Sie sind auf Seite 1von 6

UTILITIES LOCATED

WITHIN UPPER ENCLOSURE

WALL,
BY OTHERS

R
ZE
ILI
ER
ST

SS TRIM,
BY BELIMED
INSET UNIT

UTILITIES LOCATED
BEHIND WALL

WALL,
BY OTHERS

R
ZE
ILI
ER
ST

SS TRIM,
BY BELIMED RECESSED UNIT

Beli ed BELIMED MST-V 606-VS2E


DOUBLE DOOR STERILIZER WITH INTEGRATED BOILER
Dwg. No. :
CUT-606-VS2E
Rev. :
G
Infection Control Typical installation of a Belimed double door Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 MST-V 606-VS2E Sterilizer Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com N.T.S. 1 of 6
LEGEND
SERVICE SHUT-OFF VALVE, BY OTHERS

STEAM PARTICLE STRAINER, BY OTHERS

STEAM TRAP, BY OTHERS

35 1/2" AREA TO BE LEVEL


FOR DOCKING OF CART
STERILE SIDE

FINISHED WALL OPENING 40 1/2"


LINE TYPES
25 3/4"
B
4
23 3/4" FLOOR LEVEL OUTLINE
6" 20 1/4" 6" WALL
MAXIMUM TOLERANCE FOR FINISHED FLOOR LEVEL 1/4"

(BY OTHERS)
15 3/4" 3/4"
HATCH TYPES

1"
SRV3

SRV2
SRV1
DP

DR
3/4"

EL
STAINLESS STEEL ENCLOSURES, BY BELIMED

9"
11"
EL
DP

21 1/4"
SRV3 SRV3

STAINLESS STEEL TRIM, BY BELIMED

27 1/4"
DR
STERILIZER 44 1/8"

DR

31 1/4"
115" OVERALL

GENERAL NOTES
SRV2
SRV2

SRV1 SRV1

ELB
CW
CA
DI

S
R
SS
C
C

1. SEE SHEETS 5 & 6 FOR TECHNICAL SPECIFICATIONS.


2. SEE TYPICAL INSTALLATION DETAILS ON SHEET 4.
6" 4" 4" 4" 4" 4" 6 1/2"
35 1/2" AREA TO BE LEVEL
FOR DOCKING OF CART

STERILIZER 39"

CLEAN SIDE

40 1/2" AREA TO BE LEVEL


FOR DOCKING OF CART

PLAN VIEW
FOR DOCKING OF CART

SEE DETAIL D/4 FOR TYPICAL


OPTIONAL CHILLED WATER
BYPASS.

SS

WALL

CA, DI, CW, CS, CR, SS, ELB


ELB

(BY OTHERS)

EL, DP
EL

DP
CW

CR
DI

CS

SS
CA

1"

F
4
STERILE SIDE

CLEAN SIDE
ELB CONNECTION POINT 94"
ELB CONNECTION POINT 94"

CEILING HEIGHT

CEILING HEIGHT
CEILING HEIGHT

UTILITY CONNECTION POINT 84"


UTILITY CONNECTION POINT 84"

STERILIZER 95 1/4"
STERILIZER 95 1/4"

CHAMBER
WIDTH 26"
HEIGHT 26"
CHAMBER

DR DR

STERILIZER 39"
A
4 STERILIZER 44 1/8"

SEE DETAIL C/4 FOR TYPICAL


UTILITY LAYOUT.

CLEAN SIDE ELEVATION SIDE ELEVATION

Beli ed BELIMED MST-V 606-VS2E


DOUBLE DOOR STERILIZER WITH INTEGRATED BOILER
TECHNICAL DRAWING
Dwg. No. :
CUT-606-VS2E
Rev. :
G
Infection Control Typical installation of a stand alone Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 Belimed MST-V 606-VS2E Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com 1/4" = 1' 2 of 6
LEGEND
SERVICE SHUT-OFF VALVE, BY OTHERS

