Beruflich Dokumente
Kultur Dokumente
Service Manual
This manual is for DINAMAP ProCare Monitors models 100, 200, 300, and
400, with or without printers.
The model of the Monitor determines which parameters are in your monitor.
Please refer to applicable sections.
The content of this document including all figures and drawings is proprietary
information of GE Medical Systems Information Technologies, provided
solely for purposes of operation, maintenance or repair, and dissemination
for other purposes or copying thereof is prohibited without prior written
consent by GE Medical Systems Information Technologies, Tampa, Florida.
European Representative
GE Medical Systems
Information Technologies GmbH
Munzinger Straße 3-5
D-79111 Freiburg
Germany
Tel: •+ 49 761 45 43 - 0
Fax: •+ 49 761 45 43 - 233
Asia Headquarters
GE Medical Systems
Information Technologies Asia; GE (China) Co., Ltd.
24th Floor, Shanghai MAXDO Center,
8 Xing Yi Road, Hong Qiao Development Zone
Shanghai 200336, P.R. China
Tel: •+ 86 21 5257 4650
Fax: •+ 86 21 5208 2008
2009381-001 Revision A
NOTE: The information in this manual only applies to ProCare Monitor. It does not apply to earlier Monitors. Due to
continuing product innovation, specifications in this manual are subject to change without notice.
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Scope of Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.6. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Appendix B - Connectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
WARNING
Section 1 describes this volume and tells you how to use it. Information is
also provided about the physical and functional characteristics of the
Monitor, and how to get assistance in the event the unit fails to function
properly.
1.3.2 Assistance
If the product fails to function properly, or if assistance, service or spare
parts are required, contact Customer Support. Before contacting Customer
Support, it is helpful to attempt to duplicate the problem and to check all
accessories to ensure that they are not the cause of the problem. If you are
unable to resolve the problem after checking these items, contact General
Electric Medical Systems Information Technologies at 1-877-274-8456.
If repair parts or service are necessary, you will also be asked to provide:
Estimates for non-warranty repairs are provided pre-set flat rates to facilitate
prompt service. In cases where the product has external chassis or case
damage, please advise the Customer Support representative when you call.
Packing Instructions
Follow these recommended packing instructions.
Remove all hoses, cables, sensors, and power cords from the monitor
before packing.
Pack only the accessories you are requested to return; place them in a
separate bag and insert the bag and the product inside the shipping
carton.
Use the original shipping carton and packing materials, if available.
Place the product in a plastic bag and tie or tape the bag to prevent
loose particles or materials from entering openings such as hose ports.
Use a sturdy corrugated container to ship the product.
Pack with 4 to 6 in. of padding on all sides of the product.
Tape securely to seal the container for shipping.
Insurance
Insurance is at the customer's discretion. The shipper must initiate claims for
damage to the product incurred during shipping. Shipping damage is not
covered under warranty.
The customer is responsible for shipping charges to return the loaner unit to
General Electric Medical Systems Information Technologies. All loaner units
must be returned within 5 business days after receipt of your repaired unit.
Loaner and/or Rental products are available to meet your needs in non-
warranty situations.
NIBP
Nellcor™ Pulse oximetry (SpO2)
Masimo Pulse oximetry (SpO2)
Alaris™ Oral and Rectal thermometry
thermal recorder/printer
2.1 Introduction
DINAMAP ProCare Monitors provide non-invasive determination of systolic
blood pressure, diastolic blood pressure, mean arterial pressure, pulse rate,
temperature, and oxygen saturation.
1 Speaker Grille.
2 Data interface connector: Host communications port (15 pin D-type RS-
232 serial port) for use only with equipment conforming to IEC 601-1,
configured to comply with IEC 601-1.
Buttons
1 Silence button: Press to mute audible alarms. Any alarm active that is
acknowledgeable is also removed whenever this key is pressed. When
pressed after alarm sounds (silence active), the silence icon (bell) lights
to indicate that audible alarms have been silenced for 2 minutes.
2 Alarms button: Press to view or adjust parameter alarm settings
3 +/- button (Plus/Minus): Press the + button to increment an adjustable
setting and the - button to decrement an adjustable setting. This button is
active only when a user-setting mode (limit or menu) is active.
4 Menu button: Press to access menu settings that can be adjusted while in
clinical mode (i.e., ALARM VOLUME, PULSE VOLUME, INFLATE
PRESSURE; refer to Operating Modes in this section for a description of
clinical mode.)
5 SpO2 sensor connector: SpO2 sensor cable attaches here .
6 BP connector: BP cuff hose attaches here.
7 Inflate/Stop button: Press to start a manual BP determination or stop any
BP determination.
8 Temperature probe holster: Temperature probe is stored here.
9 Cycle button: Press to start Auto Cycle or STAT mode.
10 Temperature probe cover storage: Box of probe covers is stored here
11 History button: Press to activate the history mode. When activated, it
displays the most recent entries stored. Press and hold the button for 2
seconds to clear all entries stored.
12 Print button: Press to print currently displayed values or all stored entries
when in history mode.
13 On/Off button: Controls on/off state of monitor; push for power on and
push again for power off.
14 Temperature probe connector: Temperature probe cable attaches here.
Front Panel
15 Silence icon: when Silence button is pressed after alarm sounds (silence
active), silence icon (bell) lights to indicate that audible alarms have been
silenced for 2 minutes.
16 Systolic window: 3-digit red LED indicates measured systolic Blood
Pressure in mmHg.
17 Diastolic window: 3-digit red LED indicates measured diastolic Blood
Pressure in mmHg.
18 Alarm volume indicator: lights to indicate you are making a change to the
alarm volume.
19 Pulse volume: illuminates to indicate you are making a change to the
pulse volume.
20 Inflate pressure: illuminates to indicate you are making a change to the
inflation pressure.
21 Pulse Rate window: 3-digit yellow LED shows the pulse rate in beats per
minute.
22 SpO2 pulse indicator: Red LED bar flashes to indicate that real-time pulse
rate measurements are being derived from SpO2 signals.
23 SpO2 window: 3-digit red LED indicates oxygen saturation in %.
24 MAP/Cuff window: 3-digit red LED indicates measured MAP in mmHg
and shows instantaneous cuff pressure during BP determination.
25 Min window: Displays the BP mode if manual or STAT is the cycle time
when in Auto Cycle mode.
26 Battery power indicator: Green LED indicates the Monitor is operating on
battery power.
27 Low battery power indicator: Yellow LED indicates LOW charge status of
internal battery.
28 Charging indicator: Green LED indicates presence of external power
source and battery charging.
29 Temperature window: 4-digit red LED indicates measured temperature.
Right-Side Panel
30 External power socket: To be used with approved GE Medical Systems
Information Technologies AC-DC power converter ONLY.
Symbols
The following symbols are associated with the ProCare Monitor.
Note: The model of the Monitor determines which symbols appear on it.
