eject plate
B Waste bag or waste chute D Support frame
When replacing the small waste bag, place the bottom portion of the waste bag
inside the support frame, fold the top over the frame and then place the support
frame into the initialization/waste block. Dispose of either waste bag and contents
according to the appropriate local regulations.
Replacing large waste bag
1f a waste container is used remove the large waste bag from the waste container
and replace it with a new one. Close the large waste bag with a binder and dispose
of either waste bag and contents according to the appropriate local regulations.
• The waste chute and waste box need to be replaced when full and cannot be
reused.
• 1f required, replace the waste chute in the same way as the small tip waste bag.
• 1f required, fold up a new waste container as indicated in Figure 3-12 Folding up a
new waste box (p. 49)
Roche Diagnostics
System Manual Version 1.0
. 107
6 Maintenance cohas® 4800 System
Instrttnient nzaintenance Systern Manual
8 Remove the tip eject plate of the tip waste station (see Figure 6-1 Reinovingwaste
bag or waste chute (p. 107)) and clean it with disinfectant solution.
Put the clean tip eject plate back in place properly. Make sure the tip eject plate
fits horizontally on the initialization/waste block. 1f it is not placed back properly,
this may cause a run to crash with instrument malfunction.
9 Check the status of the liquid waste container. Empty the waste container ifit
requires emptying. Dispose of the liquid waste according to the appropriate local
regulations.
For additional details, see Liquid waste (p. 49)
Roche Diagnostics
108 System Manual Version 10
cobas® 4800 System 6 Maintenance
Systeni Manuni Instrument maintenance
A Open the liquid waste cap. B Lift the liquid waste cap carefully
For the capacitive liquid level detection check the needies are picked-up again.
One channel to the next is checked for the proper functioning of the liquid level
detection.
1f the automatic tightness or liquid level detection checks fail, repeat twice, then calI
Roche Service.
11 The pipetting arm and the autoloader move to the right hand side and the end of
the daily maintenance wizard is displayed.
4) by
L-z:zz]I
Roche Diagnostics
System Manual Version 1.0 109
6 Maintenance cobas® 4800 System
1nstrumeHt ninintenance Systern Manual
Weekly maintenance
1f weekly maintenance is being performed, daily maintenance is not required on that day.
Icz.:zz4
5 Remove all carriers from the autoload tray and then choose the 0K button.
The instrument deck is unloaded automatically. The procedure continues with
the tightness check of the pipetting channels. The pipetting arm moves to the
right hand side to pick up the teaching needles.
Two checks are done with the pipetting channels, the over-pressure and the
under-pressure check.
r_
Roche Diagnostics
110 System Manual Version 1.0
.
cobas® 4800 System 6 Maintenance
Systenz Mannal Instrument maintenance
For the capacitive liquid level detection check the needies are picked-up again.
One channel to the next is checked for the proper functioning of the liquid level
detection.
1f the automatic tightness or liquid level detection checks fail, repeat twice, then cali
Roche Service.
Wcofdymofntenance - deckcloartlng
• Opeftfooo&ontherkok
6 Clean all carriers with decontamination liquid and leave them to dry.
7 Open the front cover and wipe the instrument deck with a cloth saturated with
decontamination liquid. The slide blocks the guides on the autoload tray for
-
WeekLymafntennc . Upwaste
l[z2Z4
1f a small waste bag is used access the tip waste bag by lifting the instrument front
cover. Remove a full waste bag from the instrument by removing the support
frame from the initialization/waste block.
Roche Diagnostics
System Manual Version 1.0
. 111
6 Maintenance cobas® 4800 System
Instrument mainlenance Systeni Manual
C D
When replacing the small waste bag, place the bottom portion of the waste bag
inside the support frame, fold the top over the frame and then place the support
frame into the initialization/waste block. Dispose of either vaste bag and contents
according to the appropriate local regulations.
Replacing large waste bag
1f a waste container is used remove the large ivaste bag from the waste container
and replace it with a new one. Close the large waste bag with a binder and dispose
of either waste bag and contents according to the appropriate local regulations.
• Waste chute and waste container can be reused and need to be replaced only if
required.
