Sie sind auf Seite 1von 11

8-Part Eye Examination BLOCK 20

MODULE 01
Ophthalmologists

 In using the snellen’s chart, place the patient 20 feet away


SUMMARY/OUTLINE from the chart.
I. Distance Acuity Test  Remember to test each eye separately and be tested with
II. External Exam and without corrective lenses
III. Pupillary Exam  But in some clinics, there is only limited space, so they
IV. Motility Exam minimize the distance by converting them to smaller
V. Visual Fields (Confrontation Test) distances.
VI. Tonometry  The largest letter E is equivalent to 20/200. If the patient
VII. Ophthalmoscopy cannot read the letter E, move the patient 5 feet closer
VIII. Slit-lamp Biomicroscopy until the patient reaches 5 feet away. Note the distance
where the patient is able to read the letter E. If the patient
still cannot read at 5 feet, proceed to counting finger test.
DISTANCE VISUAL ACUITY
 Some clinics may use meters. Patients usually placed 6
1. Ask the patient to stand or sit at a designated testing
meters away from the chart.
distance (20 feet)
2. Occlude the left eye (testing one eye at a time) Visual acuity is scored as a fraction (eg, “20/40”). The first
3. Ask the patient to identify each letter in the chart, on number represents the testing distance between the chart
the lines of successively smaller optotypes, until patient and the patient, and the second number represents the
correctly identifies only half the optotypes on a line smallest row of letters that the patient’s eye can read. Hence
4. Note the corresponding acuity measurement shown at normal vision is 20/20 and 20/60 acuity indicates that the
the line of the chart patient’s eye can only read from 20 feet letters large enough
5. Repeat above steps for the left eye, with the right eye for a normal eye to read from 60 feet.
covered Vaugn Asbury’s General Ophthalmology 39th Edition
6. Retest acuity with the patients with low vision, e.g.
counting fingers, hand motion, light perception, etc.  For pediatric patients, you can use figures or the tumbling
E charts
Visual acuity can be tested either for distance or near,
conventionally at 20 feet (6 meters) and 14 inches (33 cms)
away, respectively, but distance acuity is the general standard
for comparison. For diagnostic purposes visual acuity is
always tested separately for each eye, whereas binocular
visual acuity is useful for assessing functional vision, such as
for assessing the eligibility to drive.
Vaugn Asbury’s General Ophthalmology 39th Edition

SNELLEN’S CHART
 Other patients can use the Jaeger chart. the patient is
scored depending on which line of sentences he can read.
J10 is the biggest and J1 is the normal acuity.

 Distance visual acuity test should be the first thing to do


during the 8 part eye exam. You should do this before
palpating, or putting any eye medication or drugs so that
visual acuity is not altered because of the prior tests
administered.

Page 1 of 11
 Example result  This is performed before studying the eye under
magnification
With Without  Gross inspection and palpation: lesions, growth,
correction correction inflammatory signs (swelling, erythema, warmth,
FAR SIGHT tenderness)
OS 20/150 20/20  Check for the ff:
OD 20/300 20/20 o position of eyelids (ptosis, lid retraction)
NEAR SIGHT o asymmetry can be quantified by measuring the
OS J8 J1 central width (in mm) of palpebral fissure (space bet
OD J10 J1 lower and upper lid margins)
o abnormal motor fxn of the lids (upper lid elevation,
COUNTING FINGERS forceful lid closure) may be due to neurologic or
 By convention, ophthalmologists test first the right eye. If primary muscular abnormalities
it is known that the other eye is “buron” maybe the patient o malposition of the globe (proptosis) that may occur
complained of it before assessment, check the “buron” eye in orbital disease
first. o bony orbital rim and periocular soft tissue
 In counting fingers, place your finger 1 ft away from the  General facial evaluation:
patient and let the patient count the fingers you are o enlarged preauricular LN, sinus tenderness, temporal
showing to him. If the patient cannot count at 1 feet away, artery prominence, skin/mucous membrane
move 1 more feet away until you reach the maximum of 3 abnormalities
feet. Record the distance where the patient can count the o You canauscultate for bruit at the temporal side of
fingers shown to him. the orbit or directly at the globe. This can help you
 After the maximum of 3 feet and the patient cannot count detect carotid sinus fistula
the fingers, proceed to light perception test.

