zTROUBLE SHOOTING PROCEDURE
zFUNCTIONS SETTING-UP SEQUENCE
Description:
1. : Weight unit
3. : Capacity
4. : Calibration weight
5. : Division
6. : Gravity compensation
zTHE PROCEDURE FOR CHANGING LOAD CELL
START
Start to calibrate.
END
zCALIBRATION PROCEDURE
A
START
Display shows 0
End
A
Note 1:
2. If the zero count value is out of range, please adjust the R1, R2 of the main board (CH-0525)
to obtain correct zero count.
3. If the span count value is out of range, please adjust the RX of the main board (CH-0525) to
obtain correct span count.
zGRAVITY COMPENSATION:
1. When the operator enables from the function setting procedure describe on page
3, the scale allows the operator to change the setting of gravity value as follows:
(1) Press and hold UNIT key 3 seconds to enter gravity compensation mode, the display will
show 9.789 (the gravity value of Taiwan factory).
(2) Enter the desired gravity value. (Press TARE key to move to the desired digit, and press
UNIT key to set the gravity of destination. At last, press TARE key to confirm the
entered gravity.)
(3) The scale will do gravity compensation automatically.
2. If the operator disables from the function setting procedure described on page 3,
the scale will not allow the operator to re-calibrate the scale by simply changing the setting of
gravity value.
z THE VALUE OF GRAVITY AT VARIOUS LOACTIONS