• 1f required, replace the waste chute in the same way as the small tip waste bag.
• 1f required, fold up a new waste container as indicated in Figure 3-12 Fo/ding up a
new waste box (p. 49)
Roche Diagnostics
112 System Manual Version 1.0
.
cobas® 4800 System 6 Maintenance
Systern Manual Instrument niaintenance
10 Remove the tip eject plate of the tip waste station (see Figure 6-5 Removing waste
bag or waste chute (p. 112)) and clean 11 with disinfectant solution.
Put the clean tip eject plate back in place properly. Make sure the tip eject plate
fits horizontally on the initialization/waste block. 1f it is not placed back properly,
this may cause a run to crash with instrument malfunction.
Roche Diagnostics
System Manual Version 1.0
.
113
6 Maintenance cobas® 4800 System
Instrument maintenance Systern Manual
11 Check the laser scanner window of the barcode reader on the autoloader and
clean it with a lint-free cloth or Q-tips lightly soaked in ethanol (70%) to prevent
unreliable barcode scanning.
12 Clean the front and side cover with a lint-free cloth soaked in disinfectant
solution and wipe dry.
13 Clean the autoload protecting ribbon with a lint-free cloth soaked in disinfectant
solution and wipe without exerting pressure.
14 Check the status of the liquid waste container. Empty the waste container if it
requires emptying. Dispose of the liquid vaste according to the appropriate local
regulations.
For additional details, see Liquid waste (p. 49)
A Open the liquid waste cap. B Lift the liquid waste cap carefully
The pipetting arm and the autoloader move to the right hand side and the end of
the weekly maintenance wizard is displayed.
WeeI,ly ,nintenfle -
rzJJ
16 Choose the 0K button.
Roche Diagnostics
114 System Manual Version 1.0
cobas® 4800 System 6 Maintenance
Systeni Manual Analyzer maintenance
Analyzer maintenance
No user maintenance is required for the analyzer. General cleaning of the analyzer
and exchanging the Xenon lamp, the ventilation dust filters, and the fuses are
des cribed below.
Electrical shock by electronic equipment
Never clean the analyzer without turning the analyzer power switch off and
WARNING
disconnecting the power cable.
Do not pour fluids into the interior of the analyzer or onto any parts of the analyzer.
Regular cleaning of the analyzer and the accessories is not required. 1f necessary,
clean the housing of the analyzer, the thermal block cycler, and the block cycler cover
with 70% ethanol.
Erroneous resuits due to dirty ventilation inlets
1f the ventilation inlets are not replaced regularly, the lense could get dirfy, and this could
WARNING cause an erroneous result.
Ensure that the ventilation dust filters are replaced twice a year.
Ventilation inlets The ventilation inlets of the analyzer should be checked regularly, to ensure an
unrestricted air flow. You should ensure that the ventilation dust filters are replaced
twice a year.
For details, see Exchanging the ventilation dust filters (p. 119).
Roche Diagnostics
System Manual Version 1.0
.
115
6 Maintenance cobas® 4800 System
Analyzer nsaintenance Systern Ivlanual
The analyzer automatically and continually measures the intensity of the Xenon lamp
to ensure optimal detection of the fluorescent signals during real time PCR. 1f the
lamp intensity falis below the minimum required intensity, a warning is issued by the
software that the Xenon lamp must be exchanged. The Xenon lamp is available as a
spare part from Roche. Please contact your local Roche representative for details.
Safety information
Only use the Xenon lamp available directly from Roche.
WARNING Never exchange the Xenon lamp without turning the analyzer power switch off and
disconnecting the power cable. Non-compliance poses the danger of electric shock
and damage to the eyes by the bright light of the lamp.
Before exchanging the Xenon lamp, make sure you have waited an appropriate period
of time (approximately 20 minutes) after you shut down the analyzer to allow the lamp
to cool. Directly after completion of a run, the lamp is hot enough to cause an
immediate burn.
In its cold state the lamp has a high internal pressure. During operation, the internal
pressure is higher than in the cold state. The lamp is extremely unlikely to explode but
the possibility cannot be entirely ruled out. Therefore, when handling the Xenon lamp
always use the safety cover supplied. When installing the lamp, remove the safety cover
and always take the following precautions: wear goggles and gloves and protect your
neck. Take the same precautions when removing the lamp.
