Beruflich Dokumente
Kultur Dokumente
The MedGraphics BREEZESUITE PFX Ultima User Manual is Part No. 142152-
001Rev C.
Caution: Federal (USA) law restricts this device to sale by or on the order of a
physician if it is to be used for diagnosis and/or prescribing of therapy.
2
INDEX
3
Stop Collect Volume............................................................................................... 29
Patient Valve Lockout ............................................................................................ 31
FRC TESTING..................................................................................................................... 32
FRC Overview .............................................................................................................. 32
Using the FRC N2 Toolbar ........................................................................................... 33
Using the FRC N2 Data Grid ....................................................................................... 34
FRC Test Procedure..................................................................................................... 35
Instructing the Patient............................................................................................. 35
Pre-Test Procedure .................................................................................................. 35
Test Procedure ......................................................................................................... 35
Repeating FRC N2 Washout Efforts ....................................................................... 36
Selecting FRC (N2) Efforts and Printing Reports....................................................... 38
Using Auto Select .................................................................................................... 38
Manually Selecting Efforts ...................................................................................... 38
Selecting a Report Loop ........................................................................................ 38
Selecting SVC Efforts Done in Spirometry............................................................. 38
Drawing Efforts ......................................................................................................... 38
Printing Reports ........................................................................................................ 38
FRC Options ................................................................................................................. 39
Configuration........................................................................................................... 39
Keystroke to Start Test ............................................................................................. 39
Perform Auto-Select................................................................................................ 39
Individual SVC Maneuvers ..................................................................................... 39
Average IC/ERV ...................................................................................................... 39
Evaluate Reproducibility ........................................................................................ 41
TLC Calculation ....................................................................................................... 41
MIP/MEP ........................................................................................................................... 42
MIP/MEP Overview ..................................................................................................... 42
MIP/MEP Test Bar ......................................................................................................... 42
MIP MEP Data Grid ..................................................................................................... 43
MIP/MEP Testing Procedure....................................................................................... 43
Instructing the Patient............................................................................................. 43
Test Procedure ......................................................................................................... 45
MIP/MEP Options......................................................................................................... 47
Keystroke to Start Test ............................................................................................. 47
Maximum Occlusion ............................................................................................... 47
Measurement Mode............................................................................................... 47
MIP/MEP Test Screen .................................................................................................. 49
VO2 Test Procedure ........................................................................................................ 50
Preparing for a Test..................................................................................................... 50
Test Procedure............................................................................................................. 50
Direct Connect preVent Pneumotach for Ventilator Patients................................. 52
Metabolic Assessment with Ventilators Review...................................................... 52
Example of a patient on Direct Connect for Metabolic/Nutritional Assessment
....................................................................................................................................... 53
Setup and Calibration of the Direct Connect preVent Pneumotach............. 53
Patient Setup ............................................................................................................... 54
Notice to PB 840 Users:............................................................................................ 55
4
References................................................................................................................... 55
Replacing the Drying Line and Cartridge................................................................... 56
Replacing the Drying Sample Line ........................................................................... 56
Replacing the Drying Cartridge ............................................................................... 56
Sample Line Maintenance............................................................................................ 58
Procedure .................................................................................................................... 58
Refreshing The O2 Cell.................................................................................................... 60
Replacing the O2 Cell .................................................................................................... 61
Cleaning CO2 Analyzer.................................................................................................. 63
Required Equipment................................................................................................... 63
Procedure .................................................................................................................... 63
Gases/Regulators/Pressures ...................................................................................... 64
5
START UP AND SHUT DOWN
System Start Up
1. Before turning the power on, plug a power conditioner into the wall
outlet. Connect the power cords to the Ultima system and PF
module, and plug the system and module into the power
conditioner. Ensure all communication cables are securely
attached between Ultima and PF module, as well as the PF arm
cable and tubing connections.
2. Push the main power rocker switch on the back of both systems.
The power light on the front of the Ultima system will turn from
amber to green when the system is completely warmed up.
3. Flip the computer power switch to “On.”
4. Turn the monitor power to “On.” The Windows™ desktop menu
displays on the screen.
5. Open Breeze.
6. Turn on the gas exchange (GX) vacuum pump in Breeze.
7. If using a drying cartridge, check to see if the blue drying crystals
are still evident before calibrating or testing.
8. Wait 30 minutes for the system to warm up and stabilize. Wait 10
minutes for vacuum pump to stabilize and 12 minutes for DLCO
helium (concurrent with the system warm up).
Note: If the system power is turned off for a short duration (less than 15 minutes),
analyzer warm-up time will be double the power-down time.
9. Turn on the printer and other external devices installed with the
system.
