Sie sind auf Seite 1von 139

GE Healthcare

System Service Manual


SYSTEM COMPAX 40E, CE Marked with MPH 50/65/80
(including XT, RS 85 and Tomo−Link

Direction 2154260−100, Revision 9


Copyright 1999 By General Electric Co.
All Rights Reserved
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


THIS PAGE INTENTIONALLY LEFT BLANK.

ii
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

− THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING − IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN
ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
− DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
− FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.

− CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


AVERTISSEMENT − SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
− NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
− LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

− DIESES KUNDENDIENST−HANDBUCH EXISTIERT NUR IN ENGLISCHER


SPRACHE.
WARNUNG
− FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
− VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST−HANDBUCH ZU RATE GEZOGEN UND VERSTANDEN WURDE.
− WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

− ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


AVISO − SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
− NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
− LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

iii
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

− ESTEMANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL


EM INGLÊS.
ATENÇÃO
− SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
− NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
− O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

− IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN


INGLESE.
AVVERTENZA
− SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
− SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
− NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

iv
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained
should be thoroughly read and understood by everyone who will use the equipment before you attempt to
place this equipment in operation. The General Electric Company, Medical Systems Group, will be glad to
assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no
practical design of equipment can provide complete protection. Nor can any practical design compel the operator to
take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents,
and representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective material and devices are available. It is urged that such materials or devices be used.
CAUTION: United States Federal law restricts this device to use by or on the order of a physician.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required. In performing all electrical work on these products,
GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the
requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 827−3468 / 8*285−3468 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
6/17/94

v
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


THIS PAGE INTENTIONALLY LEFT BLANK.

vi
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

REVISION HISTORY

REV DATE REASON FOR CHANGE

0 May. 10, 1996 Initial release.


1 Oct. 16, 1996 To add job Card IST 002
2 Feb. 26, 1997 To add RS 85 Tube Stand and Tomo−Link configurations
3 August 7, 1997 MIS Map, MIS Chart Update
4 December 7, 1997 Updated (BUCge 14603, 29344, 29457, 31763)
5 October, 1998 Updated with 115/24OV setting when used with MPH (Chapter 2)
6 November, 1998 New reference number for OM Compax 40E, SG60 and SG100
Medys Bucky instead Liebel Bucky
7 June 1999 Illustrations 7−2, 7−5 and 7−8 updated (ground plug layout)
8 July 2003 Added note about installing local language rating plate over English version.
9 November 2006 To resolve PSR 13077161: added a “Warning” in Card IST 001 to use Loctite
Threadlocker, strength 242 or 243 when installing collimators to help prevent
the screws from backing out over time.
To add “Collimator PM” and “Collimator Installation” procedures to Chapter 6.

vii
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


THIS PAGE INTENTIONALLY LEFT BLANK.

viii
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

CHAPTER 2 − INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
SECTION 1
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1-1 FOREWORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1-2 UNPACKING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1-2-1 Receipt of Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1-2-2 Inventory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1-2-3 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1-3 INSTALLATION STEERING GUIDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
1-4 CABLE CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
1-4-1 Preliminaries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
1-4-2 Mis Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
1-4-3 Mis Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
1-4-4 Mis Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

ix
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

CHAPTER 3 − DISASSEMBLY/REASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Job Card IST 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
X−Ray Tube & Collimator
Installation ON XT SUSPENSION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84

CHAPTER 4 − FUNCTIONAL CHECKS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85


SECTION 1
PRELIMINARIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
SECTION 2
OPERATIONAL TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
2-1 TABLE OPERATIONAL TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
2-2 TECHNIQUE TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
2-3 SG 60 OPERATIONAL TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
2-4 SG 100 OPERATIONAL TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
2-5 XT SUSPENSION OPERATIONAL TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
2-6 RS 85 TUBESTAND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
2-7 ULTRANET SM COLLIMATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86

x
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


2-8 ULTRANET SA COLLIMATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
SECTION 3
PERFORMANCE TESTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87

CHAPTER 5 − SYSTEM CABINET SPARE PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

CHAPTER 6 − PREVENTIVE MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95


SECTION 1
PERIODICITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
1-1 TABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
1-2 MPH GENERATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
1-3 XT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
1-4 RS 85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
1-5 SG60 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
1-6 SG100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
1-7 TOMO−LINK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
1-8 COLLIMATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
SECTION 2
OPERATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-1 TABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-2 MPH GENERATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-3 XT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-4 RS 85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-5 SG60 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-6 SG100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-7 TOMO−LINK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-8 COLLIMATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-8-1 Equipment Required for Collimator PM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2-8-2 Removing the Covers from the Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
2-8-3 Lubrication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
2-8-4 Collimator General Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
2-8-5 Check Tightness of Collimator Mounting Ring (every 12 months) . . . . . . . . . . . . . . . . . . . . 97
2-8-6 If the collimator has no play: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
2-8-7 Checking Electrical Cables and Connections, and Cable Connection Points . . . . . . . . . . . 98
2-8-8 Reassembling the Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
SECTION 3
COLLIMATOR INSTALLATION VERIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
SECTION 4
ULTRANET COLLIMATOR INSTALLATION CHECK LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102

CHAPTER 7 − SCHEMATICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103


SECTION 1
. ........................................................................................... 103
1-1 SCHEMATICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
1-2 CABLES IDENTIFICATION ACCORDING TO THE MIS MAP AND MIS CHART . . . . . . . . . . . . . 114

xi
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

CHAPTER 1 − PRELIMINARIES

SECTION 1
INTRODUCTION

1-1 Safety

ALL SERVICE OR SPARES REPLACEMENT ON THIS SYSTEM MUST BE


CARRIED OUT BY QUALIFIED TECHNICIANS ONLY. FOLLOW THE
DIRECTIONS IN THIS MANUAL CAREFULLY.

PREVENTION OF IONIZING RADIATION


− It is mandatory to wear a film badge.
− Remain as little time as possible in or near of an X−ray field.
− When working near X−rays, it is mandatory to wear a lead (Pb) apron.
− The kV and mA setting values must be as low as possible.
− The X−ray field setting must be as low as possible (a beam limiter must be used).
− The operator must stand as far away as possible from the X−ray beam.

PREVENTION OF ELECTRICAL HAZARDS


− Do not open electrical cabinets if power supply is on.
− Do not disconnect a ground wire.
− Before disconnecting HV cables, discharge them with a voltage divider.
− Be careful of capacitors which may be charged (specially those HV units).

PREVENTION OF ACCIDENTS CAUSED BY MOVING PARTS


When adjusting mechanical sub−assemblies, be careful of:
− movements starting up.
− transmission parts (chain, gears, ....).

