Beruflich Dokumente
Kultur Dokumente
MODULE 01
Ophthalmologists
SNELLEN’S CHART
Other patients can use the Jaeger chart. the patient is
scored depending on which line of sentences he can read.
J10 is the biggest and J1 is the normal acuity.
Page 1 of 11
Example result This is performed before studying the eye under
magnification
With Without Gross inspection and palpation: lesions, growth,
correction correction inflammatory signs (swelling, erythema, warmth,
FAR SIGHT tenderness)
OS 20/150 20/20 Check for the ff:
OD 20/300 20/20 o position of eyelids (ptosis, lid retraction)
NEAR SIGHT o asymmetry can be quantified by measuring the
OS J8 J1 central width (in mm) of palpebral fissure (space bet
OD J10 J1 lower and upper lid margins)
o abnormal motor fxn of the lids (upper lid elevation,
COUNTING FINGERS forceful lid closure) may be due to neurologic or
By convention, ophthalmologists test first the right eye. If primary muscular abnormalities
it is known that the other eye is “buron” maybe the patient o malposition of the globe (proptosis) that may occur
complained of it before assessment, check the “buron” eye in orbital disease
first. o bony orbital rim and periocular soft tissue
In counting fingers, place your finger 1 ft away from the General facial evaluation:
patient and let the patient count the fingers you are o enlarged preauricular LN, sinus tenderness, temporal
showing to him. If the patient cannot count at 1 feet away, artery prominence, skin/mucous membrane
move 1 more feet away until you reach the maximum of 3 abnormalities
feet. Record the distance where the patient can count the o You canauscultate for bruit at the temporal side of
fingers shown to him. the orbit or directly at the globe. This can help you
After the maximum of 3 feet and the patient cannot count detect carotid sinus fistula
the fingers, proceed to light perception test.
Page 2 of 11
evaluated for size, shape, direct light response, consensual stimulation of the right eye. As the light is swung from the
response, and near response. right to the left eye, both pupils will begin to dilate normally
SIZE as the light is moved away from the right eye and then not
In dim illumination, the average pupil diameter is 3 or 4 mm. constrict or paradoxically widen as the light is shone into the
This is ascertained by shining the light from below the left eye (since the direct response in the left eye and
patient’s nose so that the pupils are just visible to the the consensual response in the right eye are reduced
examiner; the light is not shone directly into the patient’s compared to the consensual response in the left eye and
eye. Pupils smaller than 2 mm are said to be miotic; pupils direct response in the right eye from stimulation of the right
larger than 6 mm are mydriatic. Miotic pupils may be caused eye). When the light is swung back to the right eye, both
by antiglaucoma medications, chronic iris inflammation, age, pupils will begin to dilate as the light is moved away from the
or a neurologic disorder. Abnormal mydriasis is caused by left eye and then constrict normally as the light is shone into
certain drugs, neurologic disorders, iris injury, or acute the right eye. This phenomenon is called a relative afferent
glaucoma. The pupils should be equal in size, although a pupillary defect (RAPD).
small difference (1 mm) may be a normal variation. If they are Vaugn Asbury’s General Ophthalmology 39th Edition
unequal (anisocoria), the difference between them should be
further evaluated in both dark and bright room illumination.
SHAPE
Both pupils should be round. The pupils are normally
centered or a little nasal in the iris. An eccentric pupil may be
the result of faulty embryonic development, injury,
intraocular surgery, or inflammation. In addition to being
eccentric, a pupil may also have an unusual shape.
Page 3 of 11
connective tissues, which regulates the direction of action
of the extraocular muscles by acting as their functional
mechanical origins (the active pulley hypothesis).
Landmarks to remember:
Mnemonic: "SO-4, LR-6, All the rest 3" (ie Superior Oblique by
CN 4, Lateral rectus by CN 6, and all the other EOMs by CN 3). 0mm Light reflection is at the center of the pupil
1mm Light reflection is in between the center of the pupil
TESTING OF THE VISUAL FIELD and the pupillary border
The patient is asked to follow a target with both eyes as it is 2mm Light reflection is at the pupillary border
moved in each of the four cardinal directions of gaze. The 3mm Light reflection is in between pupillary border and
examiner notes the speed, smoothness, range, and symmetry limbus
of movements and observes for unsteadiness of fixation (eg, 4mm Light reflection is at the limbus
nystagmus). Impairment of eye movements can be due to
neurologic problems (eg, cranial nerve palsy), primary KIMPSKY TEST
extraocular muscular weakness (eg, myasthenia gravis), or
mechanical constraints within the orbit limiting rotation of
the globe (eg, orbital floor fracture with entrapment of the
inferior rectus muscle). Deviation of ocular alignment that is
the same amount in all directions of gaze is called “comitant.”