Silence
Alarms
+/-
Menu
Inflate/Stop
Cycle
History
On/Off
Battery Power
Charging
Note: When using remote alarm, the ProCare Monitor should be considered
the primary alarm source. The secondary alarm is used for secondary
purposes only.
Pin # Function
1 Common
2 Inverted TTL Transmit Data
3 Inverted TTL Receive Data
4 +5 volts
5 No connection
6 No connection
7 Common
8 Remote Alarm
9 No connection
10 No connection
11 RS232 Transmit Data (TxD)
12 No connection
13 RS232 Receive Data (RxD)
14 No connection
15 No connection
Battery 633178CR
NIBP OPTIONS:
TEMPERATURE
SpO2
Nellcor**:
Masimo***:
2.6. Specifications
This product conforms with the essential requirements of the Medical Device Directive.
Accessories without the CE Mark are not guaranteed to meet the Essential requirements of
0 0 8 6 the Medical Device Directive.
The ProCare Monitor is protected against vertically falling drops of water and conforms to
IPX1 the IEC-529 standard at level IPX1. Vertically falling drops shall have no harmful effects to
the monitor.
Power Requirements
Battery 6 volt, 3.3 amp-hours. Protected by internal auto-resetting fuse and thermal protection.
Minimum Operation Time 2 hrs (5 min cycle with adult cuff at 25° C, SpO2 active at 60 bpm, temp in monitor mode,
printout of current values every 5 minutes) from full charge.
Fuses The Monitor contains four fuses. The fuses are auto-resettable and mounted within the
Monitor. The fuses protect the low voltage DC input, the battery, the remote alarm output
and the +5V output on the host port connector.
Environmental
Operating Atmospheric
700 to 1060 hecto Pascal
Pressure Range
Radio Frequency Complies with IEC Publication 601-1-2 (April 1993) Medical Electrical Equipment,
Electromagnetic Compatibility Requirements and Tests, and CISPR 11 (Group1, Class A)
for radiated and conducted emissions.
Mechanical
Mechanical
NIBP
Monitor Mode Accuracy ±0.1° C ± 0.2° F (when tested in a calibrated liquid bath; meets
ASTM E1112, Table 1 in range specified)
MAX-A,* MAX-AL* ±2
MAX-N†* ±2
MAX-P* ±2
MAX-I* ±2
MAX-R‡ ±3.5
OxiCliq A ±2.5
OxiCliq P ±2.5
OxiCliq N† ±2.5
D-YS ±3
DS-100A ±3
OXI-A/N ±3
OXI-P/I ±3
* The accuracy specification under motion conditions is ±3. For a definition of motion contact GE Medical Services
Technical Support.
† The MAX-N and the OxiCliq N were tested on patients > 40 kg.
‡ The accuracy specification has been determined between saturations of 80-100%
Pulse Rate
* The Masimo SET® SpO2 parameter with LNOP-Adt sensors has been
validated for no motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies in the range of 70-100% SpO2 against
a laboratory co-oximeter and ECG monitor. This variation equals plus or
minus one standard deviation. Plus or minus one standard deviation
encompasses 68% of the population.
**The Masimo SET® SpO2 parameter with LNOP-Adt sensors has been
validated for motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies while performing rubbing and tapping
motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive motion
between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in
the range of 70-100% SpO2 against a laboratory co-oximeter and ECG
monitor. This variation equals plus or minus one standard deviation. Plus or
minus one standard deviation encompasses 68% of the population.
†The Masimo SET® SpO2 parameter with LNOP-Neo Pt sensors has been
validated for neonatal motion accuracy in human blood studies on neonates
while moving the neonate’s foot at 2 to 4 cm against a laboratory co-
oximeter and ECG monitor. This variation equals plus or minus, one
standard deviation.Plus or minus one standard deviation encompasses 68%
of the population.
‡The Masimo SET® SpO2 parameter with has been validated for low
perfusion accuracy in bench top testing against a Biotek Index 2 stimulator
and Masimo’s simulator with signal strengths of greater than 0.02% and a %
transmission of greater than 5% for saturations ranging from 70 to 100%.
This variation equals plus or minus, one standard deviation.Plus or minus
one standard deviation encompasses 68% of the population.
3.1 Introduction
This section provides overall theory of operation and functional description
of the ProCare Monitor. The ProCare Monitor comes in six different
configurations:
The model of your monitor determines which parameters are in your monitor.
Using the ProCare monitor, a clinician can view, print and recall data that is
derived from each parameter. The Monitor is also capable of alerting the
clinician to changes in the patient’s condition. All of the main operations of
the ProCare Monitor are easy-to-use and only a button-touch away. Please
review the factory default settings and, where applicable, enter settings
appropriate for your use.
The ProCare Monitor is a portable unit that receives input power from an
internal rechargeable battery.
When the ON/OFF button is pressed, the Main Board is brought out of a
sleep mode and turns on the power regulators. The power regulators provide
conditioned power from the Lead Acid Battery. The external DC source is
used only to charge the Lead Acid Battery. Once the ProCare Monitor is
energized, a self-test is performed. The self-test automatically tests the main
functions of the ProCare Monitor. Failure of the self-test will set the ProCare
Monitor into a fail-safe mode with an audio alarm.
A reset signal to the SpO2 PWA is also provided so that the power up
sequencing is corrected. If the SpO2 circuit quits communicating to the Main
Board, the Main Board will attempt to reset the SpO2 PWA.
* Printer (optional)
Upon entering a FAILSAFE condition, the Main Board will perform the
following tasks:
The ON/OFF key can reset the Monitor and end the FAILSAFE condition.
The FAILSAFE condition will terminate automatically after 10 minutes to
preserve battery power.
The UI assembly also provides hardkey switches for the ProCare Main
Board. The primary processor asserts a HIGH on the 16 outputs of the 1-of-
16 decoder/demultiplexer one at a time and then reads at the signal on
SW_MUX. A LOW on SW_MUX indicates that the switch is asserted.
Patient data received from the finger sensor is filtered, amplified, and
analyzed on the SPO2 PWA. The information is sent to the Main Board via
the optically coupled electrically isolated serial connection. The primary
processor receives the data and routes it to the UI board for display. The
data is also sent to the printer if specified.
3.3.4 Printer
The printer receives power from the Main Board and communicates with the
primary processor. Printer presence and print head temperature is indicated
by PR_TH signal to the primary processor. When a print command is sent to
the printer from primary processor, the following will occur:
Together these signal (CONTROL DATA) cause the printer to print a graphic
hardcopy of the patient vital sign values and trend data. It also causes the
printer to print a hardcopy of error logging and service record data.
The printer has a built-in sensor to monitor the printer paper presence. When
the printer is out of paper, it sends a PAPER OUT signal to the primary
processor.