10. Open all gas tanks by turning the gas cylinder’s top pressure valve
to the left. The calibration and reference gas for the Ultima system
should be set to deliver 15 psi. Ultima WD tanks are preset to 15 psi
and have only one adjustment knob to open and close. Open the
tanks by turning the black knob counterclockwise.
11. Set the PF module gases to 135 psi for both diffusion (DLCO), and
oxygen (O2). Helium is an adjustable regulator setting between 10 –
25 psi.
12. Calibrate the pneumotach at the start of the day.
13. Automatically calibrate the gas analyzers (O2 and CO2) before
each test.
6
Shutting Down
After Gas Exchange (GX) Analyzer Calibration
Note: Gas exchange calibration (GX Autocal) calibrates the oxygen and carbon
dioxide analyzer (O2 & CO2) that is used for all GX and FRC testing. The calibration
and reference gases should be turned off only if there are no further GX or FRC
tests performed in the next two hours.
Close the Ultima gas tanks (turn the gas cylinder pressure valve to right).
For WD tanks, turn the black knob clockwise on each cylinder after
completing the gas analyzer calibration process. The calibration and
reference gases connected to the Ultima are used only for the calibration
of the O2 and CO2 analyzers for all gas exchange and pulmonary testing.
At end of test
To avoid the 30-minute gas analyzer warm-up and stabilization, allow the
system to remain turned on throughout the day. Prolong the life of the
vacuum pump and analyzers by turning the pump off whenever possible
(the software timer will display “Ready” when the gas auto calibration
can be performed). The system requires a 10-minute stabilization time
when the pump is turned on.
At end of day
The system’s main power switch may remain turned on overnight. To
prolong the life of the oxygen analyzer, turn the vacuum pump off
whenever possible. To prolong the life of the DLCO gas column, turn off the
helium flow whenever possible. Close the gas cylinders by turning the
valve (on top of the tank) clockwise until tight or turn the black knob
clockwise on the WD tanks.
7
SYSTEM CALIBRATION
Pneumotach Calibration
1. Install the pneumotach into the clip and connect it to the rubber
coupler (screen toward coupler).
Note: You must have the patient entry screen closed to select Calibration.
4. The software defaults to 3 liters for the calibration syringe size. If you
are utilizing a different volume syringe, enter this value in place of
the default value of three liters.
8
5. Enter the current temperature, barometric pressure, and humidity
at the bottom of the screen. Verify the units of measurement.
Flow Zero
Note: Within pneumotach calibration the flow should trace downward for all
withdraws and upwards for all injections. If it does not trace downward, the
pneumotach or pneumotach umbilical could be connected incorrectly or the
pneumotach may have been inserted in the syringe incorrectly.
9
Figure 3: Example of syringe
calibration
10
Syringe volume
set-up
Flow zero
11
Gas Analyzer Calibration
The CPX Ultima gas calibration uses an automatic calibration (AutoCal)
feature. AutoCal measures the signals and automatically adjusts the gain
and offset values to accommodate the signal values. Ensure the vacuum
pump has been on for a minimum of 10 minutes and displays “Ready.”
12
An out-of-range result will display in RED. Click “Retry” if a measured
result is displayed in red. If the gas analyzer calibration continues to fail,
verify the following:
Breeze uses AutoCal to measure the gain. AutoCal sets the offset to
achieve optimal min phase percentage values. Breeze may be
indicating a failing O2 cell when:
If the offsets continue to fail in AutoCal, the cal gas readings may be out
of range and may require adjustments in the manual calibration of O2
and CO2 analyzers.
13
Maximal Inspiratory and Maximal Expiratory
(MIP/MEP) Calibration
Note: The MIP/MEP reference pressure may be set to any value less than 500 cm
H2O. The default value is 20 cm H2O (see figure 8).
Figure 8: MIP/MEP
calibration
14
Figure 9: Luer fitting
15
Figure 10: Applying pressure with a manometer
Note: Lung Diffusion (DLCO) does not contain an individual calibration. The
calibration of DLCO testing is performed during the initial system pre-analysis.
16
DLCO
DLCO Overview
The ability to measure lung diffusion (DLCO) is important for a number of
reasons. Most importantly, diffusion capacity is the only non-invasive
means of determining the functional integrity of the alveolar-capillary
membrane.
Note: The helium gas must be turned on at least 12 minutes before attempting a
diffusion test. The helium gas flow is activated from the “Helium” button at the top
of the patient test screen.