Do not get finger marks, grease, paint, or similar materials on the bulb. Before using
the lamp, remove any such marks with isopropanol, ethanol, or any other suitable
agent that leaves no residues on the bulb.
The Xenon lamp does not contain any polluting material. The used lamp should be
stored in its safety cover before disposal. Where possible, the lamp should be disposed
of by a specialist waste management company. 1f this is not possible, put on protective
clothing, wrap the lamp completely in leather or thick cloth, smash the lamp, including
the discharge tube, with a suitable implement and dispose of the pieces.
Roche Diagnostics
116 System Manual Version 1.0
.
cobas® 4800 System 6 Maintenance
Systern Manual Analyzer nrnintenance
2 Move the cover of the analyzer to the right to access to the lamp unit.
------.<-
4.
1
1 cobas z 480
Roche Diagnostics
System Manual Version 1.0
. 117
6 Maintenance cobas® 4800 System
Analyzer maintenance Systeni ManHal
4 The Xenon lamp is fixed by a clip which is secured by a screw. Unscrew the clip
and open it.
-/I
k
.
0% -
,
cbas z 480
.1
5 Press the safety clip to lift the Xenon lamp out of its clamp and remove the two
conductors.
cobas z 480
Roche Diagnostics
118 System Manual Version 1.0
.
cobas® 4800 System 6 Maintenance
Systeni Manual Analyzer nzaintenance
-.\‘
:!!“ L._
oosbas z 480
1
A Conductor(-) B Conductor (+)
Figure 6-11 Remove the conductors
6 Instail the new Xenon lamp. First attach the (-) conductor and then the (+)
conductor.
7 Complete the installation by following the above mentioned steps 4 to 1 exactly in
the reverse order.
• Make sure the (-) conductor points upward while fixing the new Xenon lamp into
the clamp.
• After replacing the Xenon lamp, reset the lamp counter by choosing the Reset
values after lamp exchange button in the software. The analyzer monitors the
actual lamp intensity and lifetime. 1f either of the two expires, you will be informed
and prompted to exchange the lamp. No run can be performed unless the lamp has
been replaced.
0
Exchanging the ventilation dust filters
The electronic rack of the analyzer is cooled by ventilation. Two ventilation inlets are
located in the lower right corner of the right side of the analyzer (right beside the
block cycler compartment) and in the back of the analyzer. To avoid any
contamination of the analyzer’s interior by dust particles, these ventilation inlets
carry dust filters.
Roche Diagnostics
.
System Manual Version 1.0 119
6 Maintenance cohas® 4800 System
Analyzer maintenance Systern Manuat
Roche Diagnostics
120 System Manual Version 1.0
.
cobas® 4800 System 6 Maintenance
Systern Manual Analyzer maintenance
3 In the same way, remove the dust filter carrier from the rear ventilation inlet.
4 Remove the used dust filters from each carrier and insert new filters.
A Dust filter
5 Replace each dust filter carrier on the corresponding ventilation inlet and
reassemble the analyzer’s panel by following the above mentioned steps 3 to 1
exactly in the reverse order.
Roche Diagnostics
System Manual Version 1.0 121
6 Maintenance cobas® 4800 System
Analyzer maintenance System Manual
Exchanging fuses
The analyzer contains eight fuse types. Fuses must be exchanged when they are
blown.
Prirnaryfuses
Secondaryfuses
To access secondary fuses El — F5, remove the right panel of the analyzec
Roche Diagnostics
122 System Manual Version 1.0
cohas® 4800 System 6 Maintenance
Systeni Manual Analyzer niaintenance
Xenon lanipfuse
A Right side of Xenon lamp module, above the lamp’s mains socket
To access the Xenon lamp fuse, remove the cover of the analyzer.
0
To exchange the Xenon lamp fuse
1 Turn off the analyzer, then unplug the main cable.
2 Remove the cover of the analyzer.
3 The Xenon lamp fuse is located above the mains socket of the Xenon lamp
module.