STEAM PARTICLE STRAINER, BY OTHERS

STEAM TRAP, BY OTHERS

LINE TYPES
35 1/2" AREA TO BE LEVEL
FOR DOCKING OF CART
STERILE SIDE FLOOR LEVEL OUTLINE
FINISHED WALL OPENING 40 1/2"
25 3/4" HATCH TYPES
23 3/4"
20 1/4" WALL
STAINLESS STEEL TRIM, BY BELIMED
MAXIMUM TOLERANCE FOR FINISHED FLOOR LEVEL 1/4"

15 3/4" (BY OTHERS)

1"
SRV3

SRV2
SRV1
DR
STAINLESS STEEL ENCLOSURES, BY BELIMED
SRV3
SRV3 DR
DR
GENERAL NOTES
STERILIZER 44 1/8"
115" OVERALL

SRV2
SRV2
1. SEE SHEETS 5 & 6 FOR TECHNICAL SPECIFICATIONS.

33"
SRV1

31"
SRV1
ELB
CW

2. SEE TYPICAL INSTALLATION DETAILS ON SHEET 4.


CA
DI

6"
S
R
SS
C
C

21"
DP

EL

11"
15"
1"

6" 4" 4" 4" 4" 4" 6 1/2" 4" WALL


35 1/2" AREA TO BE LEVEL

B (BY OTHERS)
FOR DOCKING OF CART

3/4" 3/4"
STERILIZER 39" 4
FINISHED WALL OPENING 40 1/2"

CLEAN SIDE

40 1/2" AREA TO BE LEVEL


FOR DOCKING OF CART

PLAN VIEW
SEE DETAIL D/4 FOR TYPICAL
OPTIONAL CHILLED WATER
BYPASS.

SS

WALL WALL
ELB

CA, DI, CW, CS, CR, SS, ELB

EL, DP
EL

(BY OTHERS) (BY OTHERS)

DP
CW

CR
DI

CS

SS
CA

1" 1"

E
4

FINISHED WALL OPENING HEIGHT 96 1/2"


FINISHED WALL OPENING HEIGHT 96 1/2"
ELB CONNECTION POINT 94"

STERILE SIDE

ELB CONNECTION POINT 94"


CEILING HEIGHT

CEILING HEIGHT

CEILING HEIGHT

CLEAN SIDE
UTILITY CONNECTION POINT 84"

UTILITY CONNECTION POINT 84"


STERILIZER 95 1/4"

STERILIZER 95 1/4"
CHAMBER
WIDTH 26"
HEIGHT 26"
CHAMBER

DR
DR
A
STERILIZER 39" 4
STERILIZER 44 1/8"

SEE DETAIL C/4 FOR TYPICAL


UTILITY LAYOUT.

CLEAN SIDE ELEVATION SIDE ELEVATION

Beli ed BELIMED MST-V 606-VS2E


DOUBLE DOOR STERILIZER WITH INTEGRATED BOILER
TECHNICAL DRAWING
Dwg. No. :
CUT-606-VS2E
Rev. :
G
Infection Control Typical installation of a recessed Belimed Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 MST-V 606-VS2E Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com 1/4" = 1' 3 of 6
5/8" DIA. HILTI KB-TZ EXPANSION
ANCHORS (MIN. EMBED. = 4").
(x4 LOCATIONS).
WALL

3/4"

STERILIZER

#8 1-1/2" SHEET METAL


SCREW INTO STUD.

STAINLESS STEEL
TRIM, BY BELIMED.

DETAIL FOR FOR REGIONS


REQUIRING SEISMIC ANCHORAGE.

TYP. SEISMIC ANCHORING DETAIL TYP. TRIM ATTACHMENT DETAIL


VERTICAL SERIES STERILIZER VERTICAL STERILIZER
A N.T.S. B N.T.S.

UTILITY ROUGH-IN, BY OTHERS

OPTIONAL
OPTIONAL CHILLED STEAM

CW

CR
CS
SS
WATER CONNECTIONS CONNECTION
ELB
CW

CR
CA

CS

SS

EL
DI

ELB CONNECTION

.
94" A.F.F

PIPE
INSULATION,
TYPICAL
DP

PIPE
UTILITY CONNECTIONS INSULATION,

.
84" A.F.F
TYPICAL

UTILITY CONNECTIONS

84" A.F.F .