Main Board
SpO2 Probe
SpO2 Circuit
(Optional)
Isolated DC
Power Supply
Tubing
tubing Pump/Valve PT1/ PT2 Pneumatic
BP Cuff Control
Manifold NIBP Data Control
UI Board
Temp Probe Temp Control/ Data
Temp
(Optional) Control Switches
Interface
Audio
Printer Driver
Control
Control/
Data
Speaker Printer
ProCare Vital Signs Monitor -
System Block Diagram (1 of 2)
page 3-9/10
Real Isolated
Time Power
Clock Supply
External DC
Regulated DC
Power Supply
6Volt Battery
Printer
Drivers
Audio
Speaker
Control
Pneumatic
Valve Control
PT1/PT2
NIBP Data
SpO 2
Interface
Host
Communications
4.1 Introduction
This section contains general Monitor service procedures, including alarm
code interpretation, service mode operation, and periodic maintenance and
battery care. Refer to Section 5 for disassembly and reassembly procedures
and related component service information.
4.Insert the external DC Power Input connector into the DC input on the
lower right side of the monitor.
To set the date and time on the ProCare Monitor, you must first access the
configuration mode. The following table illustrates the menus and the
corresponding LCD display graphics.
SpO2
Year Systolic
Month MAP/Cuff
Day Diastolic
Hour min
Minute min
Procedures
1. With the Monitor off, press and hold the Menu button at the same time as
pressing the On/Off button for 3 seconds.
3. To set the date and time, press the Menu button to move from one setting
to another. (You will need to press the Menu button several times until the
year setting appears. Use the +/- buttons to increment or decrement the
individual settings.) Once you are finished changing a setting, press the
Menu button again to move to the next setting.
Note: For the date and time to be saved, you must advance the menu
through the min setting.
Perform a blood pressure by connecting the supplied hose and cuff together,
then attaching to the front of the ProCare Monitor. Press the Inflate/Stop on
the front to begin the NIBP cycle.
·The SpO2 sensor is an assembly consisting of two parts: the DS-100A, and
the extender cable DOC-10. Connect the cables prior to attaching to the
monitor. An SpO2 reading will be displayed within moments of attaching the
sensor to either a Nellcor simulator or to your finger.
When the pneumatic integrity of any NIBP cuff and hose is in doubt, replace
the cuff and hose, and discard the questionable accessories.
Wipe the exterior of the Monitor with a cloth slightly dampened with mild
detergent or normal hospital bactericides. Use dishwashing detergents such
as IVORY and JOY (registered trademarks of Procter & Gamble Corp.), or
PALMOLIVE (registered trademark of Colgate-Palmolive Corp.)
Clean the adult cuffs supplied for use with the monitor by hand washing in
warm, soapy water. However, take care to avoid entry of water into the cuff
and hoses at any time. If water enters the cuff, dry the cuff by passing air
through it.
The neonatal cuffs are for single patient use - discard if they become soiled.
Clean cuffs and hoses with a cloth slightly dampened with mild detergent.
Clean SpO2 sensor surface before and after each patient use. Clean SpO2
sensor with a cloth slightly dampened with a mild detergent. Wipe SpO2
sensor to ensure all detergent residue has been removed.
CAUTION: Do not apply isopropyl alcohol to the Monitor - some parts can
become marred and cracked.
For the above, follow manufacturers' recommendations for dilution rate and
use. These recommendations are not an endorsement of the manufacturers
or of the effectiveness of these materials for cleaning or disinfecting.
Elevated temperatures will shorten battery life and can lead to permanent
battery damage.
·Batteries should be charged before first use or after long periods of storage.
·The battery should be charged for 8 hours before use, as a charged battery
loses some charge when left in storage. It is possible to use the Monitor
while the Monitor is charging.
·It is normal for the battery to become warm during charging or after use.
·If the monitor is idle for extended periods, it should be fully charged once a
month to ensure optimum performance.
Note: To enter service mode press and hold the CYCLE buttons while
pressing the ON/OFF button for 3 seconds.
The absence of a ground connection on the power brick obviates the need
for any ground resistance test, such as that described in NFPA99 section 7-
5.1.3.2. (Some non-USA Power bricks are equipped with a non-conductive
pin at the ground pin location to aid mechanical retention of the Power Brick
into the receptacle.)
Hi-Pot testing of the patient connections (1500 VRMS, 60 Hz) may be done
by connecting one lead of the Hi-Pot tester to the outer shell of a spare DC
connector plug inserted into the DC input connector on the Monitor. The
other lead of the tester should be connected to the patient connection being
tested.
NOTE: Hi-Pot testing is done on every unit at the factory and should not be
repeated unnecessarily nor performed more often than required.
2. Plug an SpO2 cable into the SpO2 connector on the front of the unit.
3. Plug SpO2 probe Hi Pot adapter in to the DB9 jack at the end of the SpO2
cable.
Note: This test is written so that a technician who is familiar with the
ProCare monitor and the test equipment will be able to successfully
complete the test procedures. To enter service mode press and hold
the CYCLE button while pressing the ON/OFF button for 3 seconds.
Events that require you to manually save to the Monitor’s memory include
calibration attempts and enabling/disabling the Monitor’s parameters (refer
to section 4.6.4.) It is not necessary to manually save following a pressure
transducer calibration verification or simple calibration check.
The number of times the flash memory can be saved is limited. The fatal
alarm error 975 may be issued after this process is performed more than 90
times. Following issuance of this fatal alarm, the monitor must be returned to
the GE Medical Systems Information Technologies Service Center to have
the flash memory reset so that future saves can be performed.
4.6.1 SETUP
1. Connect your NIBP Analyzer to the ProCare Monitor.
6. After the system inflates to 200mmHg, wait for the "Wait" on the NIBP
analyzer message to be cleared
7. Press the "Start" button on the NIBP Analyzer to begin the leakage test
8. After 60 seconds the NIBP Analyzer will display the leakage rate
5. Use the inflation bulb to inflate the cuff, hose and pressure indicator setup
to 200mmHg
6. Record and verify the pressure reading on the top LED display
(SYSTOLIC)
7. Record and verify the pressure reading on the bottom LED display
(DIASTOLIC)
9. Record and verify the pressure reading on the top LED display
(SYSTOLIC)
10. Record and verify the pressure reading on the bottom LED display
(DIASTOLIC)
12. Record and verify the pressure reading on the top LED display
(SYSTOLIC)
13. Record and verify the pressure reading on the bottom LED display
(DIASTOLIC)
15. Record and verify the pressure reading on the top LED display
(SYSTOLIC)
16. Record and verify the pressure reading on the bottom LED display
(DIASTOLIC)
Note: To enter service mode press and hold the CYCLE button while
pressing the ON/OFF button.
6. Close valve on bulb and manually inflate pressure to 200 mmHg (using
the CuffLink manometer as reference).
7. Press MENU button when pressure reads 200 mmHg to save calibration
setting.