Roche Diagnostics
System Manual Version 1.0
. 123
6 Maintenance cobas® 4800 System
Analyzer niaintenance Systern Manual
4 Using forceps, press the clamps of the left and right side of the fuse holder inward.
PuIl the fuse holder out of its chamber.
5 Exchange the blown fuse with a replacement fuse and place the fuse holder back
in the chamber.
6 Re-attach the instrument cover.
Roche Diagnostics
124 System Manual Version 1.0
.
Troubleshooting
7 Troubleshooting 127
cobas® 4800 System 7 Troubleshooting
Table of contents Systeni Manual
Troubleshooting
In this chapter alarm and troubleshooting information help to deal with exceptional
situations.
Roche Diagnostics
System Manual Version
. 1.0 . 127
7 Troubleshooting cobas® 4800 System
Table of contents System Manual
Roche Diagnostics
128 System Manual •Version 1.0
cobas® 4800 System 7 Troubleshooting
System Manual Overview
Overview
The system records all problems originating from the software, instrument and
analyzer. The error records can be viewed in the alarm area at the bottom of the
screen and in the Alarms and Messages vork areas.
UwU1sbm,M
Overyrew Resuits Y.l4ttri-p Ubirties
Aiarms ,,-,—;.r* Audrt Iran
Ø2!4.R, 4MUL: ..
A .
A Iiie alarm area displays the most recent alarms and warnings that are not yet confirmed by the
operator. Select an alarm in the list and dick the Alarm button on the left to get more details
about the selected alarm.
Alarm area The color of the alarm button changes depending on the severity of the alarms listed
in the alarm area.
loon Purpose
Red icon: the alarm area contains at least one error.
1 Yellow icon: the alarm area only contains warnings but no errors.
Roche Diagnostios
System Manual Version 1.0
. 129
7 Troubleshooting cobas® 4800 System
Oi’ervieiv Systeni Mawal
Message and alarm icons The severity of the alarms and messages is indicated by the following icons:
Icon Purpose
Designates an error message (red icon). The error must be resolved before
additional samples can be run. A problem report is generated automatically.
Ï?iiJ For details about problem reports, see Problen, reports (p. 136).
Designates a warning (yellow icon) indicating an error was detected but the
system cao still be run.
Roche Diagnostics
130 System Manual Version 1.0
.
cobas® 4800 System 7 Troubleshooting
Systeni Man ual Messages ivork area
Tab Description
Alarms Contains all alarms from the analyzer or the software that have
not yet been confirmed by the operator.
Messages Contains all information, error, and warning messages.
Audit trail Shows recorded user activities, e.g. operator log on and log off.
For more information on the audit trail, see Audit (rai!
(p. 81).
Table 7-3 Tabs in Messages work area
The purge and archive function only covers run data and resuits. Messages are kept in the
software until they are deleted manually. Therefore, it is recommended to manually delete
confirmed messages from time to time to free up additional space in the database.
U5o3b04n4n49,0
024,06w Resuits 2,-0vfl.fl
Mossago; Audit tra,I
A I’’”
S nornioon 24 41002012160153 5262026 5p4146404 1100S446S4L42042MFIF40d4644fl4IWW4445244
*e**r
23000-2612145514 6252421 p4,24,gon,nsa,a4a0Èa.tt.w4624u16446uz8gnn46u
A 4100
B —
E:I
A All unconfirmed alarms are listed in the Alarms tab.
B Alarm details. Select an alarm in the list to see the alarm details.
C Choose the Report button to manually generate a problem report.
D Choose the Confirm button to confirm the selected alarm. Confirmed alarms are deleted from
the Alarms tab, but are still listed in the Messages tab.
Roche Diagnostics
System Manual Version 1.0
. 131
7 Troubleshooting cobas® 4800 System
Ivlessages ivork area System Iianual
U5eD J),:
Overv:ew Resis
Adit til
T.
A Use the sort, filter, and group functions in the column headers to customize the list of
messages.
B Choose the Confirm button to confirm the selected message.
Field Description
Severity Contains an icon to indicate whether the message is an error,
warning, or information message.
Date/time Displays the date and time when the message was generated.
Code Displays the message code.
Text Displays an abbreviated description of the error or situation that
occurred.
