Beruflich Dokumente
Kultur Dokumente
Model 3340
Laser Aerosol Spectrometer
Description of the 3
Model 3340 Model 3340
Laser Aerosol Spectrometer
Model 3340 Operation 4
Maintenance 6
Calibration 7
Troubleshooting 8
Appendixes
Manual History
Revision Date
Revision A October 2009
iii
Trademarks TSI, TSI logo are registered trademarks of TSI Incorporated.
Microsoft, Windows, are registered trademarks of Microsoft Corporation.
LabVIEW is a registered trademark of National Instrument Corporation.
HyperTerminal is a trademark of Hilgraeve, Inc.
Swagelok is a registered trademark of Swagelok Company of Solon, Ohio, USA.
Celeron is a registered trademark of Intel Corporation.
Q-tips is a registered trademark of Chesebrough-Pond's Inc.
Sensidyne and Gilibrator are trademarks of Sensidyne, Inc.
There are no user serviceable parts inside the instrument. Refer all
repair and maintenance to a qualified technician. All maintenance
and repair information in this manual is included for use by a
qualified technician.
W A R N I N G
The use of controls, adjustments, or procedures other than those
specified in this manual may result in exposure to hazardous optical
radiation.
W A R N I N G
High voltage is accessible in several locations within this instrument.
Make sure you unplug the power source before removing the cover or
performing maintenance procedures.
v
Labels
The Model 3340 has the following labels as shown in Figure 1.
Figure 1
Location of Warning and Information Labels
Caution
C a u t i o n
Caution means be careful. It means if you do not follow the procedures
prescribed in this manual you may do something that might result in
equipment damage, or you might have to take something apart and start
over again. It also indicates that important information about the operation
and maintenance of this instrument is included.
Warning
W A R N I N G
Warning means that unsafe use of the instrument could result in serious
injury to you or cause irrevocable damage to the instrument. Follow the
procedures prescribed in this manual to use the instrument safely.
Safety vii
Contents
ix
Configuration Tab ............................................................. 4-14
Calibration Tab ................................................................. 4-14
Collecting Data .................................................................... 4-15
Data File Format ............................................................... 4-16
CHAPTER 5 Theory of Operation ......................................... 5-1
Instrument Subsystems ......................................................... 5-1
Optical System .................................................................... 5-2
Flow System ........................................................................ 5-4
Analog Electronics............................................................... 5-5
Digital Electronics System ................................................... 5-6
On-board PC ....................................................................... 5-8
Particle Coincidence ............................................................. 5-10
CHAPTER 6 Maintenance .................................................... 6-1
Cleaning Optics ..................................................................... 6-1
Laser High Voltage Supply .................................................. 6-1
Laser Safety Information ..................................................... 6-2
Laser Bench Cleaning ......................................................... 6-2
Inlet Jet .............................................................................. 6-5
CHAPTER 7 Calibration....................................................... 7-1
Calibration Mode Controls ..................................................... 7-2
Configuration Tab ............................................................... 7-2
Map Tab Calibration Controls ............................................. 7-4
Calibration Tab ................................................................... 7-5
Calibration ............................................................................. 7-7
Example: Adding a Calibration Point for 0.269 µ m PSL
Particles ........................................................................... 7-7
Set up the Map ................................................................... 7-8
Initial Sampling................................................................... 7-9
Entering the Calibration Voltage ....................................... 7-10
Gain Stitches .................................................................... 7-13
CHAPTER 8 Troubleshooting ............................................... 8-1
APPENDIX A Model 3340 Specifications ..............................A-1
APPENDIX B Using Serial Data Commands ..........................B-1
Pin Connectors ...................................................................... B-1
Baud Rate .............................................................................. B-2
Format (8-Bits, No Parity) ...................................................... B-2
Stop Bits and Flow Control .................................................... B-2
ASCII Data Output ................................................................. B-2
Data File and Output Format ................................................. B-3
APPENDIX C Computer Related Issues ................................C-1
Regional Settings and LabVIEW® Software .............................. C-1
Procedure to Allow use of Regional Settings with
Non-Period Local Decimal Points ...................................... C-4
Remote Desktop Operation..................................................... C-8
Figures
1 Location of Warning and Information Labels ....................... vi
1-1 Model 3340 Laser Aerosol Spectrometer ........................... 1-1
3-1 Front Panel of the Model 3340 Laser Aerosol
Spectrometer .................................................................... 3-1
3-2 Inlet with Two Guards ...................................................... 3-2
3-3 Back Panel of the Model 3340 Laser Aerosol
Spectrometer .................................................................... 3-2
5-1 Block Diagram of Optical Particle Spectrometer ................ 5-2
5-2 Side View of Optical Block ................................................ 5-3
5-3 Top View of Optical Block ................................................. 5-4
5-4 Schematic Diagram of Flow System .................................. 5-5
5-5 Block Diagram of Analog Electronics ................................ 5-6
5-6 Example Calibration Curve ............................................... 5-9
6-1 Laser Bench Cleaning ....................................................... 6-3
6-2 Laser Bench Optiocs ......................................................... 6-3
6-3 Disconnecting Ferrule Nut ................................................ 6-4
6-4 Inlet Jet ............................................................................ 6-5
6-5 Inserting Tubing into Jet .................................................. 6-6
C-1 Distorted Calibration Curve ............................................. C-3
C-2 Normal Appearance of a Typical Model 3340
Calibration Curve (requires calibration password
to access)......................................................................... C-3
Contents xi
Tables
2-1 Accessories Packing List ................................................... 2-1
8-1 Troubleshooting Symptoms and Recommendations .......... 8-1
A-1 Specifications of Model 3340 ............................................ A-1
Purpose
This is an operation and service manual for the Model 3340 Laser
Aerosol Spectrometer.