EQUIPMENT PROTECTION
To avoid accidental damage to the equipment, it is mandatory to cut the electric power supply before:
− connecting or disconnecting a printed board or a connector.
− connecting or disconnecting a test probe.

1
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

1-2 Available Documentation

System COMPAX 40E with MPH GENERATOR, SG 60 or SG100, XT Suspension, RS 85 Tube Stand or
Tomo−Link
Type DESCRIPTION Ref. n r. or Direction Language Furnished Available

SYSTEM
ssm System COMPAX 40E 2154260−100 Eng X

TABLE
OM OPERATING MANUAL Direction 2226656−148 Eng/Fre/Spa X
COMPAX 40E CE TABLE Ger/Ita/Por

SM SERVICE MANUAL Direction 2150299−100 Eng X


COMPAX 40E CE TABLE

SCH COMPAX 40E TABLE SCHEMATICS Direction 2150299−100 Eng X

pim Pre−Installation Manual Direction 2150306−100 Eng X


COMPAX 40E CE TABLE

GENERATOR
SM MPH 50/65/80 GENERATOR Direction 2137354−100 Eng X
SERVICE MANUAL

SCH MPH 50/65/80 GENERATOR Direction 2137357−100 Eng X


SCHEMATICS

ASM ADVANCED SERVICE MANUAL Direction 2137631−100 Eng X


MPH 50/65/80 GENERATOR

pim Pre−Installation Manual Direction 2133000−100 Eng X


MPH 50/65/80 GENERATOR

OM OPERATING MANUAL Direction 2137352−100 Eng X


MPH 50/65/80 GENERATOR Direction 2137352−101 Fre
Direction 2137352−106 Spa
Direction 2137352−108 Ger
Direction 2137352−111 Ita
Direction 2137352−127 Por

ION CHAMBER
SM Ion Chamber ( UL − 3 Fields ) 826 921 P285 Eng X

PROGRAM−X Option
OM OPERATING MANUAL Direction 2103514−100 Eng X
PROGRAM−X Direction 2103514−101 Fre
Direction 2103514−106 Spa
Direction 2103514−108 Ger
Direction 2103514−111 Ita
Direction 2103514−127 Por

2
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


Type DESCRIPTION Ref. n r. or Direction Language Furnished Available

PRINT−X Option
OM OPERATING MANUAL Direction 2107167−100 Eng X
PRINT−X Direction 2107167−101 Fre
Direction 2107167−106 Spa
Direction 2107167−108 Ger
Direction 2107167−111 Ita
Direction 2107167−127 Por

RAFL−S
SM ULTRANET SA & RAFL−S (Collimator Con- DIR. Y00W20L.04 Eng X
trol)
SERVICE MANUAL

BUCKY
SI BUCKY SERVICE INSTRUCTIONS DIR. 248950 Eng X

COLLIMATOR
OM ULTRANET SM CE marking Direction 2144901−100 Eng X
Operating Manual Direction 2144901−101 Fre
Direction 2144901−106 Spa
Direction 2144901−108 Ger
Direction 2144901−111 Ita
Direction 2144901−127 Por

SM ULTRANET SM DIR Y00W21L.04 Eng X

OM ULTRANET SA CE marking Direction 2144903−100 Eng X


Operating Manual Direction 2144903−101 Fre
Direction 2144903−106 Spa
Direction 2144903−108 Ger
Direction 2144903−111 Ita
Direction 2144903−127 Por

SM ULTRANET SA DIR Y00W23L.04 Eng X

WALL BUCKY
OM SG 60 VERTICAL BUCKY STAND CE Direction 2224051−148 Eng/Fre/Spa X
Operating Manual Ger/Ita/Por

pim SG 60 VERTICAL BUCKY STAND CE Direction 2150301−100 Eng X


Pre−Installation Manual

SM SG 60 VERTICAL BUCKY STAND CE Direction 2150293−100 Eng X


Service Manual

OM SG 100 VERTICAL BUCKY STAND CE Direction 2226754−148 Eng/Fre/Spa X


Operating Manual Ger/Ita/Por

3
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


Type DESCRIPTION Ref. n r. or Direction Language Furnished Available

pim SG 100 VERTICAL BUCKY STAND CE Direction 2150305−100 Eng X


Pre−Installation Manual

SM SG 100 VERTICAL BUCKY STAND CE Direction 2150297−100 Eng X


Service Manual

X−RAY TUBE
TP Maxiray 100 X−RAY TUBE Direction 46−014427 Eng X
Technical Publications

TP Maxiray 100 X−RAY TUBE Direction 46−013804 Eng X


Technical Publications

XT SUSPENSION
OM X−RAY TUBE SUSPENSION DIR 2116790−100 Eng X
OPERATING MANUAL

SM X−RAY TUBE SUSPENSION DIR 2118733−100 Eng X


SERVICE MANUAL

pim X−RAY TUBE SUSPENSION DIR 46019618 Eng X


Pre−Installation MANUAL

RS 85 TUBE STAND
OM RS 85 CE Marked Ongoing Eng X
Operating Manual

IS Mounting Instructions RS 85 Tube Stand CE Direction 0327 7223 Eng X

SM RS 85 Interface Package for Compax 40E CE Direction 2165897−100 Eng X


Marked Systems
Service Manual

PIM RS 85 Tube Stand Direction 2172805−100 Eng X


Pre−Installation Manual

OI Operating Instructions RS 85 Tube Stand CE Direction 0327 7324 Eng X


Direction 0327 7341 Fre
Direction 0327 7361 Spa
Direction 0327 7303 Ger
Direction 0327 7351 Ita
Direction 0327 7371 Por

TOMO−LINK
OM Tomo−Link Direction 2122884−100 Eng X
Operating Manual Direction 2122884−101 Fre X
Direction 2122884−106 Spa X
Direction 2122884−108 Ger X
Direction 2122884−111 Ita X
Direction 2122884−127 Por X

SM Tomo−Link Direction 2133821−100 Eng X


Service Manual

4
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Type DESCRIPTION Ref. n r. or Direction Language Furnished Available

SM Tomo−Link Schematics Direction 2132473−100 Eng X

SM Tomo−Link Upgrade Assembly for CE Marked Direction 2182386−100 Eng X


Compax 40E Tables
Installation Guide

5
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


1-3 Manual System Composition

SYSTEM COMPAX 40E (Manual) with MPH 50/65/80

− COMPAX 40E
CLASSICAL RAD TABLE with Anticollision Kit
Core Table Bucky with Grid & Ion Chamber
− MPH 50/65/80 with AEC
− High Speed Starter
− High Impedance Adaptation

− Power Supply: 380Vac....480Vac (+/− 10%)


RST+GND
50/60 Hz

XT SUSPENSION Inboard or Outboard with:


− HV cables + Stator and Collimator cables RS 85 TUBE STAND: with:
To change mode − Manual Ultranet SM Collimator − Manual Collimator Ultranet SM
− Maxiray 100 Tube − Maxiray 100 Tube
Optional
Features

