Beruflich Dokumente
Kultur Dokumente
i
6.2 Main Assemblies Disassemble ...............................................................................6-1
6.2.1 Disassembling of the Display from the Host ..................................................6-1
6.2.2 Disassembling of the Host from the Trolley ...................................................6-2
6.2.3 Disassembling of the Arm Assembly from the Host .......................................6-3
6.2.4 Disassembling of the Humidification Assembly from the Host ......................6-4
6.3 Disassembling of Maintenance Spare Parts ...........................................................6-5
6.3.1 Disassembling of Oxygen Sensor ....................................................................6-5
6.3.2 Disassembling of Pressure Sensor Board .......................................................6-6
6.3.3 Disassembling of Supply O2 Assembly............................................................6-7
6.3.4 Disassembling of TSI Mass Flow Sensor .........................................................6-9
6.3.5 Disassembling of Differential Pressure Transmitter .................................... 6-10
6.3.6 Disassembling of Diaphragm and Scale Board ............................................ 6-12
6.3.7 Disassembling of Core Board and Main Control Board ............................... 6-13
6.3.8 Disassembling of Components in the Display ............................................. 6-14
6.3.9 Disassembling of Electric Panel Assembly................................................... 6-16
6.3.10 Disassembling of One-way Diaphragm........................................................ 6-17
6.3.11 Disassembling of Fan Filter Cotton .............................................................. 6-18
6.3.12 Disassembling of Filter Element of Gas Inlet............................................... 6-19
6.3.13 Disassembling of Filter (Part No.:130003930)............................................. 6-20
6.4 Functional Testing ............................................................................................... 6-21
ii
7.6.4 Oxygen Sensor Maintenance....................................................................... 7-14
7.7 Paramagnetic Oxygen Sensor (optional) ............................................................. 7-14
7.7.1 Paramagnetic Oxygen Sensor Calibration ................................................... 7-15
7.7.2 Technical Specifications of Paramagnetic Oxygen Sensor ........................... 7-15
7.8 Disposal ............................................................................................................... 7-16
7.9 Manufacturing techniques and materials: .......................................................... 7-16
7.10 Free Obligations: ................................................................................................. 7-17
7.11 Security, reliability and operating condition: ...................................................... 7-17
7.12 Return.................................................................................................................. 7-17
11 Troubleshooting................................................................... 11-1
11.1 Technical Error ..................................................................................................... 11-1
11.1.1 Technical Test Error Code ............................................................................ 11-1
11.1.2 Other Errors................................................................................................. 11-4
iii
11.1.3 Service Tools ................................................................................................ 11-4
iv
1 Safety Information
1 Safety Information
The operator of this ventilator must recognize their full responsibility for choosing
appropriate ventilation settings to ensure proper ventilation and patient safety. The
responsibility for the selection of the appropriate level of patient monitoring depends solely
on the equipment operator.
All the monitoring information is for reference only; it should not be used as the sole basis
for therapeutic or diagnostic decisions.
Terms
This manual uses three special indicators to convey information of a specific nature. They
include:
WARNING: Do not use the system until you have read and understood this manual
including:
WARNING: The VG70 Ventilator System is intended for use by authorized and
trained medical personnel only.
WARNING: The users must familiarize themselves with the operation and use of this
machine prior to first clinical use with a patient.
1-1
WARNING: To ensure proper servicing and avoid the possibility of physical injury,
only qualified personnel should attempt to service or make authorized modifications to the
ventilator.
WARNING: An authorized service engineer must first install the ventilator and run
Aeonmed’s installation procedure, which includes calibration of various system
components, before you connect a patient to the ventilator.
WARNING: If a fault is detected in the VG70 ventilator so that its life support functions
are no longer assured: start ventilation using an independent ventilation device
(resuscitation bag) without delay, if necessary with PEEP and/or increased inspiratory O2
concentration.
WARNING: Before activating any part of the ventilator, be sure to check the
equipment for proper operation and, if appropriate, run PUT (pre-use test) as described in
this manual,
WARNING: Ensure that inspiratory and expiratory circuits are connected to the
correct port before operation of equipment.
WARNING: The expiratory gas pathway may become contaminated with body fluids
or expired gases during normal use, and the inspiratory gas pathway may become
contaminated during fault condition, such as occlusion, breath hoses disconnection.
1-2
1 Safety Information
WARNING: Disposable breathing hoses shall not be reused. Reuse of the single use
hoses can cause cross infection.
WARNING: Assure that hoses used have the appropriate resistance and compliance
to ensure proper therapy.
WARNING: Do not disconnect the cable between the Main Control Unit and the GUI
screen while Ventilator is operating.
WARNING: To avoid risk of electric shock, this equipment must only be connected to
a supply main with protective earth.
CAUTION: The breathing circuit must not be installed whenever the VG70 powers up
and whenever a pre-use test is performed.
CAUTION: If the system test fails, do not use the system. Attempt to troubleshoot and
fix the failure. If you are unable to fix the device, ask an authorized service representative
to repair the device.
1-3
CAUTION: Check the ventilator periodically as outlined in this manual; do not use if
defective. Immediately replace parts that are broken, missing, obviously worn, distorted,
or contaminated.
CAUTION: Do not put ventilator into service until the patient setup is complete.
CAUTION: Do not use oxygen hoses that are worn, frayed, or contaminated by
combustible materials such as grease or oils. Textiles, oils, and other combustibles are
easily ignited and burn with great intensity in air enriched with oxygen.
CAUTION: Follow your hospital infection control guidelines for handling infectious
material. Aeonmed recognizes that cleaning, sterilization, sanitation, and disinfection
practices vary widely among health care institutions. It is not possible for Aeonmed to
specify or require specific practices that will meet all needs, or to be responsible for the
effectiveness of cleaning, sterilization, and other practices carried out in the patient care
setting.
CAUTION: Equipment not suitable for use in the presence of a Flammable Anesthetic
mixture with Air or with Oxygen or Nitrous Oxide.
CAUTION: To avoid an electrical shock hazard while servicing the ventilator, be sure
to remove all power to the ventilator by disconnecting the power source and turning off all
ventilator power switches.
CAUTION: To avoid a fire hazard, keep matches, lighted cigarettes, and all other
sources of ignition (e.g., flammable anesthetics and/or heaters) away from the ventilator
and oxygen hoses.
CAUTION: During operation, do not block: Speaker Holes, Exhaust Port, Air Inlet or
Cooling Fan.
1-4
1 Safety Information
CAUTION: The ventilator shall not be used with helium or mixtures with helium.
CAUTION: Tip over hazard; use care when moving ventilator mounted to cart as
device could tip over leading to injury or damage of equipment.
CAUTION: Do not use sharp objects to make selections on the LCD touch screen or
panel.
CAUTION: Do not connect a VGA or USB interface while the system is in service
CAUTION: Batteries should be removed if equipment will not be in service for more
than 6 months.
CAUTION: Do not immerse the oxygen sensor or the connector in any type of liquid.
CAUTION: Do not connect items that are not specified as part of the system.
CAUTION: The auxiliary outlet is only for the recommended humidifier; do not
connect to any other equipment or an additional multiple socket outlets.
CAUTION: When using a humidifier, user should frequently check the water trap and
look for water in the hose. If water is found in the hose, this water should be removed.