Getting Help
To obtain assistance with this product or to submit suggestions,
please contact:
TSI Incorporated
500 Cardigan Road
Shoreview, MN 55126 USA
Fax: (651) 490-3824
Telephone: 1-800-874-2811 (USA) or (651) 490-2811
E-mail Address: technical.service@tsi.com
xiii
Submitting Comments
TSI values your comments and suggestions on this manual. Please
use the comment sheet, on the last page of this manual, to send us
your opinion on the manual‘s usability, to suggest specific
improvements, or to report any technical errors.
If the comment sheet has already been used, send your comments
to:
TSI Incorporated
500 Cardigan Road
Shoreview, MN 55126
Fax: (651) 490-3824
E-mail Address: particle@tsi.com
Product Description
The Model 3340 Laser Aerosol Spectrometer, shown in Figure 1-1,
is a high sensitivity laser particle size spectrometer designed for
sampling and counting airborne particulates from 90 nm to 7.5 µm
using a patented contamination resistant Helium-Neon active cavity
laser. The spectrometer is operated via a built-in computer utilizing
the Windows® XP operating system running an executable
LabVIEW® ―Virtual Instrument (VI)‖ interface to provide instrument
control, data display, recording and output. Each instrument
includes a 10‖ color LCD flat panel display, USB keyboard and
Mouse, 40GB HDD, 3.5‖ FDD and CD-ROM drives. Also provided
are a 10/100 Ethernet interface and 9 pin RS-232 interface.
Figure 1-1
Model 3340 Laser Aerosol Spectrometer
1-1
Applications
The Model 3340 Laser Aerosol Spectrometer has application in the
following areas:
Inhalation toxicology
Atmospheric studies
Ambient air monitoring
Drug-delivery studies
HEPA/ULPA Filter testing and characterization
Indoor air-quality monitoring
Biohazard detection
Basic research
Characterization of test aerosols used in particle-instrument
calibration
Performance evaluations of other aerodynamic devices
Packing List
Table 2-1 provides a packing list of all items that should have been
shipped to you as the Model 3340 and accessory kit. Please
compare the list to the items you received. If any items are missing,
notify TSI immediately.
Table 2-1
Accessories Packing List
Qty Description
1 Model 3340 Laser Aerosol Spectrometer
1 Model 3340 Laser Aerosol Spectrometer Manual
1 Microsoft Windows® XP Distribution Disk
1 Microsoft® Office Software License
1 Line Cord
1 USB Mini Keyboard
1 USB Mouse
36 in. Tubing 1/16” × 1/8”
6 in. Tubing 1/8” × 1/4"
1 Zero Count Inlet Filter Assembly
2 Spare Fuses 5 × 20 mm 2A
1 Calibration Certificate
2-1
Mounting the Sensor
The Model 3340 Laser Aerosol Spectrometer requires no special
mounting requirements other than the ventilation requirements
(see below). The cabinet has four non-marking rubber feet that give
the instrument a good grip on clean, level surfaces and two
additional feet mounted on retractable legs at the front of the unit
that allow the unit to be angled for user convenience.
Ventilation Requirements
The Model 3340 cabinet is designed to be cooled by room air drawn
in from the bottom of the cabinet and exhausted through the back
of the cabinet.
Power Connection
Connect the AC power cord (supplied) to the IDC power input
module on the back of the Model 3340 and then into an available
power outlet. It is not necessary to select the correct voltage, the
spectrometer accepts line voltage of 85 to 260 VAC, 50 to 60 Hz,
200 W, single phase. The connection is self regulating.
This chapter describes the front panel, back panel, and internal
components of the Model 3340 Laser Aerosol Spectrometer.
Front Panel
The two components of the front panel are the color LCD display
and a combination 3½‖ floppy disk/CD-ROM/USB port unit. The
keyboard and mouse attach via the two USB ports.