Tomo−Link
− Adapting kit for 40E

SG 60/100 Vertical Bucky Stand


Right or Left hand
Grid & Ion Chamber

END

6
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

1-4 Automatic System Composition

SYSTEM COMPAX 40E (Auto) with MPH 50/65/80

− COMPAX 40E
CLASSICAL RAD TABLE with Anticollision Kit
Core Table Bucky with Grid & Ion Chamber
− MPH 50/65/80 with AEC
− High Speed Starter
− High Impedance Adaptation

− Power Supply: 380Vac....480Vac (+/− 10%)


RST+GND
50/60 Hz

XT SUSPENSION Inboard or Outboard with: RS 85 TUBE STAND: with:


− HV cables + Stator and Collimator cables − Auto Collimator Ultranet SA
To change mode − Auto Ultranet SA Collimator − Maxiray 100 Tube
Optional − Maxiray 100 Tube − RAFL−S Collimator Control
Features − RAFL−S Collimator Control

Tomo−Link
− Adapting kit for 40E

SG 60/100 Vertical Bucky Stand


Right or Left hand
Grid & Ion Chamber

END

7
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


THIS PAGE INTENTIONALLY LEFT BLANK.

8
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

CHAPTER 2 − INSTALLATION

SECTION 1
OVERVIEW

1-1 Foreword

teHLA (High Level Assembly) Program: There are two different versions of equipment shipped to the
customer, a non HLA or an HLA version.
COMPAX 40E SYSTEMS are fully assembled and calibrated at the factory (*) before shipment. Only
Tables purchased as add−ons to existing tables are non−HLA.
The aim of this factory integration, or High Level Assembly Program, is to improve the general quality of
the installation procedure on site while decreasing the installation time.
Program results:
− Increased customer satisfaction by reducing the equipment installation time and an increased
reliability and performance from its first use.
− Facilitate the technician’s work:
> by relieving him of time−consuming, tricky and repetitive jobs.
> by involving technicians in this new global process which consists of reducing installation time
and the global quality improvement of our equipment.
This reduces work load and warranty cost.
The Installation Steering Guide which follows is conceived to take maximum advantage of this High
Level Assembly program.
To avoid any duplication of information given in other documents, this manual makes a number of cross
references to Service or Installation Manuals specific to the various components that make up the system.
Any remarks improving the contents of this document would be appreciated.

(*) COMPAX 40E Systems for the US are factory checked with a XT “tool” since the suspension
is purchased locally.

9
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

1-2 Unpacking

1-2-1 Receipt of Equipment

HANDLING PACKED SUB−ASSEMBLIES


Prior to installation, store the equipment in its original packing, and handle it with care.
Do not drop, throw or stamp packages containing equipment or parts of it.

1-2-2 Inventory

Depending on the equipment ordered, check the number of packages delivered to the installation site.
The lists of packages hereafter do not include the optional components.

Crate Contents Dimensions Gross Weight


kg lb
N° 1 Compax 40E CE Table base including: 230x130x31 350 770
− Medys Bucky
− BVM Ion Chamber
− Telescopic Covers
− Compax 40E Documentation
− System Inspection Report
− System Cables
− Packing List

N° 2 Compax 40E CE Table Top 150x94x103 75 165


− Sensing Size Tray

10
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Gross Weight
Crate Contents Dimensions
kg lb
N° 3 MPH Cabinet including: 215x100x47 280 616
− I/F Room
− RAD Option
− HV Tank
− RAFL−S Collimator Control (optional)
− System cables
− C40E/MPH/XT/RS 85/SG SSM
(System Service Manual)
− Inspection Report
N° 4 MPH Console including: 72x120x62 50 110
− Program−X
− Hand Switch
− MPH Generator Documentation
− Packing List

N° 5 X−Ray Tube (Maxiray 100) 50x84x60 46 100


− X−ray Tube Unit
− Documentation

N° 6 Ultranet SA Collimator or Ultranet SM Collimator 47x39x56 30 66


− Collimator Head
− Documentation
− Tools

N° 7 SG60 or SG100 Vertical Bucky Stand including: 230x87x86 288 633


− Medys Bucky
− BVM Ion Chamber
− Size Sensing Tray
− SG60 or SG100 Documentation
− Packing List
− System cables
N° 8 Tomo−Link 109x82x88 98 216
GEMS−AM
delivered NOTE: In GEMS−AM delivered systems, the XT suspension
systems is shipped separately
only

11
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Gross Weight
Crate Contents Dimensions
kg lb
N° 8 XT Suspension including: 106x90x152 276 607
GEMS−E − Covers and Accessories
and − Documentation
GEMS−A
delivered NOTE: In GEMS−AM delivered systems, the XT suspension
systems is shipped separately
only

N° 9 XT Suspension Bridge 318x82x48 140 308


GEMS−E
and NOTE: In GEMS−AM delivered systems, the XT suspension
GEMS−A is shipped separately
delivered
systems
only

N° 10 Inboard Concealment 60’ Cables for MPH Generator 214x30x61 109 240
GEMS−E including: TBC
and − HV Cables
GEMS−A − Stator Cable
delivered − Suspension Cables
systems
only NOTE: In GEMS−AM delivered systems, the XT suspension
is shipped separately
N° 11 Stationary Rails 476x18x12 48 100
GEMS−E
and NOTE: In GEMS−AM delivered systems, the XT suspension
GEMS−A is shipped separately
delivered
systems
only

N° 12 Tomo−Link 109x82x88 98 216


GEMS−E
and NOTE: In GEMS−AM delivered systems, the XT suspension
GEMS−A is shipped separately
delivered
systems
only

12
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Crates N° 1 thru N° 7 are the same for both XT suspension and RS 85 Tube Stand Systems
Gross Weight
Crate Contents Dimensions
kg lb
N° 8 RS 85 Tube Stand including: 320x66x56 190 396
− Documentation
RS 85 Interface Package for Compax 40E System in-
cluding:
− Set of System cables
− Documentation

N° 9 Accessories 47x39x56 30 63

Nº 10 Print−X (Option) 380 x 820 x 350 20 51


− Pedestal box
− Documentation

13
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

1-2-3 Unpacking

Damage Due to Transport


On delivery of the equipment, all packages must be checked to make sure that they have not been
damaged during transport.
If damage is observed, notify the transport service and the factory transport insurance service
immediately.
The above also applies for non apparent damage due to transport discovered when unpacking the
material or during installation, WITHIN 14 DAYS MAXIMUM AFTER DELIVERY.
Generally, a transport service will not pay a bill for non−apparent damage unless an expert’s report has
been requested within 14 days of delivery.
If damage is discovered, contact the concerned services immediately, specifying the type of apparatus,
the serial number and the order number if possible, as well as a description of the damage.