Note: the number of remaining changes to the flash memory available are
displayed in the MAP/Cuff window. Following each true calibration, the
Monitor decrements by 1 the number of remaining calibrations available. If
the MAP/Cuff window is displaying a figure of less than 10 and the accuracy
of the Monitor is in doubt, contact General Electric Medical Systems
Information Technologies Technical Support at 1-877-274-8456.
b. Press and hold the MENU button until two beeps are heard (one when
9. Press ALARM button several times, verify unit cycles through all alarm
settings (i.e SYS, DIA, SPO2).
2. During the power-up self-test verify all 7 segment LED display segments
and all discrete LEDs (except CHARGING LEDs) illuminate and are the
correct color.
°F BATTERY
°C BATTERY ICON
SYS/DIA = 120/80
MAP = 90
BPM = 80
3. Record and verify systolic, diastolic, map and heart rate from the monitor
display
5. Record and verify systolic, diastolic, map and heart rate from the monitor
display
7. Record and verify systolic, diastolic, map and heart rate from the monitor
display
5. Press "Inflate/Stop" and verify that the pump does not start
5. Record and verify the reading in the TEMP display is 80.2° F ±0.2° F.
7. Record and verify the reading in the TEMP display is 102° F ±0.2° F.
8. Set the probe simulator to B.P. and verify reading is 106.0° F ± 0.2° F.
Press broken probe button down and verify the Monitor displays .
Nellcor recommends use of the SRCmax Portable Tester for use with
ProCare Monitors equipped with the Nellcor SpO2 system.
1. Connect the appropriate SpO2 simulator and cable to the SpO2 connector.
Pulse value
Saturation value
6. Verify the sound has stopped but the error display remains
Pulse Value
Saturation value
NIBP
E82 Excess Air in Cuff Residual air in cuff above threshold for successful
auto-zero.
SpO2
E23 SpO2 Sensor off finger No SpO2 signal, check or reposition the sensor
E12 Recorder cannot print Main battery voltage is too low to operate
900-
Internal Memory Errors Refer Monitor to Customer Service
999
Appendix B - Connectivity
CHANT
• When the DINAMAP monitor alarms, the Nurse Call System is triggered.
• Uses a Nurse Call Cable to attach to the System. P/N 487208CR.
Compatible Monitors: Compact, PRO 100-400 Series, ProCare Series, and
PRO 1000.
Alarm View
CIC
ApexPro
ILC-1926
ILC-1927
- Apex version 2
- CIC version 3
2. Insert a flat-head screwdriver between the top of the display cover and
the body of the monitor.
3. Twist the screwdriver so that the display cover is slightly separated from
the body of the monitor.
4. Using two hands, grasp the top-middle edge of the display cover and pull
away and down.
2. Insert the tabs at the base of the display cover into the slots underneath
the bottom row of LEDs.
3. Using two hands, apply firm pressure to the outside edges, while flexing
the upper middle-edge of the display cover away from the monitor. It is
imperative that even pressure is applied as the plastic retention tabs can
break free from the display cover.
4. Apply steady, even pressure to the opposite sides of the display cover
until an audible click is heard.
15 2010429-001 English 400 lbl kit 1. keypad labels set (LH & RH)
2010478-001 English 300 lbl kit 2. Fascia
2010479-001 English 200 lbl kit 3. Display cover
2010480-001 English 100 lbl kit For language variants reference following pages
16 Help card guides All 3 guides and for each language
or
P3V3
P5A
C47 C45
C50 100nF 25V 100nF 25V P3V3 P3V3
C48 C44
25V 100nF
100nF 25V C46 100nF 25V
R1 R117
100nF 25V 100K 100K
13 22 55 91 97
PR_TH R104
VDD1
VDD2
VDD3
VDD4
VDD5
R84 78 49 1K00
I_CHG PQA0 INT0 PUMP_ON
77 50
10K0 PQA1 INT1 DUMP_ON
76 53
V_DC_MON PQA3 INT2 LATCHED_OVP
75 56
PUMP_I PQA4 INT3 DUMP2_ON
74 57
PT1 PQB0 U9 INT4 ADULT-0
73 58
PT2 PQB1 INT5 PNEU_RESET
FPT 72
PQB2 MMC2107 INT6
