Error Codes 6 6
6.1 Introduction
Refer to the troubleshooting tables in this section to help you diagnose and correct these
types of errors.
• Those generated by the Power-On Self-Test (POST)
• Those generated by the Extended Self-Test (EST)
• Those displayed during ventilation and representing soft errors, communications errors,
and faults
Within the tables, the lettered procedures corresponding to each error code are sequenced to
correct the most probable malfunction or to present the most efficient corrective action first.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-1
6 Error Codes
6-2 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Error
Test Error Code Corrective Action
Description
POST initialization test 0 Unit enters BUV POST 1. Check +5 V output. Adjust or replace
(attempts to invoke POST): initialization power supply.
Verifies whether POST initialized and is
••••••• failure 2. Replace DCI-display controller PCB
proceeding correctly. • and/or CPU PCB.
POST kernel test 2 •ooooo•o The bus timeout 1. Check +5 V output. Adjust or replace
(connection of PCB tester): Verifies expected when power supply.
whether a PCB tester used at the factory CPU attempted 2. Replace CPU PCB.
is connected, and if so, whether it is to access tester
connected properly. memory
location did not
occur.
POST kernel test 4 •oooo•oo Watchdog timer 1. Check +5 V output. Adjust or replace
(watchdog): Verifies that watchdog failed to time power supply.
timer times out within acceptable out within 60 to 2. Replace CPU PCB.
interval. 260 ms.
POST kernel test 7 •oooo••• Expected bus 1. Check +5 V output. Adjust or replace
(bus timeout): Verifies bus timeout timeout did not power supply.
(timer 1) operation by writing to occur. 2. Replace CPU PCB.
nonexistent memory location, then
reading bus timeout flag to evaluate
subsequent interrupt sequence. This test
executed only if power fail flag indicates
adequate ac. (When power is low, a bus
timeout cannot generate a
non-maskable interrupt.)
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-3
6 Error Codes
Error
Test Error Code Corrective Action
Description
POST kernel test 8 (system clock): •ooo•ooo Integral CPU 1. Check +5 V output. Adjust or replace
Verifies system clock (timer 2) operation timer 2 (system power supply.
by initializing timer and verifying that clock) did not 2. Replace CPU PCB.
timer interrupt service routine has set time out within
timer test flag to proper value within 20 20 + 1 ms after
+1 ms. being initialized.
POST CPU PCB test 1 (POST reentry ••ooooo• More than two 1. Check ac power.
test): Checks POST reentry flag for attempts were 2. Check +5 V output. Adjust or replace
value >2. made to restart power supply.
POST. 3. Replace CPU PCB.
POST CPU PCB test 2 ••oooo•o Unable to 1. Check +5 V output. Adjust or replace
(display initialization): initialize power supply.
Initializes front panel displays, so they displays. 2. Replace CPU PCB.
can be used to relay messages. 3. Replace DCI-display controller PCB.
POST CPU PCB test 4 (digital I/O ••ooo•oo Unable to 1. Check +5 V output. Adjust or replace
initialization): Initializes digital I/O initialize digital power supply.
ports on CPU PCB, including clearing I/O ports. 2. Replace CPU PCB.
DAC values, setting all solenoids (except 3. Replace conversion PCB.
SOL6 and SOL8) off, and opening safety
and exhalation valves.
6-4 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-1: POST Error Codes (80188-Based Units) (continued)
Error
Test Error Code Corrective Action
Description
POST CPU PCB test 8 (CPU ••oo•ooo EPROM 1. Replace software EPROMs.
PCB EPROM checksum): 080x checksum 2. Replace CPU PCB.
Verifies EPROM set on CPU PCB by failure. Press
performing checksum on EPROM <++> to
contents. determine which
EPROM failed.
Press <++> again
for expected and
actual checksum
values.
POST offboard test 9 •••o•oo• Mismatch 1. Verify proper seating of and secure-
(Multibus RAM): 0901 between ness of connections between DCI-dis-
Verifies integrity of Multibus RAM by Multibus RAM play controller PCB and CPU PCB.
writing a unique pattern to each byte, pattern read and 2. Replace DCI-display controller PCB.
then reading contents, and comparing expected 3. Replace CPU PCB.
it to original pattern. pattern. Press
<++> for a
descriptive error
message.