HAND MOTION TEST PUPILLARY EXAMINATION


 Make sure that you move your hands against the light. 1. Turn off the light to decrease the room illumination
2. Ask the patient to maintain fixation on a distance target
LIGHT PERCEPTION TEST 3. Shine a bright handheld light directly into the right eye
 Make sure to turn off the lights by approaching it from the side or from below
4. Record the direct pupillary response to light in the right
The patient unable to read the largest (“20/200”) letter on a
eye in terms of briskness of the response; observe the
Snellen chart should be moved closer to the chart until that
consensual reflex by noting the response to light of the
letter can be read. The distance from the chart is then
non-illuminated pupil
recorded as the first number. Visual acuity of “5/200” means
that the patient can identify correctly the largest letter from a 5. Repeat above steps for the left eye
distance of 5 feet but not further away. An eye unable to read 6. Enumerate the steps in performing the swinging flash
any letters is tested by the ability to count fingers. “CF at 2 ft” light test and explain the clinical significance of relative
indicates that the eye was able to count fingers held 2 feet afferent papillary defect (Marcus Gunn pupil)
away but not farther away. If counting fingers is not possible,
the eye may be able to detect a hand moving vertically or Assessment of pupil function should be done before any
horizontally (“HM,” or “hand motions” vision). The next lower drops are instilled in the eye and before the cornea is
level of vision would be the ability to perceive light (“LP,” or touched (eg, applanation tension or Schirmer test).
“light perception”). An eye that is totally blind is recorded as
having no light perception (“NLP”). Examination of the pupils with a light stimulus provides
Vaugn Asbury’s General Ophthalmology 39th Edition evidence of the health of both the afferent and efferent
systems. In addition to light, the pupils also respond to
accommodation and convergence for clear and single vision
EXTERNAL EXAMINATION at near. The pupils will constrict equally when either
1. Observe the facial skin for any dermal or vascular accommodation or convergence is stimulated by a near
changes; note any lesions or evidence of trauma object. When all 3 actions—accommodation, convergence,
2. Note any significant asymmetry of facial bones and miosis—occur simultaneously, this is called the synkinetic
3. Note the lid position; assess effectiveness of eyelid near response.
closure and strength of the orbicularis muscles if
appropriate Pupil function is evaluated with a bright penlight or other
intense, small light in a dimly illuminated room. Pupil or iris
4. Palpate the bony orbit for any lesion or deformity
abnormalities found with the naked eye can then be more
thoroughly evaluated using the biomicroscope. The pupils are

Page 2 of 11
evaluated for size, shape, direct light response, consensual stimulation of the right eye. As the light is swung from the
response, and near response. right to the left eye, both pupils will begin to dilate normally
SIZE as the light is moved away from the right eye and then not
In dim illumination, the average pupil diameter is 3 or 4 mm. constrict or paradoxically widen as the light is shone into the
This is ascertained by shining the light from below the left eye (since the direct response in the left eye and
patient’s nose so that the pupils are just visible to the the consensual response in the right eye are reduced
examiner; the light is not shone directly into the patient’s compared to the consensual response in the left eye and
eye. Pupils smaller than 2 mm are said to be miotic; pupils direct response in the right eye from stimulation of the right
larger than 6 mm are mydriatic. Miotic pupils may be caused eye). When the light is swung back to the right eye, both
by antiglaucoma medications, chronic iris inflammation, age, pupils will begin to dilate as the light is moved away from the
or a neurologic disorder. Abnormal mydriasis is caused by left eye and then constrict normally as the light is shone into
certain drugs, neurologic disorders, iris injury, or acute the right eye. This phenomenon is called a relative afferent
glaucoma. The pupils should be equal in size, although a pupillary defect (RAPD).
small difference (1 mm) may be a normal variation. If they are Vaugn Asbury’s General Ophthalmology 39th Edition
unequal (anisocoria), the difference between them should be
further evaluated in both dark and bright room illumination.