The color LCD display is used in combination with the mouse and
keyboard as the main interface to the unit.
Figure 3-1
Front Panel of the Model 3340 Laser Aerosol Spectrometer
3-1
The 640 480 pixel LCD display provides continuous real-time
display of sample data, access to the various operating menus
(tabs) and the Windows® XP operating system.
Inlet
The inlet on the top of the unit is protected by two guards to avoid
damage (Figure 3-2). While these guards may look like handles, the
unit should not be picked up by these guards. Typically the unit
ships with the inlet ―zero-count‖ filter attached to one of the guards.
Figure 3-2
Inlet with Two Guards
Back Panel
As shown in Figure 3-3, the back panel of the Model 3340 allows
for power and data connections. The back panel also has a pump
exhaust port and a fan with fan guard.
Figure 3-3
Back Panel of the Model 3340 Laser Aerosol Spectrometer
Pump Exhaust
Sample aerosol is exhausted through the Exhaust Port.
Serial Port
The Serial Port is a standard RS-232 serial connection that allows
communications between the spectrometer‘s internal computer and
an external computer. In the standard ―virtual instrument‖, when
the ―run‖ button on the histogram tab is toggled, the unit
downloads at the histogram sample rate (user selectable) at
115,200 baud, 8, N, 1. The data format is given in the first two lines
of the data file, subsequent lines are unit data.
Internal Components
The location of the functional systems and electronics of the
Model 3340 are shown in Figure 3-4 and include:
Digital PC board Power Supply
Analog PC boards Laser Optical Block
Flow Strut & Filters Detector PC board
Power PC board
Figure 3-4
Internal Diagram of the Model 3340 Laser Particle Spectrometer
This chapter describes how to set up and operate the Model 3340.
These instructions assume that the Model 3340 is connected to an
AC power source, the power on/off switch on the back panel has
been switched to the ―On‖ position and that the computer has
finished booting. The instrument contains a computer which runs
the Windows® XP operating system. It is assumed that the
instrument user is familiar with the normal operation of this
operating system on a computer. The default user name entered at
the factory is TSIINC and the default password is 3340. Users may
create additional logins to suit their needs.
4-1
2. Once the virtual instrument has loaded, click on the Controls
tab (this is typically the default mode) and verify the laser
reference is between 1.0 and 2.8 volts.
With the Zero-Count filter attached to the inlet tubing, the unit
should count fewer than one particle per 5 minutes within 30
minutes of power application. Normally ―Zero-Count‖ operation
is achieved within a few minutes of power application.
Controls Tab
This tab shows the current sample flow rate , the sample flow
rate setting window , the current sheath flow rate , the current
laser reference voltage and the ambient pressure and
temperature.
The sheath flow is factory preset and is not adjustable from the
virtual instrument.
The ―Commit‖ button will flash yellow any time the map
parameters have been changed. In addition, there are graphical
illustrations of the size scale and number of channels.
Note: The default minimum and maximum sizes may vary from the
actual loaded map.
In the following example, the minimum size was set to 90 nm, the
maximum size was set to 189 nm and the ―Linear‖ button was
clicked causing the virtual instrument to generate an evenly spaced
map of 99 size channels between 90 and 189 nm (1 nm bins) and
the Commit button changed to a blinking yellow mode to
indicate changes have been made to the map but they have not
been downloaded into the digital electronics.
The map limits may be set anywhere in the units operating range
(90 to 7500 nm) such as the 90 to 189 nm range in the example
above.
Note: The minimum and maximum settings are integer numbers and
although the size bins are calculated to 2 decimal points it is
generally not useful to set bin widths less than 1nm.
When loading map files ensure that an actual map file is selected
before clicking OK as the virtual instrument does not validate the
selected file. After loading a map file, make sure to set the number
of size bins to the correct value. The number displayed in the size
channel control ( in the figure above) has to match the number of
the last size channel (31 in the figure above). If the value for the
number of size channels is not set correctly, the acquired spectrum
will not be displayed correctly when choosing size as the x-axis.
Note: Saving and Loading maps changes the virtual instrument
default map pointer to the selected map which will be
automatically loaded the next time the virtual instrument
is started.
Map files can also be created and edited using a text editor such
as Notepad.
Histogram Tab
The Histogram tab will be the most utilized screen in the virtual
instrument as it has the sample controls, histogram data display
controls, and recording control.
Record Control
This control is discussed in the Collecting Data section.
The ―Operator‖ mode is the default startup mode and will be used
for majority of the time. The ―Calibration‖ mode is accessible by
entering the calibration password into the password field then
clicking Enter Password. This is discussed in Chapter 7,
Calibration.