1-5
Also, it is important the water trap is positioned in a way such that it is lower than the
patient tubes.
CAUTION: USA Federal law restricts this device for sale by or on the order of a
physician.
NOTE: The user of this product shall have sole responsibility for any ventilator
malfunction due to operation or maintenance performed by anyone not trained by
Aeonmed.
NOTE: Usage of a filter on the expiratory side will increase the resistance of the
patient circuit.
NOTE: In non-invasive (NIV) ventilation, the exhaled volume of the patient can differ
from the measured exhaled volume due to leaks around the mask
NOTE: Do not sterilize or immerse the Mainstream CO2 Adapter in any fluids.
NOTE: All parts of the ventilator system are suitable for use within the patient
environment.
NOTE: All gas volume, flow, and leakage specifications in this manual are expressed
at STPD (standard temperature and pressure dry), except when specified with another
condition.
1-6
2 Notices
2 Notices
Do not make any service repairs on this equipment during the states warranty period. Any
unauthorized work immediately voids the warranty. Aeonmed will not be liable for any
repairs attempted by the owner. Any such attempted repairs other than specified no
warranty repairs void the warranty.
This work is protected the copyright law and is the sole property of the Company. No part
of this document permitted to be copied or otherwise reproduced, translated into other
languages, or stored in any electronic information retrieval system, without the prior
written consent of the Company.
This equipment can radiate radio frequency energy ,as well can be interfered by
electromagnetic If not installed and used in accordance with the instructions mentioned in
the manual, electromagnetic interference may result. The equipment has been tested and
found to comply with the limits set forth in EN60601-1-2 for Medical Products. These limits
provide reasonable protection against electromagnetic interference when operated in the
intended use environments described in this manual. The ventilator has been tested to
conform to the following specifications:
This ventilator is also designed and manufactured to comply with the safety requirements
of IEC 60601-1, IEC 60601-2-12, CAN/CSA C22.2 No. 601.1-M90, and UL 60601-1.
2-1
the vicinity of high-frequency surgical diathermy equipment, defibrillators, or short-wave
therapy equipment. Electromagnetic interference could disrupt the operation of the
ventilator.
The VG70 Critical Care Ventilator is intended for patients ranging from pediatric to adult,
and for use in a wide variety of clinical conditions. Specifically, the VG70 Critical Care
Ventilator is applicable for adult and pediatric patients weighing at least 5 kg for adult and
3 Kg for child (7lbs.), who require the following types of ventilatory support: Positive
Pressure Ventilation, delivered invasively (by ET or Tracheotomy tube) or non-invasively
(by mask) via Assist/Control, SIMV, CPAP and other modes of ventilation.
The VG70 Critical Care Ventilator is intended for use in hospital and hospital-type facilities.
It may be used during intra-hospital transport provided that electrical power and
compressed gas are supplied.
WARNING: The VG70 Ventilator is not designed for use in an MRI environment. Do
not use the VG70 Ventilator near an MRI machine or injury or equipment damage could
result.
This medical equipment complies with the Medical Device Directive, 93/42/EEC, and the
following Technical Standards, to which Conformity is declared:
IEC 60601-1: The device classification is: Class I, Type B applied part (ventilator breathing
circuit and mask), type BF applied part (CO2 module), ordinary enclosed equipment
without protection against ingress of liquids, continuous operation
ISO 80601-2-12
IEC 60601-1
2-2
3 Equipment Symbols
3 Equipment Symbols
Warnings and Cautions indicate all the possible dangers in case of violation of the
Instead of illustrations, symbols may be utilized. Not all of these symbols may necessarily
appear on the equipment or in this User manual. The symbols include:
Lock Manufacturer
3-1
Nebulization Expiratory hold
3-2
4 Introduction
4 Introduction
The VG70 Critical Care Ventilator System consists of two required components: a Main
Control Unit and a Graphical User Interface (GUI). Optional Components available for the
VG70 Critical Care Ventilator system are: Ventilator Cart, Gas Cylinder Transport Cart,
Patient Circuit Positioning Arm, and Hose Assemblies for O2, CO2 monitoring subsystem,
etc.
4-1
4.2 Abbreviations and Definitions
F Breath rate (frequency) in bpm, i.e. ventilation times per minute (S)
Total breath rate, i.e. the sum of breath rate f and spontaneous breath
ftotal
rate fspont(M)
Mean airway pressure. This value is updated at the end of the last
Pmean
respiratory cycle, hence, is a continuous average (M)
Ppeak Airway pressure peak value during one ventilatory cycle (M)
4-2
4 Introduction
4-3
5 Theory of Operation
5 Theory of Operation
5-1
Figure 5-1 Pneumatic Theory
5-2
5 Theory of Operation
4
1
2
3
5-3
Pneumatic Connecting Diagram
5-4
5 Theory of Operation
3 4 5
Differential pressure
1 2 Interface board 3 Inspiratory flow sensor
transmitter
5-5
Oxygen Inlet Module Overview
item Description
2 Low flow oxygen connector, connecting with low flow oxygen supplies.
3 Inlet filter, filtering impurities from the oxygen.
5 one-way valve
5-6
5 Theory of Operation
7 Connector, connecting with oxygen mixer module, supply low flow oxygen
Item Description
1 Connector, connecting with connector 7(low flow oxygen) in Figure 5-5 Supply
Oxygen Module
5-7
7 Solenoid valve V7-V8, Flow 4L/min
5-8
5 Theory of Operation
Item Description
5 Muffler
6 Turbine
7 Radiator
9 Muffler
5-9
Figure 5-8 Safety Inspiratory Module
Item Description
3 Relief valve
6 One-way valve
5-10
5 Theory of Operation
1
2
Figure 5-9
Figure 5-10
Item Description
Movement direction
2 Driving valve
Press the lock nut of the expiration valve, unlock, then pull out the
3
component.
4 Water trap
5-11
5.2 Electrical System
2 3
1 1 4
5
6
7
Figure 5-11
7 Switch power
5-12
5 Theory of Operation
Figure 5-12
5-13
5.2.3 Rear Panel View
Figure 5-13
5-14
5 Theory of Operation
Figure 5-14
Touch screen
10 NET connector 11 12 Alarm LED board
controller
5-15
5.2.5 ARM platform circuit board
Figure 5-15
5-16
5 Theory of Operation
Figure 5-16
2 VGA interface
3 USB 2 interface
4 USB 1 interface
5 NET interface
5-17
5.2.7 Interface Board
Figure 5-17
1 Signal
2 Motor controller
3 Insp. valve
4 Turbine
5 Fan
7 PEEP valve
10 O2 sensor
12 TSI sensor
13 Power in
5-18
5 Theory of Operation
2 Reserved
3 Analogy slot
5-19
6 PSB slot
10 Display connector
11 Reserved
12 Reserved
14 Reserved
5-20
5 Theory of Operation
Figure 5-20
2 DSP_JTAG
5-21
5.2.10 Main Control Board
Figure 5-21
5-22
5 Theory of Operation
Figure 5-22
5-23
5.2.12System Wiring Diagram
Expiration valve
4 122008090
heating component
Magnetic base
5 122008011
component
5-24
5 Theory of Operation
Oxygen supply
6 122007331
pressure cable
Expiration flow
7 122007327
sensor cable
5-25
No. BOM size Name Diagram
Interface board
14 122007319
power cable
Turbine driver
15 122007478
component
5-26
5 Theory of Operation
Power lamp
18 122007486
component
Mute button
20 122007483
component
5-27
No. BOM size Name Diagram
Switch mounting
22 122007455
plate component
Inspiration control
23 122007463
valve
Alarm lamp
25 122007485
component
5-28
5 Theory of Operation
W0001 cable
29 122009449
(optional)
30 122001124 W1 cable
5-29
5.3 Software System
User Interface Software Unit bears the responsibility of user interface of the whole
software system. It includes operation interface for the users, graphic interface for
showing information of the device, communication with other units, etc.