SHAPE
Both pupils should be round. The pupils are normally
centered or a little nasal in the iris. An eccentric pupil may be
the result of faulty embryonic development, injury,
intraocular surgery, or inflammation. In addition to being
eccentric, a pupil may also have an unusual shape.

DIRECT LIGHT RESPONSE


In dim room illumination, the patient is instructed to look at a
distant target (this prevents the pupillary response to a near
stimulus). The light source is presented to each eye
separately and slightly off center to avoid the near response.
Each pupil should exhibit a brisk response and constrict to MOTILITY EXAMINATION
about 2 mm. 1. Sit facing the patient. Hold finger on or small fixation
target at eye level about 10-14 inches in front of the
CONSENSUAL RESPONSE patient, with the patient looking straight ahead.
The consensual response is the simultaneous and equal 2. Ask the patient to follow the target as you move it into
response of one pupil when the other pupil is being the six cardinal fields and up and down along the midline.
stimulated by direct illumination or a near target. If the Elevate the upper eyelid with a finger on your free to
stimulated pupil constricts normally, then the consensual observe downgaze
response of the other pupil will produce equal constriction 3. Note whether the amplitude of eye movement is normal
without direct light stimulus. or abnormal in both eyes
4. Note any nystagmus that may be present
The pupils should be symmetric, and each one should be 5. Determine alignment using the Hirschberg method of
examined for size, shape (circular or irregular), and reactivity corneal light reflection test – hold a penlight in front of
to both light and accommodation. Pupillary abnormalities the patient eyes at a distance of approximately 2 feet,
may be due to (1) neurologic disease, (2) intraocular directing the light at the midpoint between the two eyes
inflammation causing either spasm of the pupillary sphincter of the patient; instruct the patient to look directly at the
or adhesions of the iris to the lens (posterior synechiae), (3) light; compare the position of the two corneal light
markedly elevated intraocular pressure causing atony of the reflections and record the estimated result
pupillary sphincter, (4) prior surgical alteration, (5) the effect
of systemic or eye medications, and (6) benign variations of The extraocular muscles include: the medial, inferior, and
normal. superior recti, the inferior oblique, and levator palpebrae
muscles, all innervated by the oculomotor nerve (III); the
SWINGING PENLIGHT TEST FOR MARCUS GUNN PUPIL superior oblique muscle, innervated by the trochlear nerve
As a light is swung back and forth in front of the two pupils, (IV); and the lateral rectus muscle, innervated by the
one can compare the reactions to stimulation of each eye, abducens nerve (VI).
which should be equal. If the neural response to stimulation
of the left eye is impaired, the pupil response in both eyes The precise action of any muscle depends on the orientation
will be reduced on stimulation of the left eye compared to of the eye in the orbit and the influence of the orbital

Page 3 of 11
connective tissues, which regulates the direction of action
of the extraocular muscles by acting as their functional
mechanical origins (the active pulley hypothesis).

Landmarks to remember:
Mnemonic: "SO-4, LR-6, All the rest 3" (ie Superior Oblique by
CN 4, Lateral rectus by CN 6, and all the other EOMs by CN 3). 0mm Light reflection is at the center of the pupil
1mm Light reflection is in between the center of the pupil
TESTING OF THE VISUAL FIELD and the pupillary border
The patient is asked to follow a target with both eyes as it is 2mm Light reflection is at the pupillary border
moved in each of the four cardinal directions of gaze. The 3mm Light reflection is in between pupillary border and
examiner notes the speed, smoothness, range, and symmetry limbus
of movements and observes for unsteadiness of fixation (eg, 4mm Light reflection is at the limbus
nystagmus). Impairment of eye movements can be due to
neurologic problems (eg, cranial nerve palsy), primary KIMPSKY TEST
extraocular muscular weakness (eg, myasthenia gravis), or
mechanical constraints within the orbit limiting rotation of
the globe (eg, orbital floor fracture with entrapment of the
inferior rectus muscle). Deviation of ocular alignment that is
the same amount in all directions of gaze is called “comitant.”
It is “incomitant” if the amount of deviation varies with the
direction of gaze.
Vaugn Asbury’s General Ophthalmology 39th Edition

