Sie sind auf Seite 1von 5

Biodegradable polymers for use in s u r g e r y -

poly(glycolic)/poly(lactic acid) homo and


copolymers:
2. In vitro degradation

A. M. Reed* a n d D. K. G i l d i n g *
Bioengineering Department, University of Liverpool, Liverpool UK
(Received 25 January 1980; revised 2 July 1980)

The degradation mechanism of a series of polyglycolic/polylactic acid, (PGA/PLA), homo and


copolymers synthesized as in Part 11, has been studied in vitro. An in vitro test model similar to that
described in a previous study 9, was used. The effects of time, temperature and pH on the rate and
mechanism of degradation were elucidated. The degree of degradation was monitored molecularly by
gel permeation chromatography (g.p.c.), tensile strength determination and mass loss measurements.
The mechanism of degradation is shown to be by hydrolysis. The copolymers of PGA and PLA are shown
to have a wide range of degradation rates governed by the hydrophilic/hydrophobic balance and
crystallinity of the respective copolymer. The effect of the glass transition temperature (Tg) of PGA on its
sensitivity to degradation is also demonstrated.

INTRODUCTION varying pH (pH 5-9), at several temperatures (25°-50°C).


Part I of this series 1 described the synthesis and structural On removal from the incubation medium the polymers
characterization of a range of homo and copolymers of were taken and the following parameters determined as a
polyglycolic (PGA) and polylactic (PLA) acid. These function of degradation time and composition:
polymers are known to undergo degradation in the (1) Loss of tensile strength (in the case of PGA, Dexon R
physiological environment 2 - 7 sutures).
This paper describes a study to elucidate the (2) Mass loss.
mechanism and kinetics of degradation of the various (3) Loss in molecular weight ( M W ) and changes in
members of this family of materials 8. This type of study molecular weight distribution ( M W D ) ,
can best be done in vitro. Although PGA has been Change in mechanical properties of PGA (Dexon R)
commercially available for several years as Dexon R sutures are the key parameters from a practical point of
surgical sutures 2 there is little basic information in the view. Mass loss is significant from the point of view of
literature. The need for the development of meaningful in absorption and removal of the polymers from the body.
vitro models for testing candidate biomaterials has been However the most sensitive parameters for studying
discussed elsewhere 9. degradation effects are M W and M W D . A significant
The data presented in Part I t (see Figure 1) show that percentage of polymer chains can be cleaved before there
PGA (Dexon R) and P-I-LA (as an annealed film) have are changes in mechanical properties. Mass loss was
crystallinities of 50 and 37~o respectively by X-ray diffrac- expected to be slower than the loss in M W and tensile
tion (XRD) and differential scanning calorimetry (d.s.c.) properties, as it is necessary to reduce both M W and
techniques. Crystallinity decreases with the increase of mechanical strength to a considerable degree before any
either comonomer, and copolymers in the range of com- material will fragment and/or dissolve.
positions between 25 and 70~o GA are amorphous. This The effect of pH on the above three parameters was
structural variation would be expected to produce an studied over the range pH 5 - 9 , for selected materials.
interesting range of properties, since the two homopo- Similarly the effect of temperature between 25 ° 50°C was
lymers represent hydrophilic (PGA) and hydrophobic also investigated.
(PLA) crystalline matrices each containing ester bonds The effect of enzymes on PGA sutures has been studied
susceptible to hydrolysis. The amorphous range of com- simultaneously and independently by Williams et al. 1°
positions represents a series of copolymers of varying and these experiments were not repeated in the present
hydrophilicity. If dl-lactic acid is used instead of I-lactic study. A series of in vivo experiments were undertaken in
acid, then the amorphous range is extended from 0-70 rats, the results from which will be compared with the in
m~o GA, since P-dI-LA is amorphous. vitro data in Part 3 lt'8 along with the histological and
The in vitro test model employed was similar to that toxicity findings.
described for the PEO/PET series of copolymers 9. In this
case the polymers under investigation were incubated for EXPERIMENTAL
predetermined periods (0-26 weeks) in buffer solutions of
On the basis of data presented in Part 11 three com-
* Present address: MilliporeCorp., Ashby Road, Bedford, MA 01730 positions were chosen for thorough study:
0032-3861/81/040494~05502.00
© 1981 IPC Business Press
494 POLYMER, 1981, Vol 22, April
Biodegradable polymers. A. M. Reed and D. K. Gilding


