K3 Unit Chair

Osstem Implant
Technical Support Division
Understanding the K3 Unit Chair

Osstem Implant
A-1
* Contents
Ⅰ K3 Product Summary

Ⅱ Overview and Operation

Ⅲ Organization of Medical Station

Ⅳ Connected Equipment

3/170
Ⅰ K3 Product Summary

4/170
Ⅰ K3 Product Summary
Description of Components

Dental Light

LED Light Device to illuminate Unit


the mouth
Device for supplying air &
water

Chair

Device that moves into


position

Foot Control Doctor Table

Operation for medical Medical instruments


Treatment and action buttons

5/170
Ⅰ K3 Product Summary
Product’s Unique Features (1/6)

6/170
Ⅰ K3 Product Summary
Product’s Unique Features (2/6)

* Attractive - 2013 GOOD DESIGN award (Korea)

Healing Tube
- Round & Cylinder shape

7/170
Ⅰ K3 Product Summary
Product’s Unique Features (3/6)

* Comfort & Safety - Ergonomic Patient Chair


Backrest
Thin & Durable
Wide/double seat(optional)

Optimal space
for leg room

Hydraulic
Z-movement
Armrest
Movable, more easily moved out of the way
*Safety Oil cylinder
Anti oil leakage device

*Emergency switch
Automatic safety shut-off

Cover-all base plate


Less arbitrary trembling

Power switch Base cover


Valley-buried for safety Anti-corrosion plastic cover *Motor Pump
High durability
8/170
Ⅰ K3 Product Summary
Product’s Unique Features (4/6)

* Easy - Displays animated information


Doctor Table 『Durable, Stable and Integrative』

9/170
Ⅰ K3 Product Summary
Product’s Unique Features (5/6)

Foot pedal 『No hand control needed』

10/170
Ⅰ K3 Product Summary
Product’s Unique Features (6/6)

* Korean World-class Product Award 2016

11/170
Ⅱ Overview and Operation

12/170
Ⅱ Overview and Operation 『CHAIR』

Chair Part Overview


• Device that supports the body of a patient lying down for treatment,
and enables the selection of a treatment position
§ Chair parts and components
Headrest
Backrest
One-touch type headrest. You can
adjust the angle after pressing the Supports the patient's body lying
button. on the chair for treatment and
adjusts the angle of the backrest
depending on the treatment
Armrest position.
Seat
Serves to relax the patient,
providing a feeling of security
Supports the patient's body lying
against the movement of the
on the chair for treatment and
treatment device and fixing the
selects the treatment position for
patient's arm during the treatment
the patient
Emergency

Protects the patient and system


automatically in an emergency by
installing a safety switch to the channel.
When there is an obstacle, the chair
operation stops and the chair moves up.

13/170
Ⅱ Overview and Operation 『UNIT』

Unit Part Overview


• Adjusts the pressure of air and water flowing in the chair,
and supplies water and air to each part
§ Parts and Components of Unit
Water outlet Spittoon

Water supply device for the Receptacle into which the patient
patient to rinse/gargle can gargle and spit after treatment

Assist Table
Cup holder
Table featuring manual chair
Device to hold a cup for
operating buttons, vacuum
rinsing/gargling
syringe, saliva ejector, 3-way
syringe, and easy-to-use
Spittoon Filter holder

Filters out impurities from the


patient's saliva to prevent
Suction Filter
them from being discharged
to the drain Filters out impurities
generated from the suction
operation in the patient's
mount to prevent them from
being discharged to the
outside
14/170
Ⅱ Overview and Operation 『Dr TABLE』

Table Part Overview


• Controls overall functions of Unit Chair
• Holds the treatment instrument (handpiece, scaler, 3way syringe) and controls its functions
• Functions as a rack to store treatment instruments

§ Parts and Components of Table


Expandable table Chart Bracket

Main Table
Control buttons

3-Way Syringe Mouse Pad

High Speed Handpiece


Slop bucket hanger
Low Speed Handpiece

Scaler

15/170
Ⅱ Overview and Operation 『LED LIGHT』

LED Light Part Overview


• Device that illuminates the patient's mouth during the treatment
• Outstanding product that does not generate shadow, and achieves almost natural
colors
§ Parts and Components of LED Light Table

Light Balance Arm

Reflector Able to control the light


deflection; can be stopped at
Function to gather any position desired by the
scattered light operator

Blue Cut Off Filter

Controls the resin


hardening speed at the
time of treatment (filter can
be turned on/off in a slide
structure)
Handle Power/Light brightness
adjustment knob
Adjusts the direction of
light during treatment Turns the power on/off and
adjusts the brightness

16/170
Ⅱ Overview and Operation 『FOOT』

Foot Control Part Overview


• Device to operate instruments such as handpiece or scaler
• Controls some functions of Unit Chair

§ Parts and Components of Foot Control

Foot Rest
Manual Chair Controller
Used to relocate the foot
controller using the foot Used to move the chair and
backrest up or down
Return Position

Moves from the spittoon


position to the position Spittoon Position
before gargling
Moves to the spittoon
position for gargling

Handpiece Operating Pedal

Used to control the handpiece RPM


(air pressure)

17/170
Ⅱ Overview and Operation 『Headrest』

Headrest Part Overview


• Device to supports the patient's head
• Device to relocate the patient's head to the optimal treatment position

§ Parts and Components of Headrest


Headrest Slide Bar

Supports the headrest and


Headrest performs ascending and
Seat descending operation

Headrest Button

Adjusts the headrest angle


to fit the patient

18/170
Ⅱ Overview and Operation 『BALANCE ARM』

Balance Arm Part Overview


• Composite medical equipment in which devices for dental treatment are combined
• Device to relocate the patient's head to the optimal treatment position

Balance Arm
Large Arm Stops the Doctor Table at
the position desired by the
Supports the balance arm operator through the
and Doctor Table operation of springs

19/170
Organization

of Medical Station

20/170
Ⅲ Organization of Medical Station
Doctor Table Instruments

§ Old Hand-Piece Line-up

ü SLOT #1 : Air driven high speed

ü SLOT #2 : Air driven high speed

ü SLOT #3 : Air driven low speed


Select
Elec. driven low speed
SLOT #1
ü SLOT #4 : Scaler
SLOT #2
SLOT #3
SLOT #4
21/170
Ⅲ Organization of Medical Station
Doctor Table Instruments

§ New Hand-Piece Line-up (K3 Power)

Classification SLOT #1 SLOT #2 SLOT #3 SLOT #4

Elec. driven Elec. driven


Air driven
Basic High speed Low speed Scaler
High speed
(200,000 rpm) (40,000 rpm)

SLOT
Air driven Elec. driven #1 SLOT
- - #2 SLOT
High speed High speed #3 SLOT
#4

Option

Air driven Air driven


- -
Low speed Low speed

22/170
Ⅲ Organization of Medical Station 『HIGH HP』

High Speed Hand-Piece


• The key instrument used for cutting teeth or for treatment
• Air type is widely used, electric type is relatively high price

§ Parts of Handpiece
Head
Body Coupling

ü Head: Section to put in the bur. Cartridge is embedded in the head, and sprays water when the bur is
operated.
ü Body: Handle section. Coupling is the generic term for the section to detach or connect the handpiece.
ü Coupling: The rear section is connected to the handpiece tubing, and the front section is connected to
the body of the handpiece. This is rotatable by 360 degrees, and reduces the burden on the wrist.

