Table of Contents
3
GDS-2000 User Manual
Measurements ..............................................64
Automatic Measurements...........................65
Go-No Go Test..........................................70
Math Operation ........................................74
Program and Play .....................................76
Trigger ....................................................78
Calibration ..................................................114
Calibrate the vertical scale ....................... 114
Compensate the probe ............................ 115
FAQ .............................................................117
Appendix.....................................................119
Specifications......................................... 119
Declaration of Conformity ........................ 122
Index ..........................................................123
4
Safety Instructions
Safety Instructions
This chapter contains important safety
instructions that you must follow when
operating GDS-2000 and when keeping it in
storage. Read the following before any operation
to insure your safety and to keep the best
condition for GDS-2000.
Fuse ........................................................ 7
Operation Environment............................... 8
5
GDS-2000 User Manual
Safety Symbols
These safety symbols may appear in this manual or on GDS-2000.
Safety Guidelines
6
Safety Instructions
opening.
7
GDS-2000 User Manual
into GDS-2000.
8
Safety Instructions
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
Green/ Yellow: Earth
Blue: Neutral
Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the colours
marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with the letter E or by the earth symbol or coloured Green
or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is
marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked
with the letter L or P or coloured Brown or Red.
9
Getting Started
Getting Started
Follow these instructions to properly setup
GDS-2000, especially if you are using it for the
first time.
11
GDS-2000 User Manual
GDS-2000 characteristics
GDS-2000 is a generic purpose digital storage
oscilloscope suitable for wide range of
applications such as production testing,
research, and field verification.
• FFT analysis
• Go-No Go test
• Built-in help
12
Getting Started
Package Contents
Check the contents before using GDS-2000. Contact your local
dealer in case there is a missing item.
Opening the
box
Ω Ω
G D S -2 1 0 4
2. Probe set
X1 X10
GDS-2062: GTP060A x 2
GDS-2064: GTP060A x 4
GDS-2102: GTP150A x 2
10M /1M
GTP-200
X10 / X1
GDS-2104: GTP150A x 4
GDS-2202: GTP250A x 2
GDS-2204: GTP250A x 4
3. Power cord
13
GDS-2000 User Manual
Power Up
Tilt stand
Red (Standby)
Green (On)
14
Getting Started
Functionality check
Before operating GDS-2000 in a new
environment, run these steps to make sure it is
functionally stable.
Ω Ω
G D S -2 1 0 4
X1
X1
0
x1 x10
15
GDS-2000 User Manual
3. Set the scale Press Auto Set key Auto Set . GDS-2000
automatically adjusts the horizontal scale,
vertical scale, and trigger level.
For Auto Set details, see page65.
Over Under
Normal
Compensation Compensation
16
Panel Descriptions
Panel Descriptions
17
GDS-2000 User Manual
Front Panel
K Vertical
position
knob
L CH1~CH4
(Vertical)
G D S -2 1 0 4
Ω Ω
menu key
M Volts/
Div knob
S Probe R Menu Q USB P Math O Ground N Input
Compensation ON/OFF Connector key Terminal Terminal
Output key CH1~CH4
I Horizontal
position
knob
G Trigger
level knob
F Trigger
menu key
H Horizontal
G D S -2 1 0 2
Ω Ω
Menu key
J Time/
Div knob
S Probe R Menu Q USB N Input P Math O Ground T External
Compensation ON/OFF Connector Terminal key Terminal Trigger
Output key CH1~CH2 Input
18
Panel Descriptions
19
GDS-2000 User Manual
R Menu On/Off Show (On) or hide (Off) the menu from the
key display. See page53.
20
Panel Descriptions
Rear Panel
21
GDS-2000 User Manual
22
Panel Descriptions
Display
G Function
A Waveforms menu
Trigger
Level
Indicator
H Input
Signal
J Channel Time/Div I Trigger Frequency
Status Status
23
GDS-2000 User Manual
I Trigger Status
(From left) Trigger source channel, trigger
type, and slope
For trigger details, see page78.
J Channel Status
(From left) Channel, Bandwidth limit,
Coupling mode, Time/Div scale
For Channel (vertical scale) details, see
page56.