Calibration Tab
The Calibration tab is visible but grayed-out as it is not accessible
in Operator mode. This is discussed in Chapter 7, Calibration.
Instrument Subsystems
The TSI Model 3340 is an optical-scattering laser-based aerosol
particle spectrometer system for accurately and precisely sizing
particles in the range from 90 nm to 7.5 µ m in diameter. It uses
fully user-specified size binning of up to 100 channels anywhere
within its size range.
5-1
4. The digital electronics system for analyzing particle signals,
binning signals according to user-specified bin mappings and
generating a histogram of number of particles in the specified
bins, and for communicating with the PC and system
monitor/control functions.
5. An onboard PC running Windows and a specialized application
GUI for instrument control, setup and data reporting and
collection.
Figure 5-1
Block Diagram of Optical Particle Spectrometer
Optical System
The Optical system consists of
The laser and associated components and optics.
The detection system, including collection optics and
photodetectors and reference monitoring.
Mechanical housing for above.
Detection System
The detection system consists of two pairs of Mangin collection
optics capable of collecting light over a large solid angle. The
Mangins image the volume of space at which the flow intersects the
laser mode onto a photodiode. There are two pairs of collecting
optics: one pair images onto an Avalanche Photo Diode (APD) for
detecting the smallest particles (the primary scattering detection
system). The other pair (located on the opposite side of the block)
images onto a low-gain PIN photodiode for detection of the upper
size range of the instrument (the secondary scattering detection
system). Each detector is amplified in a current-to-voltage stage
which feeds the analog electronics system. The system can detect
particles as small as 90 nm (≥50% efficiency, <01 count/ 5 minute
dark count rate). The system size sensitivity is limited by several
noise sources: a fundamental noise process from the photon shot
noise on the detected molecular scatter from background gas, a
fundamental noise process from the Johnson noise in the
photodiode transimpedence feedback resistor and from technical
noise of various sources.
Mechanical Housing
The laser and detection optics are built into a sealed mechanical
enclosure (the optical block, Figure 5-2 and Figure 5-3).
Figure 5-2
Side View of Optical Block
Flow System
The mechanical laser mount forms a sealed block around the laser
and the input/output jets. A pump draws on an exhaust jet pulling
flow through the inlet jet and across the laser mode. The input jet
is an aerodynamically focused assembly with a sample nozzle of
500-µ m diameter and a sheath nozzle of 760-µ m diameter. The tip
of the sheath jet sits close to the edge of the laser mode. Sample
flows are between 5 and 100 sccm and the sheath flow is typically
650 sccm. Particle velocity depends on sheath flow rate, but is on
the order of 50 to 100 m/s. The particles are confined to a region
of space whose extent is limited to a fraction of the laser mode size.
This yields a sizing resolution of approximately 5% of the
particle size.
Analog Electronics
The analog chain converts the photocurrent of the detector
photodiodes to a voltage and processes that signal (called the
particle signal). The chain is repeated for the primary and
secondary detection systems.
The particle signal is fed into two different AC gain stages, differing
in gain as specified below. In total there are four gain stages: high
and low for each of the primary and secondary detection systems.
Gain ratios:
G3/G2 = 50
G2/G1 = 20
G1/G0 = 20
Note: The gain ratios G3:G2 and G1:G0 are pure electrical
amplification gain ratios. The G2:G1 ratio is more complicated
since it involves two independent photodetectors with
independent electronics and on opposite sides of the optical
block. See the discussion in Chapter 7, Calibration.
The gain stages also provide low-pass filtering to the signal. Each
gain stage then feeds its own baseline restoration circuit, which
restores the 0 Volt base line which is disturbed by frequent particle
signals after AC coupling. The particle signal is then passed to a
peak hold circuit which tracks the rise of the photo-signal as a
particle crosses the laser and holds the peak value. The digital
system then processes the signal and when complete, issues a reset
to the baseline restore and peak hold electronics.
Figure 5-5
Block Diagram of Analog Electronics
Monitoring/Control
The digital system also provides monitoring and control of onboard
systems. It reads and sets the mass flow controller for monitoring
and control of the sample flow. It reads the electronic flowmeters for
the sheath flow (flow is controlled by a mechanically-actuated
needle valve). The laser reference from the reference photodiode is
sampled on an ADC and read in the digital electronics module.
Additional housekeeping parameters such as case temperature and
ambient barometric pressure are also monitored. All parameters
which are read by the digital system are logged with the sample
data. All parameters which are set are stored in configuration files.
Calibration
Calibration is an important process for any particle spectrometer
instrument. The Model 3340 with its high resolution and large
number of arbitrarily settable bins poses unique challenges in this
area. Several features have been added to this instrument to make
the calibration process as easy and accurate as possible.