60
TP34 R116
71 61
V_BAT_MON_PRI PQB3 INT7 TP33 100K
P3V3 80
1K00 79
VRH
VRL 7
R119 1K00 PA0 PR_DI
48 6
D_FROM_COMM RXD1 PA1 PR_CLK
R108 46 5
D_FROM_SPO2 RXD2 PA2 PR_ST1
47 4
D_TO_COMM TXD1 PA3 PR_ST2
45 3
D_TO_SPO2 TXD2 PA4 PR_ST3
2
1K00 PA5
1
PR_ST4
63
SPI_MISO MISO PA6 PR_ST5
R106 62 100
SPI_MOSI R206 MOSI PA7 PR_ST6
65
SPI_CLK SCK
R201 75R0 TP26
66
SS
17
TP39 TP40 TP41 75R0 PB0 PR_VHON
99 16
T_DE DE PB1 PR_M4
92 15
T_CLK TCLK PB2 PR_M2
93 14
T_TDI TDI PB3 PR_M3
94 11
T_TDO TDO PB4 PR_M1
95 10
T_TMS TMS PB5 PR_EOP_EN
98 9
T_TRST TRST PB6 PR_LAT
8
Place near U9 C32 PB7 PR_EOP
1K00
R103 10P
90
CLKOUT SHDN_TO_PRI
RESET_TEST-0 R62 87 27 R114
0R0 XL2 86
EXTAL PC0
26 1K00
C33 8.0MHZ 10M0 R63 XTAL PC1 SW_MUX
25 R113
10P 1K00 83
RESET
PC2 RTC_DATA
24
R90 84
RSTOUT
PC3
23
TP42
RESET_PRI-0 43
PC4 RTC_CLK
20
1K00 ICOC10 PC5 SPO2_RST-0
42 19
P3V3 R120 41
ICOC11 PC6 IPS_ON
18
40
ICOC12 PC7 CS_SEC-0
R115100K 39
ICOC13 R112
ICOC20
100K 38
ICOC21 PD0
35 1K00 FSP_FROM_PRI
ADC_PWM 37
ICOC22 PD1
34
FAST_OVP-0
CS_DISPLAY 36
ICOC23 PD2
33
DUMP_ON_MON
PWM_VOL PD3
32
D_FROM_TEMP
PWM_PITCH 59
VPP PD4
31
D_TO_TEMP
70 30
VDDH PD5 REM_ALARM-0
P5D 82
VDDA PD6
29
TEMP_PROBE_IN
81 28
52
VSSA PD7 DUMP2_ON_MON
VDDF
51
C49 85
VSSF
69 R107
C12 100NF VDDSYN PE5 FSP_TO_PRI
10N0 25V P5D P3V3 P3V3 88
VSSSYN PE6
68
TP30 1K00
64 67
VSTBY PE7
44
TEST
R118
VSS5
VSS1
VSS2
VSS3
VSS4
C53 P3V3SEC R98 P3V3 R190
C1 C52 C15 100NF C3 C16 C54 100K 10K0
C14 10N0 25V 10U0 10N0 100NF 12 21 54 89 96 10K0
10U0 10N0 100NF
25V
P5A 25V R95
1K00
PE7 needs to be high at start-up
C2 C5
10U0 100NF
25V
P3V3SEC
P3V3SEC
RESET 1
VIN
C141 3 U22 5
C35
R177 9 OSC1
U3
20K0 22P0 XL1 PIC16LC73B
4.0MHZ
C34 R35 10
OSC2
22P0 0R0 RC0 11
FSP_TO_PRI 12 RC1/CCP2 RC7/RX 18
13
SEC_ALARM RC2/CCP1 RC6/TX 17
RC5/SDO 16
RC4/SDI 15 FSP_FROM_PRI
RC3/SCK 14 PNEU_RESET
RA5/AN4/SS 7 ADULT-0
SPI_MOSI
AGND VSS
SPI_CLK
8 19
CS_SEC-0
SHDN_TO_PRI
PS_ON
SPI_MISO
C66
P5A
100nF 25V
1 VIN
LP2980-5V PNEUMATICS
U7 VOUT 5
P5D PBAT
PBAT
Pump Connector
3 ON/OFF
P5D GND C37
C39
J7
2 2U2
3 16V
220U0 D9
PBAT BAS16
1
PUMP 1 C65 TP37 TP38
2
3 P5D
PBACK_IN PBACK_OUT
4 DUMP 100nF 25V
5 P5D P5D XFPM-050
6 DUMP2 3
C28 PT1 VCC
R51 PT1
1 R36
33U0 5V1 VOUT
2
Spare Parts
DUMP_ON_MON U14 P5A P5A
R202 3 4
HCT04
100K
R9 C60
Q7 U14 10K0
SI2306DS 5 6
HCT04 100nF 25V
DUMP_ON
3
TLC2272
1
3 8
U14 1
FPT
2 9
HCT04
8 R10 2
4
DUMP2_ON_MON R203 10K0
U14 U13
11 10
100K HCT04
Q24 R6
SI2306DS C17
U14 10N0 10K0 R83
DUMP2_ON
3
13 12
649K
1
HCT04
C64
2
R204 R205 R173 R121
100K 100K 100K 100K 10N0 P5A
Q17 C61
SI2306DS R29
PNEU_OVERRIDE-0 3 4K99 100nF 25V
1 P6VMIN
VCC 14
R28 ADC_PWM
2 2 1
HCT04
5 C58 4K99 U14
Q9 6
C43 GND
7
SI9410DY 7 100nF 25V
470N
PUMP_ON 8 P5A2
4
2
LP2980-5V
1 VIN
3
U6 VOUT 5
3 ON/OFF C36
5
GND 2U2
Q8 6
2 16V
SI9410DY 7
FAST_OVP-0 4
8
R44 C63
2
1K00
3 C59 100nF 25V
PT2_SEC
C90
R87 100nF
100K
BC_5V
PBAT R39
MC78L05 301K
8 1
IN OUT
U2
C68 VR1
R149 GND 100NF
2 3 6 7 50K CW
0R10 25V
VF = 6.9 V
R150 VC = 7.2 V
0R10
R42
10K0
BC_5V
Battery Connector PBAT C71
R13
J2 4A0 10K0 100nF 25V
1
FS4
2
3 8 OPA2340
R72 D12
C98
MBR2030CTL
2
1 1 3
D14
U15
33U0 2
4 4K99 BAS16
25V
1
R11 R81
R61
249K
C117 Q15
R93 C91 R146 C118 DTC143ZKA
100NF 100NF Logic
R64
4K99
V_BAT_MON_SEC
V_BAT_MON_PRI
I_CHG
FAST_CHG
SCHEMATIC Schematic - Main Board
PWA, MAIN BOARD, PROCARE P/N 2008855 Rev B
(4 of 10) page 5-11/12
+3.3 V SUPPLY
PBAT
P3V3
MIC4680-3V3 68U0
2 VIN SW 3 1 2
U12 L2
1 SHDN FB 4
C30 C74 GND
C75
100NF C41 100NF
33U0 25V 5 6 7 8
220U0
25V
R101 25V 1
10K0 25V
D6
2
BYG22
1
68U0 2
3 P6VMIN
L3
MIC3172 D7
5 VIN VSW 7
2 1
U16 BYG22
4 EN
C93 FB 3 R27 C96 C97 C76
C31
33U0
100NF
25V
SGND GND COMP 12K1 220U0 220U0 100NF
25V
1 8 6 2 25V 25V
25V
R65
1K00
R33
C25 2K87
1U0
16V
PROGRAM CONNECTOR
J1
1 PGM_PS_ON-0
2
3
4
+5.0 V SUPPLY
5
P5D
MIC39101-5V
2 VIN VOUT 3
U5
C38 C72 1 EN
FLG 4
C73 C4
2U2 100NF
25V