23/170
Ⅲ Organization of Medical Station 『HIGH HP』

High Speed Hand-Piece

§ Parts and Components of Handpiece

Head Body Sym


Coupling Name Function descriptions
bol
When the handpiece is used,
compressed air is delivered to the rotor
Air,Water
A Spray
mounted in the head section and water
is sprayed to prevent the bur from
overheating
C Fiber optic is located inside the
F B Light
product to perform optical functions
Used to refine the shape of tooth and
C bur
remove crown
D E
Path to discharge the compressed air
D Air exhaust
inside the handpiece
G H Path to flow the compressed air inside
A B E Air drive
the handpiece
Connector Supplies current to the fiber optic to
F connection turn the head lamp on
status

G Chip Air drive Air path to spray water

H Water drive Path to supply water to the handpiece

24/170
Ⅲ Organization of Medical Station 『HIGH HP』

High Speed Hand-Piece

§ Classification by type of handpiece bearing

High speed hand piece

Image

Rotating type Air bearing Ball bearing

R.P.M 300,000~450,000

Main use For tooth cutting

Components Main body

Attachment H.P

25/170
Ⅲ Organization of Medical Station 『LOW HP』

Low Speed Hand-Piece


• The key instrument used for cutting teeth or treatment
• Generally used for dental molds (electric type and air type)

§ Parts of Handpiece

Angle Motor Motor connection

ü Angle: Section to put the bur in. Generic term for the section to detach and connect a motor section which
rotates the bur and spray water and air.
ü Motor: Connected and fixed to the motor connection. The front section is connected to the angle.
ü Connection: Connected to the motor section to supply air, water and electricity.

26/170
Ⅲ Organization of Medical Station 『LOW HP』

Low Speed Hand-Piece

§ Classification by type of handpiece motor

Air motor Electric motor

Product

RPM Average of 22,000 Maximum of 35,000 to 40,000

Operating type Air Electricity

• No heat generation from the main body • Heat generated from the main body
Characteristics • Power loss at low speed • Power loss at low speed

27/170
Ⅲ Organization of Medical Station 『LOW HP』

Low Speed Hand-Piece


• Angle is a device that is mounted on the motor, and is used for tooth cutting,
root canal treatment and producing dental molds.
Can be classified into straight angle and contra angle depending on the shape.
§ Classification of handpiece angle
Straight Angle Contra Angle

Product

vBent at about 30˚


vCA: General angle
vFG: Used to be put in the FG bur
Characteristics vI-shaped vProfi: (Pomace process after scaling)
vEndo Angle : (Deceleration angle for root canal treatment)
(engine classified by gear ratio - 10:1, 27:1, 64:1, or 128:1)
vAccelerating angle : (acceleration angle for cutting)

28/170
Ⅲ Organization of Medical Station 『SCALER』

Scaler Hand-Piece
• Device used to remove dental plague on the teeth
• Ultrasonic type and air type are available

§ Classification by operating type


Ultra sonic scaler Air scaler

Image

• Operating frequency: 20 to 60kHz • Air pressure: 2 to 3 bar


Characteristics • Additional controller required • Water supply required
• Water supply required • High scaler durability

29/170
Ⅲ Organization of Medical Station 『SCALER』

Scaler Hand-Piece

§ Scaler types
Manu Operating
Product name Type Optic Image
facturer frequency
Ultrasonic
1 Compact S dmetec 24~32(KHz) -
(Piezo)
Ultrasonic
2 Sonic buff 300 DIT 28(KHz) -
(Piezo)
Ultrasonic
4 Varios 150 28~32(KHz) Available
(Piezo)
Ultrasonic
5 S5 EMS 27~30(KHz) -
(Piezo)

Ultrasonic
6 S5L EMS 27~30(KHz) Available
(Piezo)

7 Ti-Max S950 NSK Air 6.2~6.4(KHz) Available

8 MICRON SSS Micron Air 16~17(KHz) -

30/170
Ⅲ Organization of Medical Station 『3-WAY』

3-Way Syringe
• Device used to spray water and air during dental treatment to maintain a clean and dry
state
• Consists of hose tab to provide water and air, mount to adjust the flow and nozzle to
spray water and air

§ Parts of 3-way Syringe


3-Way Syringe Tip Air spray button
Stick inserted inside the oral
Air is sprayed when the
section for cleaning. Tip can be
button is pressed
replaced depending on the
treatment type
Water spray button

Water is sprayed when the button is pressed


※ A mixture of air and water is sprayed
when both button are pressed at the same
time

31/170
Ⅲ Organization of Medical Station
Assist Table Instruments

§ Assist Table Line-up

Saliva ejector

3-Way Syringe

Vacuum syringe

32/170
Ⅳ Connected Equipment

33/170
Ⅳ Connected Equipment
Dental Unit Chair System Flow

Pipe
line
Secretion
Machine room Water purification system

Air Compressor Suction Unit

Drainage
WATER
Air supply supply

34/170
Ⅳ Connected Equipment
Air Compressor
• Device that compresses air and stores it in a tank. This compressed air is supplied to
the handpiece and syringe used in dental treatment.
CATTANI-
ACOM21 Dmega Metasys DURR Dental
Heungjin
(ACT-E160-2H50) (Mega-Air) (META Air 250) (Tornado 2)
(AC 200C)

Image

Tank
40 ℓ 40 ℓ 30 ℓ 50 ℓ 20 ℓ
capacity
Maximum
1.5 ( HP ) 2( HP ) 2( HP ) 3(HP) 2(HP)
power
Use
6.2 bar 5 to 8 bar 5 bar 5 to 7 bar 7.8 bar
pressure
Rate of flow 214 ℓ/min - 160 ℓ/min 225 ℓ/min 170 ℓ/min
Noise 71 dB 55 dB 71 dB 69 dB 70 dB
Number of
4 2 2 3 2
cylinders
Max #of
3 4 4 5 4
35/170 chairs
Ⅳ Connected Equipment
Suction Unit
• Device used to suck foreign substances or saliva from a patient's mouth during
treatment
• Classified into dry suction and wet suction
§ Classification by operating type
Dry type suction Wet type suction

Image

• Bacterial reverse infection and environmental


• Loud noise problems may occur, as water connected to form a
• Significant dust vacuum pressure is drained directly to the sewer
• More expensive than wet type • Low noise level
Charact • Patient's saliva, blood, tooth cutting debris, • Constant water supply required
eristics amalgam, mercury, etc. stored in separate • High risk of freezing and rupturing in the winter
containers to facilitate waste treatment due to water supply
• Low dust generation thanks to water supply
• Relatively quiet
36/170 • Suction power is higher than dry type
Thank you
Mechanical Theory
& Operating Principles

Osstem Implant
A-2
* Contents
Ⅰ Main components

Ⅱ Operating principles of the K3 equipment system

39/170
Ⅰ Main components

40/170
Ⅰ Main components 『Chair Part』

Chair Part

§ Chair part names and descriptions


Component
No. Image Function
name

Converts the mechanical energy supplied from the motor to


1 Motor Pump compressive energy for the operating oil in the hydraulic
system

Moves the backrest up/down by operating the piston with


2 Cylinder the pressurized oil

3 Oil tank Stores the operating oil in the hydraulic system

When electricity flows through the coil, the plunger goes up


and the valve is opened; if electricity is cut off, the valve is
4 Solenoid valve automatically closed by the weight of the plunger
Controls the ascending/descending speed of the chair and
its backrest
41/170
Ⅰ Main components 『Unit Part』

Unit Part

§ Unit part names and descriptions


Component
No. Image Function
name

1 Water Filter Filters out foreign substances in the water

Water Solenoid Opens and closes the water outlet based on the electrical
2 signal
Valve

3 Warmer Warms the cold water for the patient to gargle

Adjusts and reduces the air pressure coming from the


4 Air Regulator outside

42/170
Ⅰ Main components 『Dr. Table System』

Dr. Table Part

§ Dr. Table Part names and descriptions


Component
No. Image Function
name

Opens and closes the air supply tube based on the electric
1 Solenoid valve signal