POST offboard test 10 •••o•o•o EPROM 1. Verify that air flow sensor EPROM
(air flow EPROM): 1001 checksum (U504) and oxygen flow sensor
Verifies air flow sensor (Q2) EPROM by failure. Press EPROM (U406) are not reversed.
performing checksum on its contents. <++> for a 2. Replace Q2/T2 and air flow sensor
descriptive error EPROM (U504).
message. 3. Replace DCI-display controller PCB.
4. Replace CPU PCB.
POST offboard test 11 •••o•o•• EPROM 1. Verify that air flow sensor EPROM
(oxygen flow EPROM): Verifies oxygen 1101 checksum (U504) and oxygen flow sensor
flow sensor (Q1) EPROM by performing failure. Press EPROM (U406) are not reversed.
checksum on its contents. <++> for a 2. Replace Q1/T1 and oxygen flow sen-
descriptive error sor EPROM (U406).
message. 3. Replace DCI-display controller PCB.
4. Replace CPU PCB.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-5
6 Error Codes
Error
Test Error Code Corrective Action
Description
POST offboard test 15 •••o•••• Three 1. Check for any soft error codes
(three watchdog timeouts in 24 1501 unexpected (3000-4000 series) in memory. Refer
hour): program to that error code description.
Checks EST and power fail flags to interruptions 2. Replace power supply module(s) or
monitor watchdog timeout activities. occurred within entire power supply.
past 24 hours. 3. Replace DCI-display controller PCB.
4. Replace CPU PCB.
Warning
Error 1501 is indicative of an intermittent fault that may or may not be repeatedly detected by Total EST. The error can
be cleared and normal operation restored by running Total EST. It is important that the operator make note of this
condition and call a Puritan Bennett representative.
POST offboard test 20 ••••o•oo Possible system 1. If a system tester is connected, check
(system tester connection): Verifies 2001 tester connection between CPU PCB con-
whether a system tester used in the communication nector J100 and tester connector
factory is connected to the ventilator, link failure. J100. Ensure that the tester is opera-
and if so, whether it is connected tional before powering up the ventila-
properly tor. Replace the local bus extension
cable
(P/N 4-019694-00) if necessary.
2. If a tester is not connected, replace
CPU PCB.
6-6 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
POST kernel test 1: oo• Mismatch between 1. Check +5 V output. Adjust or replace
Verifies proper operation of CPU PCB pattern read and power supply.
RAM by writing a pattern to RAM, expected. 2. Replace CPU PCB.
reading pattern from RAM, and
comparing it to original pattern.
POST kernel test 2: o•o Failure of PIT or PIC. 1. Check +5 V output. Adjust or replace
Verifies proper operation of Intel 8253 power supply.
Programmable Interval Timer (PIT) and 2. Replace CPU PCB.
Intel 8259 Programmable Interrupt
Controller (PIC). The PIT is tested by
comparing its performance against
software-implemented timing loop.
POST kernel test 3: o•• EPROM-I/O read 1. Check +5 V output. Adjust or replace
Selectively sets and reads I/O ports of operation power supply.
Intel 8755 EPROM-I/O to perform unsuccessful. 2. Replace CPU PCB.
limited operational check.
POST kernel test 4: Checks all EPROMs •oo CPU PCB EPROM 1. Check +5 V output. Adjust or replace
on CPU PCB for proper contents and checksum failure. power supply.
positioning. 2. Reseat or replace CPU PCB EPROMs or
CPU PCB.
POST kernel test 5: •o• Mismatch between 1. Check +5 V output. Adjust or replace
Verifies proper operation of RAM by pattern read and power supply.
writing a pattern to RAM, reading what was expected. 2. Replace CPU PCB.
pattern from RAM, and comparing it to
original pattern.
POST test 060 (digital I/O): ••o Pattern read was 1. Replace CPU PCB.
Selectively sets and reads Intel 8255 0601 invalid.
Parallel Peripheral Interface I/O ports to I/O access was
perform limited operational check. unsuccessful.
POST test 080 ••o EPROM checksum 1. Check connection of/or replace
(memory PCB EPROM checksum): 0801 failure. EPROMs on memory PCB and the six
Verifies EPROM set on memory PCB by EPROMs on CPU PCB.
performing checksum on EPROM 2. Replace memory PCB.
contents.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-7
6 Error Codes
POST test 090 ••o Pattern test failure. 1. Verify that there is a RAM chip in posi-
(memory PCB RAM): 0901 tion U807 of memory PCB (If your
Verifies Multibus RAM using unit’s software has been upgraded
pattern-type test. from an “N” level to a “P” level or
higher).