GND 100NF
25V
10U0
16V 5 6 7 8
RESET_TEST-0
C79
P5D
PRINTER INTERFACE
R123 R59 R58 R57 R56 R55 R54 R122 100nF 25V
100K 100K 100K 100K 10K0 100K 100K 100K 20 19
VCC 2OE
PR_DI 2
1A0
U11 1Y0 18
4 16
1A1 1Y1
PR_ST6 6
1A2 1Y2
14
8 12
PR_ST5 1A3 1Y3
17 3
2A0 2Y0
PR_ST4 15
2A1 2Y1
5
13 7
PR_ST3 2A2 2Y2
11 9
2A3 2Y3
PR_ST2
PR_ST1 HCT240
1OE GND
PR_CLK 1 10
P5A P5A
P5A
P5D
Printer Connector
R91 R60
R97 20K0 30K1 J9
PR_EOP 1M00 TLC2272
8 5 1
7 2
P3V3 6 3
4 4
U13 R89 5
Q11 30K1 6
PR_EOP_EN DTA143XKA 7
Logic 8
R73 9
10
100R 11
PBAT 12
13
14
15
R14 Q5 16
10K0 SI2307DS
17
2
R15 1
18
10K0 PBAT_PR 19
20
3
21
C6 22
PR_VHON Q3 23
DTC143ZKA 33U0 P5A 24
Logic 25V
25
26
R26 27
15K0
PR_TH
P5D
R53
2K0
PR_LAT Q4
DTC143ZKA P3V3 C78
Logic
C77
100nF 25V
25V 100nF
2 13 5
VCC VS2 VS1
PR_M1 3 IN1 OUT1 4
LB1831M
PR_M3 7 IN2 OUT2 6
PR_M2 15 IN3 U4 OUT3 14
PR_M4 11 IN4 OUT4 12
17 19
FG5 FG7
18 20
FG6 FG8
FG1 FG2 FG3 FG4
1 8 9 16
R49
P5D 1R00
3 C
C7
R82 1 Q27 10U0
PWM_VOL
10K0 B MMBT3904LT1 16V
U10
E 2
LM386
C26 R24 C10 R41 3 6 J4
1U0 1 C42
100K 4K99 8
10N0
1
16V
2 5 2
4
7 220U0
R74
R111 C11 R71 C87 2R70 C88
10N0 100NF 100NF
10K0 15K0 25V
C8 C86 25V
Speaker Connector
10U0 100NF
25V
16V
D18
PWM_PITCH 1 3
R94
BAS16
49K9
P5D
C89
SEC_ALARM R124
10N0
10K0
R46
P5D
1K00
P5D
R178
1K00 R179
2K0
3
D15 3
D16
1
5V1 5V1 P5D
1 Q28
Q26 P5D DTC143ZKA
DTC143ZKA Logic
C85
Q25
Comms Connector
Logic
D_TO_COMM DTC143ZKA
FS2
0A3
100nF 25V
HOST COMM
Logic
J6
15 12 16 GND 1
11 T1IN T1OUT 13
External DB15
Connector Mapping
9 R1OUT R1IN 8
J6 NAME DB15
1 EN 1 INVALID 0
GND
1 GND 1
D_FROM_COMM P5D 14 2 TTL_TX 2
3 TTL_RX 3
Q13 4 5V_FUSED 4
FS3
0A3
DTA143XKA
Logic 5 GND 7
6 REMOTE_ALARM-0 8
R110
10K0 7 RS-232_TX 11
Q6 8 RS-232_RX 13
NDT3055L
1
D2
REM_ALARM-0 SM6T15CA
2
R102
10K0
C92
R22 R19 R18 R17 100NF
25V
10K0 10K0 10K0 10K0
T_TDO 3 R125
4 10K0
T_CLK 5
6
7
J5
8 1
RESET_TEST-0 9 2
T_TMS 10
SW_MUX 3
11 4
T_DE 12
CS_DISPLAY 5
13
SW_PWR-0 6
T_TRST 14
SPI_MOSI 7
SPI_CLK 8
9
R21 R133 R134 R135 CHARGING_LED 10
1K00 V_IN
499 499 499
R127 R126
100K 100K
ISOLATION BARRIER
C126
10U0
R185 C124 C149 N5VSP
PBAT SP
10U0
100NF MIC5270-5.0 SP
4K99 25V
2 GND
D21 U28-VOUT 4
1 2 5 -VIN
BYG22 DO NOT POPULATE
C151 C129
T1 P5VSP
100NF 33U0 C128 6395-TO12
25V 25V 680P 7 D20 MIC39101-5V
2 2 1 2 VIN VOUT 3
BYG22 U30
R153 8 L11 C150 C125
12 75R0 11 C123 1 EN
FLG 4
10U0
100NF
10U0
VIN L13 MCB1608S102E 25V GND
IPS_ON 15 SD-0 SW 11
5 5 6 7 8
1N0 1N0
3K15V 3K15V
ISOLATION BARRIER
T
14 4 1
Z>200/100MHZ
J16
1
DET + A
2
DET - C
3
P5VSP
INNER SHIELD
4
INNER SHIELD
5
N/C NC
TO J1 ON MASIMO MS-5 6
OUTER SHIELD 2
7
IR LED CATH
8
D35
OUTER SHIELD
9
3
BAV199
RED LED CATH
10
N/C NC
1
P3V3 P5VSP
R183
Q19
SP
DTA143XKA
U23 10K0
4N35
P3V3
SpO2 Module Connector 14 pin
Logic
R144 1 5
AN C
249 B
6
2 4
CTH E
C119 U25 4
GND
RESET
HCPL-2601 5
100nF 25V R142 GND
C 3
6
SPO2_RST-0 8
VCC
AN
2 3K01 7
+12V (N/A)
ANALOG OUT (N/A)
Q16
7
VE 1 8
6 TRANSMIT (MP506 OUT) TO J2 ON NELLCOR MP506
5
VO
CTH
3 MMBT3904LT1 B 9
RECEIVE (MP506 IN)
GND 10
2 E +5 ANALOG
11
CTS (MP506 IN)
12
D_FROM_SPO2 R136 13
GND
+5V DIGITAL
14
499 GND
R184
P5VSP SP 1K00
P3V3
SP SP
J17
Q20 U26 C120 R139 MASIMO
D_TO_SPO2 DTA143XKA HCPL-2601
1K00 1
100nF 25V +5V
Logic 2
R145 2 VCC
8
3
GND
AN 7
VE
249 6
SP 4
GND
3
CTH
VO 5 TO J3 ON MASIMO MS-5
5
GND 6
GND
7
RESET
8
SP 9
RECEIVE (MS-5 IN)
10
TRANSMIT (MS-5 OUT)
ISOLATION BARRIER SP
C130
17 AVDD DVDD 1
Temp PGM Connector
2 3 6 7 11 DIN 20
D_TO_T_ADC 6
10U0 12
AD0
D_FROM_T_ADC 5
16V AD1 DOUT 21
13 AD2 SCLK 22
SCLK_TO_T_ADC 4
3
14 AD3
DRDY-0 23
DRDY_FROM_T_ADC 2
1
15 AINCOM
T
U34 CS-0 19 DO NOT POPULATE
P5TEMP R172 P5TEMP
9 REF+
ADS1240 T J15
10 REF-
13K7 6 R156
0.1% CLK_TADC R167 DSYNC-0
3 18
P5TEMP X_IN POL 10K0
Temperature
4 24
100R X_OUT BUFEN
R155
R193 PDWN-0 7
ISOLATION BARRIER
BAS16 16 AGND DGND2 8
1
L14 R157
J14 V_THERM
1 BLM11B221SB 1K00 T T