2 Master Block Supplies water and air to the handpiece and scaler

Water Solenoid Controls the amount of water supplied to the device


3 (handpiece, scaler, 3 way syringe)
Valve

Controls the amount of water supplied to the device


4 Water valve (handpiece, scaler)

43/170
Ⅱ Mechanical operating principles

44/170
Ⅱ Mechanical operating principles『Oil Pressure System』
Oil Pressure System

§ Configuration of oil pressure system

ü Names and functions of main parts

- Motor pump Backrest


Cylinder
Generates the pressure required for the cylinder operation

- Solenoid Valve
Supplies pressure received from the motor pump to the cylinder
Oil Line
Controls the cylinder speed by adjusting the flux of the operating oil
Up/Down
- Oil Tank Cylinder

Stores the energy source oil and removes impurities

- Up/Down Cylinder
Drain Line
Moves the chair up/down

- Backrest Cylinder
Oil Motor
Moves the backrest seat up/down pump Solenoid Valve
Tank

<Hydraulic system diagram>

45/170
Ⅱ Mechanical operating principles『Oil Pressure System』
Oil Pressure System

§ Main components of oil pressure system #1

ü - Hydraulic motor pump

- Converts the mechanical energy supplied from the outside to compressive energy
for the operating oil in the hydraulic system

- Hydraulic system in K3 plays the role of controlling chair motions as intended by


the operator such as moving the chair and backrest seat up/down

Gear motor
operating Product image Features
principles
drive gear
• Simple structure, low cost
out in
• Breakdowns from foreign substances in the hydraulic oil
are rare
• Short use life due to heavy load on the bearing
Idle gear • Heavy variation in torque

46/170 <Hydraulic pump description>


Ⅱ Mechanical operating principles『Oil Pressure System』
Oil Pressure System

§ Main components of oil pressure system #2


ü - Hydraulic Solenoid Valve

- When electricity flows through the coil, the plunger goes up and the valve is
opened; if electricity is cut off, the valve is automatically closed by the weight of the
plunger.

- Controls the ascending/descending speed of chair and backrest seat


Chair Up/Down Backrest Up/Down
In/Out In/Out
in

injector
body

solenoid
Return line
return spring

needle valve Pressure In


O-ring
diffuser nozzle

Injector closed Injector open

<Hydraulic solenoid valve description> <Chair operating solenoid valve>


47/170
Ⅱ Mechanical operating principles『Oil Pressure System』
Oil Pressure System
§ Main component #3 of oil pressure system
ü Hydraulic cylinder

- Pressurized oil delivers the energy to a power unit such as a piston to operate the device

- Currently a single acting hydraulic cylinder is applied to the hydraulic system of the K3
unit chair

Operating
Product image Features
principles
heat
impact

• Features a single oil port, and operates unilaterally.


• Backward stroke is achieved by gravity or spring force.
• Saves power and is used for press or simple operating
device

<Single acting hydraulic cylinder description>


48/170
Ⅱ Mechanical operating principles 『Water System』
Water System

§ Main components of water system #1


ü Water Filter

- Filters foreign substances from the water supplied from the outside

- Water filter is needed to ensure patient hygiene at the time of treatment and
prevent damage to the device by foreign substances

Product image and configuration Features


Water supply Bronze Filter
outlet
Bronze Filter
Water IN
• Removes impurities in the liquid at temperatures ranging
from high temperature to extremely low temperature using
a filter to enable the mixture of liquids
• Applied to remove impurities in the highly viscous liquid,
Water Out Water Out
liquid metal or general gas or vapor.

<Water Filter description>

49/170
Ⅱ Mechanical operating principles 『Water System』
Water System

§ Main components of water system #2

ü Water Solenoid Valve

- Controlled by an electrical signal; delivers and cuts off the water supply

- K3 unit chair features a solenoid valve that opens and closes 2 water lines

Water IN

Water Solenoid v/v


(open/close the water)
Water IN warm warm
Warmer water water
Water Solenoid v/v
(open/close the hot water line)
<Water Solenoid Valve description>
50/170
Ⅱ Mechanical operating principles 『Water System』
Water System

§ Main components of water system #3

ü Water Valve

- Controlled by electrical signal; opens/closes the water valve by using the air

- Supplies required water to devices such as handpiece or scaler

Product image and configuration Features


Water IN

Solenoid Valve
Water Valve
Water Valve
• Although it is possible to open/close the water
supply line manually, a solenoid valve is installed
to enable automatic opening and closing by
electrical signals

Air IN Master Block 3Way-Syringe

<Water Valve description>

51/170
Ⅱ Mechanical operating principles 『Water System』
Water System

§ Main components of water system #4

ü Water Solenoid Valve


- Water is supplied to the water cutter valve and controls the water volume supplied
to the scaler and handpiece
- Knob-type adjuster can be easily used to control the water supply volume

Air IN

OFF Air Line


Sol v/v Water Line
Air IN Foot
ON Air IN
Handpiece &
Water IN Scaler

Air Sol v/v Master Block Water Cutter v/v


<Water Cutter Valve description>
52/170
Ⅱ Mechanical operating principles 『Air System』

Air System

§ Main components of air system #1

ü Solenoid Valve

- Opens/closes the air supply line through electrical signals

- Operates for each part, such as handpiece or scaler

Air IN
Air Line
OFF
Water Line
Sol v/v
Air IN Foot
ON Air IN
Handpiece &
Water IN Scaler

Air Sol v/v Master Block Water Cutter v/v


<Solenoid Valve description>

53/170
Ⅱ Mechanical operating principles 『Air System』

Air System

§ Main components of air system #2

ü Master Block

- Supplies water/air to the handpiece or scaler; easy to control the supply volume

- Made of brass which is corrosion resistant and highly durable

Air IN Air Line


Water Line
OFF

Sol v/v
Air IN Foot
ON Air IN
Handpiece &
Water IN Scaler

Air Sol v/v Master Block Water Cutter v/v


<Water Cutter Valve description>
54/170
Ⅱ Mechanical operating principles 『Air System』

Air System

§ Main components of air system #3


ü Air Regulator

- Sets and adjusts the air pressure

- Adjusts the volume of air flowing inside the unit and controls the air pressure for
each device
Product image and configuration Features
Pressure
adjusting knob
Pressure adjusting knob
Pressure
gauge • Adjusts the pressure of outside air flowing in the unit
Pressure gauge
Filter
• Checks the pressure of outside air flowing in the unit
Filter
• Removes moisture in the outside air flowing in the unit
Pull the pressure Turn the pressure adjusting
adjusting knob knob clockwise or
upwards to unlock counter-clockwise

<Air Regulator description>


55/170
Ⅱ Mechanical operating principles
Air & Water System Diagram

56/170
Thank you
Electrical Theory
& Operating Principles

Osstem Implant
A-3
* Contents
Ⅰ Basic electrical theory

Ⅱ Configuration of electrical parts in the K3 chair

Ⅲ K3 power supply system

Ⅳ Main PCB configuration and functions

Ⅴ Table PCB configuration and functions

Ⅵ Power PCB configuration and functions

Ⅶ Other PCB configuration and functions

59/170
Ⅰ Basic electrical theory

60/170
Ⅰ Basic electrical theory
Current & Voltage

• Current: Flow of charged particles. The unit representing the magnitude of current is A (amperes).

• Voltage: Difference in potential (potential energy for charge).


The unit representing the magnitude of the voltage is V (volts).

• Resistance: The degree of disturbance of current flow.


The unit representing the magnitude of resistance is Ω (ohms).

• Relations: The magnitude of the current flowing in the electrical circuit is proportional to the magnitude of the
applied voltage, and is inversely proportional to the resistance.