24
Quick Reference
Quick Reference
Calibration .............................................. 28
Trigger ................................................... 34
Utility (1 of 3) ......................................... 35
Utility (2 of 3) ......................................... 36
Utility (3 of 3) ......................................... 37
25
GDS-2000 User Manual
Operation Shortcuts
Here is the list of operations introduced in this manual and their
key shortcuts.
Symbols description
Acquisition
Select the Acquisition mode Acquire→F1~F4
Select the memory length Acquire→F5
Cursor
Select the horizontal cursor Cursor→F1~F2
Select the vertical cursor Cursor→F1, F3
Display
Freeze the waveform Run/Stop
Refresh the display view Display→F3
Select the display grid Display→F5
Select the vectors/dots waveform Display→F1
Set the display contrast Display→F4
Turn Off the display menu Menu ON/OFF
View accumulated waveform Display→F2
Horizontal
Zoom the horizontal view HORIMENU→F2~F3
Roll the horizontal view HORIMENU→F4
View in XY mode HORIMENU→F5
Vertical
Invert the waveform CH1/2/3/4→F2
Limit the frequency bandwidth CH1/2/3/4→F3
Select the coupling mode CH1/2/3/4→F1
Select the probe attenuation CH1/2/3/4→F4
Other Configurations
Select the buzzer sound Utility→F3
26
Quick Reference
Automatic Measurements
Delay automatic measurements Measure→F5→F3
Set the scale automatically Auto Set
Time automatic measurements Measure→F3→F3
View all the measurement results Measure→Measure→F1~F4
Voltage automatic measurements Measure→F1→F3
Go-No Go Test
Edit Go-No Go test template Utility→F3→F2~F3
Utility→F3→F1→F1~F4
Utility→F5→F4
Run Go-No Go test Utility→F5→F3→F4
Math Operation
Add/ Subtract/ Multiply/ Divide MATH→F1 →F2~F4
Run FFT operation MATH→F1 →F2~F5
Program and Play
Edit the program steps Program→F1 →F2~F5
Play the program Program→F1 →F2~F5
Trigger
Use Delay trigger Trigger→F1 →F2~F4→F5→
F1~F4
Use Edge trigger Trigger→F1 →F2~F3→F5→
F1~F4
Use Pulse width trigger Trigger→F1 →F2~F4→F5→
F1~F4
Use Video trigger Trigger→F1 →F2~F5
Printout
Printout display image/waveform Utility→F1 →F1
Hardcopy
Save and Recall
Quick save to USB Utility→F1→F1
Hardcopy
Save all (image/setup/waveform) Save/Recall→F5→F2→F1~F4
Save image Save/Recall→F5→F1→F1~F4
27
GDS-2000 User Manual
Remote Control
Calibration
28
Quick Reference
Menu Tree
No menu for the following keys: Auto Set, Run/Stop, Help, Auto
test/Stop, Hardcopy.
Acquire
Normal F 1
Peak
F 2
Detect
2/ 4/ 8/ 16/ 32/
Average F 3
64/ 128/ 256
Equ.Time/
Sampling F 4
Real Time
/
Horizontal F 2 Accumulate F 2 On/ Off
/
/
Vertical F 3 Refresh F 3
/
Contrast F 4
(icon) F 5 / /
29
GDS-2000 User Manual
HORI Operation F 1 +/ –/ x/ ÷
MENU
(4CH)
F 1 (CH1_CH2) F 2 CH1_CH2/
Main
CH3_CH4
(2CH)
Window F 2
CH1_CH2
Window
F 3 Position F 4 -12div ~ +12div
Zoom
XY F 5
FFT F 1
(4CH) CH1/ 2/
3/ 4
Source F 2 (2CH) CH1/ 2
Flattop/
Rectangular/
Window F 3 Blackman/
Hanning
Measure Measure
Vpp
F 1 CH1 F 1
:
Vavg
F 2 CH2 F 2
:
Frequency CH3
F 3 F 3
: (4CH model)
Duty Cycle CH4
F 4 F 4
: (4CH model)
Rise Time
F 5 OFF F 5
:
30
Quick Reference
Measure
(4CH) CH1/ 2/
Vpp 3/ 4
F 1 Source1 F 1 (2CH) CH1/ 2/
:
Vavg (4CH) CH1/ 2/
F 2 Source2 F 2 3/ 4
:
(2CH) CH1/ 2/
Frequency
F 3
: Delay Type F 3 FRR/ FRF/
Duty Cycle FFR/ FFF/
F 4 LRR/ LRF/
: (icon) F 4
Rise Time LFR/ LFF
F 5
: Back F 5
Program
Edit F 1 Play F 1
Step F 2 1 ~ 20 Cycle F 2 1 ~ 99
Menu/Memory/ 1 ~ 20
Item F 3 From:/ To: F 3
Time (From = To)
Save F 5 Start F 5
31
GDS-2000 User Manual
Save/ Recall (1 of 2)
Save/Recall Display
F 1
Refs.
