Sie sind auf Seite 1von 5

L-0016 (220319-15)

UNITY-xr  ™

Specification sheet

The UNITY-xr is an analytical thermal desorption


instrument for the simultaneous analysis of VOCs, SVOCs
and thermally labile compounds collected onto
sorbent tubes.

1. System features
• ‘Universal’ TD platform allowing analysis of • Overlap mode (desorption of a subsequent
compounds over a wide volatility range AND the sample while a previous sample is still running)
ability to select low flow path temperatures for optimises productivity.
compatibility with labile compounds: • Stand-alone injector that can be connected to
–– Highly effective retention of ultra-volatiles. any make of GC(–MS) and does not interfere
–– Quantitative recovery up to n-C44. with other GC accessories.

–– Quantitative recovery of labile compounds. • Interface to the GC typically via a direct


coupling to the analytical column. Part of a GC
–– Simultaneous analysis of volatiles and semi- inlet may be required for back-pressure-
volatiles. regulated electronic pneumatic control. This can
• Quantitative re-collection of both tube and trap be used to provide electronic carrier gas control
desorption split flow to allow repeat analysis. (ECC) through the entire TD–GC(–MS) analyser,
• Compatible with 3½″ sorbent tubes: Stainless and stabilise retention times independent of
steel, inert-coated stainless steel, and glass. split flow and other analytical settings.
Also 4½″ glass sorbent tubes (optional). • Extended standby mode reduces instrument
• Electrically-cooled focusing trap cools rapidly power consumption when not in use.
and is easy to maintain. • Small footprint for operation in mobile labs or
• Stringent, method-compliant leak test other confined environments.
(no‑flow/ambient-temperature) is carried out on • Flexible upgrade routes include:
every sample. Failed tubes are retained intact. –– One or two integrated mass flow controllers
• Splitless, single-split and double-split options (MFCs) for electronic control of split and/or
ensure compatibility with samples over a wide desorb flows.
concentration range (ppt to percent). –– ULTRA-xr for 100-tube automation, or ULTRA-
• Trap heating rates up to 100°C/s and xr + ULTRA-xr Pro for 199-tube automation,
backflush desorption combine to facilitate allowing unattended operation all weekend.
splitless operation at column flows ≥2 mL/min, –– Options for automated, multi-channel
therefore maximising sensitivity. canister/bag analysis or round-the-clock
• Pre-purge of air to vent and selective on-line air monitoring (with Kori-xr water-
elimination of water and solvents minimise management accessory).
analytical interference.

Markes International Ltd


T: +44 (0)1443 230935 F: +44 (0)1443 231531 E: enquiries@markes.com

www.markes.com
Page 2

2. System controls N.B. The tube oven heats from ambient to the
selected temperature at the start of tube
2.1 Control software desorption in order to minimise risk of flash-
vaporisation and split discrimination when
• Markes Instrument Control (MIC) allows: analysing samples with unknown water/solvent
–– Automated, unattended sequencing of tube, content.
on-line or canister/bag samples.
• Desorption time:
–– Addition, insertion or skipping of samples in
–– Range: 0 to 600.0 min.
active sequences.
–– Adjustable in 0.1 min increments.
–– Rapid set-up of methods using pre-
programmed parameters for: (a) standard 2.4 Focusing trap
methods including VDA 278, US EPA TO‑17,
US EPA 325 and PAH analysis; (b) conditioning • Quartz focusing trap: 2 mm i.d. where packed
methods for popular sorbent tubes and and 0.9 mm i.d. at the sample input/output end.
focusing traps. Collar at non-sampling end makes trap easy to
change.
–– Automated, intelligent system self-checking
diagnostics, including leak isolation. • Trap can be packed with between one and four
sorbents.
–– Preventative maintenance feedback with
usage counter – indicates when parts could • Backflush desorption ensures quantitative
be replaced to avoid instrument downtime. retention and release of wide boiling range
samples.
–– Export of sequence history to .csv file.
• Trap low temperature:
–– Set-up in English or Chinese language.
–– Range: –30°C to 50°C.
2.2 Desorption modes –– Adjustable in 1°C increments.
• Two- (or three-) stage desorption – Normal –– Temperature limits are user-settable within the
two-stage desorption of a sample, with the stated range.
additional option of an elevated-temperature
–– Uniform electrical cooling applied over full
purge.
60 mm length of sorbent bed.
• Tube conditioning – Desorption of the sample
• Trap desorption:
tube for cleaning purposes with all the effluent
directed to vent, i.e. away from the focusing trap –– Default setting is ballistic heating, which
and other important components of the sample reaches rates of 100°C/s during the first
flow path critical stages of secondary (trap) desorption.
• Trap conditioning: Desorption of focusing trap –– Alternatively, programmed trap heating rates
for cleaning purposes, or for obtaining a system from 1°C/s to 40°C/s can be selected.
blank. • Trap high temperature:
• Other modes including on-line monitoring, direct –– Range: 35°C to 425°C.
sampling, sample stacking and automation are
–– Adjustable in 1°C increments.
available when the system is configured with the
appropriate module. –– Temperature limits are user-settable within the
stated range.
2.3 Primary (tube) desorption oven
–– Uniform heating applied over full length of
• Temperature: sorbent bed.
–– Range: 35°C to 425°C. • Hold time at trap high temperature:
–– Adjustable in 1°C increments. –– Range: 0–60 min.
–– Temperature limits are user-settable within the –– Adjustable in 0.1 min increments.
stated range.