Unpacking and Inspection


Tools required for unpacking:
− a tool for removing nails, i.e. claw−hammer.
− a crowbar (minimum length 600 mm).
− a dolly for lifting and transporting the table base and SFD−fork, etc...
The components are wrapped in protective material and immobilized in the transport boxes. Each box is
numbered.
When unpacking, first find the packing list and check that all items are present. Check also that there is
no damage to the components received.
In case of missing parts, file a claim immediately mentioning the serial number of the unit as well as all
the shipping data.
Do not discard packing materials, such as envelopes or boxes, until all parts are accounted for.
− Do not place equipment cabinets in areas that are damp, poorly ventilated, too far from the console,
or on another floor.
− Avoid poor floor anchoring.
− Ensure that the screws and washers of each cover are inside plastic bags stuck on the inside of the
cover.

14
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


1-3 Installation Steering Guide

Note: Read the following paragraphs carefully before continuing with this Steering Guide.

• Pre−installation: Pre−installation work is required to prepare the room for installing the
equipment. Check that this work is complete, if not, complete
pre−installation work now.
• Inspection Report: Some of the Steering Guide steps are factory performed. Refer to Inspection
Report Dir. 2167647 TDR, 2172166 TDR or 2179651 TDR(supplied with the
system) to know exactly which steps have already been performed.
• Steering Guide: Some steps of the Steering Guide refer to other Service Manual chapters.
When each task is performed, return to the Steering Guide and proceed with
the next step.
• Clamping: Do not forget to clamp the system cables for EMC compliance. Refer
to Job Card IST 005.
• Pinning:  To facilitate critical realignment on−site, several sub−assemblies are
pre−drilled and pinned or marked during factory integration. Refer to Job
Cards IST001, IST002, IST003 and IST 004 for installation procedures
 The Undertable configuration comes with several system cables
pre−installed in the Table Rolling frame. The remaining system cables are
pre−installed in the Tubestand assy. Take care not to damage these cables.
Data Base Backup: Performed by Manufacturing on Xamine Calibration Diskette.

ÍÍ
ÍÍ
Mark each box when task is done

Step Description Done Supporting Documents

START

See PRE−INSTALLATION MANUALS


and LOCAL LAYOUT DATA Compax 40E CE, PIM
Dir. 2150306−100
CHECK
Pre−installation Requirements
ÍÍ Vertical Bucky Stand CE, PIM Dir 2150305−100 (SG100),
A1 are met

1 person 1 Hour
ÍÍ Dir 2150301−100 (SG60)
MPH50/65/80, PIM Dir 2133000−100
XT CE, PIM Dir 46−019618
RS 85 CE, PIM Dir 2172805−100

UNPACKING
Collect the Product Locator cards See section 1-2-3 of this manual.

ÍÍ
A2 in sequential order
Refer to Installation Flowchart for typical

ÍÍ
Localize Installation Flowchart main task durations

2 people 4 Hours

Continues on next page

15
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
ÍÍ
Mark each box when done

Step Description Done Supporting Documents

From previous page

No Pilot
System

Yes/No

Yes

MANDATORY: Fill out this Pilot


Feedback Form in English

ÍÍ
FEEDBACK FORMS throughout the System installation
process.

A3 PGS 017 (Appendix 2)


ÍÍ This General Service Procedure PG017 is
available for GEMSE at local GE offices
1 Person 1 Hour

 PHYSICAL INSTALLATION OF SYSTEM COMPONENTS

 XT INSTALLATION

Without XT XT

with/without

With XT

INSTALLATION If present, the ceiling suspension must be

ÍÍ
A4 installed first: carefully follow instructions
XT Ceiling Suspension given in XT SM, Dir 2118733−100;

ÍÍ
Chapter 3 or 4 (Do not install covers now)
Horizontal SID Sensors
2 people

Continues on next page

16
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Í
Step Description Done
Í Mark each box when done

Supporting Documents

From previous page


Refer to Inspection Report 2179651 TDR
or 2167647 TDR, sections 2.1 and 3.1

INSTALLATION
CAUTION: X−Ray Tube and Collimator
X−Ray Tube are factory aligned and assembled during
A5 HLA.
Collimator Head (SA/SM) The mounting ring should be not be

2 people A4 − A5=16 Hours


ÍÍ
ÍÍ
adjusted during installation .

See Job Card IST001 of this manual for


X−Ray Tube and Collimator installation
and levelling.

Without
Tomo−Link

with/without
With

INSTALLATION
Drive Unit
A6 Refer to Inspection Report 2179651 TDR,
Yoke section 5.1

Fulcrum Arm Support (Wall)

ÍÍ Refer to Tomo−Link SM 2133821−100

ÍÍ
Operator Control Console sections 1,2,5

2 People 8 Hrs

Continues on next page

17
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
Step Description Done
ÍÍ Supporting Documents

From previous page

 RS 85 INSTALLATION

without RS 85
with/without If present, RS 85 Tube Stand must be
installed first. Follow instructions given in
RS 85 PIM, Dir 2172805−100, Chapter 2.4

with

INSTALLATION Refer to Inspection Report 2172166 TDR,


sections 1.1 and 1.2
Column Cables Wall Box
A7 See SM− RS 85 Dir. 0327 7223
Mounting Instructions Chapter 1 Section
Tube Stand Rail
1.6.1 (Overall dimensions) and:
Column With Carriage

ÍÍ
Chapter 2, Section 2.3 (Adjust floor rails)
Chapter 2, Section 2.4

ÍÍ
Tube Support Arm
Counterweights Chapter 2 sections 2.5 thru2.6

2 People

INSTALLATION Refer to Inspection Report 2172166 TDR,


sections 3.1 and 2.1

X−ray Tube with Adapter See SM− RS 85 Dir. 0327 7223


Installation & Service Instructions
A8 Collimator Assembly CAUTION: X−Ray Tube and Collimator are
factory aligned and assembled during HLA.
HV Cable Routing
ÍÍ The mounting ring should be not be

ÍÍ
H adjusted during installation .

XXHrs See Job Card IST003 of this manual


2 People
HANDLE with CARE (by the support only).

Continues on next page

18
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Í
Step Description Done
Í Mark each box when done

Supporting Documents

From previous page

INSTALLATION See RS 85 Mounting Instructions, SM Dir.

ÍÍ
0327 7223. Chapter 1, Section 1.7, Chapter
A9 2 section 2.9 and Chapter 3 sections 3.1

ÍÍ
Counterbalancing & Leveling
thru 3.3 and 3.5.

1 Person

REMARK: Most of these cables are already


factory connected.