17
DLCO Screen
18
DLCO Trace
19
DLCO Data Grid
20
DLCO Toolbar
21
DLCO Test Procedure
DLCO testing is performed using the preVent pneumotach and a
mouthpiece connected to the pneumotach umbilical. The pneumotach
must also be connected to the patient circuit in the arm of the Ultima PF
or PFX.
Note: If both diffusion and nitrogen washout tests are to be conducted, the
diffusion test should be performed first. Inhaling 100% oxygen during a nitrogen
washout test may saturate the hemoglobin and decrease the patient’s diffusion
capacity values.
Pre-Test Procedure
1. Be sure you have warmed up the system and performed the daily
complete pneumotach calibration. Before testing Helium gas must
be on for twelve minutes. Warm up time is displayed on the
associated timer at the top of the screen.
Tip: The volume of He passing through the gas chromatograph column will effect
the column’s deterioration. To prolong the life of the gas chromatograph column,
turn on the He gas no more than 12 minutes before performing a diffusion study.
Turn off the He gas as soon as diffusion testing is completed.
22
The program checks the DLCO analyses generated by your system against
these criteria:
If your system meets these criteria, patient testing may begin. If your
system does not meet these criteria, see Gas Chromatograph Column
Maintenance and Troubleshooting the DLCO Trace.
Test Procedure
1. Before beginning the test, click “Zero Flow” to zero the
pneumotach. There must be no flow through the pneumotach
during this procedure.
2. If you are not testing using the Pre/Post protocol, click the
“Protocol/Log” tab and choose a test protocol from the protocol
pull down menu. Pre/Post is the default protocol for all new visits.
3. Ensure that the pneumotach is attached to the pneumotach
umbilical clip and that the unit is attached to the patient circuit in
the Ultima arm.
4. Place a nose clip on the patient and instruct the patient to begin
tidal breathing through the mouthpiece.
5. Press the spacebar to begin data collection.
6. Instruct the patient to exhale slowly and maximally to residual
volume (RV). While the patient is exhaling, or at the end of this
exhalation, click “Next” or press the spacebar. When the computer
detects zero flow, the patient valve closes for the start of
inspiration. The patient circuit opens to the source gas (diffusion
gas).
7. After the valve system is activated, instruct the patient to rapidly
inhale maximally to total lung capacity, filling the lungs with
diffusion gas. The valve will close automatically when zero
inspiratory flow is sensed.
Note: If the patient begins to inhale before the valve system is activated, the
patient’s inspired volume will not be measured correctly. The numeric data
displayed in the “Vol. Inspired” row will be lower than the trace displayed on the
volume/time graph.
8. When the valve system closes, the patient is locked out for the time
specified in the setup menu. Instruct the patient to relax against the
closed valve for this period. The end of lockout is shown as a
23
vertical dotted line on the volume/time graph. During lockout,
watch the graph to see when the valve is about to open.
9. At the end of lockout, the valve opens. Instruct the patient to
exhale rapidly until the sample is collected.
10. The test ends automatically. Have the patient relax and remove
the nose clips.
24
Automatic Gas Analysis
At the end of each effort, a gas sample is collected automatically into the
chromatograph for analysis. Analysis of the patient sample is
superimposed on the preanalysis graph and all measured values are
displayed on the data collection screen. While the sample is being
analyzed, the diffusion circuit is flushed. Once the results from the previous
effort are displayed, the system is ready for additional efforts.
25
Selecting Efforts
Using Auto Select
If the “Perform Auto-Select” option is on, the computer selects
acceptable efforts automatically. Selected efforts are indicated by a
check in the select column. The computer does not automatically save
any efforts with error messages in the codes column. If all efforts have
performance codes, no efforts are automatically saved; you must save
the desired effort(s) manually.
Drawing Efforts
Double-click on an effort’s time stamp to draw the effort. You can draw
several efforts as a montage to compare efforts of interest without
clearing the screen between selections. Double-click the graph to
enlarge it in a zoom dialog. Click to clear the review graphs.
Tip: Any time you click an effort’s time stamp to draw it, the system switches to
Review mode. You will need to click the icon to re-enter the Test mode. Also
see About Review and Test Modes.
Printing Reports
To print a report after testing is complete, select the Reports Menu. You
may choose from a list of the most recently generated report formats or
you may select Report Switchboard to select from all reports available on
your system.
26
DLCO Options
To access the DLCO test options, select the DLCO tab and choose
“Options” from the tools menu or press F5. The following options are
available from the diffusion setup menu:
Configuration
Changes made to either the test options or the test view (splitter bar
positions, column width, display variables and graph properties) can be
saved to a configuration file that can then be selected for future tests.
Several preset configurations are available for each test. You may select
one of these or create a new configuration. The default DLCO
configuration is locked at the time of installation.