5° I
o 40

_~ 3c

u
2c

c' L
// the initial weight (m0), where:

% mass loss - t o o - m e
m 0
x 100

The fragments of samples remaining from the above


mechanical and mass loss determinations were reused for
molecular characterization by g.p.c. Molecular characte-
rization by g.p.c, was carried out using chloroform at 25°C
as the eluting solvent for the polymers containing up to 70
(1)

2O 4 60 Z JO 100 m % GA. Above this level of GA, PGA type crystallinity


Mot % gtycolid¢ in copotyrner (see Figure 1) is present and hexafluoroisopropanol was
Figure 1 % Crystallinities for G A / L A copolymers as a function of required as the g.p.c, solvent 8. Two 1 m columns contain-
composition (determined by X-ray and d.s.c, measurements) ing 105 and 103 A Styragel Rwere used for this work with a
flow rate of I ml rain- 1. Although comparative M W s and
(1) PGA; in the form of Dexon R 0, O'S', 3 0 and 3-0 M W D s have been used in this study, methods of calibrat-
'S' sutures. ing the g.p.c, for use with these combinations of solvents
(2) P-1-LA; in the form of 0.5 mm thick films cast from and copolymers are discussed at length elsewhere 12.
CHCI 3 9. The effect of pH on the degradation kinetics of some of
(3) 50/50 m% G A / L A copolymer: in the form of 0.5 mm the materials was determined by incubation in buffers of
thick film, cast from CHC13 9. pH 5, 7 and 9 at 37_+ 1 C as a function of time. Tensile and
In shorter studies, to evaluate specific effects, P-dI-LA mass loss profiles were determined as previously
and the 70/'30 m% GA/LA compositions were studied. described.
In the case of PGA (Dexon ~) sutures we examined the
Polyglycolic acid (PGA) Dexon R suture dependence of the kinetics of degradation on temperature.
Approximately 50 cm lengths of PGA (Dexon R) suture The rate of degradation was followed by the de-
of each suture gauge (0 or 3-0) and each filament diameter terioration of tensile properties and mass loss.
(6 and 2 denier. 'S' denotes the finer filament) were used as Temperatures above and below the glass transition
test pieces. temperature (Tg), i.e. 3 6 C for Dexon R sutures, (measured
Each test piece was weighed and placed in a McCartney by d.s.c.), were used. The experiments were carried out at
screw capped phial containing 15 ml of buffer. The pH was pH 7 over a 12 week period.
monitored as a function of time and remained constant
indicating that the buffer capacity was not exceeded.
Degradation was carried out in a thermostatically con- RESULTS A N D D I S C U S S I O N
trolled water bath (Grant SB10, Cambridge).
The initial study was carried out over a period of 26 Polyglycolic acid (PGA) Dexon R suture
weeks in 0.2 M citrate-phosphate buffer of pH 7 at 37 The effect of degradation in 0.2 M citrate-phosphate
_+ I'~C. Citrate-phosphate buffer was used as it allowed butter o f p H 7 at 3T'C on the change in tensile properties is
the pH to be varied over a wide range (pH 2.6-7.0) without shown as a function of time in Figure 2. The data show no
the potential complication of changing to different buffer effect of suture gauge (i.e. 0 vs 3 0) or filament denier in the
systems. 2-6 denier range. In each case ~ 80% of the initial tensile
strength is maintained for two weeks. From the practical
Glycolic acid/lactic acid, (GA/IA), copolymers and poly-l- viewpoint this is important as a wound has little strength
lactic acid, (P-I-LA), polymers above that of coagulated fibrin up to l week, but is rapidly
1 cm diameter discs of P-1-LA, 50/50 m°//o G A / L A and reinforced by collagen synthesis between 1 and 3 weeks.
70/30 m% GA/LA copolymers were used for degradation Most of the suture's strength is lost over the 2-4 week
studies. These discs were cut from films cast from CHCI 3 period.
onto glass plates. Figure 3 shows the change in stress/strain curve during
As above, each test piece was weighed and placed in a this time. Elongation decreases simultaneously with ten-
McCartney screw capped phial containing 15 ml buffer. sile strength indicating that the material is becoming more
The pH was monitored as a function of time and remained brittle as degradation proceeds.
constant indicating that the buffer capacity was not Mass loss, as a function of time, is shown for two of the
exceeded. Degradation was carried out as described for Dexon R configurations (0 and O'S') in Figure 4. Again
PGA (DexonRl sutures over similar time periods at pH 7 there is no effect of filament denier size. Mass loss begins in
and 37_+ 1C. the fourth week, and is complete by 10 12 weeks provid-
On removal from the incubation buffer all samples were ing the test medium is not agitated. Agitation will cause
washed with distilled water to remove residual buffer and total mass loss within 8 weeks.
dried over P205 under vacuum. The observed asyncronization of tensile and mass loss
In the case of the PGA sutures, tensile measurements profiles indicates that the degradation process may be
were made on the samples. Between 5 and 10 replicates complex, and the molecular loss profiles for 'O' Dexon R
were used for each measurement. The tensile measure- shown in Figure 5 provide the answer. The M W loss
ments were made using a 2 cm gauge length and a 100% profiles are essentially identical for both suture gauges
/rain strain rate on a Howden Tensiometer. and both filament sizes studied. The M I47) for 0, 1, 2 and 3
Mass loss was determined for each sample by compar- week degradation periods show a steady decrease in the
ing the dry weight (me) remaining at a specific time with M Wof the peak of the distribution, with the high M Wtail