It is “incomitant” if the amount of deviation varies with the
direction of gaze.
Vaugn Asbury’s General Ophthalmology 39th Edition
Procedure:
1. Have the patient look toward a penlight held several feet
away. (33 cms according to Vaughan & Asbury’s 18th ed.
Page 244)
2. Note for the pinpoint light reflection, or “reflex,”.
The Krimsky test is essentially the Hirschberg test, but with
Note: In normal eyes, pinpoint light reflection, or “reflex,”
prisms employed to quantitate deviation of ocular
should appear on each cornea and should be centered over
misalignment by determining how much prism is required to
each pupil if the two eyes are straight in their alignment.
centre the reflex [2] The Krimsky test is advisably used for
patients with tropias, but not with phorias.
If the eye positions are convergent, such that one eye points
https://en.wikipedia.org/wiki/Hirschberg_test
inward (“esotropia”), the light reflex will appear temporal to
the pupil in that eye. If the eyes are divergent, such that one COVER TEST
eye points outward (“exotropia”), the light reflex will be - More accurate method of verifying normal ocular
located more nasally in that eye. alignment.The test requires good vision in both eyes.
Vaugn Asbury’s General Ophthalmology 39th Edition
Procedure:
1. Ask the patient to gaze at a distant target with both
eyes open. If both eyes are fixating together on the
Page 4 of 11
target, covering one eye should not affect the
position or continued fixation of the other eye.
2. Suddenly covers one eye and carefully watches to see
that the second eye does not move (indicating that it
was fixating on the same target already). If the
second eye was not identically aligned but was
instead turned abnormally inward or outward, it
could not have been simultaneously fixating on the
target. Thus, it will have to quickly move to find the
target once the previously fixating eye is covered.
Fixation of each eye is tested in turn. BRUCKNER TEST
Note!
An abnormal cover test is expected in patients with diplopia.
However, diplopia is not always present in many patients with
long-standing ocular misalignment. When the test is
abnormal, prism lenses of different power can be used to
neutralize the refixation movement of the misaligned eye
(prism cover test). In this way, the amount of eye deviation
can be quantified based on the amount of
prism power needed.
Page 5 of 11
quadrant TONOMETRY
7. Diagram the confrontation field if an abnormality is 1. Enumerate and differentiate the methods of measuring
detected intraocular pressure
Make sure to test each eye separately. When occluding globe can be thought of as an enclosed compartment
the other eye, don’t depress the eyeball because it may through which there is a constant circulation of aqueous
cause blurring of vision. humor
During the 8part eye exam, we are expected to use only this fluid maintains the shape and a relatively uniform
the manual test. pressure within the globe
Automated perimetry is done in sophisticated diagnostic tonometry is the method of measuring intraocular
centers. Is uses a machine that accurately records the pressure using calibrated instruments.
visual fields. The machine can also determine how much normal range is 10 to 21 mm Hg
light the patient can see by altering the intensity of the less corneal indentation is produced as intraocular
light being shown to the visual fields. pressure rises.
It is advisable to make a diagram as to which visual field since both methods employ devices that touch the
is blinded because it can guide you in diagnosis patient’s cornea, they require topical anesthetic and
disinfection of the instrument tip prior to use.
with any method of tonometry, care must be taken to
avoid pressing on the globe and artificially increasing its
pressure.
APPLANATION TONOMETRY
- intraocular pressure is determined by the force required to
flatten the cornea by a standard amount. The force required
increases with intraocular pressures.
- the GOLDMANN APPLANATION TONOMETER is attached to
the slitlamp and measures the amount of force required to
flatten the corneal apex by a standard amount.
- the higher the intraocular pressure, the greater the force
required.