SIMPLE TEST OF BINOCULAR ALIGNMENT/ HIRSCHBERG


METHOD

Procedure:
1. Have the patient look toward a penlight held several feet
away. (33 cms according to Vaughan & Asbury’s 18th ed.
Page 244)
2. Note for the pinpoint light reflection, or “reflex,”.
The Krimsky test is essentially the Hirschberg test, but with
Note: In normal eyes, pinpoint light reflection, or “reflex,”
prisms employed to quantitate deviation of ocular
should appear on each cornea and should be centered over
misalignment by determining how much prism is required to
each pupil if the two eyes are straight in their alignment.
centre the reflex [2] The Krimsky test is advisably used for
patients with tropias, but not with phorias.
If the eye positions are convergent, such that one eye points
https://en.wikipedia.org/wiki/Hirschberg_test
inward (“esotropia”), the light reflex will appear temporal to
the pupil in that eye. If the eyes are divergent, such that one COVER TEST
eye points outward (“exotropia”), the light reflex will be - More accurate method of verifying normal ocular
located more nasally in that eye. alignment.The test requires good vision in both eyes.
Vaugn Asbury’s General Ophthalmology 39th Edition

Procedure:
1. Ask the patient to gaze at a distant target with both
eyes open. If both eyes are fixating together on the

Page 4 of 11
target, covering one eye should not affect the
position or continued fixation of the other eye.
2. Suddenly covers one eye and carefully watches to see
that the second eye does not move (indicating that it
was fixating on the same target already). If the
second eye was not identically aligned but was
instead turned abnormally inward or outward, it
could not have been simultaneously fixating on the
target. Thus, it will have to quickly move to find the
target once the previously fixating eye is covered.
Fixation of each eye is tested in turn. BRUCKNER TEST

Note!
An abnormal cover test is expected in patients with diplopia.
However, diplopia is not always present in many patients with
long-standing ocular misalignment. When the test is
abnormal, prism lenses of different power can be used to
neutralize the refixation movement of the misaligned eye
(prism cover test). In this way, the amount of eye deviation
can be quantified based on the amount of
prism power needed.

The Brückner test is a qualitative assessment of ocular


alignment. This test is done under dark room illumination
and the direct ophthalmoscope aperture set on the largest
aperture setting so as to equally illuminate both eyes. The
examiner is viewing at about 1 meter away and observing the
relative brightness of the fundus reflex from each eye. A
whiter and brighter reflex is noted in the eye that is
strabismic. To confirm a difference in color, retest
monocularly to note any changes to the reflex. The
strabismic eye will appear whiter and brighter as a result of
the fundus reflection emanating from outside of the
pigmented macula region.
http://apps.ketchum.edu/ceonline/courseview.a...

VISUAL FIELDS EXAMINATION (CONFRONTATION TEST)


1. Seat the patient and make sure the eye not being testes
is occluded
2. Seat facing the patient at a distance of about 1m. close
your eye that is directly opposite the patient’s occluded
eye
3. Ask the patient to fixate on your nose or on your open
eye
4. Hold your hands stationary midway between yourself
and the patient is opposite quadrants about 30 degrees
from central fixation
5. Quickly extend then retract a finger or fingers on one
hand in one quadrant of the monocular field asking the
patient to state the number of fingers seen
6. Repeat all four quadrants, testing at least twice per

Page 5 of 11
quadrant TONOMETRY
7. Diagram the confrontation field if an abnormality is 1. Enumerate and differentiate the methods of measuring
detected intraocular pressure

