Sie sind auf Seite 1von 1

MEDICON GLUCOSE (UV) Reagent AU480/AU680 Reagent ID: 721

Product Code: 1418-0017 Serum / Plasma Application


Specific Test Parameters
General LIH ISE HbA1c Calculated Tests Range

Test Name: GLU ▼ ◄ ► Type: Serum ▼ Operation Yes ▼

Sample Volume 1.6 uL Dilution 0 uL OD Limit


Pre-Dilution Rate 1 ▼ Min OD Max OD
Rgt. Volume R1 (R1-1) 72 uL Dilution 64 uL Reagent OD limit
First Low -2.0000 High 2.5000
Last Low -2.0000 High 2.5000
R2 (R2-1) 24 uL Dilution 0 uL
Dynamic Range Low 0 High 600
Common Rgt. Type NoneΦ Name Φ Correlation Factor A 1.0000 B 0.0000
Wavelength Pri 340 ▼nm Sec 380 ▼nm Factor for Maker A B
Method END ▼
Reaction Slope + ▼ Onboard Stability Period 30 Day Hour
Measuring Point 1 First 0 Last 27 LIH Influence Check # ▼
Measuring Point 2 First 0 Last 10 Lipemia ▼
Linearity Limit % Icterus ▼
Lag Time Check ▼ Hemolysis ▼

Parameters Specific Test Parameters


General LIH ISE HbA1c Calculated Tests Range

Test Name: GLU ▼ ◄ ► Type: Serum ▼

Value/flag # ▼ Low High


Level Panic Value
Specific Ranges: From To Low High
Sex YearMonth Year Month L H # #
❑ 1. # ▼ # # # # # #
❑ 2. # ▼ # # # # # #
❑ 3. # ▼ # # # # # #
❑ 4. # ▼ # # # # # #
❑ 5. # ▼ # # # # # #
❑ 6. # ▼ # # # # # #
7. No Demographics # #
8. Not Within Expected Values
Unit mg/dL Decimal Places #

Parameters Calibration Parameters


Calibrators Calibration Specific STAT Table Calibration
General ISE

Test Name: GLU ▼ ◄ ► Type: Serum ▼ ❑Use Serum Cal.

Calibration Type: AB ▼ FORMULA: Y= AX+B ▼ Counts: # ▼


<Calibration Parameters> Factor Range
Calibrator OD Conc. Low High Slope Check None ▼
Point 1 # ▼ + # #
Point 2 ▼ Allowance Range Check
Point 3 ▼
Point 4 ▼ ❑ Reagent Blank
Point 5 ▼ ❑ Calibration
Point 6 ▼
Point 7 ▼ Advanced Calibration
Point 8 ▼ Operation ▼
Point 9 ▼
Point 10 ▼ Interval (RB/ACAL) ▼

<Point Cal. For No. of Correction Points ▼ Use Master Curve ▼ ❑ Lot Calibration
Master Curve> OD Range
Calibrator OD Conc. Low High Stability
Point-1 ▼ Reagent Blank 30 Day Hour
Point-2 ▼ Calibration 30 Day Hour

MB Type Factor 1-Point Calibration Point ▼ ❑ with Conc-0

+ MEDICON MEDI-CAL (p.c.: 1578-0891) or OLYMPUS System Calibrator Cat.No: 66300 (p.c.: 4478-0891).
# User defined.
Values set at mg/dL. To work in SI units (mmol/L) divide by 18.
Φ AU680

Das könnte Ihnen auch gefallen