Confirmation Displays date and time of confirmation.
date/time
Confirmed by Displays the operator who confirmed the message.
Table 7-4 Fields in Messages tab
Roche Diagnostics
132 System Manual Version 1.0
.
cobas® 4800 System 7 Troublesliooting
Systern Manual Messages work area
Roche Diagnostics
System Manual Version 1.0
.
133
7 Troubleshooting cobas® 4800 System
About resuitfiags Systeni Manual
Flags are automatically generated with resuits ifduring processing certain technical
checks were not passed, the result exceeds or does not reach predefined limits, or if
technical or mechanical problems occurred during a run.
Results that do not have flags associated with them can be considered valid. However,
not all results that have flags associated with them are invalid. Some flags are
warnings rather than error messages and do not necessarily invalidate the result(s).
Some flags are for information purposes only.
The flags may originate from either the analyzer, the software, or the result
interpretation.
Result icons In the Resuits work area, icons help you identify if a result failed, is invalid, or has a
flag.
Icon Comment
Result is invalid with one or more flags or failed.
To help you easily identify a positive result, some tests highlight positive results.
Roche Diagnostics
134 System Manual Version 1.0
.
cobas® 4800 System 7 Troubleshooting
Systern Manual About resuitfiags
E F
/ Mesogs
_s_
uti:t.j
A
-fr
1 r
B-’
IIU .
/ \
C
/ \ D
Each flag is identified by its flag code, and information on the nature of the reason
for each flag is displayed.
For a list of result flags, refer to the test-specific Operator’s Manual.
Roche Diagnostics
System Manual Version 1.0
. 135
7 Troubleshooting cobas® 4800 System
General troubleshooting Systern Manual
General troubleshooting
In this section, you find information about troubleshooting aids such as problem
reports and remote access.
Problem reports
A problem report is created by the system after every run, or when a new error
message is generated. 1f necessary, a problem report can also be created manually.
Problem reports are used by Roche Service to assist in troubleshooting cases.
Problem reports are stored as ZIP-files in a predefined folder where they can be
retrieved and sent to Roche Service when required.
lrhen the maximum number of problem reports is reached and another problem
report is added, the oldest problem report will automatically be deleted from the
folder. The maximum number of problem reports is defined under Utilities
> Configuration> System settings.
0
To manually create a problem report
1 In the software, choose Messages > Alarms.
0
Remote support
In a remote support session, the software is remotely operated by Roche Service. Call
Roche Service to initiate a remote support session.
0
LIS troubleshooting
1f you use LIS and are having problems, this will help you understand LIS statuses.
1f the connection to the LIS does not work, contact the local IT support to find out if
there is a problem with the LIS. When the problem has been solved, disable and
enable the LIS under Utilities > Configuration> System settings.
The System overview tab displays the status of the LIS connection.
Screen Status Comment
Not available LIS has been disabled under
Ab,tyS4s
Utilities> Configuration>
LI ts System settings.
Trarewr5j
4
Idie The connection to the LIS is
reay.
d
lfJ For information about the LIS host protocols, refer to the cobas 4800 System Host
Interface Manual.
1f the LIS connection does not work and a run must be processed, you can disable LIS and
use the sample editor to manually create the work orders.
Roche Diagnostics
System Manual Version 1.0
. 137
7 Troubleshooting cobas® 4800 System
Hardware troubleshooting Systent Manual
Hardware troubleshooting
Instrument
The reason for this error may be poor-quality, damaged, or missing barcode labels.
There are several options for handling barcode errors.
E! Exactly follow the procedures outlined below to resolve barcode reading errors. Deviating
from this procedure might abort the run.
Roche Diagnostics
138 System Manual Version 10
cobas® 4800 System 7 Troubleshooting
Systern Manual Hardware troubleshooting
1f the reagents have been loaded into the instrument, choose the Unload
button and wait until the carrier is unloaded. Choose and manually enter
or scan the barcode with the barcode reader. To bad the reagents back into
the instrument, choose the Reload button.
1f the connection between the software and the instrument is broken or if the instrument
was switched off the run status may not be displayed correctly until the connection has
been re-established.