In addition to the relative gains is the calibration curve, that is, the
shape of the particle signal size (in volts) and the particle size (in
nm). Once the relative gains are known the corrected response for
the entire instrument can be formed from the calibration curve. The
calibration curve has several distinct regions. Since the wavelength
of the instrument is 633 nm, it is expected that all particles below
Figure 5-6
Example Calibration Curve
In principle, if all the relative gains are known accurately, and the
calibration curve is known, the instrument need only be calibrated
in an absolute sense at one point—any point in fact. In practice it is
best to use a trusted particle or a few trusted particles. For factory
calibration, the particle sizes used are a nominal 0.100 µ m
(100 nm) for Gain 3, 0.200 µ m (200nm) for Gain 2, 0.500 µm
(500 nm) for Gain 1 and 3 µ m for Gain 0. In addition the transition
region in Gain 1 is further defined by nominal 0.9 µ m (900 nm) and
1 µ m (1000 nm) particles. The specific particles used during
calibration may be verified on the unit calibration certificate.
In some cases, the user may have preferred particles to use for
calibration. In this case as many particles as needed may be used.
If the particles do not all fall on the preset instrument calibration
curve, the calibration curve is altered slightly to ensure that the
calibration particles will return a result which is the stated size of
the particle. The data representing signal size for a given particle
size is entered in the UI and is referred to as calibration points.
Note that alteration of the calibration curve from the preset may be
Particle Coincidence
Particle coincidence is typically defined as more than one particle in
the viewing volume of the particle counter creating a signal that
causes the counter to incorrectly classify the particles as a single,
mis-sized particle. Coincidence typically increases somewhat
linearly with particle concentration until the saturation limit is
reached. Above this point instrument operation is unpredictable.
This, along with particle transit time and processing time is
factored into the 3000 particle/second count limit. Coincidence
may be reduced by reducing the instrument flow rate and/or
diluting the sample.
Note: Changing the lower limit of the bin map to exclude small
particles from the histogram does not affect the total number of
particles the instrument has to process and has no effect on
coincidence. This is also why the count limit flag may be
displayed even though relatively few particles are being
counted in the histogram.
The Model 3340 does not normally require maintenance beyond the
occasional optical cleaning. The optical system is designed to
minimize contamination and it is not unusual for units to operate
for years without any significant contamination. In addition, the
unit is designed with automatic gain control (AGC) which allows
continued accurate particle sizing with varying laser power.
Cleaning Optics
Before starting this operation, read the following safety information
and become familiar with the warning and caution labels found on
the instrument.
6-1
The anode wire from this supply is attached to the laser tube anode
connector. This part of the laser is covered. There is a sticker on the
anode cover denoting the danger here.
To clean the optics you will need the following tools and supplies:
3/32 inch hex driver for cleaning port screws
PH1 (Phillips #1) screwdriver for lid screws
Cotton swabs (“Q-tips®” or Medical Grade recommended)
Reagent, Spectroscopic or Analytic Grade acetone
Maintenance 6-3
Cleaning Procedure
1. Remove the 16 flat Phillips-head cover screws and remove the
unit cover.
2. Attach the zero count filter to the inlet jet tubing.
3. Put the virtual instrument in the controls tab. Monitor the
laser reference display during the optics cleaning procedure.
4. Disconnect the Ferrule nut that connects the input tubing of
the flow assembly to the exhaust jet of the optical bench.
Figure 6-3
Disconnecting Ferrule Nut
Inlet Jet
The alignment of the inlet jet is critical to the performance of the
Model 3340. While it is ―locked down‖ it is still possible to move it
under the right circumstances. For this reason, the inlet jet is
completely recessed inside the unit in order to protect it from
bumps, scrapes overtorquing and mishandling.
Figure 6-4
Inlet Jet
The inlet is brought out through the cover port with the supplied
0.063‖ ID/0.125‖ OD tubing. Additional 0.125‖ tubing is supplied
with the instrument along with 0.125‖ ID/0.25‖ OD tubing for
making splices.
In the event that the user needs to change the inlet tubing they will
need to remove the 16 cover screws and the cover in order to access
the Swagelok® fitting on the inlet jet.
Maintenance 6-5
The tubing MUST be fully inserted into the jet, approximately 1.3‖
(3.3 cm) from the front of the ferrule as shown below before the
knurled ferrule nut is tightened or a ―particle trap‖ may develop in
the jet.
Note: The knurled ferrule nut should be tightened “Finger tight” only
to avoid potential misalignment of the jet.
Figure 6-5
Inserting Tubing into Jet
In the highly unlikely event that the inlet jet becomes misaligned
TSI recommends factory service. Inlet jet alignment requires the
availability of a monodispersed particle generator, 100 nm
calibration particles and knowledge of their use. Additionally, an
oscilloscope may be required in some cases.