If the patient has bitemporal hemianopia (number 2)
- Goldmann applanation tonometer is a more accurate
what possible disease can lead to such problem knowing
method than Schiotz tonometry
that the optic chiasm is the structure affected? Pituitary
- following topical anesthesia and instillation of fluorescein,
tumors.
the patient is positioned at the slitlamp and the tonometer is
In automated perimetry:
swung into place. To visualize the fluorescein, the cobalt blue
filter is used with the brightest illumination setting. After
grossly aligning the tonometer in front of the cornea, the
examiner looks through the slitlamp ocular just as the tip
contacts the cornea. A manually controlled counterbalanced
spring varies the force applied by the tonometer tip.Upon
contact, the tonometer tip flattens the central cornea and
produces a thin circular outline of fluorescein. A prism in the
tip visually splits this circle into two semicircles that appear
green while viewed through the slitlamp oculars. The
tonometer force is adjusted manually until the two
semicircles just overlap, as shown in Figure 2–10. This visual
end point indicates that the cornea has been flattened by the
set standard amount. The amount of force required to do this
is translated by the scale into a pressure reading in
millimeters of mercury.
Page 6 of 11
It can be used in any clinic or emergency room setting, at the
hospital bedside, or in the operating room, but it requires
greater expertise and has generally been superseded by
applanation tonometers.
NONCONTACT TONOMETRY
- noncontact (“air-puff”) tonometer is not as accurate as
applanation tonometers.
- small puff of air is blown against the cornea.
- air rebounding from the corneal surface hits a pressure-
sensing membrane in the instrument.
- does not require anesthetic drops, since no instrument
touches the eye. Thus, it can be more easily used by
optometrists or technicians and is useful in screening
programs.
Page 7 of 11
glasses, they can do so, and effectively they will need a zero into the patient’s eye and eventually reach the retina. It
lens in the eyepiece. The patient’s refraction must also be must be stressed that the head of the ophthalmoscope
taken into account and the relevant lens dialed into place. must be held very close to the patient’s eye in order to
With highly myopic or hypermetropic patients, their glasses gain the maximum field of view.
can be left on and used to nullify the effect of the refractive Clinical Skills for the Ophthalmic Examination: Basic Procedures, 2nd Edition
variation. Alternatively plus and minus 10 or 20 D lenses can
be positioned in the sight aperture to take account of very *For parts of the opthalmoscope: see last page
high hypermetropia or myopia. The size and brightness of the
illumination spot can be varied with the appropriate controls.
Additional features vary among the different models but
include a slit filter, producing a vertical slit of light which can
be used to examine contours or elevations on the fundus, a
grid for assessing the size of a fundus lesion, and a green filter
for red-free viewing. This latter filter will make red features,
such as haemorrhages, stand out due to increasing contrast
between the various shades of red and orange which reflect
from the fundus. Some ophthalmoscopes also include a
cobalt blue filter for use with fluorescein dye.
METHOD OF USE
1. Inform patients that you are going to look at their eye
with a bright light and that you will have to get very close
to their face. Instruct them to breathe normally.
2. The instrument is held to the examiner’s eye with the
illumination system switched on and for steadiness and
ease of use a hand can be placed on the patient’s
shoulder.
3. The examiner’s right eye is used for the patient’s right
eye and the examiner’s left for the patient’s left eye. If
the examiner finds it difficult to close one eye, or the
other, then it can be left open – with practice the brain
manages to ignore the image from the non-examining
eye.
4. The correct lens, as described above, is dialled into the
aperture.
5. The patient is asked to fix on a distant object and is told
to maintain that fixation, regardless of whether the INDIRECT OPHTHALMOSCOPY
examiner gets in the way. The examiner thus knows
roughly where the patient’s macula is situated and the
optic disc will be just nasal to this.
6. The examiner then points the instrument’s illumination
beam into the patient’s pupil and obtains a red reflex
from a distance of about half a metre and slowly moves
towards the patient. At this point media opacities such as
cataract can be seen as black features against the red
reflex. The rheostat is used to adjust the brightness of
the light for the patient’s comfort. If required, the front
of the eye, cornea, iris and lens can be examined with a
+10 lens dialed into the instruments lens bank.