Default
F 1 Ref A F 2 On/ Off
Setup
Display
F 2 Ref B F 3 On/ Off
Refs.
More F 5
Save Setup F 1
Ref: A/ B Save F 4
Destination F 3 Memory: 1 ~ 20
USB File Utilities
F 5 To File Utilities
(USB only)
Save F 4
File Utilities
F 5 To File Utilities
(USB only)
File Utilities
Enter
Select F 1 F 1
Character
Back
New Folder F 2 F 2
Space
Rename F 3
Delete F 4 Save F 4
Previous Previous
F 5 F 5
Menu Menu
32
Quick Reference
Save/ Recall (2 of 2)
To File Utilities
Recall Recall
F 1 F 1
Waveform Setup
Memory: 1 ~ 20 Memory: 1 ~ 20
Source F 2 Source F 2
USB USB
(4CH) Ref A/ B/
Destination F 3
C/ D
(2CH) Ref A/ B
Recall F 4 Recall F 4
33
GDS-2000 User Manual
Trigger
MENU (Trigger)
Auto/ Normal/
Mode F 3 Rejection F 3 LF/ HF/ Off
Single/ Auto Level
Noise
F 4 On/ Off
Rejection
Slope/ Previous
F 5 F 5
Coupling Menu
Video F 1
(4CH) CH1/ CH2/
CH3/ CH4
Source F 2 (2CH) CH1/ CH2
NTSC/
Standard F 3
SECAM/ PAL
Polarity F 4 /
Field 1/ Field 2
Line F 5
1~263 (NTSC) 1~313 (SECAM/ PAL)
Pulse F 1 Slope F 1 /
(4CH) CH1/ 2/ 3/ 4
Source F 2 (2CH) CH1/ 2 Coupling F 2 /
Auto/ Normal/
Mode F 3 Rejection F 3 LF/ HF/ Off
Single/ Auto Level
>/ </ =/ ? Noise
When F 4 F 4 On/ Off
20ns~200us Rejection
Slope/ Previous
F 5 F 5
Coupling Menu
Delay
F 1 Slope F 1 /
(2CH model)
34
Quick Reference
Utility (1 of 3)
Utility
Hardcopy SaveImage/
F 1 Function F 1
Menu SaveAll/ Printer
Interface
F 2 Ink Saver F 2 On/ Off
Menu
Type Color Landscape/
(Sound) F 3 F 3 Color Portrait/
(Printer only)
Gray Landscape/
/ Gray Portrait
/ Off Ratio
F 4 5 ~ 95
(Printer only)
Language
F 4 Previous
English F 5
Menu
More F 5
Type F 1 RS232/ USB/ GPIB
Parity
F 4 Odd/ Even/ None
(RS232 only)
Previous
F 5
Menu
Self CAL
F 1 Vertical F 1
Menu
System Trigger
F 2 F 2
Info. Level
Go-No Go Delay
F 3 F 3
Menu Time
No Go
F 4
When
More F 5
35
GDS-2000 User Manual
Utility (2 of 3)
Utility
More F 5 More F 5
Template
F 1
Edit
(4CH) CH1/ 2/ 3/ 4
Source F 2
(2CH) CH1/ 2
STOP
Violating F 3
STOP+
Continue
Continue+
Ratio: F 5
36
Quick Reference
Utility (3 of 3)
37
GDS-2000 User Manual
Default Settings
Save/Recall
These are the factory installed settings that
appear when pressing Save/Recall key→F1
Default
F 1 (Default Setup).