2. Replace memory PCB.
POST test 100 ••o Checksum invalid. 1. Replace Q2/T2 and air flow sensor
(air flow sensor EPROM): Verifies air 1001 EPROM, U403.
sensor EPROM by performing checksum 2. Replace memory PCB.
on EPROM contents.
POST test 110 ••o Checksum invalid. 1. Replace QI/T1 and oxygen flow sensor
(oxygen flow sensor EPROM): 1101 EPROM, U302.
Verifies Q1 EPROM by performing 2. Replace memory PCB.
checksum on EPROM contents.
POST test 120 ••o Reference or ground 1. Check ±15 V output. Adjust or
(A/D reference voltages): Verifies 1201 channel out of replace power supply.
ground and reference channels of A/D range. 2. Replace conversion and/or interface
converter on conversion PCB. Tests for PCB.
ground tolerance of ±14 LSB (±34 3. Install surge suppressor if not present.
mV) and reference tolerance of 14 LSB 4. Check connection of interconnect
(±34 mV). cable between CPU PCB and conver-
sion PCB. Replace cable if required.
POST test 130: ••o Voltage comparison 1. Check +5 V output. Adjust or replace
Tests A/D and D/A converters’ 1301 between DAC power supply.
breakpoint values throughout their output and A/D 2. Replace conversion PCB
voltage range. Channels DAC output to input out of range. 3. Replace interface PCB.
the A/D converter and ramps through 4. Replace CPU PCB.
its voltage range. The analog meter or
bar graph display will respond during
this test.
POST test 140: ••o Battery-backed RAM 1. Check secureness of battery connec-
Verifies performance and contents of 1401 pattern test failed. tions.
battery-backed RAM. Performs check- 2. Check continuity of CR2 on mother-
sum of battery-backed RAM contents board. Replace motherboard.
and determines battery condition. 3. Replace fuse in utility panel harness.
4. Replaces batteries.
NOTE: 5. Replace memory PCB.
This code will appear if you have
disconnected the batteries for any
reason.
6-8 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-2: POST Error Codes (8088-Based Units) (continued)
POST test 150: ••o Three unexpected 1. Check for any soft error codes
Determines whether ventilator has had 1501 program (3000-4000 series) in memory. Refer
three unexpected system interruptions interruptions to that error code description.
within past 24 hours. occurred within past 2. Replace power supply module(s) or
24 hours. entire power supply.
3. Replace display controller PCB.
4. Replace CPU PCB.
Warning
Error 1501 is indicative of an intermittent fault that may or may not be repeatedly detected by Total EST. The error can
be cleared and normal operation restored by running Total EST. It is important that the operator make note of this
condition and call a Puritan Bennett representative.
Display controller and/or front panel ••o Indeterminate. 1. Verify security of display cable connec-
display PCB test No tions.
message 2. Replace display controller PCB.
3. Replace front panel display PCB.
Test 02x: 02xx System error; 1. Replace CPU PCB or memory PCB.
Prepares ventilator to run EST. problem occurred
between time <EST>
button was pressed
and POST began.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-9
6 Error Codes
6-10 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-3: Soft Error (All Units) (continued)
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-11
6 Error Codes
6-12 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units)
Test 521: 5211 P1 failed to attain 1. Check pressure transducer PCB con-
Autozeroes P1 by venting transducer to zero level. nector J4.
atmosphere by energizing SOL8. 2. Check connection of SOL8.
3. Replace SOL8.
4. Replace pressure transducer PCB.
5. Replace interface PCB.
Test 522: 5221 P2 failed to attain 1. Check pressure transducer PCB con-
Vents P2 to atmospheric pressure. zero level. nector J4.
Autozeroes P2, de-energizes SOL5. 2. Check SOL5 connection.
3. Replace SOL5.
4. Replace pressure transducer PCB.
5. Replace interface PCB.
Test 523: 5231 DP failed to attain 1. Check pressure transducer PCB con-
Autozeroes DP. Energizes SOL6. zero level. nector J4.
2. Check SOL6 connection.
3. Replace SOL6.
4. Replace pressure transducer PCB.
5. Replace interface PCB.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-13
6 Error Codes
Test 524: 5241 Oxygen supply not 1. Check that oxygen supply is
Verifies that oxygen supply is connected connected to the connected to ventilator and supply
to ventilator. Prompts operator to ventilator. pressure is at least 35 psig
connect oxygen. (241.33 kPa).