1f the connection between the control unit and the instrument is interrupted (i.e. the
USB connector is disconnected), the instrument will immediately abort the run. The
status of the instrument changes to Error, however the status of the run in the
software will only be updated after the connection is re-established. The same is true
for the display of the associated error message in the alarm area. An error message
will only be issued after the connection to the instrument is re-established.
After a connection error the instrument has to be unloaded manually.
For details, see Unloading the instrument deck nianually (p. 140)
Roche Diagnostics
System Manual Version 1.0
. 139
7 Troubleshooting cobas® 4800 System
Hardware troubleshooting System Manual
Run recovery
1f the run control is lost by the software an automatic run recovery process is started.
Depending at what time point in a run the run control was lost the recovery process
continues the run or aborts it. Run control can be lost for reasons like power drops,
shutdown of the control unit, or connection or hardware problems.
Results of such runs are flagged with flag M6 (analyzer) or MiO (instrument). The
flag only indicates that run recovery was initiated but does not indicate if the run
could be recovered by the system or not. To decide if the result is reportable analyze
the result itself and all associated flags as you would do with any result.
Remote support
In a remote support session, the software is remotely operated by Roche Service. Call
Roche Service to initiate a remote support session.
0
Unloading the instrument deck manually
All carriers have to be manually unloaded from the instrument deck ifa run was
aborted and the carriers are not automatically unloaded.
Safety information
Take adequate precautions if the instrument deck has to be unloaded manually. The
instrument deck or the carriers could be contaminated.
Make sure that you have read and understood the chapter General safety information
(p. 13). The following safety messages in particular are relevant:
Warning messages:
• Infection by samples and associated materials (p. 19)
• Infection and injury due to sharp objects (p. 19)
Caution messages:
• Skin inflammation orinjuly caused by reagents (p. 21)
Roche Diagnostics
140 System Manual Version 1.0
.
cobas® 4800 System 7 Troubleshooting
Systeni Manual Hardware troubleshooting
3 Clear all obstructions and remove carefully all carriers from the instrument deck.
Be careful not to break the stop hooks. Unloading carriers manually causes stop
hooks to break easily.
Call Roche Service ifthere are any obstructions that can not be cleared.
4 Clean the instrument deck.
5 Switch on the instrument.
Check before switching on the instrument that no obstructions are left on the
instrument deck.
6 Perform daily maintenance.
Do not manually remove tips from the pipetting channels. Cail Roche Service if
there are stili tips on the pipetting channels after daily maintenance.
For details, see Daily maintenance (p. 105)
0
Take out of operation
Perform the following take out of operation procedure if the system is not used for a
longer period of time.
2 Switch off the instrument and unplug the power cord and the USB cable from the
instrument.
3 Switch off the heater / shaker unit and unplug the power cord from the heater /
shaker unit.
4 Log off the software, switch off the control unit, and unplug the power cord from
the control unit.
0
Analyzer
Roche Diagnostics
System Manual Version 1.0
. 141
7 Troubleshooting cobas® 4800 System
Hardware troubleshooting Systeni Manual
0
Red status LEDs
For error messages and flags, refer to the test-specific Operator’s Manual.
Roche Diagnostics
142 System Manual Verson 1.0
.
Glossary
cobas® 4800 System
Systern Manual Arnp4fïcation Saniple carrier
-
Glossary
The glossary explains key terms used in this 115 Laboratnry information system. A computerized
documentation. system for entering, managing and reporting laboratory
information. Information includes but is not limited to
Amplification The process of producing many DNA patient demographics, test orders, and test results.
copies from one original DNA or RNA target region.
Loading deck The external shelfat the front of the
PCR is a nucleic acid amplification technique.
instrument. It provides a surface onto which carriers can
Anneal The biochemical process of hybridizing or be placed before the autoloader draws them into the
binding two segments of complementary nucleic acid. instrument.
Autoloader The hardware assembly on the instrument Magnetic glass particles Used in combination with
that enables automatic loading of carriers placed on the other reagents to bind nucleic acids, allowing separation
loading deck. It consists of a loading head movable in the of impurities during washing.
Y direction, which draws the items into the instrument
MGP see Magnetic Glass Particles.
and can read the barcodes on them.
