Radiography Table

User Manual

1
1 Product structure

4 5
2
1 12

16
3 14
13
11
6

7 10 15

1、X-ray tube 2、Console 3、Collimator


4、 Cross arm 5、Stand column 6、Power switch of table
7、Release device 8、Sliding base 9、Rotating position
10、Back carriage 11、Table 12、Lathe table
13、Grid 14、Mobile table frame 15、Foot switch
16、Focus of scale

Main composition diagram of the photographic table

2
2 Product technical parameters

2.1 Power condition


a) Power voltage:Single AC220V±22V;
b) Power frequency:50Hz±1 Hz;
c) Input power:≥1kVA。

2.2 Table surface device

2.2.1 Size of table surface


a) Size:2100mm×800mm,±5%;
b) Height:650mm,±5%.

2.2.2 Mobile table surface performance


a) Horizontal movement range: 0~250mm,±5%;
b) Vertical movement range: 0~900mm,±5% 。

2.2.3 Table locking


a) Locking way:Electromagnetic lock;
b) Locking control:Foot switch;
c) The attenuation equivalent of the table panel is less than1.2mmAL.

3 Symbol explain

Symbol Explain Symbol Explain

Hazard voltage The B-type application part

Physiological effect To protect ground (earth)

Represents the control box Brake keys of the X ray tube


power switch key components rotate
Brake keys of the movement
Brake keys of the column
distance between X-ray tube
moves sideways
focus and detector

3
4 Set up

4.1 Power requirement


Power supply 220VAC 50Hz ,1000VA,R≤10Ω ,and supply from X-ray machine

4.2 Step
1) Carry the table to a flat surface of earth.
2) Remove the fixed device of the horizontal and vertical and Line filter
3) Set X-ray tube ,Collimator then console,high voltage wire and control
wire at the end.

When the table frame is in the level then install the stand column

5 Testing

5.1 Balance adjustment of stand column


If the stand column cannot be balanced in the range of up and down moving,then
need to adjust the weight of the inside immediately.need to add weight if stand
column going down,need to lose weight if stand column going up.

5.2 Connect line

5.2.1 The connect line between radiography table and high voltage generator

Radiography table High voltage generator Mark

L (a.c 220V) a.c 220V


AC power
N (0V) 0V

Protective
ground wire

4
 Protective ground wire between radiography and console must be
good in case of Electric shock risk

 Turn off the power of radiography table after radiography,remove


the brake,in case of damage of the electromagnetic brake has a
long term power.

 Always checking the wire rope, if the damage must be replaced


in time
 No using the radiography table for a period of time.Before using
it again,make sure the machine is normal and safe after checking
the wire rope.
6.Maintenance

Fault phenomenon Reason Solution


1.The brake
1.The brake coil does not coil is broken 1.Replace brake coil
absorb 2.The fuse of 2.Change the fuse
console have burn。
Adjust the distance
2.The unusual loud noise of The brake coil is not adjusted
between the brake coil and
brake coil. to the suction surface
the suction surface.
3 . Moving parts and Adjust the bearing
Check whether the bearing is
rotating parts have position or replace the
damaged or not.
jammed bearing.
Adjust the weight of the
4 . The column is not The balance weight of arms and
inside of the column to
balanced and going down. columns have problem
balance it with the arm.

5
7 Radiography table schematic diagram, see below

Each interface is explained


P 1’s 1 foot connect power button NO port,2 foot connect COM port,3,4 foot connect
220VAC input,
5,6 foot connect power button both port.
X1’s 1,2 foot connect column electromagnet,3,4 foot connect power of collimator,
5,6,7 foot connect electromagnet of casstte
X2’s 1,2 foot connect foot brake,3,4 foot connect vertical motion electromagnet of table,
5,6,7 foot connect horizontal motion electromagnet of table.

6
8、Rotate the focus radiance diagram

W1 W2