2. Make sure pneumatic harness is
securely connected to motherboard.
3. Check connections to PS1.
4. Replace interface PCB.
5. Replace conversion PCB interconnec-
tion cable (P/N 4-019231-00) con-
nected to J4 on interface PCB.
6. Replace pneumatic harness
(P/N 4-019238-00).
7. Replace PS1 (P/N 4-019068-00).
Test 525: 5251 Air supply not 1. Check that air supply is connected to
Verifies connection of air supply connected to ventilator and supply pressure is at
ventilator. least 35 psig (241.33 kPa).
2. Check pneumatic harness connection
J10 on motherboard.
3. Check connections to PS2 and PS3.
4. Replace interface PCB.
5. Replace conversion PCB interconnec-
tion cable
(P/N 4-019231-00) connected to J4
on interface PCB.
6. Replace pneumatic harness
(P/N 4-019238-00).
7. Replace PS2 and PS3.
6-14 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
Test 531: 5311 PEEP pressure of 1. Operator did not turn PEEP/CPAP con-
Verifies that an operator-requested PEEP zero was not trol to zero within 15 seconds. Refer to
pressure of zero was established. This established. This test Section 5 to continue
test assumes that test 521 has passed assumes the PEEP Total EST.
and P1 can be autozeroed. autozero test was 2. Block patient wye.
passed. 3. Replace REG5.
4. Replace pressure transducer PCB.
5. Replace SOL8.
6. Replace interface PCB.
Test 532: 5321 Flow reading 1. If value in peak flow setting window is
Before Total EST continues with active discrepancy constant (not fluctuating), take these
portions of testing, patient connection is between Q2 and corrective actions:
evaluated. The air flow subsystem is set Q3. Air flow a. Check to see if patient system is
to establish 10 lpm flow, and measured controller unable to properly connected and Q3/T3
air flow (Q2) is compared to Q3 exhaled establish desired properly inserted. Check for
flow. The patient must be disconnected, flow rate after 30 obstructions.
the wye properly plugged, and Q3 seconds. b. Check air supply for given flow rate.
properly connected.
c. Check for leaks in
Q2/T2 couplings.
d. Adjust REG2 to between 10 and 11
psi (68.95 to 75.85 kPa) or replace.
e. Replace Q3/T3.
f. Replace Q2/T2 and air flow sensor
EPROM, U504 (80188) or U403
(8088).
g. Replace Q3/T3 harness.
h. Replace CV6 and CV7.
i. Replace interface PCB.
j. Replace proportional valve group.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-15
6 Error Codes
Test 532: (continued) 5321 5. If test fails only when operating from
compressor, take these corrective
actions:
a. Check for leak at air inlet while
ventilator is operating from
compressor. If a leak is detected,
replace REG2 assembly.
b. Replace F6.
c. Replace compressor shock mounts.
d. Replace compressor (perform
10,000-hour preventive
maintenance).
6-16 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-17
6 Error Codes
Test 541: 5411 Patient system failed 1. Check patient tubing for leaks, check
System leak test. Pressurizes patient to achieve desired exhalation valve operation and
system to 90 cmH2O using air flow rate pressure. humidifier connection, check patient
of 10 lpm and checks for leaks. This test bacteria filter, and check patient wye
assumes that patient detection tests for blockage.
533, 534, and 535 passed; that patient 2. Check that patient system is con-
wye is blocked; and that patient tubing nected and patient pressure water
system is properly installed. trap is not leaking.
3. Replace exhalation valve.
4. Check SOL6 actuation.
5. Check SOL4 actuation.
Test 542: 5421 Patient system failed 1. Check that the patient system is prop-
Monitors 90 cmH2O pressure obtained to maintain erly connected and free of leaks.
in test 541 for leaks and verifies that constant pressure. 2. Check exhalation valve operation and
patient tubing system does not lose The system pressure connection.
more than 15 cmH2O over 10 seconds. at which leak was 3. Check the operation of SOL6.
detected is 4. Check outlet connector (P/N
displayed beside 4-019353-00) and attached flex tube.
error code. A 5421 5. Check the operation of SOL4.
error that occurs 6. Check for leaks in CV7.
during Quick EST 7. Check the operation of SV/CV3.
does not cause Total
EST to fail.