Setup
38
Configure the Settings
View in XY mode...................................... 55
39
GDS-2000 User Manual
40
Configure the Settings
Acquisition
Acquisition process samples the analog input signal and converts
it into digital format, later to be reconstructed into waveform.
Panel operation
Normal F 1
Peak
F 2
Detect
Acquire
2/ 4/ 8/ 16/ 32/
Average F 3
64/ 128/ 256
Equ.Time/
Sampling F 4
Real Time
41
GDS-2000 User Manual
is recorded.
Equ. Time Equivalent Time sampling.
GDS-2000 draws the waveform by
accumulating the sample records.
Use this only for repetitive signal.
Real Time Real Time sampling. GDS-2000
draws the waveform from a single
sample record.
Example
Average mode
42
Configure the Settings
500/25000 (1CH)
Panel operation Acquire F 5 500/12500 (2CH)
Mem Leng
500/5000 (4CH)
43
GDS-2000 User Manual
Note that the display always shows 250 points (300 points when
menu is turned off).
44
Configure the Settings
Cursor
/
Horizontal F 2
/
45
GDS-2000 User Manual
Range Source
CH1~CH4 Channel1~Channel2 waveform
(4CH model)
46
Configure the Settings
/
Vertical F 3
/
47
GDS-2000 User Manual
Range Source
CH1~CH4 Channel1~Channel4 waveform
(4CH model)
48
Configure the Settings
Display
Panel operation
Display Type F 1 Vectors/ Dots
Example
Vectors Dots
49
GDS-2000 User Manual
Panel
Display Accumulate F 2 On/ Off
operation
Refresh F 3
Example
50
Configure the Settings
Panel operation
Display Contrast F 4
51
GDS-2000 User Manual
Panel operation
Display (icon) F 5 / /
Example
52
Configure the Settings
Panel operation
Example
53
GDS-2000 User Manual
Horizontal View
Panel operation
Window F 2
HORI
MENU
Window
F 3
Zoom
Range 1ns~1ms
54
Configure the Settings
View in XY mode
55
GDS-2000 User Manual
Vertical
Panel operation
CH1 Coupling F 1 / /
Range AC coupling
DC coupling
Ground coupling
DC Coupling AC Coupling
56
Configure the Settings
Panel operation
CH1 Invert F 2 On/ Off
Example
57
GDS-2000 User Manual
Panel operation
CH1 BW Limit F 3 On/ Off
Panel operation
CH1 Probe F 4 x1/ x10/ x100
Range x1 No attenuation
58
Configure the Settings
Other Settings
Panel operation /
Utility F 3
/ Off
59
GDS-2000 User Manual
Range English
Chinese Traditional
Chinese Simplified
60
Configure the Settings
Panel operation
Time Set
F 2 Date F 1
Menu
Save F 4
Cancel F 5
61
GDS-2000 User Manual
Panel operation
Time Set
F 2 Time F 1
Menu
Save F 4
Cancel F 5
62
Configure the Settings
System
Utility More F 5 F 2
Info.
When not in use Take the batteries out of the unit to prolong the
battery life.
63
GDS-2000 User Manual
Measurements
64
Measurements
Automatic Measurements
Auto Set
Trigger Coupling: DC
Position: Center
Slope: Positive
Type: Edge
Source: Highest frequency
Level: Midpoint of data for the trigger
source
65
GDS-2000 User Manual
Panel operation
Measurement
Edit mode
Result
Vpp (4CH) CH1/ 2/ 3/ 4
F 1 Source1 F 1
: (2CH) CH1/ 2
Vavf (4CH) CH1/ 2/ 3/ 4
F 2 Source2 F 2
: (2CH) CH1/ 2
Frequency Vpp, Vmax, Vmin,
Measure F 3 Volt Type F 3
: Vamp, Vhi, Vlo,
Dutycycle Vavg, Vrms,
F 4 (Icon) F 4 ROVShoot,
:
RiseTime FOVShoot,
F 5 Back F 5 RPREShoot,
:
FPREShoot
66
Measurements
Range Source1, 2
(4CH) CH1~CH4 (4CH model) Channel1~Channel2
Volt type
Vpp Difference between positive and
negative peak voltage.