Microweil plate Plate used for amplification and
Control unit A personal computer that runs the
detection. The 96-well plate is barcoded and has to be
Microsoft Windows operating system and the software.
sealed with a special sealing film before inserting into the
Controls Reagent formulated to produce known resuits analyzer.
that are processed like samples. The software monitors
PCR Polymerase Chain Reaction. The in vitro process
control resuits. Each run requires a positive and a
used to amplify short specific target nucleic acid
negative control.
sequences. PCR is performed by cycling the temperature
Ct value The amplification/detection cycle where the of the amplification mixture according to a set profile.
growth curve crosses a specific fluorescence level and the The profile generally consists of denaturation and
growth rate can be observed. annealing.
Deepweli plate Plate used for extraction on the Pipetting channel A hardware assembly mounted on
instrument. the pipetting arm. It can pick up and eject tips, aspirate
and dispense liquid, and detect liquid level in tubes.
Denaturation The process of separating double
There are 8 pipetting channels working in parallel.
stranded DNA into single strands by breaking the
hydrogen bonds. Pipetting tip A disposable tip for pipetting. The tips
used by the instrument are capable of transferring up to
Detection Obtaining measurements to determine
1000 microliters of liquid per pipetting operation. The
whether a sample is reactive for the target analyte.
tip includes a buik-in filter that prevents the possibility
Fluorescence measurements are made at selected
of liquid inside the tip entering the pipetting channel
temperatures and times during the amplification process.
hardware.
When the run is complete data are analyzed to determine
the presence of the amplified products from the target Plate carrier Carrier for deepwell plate and microweil
and internal control nucleic acid sequences. plate on the instrument.
DNA Deoxyribonucleic Acid (DNA) is the genetic Reagent reservoir The reagent reservoirs hold the
material that is passed from parent to progeny and reagents used for sample preparation. The reagent
propagates the characteristics of the species in the form reservoirs are barcoded and are filled manually by the
of genes it contains and the proteins for which it codes. operator (scan-scan-pour-place principle) for each run.
DNA contains the following four nucleotides: dATP, Reagent reservoirs are available in two sizes: 200 mL and
dCTP, dTTP, and dGTP. 50 mL.
Instrument deck The work surface of the instrument, Sample carrier Used to bad unprocessed samples
where pipetting takes place. It is organized as a left-to onto and to unload processed samples from the
right series of tracks, into which the autoloader draws instrument.
carriers.
LAN Local Area Network. The system control unit is
con nected to the instrument and the analyzer.
Roche Diagnostics
System Manual Version 1.0
. 145
cobas® 4800 System
Scan-scan-pour-placeprinciple XML
- Systern Manual
Roche Diagnostics
146 System Manual Version 1.0
.
Index
cobas® 4800 System Index
Systern Manual
Index
A — overview, 35
—x480, 58,64
abbreviations, 8 —z480, 68
alarms contact addresses, 3
— confirming, 132 context menus, 77
—tab, 131 control unit, 69
— viewing, 132 controls, specifications, 62
approvals, 2 conventions used in manual, 7
archiving copyrights, 2
— parameters, 91 covers, x 480, 40
— settings, 90 creating
—tab, 90 — problem report, 136
audit trail, 81 cross contamination, 22
autofilter, 78
autoload unit, 43
D
evaporation of samples, 21 1
exchanging
—fusesonz480, 122 labels, 25
—ventilation dust filters on z 480, 119 LEDs
—xenon lamp on z 480, 116 —x480, 41
explosion, safety, 20 —z480, 67
exporting LIS, 34
— support data, 97 LIS statuses, 137
lists
— advanced filtering, 79
F
— filtering, 79
— organizing, 76
filtering information
— sorting, 77
— advanced features, 79
— use, 76
— autofilter, 78
— drop-down list, 78
filters, for dust, 119 M
fire, safety, 20
flags magnet plate, 57
— displaying, 134 maintenance
fluid waste, disposing, 108, 114 —daily for x 480, 105
fuses —weeklyforx 480, 110
— exchanging on z 480, 122 —x480, 104
— safety, 24 —z480, 115
managing
—passwords, 92
G
—user accounts, 93
Manual
general cleaning
—version, 2
—x480, 105
measurements
—z480, 115
—x480, 62
global action buttons, 75
—z480, 68
messages, 131
H —work area, 131
microwell plates, 60
hardware, 37
—x480, 39
0
—z480, 65
heater unit, 56
operating con ditions, 16
heater/shaker unit
operator qualifications, 16
— specifications, 64
optical safety, 18
hiding columns, 77
overview
—cobas 4800 user interface, 73
—cobas x 480, 32, 39
—cobasz480, 33,65
installation, 16 —control unit, 34
—x480, 62 —system, 31
instrument approvals, 2
instrument deck, x 480, 41
intended use, 2
p
interfaces
passwords
—x480, 63
— changing, 92
—z480, 68
—managing, 92
ISWAP, 46
—rules, 96
physical dimensions, 67
—x480, 62
Roche Diagnostics
150 System Manual Version 1.0
.