5-30
5 Theory of Operation
5-31
5.3.3 Hardware and Software Interface
5-32
6 Disassemble and Functional Testing
6.1 General
The ventilator is packaged with two cartons. One carton filled with cart. Another carton
filled with ventilator main units, display, and accessories crate.
Figure 6-1
To separate the display from the host: pull out the display (1) along the guide (2).
6-1
6.2.2 Disassembling of the Host from the Trolley
Figure 6-2
c. Remove the host installing piece (3) from the tray connector (4).
NOTE: When installing the host to the trolley, please align the host installing piece
with the tray connector.
6-2
6 Disassemble and Functional Testing
Figure 6-3
6-3
6.2.4 Disassembling of the Humidification Assembly from
the Host
Figure 6-4
To separate the humidification assembly from the host: remove the humidification
6-4
6 Disassemble and Functional Testing
Figure 6-5
6-5
6.3.2 Disassembling of Pressure Sensor Board
Figure 6-6
a. Remove the six screws (2) (Cross recessed Pan head screw M4x8);
d. Remove the two screws (3) (Cross recessed Pan head screw M3x6);
Test: When you replaced the oxygen pressure sensor, turn on the machine to check if the
oxygen pressure value is back to normal.
6-6
6 Disassemble and Functional Testing
Figure 6-7
a. Remove the screws (Hexagon socket head cap screws M4×16) of the upper
cover, and remove the upper cover (1);
b. Remove the inlet link block of the oxygen inlet module, and remove the filter cover
of the air inlet.
c. Remove the screws (Cross recessed Pan head screw M3x6) of the lower rear
cover, and remove the lower rear cover (2);
1
Figure 6-8
d. Remove the screws of the electrical box, pull out the three connectors(2, 3 4) of
supply O2 module from the mother board of the electrical box (5), and then remove the
electrical box (1);
6-7
1 1
Figure 6-9
e. Remove the four screws (1) (Cross recessed Pan head screw M3x25);
f. Remove supply O2 module (2).
6-8
6 Disassemble and Functional Testing
Figure 6-10
a. Remove the six screws (2) (Cross recessed Pan head screw M4x8);
d. Drag out the TSI mass flow sensor along “ ” direction (3).
Test: When the TSI mass flow sensor is replaced, the inspiratory valve must be calibrated.
Please refer to Chapter 10.1.
6-9
6.3.5 Disassembling of Differential Pressure Transmitter
1
Figure 6-11
a. Remove the handle (1), and then remove the host from Cart assembly;
b. Remove screws, and then remove the two side covers (2);
6-10
6 Disassemble and Functional Testing
Figure 6-12
d. Remove the Electrical box components backwards to the right place (1);
Test: When the differential pressure transmitter is replaced, the inspiratory valve must be
calibrated. Please refer to Chapter 8 and Chapter 10.1.
6-11
6.3.6 Disassembling of Diaphragm and Scale Board
To remove the diaphragm and the scale board from expiratory module assembly:
c. Remove the end cover (1 in Figure 6-14) and the gasket (2 in Figure 6-14);
d. Remove the scale board (3 in Figure 6-14) and the diaphragm (4 in Figure 6-14).
Test: When the diaphragm and scale board are replaced, the inspiratory valve must be
calibrated. Please refer to the Chapter 10.1; Also perform a pre-use test and verify all test
pass Refer to Chapter 8.
6-12
6 Disassemble and Functional Testing
Board
Figure 6-15
a. Remove the four screws(1) (Cross recessed Pan head screws M3×6);
6-13
c. Pull out the core board (3);
Test: when you replaced the core board and main control board. Cut off the battery,
connect the alternating current, and check if the machine operates normally.
a. Remove the four screws (1) (Cross recessed countersunk head screw M4×8), take off
the rear shell hanger and hanging block interface board (Figure 6-16);
Figure 6-16
b. Pry the back-shell of display open with the tool. You can use the tool to pry the black
circle signed places (Figure 6-17).
Figure 6-17
6-14
6 Disassemble and Functional Testing
d. Remove the six screws (2) (Cross recessed Pan head screws M3×6), and then
remove the shielding mask (1) (Figure 6-18);
Figure 6-18
6-15
6.3.9 Disassembling of Electric Panel Assembly
Figure 6-19
a. Remove the four screws (1) (Cross recessed Pan head screws M3×6);
Test: When any circuit board has been removed and or replaced, you must perform ad
calibration on all sensors and perform a pre-use check prior to use. Please refer to
Chapter 8 and Chapter 10.1.
6-16
6 Disassemble and Functional Testing
Figure 6-20
Test: When the one-way diaphragm has been replaced, the inspiratory valve must be
calibrated. Please refer to chapter 8 and chapter 10.1.
6-17
6.3.11 Disassembling of Fan Filter Cotton
Figure 6-21
6-18
6 Disassemble and Functional Testing
Figure 6-22
To remove filter element of gas inlet:
6-19
6.3.13 Disassembling of Filter (Part No.:130003930)
Figure 6-23
a. Remove the two screws (1) (Hexagon socket head cap screws M4×10);
6-20
6 Disassemble and Functional Testing
Note: Before the functional testing you need to set ventilator in standard states.
PEEP: 3 cmH2O;
6-21
In standard work status, breathing test device
is connected to ventilator inspiratory port in
VCV mode, after the output tidal volume is Set the value of 20mL, the
stable, record Vti monitoring values of detection value is 10 ~
breathing test equipment and ventilator. Tidal 30mL;
Tidal
volume settings as the following table:
volume
Set the value of 500mL, the
Volume Inspiratory Frequency detection value should be
(mL) time(s) (bpm) 450 ~ 550mL;
Child 20 0.2 60
Adult 500 2 10
6-22
6 Disassemble and Functional Testing
NOTE: These items relate to the external battery, the application of optional
external battery, or NA.
6-23
7 User Maintenance
7 User Maintenance
• Do not inhale fumes that may result from any disinfection process.
WARNING: Movable and removable parts may clamp or even crush your hand. Use
caution when moving or replacing system components.
• Do not inhale fumes that may result from any disinfection process.
WARNING: Do not use talc, zinc stearate, calcium carbonate, corn starch or similar
material to prevent sticking of the bellows, as these materials may enter the patient’s
lungs or airway, causing irritation or injury.
7-1
• Never use organic, halogenated or petroleum-based solvents, anesthetic, glass
cleaning agents, acetone or other irritant agents.
• Never use abrasive agents (i.e. steel wool or silver polish) to clean components.
• All cleaning solutions used must have a pH between 7.0 and 10.5.
CAUTION: Never immerse the oxygen sensor or its connector in any type of liquid.