 Make sure to test each eye separately. When occluding  globe can be thought of as an enclosed compartment
the other eye, don’t depress the eyeball because it may through which there is a constant circulation of aqueous
cause blurring of vision. humor
 During the 8part eye exam, we are expected to use only  this fluid maintains the shape and a relatively uniform
the manual test. pressure within the globe
 Automated perimetry is done in sophisticated diagnostic  tonometry is the method of measuring intraocular
centers. Is uses a machine that accurately records the pressure using calibrated instruments.
visual fields. The machine can also determine how much  normal range is 10 to 21 mm Hg
light the patient can see by altering the intensity of the  less corneal indentation is produced as intraocular
light being shown to the visual fields. pressure rises.
 It is advisable to make a diagram as to which visual field  since both methods employ devices that touch the
is blinded because it can guide you in diagnosis patient’s cornea, they require topical anesthetic and
disinfection of the instrument tip prior to use.
 with any method of tonometry, care must be taken to
avoid pressing on the globe and artificially increasing its
pressure.

APPLANATION TONOMETRY
- intraocular pressure is determined by the force required to
flatten the cornea by a standard amount. The force required
increases with intraocular pressures.
- the GOLDMANN APPLANATION TONOMETER is attached to
the slitlamp and measures the amount of force required to
flatten the corneal apex by a standard amount.
- the higher the intraocular pressure, the greater the force
required.
 If the patient has bitemporal hemianopia (number 2)
- Goldmann applanation tonometer is a more accurate
what possible disease can lead to such problem knowing
method than Schiotz tonometry
that the optic chiasm is the structure affected? Pituitary
- following topical anesthesia and instillation of fluorescein,
tumors.
the patient is positioned at the slitlamp and the tonometer is
 In automated perimetry:
swung into place. To visualize the fluorescein, the cobalt blue
filter is used with the brightest illumination setting. After
grossly aligning the tonometer in front of the cornea, the
examiner looks through the slitlamp ocular just as the tip
contacts the cornea. A manually controlled counterbalanced
spring varies the force applied by the tonometer tip.Upon
contact, the tonometer tip flattens the central cornea and
produces a thin circular outline of fluorescein. A prism in the
tip visually splits this circle into two semicircles that appear
green while viewed through the slitlamp oculars. The
tonometer force is adjusted manually until the two
semicircles just overlap, as shown in Figure 2–10. This visual
end point indicates that the cornea has been flattened by the
set standard amount. The amount of force required to do this
is translated by the scale into a pressure reading in
millimeters of mercury.

Page 6 of 11
It can be used in any clinic or emergency room setting, at the
hospital bedside, or in the operating room, but it requires
greater expertise and has generally been superseded by
applanation tonometers.

NONCONTACT TONOMETRY
- noncontact (“air-puff”) tonometer is not as accurate as
applanation tonometers.
- small puff of air is blown against the cornea.
- air rebounding from the corneal surface hits a pressure-
sensing membrane in the instrument.
- does not require anesthetic drops, since no instrument
touches the eye. Thus, it can be more easily used by
optometrists or technicians and is useful in screening
programs.