Lock-out Time
Enter the desired time (in seconds) to specify the period for which the
patient is locked out of the system for breath holding. The lockout time
may be changed to account for slow exhalations by obstructed patients.
Lockout time should be approximately six to eight seconds.
DL/VA
The DL/VA option allows for the VA used in the DL/VA calculation to be
either BTPS or STPD (the default is BTPS).
27
Inspiratory Time
There are two options available for the evaluation of inspiratory time:
1. ATS measures the time required to inspire 90% of the final inspired
volume.
2. STD uses the entire inspiratory time (Default is ATS).
Note: The “Inspiratory Time” setting refers to the evaluation of inspiratory time for
acceptability and does not affect the calculation of breath-holding time.
Perform Auto-Select
If Effort Analysis is on, all acceptable and reproducible efforts are
averaged and saved automatically for generated reports.
Evaluate Reproducibility
When selected, new efforts are evaluated for reproducibility. Select which
variable to use as criteria for reproducibility. Use the spinner control to
adjust the margin of reproducibility.
Pressure Indicator On
Displays mouth pressure during lock out.
28
Collection Protocol
The following four options are available for collection protocol: Jones-
Meade (ATS), Ogilvie (ATS), ESP, and User-Defined:
1. Using Jones-Meade, total breath-holding time consists of: two-thirds
of the inspiratory time; the entire lockout time; and one-half of the
collection time.
2. Using Ogilvie, breath-holding time is calculated from the onset of
inspiration to the onset of collection.
3. Using ESP, breath-holding time begins when half of the inspired
volume has been inhaled and ends at the onset of collection.
4. User-defined protocols allow you to set the volume to be used for
determining the start and stop collect volumes. If the user-defined
configuration is selected, the start and stop collect volumes
become accessible from the setup menu. (default value is Jones-
Meade, ATS)
Hemoglobin Author
The start collect volume defaults to 0.75 liters if the patient’s VC is greater
than 2.00 liters. This volume defaults to 0.50 liters if the patient’s VC is less
than 2.00 liters (ATS recommendation).
29
The default volume is 1.25 liters if the patient’s VC is 2.00 liters or more. The
default value decreases to 1.00 liter if the patient’s VC is less than 2.00
liters (ATS recommendation).
30
Patient Valve Lockout
Set to either Automatic or Manual mode to specify how the patient is
switched-in to breathing the diffusion gas mixture and how the lockout
period is initiated during testing:
Note: In Manual mode, press the spacebar to trigger switch-in and lock out while
conducting the test.
31
FRC TESTING
FRC Overview
The Ultima PF and PFX use an open circuit nitrogen washout technique,
functional residual capacity (FRC), to measure lung volume. The nitrogen
washout procedure calculates lung volume by (1) assuming an initial
concentration of nitrogen in the lungs, (2) removing the nitrogen by
“washing it out” with pure oxygen, (3) measuring the volume of air expired
during “washout,” (4) measuring nitrogen concentration in that volume to
determine the actual volume of the lung and (5) using O2 and CO2 to
measure the oxygen and carbon dioxide gas concentrations of each
exhaled breath. Nitrogen gas concentrations are then indirectly
calculated by subtracting the O2 and CO2 concentrations from 100%. The
remaining gas would be N2.
All other calculations and test considerations would remain the same as
nitrogen washout testing.
32
Using the FRC N2 Toolbar
33
Using the FRC N2 Data Grid
34
FRC Test Procedure
Instructing the Patient
Standard testing procedures begin with instructing the patient and
demonstrating proper technique. The greatest potential source for error is
the failure of the patient to perform the test properly. If the patient is
relaxed, the end-expiratory volume will represent true resting end-tidal
volume (FRC).
Pre-Test Procedure
1. Be sure you have warmed up the system and performed the daily
complete pneumotach calibration. Before testing, the GX vacuum
pump must be on for 10 minutes. Warm-up time is displayed on the
associated timer at the top of the screen.
2. Click to create a new patient or click to open an existing
patient. Enter the appropriate information. Required fields are
accented with color. See Opening a Patient File and Starting a
Visit.
3. When you have finished entering patient information, click to
enter a new visit and begin a new test session.
4. Fill out the visit demographics and patient history tabs.
5. Select the FRC tab.
6. The system samples room air. If the room air N2 level is not within the
system’s allowable range (79.6% +2%), flush the system with room
air using a calibration syringe.
7. The GX AutoCal is all that is required to be performed before a N2
Washout Test (FRC).
8. Attach the patient circuit to the arm of Ultima PF or PFX. Connect
the pneumotach and umbilical clip to the patient circuit so the
umbilical tubing is in a straight up (12 o’clock) position.