TO DRAIN
MANUAL SERVICE VALVES, BY OTHERS.
MANUAL SERVICE VALVES, BY OTHERS.
EL - ELECTRICAL (STERILIZER) 1. REFER TO EQUIPMENT PLUMBING,
ELB - ELECTRICAL (BOILER) MECHANICAL AND ELECTRICAL
CA - COMPRESSED AIR PLANS FOR UTILITY LOCATIONS.
CW - COLD WATER
DI - DEIONIZED WATER
CW - COLD WATER 2. SEE TECHNICAL SPECIFICATIONS CS - CHILLED WATER SUPPLY
SS - STEAM (OPTIONAL) FOR UTILITY SIZING, INPUT CR - CHILLED WATER RETURN
CS - CHILLED WATER SUPPLY (OPTIONAL, SEE PLANS) REQUIREMENTS, AND FINAL
CR - CHILLED WATER RETURN (OPTIONAL, SEE PLANS) CONNECTIONS.

TYPICAL UTILITY LAYOUT TYP. (OPTIONAL) BY PASS DETAIL


VERTICAL STERILIZER - (ELECTRIC) CHILLED WATER CONNECTION
C N.T.S. D N.T.S.

WALL, BY OTHERS.

CHANNEL SECTION
ATTACHMENT ANCHOR,
BY OTHERS

CEILING, BY OTHERS. UNISTRUT CHANNEL


SECTIONS, BY OTHERS
UNISTRUT CONDUIT PIPE
CLAMP, (TYPICAL) BY OTHERS.
CHANNEL SECTION THREADED ROD ,
ATTACHMENT ANCHOR, BY OTHERS.
UTILITIES, BY BY OTHERS
OTHERS. UNISTRUT CONDUIT PIPE
CLAMP, (TYPICAL) BY OTHERS.
UNISTRUT CHANNEL
SECTIONS, BY OTHERS

UTILITIES, BY
OTHERS.
EQUIPMENT
WALL OPENING

1. UTILITY ATTACHMENT ASSEMBLY 1. UTILITY ATTACHMENT ASSEMBLY


SHALL FALL WITH IN THE FOOTPRINT
SHALL FALL WITH IN THE FOOTPRINT
OF THE SPECIFIED EQUIPMENT.
OF THE SPECIFIED EQUIPMENT.
2. REFER TO EQUIPMENT PLUMBING,
2. REFER TO EQUIPMENT PLUMBING,
MECHANICAL AND ELECTRICAL
MECHANICAL AND ELECTRICAL
NOTE: REFER TO LOCAL BUILDING PLANS FOR UTILITY LOCATIONS.
NOTE: REFER TO LOCAL BUILDING PLANS FOR UTILITY LOCATIONS.
CODES AND/OR SEISMIC
CODES AND/OR SEISMIC
ANCHORAGE REQUIREMENTS FOR ANCHORAGE REQUIREMENTS FOR 3. SEE TECHNICAL SPECIFICATIONS
3. SEE TECHNICAL SPECIFICATIONS
FINAL ATTACHMENT METHOD. FOR UTILITY SIZING, INPUT
FINAL ATTACHMENT METHOD. FOR UTILITY SIZING, INPUT
REQUIREMENTS, AND FINAL
REQUIREMENTS, AND FINAL
CONNECTIONS. CONNECTIONS.

RECOMMENDED TYPICAL WALL HUNG RECOMMENDED TYPICAL CEILING HUNG


UTILITY ATTACHMENT DETAIL (SIDE VIEW) UTILITY ATTACHMENT DETAIL (SIDE VIEW)
E N.T.S. F N.T.S.