Please contact TSI if you feel that the inlet jet may be misaligned.
7-1
Calibration Mode Controls
Configuration Tab
The software starts in the operator mode which does not allow
changes to be made in the calibration. To enter the calibration
mode type the calibration password obtained from TSI in the
password window and click the Enter Password button with
the mouse. Contact TSI Customer Service for the calibration
password.
Note that the User Mode window now indicates ―Calibration‖ and
the former ―Enter Password‖ button is now labeled ―Revert
Mode‖. The ―Save‖, ―Load‖ and ―Revert to Factory Configuration‖
buttons are now active.
Load Button
The ―Load‖ button is used to open an alternate configuration file.
One use for this is when a specific alternate calibration is desired
for a specific test without altering the normal calibration. An
alternate configuration file can be created and loaded for specific
tests. When the tests are complete, the original configuration can
be loaded which will then be loaded when the virtual instrument
is started.
Calibration 6-3
Revert to Factory Configuration Button
―Revert to Factory Configuration‖ is a way to recover if a
configuration file has become corrupted or was saved with
parameters that cause the unit to function incorrectly. A duplicate
copy of the ―as shipped‖ configuration file named factory.cfg is
loaded when this button is clicked. When using this feature it is
very important that you immediately change the filename from
factory.cfg by clicking on the Save button and entering a new file
name. Otherwise, the original settings in the C:\PH\factory.cfg file
will be overwritten and lost when saving subsequent changes.
Calibration 6-5
The above screens are illustrations of the saturation points (red)
that separate the gain stages G3 - G0 and the calibration points
(white) that are displayed in the calibration curve. The bottom two
calibration points in the curve are calculated and not changeable
from the virtual instrument.
Calibration 6-7
Set up the Map
The map is typically set to bins ≈1% of the stated particle size with
the map limits set such that the test particle will be near the center
of the map, keeping in mind that the lowest map setting is 90 nm.
Typically all 99 channels are used although this is sometimes
reduced for particles >2 µ m. In the case of a 269 nm (0.269 µ m)
particle we would use 2.5 nm bins. If we start at 125 nm the
maximum size limit would be 125 + (99 × 2.5) = 372.5 but as we
are limited to integer sizes the upper limit will be set to 373 as
shown here. The map could as easily be set to 130-378 etc.
Calibration 6-9
We will look for the approximate center of the main peak rather
than the channel with the maximum number of counts. Here, the
center of the peak is ≈5000 mV.
To finish creating the calibration point the gain stage, voltage (as
determined by particle data from the voltage histogram) and
particle size will be entered into the appropriate calibration
windows as shown. When finished, the calibration curve should be
(relatively) smooth and the commit button may be pressed to
download the new curve values.
Calibration 6-11
The particles will then be sampled again with the appropriate size
map to verify that the voltage entered was correct. If the peak is not
in the desired location then the voltage is adjusted until the peak is
in the desired location. In this case, the peak was in the desired
location on the first pass.
If the peak channel was too high, the calibration voltage would be
increased until the peak was in the right channel. If the peak was
too low, the calibration voltage should be reduced. This is a trial-
and-error process.
Calibration 6-13
At this point, the ―Commit‖ button can be clicked to download the
new stitch to the electronics.
Note: The “Align” button is for factory technician use and is not
described here.
This shows the same stitch run after completion. Note that the ratio
and offset have changed from the earlier example.
Calibration 6-15
When the stitches have been completed, the results may be saved
by clicking the configuration tab and clicking Save as previously
described.
Table 8-1
Troubleshooting Symptoms and Recommendations
Symptoms Recommendations
Unit does not turn on. Check for good contact between the power cord and
the wall outlet. Check for power at the outlet. Check
fuses (2A 5 × 20) in power inlet.
Fan comes on but Contact TSI for instructions.
computer does not boot.
LabVIEW® virtual Check for recent changes in the operating system
instrument displays error and software installations. Check if any files have
messages on startup. recently been moved, deleted, etc.
The “Warning! The This is an indicator that the concentration of aerosol
particle Count exceeds that the instrument is sampling is too high to
the maximum allowed” accurately measure. This flag is activated when the
flag is active. count rate exceeds 3000 particles/sec. Although the
3340 can measure aerosols at concentrations
greater than this value, concentration errors due to
coincidence will increase and some of the particles
will not be counted. To correct this problem the user
may either reduce the flow rate or dilute the sample.
Note: This flag can be ignored when running
stitches (Calibration mode).
“NAN” is displayed in “NAN” indicates “Not A Number” which is usually
display windows such as indicative of a problem with the configuration file.
pressure, temperature, Check the configuration tab to verify that a valid
etc. configuration file is loaded.
Unit reports low or no See “Cleaning Optics” section for cleaning
reference voltage. instructions. If this does not help or if laser is not
ionizing (pink/red glow visible through slot in anode
cover), contact TSI.