Note: In order to calculate RV and TLC, an SVC test should be conducted prior to
performing a nitrogen washout test. An optional SVC test may be performed
within the Nitrogen Washout test. Select “Individual SVC Maneuver” from the FRC
N2 options tab to allow SVC efforts within FRC N2 testing.
Test Procedure
1. Before beginning the test, click “Zero Flow” to zero the
pneumotach. There must be no flow through the pneumotach
during this procedure.
2. If you are not testing using the pre/post protocol, click the
protocol/log tab and choose a test protocol from the protocol pull
down menu. Pre/post is the default protocol for all new visits.
3. Ensure the preVent pneumotach with rubber mouthpiece and clip
assembly are attached to the patient circuit in the arm of the
35
Ultima PF or PFX. The umbilical tubing should be positioned in a
straight up (12 o’clock) position.
4. Place nose clip on patient and instruct patient to breathe normally
on the mouthpiece. Press the spacebar to begin testing.
5. Instruct the patient to relax as much as possible and, breathing
normally, perform a minimum of four stable tidal breaths. The
volume/time graph displays the patient’s breathing pattern while
the patient is breathing on the system.
Note: If the patient’s breathing pattern is unstable, have the patient continue to
breathe tidally beyond the minimum four breaths.
6. After four stable tidal breaths, press the spacebar or click “Next”
during an exhalation to begin the washout process. At the end of
that exhalation, the computer automatically switches the patient
to 100% oxygen. The graph on the right of the screen displays the
percent nitrogen of each breath the patient exhales.
Note: After switching to 100% O2, observe the switch-in error and N2 indicators on
the screen. The nitrogen concentration of each breath is displayed in real time.
Note: The less uniform the distribution of ventilation, the more time will be required
to wash out the nitrogen from the lungs. Chronic obstructive patients may not be
able to obtain a nitrogen reading less than 1.7%. In these cases your lab should
have a protocol to end the test at different nitrogen percentages or after a
specified time (usually 7 to 8 minutes).
36
37
Selecting FRC (N2) Efforts and Printing Reports
Using Auto Select
If “Perform Auto-Select” is on, the computer selects acceptable efforts
automatically. The computer does not automatically save any efforts with
error messages in the codes column. If all efforts have performance
codes, no efforts are automatically saved; you will then have to save the
desired effort(s) manually.
Drawing Efforts
Double-click on an effort’s time stamp to draw the effort. You can draw
several efforts as a montage to compare efforts of interest without
clearing the screen between selections. Double-click on the graph to
enlarge it in a zoom dialog. Click to clear review graphs.
Tip: Any time you click an effort’s time stamp to draw it, the system switches to
Review mode. You will need to click the icon to re-enter Test mode. Also see
About Review and Test Modes.
Printing Reports
When testing is complete, select the reports menu. You may choose from
a list of the most recently generated report formats, or select “Report
Switchboard” to select from all reports available on your system.
38
FRC Options
To access the FRC test options, select the FRC tab and choose “Options”
from the tools menu or press F5. The following options are available from
the FRC Setup menu:
Configuration
Changes made to either the test options or the test view (splitter bar
positions, column width, display variables, and graph properties) can be
saved to a configuration file that can then be selected for future tests.
Several preset configurations are available for each test. You may select
an existing configuration file or create a new configuration. Configuration
changes are automatically saved unless they are locked.
To create a new test configuration:
1. Open a patient visit and select the desired test tab.
2. From the Tools menu, select Options.
3. In the Options menu, click the arrow in the Configuration area and
select Create New.
4. Enter a name for the configuration and click OK.
5. Modify the test configuration as necessary. Your changes are
automatically saved.
Perform Auto-Select
If selected, the software automatically selects efforts that are both
acceptable and reproducible.
Average IC/ERV
39
When selected, the RV and TLC values for average efforts are calculated
from the average IC and ERV and the average FRC. This option is not
available if “Individual SVC Maneuvers” is not selected.
40
Evaluate Reproducibility
When selected, new efforts are evaluated for reproducibility. Select which
variable to use as criteria for reproducibility. Use the spinner control to
adjust the margin of reproducibility.
TLC Calculation
Click on the arrow to select from a list of options for calculating TLC.
41
MIP/MEP
MIP/MEP Overview
Patients can be tested to determine their Maximal Inspiratory Pressure
(MIP) and Maximal Expiratory Pressure (MEP) using the BREEZESUITE software.
The graph on the left of the data collection screen displays the
volume/time and pressure tracing as the patient is breathing on the
system. The graph on the right displays pressure vs. %TLC for each patient
effort. The yellow highlighted message bar in the upper-right displays
operating messages throughout the test. The table beneath the message
bar displays the predicted value and the values obtained as each effort is
performed.