CONNECTIONS See RS 85 Interface Kit for Compax 40E CE;


SM Dir.2165897−100, Chapter 1 Installation,
 Collimator cable Chapter 2 MIS MAP and Chapter 3 MIS
CHART
A 10  Lock Cable See Job Card IST 005 for cable clamping

Column Cable concealment


1 Person A7 − A10= 8 Hours
Í See MIS MAPS, Chapter 5 of this manual, and
MIS CHART in section 1-4-3 this chapter.
MIS No. 27304
See Job Card IST003 for Cable concealment

COMPAX 40E TABLE INSTALLATION

Refer to Inspection Report 2179651 TDR


INSTALLATION or 2167647 TDR sections 4.1 and 4.3
or

Í
Inspection Report 2172166 TDR
Table Base Installation and leveling sections 4.1 and 4.8
A 11
Tabletop Installation

2 People
Í See SM COMPAX 40E CE, Dir. 2150299−100,
Chapter 1, Section 3.

ALIGNMENT
Table Base Longitudinal Axis Alignment with:
NOTE: Precise alignment is extremely
XT Longitudinal Rails important for Tomographic Performance.
A12

ÍÍ
See SM COMPAX 40E CE, Dir. 2150299−100,
(or) RS 85 Longitudinal Rails
Chapter 1, Section 3.2
2 People A11 − A12 = 4 Hours

Continues on next page

19
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
ÍÍ
Mark each box when done

Step Description Done Supporting Documents

From previous page

WALL BUCKY INSTALLATION

without WALL
BUCKY See SM Vertical Bucky Stand SG 100 CE −
with/without Dir 2150297−100, Chapter 1 or SM Vertical
Bucky Stand SG60, Dir 2150293−100
Chapter 1.
with

without REMARK: Use of the Wall Mounting Kit is optional


A17 but strongly recommended with certain kinds of
WALL Hospital wall construction (Unistrut wall). Refer to
MOUNTING SG60 PIM 2150301−100 or SG100 PIM 2150305−100
KIT .

with

INSTALLATION

ÍÍ
Mounting Hardware Kit See SG60 SM 2150293−100, Job Card IST001
A16

ÍÍ
or SG100 SM 2150297−100, Job Card IST001

1 Person 2 Hours

Refer to Inspection Report 2179651 TDR


INSTALLATION or 2172166 TDR sections 6.1, 6.2 and 6.3
or
SG60/SG100 Installation Inspection Report 2167647 TDR,
A17 sections 5.1, 5.2 and 5.3
Vertical Leveling
REMARK: Most of these cables are already factory
Route System Cables from SG to MPH connected. See MIS MAP / MIS CHART of this
Cabinet location Manual (section 1−4)
.
 Connections to be done during Installation:

ÍÍ
Lateral alignment across the wall to Table
Longitudinal Axis SG60: MIS 5.1
SG100: MIS 5.1−9
2 People 2 Hours

Continues on next page

20
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
ÍÍ
Mark each box when done

Step Description Done Supporting Documents

From previous page

MPH SYSTEM CABINET INSTALLATION


See Pre−installation Manual (Layout).
See SM MPH 50/65/80 dir. 2137354−100,
Chapter 1 Installation ; section 2,
Job Card IST002. Use Securing Kit (Part
INSTALLATION 2175309 provided with Upper Cabinet

Cabinet Positioning

Í
REMARK:Verify PCB have not moved
during transportation, particularly the

Í
A18 Cabinet Securing Power PCB in RAFL−S rack.
If the Wall Attachment Kit* is available, follow
2 People 2 Hours instructions supplied with the Kit *(Part #
2164271, S16201EB)

Refer to Inspection Report 2179651 TDR or


2167647 TDR, section 1.2 and 3.3
CONFIGURATION or
Inspection Report 2172166 TDR, section 1.2
A19 Generator Power Connection and 3.4

See SM MPH Dir 2137354−100, Chapter 1,


Line Voltage and Frequency: Installation: section 2, Job Card IST002
Dependent Settings
WARNING MAINLY FOR USE IN THE USA:
AC Voltage Supply for

XT*, RS85 Power Supply*


Buckies*
ÍÍ 1. Refer to Inspection Report and/or to orange
labels fixed to the buckies

2. With MPH the XT is ALWAYS powered with


1 Person 230VAC.
1 Hour
3. XT is shipped separately for the US.
Default setting is 110V.
Refer to XT Service Manual Dir
2118733−100 for 220V setting.

Refer to Inspection Report 2179651 TDR or


INSTALLATION 2172166TDR or 2167647 TDR, section 1.1

Console Generator
A20 PRINT−X Console (option) See Pre−Installation manual (layout).

ÍÍ See SM MPH 50/65/80 dir. 2137354−100, Chapter 1

ÍÍ
PROGRAM−X (option)
Installation; section 2, Job Card IST002.
PRINT−X
1 Person 1 Hour
CAUTION: Ensure proper tightening of wire
connections and setting of connectors

Continues on next page

21
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
ÍÍ Mark each box when done

Step Description Done Supporting Documents

From previous page

 SYSTEM CONNECTIONS

Refer to Inspection Report 2179651 TDR or


POWER AND GROUND 2172166 TDR or 2167647 TDR, section 1.2

CABLE CONNECTIONS
REMARK: Most of these cables are already factory
connected. See MIS MAP / MIS CHART of this
POWER WIRING CONNECTIONS Manual (section 1−4)
.
 Connections to be done during Installation:
 Route Power and Ground cables
A21 MIS 3, 4, 6, 9, 11

Í
MIS 1−1, 21, 3−1, 4−1, 5−1, 6−1, 7.1
 Connect cables onto MPH

2 People
Í See SM MPH: Dir 2137354−100
Chapter 1, Job Card IST002

SYSTEM CABLES CONNECTION

REMARK: Most of these cables are already factory


connected. See MIS MAP / MIS CHART of this
SYSTEM WIRING CONNECTION Manual (section 1−4)
.
 Connections to be done during Installation:
MIS 1, 2, 6, 7, 8, 10, 12 to 23
 Route remaining System cables
A22
ÍÍ NOTE: For connection of DOOR OPEN switch

2 People A21 − A22 = 4 Hours


ÍÍ and EXPOSURE ON signal lamp refer to Jov
Card IST004 of this manual
See Job Card IST 005 for cable clamping

Continues on next page

22
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
Step Description Done
ÍÍ Mark each box when done

Supporting Documents

From previous page

WARNING
MAINLY FOR USE IN THE USA:

CONFIGURATION CHECK*
• For Bucky: refer to Inspection Report 2179651
Bucky Voltage Settings: TDR, section, 4.2 for 40E, section 6.3 for SG
or to Inspection Report 2167647 TDR or
A23  110 or 220AC for SG60−100 Bucky 2172166 TDR, section 4.2 for 40E; section
5.3 for SG
 110 or 220AC for Compax 40E Bucky Refer also to orange label fixed to the Bucky

220V for XT Suspension, RS85, C40E,


ÍÍ See IST 004
TLK, SG
ÍÍ • For components: refer to Inspection Report
2179651 TDR or 2167647 TDR or 2172166
1 Person
TDR

VERY IMPORTANT: Read Inspection Report (IR) ( Dir. 21722166TDR for Compax 40E/MPH/SG/RS 85; Dir.
2167647TDR for Compax 40E/MPH SG/XT; Dir. 2179651TDR for Compax 40E/MPH/SG/Tomo−Link) through
carefully before eventually proceeding with the MPH software configuration, Step A24 of this Steering Guide.