Current V

The slope of the


Lamp
straight line
represents
resistance
I
Example Circuit
Voltage-Current chart
61/170
Ⅰ Basic electrical theory
DC & AC

• Direct Current (DC): State in which current flows unilaterally in an electrical circuit
• Alternating Current (AC): State in which current reverses its direction many times at
regular intervals

Current

Current flow
Direct
current (DC)

Time DC

Current

Alternating Current flow

current (AC) Time

AC

62/170
Ⅰ Basic electrical theory
PCB & PBA

• PCB: Printed Circuit Board


- Item on which separate electrical parts are laid out and supported, and those parts
are electrically connected
• PBA: Printed Board Assembly: PCB with parts mounted
- Refers to a fully operational PCB with electrical parts assembled
- Although a PBA is distinct from a PCB, a PBA can often be referred to as a PCB

PCB PBA
63/170
Ⅰ Basic electrical theory
HARNESS

• HARNESS: Bundle of insulated wires with various lengths


- Path for supplying the required electricity and signals to each part of the Unit Chair,
functioning like nerve tissue or blood vessels in the human body
- In the practice, it is also referred to as harness assembly or harness ass’y
(contraction), although such terms are not correct English

64/170
Ⅱ Configuration of Electrical Parts in the K3 Chair

65/170
Ⅱ Configuration of Electrical Parts in K3 Chair
K3 Electrical Parts
11 proprietary 3 modules from
8 peripheral parts
PCBs third parties

LED Controller PCB


Air Sol. Valve
LED Sensor PCB

LED Module PCB Elec. Low H/P Module

Scaler Module
Assist PCB

Sensor PCB
Warmer Table PCB
Water Sol. Valve
LCD Module

Oil Sol. Valve

Oil Motor USB PCB

Transformer
Main PCB
Main Switch
Light PCB
Noise Filter

Power PCB
Foot PCB

66/170
Ⅱ Configuration of Electrical Parts in K3 Chair
K3 Controller Flow

Input Signal processing Operating

67/170
Ⅱ Configuration of Electrical Parts in K3 Chair
K3 UNIT

Sensor PCB Assist Membrane

Assist PCB

Main PCB

Light PCB
USB PCB

Noise Filter
*different options for
Foot PCB export models

68/170
Ⅱ Configuration of Electrical Parts in K3 Chair
K3 CHAIR

Transformer ①

Back-rest
Oil Solenoid Valve ② Potentiometer ①

Noise Filter ③
Emergency
※ Domestic specification’s Noise filter is placed here
(For chairs for export, it is mounted on the Unit)
Switch ②

Power PCB ④
Chair
Oil Motor ⑤ Potentiometer ③

69/170
Ⅱ Configuration of Electrical Parts in K3 Chair
K3 Dr. TABLE

Elec. Low H/P Module ① Table PCB ② Air Sol. Valve ③ Scaler Module ④

LCD Module

Dr. Table Membrane


Cart Type

70/170
Ⅱ Configuration of Electrical Parts in K3 Chair
K3 LED LIGHT

Shade-less Reflector LED Controller

LED Sensor
LED Module
(inside here, not seen)

Blue-cut control

ON/OFF Switch

Autoclave Handles
71/170
Ⅲ K3 Power Supply System

72/170
Ⅲ K3 Power Supply System
K3 Power Diagram

AC Input Noise Filter Transformer Rectifier Sol. Valve Sol. Valve

DC-DC
AC24V Converter DC24V DC24V
(LM2596)

Water level Control

Voltage Voltage
Regulator DC12V Regulator DC5V
(LM7812) (LM7812)

Rectifier Control Optic

Voltage Voltage
DC
Regulator DC5V Regulator
(LM2596) (LM2596) 1.6~4V

Warmer Elec. H/P & Scaler

AC AC
AC24V
24V 24V

Light PCB Light

AC
AC12V
12V
73/170
Ⅲ K3 Power Supply System
Transformer
Standard for export Optional for export Domestic
(280VA) (350VA) (scheduled to be discontinued)
MODEL : DN-BET 0.28K MODEL : DN-BETA 0.35K

Water separator
MODEL : CS1

※ The type can be distinguished by the color of the label attached to the top of the transformer. The
74/170 optional product for export (350VA) is used when water separator(CS1) is mounted!
Ⅲ K3 Power Supply System
Type Plug & Socket Spec. Type Plug & Socket Spec. Type Plug & Socket Spec.
220~240V
Voltage 100~127V 16A 220~240V
Voltage Voltage
Current 15A 2pin 16A
Current Current
Pins 3pin Yes 3pin
Pins
A GND Yes F GND
Korea, Indonesia, K Pins
Yes
Russia, Kazakhstan, GND
Country : United States, Country
Jordan, Iran, Country
Canada
Vietnam, Turkey,
Romania
100~127V, 220~240V
230V(Thailand) 13A
Voltage Voltage 220~240V
15A Voltage 3pin
Current Current 10A & 16A
3pin Current Yes
B Pins
Yes G Pins Hong Kong, L Pins 3pin
GND GND Yes
Japan, Taiwan, GND Singapore, Malaysia,
Country Country Italy, Chile
Thailand, Mexico, Country Bangladesh,
the Philippines Cambodia, UAE
(Dubai), UK
220~240V
Voltage Voltage
220~240V Voltage 220~240V 15A
Current Current
2.5A 3pin
C Pins
2pin
Current 16A M Pins
Yes
GND
No H Pins 3pin GND
Country: GND Yes Country:
North Korea Country: Israel

220~240V
Voltage Voltage 220~240V
5A Voltage 220~240V
Current Current 10 & 20A
3pin
D Pins
Yes
Current 10A N Pins 3pin
GND
India I Pins 2 or 3pin GND Yes
Country GND Yes Country : Brazil
Country Australia, China

220~240V 220~240V
Voltage 220~240V Voltage
16A Voltage 16A
Current 10A Current
2pin 3pin
E Pins
Yes
Current
3pin O Pins
Yes
GND
Mongolia, France, J Pins
Yes GND
Country: GND Country
Belgium Switzerland,
Country
Rwanda

75/170
Ⅲ K3 Power Supply System
Transformer Wiring Diagram
100V 120V 220V 230V 240V

120V 100V 0V 130V 120V 120V 100V 0V 130V 120V 120V 100V 0V 130V 120V 100V 0V 130V 120V 120V 100V 0V 130V
100V 0V 100V 0V 120V 100V 0V 100V 0V 120V 100V 0V

0V 24V 0V 12V 0V 24V 0V 12V 0V 24V 0V 12V 0V 24V 0V 12V 0V 24V 0V 12V
0V 24V 0V 24V 0V 24V 0V 24V 0V 24V

Japan United Korea North Australia Cambodia Malaysia


States China Korea Thailand UAE (Dubai) Kazakhstan
Taiwan Hong Kong Singapore United
Canada Vietnam Indonesia Kingdom
Bangladesh India Iran
Russia Mongolia Australia
The Jordan Turkey
Philippines Rumania

76/170
Ⅲ K3 Power Supply System
Grounding diagram

Building
ground
4

2
1

6
7
8

77/170
Ⅳ Main PCB configuration and functions

78/170
Ⅳ Main PCB configuration and functions
Main PCB Configuration

§ Block Diagram

Assist Membrane Light PCB


UNIT UNIT
Sensor PCB Water Sol. Valve
Water volume
Spittoon Sol. Valve
adjustment
Chair VR/Limit
CHAIR Main PCB
B.Rest VR/Limit

Emergency SW

Trans AC power BASE BASE


Power PCB
Foot Pedal

Bilateral communication (USART)