42
MIP MEP Data Grid
43
44
Test Procedure
1. Autozero the pneumotach.
2. Ensure that the pneumotach is attached to the pneumotach clip
and umbilical and that this unit is attached to the patient circuit in
the Ultima PF or PFX arm.
3. Place nose clips on the patient and instruct the patient to breathe
normally through the mouthpiece.
4. Press the spacebar or click “Next” to display the tracing of the
patient’s tidal breathing.
The graph on the left of the Data Collection screen displays the
volume vs. time tracing of the patient’s breathing. After a minimum of
three tidal breaths, a MIP or a MEP pressure measurement can be
obtained with one of the following steps:
5. MIP:
• Instruct patient to exhale slowly to residual volume (RV).
• When the patient appears to be at RV, press the spacebar
or click "Next." The Distal occlusion piston will immediately
close.
• When the piston closes, have the patient inhale as hard as
possible to produce maximal negative pressure. The
occlusion piston reopens in four seconds, ending the test.
• To open the piston prior to this time, press the spacebar or
click "Next" as soon as pressure measurement is made.
6. MEP:
• Instruct the patient to inhale slowly to total lung capacity
(TLC).
• When the patient appears to be at TLC, press the spacebar
or click "Next." The distal occlusion piston will immediately
close.
• When the piston closes, have the patient exhale as hard as
possible to produce a maximal positive pressure.
• The occlusion piston reopens in four seconds, ending the
test. To open the piston prior to this time, press the spacebar
or click "Next" as soon as the pressure measurement is made.
The graph on the left half of the screen also displays the pressure vs.
time tracing for each patient effort. The graph’s lower half is negative
pressure (inspiratory maneuver); the top half is positive pressure
(expiratory maneuver).
45
Note: Pressure measurements can be obtained at any lung volume.
Measurements do not have to be taken at RV or TLC.
46
MIP/MEP Options
Keystroke to Start Test
When “Keystroke to Start Test” is not checked, breathing on the system
initiates the start of data collection. When “Keystroke to Start Test” is
checked, the spacebar must be pressed to start a test.
Maximum Occlusion
Displays the amount of time before the shutter will reopen when making a
MIP/MEP measurement. The shutter can be opened manually by pressing
the spacebar before the maximum occlusion time has elapsed (default is
4 seconds).
Measurement Mode
Displays the method used to measure the MIP and MEP pressures:
• Peak: takes the greatest value, negative or positive.
• Peak after one second: takes the greatest value after 1 second,
negative or positive.
• Sustained one second: Takes the pressure measurement that is
sustained for 1 second.
47
48
MIP/MEP Test Screen
49
VO2 TEST PROCEDURE
Test Procedure
1. Select the Test tab.
2. Select PWave from the Tools menu to check the physiologic
waveforms to verify that the equipment is set up and operating
properly. Close the PWave window after verifying the waveforms.
3. Click Start on the Test Pad to begin resting data collection.
4. Click Exercise to manually activate exercise testing. Exercise testing
begins when starting the protocol by clicking Exercise. While testing,
use the Test Pad to:
50
5. Click Recovery to end exercise and begin the recovery phase of
testing.
6. Click Stop to end data collection.
7. Review Anaerobic Threshold achieved from the AT tab.
8. Select and print report.
51
DIRECT CONNECT preVent PNEUMOTACH FOR
VENTILATOR PATIENTS
52
Example of a patient on Direct Connect for
Metabolic/Nutritional Assessment
53
Figure 21: Direct connect
setup
Patient Setup
1. Connect the Direct Connect pneumotach into the ventilator
circuit.
2. Use the Direct Connect ventilator script. Scripts are located in
the Settings Tab under the GX tab.
3. Select the Test tab and begin testing.
Note: In order to test with a standard preVent Pneumotach after testing with the
Direct Connect, a 3L pneumotach calibration must be completed.
54
Figure 22: Direct connect
setup
While there have been no reported incidents of nosicomial infections or
cross contamination related to the metabolic cart, these options may be
considered for issues of infection control.
Option 2: Use HME Filter MGC P/N 536606-001 See Figure 22.
Puritan Bennett has recommended the use of pre-blended gas, turning off
the pressure support during the study, or using alternative modes of
ventilation.
References
1. Benotti PN, Bistrian B: Metabolic and nutritional aspects of weaning
from mechanical ventilation, Critical Care Medicine 1989; 17:2 181-
185.