POWER ON AUTHORIZED

 CALIBRATION OF SYSTEM COMPONENTS

IR
meets Reconfiguration is complete.
expectations? If more reconfiguration is necessary, proceed to step A24
Yes

No

Continues on next page

23
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
ÍÍ
Mark each box when done

Step Description Done Supporting Documents

From previous page

MPH CONFIGURATION CHECKS

Refer to Inspection Report 2179651 TDR or to


Inspection Report 2167647 TDR, or 2172166 TDR
CONFIGURATION section 1.3

ÍÍ
A24 See MPH SM 2137354−100, Job Cards
Re−do MPH necessary Configuration LG001, RG001 and RG002

1 Person ÍÍ NOTE: Connect Service Terminal as described


in Job Card LG 001

Refer to Inspection Report 2167647 TDR, or 2172166 TDR


section 4.3 for C40E; section 5.3 for SG; section 6.2 for
RAFL−S
CONFIGURATION or
Refer to Inspection Report 2179651 TDR, section
4.3 for 40E; 6.3 for SG; 7.2 for RAFL−S

D Re−do necessary Sub−System

ÍÍ
Components Configuration
A25
See C40E SM 2150259−100 (Chapter 2)

1 Person
ÍÍ SG60 SM 2150258−100 (Chapter 2), SG100 SM
2150257−100 (Chapter 2), XT SM 46−019803 or
2118910−100 (section 12), RS 85 SM Mounting
Instructions 0327 7223 (section 3)

Refer to Inspection Report 2179651 TDR or Inspection


Report 2167647 TDR, or 2172166 TDR section 2.2
RUN−IN & STABILITY CHECK
A26

ÍÍ
Rotating anode Tube REMARK: The Run−in & Stability check is

ÍÍ
performed at the factory. See Technical
Publications, Dir. 46−013804
1 Person 1 Hour
To redo on site ONLY if necessary (refer to
System Inspection Report Completion date).
Use procedure for MVP generators.

Continues on next page

24
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
ÍÍ
Mark each box when done

Step Description Done Supporting Documents

From previous page

REMARK: The MPH calibration is


MPH CALIBRATION performed at the factory but it is necessary
to re−check it.

Refer to Inspection Report 2167647 TDR or


2179651 TDR or 2172166 TDR.section 2.3
Yes and proceed with required calibration
checks marked in grey:
kV, mA, mAs
Calibration Checks
A27

ÍÍ
OK

1 Person
No
1 Hour
ÍÍ Make several exposures at different kV, mA
and mAs stations. Read Post Display on
MPH Console and if out of range, proceed
to step A28 otherwise, skip to step A29

NOTE: kV must be measured using a


Keithley 3508 kV pic meter.

CALIBRATION Refer to Inspection Report 2167647 TDR, or


2172166 TDR, section 2.2−2.3

A28

ÍÍ
Run necessary MPH Calibrations

1 Person XX Hrs ÍÍ Connect Service Terminal. Refer to the MPH


SM Dir. 2137354−100 Chapter 1 section 3
and proceed with calibrations in that section.

Continues on next page

25
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
Step Description
ÍÍ
Done
Mark each box when done

Supporting Documents

From previous page

COMPAX 40E, XT, RS−85 CALIBRATION


(Factory Calibrated)

Refer to Inspection Report 2179651 TDR or


CALIBRATION 2167647 TDR, section 4.9 for 40E
and XT
 Perform necessary calibrations on:
Refer to Inspection Report 2172166 TDR,
section 4.10 for RS 85 and 40E
A29 Compax 40E Table
See SM COMPAX 40E− Dir. 2150299−100,
XT Suspension Chapter 3 − Calibration,

See SM. XT Radiographic Suspension Dir

ÍÍ
RS 85 2118733−100, Chapter 5 Service
Adjustments.
1 Person
ÍÍ See RS 85 SM Mounting Instructions 0327
7223, section 3

X−RAY BEAM ALIGNMENT/SID ADJUSTMENT

PERPENDICULAR ADJUSTMENT
Perpendicularity of X−Ray beam with
See C40E SM; Dir. 2150299−100
A30 Table Chapter 3, section 3 (use RMI 161 B

ÍÍ
tool)
Bucky Stand.

ÍÍ
1 Person 1/2 Hour

Continues on next page

26
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
Step Description Done
ÍÍSupporting Documents

From previous page

CENTERING ADJUSTMENT
X−ray beam and Light beam alignment check See Ultranet SA + RAFL−S sm
A31 Dir. Y00W20L.04 or Ultranet SM sm
Dir. Y00W1L.04 Chapter 4.
Center X−Ray beam on Table

Center X−Ray beam on Bucky Stand

1 Person 1 1/2 Hours


ÍÍ
ÍÍ
See C40E SM: Dir 2150295−100,
Chapter 3, section

SID DETENT ADJUSTMENT


(FINAL POSITIONING) See XT SM Dir 46−019803, section 8

Í
or Dir 2118910−100, section 12
Horizontal SID Detent

Í
or RS−85 SM Mounting Instructions
A32 Horizontal SID Indicators Dir. 2172805−100 Chapter 2.

1 Person 1/2 Hour and RS 85 Interface SM 2165897−100,


Chapter 1, section 5.

COLLIMATOR CALIBRATION

Refer to Inspection Report 2179651 TDR or


2167647 TDR, section 3.3
VERIFICATION or
Inspection Report 2172166 TDR section 4.7

ÍÍ
A33 (SA or SM) Collimator lamp voltage −For XT see Job Card IST002
of this manual

1 Person 1/2 Hour


ÍÍ −For RS 85 see Job Card IST003 of this
manual

Continues on next page

27
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
Step Description
ÍÍ
Done Supporting Documents

From previous page

Manual (SM)
Ultranet
Collimator
Auto/Manual
Auto (SA)

Refer to Inspection Report 2179651 TDR,


section 7.3, or 2167647 TDR or 2172166 TDR,
section 6.3
Proceed with required calibrations labelled:


CALIBRATION
Ultranet SA Collimator + RAFL−S See ULTRANET SA + RAFL−S SM
Dir. Y00W20L.04, Chaps 3, 4 and 6.