Table PCB
Input device Output device

79/170
Ⅳ Main PCB configuration and functions
Main PCB Power Supply
• Supplied with 2 types of AC 24V from trans located at the base.
(2 types : PCB power supply and operation of peripherals)

§ Main PCB power supply

Main PCB
Base end trans

80/170
Ⅳ Main PCB configuration and functions
Inter-Power Supply Configuration on Table PCB
• Supplies AC 12V (PCB) and AC24V (Scaler, Bien-Air Low Elec. HP) power
from Main PCB to Table PCB

§ Table PCB power supply

Table PCB

ü If in the Main PCB, the squared section ( ) and the dotted circled section ( ) are switched
in connection, "Communication Error" will be displayed on the LCD and the Main DC5V power
supply will be cut.
81/170
Ⅳ Main PCB configuration and functions
Water Sol. Valve Electricity Control Signal
• Controls the water supply to the water outlet and spittoon by receiving solenoid valve
control operation signals and applying DC 24 power

§ Water solenoid valve connection

DC 24V power for solenoid valve

Main PCB

82/170
Ⅳ Main PCB configuration and functions
Warmer Electric Signal
• Supplies power to warm the water supplied to the cup (based on 32℃)
§ Warmer connection

AC 24V power for warmer


Main PCB

ü The function to turn the warmer power on/off is located outside the UNIT, and the embedded
bimetal stops the temperature rise once the temperature reaches a set value.

83/170
Ⅳ Main PCB configuration and functions
FUSE Types and Capacities
Warmer AC 24V 3.15A - warmer(2.5)

B. AC24V 1.6A - Main PCB(1.5) + Table PCB(1.5) + LCD PCB(1)

Size : 5X20mm

Warmer AC24V B. AC24V

Main PCB

84/170
Ⅳ Main PCB Configuration and Functions
vChair and Backrest Ascending/Descending Limit Switch and Variable Resistance
• Limit Switch : When the chair or backrest moves to the upper/lower threshold value,
the switch is closed (prevents any additional operations)
• Variable resistance : Detects the current position of chair or backrest
during ascending/descending
• As the number of connector pins is identical, it is important to take care not to insert
incorrectly when connecting using a wire harness

B.REST

CHAIR

Main PCB

85/170
Ⅳ Main PCB configuration and functions
Water Volume Adjustment for Water Outlet and Spittoon
• Auto cup water level adjustment (with the internal timer function, up to 10 seconds)
• Adjusts the duration for which water is out from the spittoon (Min : 1sec. Max : 10sec)

UNIT
Main PCB

86/170
Ⅳ Main PCB configuration and functions
Water Volume Adjustment for Water Outlet and Spittoon
• Judges whether DC5V power is supplied to the sensor PCB, and whether a cup is placed
• Identifies whether the water level has reached the set position

Sensor PCB

UNIT
Main PCB

87/170
Ⅳ Main PCB configuration and functions
v Water Supply Cup Recognition and Water Volume Recognition

Front Rear
Light
emitting
sensor

Sensor signal
Light receiving Side
sensor
level signal
[Cup recognition type] [Cup water level]

Voltage for adjusting supplied water volume (4 to 12V)

CPU Potential Water level recognition

Level voltage
+ L/H
Ref. voltage L H
- Ref.
(2V)
Voltage

Volume

[Cup water volume detection circuit]

88/170
Ⅴ Table PCB configuration and functions

89/170
Ⅴ Table PCB configuration and functions
Table PCB Configuration

§ Block Diagram

Main PCB Soll. valve signal


UNIT Table
Foot RPM Fiber Optic signal

Trans AC power High H.P 1,2

ELEC. H.P
Table PCB Scaler
Table Membrane Table

H.P Slot Micro SW SLOT LED

Speaker
Full Duplex Table
Bien-Air Module Scaler PCB
Communication

Full Duplex communication


Table
LCD Panel
LED Module PCB
Input device Output device

90/170
Ⅴ Table PCB configuration and functions
Table PCB Power Supply Unit Connection and Status Check
• When the applicable power is input properly, the LED is turned ON
§ Power connection to Table PCB

Main PCB
AC24,DC24V

DC 5V LED
Power supply to Table PCB

DC 24V LED
Power supply to Sol. valve

91/170
Ⅴ Table PCB configuration and functions
Table PCB Power Supply Unit Connection and Status Check

Elec. Low HP(Bien Air) Scaler HP(Bien Air)


AC 24V for power supply AC 24V for power supply

Air Solenoid Valve


For power supply to PCB DC 24V for power
DC5V supply

92/170
Ⅴ Table PCB configuration and functions
Connection Configuration Between Table PCB and LCD PCB
• Device to display information on the LCD window to let the user know the
operational status or commands of the chair
§ Power connection to LCD PCB

LCD PCB

ü Necessary to check whether the copper plate is properly shielded in the event of any abnormal
phenomenon, such as the LCD window in use being turned off or whitened

93/170
Ⅴ Table PCB configuration and functions
Connection Configuration Between Table PCB and Membrane
• Membrane-type key interface, with 23 embedded dome switches
• The rear side is attached with a double-sided adhesive tape
(care should be taken when attaching to the panel)
• Matrix type is applied to reduce I/O ports when connecting to the Table PCB

94/170
Ⅴ Table PCB configuration and functions
Connection Configuration Between Table PCB and Bien air

Table

Bien-
Power
Air

Bien-Air Controller

1 2 3 4 5 6 ON

Serial Mode Communication Disabled


error check (fixed)
95/170
Ⅴ Table PCB configuration and functions
Operation Check Between Table PCB and Bien air

K3 Table Board

K3 Foot
Pedal On/Off Controller
,Pedal AD
Power Supply
24Vac ±10%
RS-232C
(UART0) 1
2
K3 Main Board 3
4
5
6
Signal GND

3
2 (-)
RXIN

32Vdc ± 10%
1
(+)
TXOUT

5
4
3
2
1

Bien-Air Controller

96/170
Ⅴ Table PCB configuration and functions
Operation Check Between Table PCB and Scaler
• AC 24V power supply from Table PCB to Scaler PCB
• Start signal delivered from Table PCB to Scaler PCB

AC24V

Table Start
Signal

SCALER

97/170
Ⅵ Power PCB configuration and functions

98/170
Ⅵ Power PCB configuration and functions
Power PCB Configuration

§ Block Diagram

Table Membrane Hydraulic motor

Keys Hydraulic solenoid


Base valve
UNIT (chair operating valve)
Main PCB

Key, Suction Power PCB


Suction start signal
Assist Membrane
Assist Micro SW
Foot Controller

Input device Output device

99/170
Ⅵ Power PCB configuration and functions
Configuration of Harness Connectors (5EA)

§ HARNESS configuration

⑤ Hydraulic solenoid valve

④ Hydraulic Motor ① MAIN PCB

② Suction device
③ POWER SWITCH

100/170
Ⅵ Power PCB configuration and functions
Check Power PCB Operation via LED

T
CD
Hydraulic motor CU
2 3 BU
BD

AC 220V Input 1

Control
Basic operation Checking operation Checklist
function
Operates when
Hydraulic Measure the AC 220V voltage Check the motor LED
chair goes up
motor when the motor operates illumination 1
or backrest goes up
Each operating solenoid
Check the operating voltage of each solenoid valve
Hydraulic Chair moves up/down valve
Ex) When the chair goes down,
Solenoid Valve Back moves up/ down Check the signal LED
check the 220V voltage at T & CD test point
illumination
2

Perform short check on the output section


Check the CVS LED
Suction(C.V.S) On/Off to check whether CVS signal is properly controlled
illumination 3
(Suction On : short / Off : open)

101/170
Ⅵ Power PCB configuration and functions
Connection with Main PCB
• Sends chair operation signal from Main PCB
• Assist PCB relays the suction operation signal via Main PCB to Power PCB

§ Operation signal connection to harness

Power PCB

Main PCB

102/170
Ⅵ Power PCB configuration and functions
Connection with Suction Start Signal
• Delivers start signal for the suction device in the dental clinic

Power PCB
Suction pump

103/170
Ⅵ Power PCB configuration and functions
Connection Between Power PCB and Power Supply (AC220V) Line
• AC 220V power received from the outlet via the Main Switch is supplied to operate
the hydraulic motor and AC solenoid valve, etc.