cobas® 4800 System Index
Systern Manual
—z480, 67 — biohazards, 19
pipetting arm, x 480, 45 — carryover, 22
plate carriers, 55 — data security, 23
plates — disposal of analyzer, 27
— deepweil, 59 — disposal of instrument, 27
— microweli, 60 — electrical, 18
power requirements — electromagnetic fields, 17
—x480, 62 — explosion, 20
—z480, 67 — fire, 20
precautions, 16 — fuses, 24
principles — hot surfaces, 22
—x480, 62 — incorrect results, 22
—z480, 68 — labels, 25
problem reports, 136 — labels on the system, 25
product names, 9 — moving parts, 20
protective equipment, 16 — operating conditions, 16
— optical, 18
— power failure, 17
Q —precautions, 16
— result accuracy, 16
qualifications, 16
—samples, 21
—use, 16
R —waste, 19
safety classification, 15
reading errors, 138 sample carriers, x 480, 50
reagent carriers, 53 samples
reagent reservoirs, 58 —safety, 21
reagents — specifications, 62
— carriers, 53 sealing film, 60
remote support, 136, 140 settings, for archiving, 90
replacing shaker unit, 56
—fusesonz480, 122 software version, 2
— ventilation dust filters on z 480, 119 sorting list entries, 77
—waste bags, 107, 111 stationary carrier, 56
—xenon lamp on z480, 116 support data, 97
reporting, 88 symbols, 7
reports system
— for problems, 136 — settings cobas 4800, 87
result counters, 98
result flags, see flags
T
result icons, 134
resuits
tabs
—accuracy, 16
— Alarms, 131
retrieving
— Audit trail, 81
— problem report, 136
— Configuration, 87
reviewing
— Export support data, 97
— alarms, 132
— in user interface, 74
revision history, 2
— Messages, 132
rules
— Users, 93
for passwords, 96
Workpiace, 74
—
— disabling, 89 —LEDs, 41
— enabling, 89 — maintenance, 104
third-party software, 17 — manual unloading, 140
tip rack carriers, 55 — pipetting arm, 45
track positions, x 480, 42 — sample carriers, 50
trademarks, 2 — teaching needies, 46
— technical specifications, 61
track pos itions, 42
u —
— waste container, 48
— waste station, 47
unloading
xenon lamp, exchanging on z 480, 116
— x 480 manually, 140
unlocking user accounts, 95
use of manual, 7 z
user accounts
—changing, 95 z480, 65
— defining, 94 —block cycler unit, 65
— deleting, 95 — detection unit, 65
— management, 93 — LEDs, 67
—rules, 96 — technical specifications, 67
— unlocking, 95
user interface
— colors, 74
—lists, 76
—tabs, 74
user management, 92
w
waste
—consumables, 61
—emptying fluids, 108, 114
—replacing bags, 107, 111
—safety, 19
—x480, 63
waste bags, replacing, 107, 111
waste container, x 480, 48
waste station, x 480, 47
weekly maintenance, for x 480, 110
wizards, 80
x
x480, 39
—autoload unit, 43
—carriers, 50
—consumables, 58
—covers, 40
—heater/shaker unit, 56
—instrument deck, 41
—ISWAP, 46
Roche Diagnosties
152 System Manual Version 1.0
.
Revisions