NOTE:
If ventilator continues to
pressurize after achieving 90
cmH2O, replace proportional
valve group.
6-18 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
Test 551: 5511 Flow rate 1. Check air supply for given flow rate.
Compares Q2 reading with Q3 reading comparison out of 2. Check for leaks in Q2/T2 couplings.
at steady-state flow of 0 lpm. range (+0.6 lpm) or 3. Replace Q2/T2 and air flow sensor
air flow controller EPROM, U504 (80188) or U403
unable to establish (8088).
desired flow rate 4. PSOL2 leaks at zero flow. Replace pro-
after 15 seconds. portional valve group.
5. Replace interface PCB.
6. Check for leaking CV6 at 0 lpm.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-19
6 Error Codes
6-20 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
Test 557: 5571 Unable to autozero 1. Check Q1/T1 and Q2/T2 connections
Autozeroes PSOL1 and PSOL2 with flow PSOL1 and PSOL2 (J5 and J6) on interface PCB.
rate of 1 lpm and stores current required with flow rate of 1 2. If current value displayed in PEAK
to establish flow in battery-backed RAM. lpm. FLOW window is outside of range
If current used is outside of range 25-380 mA, replace Q1/T2 and
25-380 mA, the test does not fail, but a oxygen flow sensor EPROM, U406
default current value of 130 mA is used (80188) or U302 (8088). If value in
in the calculations. TIDAL VOL window is outside of same
range, replace Q2/T1 and air flow sen-
sor U504 (80188) or U403 (8088).
3. Check proper adjustment of REG1 (if
Q1/T1 value is outside of range) or
REG2 (if Q2/T2 value is out of range).
4. Replace proportional valve group.
5. Replace REG1 or REG2,
as appropriate.
6. Replace interface PCB.
7. Replace compressor.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-21
6 Error Codes
Test 557 (continued) 5577 Oxygen flow 1. Check Q1/T1 connection (J5) on inter-
reading from Q1 face PCB.
invalid. 2. Replace Q1/T1 and oxygen flow sen-
sor EPROM, U406 (80188) or U302
(8088).
3. Replace interface PCB.
Test 558: 5581 Flow rate 1. Check Q3/T3 electrical and tubing
Flow-by offsets calibration. Compares comparison out of connections.
Q2 reading with Q3 reading at range or air 2. Replace Q3/T3.
steady-state flow of 10 lpm. Stores air controller unable to 3. Replace Q2/T2.
and exhaled flows in establish flow within 4. Adjust or replace REG2.
battery-backed RAM. 15 seconds. 5. Replace interface PCB
6-22 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-23
6 Error Codes
6-24 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-25
6 Error Codes
6-26 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-27
6 Error Codes
Test 572: 5721 System compliance 1. Verify that patient wye is blocked and
Derives system compliance as a reading invalid or tests 531 through 535 passed.
weighted average of three points along unable to establish 2. Check for patient tubing system leaks.
volume/pressure curve. A flow rate of 5 5 lpm flow. Unable Verify that tests 541 through 543
lpm is established and delivered volume to pressurize to passed.
required to obtain 30, 60, and 85 desired value 3. Replace patient tubing system with
cmH2O pressure is used to calculate within15 seconds. Puritan Bennett-approved system.
system compliance. This test assumes
tests 561 through 566 were passed. 5726 Patient system failed 1. Verify that patient wye is properly
to achieve desired blocked, exhalation valve is active,
If test fails, ventilator uses
pressure. and tubing system does not leak.
a default value of 0.01
2. Check pressure transducer PCB con-
mL/cmH2O.
nector J4.
3. Replace pressure transducer PCB.
4. Replace interface PCB.
Test 572: (continued) 572A PSOL1 drive current 1. See error 561A.
out of range.
6-28 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
Test 581: 5811 P1/DP comparison 1. Adjust or replace REG5, and replace
Cross-checks P1 and DP at 0 cmH2O out of range. jet venturi.
with system pressure relieved. This test 2. Check pressure transducer
assumes test 572 was passed and no PCB connector J4.
leaks have been found. 3. Replace pressure transducer PCB.
4. Replace interface PCB.
5. Replace SOL7.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-29
6 Error Codes
Test 582: 5821 PEEP did not achieve 1. Operator did not rotate
Prompts operator to set PEEP to 0 value of 0 cmH2O. <PEEP/CPAP> control to fully counter-
cmH2O and verifies pressure reading. clockwise position. Continue Total EST
as described in Section 5, making sure
PEEP pressure reading is zero during
next test pass.