2. Anderson CF, Loosbrock LM, Moxness KE: Nutrient Intake in the
Critically Ill Patients: Too Many or Too Few Calories, Mayo Clinic
Proceedings 1986; 61:853-858.
3. D’Attellis NP, Bursztein S, Askanazi J, Kvetan V: Tailoring nutritional
support: What, when, and why, Journal of Critical Illness 1988; 3: 49-68.
4. Twomey PL, Patching SC: Cost Effectiveness of Nutritional Support,
JPEN 1985; 9:3-10.
55
REPLACING THE DRYING LINE AND CARTRIDGE
Note: When installing drying sample line tubing into the SAMPLE fitting, loosen
compression screws and gently insert the tubing into the tubing fitting until the
tubing cannot be inserted any further. Tighten compression screws (finger tight)
and tug gently on the tubing to ensure a proper seal.
Note: Use a twisting motion when “pushing” the clear acrylic end plugs (part of
the “drying cartridge holder”) onto the drying cartridge. Pushing plugs straight
into the cartridge without twisting may damage the O-ring on the acrylic plugs.
3. Place the drying cartridge assembly into the clip, keeping the end
with attached tubing towards the bottom of the Ultima. Attach the
other end of the silicone tubing to the “OUTLET” port on the front of
the Ultima.
4. Obtain a drying sample line (750020-002). Install the end of the
drying sample line with the short piece of silicone tubing onto the
barb fitting on top of the drying cartridge assembly.
5. Attach the umbilical luer fitting to the drying sample line luer lock
ring. Install the remaining length of tubing into the “SAMPLE” fitting.
Lightly secure the drying sample line to Ultima with the provided
cable tie.
56
Figure 23: Drying sample line
57
SAMPLE LINE MAINTENANCE
The sample line from the system to the pneumotach must be 7 feet long
and must be purged or replaced when saliva enters and clogs the sample
line.
Procedure
1. Unscrew the luer fitting between the pink sample line and the
drying line.
2. Ensure the Ultima pink sample line assembly (# 701068-001) is
available as a replacement.
3. Pull back the rubber strain relief from the clip by pushing in on the
side tabs to reveal the sample line.
4. Loosen the compression screw that holds the sample line inside the
clip and pull the pink sample line out of the fitting.
5. Attach the end of the new line to the luer fitting of the old sample
line and pull both lines through the protective sleeve from the boot
end to the clip.
6. Insert the new pink sample line into the compression screw,
ensuring it is inserted through o-rings inside of the clip.
7. Reattach the luer fitting on the drying line fitting.
The design of the umbilical is such that the pink sample line is located
inside a protective sleeve to minimize any kinking or crimping of the
sample line. The purge pump is used to clear the lines of any saliva or
debris that may have seeped into the sample line.
58
The connection between the drying line and the pink sample line is a luer
fitting. The sample line filter is located inside the luer fitting. Use a blunt
needle to extract the filter for replacement. A quality control check
should be performed on the sample line every month, along with any
necessary maintenance or filter maintenance.
59
REFRESHING THE O2 CELL
60
REPLACING THE O2 CELL
Replace the O2 cell when the cell refresh process no longer provides
acceptable performance.
Caution: The O2 cell components can be very hot if not properly cooled. Use
caution when removing them. Place items on a non-combustible surface.
Caution: Turn the system power off and allow the system to cool for at least one
hour. Always unplug the power cord when working inside the system.
1. Ensure the system power has been off for at least one hour.
2. Unplug the device from the wall outlet.
3. Locate the O2 assembly cell plate behind the access panel.
4. Loosen four large screws (1/4 turn) from the corners of the O2 cell
assembly plate with a large standard-head screwdriver.
5. Carefully slide the assembly out of the cavity. Ensure the entire O2
cell, including both contact rings, slides out. If one gold contact
ring remains in the cavity, bend a straightened paperclip into a
hook shape and carefully remove the remaining contact ring. The
white ceramic extension manifold may remain in place when the
O2 cell is removed. If the extension manifold remains inside the O2
cell when the O2 cell and heater core are removed:
• Place the outer (small) gold contact ring over the open end of
new O2 cell.
• Notice there is a small flat area on the rim of the cell contact.
Place this part of the gold contact onto the cell first. Use the
thumb to firmly press the gold contact ring onto the cell until the
contact is fully seated onto the cell.
Note: The small gold contact ring must be replaced whenever the O2 cell is
removed. The seal is formed when the O2 cell is seated. Do not reuse the small
gold ring.
61
8. Place the large gold contact ring over the conical end of the new
O2 cell and slide the ring down until the shoulder of new O2 cell fits
tightly inside larger gold contact ring.