- accuracy of intraocular pressure measurement is affected by


OPHTHALMOSCOPY
central corneal thickness. The thinner the cornea, the more
easily it is indented, but the calibration of tonometers 1. Position the patient about 2 feet away
generally assumes a cornea of standard thickness. If the 2. Turn off the light to dim the room illumination
cornea is relatively thin, the actual intraocular pressure is 3. Set the focusing lens of the ophthalmoscope to zero
higher than the measured value, and if the cornea is relatively 4. Check the red reflex from a distance of 2 feet
thick, the actual intraocular pressure is lower than the 5. Approach the patient’s eye; the instrument is steadied
measured value. Thus ultrasonic measurement of corneal against the patient’s face by resting the ulnar border of
thickness (pachymetry) may be helpful in assessment of the hand holding the instrument against the patient’s
intraocular pressure. The Pascal dynamic contour tonometer, cheek; the thumb of the free hand raises the upper lid
a contact but non-applanating technique, measures 6. Instruct the patient to stare into the distance
intraocular pressure independent of corneal thickness. 7. Dial the ophthalmoscope’s focusing lenses into place to
- other applanation tonometers are the Perkins tonometer, a clarify the fundus image
portable mechanical device with a mechanism similar to the 8. Find the optic disc by following a retinal blood vessel
Goldmann tonometer, the Tono-Pen, a portable electronic 9. Examine the peripapillary retina
applanation tonometer that is reasonably accurate but 10. From the optic disc, follow the blood vessels outward to
requires daily recalibration, and the pneumatotonometer, examine the four quadrants around the posterior pole
which is particularly useful when the cornea has an irregular
11. Check for foveal reflex
surface. The Perkins tonometer and Tono-Pen are commonly
used when examination at the slitlamp is not feasible, for
THE DIRECT OPHTHALMOSCOPE
example, in emergency rooms in cases of orbital trauma with
This instrument consists of a single aperture through which
retrobulbar hemorrhage and in operating rooms during
light is projected into the subject’s eye and the examiner
examinations under anesthesia.
views the eye. It provides a magnified image (×15) and a field
of view of some 6.5–10 degrees. A set of corrective lenses can
SCHIOTZ TONOMETRY
be dialled into the aperture. These enables the focal point of
- now rarely used, measures the amount of corneal
the instrument to be adjusted. The rack of lenses usually
indentation produced by preset weights
contains equal numbers of positive and negative spheres
- advantage of this method is that it is simple, requiring only a
which can be dialled up to take account of the patient and/or
relatively inexpensive, easily portable hand-held instrument.
examiner’s refractive status. If examiners wish to wear their

Page 7 of 11
glasses, they can do so, and effectively they will need a zero into the patient’s eye and eventually reach the retina. It
lens in the eyepiece. The patient’s refraction must also be must be stressed that the head of the ophthalmoscope
taken into account and the relevant lens dialed into place. must be held very close to the patient’s eye in order to
With highly myopic or hypermetropic patients, their glasses gain the maximum field of view.
can be left on and used to nullify the effect of the refractive Clinical Skills for the Ophthalmic Examination: Basic Procedures, 2nd Edition
variation. Alternatively plus and minus 10 or 20 D lenses can
be positioned in the sight aperture to take account of very *For parts of the opthalmoscope: see last page
high hypermetropia or myopia. The size and brightness of the
illumination spot can be varied with the appropriate controls.
Additional features vary among the different models but
include a slit filter, producing a vertical slit of light which can
be used to examine contours or elevations on the fundus, a
grid for assessing the size of a fundus lesion, and a green filter
for red-free viewing. This latter filter will make red features,
such as haemorrhages, stand out due to increasing contrast
between the various shades of red and orange which reflect
from the fundus. Some ophthalmoscopes also include a
cobalt blue filter for use with fluorescein dye.

The view obtained with this instrument has a narrow angle of


view and a high magnification. The more myopic the patient,
the more effective the magnifying effect. This is useful for  Macula is usually at temporal side while optic disc is on
examining the optic nerve head; however the view is
the temporal side.
monocular and two-dimensional.

METHOD OF USE
1. Inform patients that you are going to look at their eye
with a bright light and that you will have to get very close
to their face. Instruct them to breathe normally.
2. The instrument is held to the examiner’s eye with the
illumination system switched on and for steadiness and
ease of use a hand can be placed on the patient’s
shoulder.
3. The examiner’s right eye is used for the patient’s right
eye and the examiner’s left for the patient’s left eye. If
the examiner finds it difficult to close one eye, or the
other, then it can be left open – with practice the brain
manages to ignore the image from the non-examining
eye.
4. The correct lens, as described above, is dialled into the
aperture.
5. The patient is asked to fix on a distant object and is told
to maintain that fixation, regardless of whether the INDIRECT OPHTHALMOSCOPY
examiner gets in the way. The examiner thus knows
roughly where the patient’s macula is situated and the
optic disc will be just nasal to this.
6. The examiner then points the instrument’s illumination
beam into the patient’s pupil and obtains a red reflex
from a distance of about half a metre and slowly moves
towards the patient. At this point media opacities such as
cataract can be seen as black features against the red
reflex. The rheostat is used to adjust the brightness of
the light for the patient’s comfort. If required, the front
of the eye, cornea, iris and lens can be examined with a
+10 lens dialed into the instruments lens bank.
7. Following this part of the examination the lens dial is
progressively turned towards zero to focus further back

