Sie sind auf Seite 1von 32

SERVOMEX OxyDetect

CERTIFICATION MANUAL

Part Number: 05311008B


Revision: 4
Language: UK English
Blank
CONTENTS

1 INTRODUCTION

2 EU DECLARATION OF CONFORMITY

3 ATEX CERTIFICATION (SGS BASEEFA)

4 IECEx INTERNATIONAL CERTIFICATION (SGS BASEEFA)

5 NORTH AMERICAN HAZLOC CERTIFICATION (SGS)

6 NORTH AMERICAN GP CERTIFICATION (SGS)


Blank
SECTION 1 INTRODUCTION

The 05311B1 OxyDetect has been approved by various authorities for installation in
hazardous and non-hazardous areas. The certification offered depends upon the analyser
specification and country of delivery. See the certification labels attached to the analyser
for exact approval details. The following information summarises the approvals that have
been obtained.

This certification manual covers:

1. 05311B1 OxyDetect

The installer and/or the user of the analyser must be satisfied that the installation complies
with any “Special Conditions for Safe Use” or “Schedule of Limitations” contained in the
following Certificates.

If the analyser is installed in a hazardous area then electrical power and other connecting
cables should be installed and protected according to conditions on the relevant
certificates, national wiring regulations, hazardous area codes of practice and local
authority regulations.

About this manual

Ref: 05311/008B/4

Order as part no: 05311008B


Blank
SECTION 2 EU DECLARATION OF CONFORMITY
Blank
Blank
SECTION 3 ATEX CERTIFICATION

Certificate No.: Baseefa15ATEX0165X

05311B OxyDetect has been certified for:

Rated Supply Voltage 10 - 28Vdc

II 3G Ex ic ec nC IIC T4 Gc (-5°C ≤ Ta ≤ +45°C)

Special Conditions of Safe Use:

1. Possible electrostatic risk - clean only with a damp cloth.


2. The equipment is not capable of passing the dielectric strength test prescribed by
the standards, and so this must be taken into account during installation.
3. The equipment shall be used in an area not exceeding pollution degree 2 as defined
in IEC 60664-1.
Blank
Blank
SECTION 4 IECEx INTERNATIONAL CERTIFICATION

Certificate No.: IECEx BAS 15.0112X

05311B OxyDetect has been certified for:

Rated Supply Voltage 10 - 28Vdc

Ex ic ec IIC T4 Gc (-5°C ≤ Ta ≤ +45°C)

Special Conditions of Safe Use:

1. Possible electrostatic risk - clean only with a damp cloth.


2. The equipment is not capable of passing the dielectric strength test prescribed by
the standards, and so this must be taken into account during installation.
3. The equipment shall be used in an area not exceeding pollution degree 2 as defined
in IEC 60664-1.
Blank
SECTION 5 NORTH AMERICAN HAZLOC CERTIFICATION (SGS)

Certificate No: SGSNA/19/SUW/00043

05311B OxyDetect has been certified for:

Model Number: 05311B1

Rated Supply Voltage 10 - 28VDC, 3W Max

Class I, Division 2, Groups A-D, T4


Class I, Zone 2 Group IIC, T4
Ambient temperature range -5°C to +45°C,
IP66
Blank
SECTION 6 NORTH AMERICAN GP CERTIFICATION (SGS)

Certificate No: SGSNA/19/SUW/00067

05311B OxyDetect has been certified for:

Model Number: 05311B1

Rated Supply Voltage 10 - 28VDC, 3W Max

Ambient temperature range -5°C to +45°C

IP66
Blank