(=Vmax−Vmin)
Time Type
Freq Frequency of the waveform.
67
GDS-2000 User Manual
Delay Type
68
Measurements
Vpp
F 1 CH1 F 1
:
Vavf
F 2 CH2 F 2
:
Frequency CH3
Measure F 3 F 3
: (4CH only)
Dutycycle CH4
F 4 F 4
: (4CH only)
RiseTime
F 5 OFF F 5
:
69
GDS-2000 User Manual
Go-No Go Test
Panel operation
Go-No Go
F 3
Menu
Utility More F 5
No Go
F 4 /
When
Template
F 1 Template F 1 Max/ Min/ Auto
Edit
Max:RefA/M1~20
Source F 2 CH1/ CH2 Source F 2 Min:RefB/M1~20
Auto:CH1/CH2
STOP
Violating F 3 STOP+
Position/
F 3 Max:-12~+12Div
Continue Tolerance Min:-12~+12Div
Cont.+ Save & Auto:0.04~4Div/
F 4 0.4%~40%
Create
Previous
F 5
Menu
70
Measurements
Template
Max Sets the maximum side of the
template.
Template source
RefA: Reference waveform that has to
be stored in advance.
M1~20: Templates stored in advance
in internal memory.
To store a waveform (template)
internally, see pageError! Bookmark
not defined..
Template position
±12/Div
Template source
RefB: Reference waveform that has to
be stored in advance.
M1~20: Templates stored in advance
in internal memory.
To store a waveform (template)
internally, see pageError! Bookmark
not defined..
Template position
±12/Div
71
GDS-2000 User Manual
Tolerance
Source signal
CH1 Channel1 as the subject of test
Violation Condition
Stop The test stops in case of violation.
72
Measurements
Go-No Go
Utility More F 5 F 3 Go-No Go F 4
Menu
On/ Off
Ratio
F 5
0:0
73
GDS-2000 User Manual
Math Operation
Panel operation
(+) F 1 +/ -/ x/ ÷
(2CH) CH1_CH2
(CH1+CH2) F 2 (4CH) CH1_CH2/
MATH CH3_CH4
Channel Pair
CH1_CH2 Math operation between Channel1
and Channel2
CH3_CH4 Math operation between Channel3
and Channel4
(only for 4CH model)
Position
-12Div~+12Div
74
Measurements
Panel operation
FFT F 1
Flattop/Rectangular
MATH Window F 3
Blackman/Hanning
Position
-12Div~+12Div
Amplitude scale
1, 2, 5, 10, 20 dB/Div
75
GDS-2000 User Manual
Panel operation
Edit F 1
Step F 2 1 ~ 15
Menu: Cursor/
Program Item F 3
AutoMeasurement
Memory: M1~20
Time: 1~99 (sec)
Save F 5
Item
Menu “AutoMeasure” or “Cursor”.
Memory M1~M20 predefined waveforms.
To store waveforms, see pagexx.
Time 1~99 seconds for each step.
76
Measurements
Panel operation
Play F 1
Cycle F 2 1 ~ 99
Start F 5
77
GDS-2000 User Manual
Trigger
Panel operation
(2CH)
CH1/2/
External/
Edge F 1 Line Slope F 1 /
(4CH)
CH1/2/3/4/
(Trigger) Source F 2 Line Coupling F 2 /
MENU Auto/
Mode F 3 Rejection F 3 LF/ HF/ Off
Normal/
Single/ Noise
Auto Level F 4 On/ Off
Rejection
Slope/ Previous
F 5 F 5
Coupling Menu
78
Measurements
Trigger mode
Auto GDS-2000 generate an internal
trigger if there is no trigger event.
Select this mode when viewing
rolling waveform at slower
timebase, maximum 10s/div.
Normal GDS-2000 acquire waveform in a
trigger event.
Single GDS-2000 acquire waveform only
once in a trigger event. Press
Run/Stop key to acquire again.
Auto Level GDS-2000 automatically adjust the
trigger level indicator to the center
part of the waveform.
Slope
Rising edge
Falling edge
Coupling
AC coupling
DC coupling
(Frequency) Rejection
LF Low Frequency rejection. Rejects
frequency below 50kHz.
HF High Frequency rejection. Rejects
frequency above 50kHz.