Discard one time use components after using. Don’t use hard brushes or other sharp tools
in cleaning to avoid damage to parts.
• Clean components with warm water and light detergent.
• After a component is replaced, test the ventilator first prior to putting the ventilator
into service.
CAUTION: Before the first use clean, disinfect, and sterilize the ventilator.
Disposable components must be disposed in accordance with local regulations. Don’t use
hard brushes or other sharp tools in cleaning to avoid damage to parts
7-2
7 User Maintenance
Agent Classification
Detergent
• Mild dishwashing detergent
Detergent
• Soapy water with detergent ph between
7.0 and 10.5
Different parts of the ventilator have their respective cleaning and disinfection methods.
The following categories are defined for the parts noted in Table 7-1. The parts need to be
cleaned, disinfected and thoroughly dried before reassembly.
B: Machine washing: Automatic washing with washer and disinfecting with disinfection
machine.
High-temperature disinfection does not have any cleaning effect. It should only be used on
components that have already been cleaned by hand or machine and then thoroughly
dried.
7-3
Table 7-1
Using a soft cloth with a water-soluble detergent or disinfectant wipes, clean the housing,
gas supply hoses and power cord.
(1) Disassembly
7-4
7 User Maintenance
Figure 7-3
f. Pull out the expiratory valve core component (2 in Figure 7-1, Figure 7-2);
g. Remove the end cover(1 in Figure 7-3) and the gasket (2 in Figure 7-3);
h. Remove the scale board(3 in Figure 7-3)and the diaphragm(4 in Figure 7-3);
j. Pull out one-way valve core(6 in Figure 7-3), rubber o-sealing ring(7 in Figure 7-3)
and one-way diaphragm(8 in Figure 7-3);
(2) Cleaning
(3) Disinfection
NOTE: Ensure that all the components have been cleaned before disinfecting.
Using the Gluteraldehyde disinfection solution, follow the manufacturer’s instructions for
high level disinfection and rinsing of all components while adhering to facility policies and
procedures.
7-5
All the components can also be high temperature and pressure disinfected. Using an
autoclave, follow the manufacturer’s instruction for high level disinfection of all the
components while adhering to facility procedures.
(4) Assembly
After installation, please perform a pre-use test and verify all tests passed.
Do not use a faulty machine. Ask an authorized agent of our company to carry out all
necessary maintenance tasks. Test the machine after maintenance for normal operation.
Every parameter should meet specifications.
In order to ensure the reliability of the machine, all maintenance and repair work should be
carried out by an authorized agent of our company. In case such person is unavailable, a
qualified person with similar maintenance experience is acceptable for the work.
Only use products provided by Aeonmed to replace damaged items. Verify and test all
items ensuring they meet specifications.
Contact local service agent of our company in case support is needed. In all cases,
maintenance fee is the current component price plus reasonable labor cost, except for
those within warranty period.
Minimum maintenance
Task
interval
7-6
7 User Maintenance
Minimum maintenance
Task
interval
WARNING: If the Ventilator will not be in use for a period of more than 6 months, the
internal batteries must be disconnected or removed to prevent possible damage to the
equipment or risk to users or service personnel.
WARNING: The ventilator must not use, nor be connected to, any anti-static or
electrically conductive hoses and tubing.
Service Life is defined as the time that the manufactured device can be expected to be
'serviceable' or supported by our company and the maximum time the device can be used
safely. Some items within the device may require maintenance, repair or replacement
within this time. Such items will be available from our company for the service life of the
product. The calculation of the service life begins at the installation of the product at the
customer site. Our company recommends for safe use that each device is replaced after
its service life is completed.
7-7
Mask, patient pipeline Sterilize for 20 times at 121℃
Machine 8 years
During transportation of the ventilator with or without a patient connected, make sure that
the following conditions are fulfilled:
Gas cylinders are connected with a sufficient amount of gas, the Battery module is
functioning. Follow the hospital guidelines.
Use the handles on the Mobile Cart. Transport the bed and the ventilator slowly, and
watch the patient connection carefully to see that no pulling or other movement
occurs.
Be careful not to tip the Mobile Cart when crossing an obstacle such as a doorstep.
7.3.1 Transportation
WARNING: Tip over hazard – use care when moving Ventilator mounted to Cart as
device could tip over leading to injury or damage to the equipment and possible
subsequent patient injury.
User can carry packaged machine while riding vehicle, plane and train. Impact, severe
shock and damp should be avoided during transportation (other conditions should accord
with purchase contract), with ambient temperature -20°C~+60°C and relative humidity not
more than 95%. In case transportation conditions don’t agree, put the machine in
specified operating environment at least 24h before using.
7-8
7 User Maintenance
7.3.2 Storage
The machine should be stored in a room with temperature -20°C~+60°C and relative
humidity –not more than 95% non-condensing, with ventilation and no corrosive gas.
CAUTION: The device should be stored in a room that is drafty where no corrosion
gas exists.
CAUTION: When the storage conditions are beyond the requirements of operational
environment, and the storage state is transferred into operation state, the product only can
be used after being stored in environment for over 24 hours.
WARNING:When replacing fuse, make sure the new fuse is the same type and size
as the old one; otherwise, the ventilator may be damaged.
CAUTION: The fuse is a damageable part, and care must be taking when replacing it
so no damage occurs.
1. AC Circuit Fuse
Fuse replacement steps:
• Insert screwdriver into the trench (2) of end of fuse box, see Figure 7-4.
• Pull out fuse holder (1).
• Remove the fuse (3).
• Load new fuse.
• Push the new fuse into the original position gently.
• Connect AC power, and then start the ventilator to test.
7-9
Figure 7-4
2. DC Power Fuse
Fuse replacement steps:
• Insert screwdriver into the trench (1) of end of fuse box, see Figure 7-2.
• Pull out fuse holder (2).
• Remove the fuse (3).
• Load new fuse.
• Push the new fuse into the original position gently.
Figure 7-5
Battery module
- DC12V, 6.6AH, 14.4V lithium-ion battery
When the main power supply voltage is too low or the main power supply fails, two backup
batteries (one is necessary, and one is optional) can protect the ventilator. When having a
power failure, the ventilator can switch to battery supply automatically, and can normally
work without pneumatic power supply failure. The two batteries are usually available for
the ventilator working for 4 hours.
7-10
7 User Maintenance
7.5.2 Precautions
Charge: When operating with AC power supply on, the system will maintain the battery
automatically. Charge time is less than 3.5h.
In case of low battery condition, an alarm message “Low battery” will appear, notifying the
user to restore AC power supply to charge batteries, otherwise the batteries will be
depleted and alarm “Limited Battery Capacity” will be displayed, and eventually the
system will be shut down. (For safety reason, manual power-on is required to start the
machines again after an automatic shutdown).
Before the machine is put into patient use, or disconnecting AC power for transport, or
other purposes, check the battery power. If the battery is not fully charged, connect the
ventilator to AC power for at least 3.5hours and recharge the battery until the power
reaches 80%~100%.
In case the battery is to be stored for a long time, charge it fully prior to storage.
To keep the battery power and prolong the battery life, please ensure that the ventilator is
connected to the main power. Charge the battery every six months. Actual charging time
depends on the storage environment.
High humidity and high temperature environments should be avoided for storage.
Same model battery with CE certification is required. Make sure AC power supply is
disconnected before replacing.