9. Insert the new O2 cell, with the gold contact rings attached, onto
the manifold extension bracket until the sensor touches the
manifold.
10. Install the O2 heater assembly gently over the O2 cell inside the
heater enclosure, ensuring the card-edge connector of the O2
heater PCB aligns with the mating connector on the O2 heater
assembly.
11. Sequentially tighten the four screws in an “X” pattern until the plate
is secure.
12. Plug in the power cord. Turn the power to on. Wait at least 30
minutes for stabilization before calibrating. Turn the vacuum pump
to on 10 minutes prior to calibrating system.
13. Calibrate the analyzer after the system has stabilized. If there is any
difficulty in calibration, the cell still may not be properly sealed. If
resealing is necessary, use the provided spare gold contact rings.
62
CLEANING CO2 ANALYZER
Required Equipment
• 1 Small beaker or other container.
• 1 Medium-sized flat-blade screwdriver.
• 1 25-50 cc syringe with hypodermic needle.
• Distilled water.
• Absolute ethanol or 98% isopropyl alcohol.
• 1 Small gold contact seal for O2 cell (P/N 535093-001).
• Mild cleaning solution (this may be any gentle household product such
as Ivory dish soap or baby shampoo. Any product containing bleach
or ammonia should NOT be used).
Procedure
1. Turn system power to OFF and allow the system to cool for at least one
hour. Make sure power cord is unplugged.
2. Remove the sample line bulkhead by unscrewing both sections from
the sample port. The metal sample tube should be exposed.
3. Release ¼-turn screws from O2 assembly plate behind the front access
panel.
Caution: The O2 cell components can be VERY HOT if not properly cooled. Use
caution when removing them. Place items on a non-combustible surface.
4. Carefully slide the assembly out of the cavity and ensure the entire cell
(both gold ring contacts and the extension manifold) is removed.
5. Place a small beaker or other container under the sample needle
protruding from inside O2 cell cavity.
6. Inject at least 20 cc of mild cleaning solution into the exposed metal
sample tube using a syringe fitted with hypodermic needle.
7. Repeat Step 6 using distilled water.
8. Repeat Step 6 using absolute ethanol, 98% isopropyl alcohol, or any
alcohol that does not leave film or crystal residue.
Caution: Alcohol may be combustible. Purge all alcohol from the system before
subsequent operation.
9. Place the reference gas hose over the exposed metal sample tube
and run the reference gas through system at approximately 15 psi for
minimum of 2 minutes to dry sample chamber.
10. Prepare the new O2 cell, if replacing the O2 cell, as follows:
• Place the large gold ring contact over the pointed end of new
O2 cell and slide the ring down until the contact fits snugly onto
the cell.
63
• Place the small gold ring contact over the open end of cell until
it fits snugly.
• Remove the defective O2 cell and insert the replacement cell
as described in Steps 11 through 14.
• Replace the manifold extension center over the stainless steel
tube coming from O2 manifold.
Note: The small gold ring contact must be replaced when the O2 cell is removed,
even if the cell is not replaced. When a seal is created but will no longer form a
tight seal, the ring is deformed.
12. Slide the O2 cell with gold ring contacts attached onto the extension
manifold.
13. Insert the O2 heater into the O2 cavity by gently sliding it over the O2 cell,
(be sure to align the screws to the holes). The metal plate must be flush
against the sheet metal for proper seating of the O2 cell.
14. Tighten ¼-turn screws in “X” pattern (left screw followed by lower right,
lower left, and upper right screw).
15. Reconnect the power cord and sample line. Restore power to the system
and allow approximately 30 minutes for stabilization.
16. Check the vacuum pump gauge on the front of system to ensure proper
vacuum level after the stabilization light on the front of the system has
changed from amber to green.
17. Calibrate the analyzer when a good seal is established. A problematic
calibration indicates the O2 cell may not be properly sealed.
Gases/Regulators/Pressures
CO2 Calibration Gas Specifications: 5.00% +/- 40% (3.00 – 7.00). Mixture purity
0.05% of label Marking. . Regulator delivery pressure 15 psi
CO2 Reference Gas Specifications: 0.0% CO2. Mixture purity 0.05% of label
marking. Regulator delivery pressure 15 psi
64
Oxygen (O2): 99.995% Medical Grade. Regulator delivery pressure 135 psi.
Diffusion (DLCO): 0.5 Neon (NE) +/- 0.05% purity. 0.03 Carbon Dioxide (CO) +/-
0.15% purity. Regulator delivery pressure 135 psi.
Helium (HE): 99.95% Medical Grade. <50 ppm CO2, <50 ppm CO, <15 ppm H2O.
Regulator delivery pressure 10 – 30 psi.
65