§ AC 220V power supply

Main Switch
(AC220V)

Power PCB

104/170
Ⅵ Power PCB configuration and functions
Power PCB Operation

• Turns the solenoid On/Off through the AC Sol_valve control signal received from Main PCB
• This moves up or down the chair and backrest to a desired position

§ Connection with AC Solenoid Valve

Hydraulic solenoid
valve

Power PCB

ü Caution
è Take care to avoid an electric shock, as AC220V power is directly applied to the solenoid valve
105/170
Ⅵ Power PCB configuration and functions
Power PCB Operation
• Turns the hydraulic motor On/Off based on the control signal received from Main PCB
• When a command to move the chair and backrest up is input, the hydraulic motor
works to provide the hydraulic pressure
§ Connection between hydraulic motors

Power PCB

ü Caution
è Take care to avoid an electric shock, as AC220V power is directly applied to the solenoid valve
106/170
Ⅶ Other PCB configuration and functions

107/170
Ⅶ Other PCB configuration and functions
Foot PCB
• There are buttons for moving the chair and backrest up/down, rinsing, and returning to
the original position
• For Elec. HP, you need to connect the 3 pin harness of the foot pedal to the Main PCB

§ Connection between Main PCBs

Foot Pedal Foot PCB

Main PCB

108/170
Ⅶ Other PCB configuration and functions
Foot PCB
• Able to control the functions required for the operation of chair position, and operate the
treatment chair both manually and automatically
• Able to control the pedal at both high and low speeds
• Foot emergency: If there is an emergency during chair operation, press the pedal and the
chair stops
Numb
Function How to operate
er
The backrest goes down while the controller is
Descend backrest
pushed to the left
The backrest goes up while the controller is
Ascend backrest
pushed to the right

The chair goes up while the controller is
Ascend chair
pushed backwards
The chair goes down while the controller is
Descend the chair
pulled forward

② Return Position Moves to the position before gargling

③ Rinsing Position Moves to the gargling position

④ Pedal Operates the handpiece and scaler

⑤ Footrest Place the foot to move


109/170
Ⅶ Other PCB configuration and functions
Light PCB
• AC 12V power is applied from the Base Transa and delivered to the light.
§ Position and connection

AC 12V Input
Light PCB connection

Light PCB position

110/170
Ⅶ Other PCB configuration and functions
Light PCB
• Supplies operating power (DC 12V) to the Light PCB
• Delivers Light Relay ON/OFF signal

§ Light Control Harness Connection

Main PCB

111/170
Ⅶ Other PCB configuration and functions
Light PCB

§ Production of new Light PCB for water sanitation

• Light signal
à yellow 5p

• Light output • Solenoid Valve


à white 2p à For 3-way syringe
control

• Light input
à blue 2p

[Functions for each connector] • Sol Valve power (Main PCB)


à DC24V red 4p

112/170
Ⅶ Other PCB configuration and functions
Assist PCB

§ Position and Components

Assist Membrane

Syringe Suction Small


Suction Large
Assist PCB

113/170
Ⅶ Other PCB configuration and functions
Assist PCB
• Activated when either of them is lifted - parallel circuit
§ Micro-switch connection and operation

DC24V

Assist PCB

114/170
Ⅶ Other PCB configuration and functions
Assist PCB

§ Micro-Switch + Suction Operation

How to check the suction operation LED indicator

- On/Off: Displayed 1

Suction
Assist PCB
Main PCB Power PCB
115/170
Ⅶ Other PCB configuration and functions
USB Extension PCB
• When using various kinds of USB type dental equipment, it can be inconvenient when the
device is a long distance from the PC
• Chair can be relayed to the PC using USB extension PCB
§ USB connection available

USB 3m

USB – Portable X-ray

USB 3m

116/170
Thank you
How to Use the K3 Unit Chair

Osstem Implant
A-4
* Contents
Ⅰ K3 UNIT CHAIR

Ⅱ Description of each part

Ⅲ How to clean

119/170
Ⅰ K3 UNIT CHAIR

120/170
Ⅰ K3 UNIT CHAIR
Unit Chair Overview
• Composite medical equipment in which devices for dental treatment are combined
• Equipment for patient treatment by dentist and dental hygienist

§ K3 UNIT CHAIR components

Monitor Arm
Light Pole
Dental Light

HANARO Console Balance Arm

Head rest
Chart bracket

Chair

Arm rest

Doctor Table

Foot Control
Unit
121/170
Ⅰ K3 UNIT CHAIR
Unit Chair Overview

§ Part names and descriptions of their functions


Part Name Description
Dental Light Illuminates the patient's mouth during treatment

HANARO Console Combines the functions of a cup dispenser, tissue holder and belongings cradle

Headrest Supports the patient's neck and head

Chair Supports the patient's body, providing the optimal patient posture for the treatment

Relaxes the patient by providing the feeling of security against the movement of the treatment
Armrest device and fixing the patient's arm during the treatment

Foot Control Device that controls some functions of the Unit Chair

Monitor Arm Supports and fixes the medical monitor

Light Pole Supports the dental light, and fixes the console and the monitor arm

Balance Arm Supports the Dr.Table and allows it to be moved up/down/left/right by the user

Chart bracket Holds the patient's medical chart

Doctor Table Holds the treatment instrument (handpiece, scaler, 3-way syringe) and controls its functions

UNIT Controls the pressure of the air and water input to the chair and supplies it to each part of the
chair
122/170
Ⅱ Description of each part

123/170
Ⅱ Description of each part 『 Main switch 』

Main switch

§ How to turn the chair ON/OFF


1. Connect the power cable 2. Turn the main switch On

Connect the K3 power cable to the main outlet Turn the power on. The main switch will turn
green and the system will start (check the LCD
screen)

Domestic Overseas

| Indicates that the power supply is turned [ON]


○ Indicates that the power supply is turned [OFF]

124/170
Ⅱ Description of each part 『Doctor Table Panel

Doctor Table Panel

§ Dr.Table button functions

System power LCD Function buttons


button

Film viewer Treatment position


button selection button
『Configuration of Dr.Table』

125/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ System power button


Button Function LCD screen Description

System Power
Turns the system power ON/OFF
ON/OFF

§ Film viewer button


Button Function LCD screen Description

FILM VIEWER
Film Viewer Allows the operator to check the patient's x-ray image

126/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ Treatment position selection button (manual movement)


Button Function LCD screen Description

The chair moves up while the button is pressed and held


Raise the chair
down

The chair moves down while the button is pressed and held
Lower the chair
down

Raise the The backrest moves up while the button is pressed and
backrest held down

Lower the The backrest moves down while the button is pressed and
backrest held down

127/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ Treatment position selection button (memory-based relocation)


Button Function LCD screen Description

Moves to the spittoon position to allow the patient to


Rinsing Position
gargle after treatment

Return Position Returns to the treatment position after patient gargles

Moves the chair to the initial position (lowest position) to


Start Position make it easy for the patient to get out of the chair

Automatically moves to the saved Memory 1 position

Memory 1
Set a desired chair position and press and hold the
Memory 1 button for at least 3 seconds to save the
position