2. Verify that tests 521 through
524 passed.
Test 583: 5831 Unable to establish 1. Check oxygen supply for given
Calculates average area 2 lpm oxygen flow; flow rate.
ratio [(P1 + DP) / P1] applied to or area ratio outside 2. Check for leaks in Q1/T1 couplings.
exhalation valve over PEEP range of 0 to range 0.90 through 3. Replace Q1/T1 and oxygen flow sen-
36 cmH2O. Writes ratio into 9.0. sor EPROM, U406 (80188) or U302
battery-backed RAM. The operator is (8088).
prompted to raise PEEP, then to lower it 4. Adjust or replace REG1.
to various values. This test assumes tests 5. Replace proportional valve group.
561 through 566 were passed. 6. Replace interface PCB.
If test fails, ventilator uses default value 7. See error 5621.
of 1.35. 8. Adjust or replace REG5.
9. Replace exhalation valve.
6-30 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
Test 592: 5921 Nebulizer flow at Q3 1. Check that <PEEP/CPAP> control is set
Directs nebulizer output through or Q1 out of range. to off position.
oxygen nebulizer solenoid SOL1 and 2. Empty liquid from nebulizer vial.
oxygen supply line. This test assumes 3. Check proper nebulizer connection
oxygen supply line and nebulizer are and clear flow in nebulizer venturi.
attached correctly and that flow Verify that approved patient tubing
through Q1 is within tolerance of that system is in use.
measured by Q3. This test assumes 4. Check Q3/T3 electrical and tubing
proper operation of Q1 and Q3, i.e., connections.
successful completion of tests 561 5. Replace SOL1.
through 566. 6. See error 5611.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-31
6 Error Codes
Test 593: 5931 Nebulizer flow at Q3 1. Check that <PEEP/CPAP> control is set
Directs nebulizer output through or Q2 out of range. to off position.
nebulizer air solenoid SOL2 from air 2. Empty liquid from nebulizer vial.
supply line. This test assumes that air 3. Check proper nebulizer connection
supply line and nebulizer are attached and clear flow in nebulizer venturi.
correctly and that Q2 and Q3 are Verify that approved patient tubing
operating properly, i.e., successful system is in use.
completion of tests 551 through 556. 4. Check Q3/T3 electrical and tubing
connections within exhalation com-
partment.
6-32 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
Test 602: 6021 Air flow measured at 1. Check that compressor inlet and out-
Response to the compressor is verified Q2 failed to achieve let filters, F5 and F6, are clear
by the previous Test 601. Verifies 120 lpm. of obstruction.
compressor flow by allowing a 120 lpm 2. Test for leaks between compressor
flow through PSOL2. This test assumes outlet and ventilator base. Check for
tests 551 through 556 were passed. leaks around Q2/T2 couplings.
3. Verify that no flow is coming out of air
inlet port.
4. Adjust or replace REG3.
5. Check pulsation damper connection
and condition.
6. Replace compressor.
7. Replace compressor shock mounts.
Test 611: 6111 BUV rate less than 8 1. Adjust BUV volumes.
Invokes BUV and verifies that bpm. 2. Replace interface PCB.
microprocessor is unable to control 3. Replace PS4.
safety valve, PSOLs, SOL8, SOL3, and 4. Replace pressure transducer PCB.
SOL4. This test assumes proper 5. Replace conversion PCB.
functioning of flow sensors Q1 and Q2
(tests 551 through 561 passed) as well 6116 DP reading less than 1. Replace pressure transducer PCB.
as DP. or equal to 13 2. Replace interface PCB.
cmH2O during BUV
inspiration and
greater than or
equal to 5 cmH2O
during BUV
exhalation.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-33
6 Error Codes
Test 611: (continued) 6119 Q3 exhaled flow 1. Check Q3/T3 electrical connection
reading less than or within exhalation compartment.
equal to 5 lpm. 2. Check Q3/T3 connection (J1)
on interface PCB.
3. Replace Q3/T3.
4. Replace PS4.
5. Replace interface PCB.
6. Replace conversion PCB.
Test 612: 6121 SOL3 failed to meet 1. Check that oxygen supply has
Tests operation of crossover solenoid 6123 specified pressure been disconnected and back pressure
valve SOL3 after prompting operator to 6126 limits at patient bled from line. Check electrical and
disconnect oxygen supply. Establishes 6128 wye. Air flow pneumatic connections to SOL3.