POLYMER, 1981, Vol 22, April 495


Biodegradable polymers." A. M. Reed and D. K. Gilding
100~ Poly-I-lactic acid (P-I-LA)
a
I P-1-LA represents a more complex material than the
801
50/50 PGA/PLA copolymer as this is a 37% crystalline
6O hydrophobic matrix. The mass and molecular loss profiles
are shown in Figure 7. The results show that over a 16
4o !
week period, only 10-15% of the mass of the 0.5 mm thick
20 disc is lost, whereas Mw is reduced to ~ 50% of its initial
value. Although tensile properties were not measured,
6 gross observations indicated relatively small changes in
1oo
mechanical properties over a 26 week period. Over this
80 period of time there is no change in crystallinity (as
6c
4c
"F_
~ 20
x: © w

N~oo
C
1'5 ' ~ S
~ 80
~ 6c
4C

20
xl.O
0
100 %
80 v ' weeks

60

40 0"5
2C

o i 2 3 4 s 7
Time (weeks)
Figure 2 In-vitro tensile data, Dexon R sutures, (PGA) (Degraded '6' weeks
at pH 7 and 37 o C). (a), '0' Dexon R ;(b), "0' Dexon R %';(e),'30'
DexonR; (d), '30' Dexon R %' o 16 2b ab
°/o elongation
Figure 3 Stress/strain curves for degraded Dexon R, (PGA),
sutures ('0' Dexon R) (Degraded at 37°C at pH7)