128/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ Treatment position selection button (memory-based relocation)


Button Function LCD screen Description

Automatically moves to the saved Memory 2 position

Memory 2
Set a desired chair position and press and hold the
Memory 2 button for at least 3 seconds to save the
position

Automatically moves to the saved Memory 3 position

Memory 3
Set a desired chair position and press and hold the
Memory 3 button for at least 3 seconds to save the
position

129/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ Function buttons
Button Function LCD screen Description

Dental Light
Turns the dental light ON/OFF
On/Off

Handpiece water
injection Turns the handpiece water injection function ON/OFF
On/Off

Handpiece optic
Turns the handpiece optic ON/OFF
On/Off

Normal direction, Selects the rotating direction of the low speed handpiece
Reverse direction (normal or reverse)

Turns the timer ON/OFF (press and hold for at least 3


Timer
seconds to reset)

130/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ Function buttons (continued)


Button Function LCD screen Description

Adjusts the speed to a value saved in Low speed Handpiece


RPM1
RPM 1
Select a desired speed and press and hold the RPM1
button for at least 3 seconds to save

Adjusts the speed to a value saved in Low speed Handpiece


RPM2
RPM 2
Select a desired speed and press and hold the RPM2
button for at least 3 seconds to save

Adjusts the speed to a value saved in Low speed Handpiece


RPM3
RPM 3
Select a desired speed and press and hold the RPM3
button for at least 3 seconds to save

131/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ Function buttons (continued)


Button Function LCD screen Description

Minutes/Seconds/
Function to change the mode for Minutes/Seconds/RPM
RPM -
adjustment
Mode Selection

Increase
Minutes/Seconds/ - Increases Minutes/Seconds/RPM when selected
RPM

Decrease
Minutes/Seconds/ - Decreases Minutes/Seconds/RPM when selected
RPM

132/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ Doctor Table Error Messages


Code name Description LCD Screen

[ET1] Dr. Table Key error

[EF1] Foot Key error

[EF2] Foot Pedal error

[EA1] As. Table Key error

133/170
Ⅱ Description of each part 『Doctor Table Panel』

Doctor Table Panel

§ Doctor Table Error Messages [continued]


Code name Description LCD Screen

[ES1] Water supply volume detection sensor error

[EM1] Emergency switch error

[EA2] Auxiliary table forced water supply (cup) error

[EA3] Auxiliary table forced water supply (spittoon) error

134/170
Ⅱ Description of each part 『Assistant Table Panel』

Assistant Table Panel

§ Assistant Table Button Functions

Function buttons Treatment position


selection button

『Configuration of assistant
table』

135/170
Ⅱ Description of each part 『Assistant Table Panel』

Assistant Table Panel

§2-pole selection switch


Button Function LCD screen Description

Raise chair Moves the chair up while the button is held down

Lower chair Moves the chair down while the button is held down

Raise backrest Moves the backrest up while the button is held down

Lower backrest Moves the backrest down while the button is held down

136/170
Ⅱ Description of each part 『Assistant Table Panel』

Assistant Table Panel

§ Function buttons
Button Function LCD screen Descriptions

Moves to the spittoon position to let the patient gargle


Rinsing Position
after treatment

Return Position Returns to the treatment position after gargling

Moves the chair to the initial position (lowest position) to


Start Position
make it easy for the patient to get out of the chair

Water supply
Supplies water to the cup while the button is held down
( to cup )

137/170
Ⅱ Description of each part 『Assistant Table Panel』

Assistant Table Panel

§ Function buttons (continued)


Button Functions LCD screen Descriptions

Memory 1 Automatically moves to the saved Memory 1 position

Memory 2 Automatically moves to the saved Memory 2 position

Dental Light
Turns the dental light ON/OFF
On/Off

Water supply to Supplies water to the spittoon while the button is held
spittoon down

138/170
Ⅱ Description of each part 『Instruments』

Instruments

§ Check the LCD display when operating Dr.Table devices

High handpiece
Low handpiece
Scaler

139/170
Ⅱ Description of each part 『Instruments』

Instruments

§ High Speed handpiece


High Speed Handpiece State LCD screen

Handpiece water injection Off, Handpiece optic Off

Handpiece water injection On, Handpiece optic Off

Handpiece water injection Off, Handpiece optic On

Handpiece water injection On, Handpiece optic On

140/170
Ⅱ Description of each part 『Instruments』

Instruments

§ Low Speed Handpiece


Low Speed Handpiece State LCD screen

Handpiece water injection On, Handpiece optic Off

Handpiece water injection Off, Handpiece optic On

Handpiece water injection On, Handpiece optic On

Handpiece water injection Off, Handpiece optic Off

141/170
Ⅱ Description of each part 『Instruments』

Instruments

§ Scaler
Scaler state LCD screen

Scaler water injection On

Scaler water injection Off

142/170
Ⅱ Description of each part 『Safety Switch』

Safety Switch

§ Check how to use the chair’s safety switch

• The safety switch installed in the channel is used to protect the patient and system
automatically in an emergency. When an obstacle is touched, the chair operation
stops and the chair moves up.

Center Piece
Lower Channel

『Position of safety switch』 『LCD screen when the safety switch activates』

143/170
Ⅱ Description of each part Adjust water volume for handpiece

Water Volume Adjustment for Devices

§ How to adjust the water volume for Dr.Table devices

• Water volume adjustment knob: Turning the knob clockwise increases the water
volume; turning it counter-clockwise decreases water volume

Position of water volume adjustment knob

144/170
Ⅱ Description of each part How to operate Dr.Table devices

How to Operate Devices on Dr. Table

§ How to operate high/low speed handpiece

• Take the handpiece out from the Dr. Table holder and press the foot control pedal
and the handpiece starts operating

High handpiece
Low handpiece

145/170
Ⅱ Description of each part How to operate Dr.Table devices

How to Operate Devices on Dr. Table


§ Hot to operate scaler
• Take the scaler out from the scaler holder and press the foot control pedal and the
scaler starts operating.
To stop the scaler, press the foot control once again or place the scaler on the
holder, and the power will automatically turn off.
• Turning the knob clockwise increases the scaler power; turning the knob counter-
clockwise decreases the scaler power.

Scaler power control knob

Scaler

146/170
Ⅱ Description of each part How to operate Dr.Table devices

How to Operate Devices on Dr. Table

§ How to operate 3-way syringe

• Air spray button: Press the button to spray air


• Water spray button: Press the button to spray water
• Spray button: Press the water and air discharge buttons at the same time and a
mixture of air and water will be sprayed
• 3-way syringe tip: Replace the tip depending on the treatment type

Air spray button

Water spray button

3-way syringe tip


3-way syringe

147/170
Ⅱ Description of each part Dental Light

Dental Light

§ How to operate dental light


• Press the Dr.Table button and the light turns on.
• You can adjust the brightness in 6 levels using the brightness knob at the bottom of the
light head.
• If the chair moves to a set auto position, the dental light turns on automatically.
• The angle of the head can be freely adjusted in an up/down/left/right/rotating direction

Light Balance Arm

Reflector
Blue Cut Off Filter function

Handle

ü Blue cut off function: Slide it to the right and the Blue Cut Off Filter function is activated
ü When it returns to the original position, Blue Cut Off is disabled

148/170
Ⅱ Description of each part Dental Light

Dental Light

§ How to change the dental light position

• Grab the light handle and turn left, right, up or down

149/170
Ⅱ Description of each part 『Automatic water supply device』

Automatic Water Supplier

§ How to operate the automatic water supply device

• Water volume adjustment: Turn the knob clockwise to increase the water volume
• Warmer: You can turn the warmer switch On/Off

Light Sensor
Warmer On/Off
Cup water volume
adjustment knob

Water outlet table water


volume adjustment knob
(Spittoon)
[User control section]
ü Auto cup: Water level is controlled based on an optical sensor, which stops supplying water
when the water level reaches a certain point
ü Function to supply warm water (25 to 41℃) available
ü When the cup is removed during water supply, water supply is suspended and the spittoon
cleaning operation begins automatically after 5 seconds.
150/170
Ⅱ Description of each part 『HEADREST』

Head-Rest

§ How to operate the headrest

• One-touch headrest is up-down adjustable to adjust to patient's body


• Press the button to set a desired angle and release the button to fix the position
• To adjust the height of the headrest, move the entire headrest up or down.

