6-1. Stand
95
Disassembly and Replacement Procedures
JQ7
(7) Tilt the main body and pull out the cables through
the hole in the bottom of the main body.
Cable
96
Disassembly and Replacement Procedures
Stand
97
Disassembly and Replacement Procedures
(5) Remove the relay board of the pedals for height ad-
justment. Double sems screw (M3x8)
Remove the double sems screw (M3 x 8) with a
Phillips screwdriver and a wrench (distance between
the opposite sides: 5.5mm) and then remove the
connector of the foot switch. Relay board of pedals
For details, see “[8] Wiring Diagram.” for height adjustment
98
Disassembly and Replacement Procedures
6-1-4. Replacing the Relay Board of the Pedals for Height Adjustment
(1) Follow the steps described in “6-1-3. Replacing the
Foot Switch” except for (3).
(2) Remove the relay board of the pedals for height ad-
justment. Remove the double sems screw (M3 x 8)
with a Phillips screwdriver and a wrench (distance
between the opposite sides: 5.5mm) and then re-
move all the connectors from the relay board of the
pedals for height adjustment. For details, see “[8]
Wiring Diagram.” Connector
Stand
Special tool
Caster
99
Disassembly and Replacement Procedures
6-2. Hood
100
Disassembly and Replacement Procedures
Flat head
screw (M4x12)
Protective sticker
Claw
6-2-6. Replacing the Admittance Panel Hinge (with X-ray Cassette Tray Door)
(1) Remove the admittance panel hinge cover.
Remove the protective stickers affixed to the admit-
Double sems screw (M3x10)
tance panel hinge cover and then remove the six
double sems screws (M3 x 10) with a Phillips screw-
driver.
(2) Remove the rotary dampers (15kg) R/L and the ro-
tary shaft of the admittance panel.
Slide horizontally the rotary dampers (15kg) R and
L inside the admittance panel hinge R on your right
and remove them from their bearings.
Rotary damper (15kg) R
Turn the rotary shaft 180 degrees in such a way that the
CAUTION dot (●) is directly on the opposite side of the notch.
<Bear in mind the following point when attaching
the rotary dampers (15kg) R/L>
쎲 Before attaching a new rotary damper, be sure to
turn the rotary shaft 180 degrees.
103
Disassembly and Replacement Procedures
104
Disassembly and Replacement Procedures
105
Disassembly and Replacement Procedures
106
Disassembly and Replacement Procedures
6-3-1. Removing the Upper Section of the Main Body from the Main Body
(1) Remove the rear panel of the main body and pull
out the power box.
See (1) ~ (5) in “6-1-1. Removing the Main Body
from the HL Stand.”
(2) Remove the connectors and the ground wires from
the control board and the heater control board.
If the unit contains the SpO2 module, see (2) and (3)
in “6-4-1. Replacing the Control Board (for the Dual
Incu i)” and remove the slide plate.
Remove the connector (JC1) from the heater con-
trol board (for the Incu i), and the connectors (JA5,
JA8, JA19 and JA21) from the control board (for the
JC1
Dual Incu i).
Remove the sems screws (M4 x 8) fixing the ground
terminals of the upper section of the main body, the
107
Disassembly and Replacement Procedures
Steam gate
CAUTION
쎲 Be sure to attach a new packing when replacing
the heater.
Fan assembly
Connector
109
Disassembly and Replacement Procedures
CAUTION
쎲 Be sure to attach a new packing when replacing
the motor assembly.
CAUTION
쎲 Be sure to attach a new packing when replacing
the high temperature sensor assembly.
110
Disassembly and Replacement Procedures
(3) Remove the control box cover from the upper sec-
tion of the main body.
Remove the two double sems screws (M3 x 10) from
the upper section of the main body with a Phillips
screwdriver.
Power box
111
Disassembly and Replacement Procedures
Nut (M3)
112
Disassembly and Replacement Procedures
Setscrew (M6)
6-3-9. Replacing the Position Detection Board (for the Detection of the Middle Board)
(1) Remove the rear panel of the main body.
See (1) ~ (3) in “6-1-1. Removing the Main Body
from the HL Stand.”
(2) Remove the filter assembly.
Remove the four double sems screws (M3 x 10) with
a Phillips screwdriver.
Remove the pipe of the oxygen supply valve and the
Double sems screw (M3x10)
connector of the position detection board of the fil-
ter assembly.
113
Disassembly and Replacement Procedures
114
Disassembly and Replacement Procedures
JC2
115
Disassembly and Replacement Procedures
Humidity chamber
cover packing C
116
Disassembly and Replacement Procedures
6-3-16. Replacing the Position Detection Board (for the Detection of the Boiler Cap)
(1) Remove the humidity chamber cradle from the up-
per section of the main body.
See (1) ~ (3) in “6-3-12. Replacing the Humidity
Chamber Cradle.”
(2) Remove the position detection board.
Remove the washer head (M3 x 8) with a Phillips Washer head (M3x8)
screwdriver and then remove the connector.
117