No reference or no Turn the unit off and remove the lid. Check for loose
particle data above monitor cable(s) inside unit. NEVER plug/unplug
Gain 2. internal cabling with the power on! Rarely occurs,
but most likely to occur after rough handling of unit.
8-1
Symptoms Recommendations
Sample flow rate is Verify the external flow measurement has been
incorrect. corrected for pressure and temperature.
Compare inlet flow vs. outlet flow. Differences
greater than ≈0.5 sccm may indicate a leak has
developed. Contact TSI if this occurs.
Possible corrupt configuration file. Enter the
calibration password on the Configuration tab. Note
the current configuration file path and name then
click the Revert to Factory Configuration button
and re-measure flow. If this is correct, click save
and rename the configuration file. If the flow is still
incorrect click load and re-load the original
configuration file and contact TSI.
No sample or sheath Possible Pump failure or Flow PC board failure.
flow. Contact TSI
Table A-1
Specifications of Model 3340
Measurement Technique ...................... Light scattering
Particle Type ........................................ Airborne solids and liquids
Particle Size Range .............................. 0.90 to 7.5 µm optical size (PSL equivalent)
Maximum Particle Concentration .......... 3000 particles per second
Display Resolution ................................ Up to 100 user-defined channels
Resolution ............................................ Within 5% at 0.1 m diameter
Sampling Time ..................................... User Selectable from 1 second to 60 hours, 60 min, 60 sec
Flow Rates ........................................... Total flow: 750 ccm, Sheath flow 650 ccm ±25 ccm, Aerosol Sample
10 to 100 sccm ±2.5% @ 50 sccm. Inlet vs. Outlet flow match ±2.5% @ 60ccm
Atmospheric Pressure Correction ......... Sample flow automatically corrected by internal flow controller
Operating Temperature ........................ 10 to 30°C (50 to 86°F)
Operating Humidity ............................... 10 to 90% RH non-condensing
Operating Altitude ................................. Sea level to 4000 meters (13,000 ft)
Laser Source ........................................ >1W intercavity power 633 nm Helium-Neon gas laser
Detectors .............................................. Avalanche Photo Diode (APD) and PIN photodiode
Front Panel Display .............................. 10” Color, 640 by 480 pixels
Power ................................................... 100 to 240 VAC, 50 to 60 Hz, 200 W, single phase
Communications ................................... 10/100 Ethernet (RJ45 Jack), RS232 (9-pin) port (Output only in standard
operating mode)
Outputs ................................................. RS-232 and Ethernet ports
Dimensions (HWD) ............................... 25.4 cm 43.2 cm 55.9 cm (10 in. 17 in. 22 in.)
Weight .................................................. 24 kg (53 lb.).
Fuse .................................................... 2, 2A 250V 5 20mm in power entry module
Internal fuse not accessible to user 3.5A 250V 3AG hard wired in power supply
TSI and TSI logo are registered trademarks of TSI Incorporated.
A-1
APPENDIX B
Using Serial Data
Commands
This appendix contains information you need if you are writing your
own software for a computer or data acquisition system.
Information includes:
Pin connectors
Baud rate
Parity
Command definitions and syntax.
Pin Connectors
The Model 3340 has a single 9-pin, D-subminiature connector port
on the back panel labeled SERIAL PORT (See Figure 3-3 and
Figure B-1). This communication port is configured at the factory to
work with RS-232 type devices. Table B-1 provides the signal
connections.
5 4 3 2 1
9 8 7 6
Figure B-1
SERIAL PORT Pin Designations
B-1
Table B-1
Signal Connections for RS-232 Configurations
Pin Number RS-232 Signal
1 —
2 Transmit Output
3 Receive Input
4 —
5 GND
6 —
7 —
8 —
9 —
Baud Rate
The baud-rate setting is the rate of communication in terms of bits
per second (baud). The Model 3340 uses a baud rate setting of
115,200. For proper communications, make sure that all software
used with the instrument is set at the appropriate rate.
A typical data file opened in Notepad with extra tabs inserted in the
two header rows to make the columns align correctly (Note: word-
wrap is off). It is generally easier to open data files in a spreadsheet
such as Microsoft® Excel® software but Notepad illustrates the exact
recorded time format whereas Excel® will often change the time
format.
Field Description
Date Date at the end of the record interval
Time Time at the end of the record interval
Accum Secs. Sample Accumulation Time
Scatter Volts Background Light Level
Current Volts N/A in the Model 3340
Sample sccm Sample Flow Rate (measured)
Ref. Volts Laser Reference Voltage
Temp. Volts N/A in the Model 3340
Sheath sccm Sheath Flow Rate (measured)
Diff. Volts N/A in the Model 3340
Box K Internal Temperature Sensor (°K)
Purge sccm N/A in the Model 3340
Pres. kPa Internal Ambient Pressure Sensor
Aux. Volts N/A in the Model 3340
Flow sscm N/A in the Model 3340
90.00 94.11 Size Range & counts in 1st data channel
9411 98.41 Size Range & counts in 2nd data channel
98.41 102.91 Size Range & counts in 3rd data channel
etc.