7-11
7.5.5 Battery Charging and Calibration
Use the battery charger supplied by our company to charge or calibrate the battery. After
calibrating the battery, the ventilator can read the residual battery capacity accurately.
Please charge or calibrate the battery according to the instructions of the battery charger.
CAUTION: When ‘low battery’ alarm occurs, charging should be done immediately.
The VG70 Ventilator System will shut off in several minutes automatically.
The oxygen sensor can be used to measure the local oxygen concentration when it is
connected to the ventilator or other equipment. The oxygen sensor is suitable for adult
and child.
Disassembly
Step 2: Pull up the crystal joint of the oxygen sensor, turn the oxygen sensor (3)
anticlockwise, and then remove it.
7-12
7 User Maintenance
Assembly
Inspect the oxygen sensor for damage and replace as necessary. Then reassemble the
oxygen sensor.
7-13
produces a current with an amplitude proportional to the oxygen
partial pressure indicated by the electrode sensor. The base
metal electrode is gradually exhausted during the oxidation
process.
The voltage of the sensor is influenced by the temperature of
the monitoring gas mixture. The surgical thermosensitive
resistor of the sensor automatically compensates temperature
change in the sensor.
The O2 monitor converts the sensor signal into the
corresponding oxygen percentage value by using signal
processing and circuit analysis. The system displays the value
and compares it with the stored alarm limit value. If the value
exceeds the threshold, the monitor will alarm.
NOTE: After being in a condensing atmosphere, the oxygen sensor shall be stored for
more than 24 hours in an environment equivalent to operating humidity.
The oxygen sensor should be regularly calibrated. For the calibration interval, refer to
section 7.2.2.
To improve the life time of the oxygen sensor, when the ventilator is not in use, the oxygen
sensor should be avoided contact with the high-concentration oxygen.
The oxygen sensor is consumptive, and the period of valid is ordinarily 12 months. So the
user should pay attention to the use of the oxygen sensor. When the oxygen sensor fails,
replace it.
The paramagnetic oxygen sensor can be used to measure the local oxygen concentration
when it is connected to the ventilator or other equipment. The oxygen sensor is suitable
for adult and child.
7-14
7 User Maintenance
Sensor
Performance
Technology Paramagnetic
Range 0-100% O2 (with over range -15% O2 to 200% O2)
Accuracy (Intrinsic <± 0.2% O2
Error)
Linearity <± 0.2% O2
Repeatability <± 0.2% O2
Zero Drift <± 0.4% O2 in first 24 hours, then <± 0.2% O2 /week, then <
± 0.2% O2 /month
Response Time (T10 – 8 to 20 seconds dependent on application and filter selection
T90) (biological filter on request)
Outputs/Controls
Signal Output Digital UART or linear mV output (0.5mV or 10mV per % O2)
Physical
Weight 70g (2.47oz)
Dimensions Molex low profile connector: 33.5×30.0×46.1mm
(1.32×1.18×1.81”)
Diffusion Port Aperture diameter: 15.5mm (0.61”)
Sample condition
Sample Gas Condition Clean dry gas, free of entrained oil, particulates <3μm,
non-condensing
Gas Exchange Diffusion
Ambient conditions
Operation Temperature 5℃ to 50℃ (41℉ to 122℉)
Storage Temperature -30℃ to 70℃ (-22℉ to 158℉)
Temperature Coefficient Within a range of 0℃ to 50℃: Zero <± 0.5% O2 /10℃, Span:
<±0.5% O2 /10℃
Operating Pressure ±33kPag (±5psig)
Range
7-15
Ambient Humidity 0 to 95% non-condensing
External Power Supply 5V dc, 70mA nominal
Power Consumption 350mW
7.8 Disposal
This product must not be disposed of with your other waste. Instead, it is your
responsibility to dispose of your waste equipment by handing it over to a designated
collection point for the recycling of waste electrical and electronic equipment, or by
returning it to our company for reprocessing. The separate collection and recycling of your
waste equipment at the time of disposal will help to conserve natural resources and
ensure that it is recycled in a manner that protects human health and the environment. For
more information about where you can drop off your waste equipment for recycling,
please contact your local city office, your waste disposal service, or your product
distributor or retailer.
Follow all local regulations with respect to environmental protection when disposing
of batteries and O2 sensor. These products contain toxic compounds irrespective of
physical condition. They should be disposed of according to local waste management
requirements and environmental legislation. They should not be burned since they may
give off toxic fumes.
Do not throw into fire! Risk of explosion.
Do not force open! Danger of bodily injury.
For a period of one year from the date of original delivery, the components and
assemblies of this product are warranted to be free from defects manufacturing
techniques and materials, provided that the same is properly operated under the
conditions of normal use and regular maintenance. The warranty period for other parts is
three months. Expendable parts are not included. Aeonmed’s obligation under the above
warranties is limited to repairing free of charge.
7-16
7 User Maintenance
• Aeonmed’s obligation under the above warranties does not include the freight and other
fees;
• Aeonmed is not responsible for any direct, indirect or failed product or delay which
result from improper use, alteration by using unapproved assemblies and maintenance by
anyone other than Aeonmed;
Aeonmed is not responsible for the security; reliability and operating condition of this
product in case that:
• The assemblies are disassembled, extended, altered and/or readjusted
• This product is not operated correctly in accordance with the manual instruction.
The power supply used or operating environment does not follow the requirements
in this manual.
7.12 Return
Contact customer service of Aeonmed by informing them of the serial number and type of
the product. The number is marked on the surface of the product. Return is unacceptable
if the number cannot be identified. Enclose a statement of the number, type and the
reason of return.
2、Transportation charges
Transportation and insurance charges must be prepaid by the user for transporting the
product to Aeonmed for repairing.
7-17
8 Pre-Use Test
8 Pre-Use Test
After power on, the machine enter Pre-use test. The user can also do all the pre-use test
items on Standby screen.
Gas supply test Test will proceed when the hyperbaric oxygen is functional.
This test requires that oxygen supply is available. If oxygen
Oxygen sensor test. source is not available then a message “oxygen source is
inadequate” and the test cannot be carried out.
Leak test Internal leakage test.
Methods for testing the function of the alarm system for conditions are specified by
IEC60601-2-12. These are to be performed at the user’s discretion.
Perform the following procedure to prove operation of the Low MVe and High Airway
pressure alarms:
8-1
3. Press Start Ventilation and ventilate with default settings, set O2 to 21%.
5. Set the MVe low alarm limit to a value greater than the observed MVe reading.
6. Verify that a low level MVe low alarm is present on the 3rd breath.
9. Set the PAW upper alarm limit to a value lower than the observed Ppeak reading.
10. Verify that a low level High Airway Pressure alarm is present after 1 breath and that a
high level High Airway Pressure alarm is present at the start of the 4th breath.
If the pre-use test failed, you need to follow the way to troubleshoot.
8-2
8 Pre-Use Test
8-3
9 Network
9 Network
9.1 Overview
The ventilator can export patient monitoring data in HL7 format over a local area network
(LAN) using the TCP/IP transport protocol. The data can be sent to a server and then
forwarded to an Electronic Medical Records (EMR) system or sent directly to the EMR
system.