Markes International Ltd


T: +44 (0)1443 230935 F: +44 (0)1443 231531 E: enquiries@markes.com

www.markes.com
Page 3

2.5 Sample flow path • An additional carrier gas pre-purge can be


carried out at elevated temperature to remove
• Temperature range:
water or other interfering solvent if required.
–– Valve: 50°C to 210°C.
• The focusing trap can be selected to be in or out
–– Transfer line: 50°C to 250°C. of line during either of the pre-purge stages.
–– Both adjustable in 1°C increments. • The split can be selected to be open or closed
–– Temperature limits are user-settable within the during either of the pre-purge stages.
stated range. • Tube dry-purge: This is an alternative to pre-
–– Uniform heating. purge. The sample tube is purged with dry
carrier gas in the sampling direction to eliminate
• Constructed entirely of inert materials: PTFE,
water and oxygen from the back of the tube. The
quartz, inert-coated stainless steel and
purged air is directed away from the analytical
uncoated, deactivated fused silica.
column and is sent to vent.
2.6 Pneumatics • Tube dry-purge time:
• Requires a pressure-controlled 0–60 psig –– Range: 0–60.0 min.
(0–415 kPa) supply of helium or nitrogen carrier –– Adjustable in 0.1 min increments.
gas under manual or electronic control.
• Tube dry-purge flow rate (when MFC is fitted):
• Electronic mass flow control (option) is settable
between 2–500 mL/min (helium), and –– Range: 2–500 mL/min.
2–250 mL/min (nitrogen). –– Adjustable in 1 mL/min increments.
• Requires a pressurised supply of dry air or • Trap dry-purge: The focusing trap can be
nitrogen (dewpoint below –50°C) at 50–60 psig optionally purged with dry carrier gas after
(340–415 kPa). The dry gas is used for both primary (tube) desorption and before the trap is
pneumatic actuation of the valve and for purging desorbed. The purge flow is then directed
the focusing trap box. through the focusing trap, in the trapping
(sampling) direction, to sweep any remaining
N.B. Helium cannot be used as the dry gas
interferences to vent.
supply.
• Trap dry-purge time:
2.7 Pre-desorption checks and controls –– Range: 0 to 60.0 min.
• Leak test: Each tube is pressurised and –– Adjustable in 0.1 min increments.
subjected to a stringent, ambient-temperature • Trap dry-purge flow (when MFC is fitted):
leak test without carrier gas flow. Failed tubes
are not desorbed, but are preserved intact for –– Range: 2–500 mL/min.
operator attention. –– Adjustable in 1 mL/min increments.
• Pre-purge: Each tube can be optionally purged • Trap dry-purge temperature:
with carrier gas (in the desorption direction) at –– Range: –30°C to 50°C.
ambient temperature to remove oxygen before
desorption. The air is purged to vent and none of –– Adjustable in 1°C increments.
it is allowed to reach the analyser (e.g. GC–MS). –– Temperature limits are user-settable within the
• Pre-purge time: stated range.
–– Range: 0–60.0 min. 2.8 Sample splitting and quantitative
–– Adjustable in 0.1 min increments. re‑collection for repeat analysis
• Pre-purge flow rate (when MFC is fitted): • The UNITY-xr split can be operated in the
following ways:
–– Range: 2–500 mL/min.
–– During primary (tube) desorption only (inlet
–– Adjustable in 1 mL/min increments.
split).

