1 How much blood does heart pump every day? 7000 L/Day
2 The Upper Chambers of the heart called. Atria
3 Mitral Valve is located at. Between LA and LV
4 Endocardium has specialized Cardiac Fibers called. Purkinje Fibers
5 Which Node initiates Electric Conduction for stimulating SA Node
Blood flow in Heart ?
6 During Ventricular Contraction this valve prevents blood Tricuspid Valve
from moving from right ventricle into right atrium
7 During Ventricular Relaxation, this valve prevents blood Pulmonary Valve
from moving from pulmonary trunk into right ventricle.
8 The Only Normal conduction pathway between the atrial AV Node
and ventricular syncytia is provided by this node.
9 The most important ions that influence the heart action Potassium and Calcium
are.
10 The conduction of the electrical impulses throughout the P wave.
atria is seen on the ECG as the _____ Wave.

A B C D
1 The delay in the AV node forms much of the PR segment P Wave QRS T Wave
______ pattern on the ECG waveform. segment
2 The spread of electrical activity through the QRS PR PQ T Wave
ventricular myocardium forms the _______ complex  segment Segment
pattern on the ECG waveform.
3 Atrial repolarization and Ventricle QRS Q Wave R Wave P Wave
depolarization occurs during this waveform complex
pattern of ECG wave
4 Which Lead has Following Connection? LA Lead -I Lead-II Lead-III None
connection to non inverting amp. Input and RA
connecting to inverting amp. Input
5 Which Lead has Following Connection? LL Lead -III
connected to non inverting input LA
connected to inverting input.
6 Which Lead has Following Connection? LA Lead aVL
connected to non-inverting input while RA and
LL are summed at inverting input augmented
(amplified) Voltage for Left arm.
7 Which Lead has Following Connection? RA Lead aVR
connected to non-inverting input while LA and
LL are summed at inverting input augmented
(amplified) Voltage for Right arm
8 Wilson’s Central Terminal configuration is Unipolar
used with these leads. Chest Leads
9 In Interdigital ECG Signal, Signal is taken from Signal
taken
between 2
fingers
10 Heart Problem that results into 3rd Degree of Acute
Block or complete failure of conduction Myocardial
between atria and ventricles. Infarction(A
MI)
11 Right leg driver circuit is used in a feedback 60 Hz noise
configuration to reduce and drive
noise on
patient to a
lower level.
12 Notch Filter is used to reduce 60 Hz noise
13 Isolation Amplifier used mainly for Isolation of
Patient
from
Electric
Shock
14 ECG Signal Frequency Range is DC to 500
Hz
15 Unipolar Limb Leads are also called Augmented
Limb Leads

1. How much blood does heart pump every day?