of the distribution being reduced. The low MWtail of the


MWD increases up to 3 weeks, and begins to decrease 100 - : : :
from that point, indicating that mass loss is taking place.
The material remains highly crystalline over the initial 6 80
week period indicating, as expected, that degradation is 6C
taking place in the amorphous regions and the remaining
mass is contained in small crystallites. At 56 days (Figure 4C
5) the majority of the Dexon R is present as oligomers.
i~ 2oi
o
50/50 tool % glycolic acid~lactic acid (GA/LA) copolymer L.
~ lOOI - _._.... b
This copolymer is amorphous and more hydrophobic
than the PGA. The mass and MWloss profiles are shown £ 8O
in Figure 6 and are found to be virtually superimposable,
with the major loss in each occurring over the 2 4 week ~ 60'
period. This arises due to the amorphous nature of the 40
matrix, and confirms, as was suggested in Part 1 x, that this
composition should be an excellent sustained drug 20
release matrix, as it should disappear from the body
within 3 months after the drug has been delivered. As the O 1 3 4 5 6
material is inherently hydrophilic 1, i.e. the equilibrium Time (weeks)
water level is ~25%, it will probably be useful for the Figure 4 In vitro mass loss data, Dexon R sutures (Degraded at
more hydrophobic drugs. pH 7 and 37°C). (a) '0' DexonR; (b), '0' Dexon R %'

496 POLYMER, 1981, Vol 22, April


Biodegradable polymers. • A. M. Reed and D. K Gilding

/--\. 56 days 14 days gradations similar to those described previously for PGA
// \'\ 7 days. X at pH 7and 37~C were set up at 2 5 , 3 0 , 3 5 , 4 0 , 4 5 ' and
50C. Three suture sizes, '0', ' 4 0)' and '5 0', 6 denier/fila-
ment Dexon R were used.
The tensile and mass loss profiles over 12 weeks are
shown for '0' Dexon R as a function of temperature in
Low MW High MW Figures 8a and h respectively. As expected, temperature
Figure 5 In vitro molecular weight distributions for PGA ('0'
Dexon R) sutures, (expressed as a function of degradation time.)
Degradation at pH 7 and 37°C. Degradation time expressed in
days
c~
c
8O " ~
(]
1001~.~. E

60 ~

c~
c 40
E3
E
P
20
(3
E

O 2 4 6 8 10 12 14
Time(weeks)
Figure 7 In vitro mass and molecular weight loss for P - 1 - - L A
polymer (Degraded at pH 7 and 37°C), ( - - I - ) % mass remaining;
(-- - - A - - ) % MW remaining in vitro

0 2 4 6 8 10 12
Ti me (weeks) loo a!
Figure 6 In vitro mass and molecular weight loss for 50/50 GA/LA c~
_c
copolymer (Degraded at pH 7and 37°C). ( - Q - ) % mass remaining, .~ 80
(-- --A-- --) % MW remaining in vitro O

25°C
60
measured by d.s.c.) indicating that the mechanical proper-
U3
ties are controlled by the crystalline phase. 40 mperatures

EJ]'ect qt' pH on degradation kinetics


I//\ \
In vitro degradation was carried out in citrate-
phosphate buffers at pH 5 and pH 7 and boric acid borax
buffer at pH 9 (ref 8). In each case there was no significant
0 2 4 6 8 10 12
change in degradation kinetics with change in pH 8. This
Time (weeks)
was surprising as polyesters are known to be base
sensitive. We considered that the reason accelerated Figure 8a Tensile loss profiles for PGA sutures ('0' Dexon R) as a
function of time
degradation was not observed at pH 9 may have been due
to the combination of hydrophobicity and crystallinity.
The 70/30 GA/LA copolymer has the most hydrophilic 100
composition of the amorphous structures, and it was
expected to show greater sensitivity to pH than any other 8O
in the family. A limited study showed the 70/30 GA/LA c
copolymer beginning to fragment at 12 days, and undergo
1007,,~ mass loss within 4 weeks at pH 9 (ref 8). This
confirms that any sensitivity of these copolymers to pH is
dominated by the combined effects of crystallinity and
6o
\\\
hydrophobicity, and is relatively unimportant, which is
converse to the P E O / P E T system described previously 9.