7. Following this part of the examination the lens dial is
progressively turned towards zero to focus further back
Page 8 of 11
light to the right is reflected off of the cornea (C), while the
Binocular Indirect Ophthalmoscopy (BIO) is a technique that slit to the left is reflected off of the iris (I). As the latter slit
provides thorough view of the retina and vitreous through a passes through the pupil, the anterior lens (L) is faintly
dilated pupil in order to evaluate the health of the interior of illuminated in cross section.
the eye and to identify structural abnormalities that may be The patient is seated while being examined, and the head is
associated with reduced visual acuity thereby aiding the stabilized by an adjustable chin rest and forehead strap.
diagnosis of amblyopia Vaugn Asbury’s General Ophthalmology 39th Edition
optometry.osu.edu
Parts of Slit lamp biomicroscopy
Comparison Between Direct and Indirect Ophthalmoscopy Viewing Arm.
DIRECT INDIRECT The binocular eyepieces provide stereoscopic vision and can
Magnified image Not magnified be adjusted to accommodate the examiner's interpupillary
Can see only small area Lets you see bigger area distance. The focusing ring can be twisted to suit the
If with cataract, cannot see Ideal if with cataracts examiner's refractive error.
One handheld apparatus Uses a head gear and a
handheld condensing lenses The magnification element can be adjusted with the side dial.
SLIT-LAMP BIOMICROSCOPY
1. Identify the different parts of the slit-lamp biomicroscope
2. Enumerate the different uses of the slit-lamp
Page 9 of 11
Just below the slit lamp table on the left is the ON switch and
The chin rest height can be adjusted with the knob just below provides high or low options in light intensity.
it.
Definition of Terms
Conjugate movement: Movement of the eyes in the same
direction at the same time.
Deviation: Magnitude of ocular misalignment, usually
measured in prism diopters but sometimes measured in
degrees.
Comitant deviation: Deviation not significantly affected by
which eye is fixing or direction of
gaze, typically a feature of childhood (nonparetic) strabismus.
Incomitant deviation: Deviation varies according to which
eye is fixing and direction of gaze,
Page 10 of 11
usually a feature of recent onset extraocular muscle paresis arc equals approximately 1.7 PD.
and other types of acquired Secondary deviation: The deviation measured with the
strabismus. paretic eye fixing and the
Primary deviation: Incomitant deviation measured with the normal eye deviating.
normal eye fixing). Torsion: Rotation of the eye about its anteroposterior axis
Secondary deviation: Incomitant deviation measured with Intorsion (incycloduction): Rotation of the 12 o’clock meridian
the affected eye fixing. of the eye toward the midline of
the head.
Ductions: Monocular rotations with no consideration of the Extorsion (excycloduction): Rotation of the 12 o’clock
position of the othereye. meridian of the eye away from the
Adduction: Inward rotation. midline of the head.
Abduction: Outward rotation. Vergences (disjunctive movements): Movement of the two
Supraduction (elevation): Upward rotation. eyes in opposite directions.
Infraduction (depression): Downward rotation. Convergence: The eyes turn inward.
Fusion: Formation of one image from the two images seen Divergence: The eyes turn outward.
simultaneously by the two eyes.Fusion has two aspects. Versions: Binocular rotations of the eyes in qualitatively the
Motor fusion: Adjustments made by the brain in innervation same direction.
of extraocular muscles in order to
bring both eyes into bifoveal and torsional
alignment.
Sensory fusion: Integration in the visual sensory
areas of the brain of images seen with the two
eyes into one picture.
Heterotropia (tropia):
Strabismus: Manifest deviation of the eyes that cannot be
controlled by binocular vision.
Esotropia: Convergent manifest deviation (“crossed eyes”).
Exotropia: Divergent manifest deviation (“wall eyes”).
Hypertropia: Manifest deviation of one eye upward.
Hypotropia: Manifest deviation of one eye upward. By
convention, in the absence of specific
causation to account for the lower position of one eye,
vertical deviations are designated by the
higher eye (eg, right hypertropia, not left hypotropia, when
the right eye is higher).
Incyclotropia: Manifest rotation of the 12 o’clock meridian of
one eye about its anteroposterior
axis toward the midline of the head.
Excyclotropia: Manifest rotation of the 12 o’clock meridian of
one eye about its anteroposterior
axis away from the midline of the head.
Orthophoria: The absence of any tendency of either eye to
deviate when fusion is suspended.
This state is rarely seen clinically. A small phoria is normal.
Prism diopter (PD): The unit of angular measurement used to
characterize ocular deviations. A
1 diopter prism deflects a ray of light toward the base of the
prism by 1 cm at 1 m. One degree of
Page 11 of 11