2 L9 C152 C136
3 100NF 100NF
3
D23 25V 25V
4
BLM11B221SB 1
BAS16 P5TEMP P3V3
5
6
T T C138
R154
T
100nF 25V
T P5TEMP U33
4N35 10K0
C133 T
P5TEMP 1
AN C
5
D_FROM_TEMP
P5TEMP 100nF 25V R168 2
B
6
4
CTH E
T
2K0
Q21 R171 R191
3
D28 R194 R195 R196 R197 R164 SI2307DS
15 14 2K21 Q23
30R 30R 30R 30R 499K 2 1K00 VDD1 RB7
DTA143XKA D_TO_TEMP
BAS16 R192 16 13
L7 1 1 VDD2 RB6
4 12 Logic
MCLR-0 RB5
10K0 11
BLM11B221SB
3
19
RB4/PGM
10 T U32 R169
RA0/AN0 RB3
20
RB2/TX
9
4N35
1K00
3 RA1/AN1
D29 1
RA2/AN2/VREF RB1/RX
8 R198 R188 5
C AN
1
BAS16 2
RA3/AN3/CMP1 RB0/INT
7 10K0 100K 6
1
R160 MON_HTR 3 4
B
2
RA4/T0CKI/CMP2 E CTH
10K0
C134
18
OSC1
U35
R187
T T T
100NF PIC16F628
25V 17 10K0
OSC2/OUT
P5TEMP
VSS1 VSS2
T C132 5 6
T
3
D22 22P0 R162
BAS16
XL4
1 4.0MHZ 0R0 T T
L10 C131
BLM11B221SB 22P0
3
D25 TEMP_PROBE_IN
1
BAS16 Q22 T
SI2306DS
3
P3V3
Temp Probe Switch Connector
R158 R189 R132 P5D
10K0
T
EN_HTR 10K0 J13
499K
V_JUMPER C142 1
2
2U2
C135 16V
3
100NF R165
25V
499K T
SCHEMATIC
PWA, MAIN BOARD, PROCARE P/N 2008855 Rev B
(10 of 10) page 5-23/24
P5D
C3
PBAT
P5D 100NF
C2
24 PDG_L[0:12]
100NF VCC 11 PDG_L0
Y0
20 9 PDG_L1
Y1
VCC 7 DIG_MUX0 2 10 PDG_L2
IO0 A0 Y2
R127 11 13 DIG_MUX1 3 8 PDG_L3 R1 R13 R4 R3 R5 R6 R7 R8 R9 R10 R11 R12
DS0 IO1 A1 Y3
100R 12 6 DIG_MUX2 21 7 PDG_L4 10K0 10K0 10K0 10K0 10K0 10K0 10K0 10K0 10K0 10K0 10K0 10K0
CLK IO2 A2 Y4 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1%
8 14 DIG_MUX3 22 6 PDG_L5
Q0 IO3 A3 Y5 Q15 Q16 Q17 Q18 Q19 Q20 Q21 Q22 Q23 Q24 Q25 Q26
2 HCT299 5 5 PDG_L6
OE1 U5 IO4 Y6 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188
P5D 3 15 4 PDG_L7
OE2 IO5 P5D Y7
1 4 18 PDG_L8 R41 R53 R52 R43 R45 R44 R49 R48 R47 R46 R51 R50
DG11
DG10
DG0
DG1
DG2
DG3
DG4
DG5
DG6
DG7
DG8
DG9
S0 IO6 Y8
19
19 16 17 PDG_L9
180R 180R 180R 180R 180R 180R 180R 180R 180R 180R 180R 180R
PDG_L11
PDG_L10
S1 IO7 Y9
PDG_L0
PDG_L1
PDG_L2
PDG_L3
PDG_L4
PDG_L5
PDG_L6
PDG_L7
PDG_L8
PDG_L9
17 R60 20 PDG_L10
Q7 10K0 HEF4514 Y10
P5D 18 1% U2 19 PDG_L11
CLR GND DS7 Y11 L Q1 L Q3 L Q4 L Q5 L Q6 L Q7 L Q8 L Q9 L Q10 L Q11 L Q12 L Q14
9 1 14 PDG_L12 DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA
EL Y12
9 10
23 13
E Y13
16
Y14
R61 15
Y15
P5D P5D 10K0 GND
D1 HCT132 1% DG[0:12]
12
GND R2
U1 10K0
1%
MAP PULSE
BAT54C
R39
Q2
2SB1188 SYSTOLIC
200R LED13 LED10 LED16
DMR20A-1 RED DMR20A-1 RED DMR20A-1 RED LED29 LED28
R42 DG12
180R SA0 2 SA0 2 SA0 2 LTC2621JD RED LTC4624Y YELLOW
C1 A A A
PDG_L12
SA1 15 SA1 15 SA1 15 S8 R104 15 S0 R111 12
R40 100NF B a B a B a A/L1 A/L1
470K SA2 13 SA2 13 SA2 13 S9 R98 620 12 S1 R105 200 11
C C C B/L2 a a a B/L2 a a a
50V L Q13 SA3 11
f b
SA3 11
f b
SA3 11
f b
S10 620 R103 6 L3 S2 200 R110 3 L3
D1 D1 D1 C/L3 C/L3
DTC143ZKA 9 g 9 g 9 g S11 R99 620 1
f b f b f b
S3 R106 200 4
f b f b f b
D2 D2 D2 D L1 D L1
5 5 5 S12 620 R102 4 g g g S4 200 R109 2 g g g
E1 e c E1 e c E1 e c E E
SA4 7 SA4 7 SA4 7 S13 R100 620 16 S5 R107 200 15
E2 E2 E2 F e c e c e c F e c e c e c
P5D P5D d d d 620 R101 200 R108
CONN_S_2X5_HDRF SA5 3 SA5 3 SA5 3 S14 7 L2 S6 8 L2
PBAT C9 F F F G G
HCT132 DP DP DP
d d d d d d
SA6 14 SA6 14 SA6 14 620 3 200 6
V_IN G G G DP DP
DP DP DP DG6 1 DP DP DP
10 100NF 10 10 10 DG0 2 DG1
PBAT DP DP DP DIG1 DG7 5 DG2
9 GND 4 4 4 DG1 5
SPI_CLK P5D U1 COM COM COM DIG2 DG8 7
12 12 12 DG2 138 DIG3
8 HCT132
HCT132P5D
SPI_MOSI COM COM COM DIG3 14
R55 L1,L2,L3 L1,L2,L3
7 17 17 17
SW_PWR COM COM COM
6 R131 100R S0 R14 SA0
DG0
DG1
DG2
CS_DISPLAY GND
5 100R U1 GND S1 R23 330 SA1
P5D P5D U1
4 S2 330 R24 SA2
SW_MUX SA[0:6]
3 S3 R25 330 SA3
GND
2 S4 330 R26 SA4 LED9 GURX10D
RED
1 S5 R27 330 SA5 DIASTOLIC SPO2
LED17 LED18 LED19 1 20
J1 S6 330 R28 SA6
DMR20A-1 RED DMR20A-1 RED DMR20A-1 RED LED27
330 LTC4624JD RED 2 19
SA0 2 SA0 2 SA0 2
A A A S0 R97 12
SA1 15
B a SA1 15
B a SA1 15
B a
A/L1 3 18
S1 R37 330 11
P5D SA2 13 SA2 13 SA2 13 B/L2 a a a
C
f b
C
f b
C