Page 8 of 11
light to the right is reflected off of the cornea (C), while the
Binocular Indirect Ophthalmoscopy (BIO) is a technique that slit to the left is reflected off of the iris (I). As the latter slit
provides thorough view of the retina and vitreous through a passes through the pupil, the anterior lens (L) is faintly
dilated pupil in order to evaluate the health of the interior of illuminated in cross section.
the eye and to identify structural abnormalities that may be The patient is seated while being examined, and the head is
associated with reduced visual acuity thereby aiding the stabilized by an adjustable chin rest and forehead strap.
diagnosis of amblyopia Vaugn Asbury’s General Ophthalmology 39th Edition
optometry.osu.edu
Parts of Slit lamp biomicroscopy
Comparison Between Direct and Indirect Ophthalmoscopy Viewing Arm.
DIRECT INDIRECT The binocular eyepieces provide stereoscopic vision and can
Magnified image Not magnified be adjusted to accommodate the examiner's interpupillary
Can see only small area Lets you see bigger area distance. The focusing ring can be twisted to suit the
If with cataract, cannot see Ideal if with cataracts examiner's refractive error.
One handheld apparatus Uses a head gear and a
handheld condensing lenses The magnification element can be adjusted with the side dial.

SLIT-LAMP BIOMICROSCOPY
1. Identify the different parts of the slit-lamp biomicroscope
2. Enumerate the different uses of the slit-lamp

The slitlamp is a table-mounted binocular microscope with a


special adjustable illumination source attached. A linear slit
beam of incandescent light is projected onto the globe,
illuminating an optical cross section of the eye. The angle of
illumination can be varied along with the width, length, and
intensity of the light beam. The magnification can be adjusted
as well (normally 10× to 16× power). Since the slitlamp is a Illumination Arm
binocular microscope, the view is “stereoscopic,” or three- The illumination arm can be swung 180 degrees side to side
dimensional. on its pivoting bases allowing the examiner to direct the light
beam anywhere between the nasal and temporal aspect of
the eye examination.The dimension of the light beam can be
varied in height and width with these levers. It can provide
diffuse or focal illumination as an optical cross-section of the
anterior segment.Cobalt blue, or green filters can be selected
with this lever.

The Patient Positioning Frame


The patient positioning frame consist of two upright metal
Slitlamp photograph of a normal right eye. The curved slit of rods to which are attached a forehead strap and a chin rest.

Page 9 of 11
Just below the slit lamp table on the left is the ON switch and
The chin rest height can be adjusted with the knob just below provides high or low options in light intensity.
it.

Uses of Slit lamp


1. To visualize the anterior half of the globe—the
“anterior segment.
2. To study the details of the lid margins and lashes,
the palpebral and bulbar conjunctival surfaces, the
tear film and cornea, the iris, and the aqueous can
be studied.
3. Through a dilated pupil, the crystalline lens and the
anterior vitreous can be examined as well.
4. Because the slit beam of light provides an optical
cross section of the eye, the precise anteroposterior
location of abnormalities can be determined within
each of the clear ocular structures (eg, cornea, lens,
vitreous body).
5. The highest magnification setting is sufficient to
show the abnormal presence of cells within the
aqueous, such as red or white blood cells or pigment
granules. Aqueous turbidity, called “flare,” resulting
from increased protein concentration can be
The Joystick detected in the presence of intraocular
The joystick allows for focusing by shifting forward, inflammation. Normal aqueous is optically clear,
backward, laterally or diagonally. The joystick can also be without cells or flare.
rotated to lower or elevate the light beam. Vaugn Asbury’s General Ophthalmology 39th Edition
The locking screw located at the base secures the slit lamp
from movement when it is not in use. Other uses of slit lamp biomicroscope:- Internet source.
1. Routine observation of ocular adnexia
2. Routine investigation of posterior segment
3. Monitoring signs and symptoms of anterior segment
conditions
4. Further "special eye" investigations

