Beruflich Dokumente
Kultur Dokumente
WALL,
BY OTHERS
R
ZE
ILI
ER
ST
SS TRIM,
BY BELIMED
INSET UNIT
UTILITIES LOCATED
BEHIND WALL
WALL,
BY OTHERS
R
ZE
ILI
ER
ST
SS TRIM,
BY BELIMED RECESSED UNIT
(BY OTHERS)
15 3/4" 3/4"
HATCH TYPES
1"
SRV3
SRV2
SRV1
DP
DR
3/4"
EL
STAINLESS STEEL ENCLOSURES, BY BELIMED
9"
11"
EL
DP
21 1/4"
SRV3 SRV3
27 1/4"
DR
STERILIZER 44 1/8"
DR
31 1/4"
115" OVERALL
GENERAL NOTES
SRV2
SRV2
SRV1 SRV1
ELB
CW
CA
DI
S
R
SS
C
C
STERILIZER 39"
CLEAN SIDE
PLAN VIEW
FOR DOCKING OF CART
SS
WALL
(BY OTHERS)
EL, DP
EL
DP
CW
CR
DI
CS
SS
CA
1"
F
4
STERILE SIDE
CLEAN SIDE
ELB CONNECTION POINT 94"
ELB CONNECTION POINT 94"
CEILING HEIGHT
CEILING HEIGHT
CEILING HEIGHT
STERILIZER 95 1/4"
STERILIZER 95 1/4"
CHAMBER
WIDTH 26"
HEIGHT 26"
CHAMBER
DR DR
STERILIZER 39"
A
4 STERILIZER 44 1/8"
LINE TYPES
35 1/2" AREA TO BE LEVEL
FOR DOCKING OF CART
STERILE SIDE FLOOR LEVEL OUTLINE
FINISHED WALL OPENING 40 1/2"
25 3/4" HATCH TYPES
23 3/4"
20 1/4" WALL
STAINLESS STEEL TRIM, BY BELIMED
MAXIMUM TOLERANCE FOR FINISHED FLOOR LEVEL 1/4"
1"
SRV3
SRV2
SRV1
DR
STAINLESS STEEL ENCLOSURES, BY BELIMED
SRV3
SRV3 DR
DR
GENERAL NOTES
STERILIZER 44 1/8"
115" OVERALL
SRV2
SRV2
1. SEE SHEETS 5 & 6 FOR TECHNICAL SPECIFICATIONS.
33"
SRV1
31"
SRV1
ELB
CW
6"
S
R
SS
C
C
21"
DP
EL
11"
15"
1"
B (BY OTHERS)
FOR DOCKING OF CART
3/4" 3/4"
STERILIZER 39" 4
FINISHED WALL OPENING 40 1/2"
CLEAN SIDE
PLAN VIEW
SEE DETAIL D/4 FOR TYPICAL
OPTIONAL CHILLED WATER
BYPASS.
SS
WALL WALL
ELB
EL, DP
EL
DP
CW
CR
DI
CS
SS
CA
1" 1"
E
4
STERILE SIDE
CEILING HEIGHT
CEILING HEIGHT
CLEAN SIDE
UTILITY CONNECTION POINT 84"
STERILIZER 95 1/4"
CHAMBER
WIDTH 26"
HEIGHT 26"
CHAMBER
DR
DR
A
STERILIZER 39" 4
STERILIZER 44 1/8"
3/4"
STERILIZER
STAINLESS STEEL
TRIM, BY BELIMED.
OPTIONAL
OPTIONAL CHILLED STEAM
CW
CR
CS
SS
WATER CONNECTIONS CONNECTION
ELB
CW
CR
CA
CS
SS
EL
DI
ELB CONNECTION
.
94" A.F.F
PIPE
INSULATION,
TYPICAL
DP
PIPE
UTILITY CONNECTIONS INSULATION,
.
84" A.F.F
TYPICAL
UTILITY CONNECTIONS
84" A.F.F .
TO DRAIN
MANUAL SERVICE VALVES, BY OTHERS.
MANUAL SERVICE VALVES, BY OTHERS.
EL - ELECTRICAL (STERILIZER) 1. REFER TO EQUIPMENT PLUMBING,
ELB - ELECTRICAL (BOILER) MECHANICAL AND ELECTRICAL
CA - COMPRESSED AIR PLANS FOR UTILITY LOCATIONS.
CW - COLD WATER
DI - DEIONIZED WATER
CW - COLD WATER 2. SEE TECHNICAL SPECIFICATIONS CS - CHILLED WATER SUPPLY
SS - STEAM (OPTIONAL) FOR UTILITY SIZING, INPUT CR - CHILLED WATER RETURN
CS - CHILLED WATER SUPPLY (OPTIONAL, SEE PLANS) REQUIREMENTS, AND FINAL
CR - CHILLED WATER RETURN (OPTIONAL, SEE PLANS) CONNECTIONS.
WALL, BY OTHERS.
CHANNEL SECTION
ATTACHMENT ANCHOR,
BY OTHERS
UTILITIES, BY
OTHERS.
EQUIPMENT
WALL OPENING
Additional Information:
- MNPT = Male pipe thread -FNPT = Female pipe thread
- Others must provide service valves and strainers for steam and water supply.
- All utility supplies and connections are customer's responsibility.
TS-606-VS2E_RevJ