f b S2 330 R96 3
C/L3
L3 4 17
needs pull up on main board SA3 11 SA3 11 SA3 11 f b f b f b
C4 D1 D1 D1 S3 R57 330 4
9
D2
g 9
D2
g 9
D2
g D L1 5 16 R119 S15
S4 330 R95 2 g g g
100NF 5 5 5 E 200
E1 e c E1 e c E1 e c S5 R58 330 15
F
6 15 R120 S14
SA4 7 SA4 7 SA4 7 e c e c e c
E2 E2 E2 S6 330 R94 8 L2 200
R56 SA5 3
F
d SA5 3
F
d SA5 3
F
d G
d d d
7 14 R121 S13
20 DP DP DP 330 1 6
100R DP 200
SA6 14 SA6 14 SA6 14 DG9 DP DP DP
P5D VCC
IO0
7 PS0 G G G 5 DG1 8 13 R122 S12
10 10 10 DG10 DG2
11 13 PS1 DP DP DP 7 200
DS0 IO1 4
COM
4
COM
4
COM
DG11
DIG3
9 12 R123 S11
12 6 PS2 14
CLK IO2 12 12 12 200
8
Q0 IO3
14 PS3 COM COM COM L1,L2,L3 10 11 R124 S15
HCT299 17 17 17
2 5 PS4 COM COM COM 200 S[0:15]
OE1 U4 IO4
DG11
DG3
DG4
DG5
DG6
3 15 PS5
OE2 IO5
1 4 PS6
19 S0 IO6
19 16 PS7
S1 IO7 AUTO/TIME
Q7
17
TEMP LED11
CLR 18 LED30
GND DS7 LTC2623JD RED
P5D 9 LTC4627JD RED
9 10 S8 R118 13
S8 R36 14 A/L1
A S9 R112 620 15
S9 R29 330 16 B/L2 a a a a
B a a a a S10 620 R117 2 L3
S10 330 R35 13 C/L3
C R113 620 f b f b f b f b
f b f b f b f b S11 6 L1
S11 R30 330 3 D
P5D D S12 620 R116 5 g g g g
C5 S12 330 R34 5 g g g g E
E S13 R114 620 16
S13 R31 330 11 F e c e c e c e c
100NF F e c e c e c e c S14 620 R115 7 L2
S14 330 R33 15 G
20 G 620 3 d d d d
S15 R32 330 7 d d d d DP DP DP DP DP
VCC 7 PS8 DP DP DP DP DP DG3 1
P5D IO0 330 DG7 1 DIG1
11 13 PS9 DIG1 DG4 14
DS0 IO1 DG8 2 DIG2
12 6 PS10 DIG2 DG5 11
CLK IO2 DG9 6 DIG3
8 14 PS11 DIG3 DG6 8
Q0 IO3 DG10 8 DIG4
2 HCT299 5 PS12 DIG4 DG11 12
OE1 U3 IO4 L1,L2,L3
3 15 PS13
OE2 IO5
1 4 PS14
19 S0 IO6
19 16 PS15
S1 IO7
17
Q7
CLR 18
GND DS7
P5D 9
9 10
PS[0:15]
DTC143ZKA is a digital transistor with a 4K7 base resistor and a 47K base pull-down resistor
DTD123EK is a digital transistor with a 2K2 base resistor and a 2K2 base pull-down resistor S[0:15]
S10
S9
S11
S12
S13
S14
S15
S1
S2
S3
S4
S5
S6
S7
S8
S0
Schematic -
L Q27
DTD123EK
L Q28
DTD123EK
L Q29
DTD123EK
L Q30
DTD123EK
L Q31
DTD123EK
L Q32
DTD123EK
L Q33
DTD123EK
L Q34
DTD123EK
L Q35
DTD123EK
L Q36
DTD123EK
L Q37
DTD123EK
SCHEMATIC L Q38
DTD123EK
L Q39
DTD123EK
L Q40
DTD123EK
L Q41
DTD123EK
L Q42
DTD123EK
User Interface Board
P/N 2008741 Rev B
PS10
PS11
PS12
PS13
PS14
PS15
PS0
PS1
PS2
PS3
PS4
PS5
PS6
PS7
PS8
PS9
(1 of 2) page 5-25/26
PROCARE USER INTERFACE BOARD
PCB BOARD SWITCH LOCATIONS
SILENCE 2 5 INFLATE
PBAT P5D
ALARM 1 3 CYCLE
50V
100NF
50V
100NF
50V
47U0
16V
47U0
16V
47U0
16V
- 6 8 PRINT
MENU 4 9 ON/OFF
V_IN
DG[0:12]
DG3
LED37 DG4
PDG_L[0:11] APTD3216QYC
GREEN
CHARGING DG5
R126
10K0 R135
1% 10K0
LED21 1% R136
APTD3216QYC 10K0
GREEN 1%
LED56
APTD3216SYCK LED54 LED15 LED1
SW9 YELLOW AP3216SURC APTD3216QYC APTD3216QYC
SW_PWR A B BATT BATT RED 2 GREEN GREEN
LED66 LOW LOW BELL BATTERY BATTERY
LEFT RIGHT BATTERY MIDDLE LEFT
ON/OFF APTD3216QYC RIGHT
GREEN LED62 LED61
AP3216SURC AP3216SURC LED69 LED60 LED55 LED68 LED57 LED5
CHARGING RED RED APTD3216SYCK AP3216SURC 1 APTD3216QYC APTD3216QYC
APTD3216SYCK APTD3216QYC
degF degC YELLOW YELLOW RED GREEN GREEN GREEN
LED65
APTD3216QYC
GREEN R20 R18 R75
SW_MUX 38R3 38R3 R133 R77 R134 R16 R74
38R3 22R 38R3
22R 22R 22R
SW1
BAT54C
A B
PDG_L1
ALARM DG12
SW2
A B LED53 LED40 LED52 LED50 LED49 LED46 LED44 LED42
D6 SILENCE PDG_L2 LED67 AP3216SURC AP3216SURC AP3216SURC AP3216SURC AP3216SURC AP3216SURC AP3216SURC AP3216SURC
AP3216SURC RED RED RED RED RED RED RED RED
RED PULSE PULSE SpO2 SpO2 DIA DIA SYS
LOW HIGH LOW HIGH LOW HIGH LOW SYS
minus HIGH
R125
10K0 LED38 LED39 LED41 LED51 LED48 LED47 LED45 LED43
BAT54C SW3 1% AP3216SURC AP3216SURC AP3216SURC AP3216SURC AP3216SURC AP3216SURC AP3216SURC AP3216SURC
RED RED RED RED RED RED RED RED
A B PDG_L3 R64
CYCLE 38R3
SW5
BAT54C A B
INFLATE PDG_L5
DG0
DG1
SW6
DG2
A B
D4 PDG_L6 DG12
MINUS
Schematic -
User Interface Board
P/N 2008741 Rev B
(2 of 2) page 5-27/28
PROCARE USER INTERFACE BOARD
0086
2009381-001 A
gemedical.com