Headrest Seat

Headrest Slide Bar


Headrest Button

[Headrest adjustment function]

151/170
Ⅱ Description of each part 『ARMREST』

Arm-Rest

§ How to operate armrest

• The position of the armrest is manually adjustable to allow the patient to get in and out
of the chair
• Press the PUSH button and the armrest falls down due to its weight
• To return it to the original position, grab the armrest and pull it to the original position

Push Button

Armrest operational function is only


available on the right side
[Armrest adjustment function]

152/170
Ⅱ Description of each part 『FOOT CONTROLLER』

Foot Controller

§ How to operate the foot controller

Pedal

Manual chair controller

Return position Rinsing position

Handpiece RPM controlling pedal

[Foot Control Adjustment Function]

※ Emergency stop function (ESB: Emergency Stop Button)


Foot controller pedal features an emergency stop function, which stops the chair’s operation when
you press the pedal in the event of an emergency during automatic chair operation

153/170
Ⅱ Description of each part 『FOOT CONTROLLER』

Foot Controller

§ How to operate foot controller [continued]


Functions How to operate
Descend
The backrest goes down when you push the controller forward
backrest
Ascend
The backrest goes up when you pull the controller backward
backrest
Manual chair controller
Ascend chair The chair goes up when you push the controller to the left

Descend chair The chair goes down when you push the controller to the right

Rinsing position - Moves to the position that allows the patient to gargle at the spittoon

Return position - Returns to the position before gargling

Handpiece RPM
- Adjusts the handpiece RPM and activates the scaler
controlling pedal

Pedal - Used to relocate the foot controller using the foot

154/170
Ⅱ Description of each part 『MONITOR ARM』

Monitor Arm

§ How to operate monitor arm

• A : POLE ARM – part mounted on the LIGHT POLE


• B : MAIN ARM – supports and rotates the monitor
• C : MONITOR BRACKET – part on which the monitor is mounted
C

B
A A B C

D A, B and C section rotatable up to


300˚ each
E

Joint rotation range can be adjusted by


unfasten D bolt and fasten E nut (lock nut)

155/170
Ⅲ How to clean

156/170
Ⅲ How to clean 『UNIT Cleaning』

UNIT Cleaning

§ How to clean the unit

• Detach the spittoon and the filter, clean them at least once a day
[Remove the cup holder, lift the filter cap of the spittoon and take out the filter to
remove impurities and clean the filter]
• Take the suction filter out once every other day to remove impurities
[Remove the cap in the filter section, and remove any impurities]
Cup holder

Spittoon
Suction Filter
Spittoon Filter Cap
Suction Cap

Spittoon Filter Cap

157/170
Ⅲ How to clean 『Water Sanitation』

Water Sanitation

§ Before leaving the office (chemical cleaning)


• Attach a bottle containing washing solution
• Turn the bottle switch ON
• Press the water sanitation mode start button
(F + Inject Water)
• Mount the water sanitation cover on the spittoon
• Insert a hose (syringe, handpiece, etc.) into the water sanitation cover
• Press the Start Washing button (-) and then the washing solution
will be discharged from the pipe and the washing process
begins (takes about 7 minutes)
• After the washing process is over, a melody will play. F/V WET - Key
Then, turn the bottle switch Off and turn the main power Off
After (Bottle + Off), turn the main power off
• Leave the office

158/170
Ⅲ How to clean 『Water Sanitation』

Water Sanitation

§ After arriving at the office (water washing)

• Arrive at the office


• Turn the main power of the unit chair ON
• Press the water sanitation mode start button
(F + Inject Water)
• Set the bottle switch (City + Off)
• Press the Start Washing button (+) and water will be
discharged from the pipe and the washing process
will begin (takes about 10 minutes)
• After the washing process is over, a melody will play. F/V WET MODE + key
Then, return the tubing and water sanitation cover
to the original position
• Press the End Washing button (MODE)
• Cleaning of the water sanitation cover and the spittoon is over
• Start treatment

159/170
Ⅳ K3 Upgrade

160/170
Ⅳ K3 Upgrade 『SANITATION』

v K3 Distilled water tank & water sanitation system


ü Reason for Development

- Sales have increased, as the product satisfies the specifications of dental clinics
- Water tube cleaning system added to improve K3 product quality
- Distilled water tank mounted to supply distilled water to each treatment instrument

Bio film in the unit chair water line removed

161/170
Ⅳ K3 Upgrade 『SANITATION』

v K3 Distilled water tank & water sanitation system


ü OSSTEM Distilled water tank & water sanitation system (development completed)
- Water tube cleaning using a wash liquid (for domestic use: 2 to 3% hydrogen peroxide)
- A single holder that is compatible with multiple treatment instruments
- Fully automated production (hot key used in the Dr. table)

162/170
Ⅳ K3 Upgrade 『SANITATION』

v K3 Distilled water tank & water sanitation system


ü Product Application Plan

- Currently permission is pending (~ 2017.05)


- Pilot application at domestic university hospitals
(Yonsei University/Kyunghee University)

163/170
Ⅳ K3 Upgrade 『WATER SOLUTION』

v Dental Air & Water Solution (Compressor & Suction + Separator & Valve)

Chair Unit

Dental Office

Separator
& Valve

Compressor
Machine room & Suction Dental Air & Water

Compressor Suction Solution


164/170
Ⅳ K3 Upgrade 『WATER SOLUTION』

v Dental Air & Water Solution (Compressor & Suction + Separator & Valve)
ü Water Separator Product (CS1)
- Due to the need for infection control and environmental regulation, in European and
other overseas countries where dry suction is used, the dental system must separate
water from the air to ensure that only air is sucked.

CS1

Water Separator

165/170
Ⅳ K3 Upgrade 『WATER SOLUTION』

v Dental Air & Water Solution (Compressor & Suction + Separator & Valve)
• CS1 (Water Separator) Mounting

K3 Unit System Configuration K3 Unit Internal Mounting(Singapore)

Spittoon
Suction
Wastewater

Air mixture

Dry
Suction
Air

T-pipe

K3 Unit Drainage
Embedded

166/170
Ⅳ K3 Upgrade 『WATER SOLUTION』

v Dental Air & Water Solution (Compressor & Suction + Separator & Valve)

ü Spittoon V/V + Suction V/V Product


- Blocks the back flow of odor and germs from the stagnant ammonia pollutants in
the sewer and pipe

Connector Spittoon V/V Suction V/V

167/170
Ⅳ K3 Upgrade 『WATER SOLUTION』

v Dental Air & Water Solution (Compressor & Suction + Separator & Valve)
• Spittoon V/V + Suction V/V Mounting

K3 Unit System Configuration K3 Unit Internal Mounting(Australia)

Spittoon
Suction
Wastewater

Air mixture

Embedded unit
Drainage

168/170
Ⅳ K3 Upgrade 『WATER SOLUTION』

v Dental Air & Water Solution (Compressor & Suction + Separator & Valve)
ü Product Application Plan
- Currently permission is pending (~ 2017.05)
- Will go into production after field monitoring (2017.06 ~)

169/170
Thank you