Off Rejection disabled
Noise Rejection
ON Uses DC coupling with low
sensitivity to reject noise.
OFF Noise rejection disabled
79
GDS-2000 User Manual
Panel Video F 1
operation
(4CH) CH1/ 2/ 3/ 4
Source F 2
(2CH) CH1/ 2
MENU
Standard F 3 NTSC/ SECAM/ PAL
Polarity F 4 /
5. To select the trigger field line, press F5. Then use the Variable
knob .
Video standard
NTSC National Television System
Committee video standard.
PAL Phase Alternative by Line video
standard.
SECAM SEquential Couleur A Memoire
video standard.
Polarity
Positive pulse
Negative pulse
Video Field
1 ~ 263 For NTSC
1 ~ 313 For PAL/ SECAM
80
Measurements
Panel operation
81
GDS-2000 User Manual
Trigger mode
Auto GDS-2000 generates an internal
trigger if there is no trigger event.
Normal GDS-2000 acquires waveform in a
trigger event.
Single GDS-2000 acquire waveform only
once in a trigger event. Press
Run/Stop key to acquire again.
Auto Level GDS-2000 automatically adjusts
the trigger level indicator to the
center part of the waveform.
Slope
Trigger on the positive pulse width
Trigger on the negative pulse width
Coupling
AC coupling
DC coupling
(Frequency) Rejection
LF Low Frequency rejection. Rejects
frequency below 50kHz.
HF High Frequency rejection. Rejects
frequency above 50kHz.
Off Rejection disabled
Noise Rejection
On Uses DC coupling with low
sensitivity to reject noise.
Off Noise rejection disabled
82
Measurements
Panel operation
Delay F 1 Slope F 1 /
100ns~
By Time F 2 Coupling F 2 /
1.3ms
TTL:1.48V Noise
Ext: F 4 ECL:-1.35V F 4 On/ Off
Rejection
User:±12V
Slope/ Previous
F 5 F 5
Coupling Menu
3. To set the delay time, press F2. Then use the Variable knob .
4. To set the number of trigger event, press F3. Then use the
Variable knob .
By Event
2 ~ 65000
83
GDS-2000 User Manual
Slope
Rising edge
Falling edge
Coupling
AC coupling
DC coupling
(Frequency) Rejection
LF Low Frequency rejection.
Rejects frequency below
50kHz.
HF High Frequency rejection.
Rejects frequency above
50kHz.
Off Rejection disabled
Noise Rejection
On Uses DC coupling with low
sensitivity to reject noise.
Off Noise rejection disabled
84
Measurements
D
A: Start Trigger (External)
A
B: Main Trigger (CH1 or 2)
1 2 3 C: Start point of External
B trigger count
D: Trigger
C E E: Trigger point
85
GDS-2000 User Manual
Printout/ Data
Transfer
86
Printout/ Data Transfer
Printout
Panel operation
Hardcopy Function
Utility F 1 F 1
Menu Printer
Color Landscape
(Color) F 3 Color Portrait
Gray Landscape
Gray Portrait
Ratio F 4 5 ~ 75
HardCopy
4. To select the image size, press F4. Then use the Variable
knob .
87
GDS-2000 User Manual
Color/ Portrait
Color
Landscape
Color Portrait
Gray
Landscape
Gray Portrait
Ratio (Image size)
5~75
The following printers have been confirmed.
Confirmed HP Deskjet 970CXI
Printers HP LaserJet 1010/ 1015/ 1300
Epson AL-C8600
Use the printout function in the software. In this way, you can
print out the image to any printer as long as your PC recognizes it.
See page97 for software setup.
88
Printout/ Data Transfer
Save/ Recall
Panel operation
Hardcopy Save Image/
Utility F 1 Function F 1
Menu Save All
HardCopy
4. Connect the USB flash drive to the front or rear panel USB
connector. (Make sure the other USB connector is not used.
The two cannot work together at the same time.)