Markes International Ltd


T: +44 (0)1443 230935 F: +44 (0)1443 231531 E: enquiries@markes.com

www.markes.com
Page 4

–– During secondary (trap) desorption only (outlet • Constructed of stainless steel, inert-coated
split). stainless steel or glass.
–– During both desorption stages, i.e. double- • With or without sorbent packing.
split operation (inlet and outlet split). • With or without TubeTAG RFID tags.
–– During neither desorption stage, i.e. splitless
operation. 3.3 Ambient operating conditions
• The split can be turned on or off during system • Temperature: 15°C to 30°C.
standby. • Relative humidity: 5–95% RH (non-condensing).
• Split and desorb/trap flows are controlled by
3.4 Power requirements
needle and solenoid valves downstream of the
sample flow path. • 100–240 V, 50/60 Hz, 650 W (UNITY-xr self-
–– Optional mass flow controllers provide adjusts to local voltage input).
electronic control of split and desorb/trap 3.5 Gas consumption
flows.
–– Configurations including Air Server-xr or • Dry air or nitrogen: ~100 mL/min.
CIA Advantage-xr modules have mass flow • Carrier gas consumption is method-dependent
controllers configured as standard for (typically 5–200 mL/min).
controlling both split and trap flows.
3.6 Minimum PC specification
• The split vent line contains a charcoal filter in
front of the control valves (and MFC) to prevent For TD control:
contamination of the valves/MFC and laboratory
• CPU: 1 GHz 64-bit dual-core or better.
atmosphere. The charcoal filter has the same
external dimensions as a standard sorbent tube. • RAM: 4 GB.
The charcoal filter is connected to the main • Hard disk space: 2 GB.
heated valve via a short, inert, heated flow path.
• Graphics card: DirectX 9 or later.
• When required, the charcoal filter can be
• Display: 1024 × 768 display.
replaced with a conditioned sorbent tube to
quantitatively re-collect the split effluent from • Operating system: Windows 7, 8.1 or 10, 64-bit,
tube and trap desorption (inlet and outlet split). English.
This capability allows repeat analysis, method/ • Other requirements: Windows-compatible
data validation and archiving of critical samples. keyboard and mouse; one free USB connection
for UNITY-xr communication with PC.
3. System specification
3.7 Safety and regulatory certifications
3.1 Dimensions and weight • The instrument is designed and manufactured
• Height: 46 cm (18.1″). under a quality system registered to ISO 9001.
• Width: 16 cm (6.3″). • The instrument complies with the essential
requirements of the following applicable
• Depth: 54 cm (21.3″). European Directives, and carries the CE mark
• Weight: 16 kg (35 lb). accordingly:
3.2 Tubes accommodated –– Low Voltage Directive 2014/35/EU
–– EMC Directive 2014/30/EU.
• 3½″ (89 mm) long × ¼″ (6.4 mm) o.d. tubes.
–– ROHS Directive 2011/65/EU.
• 4½″ (114 mm) long × 6 mm o.d. tubes (optional).
• The instrument conforms to the following
N.B. Kits are available to allow users to product safety standards:
interchange between the 3½″ and 4½″ versions.
–– IEC 61010-1:2010/EN 61010-1:2010.
–– IEC 61010-2-010/EN 61010-2-010:2014.

Markes International Ltd


T: +44 (0)1443 230935 F: +44 (0)1443 231531 E: enquiries@markes.com

www.markes.com
Page 5

–– IEC 61010-2-081/EN 61010-2-081:2015. Trademarks


–– Canada: CSA C22.2 No.61010-1:2012.
UNITY-xr ™, Air Server-xr ™, CIA Advantage-xr ™,
–– USA: ANSI/UL 61010-1:2012. TubeTAG™, ULTRA‑xr ™ and ULTRA-xr Pro™ are
• The instrument conforms to the following trademarks of Markes International.
regulation on electromagnetic compatibility
DirectX® and Windows® are trademarks of
(EMC):
Microsoft Corporation.
–– IEC 61326-1/EN 61326-1:2013.
3.8 GC remote cable connections
Markes International shall not be liable for errors contained in
• UNITY-xr includes a GC interface cable that this specification or for incidental or consequential damages
in connection with the supply, performance or use of this
connects to the ‘ready’ output and ‘start’ input
equipment.
of the GC(–MS) and data-handling systems.
Markes International reserves the right to change information,
• The cable supports automatic start of the entire descriptions and specifications within this document without
analytical system when the UNITY-xr focusing notice.
trap desorbs, and allows UNITY-xr to check the
‘ready’ status of the analyser and associated data
handling.
• The UNITY-xr focusing trap will not desorb unless
and until it receives a ‘ready’ signal from the
GC(–MS) system.

4. System options
Accessory and upgrade options include:
• Integrated electronic mass flow control of split
and/or desorb flow. MFCs available with flow
range between 2–500 mL/min (helium), and
2–250 mL/min (nitrogen). Allows split ratios from
zero to 125,000 : 1 to be used with standard
(60 m × 0.25 mm i.d.) capillary columns.
• ULTRA-xr 100-tube autosampler and automated
outlet split re-collection.
• ULTRA-xr plus ULTRA-xr Pro 199-tube
automation and automated inlet and outlet split
re-collection.
• Air Server-xr options (3- or 8-channel) for
continuous on-line air/gas monitoring or analysis
of multiple canisters/bags.
• CIA Advantage-xr options (T or HL models) for
method-compliant analysis from whole-air/gas
samples, collected using multiple canisters/bags
or on-line monitoring.

For more information about our products and


services, please visit www.markes.com.

Markes International Ltd


T: +44 (0)1443 230935 F: +44 (0)1443 231531 E: enquiries@markes.com

www.markes.com

Das könnte Ihnen auch gefallen