10 lpm air flow. Energizes SOL3, causing 612B controller unable to 2. Verify that no flow is coming out of
oxygen supply to pilot safety valve, and establish 10 lpm top of SV/CV3.
resulting in all flow being vented and flow after 30 3. Replace SOL3.
pressure rise at patient wye being seconds. 4. Replace pressure transducer PCB.
prevented (less than 5. Replace SV/CV3.
2 cmH2O). 6. Replace conversion PCB.
De-energizes SOL3, causing air supply 7. Replace interface PCB.
to pilot safety valve and increasing 8. Replace motherboard.
patient pressure above 30 cmH2O. This 9. Check for correct orientation of PS4.
test assumes tests 541 through 543 and 10. Replace CPU PCB.
tests 551 through 556 were passed. 11. Replace PS4.
12. Clean R1 and R2.
Test 613: 6131 BUV air volume out 1. Adjust BUV volumes.
Invokes BUV and monitors delivered air of range 375 to 625 2. Check air supply for given flow rate.
volume using Q2. The volume of mL. 3. Check for leaks in Q2/T2 couplings.
delivered air is displayed in PEAK FLOW 4. Replace Q2/T2 and air flow sensor
lpm display. EPROM, U504 (80188) or
U403 (8088).
5. Adjust or replace REG2.
6. Replace proportional valve group.
7. Replace interface PCB.
8. Replace pressure transducer PCB.
9. Replace conversion PCB.
10. Replace PS4.
6-34 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-4: EST Errors (All Units) (continued)
Test 621: 6211 Invalid key read. 1. Incorrect keystroke by operator. Press
Keyboard test for <ALARM RESET> key. This error condition <ENTER> to restart Total EST at next
Operator has two chances to press could be caused by test or press <ALARM RESET> to abort
correct key. an operator timeout Total EST.
or a short in 2. Check connection of keyboard to
keyboard. front panel display PCB.
3. Replace DCI-display controller or dis-
play controller PCB.
4. Replace keyboard.
5. Replace front panel display PCB.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-35
6 Error Codes
Test 631: 6311 Channel A transmit/ 1. Check that DCI-display controller PCB
Communication channel A (host/ receive error. (80188) or memory PCB (8088) is
CliniVision port) test. A pattern of A-Z properly seated.
and 0-9 is sent in loopback mode to 2. Replace DCI-display controller PCB or
RS-232 channel. memory PCB.
Test 634: 6341 Channel C transmit/ 1. Check that DCI-display controller PCB
Channel C (7202) port test. receive error. is properly seated.
2. Replace DCI-display controller PCB.
6-36 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-5: Communications Errors (All Units)
Test 73x: 73xx Ventilator unable to 1. Verify proper operation of 7202 Dis-
Executes DCI tasks related to 7202 communicate with play (including DIP switch settings
Display. 7202 Display. and cabling).
2. Verify proper DIP switch settings
on DCI-display controller or mem-
ory PCB.
3. Check connections of DCI interface,
including seating of DCI-display
controller PCB or memory PCB.
4. Replace DCI-display controller PCB
or memory PCB.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-37
6 Error Codes
6-38 7200 Series Ventilator System Service Manual 066317 Rev. A (03/04)
Error Codes 6
Table 6-6: Faults (All Units) (continued)
Test 99x (continued) 9908 BUV is active without a 1. If fault is intermittent (cannot be
detected POST or duplicated), do the following:
system fault. a. Make these ventilator control
settings:
<PEEP/CPAP> knob - Maximum
<NEBULIZER> - On
Mode - <CPAP>
<SENSITIVITY> - 2 cmH2O
<O2%> - 60%
Apnea interval - 60 seconds
b. Disconnect main flow bacteria
filter/patient circuit from
ventilator outlet.
c. Monitor +12 V at J14
(Section 7.6.7).
d. Press <MANUAL INSPIRATION>.
e. If 9908 error is present and
voltage is low, replace +12 V
supply.
f. If 9908 error is present and unit
has a 7202 Display, retest
without Display. If 9908 is no
longer present, replace EL PCB in
Display.
066317 Rev. A (03/04) 7200 Series Ventilator System Service Manual 6-39