Network completes ventilator local IP and port settings, specifies IP (common PC) of node
and port number that connect and send data. After doing the above settings can establish
a connection to the ventilator. This machine data is sent to the remote destination node.
Currently send data by two ways. One is sent cyclically, the period is 1second, and
sending waveform data and monitoring values. Another is sent immediately, send alarm
information, ventilation parameters, alarm limit settings and so on at the beginning of
connection. When the information is changed that send synchronal information. Detailed
data communication protocols refer to "HL7 agreement."
Using data transmission functions require checking whether connection is open. Check by
following steps:
Connection can be achieved by two ways: First, the ventilator access network equipment,
such as Hub, switches and so on. Second, connect the ventilator and the destination PC's
network card with a network cable. Two connections use ordinary network cable.
2) Ensure that the ventilator set the IP address manually, not automatically obtain IP.
USB key board is connected to the ventilator while click two keys Win and
Esynchronously, then appear “My Device” interface, press“ control Panel”“Network and
Dial-up Connections”“PCI-E100E1”, enter the connection settings interface. Obtaining
the IP address is set to: Specify an IP address, and the ventilator is specified an IP
address.
Set IP address of ventilator and destination PC in the same network segment. Such as
9-1
PC's subnet mask is 255.255.255.0, subnet mask of ventilator also performs the same
setting, first three digits setting of ventilator IP address is the same as PC's IP address
and ensure that the final one setting is different other hosts on the network segment. Then
click “win” button, and select “reboot”, hot restart ventilator.
After this operation, ventilator enter the local IP and remote IP and click the Connect
button to complete the local IP settings by the ventilator settings →UI interface network
interface. You can enter the ventilator control panel to set without keyboard each time
Use “Ping” command to verify network connectivity on the PC. “Ping” open indicates the
network is available, you can click connection button to start data transmission. Click the
Connect button, if the ventilator displays IP settings are different the current IP, change
interface IP settings for ventilator IP, or retain the current setting. Then start data
transmission.
4) The PC starts data receiving test software to check whether the data is sent correctly.
The PC starts data receiving test software, input ventilator IP, click “connection”.
Ventilator system setting web interface, complete the above 1), 2), 3) operation, click
the “connection” that starts data transmission. Received data can be seen by PC-data
receiving software.
Click “Configurations” to enter the configurations interface. On this page, the user needs
to enter the password, as shown in Figure 9-1.
Figure 9-1
9-2
9 Network
Input the correct password to enter, Click “Site Configuration” to enter the site
configuration interface. As shown in Figure 9-2.
Figure 9-2
Figure 9-3
1) Local IP Address. Each number is from 1 to 255. If the input is outside this range, a
message is displayed: “* is invalidated. Please insure the value is between 1 and 255.” If it
is not set then display 0.0.0.0.
2) Local subnet mask. If it is not set then display 0.0.0.0. This is generally filled
255.255.255.0 or 255.255.0.0. This set is the same as subnet mask of destination PC.
3) Local port. Ventilator sends data through the port. Receiver gets data from the
ventilator through the port. Port 9101 is used by default. It can use from 65535 to 5000 as
9-3
the port number, If the input is outside this range, a message is displayed:” * is invalidate.
Please insure the value is between 5000 and 65535.”
4) Remote IP address. It is the IP address of the data receiving end. Once connected, the
ventilator only sends data to this address of the host. Usually the IP and ventilator IP is in
the same segment.
5) Remote port. It is data receiving end port. Ventilator establishes a connection with the
receiver through the port, receiving end receives data through the port. This machine is
generally associated with Ventilator port. Port 9101 is used by default. It can use from
65535 to 5000 as the port number, If the input is outside this range, a message is
displayed:” * is invalidate. Please insure the value is between 5000 and 65535.”
Data is packaged in HL7 format and transmitted by UDP packets. You can set up UDP
listening service to receive data, unpack the data according to the rules of the relevant
agreements. Detailed rules refer to "HL7 Protocol."
1) Waveform data. The period of Waveform is 1second. Receiver receives the waveform
data and saves into the buffer, and then it is drew on the interface according to the
sampling frequency (50ms). Drawing frequency and sampling frequency may be different.
It is adjusted according to the number of data in the buffer. Since the network delay and
buffer, there will be a delay drawing, the maximum delay is about 7 seconds.
2) Monitoring data. The transmission period is 1 second. Only the network latency is affect
monitoring data updates, it is generally not delay in the open network.
3) Alarm information. It sends all current alarm information when connection is established
and ventilation is started. It will also send the appropriate information when the new alarm
generates or alarm disappears.
4) Mode and ventilation parameters. It sends all current mode and ventilation parameters
when connection is established and ventilation is started. It will send the change mode
and ventilation parameters settings in the mode transitions.
5) Alarm limit. It sends all current alarm limits when connection is established and
9-4
9 Network
ventilation is started. It will send the change value when alarm limit is changed.
9.6 Troubleshooting
Right transmission data requires that network settings of ventilator is correct, receiver test
software can start normally, and network port of ventilator is installed correctly, and so on.
2) Receiver test software can move in windows XP and windows 7 system, if it can’t
receive data and ventilator and network cable are normally, network configuration
information may be incorrect. In this case, you can use the command:”netsh winsock
reset” that reset the network configuration and reboot the computer.
3) Ventilator network card. The above case is correctly, the network is still blocked,
namely, ping is failure. You should check that ventilator network card is properly
installed and connected. Use correct network port which is connected to the ventilator,
if you can Unicom network, show the network adapter or interface have problems.
Can replace card or connect afresh.
4) The PC's network connection is disconnected if you connect the ventilator and
computer with network cable, you should check whether the line is connected properly.
Make sure the cable is normal, then should open the ventilator screen, check the card
is properly connected.
5) If uld op is fail, make sure “remote IP Address” is the same as PC IP. If the PC uses a
wireless connection that can be disconnected, then verify properly connected.
9-5
10 Service Menu
10 Service Menu
CAUTION: On “System” of “Main Menu”, “Service” tab is reserved for service only.
Press “Service” to enter the “Service” page. On this page, the user needs to enter the
password, as shown in Figure 10-1.
Figure 10-1
Input the correct password to enter, There are six choices on this page: Calibration,
Event/Alarm Log, Machine Information, Language, Test Page and Update. The default
page is Calibration. See Figure 10-2.
Figure 10-2
10-1
10.1 Calibration
Figure 10-3
Press “Start” to start pressure sensor calibration. A progress bar and a message
“Calibration in progress. Please Wait” will be displayed as shown in Figure 10-4.
NOTE: During this period no other operation can be performed. Pressing other areas
will have no response.
10-2
10 Service Menu
Figure 10-4
2) Flow Sensor Calibration: Press “flow sensor” to enter the interface. A message is
displayed: “This step is to calibrate the flow sensor. Please connect the insp. Port and Exp.
Port directly with a tube” as shown in Figure 10-5.
Figure 10-5
Press the “Start” button to start flow sensor calibration, the remaining procedure is the
same as the pressure sensor calibration.
3) O2 Sensor Calibration: Press “O2 sensor” to enter the interface. A message displayed:
“Please verify that the oxygen source are connected correctly. Verify that the gas inlet
pressure is within specification.” A legend will also be shown. There are two keys below
the legend: “Start 21%” and “Start 100%”. Choose the needed one and Press, as shown in
Figure 10-6. The remaining procedure is the same as described above.