ÍÍ
A34 C40E (Use extension cable/support provided to
SG60/SG100 connect RAFL keyboard and fix it on front

ÍÍ
of cabinet − Must be removed after the
1 Person CALIBRATION is completed).
2 Hours
See Job Card IST005 to open the
RAFL−S rack

TOMO−LINK CALIBRATION

CALIBRATION Refer to Inspection Report 2179651 TDR

ÍÍ
A35 section 5.2
 Perform C1, C2, C3, C4, C5 ,C7

ÍÍ
Calibration Procedures
 Drive Power amp adjustment See SM Tomo−Link Dir. 2133821−100,
Section 16 and 20.
1 Person 1 Hour

Continues on next page

28
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Í
Step Description Done
Í Supporting Documents

From previous page

AEC CALIBRATION

FILM PROCESSING
NOTE: Film processor must be
fully operational at this stage to
Processor ready and operational proceed with AEC calibration.
A36
Film/Screen characteristics defined Image quality tools (sensitometer,

ÍÍ etc.) must also be available.

ÍÍ
Film Image Quality available

CALIBRATION

AEC for Table Bucky Technique


A37
AEC for Wall Bucky Technique
See SM MPH 50/65/80, X−ray unit, Dir.

ÍÍ
2137354−100, Chapter 1, Installation, Job
AEC for Tomo Technique Card RG005.

2 People 6 Hours ÍÍ

SID SCALES

XT: Refer to SM 2118733−100 Dir


POSITIONING 46−019803, section 8.6
or Dir 2118910−100, section 12.6
XT Suspension RS−85: Refer to Interface SM
2165897−100 (Chapter 1, section 4 and 5)
A38
RS 85 Tubestand
SG60: Refer to SM 2150293−100
SG100: Refer to SM 2150297−100
SG60/SG100 Chapter 1, section 9

1 Person 1 Hour

Continues on next page

29
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
Step Description Done
ÍÍ
Supporting Documents

From previous page

Refer to Inspection Report 2167647 TDR,


 FUNCTIONAL CHECKS sections 1.2 & 1.5 (MPH), 4.11 (40E), 3.4
(XT/Collimator), 5.4 (SG60/100), 7 (general system
functional check)
or
Refer to Inspection Report 2179651 TDR,
sections 1.2 & 1.5 (MPH), 4.11 (40E), 3.4
SYSTEM CHECKS (XT/collimator), 5.3 (Tomolink), 6.4 (SG60/100),
8 (general system functional check)
or
Inspection Report 2172166 TDR, sections 3.3
(RS 85/collimator), 4.12 (40E), 1.2 & 1.5 (MPH),
SYSTEM COMPONENT 5.4 (SG60/100)
FUNCTIONAL CHECKS
SG60/SG100 Dir 2150297−100
SG60/SG100
ULTRANET Collimator SM Dir.Y00W1L.04;

A39 Collimator SA/SM SM COMPAX 40E CE, Dir. 2150299−100,


Chapter 4, Section 3−4.;
Compax 40E
XT sm Dir. 2118733−100;
XT/RS 85
RS 85 sm Mounting Instructions 03277223;
MPH 50−65−80 MPH SM Dir.2137354−100

ÍÍ
Tomo−Link SM ULTRANET SA & RAFL−S, Dir. Y00W20L.04,
Chapter 4;
Program−X
SM Tomo−Link 2133821−100 (section 18)
2 People 4 Hours

 IMAGE QUALITY

CHECKS

SG60−100 Film Resolution Refer to SG60 SM Dir 2150293−100, Job Card


VF001,
Tomo−Link Image Quality

ÍÍ
A40 SG100 SM Dir 2150257−100, Job Card VF001,
40E Film Resolution

2 People 4 Hours ÍÍ Tomo−Link SM Dir 2133821−100 section 18−10


Refer to 40E Service Manual 2150299−100,
Jobcard VF001.

Continues on next page

30
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9
ÍÍ DIRECTION 2154260−100

ÍÍ
Step Description Done Supporting Documents

From previous page

HHS TEST (If Required)

Note: HHS test or any local, in−house,


No Test regional or governmental acceptance tests.
Step A28
NOTE: If your country requires the product
OK rating plate label to be in the language of that
Yes country, follow the instructions on document
#45474336. Paste the appropriate language
1 Person 4 Hours label over the English rating plate. The rating
plate labels are: 2379361 (Spanish),
2379361−2 (German), 2379361−3 (French),
2379361−4 (Italian).

COVER INSTALLATION
Refer to Inspection Report 2167647 TDR,
4.12 (40E) and 5.5 (SG60/100)

Refer to Inspection Report 2179651 TDR,


sections 4.12 (40E) and 6.5 (SG60/100)
Compax 40E
Refer to Inspection Report 2172166 TDR,
A41 SG60−100 sections 4.13 (40E) and 5.5 (SG60/100)

Tomo−Link See SM COMPAX 40E CE, Dir. 2150299−100,


Chapter 1, Installation, Section 7.
MPH See SM SG60 Dir 2150253−100 Chapter 1
See SM SG100 CE Dir 2150297−100 Chapter 1
XT Suspension
See SM XT RADIOGRAPHIC SUSPENSION
RS−85 Dir 2118733−100; Chapter 3 or 4, Installation.

ÍÍ
ÍÍ
MPH 50/65/80: See SM 2137354−100,
2 People Job Card IST 002
2 Hours
Note: GRN wires must be connected to
metal covers when available.

Continues on next page

31
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Í
Í
Mark each box when done

Step Description Done Supporting Documents

From previous page

FINAL SYSTEM CHECKS

MOVEMENTS
COMMENT: Check for any unusual noise
A42 and on vibration during components
Compax 40E, SG60−100, XT, RS 85 movement. Make appropriate corrections

X−Ray Tube Rotor noise


Í
1 Person 1 Hour
Í

 MISCELLANEOUS

PRODUCT LOCATOR
A43 Refer to COM 3.3 or COM 3.4 ISERV (on Profs)
for Administrator Contact
Please complete and send in cards to
Product Locator Administrator
ÍÍ
ÍÍ
1 Person 1 Hour

GIF & DOA


Refer to COM 3.3 or COM 3.4 ISERV (on Profs)
for more details.
Please complete and send in green card

ÍÍ
A44 to GIF Administrator

DOA tag must be returned with parts


found Dead On Arrival ÍÍ
Continues on next page

32
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ÍÍ
ÍÍ
Mark each box when done

Description Done Supporting Documents


Step

From previous page

PILOT SYSTEM ONLY


(PGS017 FORM)
Refer to PGS017 Form and to COM 3.3 or
COM 3.4 ISERV (on Profs) for USD Contact.
A45
Complete and return the PILOT FEEDBACK
FORM to USD/RAD (BUC)