Beli ed BELIMED MST-V 606-VS2E


DOUBLE DOOR STERILIZER WITH INTEGRATED BOILER
TYPICAL DETAILS
Dwg. No. :
CUT-606-VS2E
Rev. :
G
Infection Control Typical installation details for a Belimed Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 MST-V 606-VS2E Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com N.T.S. 4 of 6
MST-V 606VS2E STERILIZER (W/ INTEGRATED BOILER) - TECHNICAL SPECIFICATIONS
REQUIRED INPUTS

Symbol Description Utility Notes Connection Final Utility Conn. By:


Value
Method Others Belimed
Utility: Deionized Water Do eznieid wR oa
ertO w u
ssia e
o
ptd
roduce puer
Connection Size/Type: 1/2 inch / MNPT e
stam n
hewte
g
ia erite
d
stam geneb )(a ro
ltri .
DI Pressure: Min: 30 Max: 70 psig :snoi taci f iScepyt iQlaumW
rm
e
uinMtai F
exl X
m
p2 <:se diroC lh
Peak Consumption: 0.43 gal/min m
p9.3<T D S
:
*Consumption per Load: 3 gal S
u
0.5
1d
n
aS
u
0.1wn
ete
b :yt ivi tC cu
d
no
Utility: Cold Water Cdw ouslsia
m
eoptd
rihetcondensaeptumpandcool
Connection Size/Type: 1/2 inch / MNPT htecondensO aemt.epitm alpe 6saunr0eitcF
°o
(m ngi
CW Pressure: Min: 30 Max: 70 psig w eta
m
etrpmb ue s7t0
eF °slo.)r TDS
ob<t4e00ppm. F
exl X
Peak Consumption: 3 gal/min Wsaehtrodb ue lsotnfeo0d
6t-0ppm Boaw l.cfk
*Consumption per Load: 42 gal pervoenmitb ue stuepdbipcyluostm aopcselo
,ardle.
Utility: Compressed Air
Connection Size/Type: 1/4 inch / MNPT
Air must be dry and oil free. Used to operate
CA Pressure: Min: 70 Max: 100 psig F
exl X
pneumatic control system.
Peak Consumption: 1 CFM
*Consumption per Load: 10 CF
Utility: Electric
Connection Type: Receptacle
Voltage (Nominal): 208 VAC
Voltage (Phase): Junction Box with Receptacle. NEMA
3 Phase
EL Voltage (Frequency): 60 Hz
15-20R to NEC local codes. Grainger part Recep
aetcl X
No. 5C793.
Circuit Protection: 10 Amps
Peak Consumption: 2.3 kW
*Consumption per Load: 0.9 kWh
Utility: Electric for boiler
Connection Type: 1 1/4 inch / EMT
Voltage (Nominal): 480 VAC
Voltage (Phase): 3 Phase Service disconnect provided by others per
ELB Voltage (Frequency): 60 Hz local code.
Had
W
rerdi X
Circuit Protection: 50 Amps
Peak Consumption: 30 kW
*Consumption per Load: 12 kWh
Utility: Data Port
Connection Method: TCP/IP Ethernet to hospital network X
Connection Quantity: One (1) RJ45 Ethernet port per sterilizer X
Facility Wiring: *Cat-5, Cat-6 or Cat-6e wire Location: Ethernet port should be located X
NIC/Com Speed: 10/100 Mbps Networking Port (RJ45) within 5 ft of machine within the enclosed X
DP Broadcast Method: Unicast service area of the sterilizer. X
MTU Size: 1500 X
Belimed provided patch cables for * (By others)
Final Connection:
final connection
X
Port IP Setup: Each RJ45 Ethernet Port with
individual static IP address X
Utility: Equipment Area Exhaust Provide sufficient air exhaust to remove n
erId
cti
Connection Type: Exhaust return grille 2 kW heat rejection. Exhaust X
Utility (Chamber): Safety Pressure Relief Valve Provide connection to safety relief valve and pipe
SRV1 Had
P
rpe
i X
Outlet size: 1 inch / FNPT to drain, atmosphere, or as required by local code.
Utility (Jacket): Safety Pressure Relief Valve Provide connection to safety relief valve and pipe
SRV2 Had
P
rpe
i X
Outlet size: 1 inch / FNPT to drain, atmosphere, or as required by local code.
Utility (Boiler): Safety Pressure Relief Valve Provide connection to safety relief valve and pipe
SRV3 Had
P
rpe
i X
Outlet size: 1 inch / FNPT to drain, atmosphere, or as required by local code.
Optional utilities for chilled water cooling (used for reduction of cold water consumption).
Utility: Chilled Water Supply
Connection Size/Type: 1/2 inch / MNPT
nIsa
e
utd
lHadr
CS Pressure: Min: 30 Max: 90 psig X
Ppei
Peak Consumption: 96 kBTU/hr
*Consumption per Load: 21 kBTU
Utility: Chilled Water Return Pressure difference P CS/CR must be nIsa
e
utd
lHadr
CR Connection Size/Type: 1/2 inch / MNPT greater than 7 PSIG. Ppei X
Utility: Cold Water Cold water is used to prime the vacuum
Connection Size/Type: 1/2 inch / MNPT pump. TDS to be < 400 ppm. Water should
CW Pressure: Min: 30 Max: 70 psig be softened to 0-60 ppm. Back flow F
exl X
Peak Consumption: 2.2 gal/min prevention must be supplied by customer as
*Consumption per Load: 2.6 gal per local code.
Optional Steam Bypass CONTINUED ON PAGE 6