10-3
Figure 10-6
4) CO2 Sensor Calibration: Press “CO2 Sensor” to enter the interface, as shown in
Figure 10-7. A message displayed: “Disconnect the CO2 sensor with the adapter from
breathing circuit and ensure it is in ambient air. Wait 1 minute for warm up after the unit is
powered on or after installing an airway adapter. Press “Zero” when the State-Area turns
green”. Please follow the prompt message to calibrate.
Figure 10-7
Click the “Start” button to start inspiratory valve calibration, the remaining procedure is the
same as the pressure sensor calibration.
10-4
10 Service Menu
Figure 10-8
Click the “Start” button to start expiratory valve calibration, the remaining procedure is the
same as the pressure sensor calibration.
Figure 10-9
7) Touch Screen Calibration: Press “Touch screen” to enter the calibration interface. A
message is displayed: “This step is to calibrate the touch screen. The ventilator’s screen
will disappear during the calibration. Please follow the instruction in the calibration
program.” as shown in Figure 10-10.
10-5
Figure 10-10
Press the “Start” key to start, the remaining procedure is as described in steps above.
8) Leakage Test Calibration: Press “Leakage Test” to enter the test interface. A
message is displayed: “This step is to test the internal leakage of ventilator. Please
connect the Insp. Port and Exp. Port directly with a tube”, as shown in Figure 10-11.
Figure 10-11
Press “Start” to start the leakage test. A progress bar and a message “Test in progress,
please wait” will be displayed as shown in Figure 10-12.
10-6
10 Service Menu
Figure 10-12
9) Breath Circuit Test Calibration: Press “Breath Circuit Test” to enter the test interface.
A message is displayed: “This step is to test the compliance and leakage of breathing
circuit. Please connect the patient circuit to the T-piece, and plug up the patient end of the
T-piece”, as shown in Figure 10-13.
Before starting the test, ensure the patient circuit has been connected to the T-piece and
the patient end of the T-piece has been plugged up. Press the “Start” button to start the
breath circuit test.
Figure 10-13
10-7
10.1.2Calibration failure analysis
10-8
10 Service Menu
Press “Event/Alarm Log” to enter “Event/Alarm Log” page, as shown in Figure 10-14.
Figure 10-14
The middle area of this page is the message area. It can store up to 1000 messages,
including event messages and alarm messages. All messages will be listed in time
sequence. The top is the latest event or alarm message, and the bottom is the oldest. Use
the scroll bar to check all the messages. An asterisk (*) in front of an alarm message
indicates that alarm message was not displayed in the alarm message area. As shown in
Figure 10-15. The event/alarm log records all alarms and most actions.
Figure 10-15
Below is the message area of the Settings area resulting from highlighting an alarm. All
settings will be given here, as shown in Figure 10-16.
Figure 10-16
NOTE: Alarm/Event Log data is retained during a power interruption and can be viewed
when power returns.
10-9
10.3 Machine Information
Press the “Machine information” key to enter the Machine information page, this area
includes the following information, as shown in Figure 10-17.
1. Software Version:
a. UI
b. BDU
c. Power Supply
2. Runtime Hours
3. O2 Source Pressure
4. Atmospheric Pressure
Figure 10-17
10.4 Language
Press “Language” to enter the language screen. English and other languages are
available for the user to choose.
Press “Test Page” to enter the test page. There are five choices on this page: Demo, Vlt.
Monitor, Schematic, Cali. Data, Service Timer, Error Code and PT100 Cali. as shown in
Figure 10-18.
10-10
10 Service Menu
Figure 10-18
10.5.1Demo
Press “Demo” to enter the demo setting page, as shown in Figure 10-19. In this page, the
operator can turn on or turn off the demo mode, and test the broken pixel of the screen.
Figure 10-19
10.5.2Vlt. Monitor
Press “Vlt. Monitor” to enter the vlt. monitor page, as shown in Figure 10-20. This page
displays all the BDU and PS circuit monitoring value.
10-11
Figure 10-20
10.5.3Schematic
Press “Schematic” to enter the schematic page. The schematic is as shown in Figure
10-21. In this page, the operator can control all valves and read all sensors state.
The valves (#1 to #9) and the Exp. Valve Heater shall have two states: on and OFF; The
Turbine voltage range: 0-5, the units: V, the resolution: 1; The Insp. Valve range: 0-180,
the units: LPM, the resolution: 0.5; The IP Valve range: 0-100, the units: cmH2O, the
resolution: 1.
Figure 10-21
10-12
10 Service Menu
10.5.4Cali. Data
Press “Cali. Data” to enter the cali. data page, as shown in Figure 10-22. This page
displays the calibration data of Inspiration /Expiration Sensor.
Press “Refresh” button to update the calibration data and the zero AD value of the sensor.
Press “Manual Cali.” button to display the manual calibration page; in the page, the
operator can manually calibration the Insp/Exp sensor.
Figure 10-22
10.5.5Service Timer
Press “Service Timer” to enter the service timer page, as shown in Figure 10-23. This
page displays the service time about the ventilator, the turbine runtime, and the int./opt.
battery charging/discharging circle count.
Figure 10-23
10-13
10.5.6 Error Code
Press “Error Code” to enter the Error Code interface, as shown in Figure 10-24.
Figure 10-24
Press “PT100 Cali.” to enter the PT100 calibration interface, as shown in Figure 10-25.
Figure 10-25
10-14
10 Service Menu
Step 1: Press “Stop Heater” and wait for 40min to make the expiration valve to cool to
ambient temperature. Then measure and record the temperature of the expiration valve.
Step 2: Press “ ” turning the key background color to yellow, adjust the value equal to
the temperature recorded at Step 1, then read the value “ ” and record the
value in Textbox 1.
Step 3: Press “ ” turning the key background color to yellow, adjust the value equal to
the setting temperature, then read the value “ ” and record the value in
Textbox 3.(The setting temperature: the highest temperature of the expiration value
Step 4: In turn, press “Write”, “Read”, “Set Point”, “Read Point”, then determine that the
value of Textbox 2 is equal to the value of Textbox 1 and the value of Textbox 4 is equal to
the value of Textbox 3. If same, the calibration is finished; if not, carry out Step 5.
Step 5: Repeat the above steps.
10.6 Update
Press “Update” to enter the update page, as shown in Figure 10-26. In this page, the
operator can update the software.
Figure 10-26
10-15
10.7 Optional
Press “Optional” to enter the optional page, as shown in Figure 10-27. In this page, the
operator can choose configurable functions.
Figure 10-27
10-16
11 Troubleshooting
11 Troubleshooting
The error code is made up by 8 digits hexadecimal numbers, which are four byte using
binary system. The highest two bytes are internal AD test result flag, the lowest byte are
external AD test result flag. The meaning of each bit is shown in the below table.
External
AD##(0xF Test Point Possible Reason and Recommended Action
FFF00xx)
1. Remove the test lung and breath circuits,
perform the test again.
2. Don’s start PUT too quickly after quit from
ventilation.
Zero point of 3. Calibrate the pressure, flow rate sensor, Insp.
Bit 0
expiratory flow rate valve and Exp valve.
4. Proportional solenoid valve can’t close
completely. Replace proportional solenoid
valve.