89
GDS-2000 User Manual
Panel operation
Save Setup F 3
Save Save
Save/Recall F 4 F 1
Waveform Image
InkSaver
F 2 On/ Off
(setup,image)
(4CH) CH1/ 2/ 3/ 4,
Math, RefA/ B/ C/ D
Source (2CH) CH1/ 2,
F 2
(Waveform) Math, RefA/ B
90
Printout/ Data Transfer
Source
CH1~CH4 Channel1 ~ Channel4 waveforms
Destination
91
GDS-2000 User Manual
Panel operation
Enter
Save/Recall ... F 1 Select F 1 F 1
Character
Back
... F 2 New Folder F 2 F 2
Space
... F 3 Rename F 3
Previous Previous
File Utilities F 5 F 5 F 5
Menu Menu
See the folder 1. Use the Variable knob to select the folder.
contents
2. To enter the folder, press F1.
Folder
Selected
92
Printout/ Data Transfer
93
GDS-2000 User Manual
Panel operation
Recall
F 3
Setup
Save/Recall More F 5
Recall
F 4
Waveform
Recall F 1
Memory: M1 ~ 20
Source F 2
USB
Destination (4CH) RefA/ B/ C/ D
F 3
(waveform) (2CH) RefA/ B
Recall F 4
File Utilities
F 5
(USB only)
Ref C
F 4 On/ Off
(4CH only)
Ref D
F 5 On/ Off
(4CH only)
4. (Recalling from USB flash drive) Connect the flash drive to the
front or the rear USB connector. (Make sure the other USB
94
Printout/ Data Transfer
Source
Memory M1~M20 internal memories
Destination
Ref A/B/C/D (4CH Reference waveforms
models) stored internally.
Ref A/B
(2CH models)
95
GDS-2000 User Manual
96
Remote Control
Remote Control
Use PC Software Configure the interface ............................. 97
Use PC Software
Interface Icon
97
GDS-2000 User Manual
Download
98
Remote Control
Full
Screen
button
Quit
button
99
GDS-2000 User Manual
Connection tab
2. Capture the Click the Play mode tab. Then click the Play
waveform button. GDS-2000 display appears on the
screen. Command history pane shows the record
of issued commands.
Play mode tab
Command
history pane Play button
100
Remote Control
101
GDS-2000 User Manual
Panel operation
Interface
Utility F 2 Type F 1 USB/ RS232/GPIB
Menu
Address
F 2 0 ~ 30
(GPIB only)
Baud Rate 2400/ 4800/ 9600/
F 2
(RS232 only) 19200/ 38400
Stop Bit
F 3 1/ 2
(RS232 only)
Parity
F 4 Odd/ Even/ None
(RS232 only)
102
Remote Control
1 2 3 4 5
6 7 8 9
*idn?
Parity (RS232C)
Odd/ Even/ None
Address (GPIB)
1~30
103
GDS-2000 User Manual
Command Syntax
Example
command
1: Command Header
2: Single space
3: Parameter
4: Message Terminator OR Message Separator
Parameter <0/1> 0 or 1.
example
<1~4> Integer between 1, 2, 3, or 4.
104
Remote Control
Command Set
Acquisition
Automatic measurements
105
GDS-2000 User Manual
channel first)
:meas:ris? Returns the first pulse rising edge timing. (select
the channel first)
:meas:sour Sets the measurement channel.
<1~4> Range: 1~4 (Channel1~Channel4)
:meas:vamp? Returns the voltage difference between positive
and negative peak. (select the channel first)
:meas:vav? Returns the average voltage. (select the channel
first)
:meas:vhi? Returns the global high voltage. (select the
channel first)
:meas:vlo? Returns the global low voltage. (select the
channel first)
:meas:vmax? Returns the maximum amplitude. (select the
channel first)
:meas:vmin? Returns the minimum amplitude. (select the
channel first)
:meas:vpp? Returns the difference between maximum and
minimum amplitude. (select the channel first)
:meas:vrms? Returns the root mean square voltage. (select
the channel first)
Cursor
106
Remote Control
Display
Go-No Go test
107
GDS-2000 User Manual
Horizontal scale
108
Remote Control
Trigger
109
GDS-2000 User Manual
110
Remote Control
111
GDS-2000 User Manual
Registers manipulation
112
Remote Control
113
GDS-2000 User Manual
Calibration
Panel operation
Self Cal
Utility More F 5 F 1 Vertical F 1
Menu
Trigger
F 2
Level
Delay Time F 3
Ω Ω
G D S -2 1 0 4
114
Calibration
Panel operation
Probe Comp
F 1 Wave Type F 1 / /
Menu
Frequency F 2 1k ~ 100k
Default 1k F 4
Ω Ω
G D S -2 1 0 4
X1
X10
x1 x10
115
GDS-2000 User Manual
Over Under
Normal
Compensation Compensation
Frequency
1k~100k 1k step.