Definition of Terms
Conjugate movement: Movement of the eyes in the same
direction at the same time.
Deviation: Magnitude of ocular misalignment, usually
measured in prism diopters but sometimes measured in
degrees.
Comitant deviation: Deviation not significantly affected by
which eye is fixing or direction of
gaze, typically a feature of childhood (nonparetic) strabismus.
Incomitant deviation: Deviation varies according to which
eye is fixing and direction of gaze,

Page 10 of 11
usually a feature of recent onset extraocular muscle paresis arc equals approximately 1.7 PD.
and other types of acquired Secondary deviation: The deviation measured with the
strabismus. paretic eye fixing and the
Primary deviation: Incomitant deviation measured with the normal eye deviating.
normal eye fixing). Torsion: Rotation of the eye about its anteroposterior axis
Secondary deviation: Incomitant deviation measured with Intorsion (incycloduction): Rotation of the 12 o’clock meridian
the affected eye fixing. of the eye toward the midline of
the head.
Ductions: Monocular rotations with no consideration of the Extorsion (excycloduction): Rotation of the 12 o’clock
position of the othereye. meridian of the eye away from the
Adduction: Inward rotation. midline of the head.
Abduction: Outward rotation. Vergences (disjunctive movements): Movement of the two
Supraduction (elevation): Upward rotation. eyes in opposite directions.
Infraduction (depression): Downward rotation. Convergence: The eyes turn inward.
Fusion: Formation of one image from the two images seen Divergence: The eyes turn outward.
simultaneously by the two eyes.Fusion has two aspects. Versions: Binocular rotations of the eyes in qualitatively the
Motor fusion: Adjustments made by the brain in innervation same direction.
of extraocular muscles in order to
bring both eyes into bifoveal and torsional
alignment.
Sensory fusion: Integration in the visual sensory
areas of the brain of images seen with the two
eyes into one picture.

Heterophoria (phoria): Latent deviation of the eyes held


straight by binocular fusion.
Esophoria: Tendency for one eye to turn inward.
Exophoria: Tendency for one eye to turn outward.
Hyperphoria: Tendency for one eye to deviate upward.
Hypophoria: Tendency for one eye to deviate downward.
(See Hypotropia.)

Heterotropia (tropia):
Strabismus: Manifest deviation of the eyes that cannot be
controlled by binocular vision.
Esotropia: Convergent manifest deviation (“crossed eyes”).
Exotropia: Divergent manifest deviation (“wall eyes”).
Hypertropia: Manifest deviation of one eye upward.
Hypotropia: Manifest deviation of one eye upward. By
convention, in the absence of specific
causation to account for the lower position of one eye,
vertical deviations are designated by the
higher eye (eg, right hypertropia, not left hypotropia, when
the right eye is higher).
Incyclotropia: Manifest rotation of the 12 o’clock meridian of
one eye about its anteroposterior
axis toward the midline of the head.
Excyclotropia: Manifest rotation of the 12 o’clock meridian of
one eye about its anteroposterior
axis away from the midline of the head.
Orthophoria: The absence of any tendency of either eye to
deviate when fusion is suspended.
This state is rarely seen clinically. A small phoria is normal.
Prism diopter (PD): The unit of angular measurement used to
characterize ocular deviations. A
1 diopter prism deflects a ray of light toward the base of the
prism by 1 cm at 1 m. One degree of

Page 11 of 11

Das könnte Ihnen auch gefallen