Effect of temperature on degradation kinetics


It was shown by d.s.c, in Part I that the glass transition
temperature (7~) for PGA is 36°C. As temperature coef- O 2 4 6 8 10 12
ficients of diffusion and hydrolysis rate constants would Time (weeks)
be expected to show a discontinuity at ~, an experiment Figure 8b Mass loss profiles for PGA sutures ('0" Dexon R) as a
was designed to investigate this phenomenon. Six de- function of time

POLYMER, 1981, Vol 22, April 497


Biodegradable polymers. A. M. Reed and D. K. Gilding
10( 12

8c lO

6c 8
F-
o

CD
E
-6 4 0 o 6
.-E
.~
E
"6
~ 2o -6 4 40
t-

L_

F--
2 20
0 25 30 35 40 45 50
Temperature of incubation (°C) at pH 7
E
Figure 9 Slopes of the linear portions of the tensile loss profiles 0 -- L I 4'0 I -n j 0
plotted as a function of temperature PGA suture: '0' Dexon R 25 30 35 45 50
Temperature of incubation (oC)at pH 7
Figure 10 Threshold time to mass loss ( - o - ) and rate of mass
has a dramatic effect on the rate of loss of tensile strength loss (-- --e-- --) plotted as a function of temperature PGA suture:
and mass. '0' Dexon R
The slopes of the linear portions of the tensile loss
profiles are plotted as a function of temperature in Figure
9. The rates are expressed as ~o initial strength loss per The effect of the To of PGA on its sensitivity to
week and show the classical degradation behaviour of a degradation has been demonstrated.
These simple in vitro modelling experiments will be
constant temperature coefficient above and below To with
compared with in vivo behaviour in Part 3 (ref I I).
a clear discontinuity at TO. The temperature coefficient
above Tg is ~ 7 times greater than that below. This
indicates that the glassy state protects PGA from hy- ACKNOWLEDGEMENT
drolysis as all short term chain motions are frozen. Above
We are grateful to the Science Research Council and ICI
TO,water diffusion, and therefore hydrolysis, is more facile.
Corporate Laboratory for the CASE award to A.M.R.
The mass loss profiles in Figure 8b shows that not only
is the rate of mass loss dependent on temperature, but also during the course of this work.
the threshold time before its onset. Figure 10 shows
threshold time prior to mass loss, and rate of mass loss, as REFERENCES
a function of temperature. Clear discontinuities are seen
1 Gilding, D. K. and Reed, A. M. Polymer 1979, 20, 1459
at 36.5°C for each of the functions indicating the position 2 Frazza, E. and Schmitt, E. E. J. Biomed. Mater. Res. 1971, 1, 43
of TO. The differences in slope represent differences in the 3 Dardik, H., Dardik, I. and Laufman, H. Am. J. Surg. 1971, 121,
diffusion constants of the degradation products in the 656
glassy and rubbery states. 4 Miln, D. C., O'Connor, J. and Dalling, R. Scott. Med. J. 1972, 17,
108
5 Jackanicz, T. M., Nash, H. A., Wise, D. L. and Gregory, J. B.
CONCLUSIONS Contraception 1973, 8, 227
6 Kulkarni, R. K., Pani, K. C., Neuman, C. and Leonard, F. J.
This study has shown that the glycolic acid/lactic acid Biomed. Mater. Res. 1971, 5, 169
family of homo and copolymers has a wide range of 7 Miller, R. A., Brady, J. M. and Cutright, D. E. J. Biomed. Mater.
Res. 1977, 11,711
degradation rates that are governed by both hydro- 8 Reed, A. M. Ph.D. Thesis, University of Liverpool, 1978
phobic/hydrophilic balance and crystallinity. The 9 Reed, A. M. and Gilding, D. K. Polymer 1981 (to be published)
mechanical properties and property loss profiles during 10 Williams, D. F. and Mort, E. J. Bioeng. 1977, 1,231
degradation of both homopolymers are governed by 11 Wilson, J. G., Gilding, D. K. and Reed, A. M. 1980 (to be
the crystalline phase. published)
12 Gil.ding, D. K., Reed, A. M. and AskilI, I. N. Polymer 1981 (to be
The degradation mechanism is one of simple hy- published)
drolysis, which is insensitive to pH for most compositions. 13 Wasserman, D. US Pat 3839297, 1975

498 P O L Y M E R , 1 9 8 1 , Vol 22, A p r i l