C-1
Note: Number format is: 123 456 789,00.
Figure C-2
Normal Appearance of a Typical Model 3340 Calibration Curve (requires calibration
password to access)
2. Select Edit>Replace.
5. Save the resulting file using the same name and directory as the
original file, i.e., C:\PH\config.cfg.
6. Open the file C:\PH\LAS-3340_3p3_090519.ini. Highlight the
line useLocalDecimalPt=False as shown here and delete it
(alternatively, you could edit this to read
―useLocalDecimalPt=true‖). Save the file and exit.
For the illustrations here HyperTerminal® was also set to echo locally,
append linefeeds ,and wrap lines.
Start Command
Format: start n <Enter>.
The sample was allowed to finish, then the status command was
repeated. This time the ―Not running‖ response indicates the
sample is finished.
Distribution Command
Format: distribution <Enter>
This command causes the unit to send the sample data to the
control computer in a comma delimited format. In this example
―458‖ is the number of counts in the first channel, ―424‖ is the
number of counts in the second channel, etc. At the end of the
channel data the unit sends the following data: set sample time,
date and time the sample was started, sample flow, sheath flow,
Reference Voltage, Pressure, and Temperature.
Map Command
Format: map <Enter>
This command causes the unit to send the bin map currently in
use to be sent to the control computer.
Invalid Commands
Any command that is not recognized generates an ―?Invalid
Command!‖ response as demonstrated by entering a misspelled
stop command in the example below.
Index-1
laser aerosol spectrometer, 1-1 relative gain, 5-10
laser optical block, 3-4 RS-232
laser reference voltage, 4-4 signal, B-2
laser safety, 6-2 run button, 4-2, 4-10
LCD display, 3-1
load control, 4-6 S
location of warning labels, vi safety, v
sample flow, 4-3
M–N sampling controls, 4-10
maintenance, 6-1 save control, 4-6
calibration, 7-1 schematic, 5-5
manual history, ii sensor, mounting, 2-2
map command, C-19 serial data commands, B-1
map controls, 4-4 baud rate, B-2
map limits, 4-6 format, B-2
map tab, 4-4 pin connectors, B-1
mechanical housing, 5-3 stop bits and flow control, B-2
monitoring, 5-7 serial port, 3-3, B-1
mounting sensor, 2-2 designations, B-1
mouse, USB, 2-1 service policy, iii
setting up, 2-1
O sheath flow, 4-3
on-board PC, 5-8 signal connections for RS-232 configurations, B-2
operation, 1-2, 4-1 solid-state, 6-1
operator mode, 4-14 specifications, A-1
optical block start command, C-17
side view, 5-3 status command, C-17
top view, 5-4 stop bits and flow control, B-2
optical system, 5-2 stop button, 4-2
optics, 5-2 stop command, C-20
cleaning, 6-1, 6-3 submitting comments, xiv
cleaning procedure, 6-4 subsystems, 5-1
over control, 4-8
overview, 1-1 T
temperature sensor, 4-4
P test computer settings, C-16
packing list, 2-1 theory of operation, 5-1
particle coincidence, 5-10 trademarks, iv
particle signal, 5-5 troubleshooting, 8-1
password symptoms and recommendations, 8-1
default, 4-1
entry window, 4-14 U
peak height analysis, 5-7 under control, 4-8
pin connectors, B-1 unit controls, 4-3
power connection, 2-2 unpacking, 2-1
power PC board, 3-4 USB keyboard, 2-1
power supply, 3-4 USB mouse, 2-1
pressure sensor, 4-4 USB port, 3-1
product description, 1-1, 3-1 user mode, 4-14
product registration, iii user name
pulse height analysis, 5-6 default, 4-1
pump exhaust, 3-3
purpose of manual, xiii V
ventilation requirements, 2-2
Q vi. (see virtual instrument)
quick start guide, 4-1 virtual instrument, 1-1, 4-1
R
record control, 4-13
related product literature, xiii
Z
zero-count filter, 4-2
Index Index-3
Reader’s Comments
Please help us improve our manuals by completing and returning this questionnaire to the address
listed in the ―About This Manual‖ section. Feel free to attach a separate sheet of comments.
Manual Title Model 3340 Laser Aerosol Spectrometer P/N 6002729 Rev. A
Contact your local TSI Distributor or visit our website www.tsi.com for more detailed specifications.