Beli ed BELIMED MST-V 606-VS2E


DOUBLE DOOR STERILIZER WITH INTEGRATED BOILER
TECHNICAL SPECIFICATIONS
Dwg. No. :
CUT-606-VS2E
Rev. :
G
Infection Control Technical Specifications for the installation Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 of a Belimed MST-V 606-VS2E Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com NONE 5 of 6
*Consumption per Load:
Optional Steam Bypass
Utility: Steam
Steam must be dry, saturated, and free from
Connection Size/Type: 3/4 inch / MNPT
all impurities such as gases, oil, dirt, or other
Pressure: Min: 40 Max: 65 psig
SS contaminants. Provide adequate steam F
exl X
Peak Consumption: 1.7 lbs/min strainer typical for sterilizers. All steam drops
*Consumption per Load: 24 lbs must be piped from top of steam supply line.
Consumption while Idle: 0.05 lbs/min
ESTIMATED EMISSIONS
Utility: Open Drain
Connection Size: 1-1/4 inch Provide flush mounted floor sink and trap 1C
4
-/"opper
DR Discharge Flow Rate: 4 gal/min per local code. pei
X
*Consumption per Load: 45 gal
Medium: Noise Radiation In accordance with ISO3746: 1995 Testing
Radiation (Net): 65 dBA Standards.
Medium: Heat Radiation Heat dissipated into service room with
Radiation (Net): 2 kWh ambient temperature of 86°F.
PHYSICAL SPECIFICATIONS
Parameter: Carry-In Dimensions
Height: 81 in Machine Alone
Width: 39 in (i.e. removed from packaging)
Length: 44.1 in
Parameter: Chamber Usable Dimensions
Height: 26 in
Width: 26 in
Length: 27.5 in
Parameter: Weights
Transport Weight: 2000 lbs Transport Weight = Machine + Crating
Operating Weight: 2680 lbs Operational Weight = Machine + Load
OSHPD NUMBER: OPM-0131-13
*Consumptions calculated under pre-vac cycle at 270°F and water temperature of 63°F. Average load/cycle time one hour.

Additional Information:
- MNPT = Male pipe thread -FNPT = Female pipe thread
- Others must provide service valves and strainers for steam and water supply.
- All utility supplies and connections are customer's responsibility.
TS-606-VS2E_RevJ

Beli ed BELIMED MST-V 606-VS2E


DOUBLE DOOR STERILIZER WITH INTEGRATED BOILER
TECHNICAL SPECIFICATIONS
Dwg. No. :
CUT-606-VS2E
Rev. :
G
Infection Control Technical Specifications for the installation Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 of a Belimed MST-V 606-VS2E Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com NONE 6 of 6