Duty Cycle
5%~95% 5% step.
116
FAQ
FAQ
● I pressed the Power (On/Standby) key on the front panel but
nothing happens.
I pressed the Power (On/Standby) key on the front panel but nothing
happens.
117
GDS-2000 User Manual
Make sure you are not using the front and the
rear USB connector at the same time.
Disconnect either of the USB device and try
again.
The accuracy does not match the specification.
118
Appendix
Appendix
Specifications
The specifications apply under the following conditions: GDS-2000
is powered on for at least 30 minutes, within +20°C~+30°C.
GDS-2062/64 GDS-2102/04 GDS-2202/04
Channels 2/4 2/4 2/4
Bandwidth DC~60MHz DC~100MHz DC~200MHz
(−3dB) (−3dB) (−3dB)
Rise Time 5.8ns approx. 3.5ns approx. 1.75ns approx.
GDS-2062/2064/2102/2104/2202/2204
119
GDS-2000 User Manual
120
Appendix
Accuracy: ±2%
Signal source: All available trigger
source except the Video trigger
121
GDS-2000 User Manual
Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 95-11, Pao-Chung Rd., Hsin-Tien City, Taipei Hsien, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69, Lushan Road, Suzhou New District Jiangsu, China
declare that the below mentioned product
GDS-2000
is herewith confirmed to comply with the requirements set out in the
Council Directive on the Approximation of the Law of Member States relating
to Electromagnetic Compatibility (89/336/EEC, 92/31/EEC, 93/68/EEC)
and Low Voltage Equipment Directive (73/23/EEC, 93/68/EEC). For the
evaluation regarding the Electromagnetic Compatibility and Low Voltage
Equipment Directive, the following standards were applied:
◎ EMC
EN 61326-1: Electrical equipment for measurement, control and laboratory use
–– EMC requirements (1997+A1: 1998+A2: 2001)
Conducted and Radiated Emissions Electrostatic Discharge
EN 55011: 1998 class A EN 61000-4-2: 1995+A1:1998
Current Harmonic Radiated Immunity
EN 61000-3-2: 2000 EN 61000-4-3: 1996+A1:1998
Voltage Fluctuation Electrical Fast Transients
EN 61000-3-3: 1995 EN 61000-4-4: 1995
------------------------- Surge Immunity
EN 61000-4-5: 1995
------------------------- Conducted Susceptibility
EN 61000-4-6: 1996
------------------------- Power Frequency Magnetic Field
EN 61000-4-8 : 1993
------------------------- Voltage Dips/ Interrupts
EN 61000-4-11: 1994
◎ Safety
Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC
Safety Requirements
IEC/EN 61010-1: 2001
122
Index
Index
EN61000 ............................................... 122
A
EXT input
acquisition ...............................................41
description .......................................... 18
Auto set
trigger ................................................. 83
configuration.......................................65
F
FAQ.................................................... 117
functionality check .............................16 FAQ ....................................................... 116
automatic measurement fft operation ............................................ 75
configuration....................................... 66 frequency limit........................................ 58
view result...........................................69 front panel .............................................. 18
functionality check ................................. 15
B
fuse
battery
safety instruction ................................. 7
FAQ............................................ 116, 117
specification...................................... 121
maintenance........................................63
G
C
GDS-2000
Calibration ............................................ 113
cleaning............................................. 5, 7
coupling
feature list .......................................... 12
channel ................................................56
specifications .................................... 119
trigger..................................................78
go-no go
cursor
edit ...................................................... 70
horizontal ............................................45
run....................................................... 73
vertical.................................................47
GPIB
D IEEE remote interface..................... 101
display H
contents description ...........................23
help, built-in ........................................... 59
contrast ...............................................51
horizontal
FAQ.................................................... 116
cursor .................................................. 45
grid ......................................................52