5. Replace expiratory flow rate sensor.
1. Remove the test lung, connect Insp. port and
Exp. port directly with a tube.
2. Don’s start PUT too quickly after quit from
ventilation.
Zero point of
Bit 1 3. Calibrate the pressure, flow rate sensor, Insp.
inspiratory flow rate
valve and Exp valve.
4. Proportional solenoid valve can’t close
completely. Replace proportional solenoid valve.
5. Replace mother board.
1. Check the cable between the mother board to
the temperature sensor
2. The interface IC on the mother board is
Inspiratory broken. Replace mother board.
Bit 2
temperature 3. The input port of external AD chip on the
mother board is broken. Replace mother
board.
4. Replace temperature sensor.
11-1
Bit 3 -- --
1. Remove the test lung, connect Insp. port and
Exp. port directly with a tube.
2. Don’s start PUT too quickly after quit from
ventilation.
3. Calibrate the pressure, flow rate sensor, Insp.
valve and Exp valve.
Zero point of
Bit 4 4. Proportional solenoid valve can’t close
expiratory pressure
completely. Replace proportional solenoid valve.
5. The amplifier on the mother board is broken.
Replace mother board.
6. The input port of external AD chip on the mother
board is broken. Replace mother board.
7. The sensor is broken. Replace mother board.
1. Remove the test lung, connect Insp. port and
Exp. port directly with a tube.
2. Don’s start PUT too quickly after quit from
ventilation.
3. Calibrate the pressure, flow rate sensor, Insp.
valve and Exp valve.
Zero point of
Bit 5 4. Proportional solenoid valve can’t close
inspiratory pressure
completely. Replace proportional solenoid valve.
5. The amplifier on the mother board is broken.
Replace mother board.
6. The input port of external AD chip on the mother
board is broken. Replace mother board.
7. The sensor is broken. Replace mother board.
1. Proportional solenoid valve can’t close
completely. Replace proportional solenoid
valve.
2. The amplifier on the mother board is broken.
Bit 6 Barometric pressure Replace mother board.
3. The input port of external AD chip on the mother
board is broken. Replace mother board.
4.
The sensor is broken. Replace mother board.
Bit 7 -- --
Internal
AD(0xxxx -- --
x00FF)
Bit 15 -- --
Bit 16 -- --
Bit 17 -- --
Bit 18 -- --
11-2
11 Troubleshooting
Error Test
Possible Reason and Recommended Action
Code Point
1000 T12 DC The A5V out is out of range. Replace PS board.
1001 T1 DC The W24V output is out of range. Replace switch PS.
1002 T13 DC The P12V output is out of range. Replace PS board.
1004 T19 DC The N24V output is out of range. Replace PS board.
1008 T24 DC PS board failure. Replace PS board.
1009 T14 DC The A12V output is out of range. Replace PS board.
1013 -- During expiratory valve heating the temperature is out of range.
Replace PS board.
2000 T31 DC During expiratory valve heating the output voltage is out of range.
Replace PS board.
2001 T30 DC During expiratory valve heating the output current is out of range.
Replace PS board.
3006 -- PS board communication failure. Replace PS board.
3005 -- BDU communication failure. Replace core board.
3002 -- Software mismatch failure. Update software.
11-3
11.1.2Other Errors
11.1.3Service Tools
Sequence
Name Quantity Remark
number
11-4
12 Appendix A
This service manual provides procedures for testing and maintaining the Ventilator VG70.
It is not intended to be a complete maintenance document; therefore, it contains no
detailed disassembly or reassembly instructions.
Refer any repairs or adjustments that exceed the scope of this manual to the Service
Center of Aeonmed by calling 800-810-8333.
This manual contains proprietary information. It is intended for use only by individuals
qualified in the installation and maintenance of the Aeonmed ventilator. Receipt, purchase,
or possession of this document in no way confers or transfers any other rights for the use
of this information. Disclosure or reproduction of the enclosed, without the written
permission of Aeonmed is prohibited.
This manual is intended for use only by technicians who have successfully completed
Aeonmed training on this product.
Aeonmed believes the information herein is accurate but accepts no responsibility for
errors, omissions, or misrepresentation.
12-1
13 Appendix B
Pneumatic diagram
13-1
14 Appendix C
14 Appendix C Specifications
System
Operating Conditions
Non-operating Conditions
Power Supply
Ventilator
General Dimensions:
Ventilation Delivery Unit: 322 wide x 375 deep x 366 high (mm)
System with Cart (optional): 547 wide x 675 deep x 950 high (mm)
14-1
Cart: 25kg
Gas Supply
Supplied gases must be free of water, oil and particles
DISS, NIST
User Interface
Attaches to the cart, a rail, or other mounting system.
Room temperature: 20 ℃
Inspiratory Channel
Pressure Drop
Compliance
Maximum 2 ml/cmH2O (with Fisher &Paykel MR 810 humidifier and reusable silicone adult
patient circuit)
Flow Range:
14-2
14 Appendix C
0 to 60 liters/min. (Child)
Adult: 60liters/min
Child: 30liters/min
Expiratory Channel
Pressure Drop (Resistance)
Compliance
Maximum 2 mL/cmH2O (with Fisher & Paykel MR 810 humidifier and reusable silicone
adult patient circuit)
PEEP Regulation
Monitoring
Expiratory Minute Volume
Range: 0 – 60 liters/min.
Range: 0 – 4000 mL
Resolution: 1 mL
O2 Concentration
Range: 18 – 100%
14-3
Resolution: 1%
Airway Pressure
Resolution: 1 cmH2O
14-4
15 Appendix D
Replacement
No. Code Description Remarks
period
1 122007310 NBP mother board As needed ——
2 122007306 NBP power supply board As needed ——
3 122007308 NBP main control board As needed ——
4 122007309 NBP core board As needed ——
5 122007311 NBP interface board As needed ——
Gas supply pressure
6 122005527 As needed ——
board
Encoder knob
7 122007482 As needed ——
component
8 122007483 Mute button component As needed ——
NBP display interface
9 122007304 As needed ——
board
Common display platform
10 122007303 As needed ——
core board
Common display platform
11 122007305 As needed ——
mother board
12 122007485 Alarm lamp component As needed ——
NBP indicate lamp
13 122007486 As needed ——
component
Differential pressure
14 122010401 As needed ——
transmitter board
15 210003677 Turbine As needed ——
Flower sensor TSI
16 210002403 As needed ——
840201
17 122007463 Inspiration control valve As needed ——
18 240000089 Check valve As needed ——
Electromagnetic valve
19 240000252 As needed ——
SY114-6G
Mini-electromagnetic
20 240001020 As needed ——
valve N332/2B
High rupture fuse
21 210003856 As needed ——
T250V/3.15A
15-1
Replacement
No. Code Description Remarks
period
22 210003676 Brushless motor driver As needed ——
23 210003693 Touch panel AMT9542 As needed ——
24 210003675 LCD As needed ——
Exhalation valve
25 122007471 As needed ——
components
Exhalation valve core
26 122007890 As needed ——
assemble
Switch power supply
27 210003491 As needed ——
SNP-G169-M
28 210003678 Switch power supply As needed
29 122007322 Fan cable As needed ——
Consumables list:
15-2
This manual No.: 130015826
Directive 93/42/EEC
concerning Medical Devices
Edition 01.04
May. 2018