Complete and return the Installation


Flowchart to ISS (BUC)

ÍÍ
1 Person XX Hours ÍÍ

SOFTWARE BACKUP Refer to MPH SM Dir. 2137354−100,


Job Card IST013
A46  Record MPH set up parameters on the

ÍÍ
calibration diskette

1/2 Hour
ÍÍ
1 Person

END

33
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


1-4 Cable Connections
1-4-1 Preliminaries
1. The following interconnection cable numbers refer to the MIS MAP and GROUNDING
INTERCONNECTING MAP included at the end of this manual.
2. The reference numbers between brackets ( ) refer to internal cables also described in the Service
Manuals.
1-4-2 Mis Chart

System MPH 50/65/80 with Compax 40E / SG 60 or SG100 / XT Suspension


Note: Some of the cables, or set cables, are factory connected at one end. In the following MIS Charts:

− Numbers on the left presented in grey squares are HLA numbers (CABLES #).
− The  symbol is used to identify those connections which remain to be done during on site
installation.
− The symbol indicates where cable clamping is necessary (refer to Job Card IST006 of this Manual
for EMC rules requirements.

DO NOT FORGET TO CLAMP CABLES ENTERING THE MPH CABINET. IT IS


MANDATORY TO MAINTAIN THE SYSTEM EMC COMPLIANT.

34
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


HV TANK X−RAY TUBE

L
MIS 107
CABLE #
 
L
C C
TUBE−I (+) S 46−195120G4 S ANODE
HV ANODE M M X−RAY TUBE HV
1 TERMINALS TUBE−I (−)
CATHODE
L
C
 46−195120G4
 L
C CATHODE
CONNECTIONS
S
S M
M
MPH 50/65/80 CABINET

ST1−1 Black Black (8)−Terminal Strip


ST1−3 White MIS 10680 White (6)−Terminal Strip
2 ST1−5
ST1−7
Green
Brown
 2128497  Green
Brown
(7)−Terminal Strip
(5)−Terminal Strip
MPH A1 A2 MODULE X−RAY TUBE
(Stator Connector ST1−10 Blue Blue (4)−Terminal Strip (Maxiray−75/100)
Panel) STATOR
ST1−11 Orange CABLE Orange (2)−Pressure Switch
ST1−12 Yellow Yellow (1)−Pressure Switch
Clamped Shield Shield GND Screw

PRINTER − X

3 W34

MPH A6 A3 MODULE
XJ1A/B−1 120/240Vac
XJ1A/B−2 0Vac
 2104204
120/240Vac TSxx−L
0Vac TSxx−N MPH A9 MODULE
(Distribution Board) Ground Plug Ground PRINTER−X Power Supply Ground GND STUD
CABLE

CAUTION VERIFY THE 240V XT POWER SETTING XT SUSPENSION

4 MIS10707
MPH A1 MODULE
F15−2 240V AC
F14−2 0Vac
 46−329288G2  240Vac J100−10
0Vac J100−12
(AC Distribution (Control Board 3A1)
Panel) Gnd XT Suspension Power Supply CABLE Gnd J100−6
Ground Plug
J100−5
jumper
J100−1

35
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET


MAIN POWER

CABLE # MIS 27235


(OPTIONAL)
5 P1 Phase 1 (R)

MPH A1 A1 MODULE
P2
P3
 2138905  Phase 2 (S)
Phase 3 (T) LINE SAFETY SWITCH
(AC Distribution Panel)
POWER SUPPLY CABLE
TS−GND GROUND

(PU3) MADRID CONSOLE PRINTER−X


(2125971)
RX RX

XJ4−2 MPH A9−2
2125971 TX
MPH A7 MODULE TX
6 (Control Console)
XJ4−3
Gnd Gnd
MPH A9−3 MPH A9 MODULE
XJ4−5 PRINTER−X Control CABLE MPH A9−7
Shield Shield
Connector Support Connector Support

MPH 50/65/80 CABINET (PU3) MADRID CONSOLE

XJ10−13
XJ10−32 Gnd Gnd
XJ1−4
Gnd Gnd
7 XJ10−9
5Vdc 5Vdc
XJ1−4
XJ10−24 XJ1−24
XJ10−11
TXA W32 TXA
XJ1−11
TXB TXB
XJ10−30
RXA  2126153  RXA
XJ1−30
XJ10−12 XJ1−12
RXB RXB
XJ10−31 XJ1−31
MPH A6 A3 MODULE PROGRAM−X Control CABLE
19V 19V MPH A8 MODULE
(Distribution Board) XJ10−17 XJ1−17
(Program−X Control)
XJ10−18 XJ1−18
XJ10−36 XJ1−36
XJ10−37 XJ1−37
0 / 19V 0 / 19V
XJ10−14 XJ1−14
XJ10−15 XJ1−15
XJ10−33 XJ1−33
XJ10−34 XJ1−34
XJ10−20 XJ1−20
Connectorsupport/clamp Connectorsupport/clamp
Shield Shield

FL switch 2
XJ12−21 FL switch 2
C. switch XJ1−21
XJ12−22 C. switch
OFF XJ1−22
XJ12−24 OFF
8 ON (2126152) XJ1−24

MPH A6 A3 MODULE
XJ12−25
XJ12−26
TXB cons.  2126152  ON
TXB cons.
XJ1−25 MPH A7 MODULE
(Distribution Board) RXB cons. XJ1−26 (Control Console)
XJ12−27 CONSOLE Control CABLE RXB cons.
TXB TAV XJ1−27
XJ12−28 TXB TAV
RXB TAV XJ1−28
XJ12−29 RXB TAV
TXB print XJ1−29
XJ12−30 TXB print
XJ1−30
XJ12−31 RXB print
RXB print XJ1−31
Shield
clamp Shield clamp

SG100 VERTICAL BUCKY STAND


MIS 27225
230Vac 230Vac
TS2−L
TS2−N O Vac 2151729  F11−2
O Vac F10−2
MPH A1 MODULE
9 I / O PANEL
GND STUD GND SG100 Power Supply CABLE GND Ground Plug
(AC Distribution Panel)

36
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SG60/100 VERTICAL
MPH 50/65/80 CABINET BUCKY STAND

CABLE # XJ3−7
O Vac
MIS 27221 O Vac J42−X
XJ3−6
10  
Bucky Start Bucky Start J42−N
XJ3−4 2145641
MPH A6 A4 MODULE 115/230Vac 220Vac J42−D
XJ3−9
(Wall Bucky Board) I / O PANEL
SG60/100 BUCKY CABLE
XJ3−2 RTN RTN J42−d
XJ3−1 Exp Order Exp Order
J42−b
clamp shield shield connector plate

CAUTION VERIFY THE BUCKY SUPPLY 115/240 SETTING


COMPAX 40E TABLE

MIS 27227
11 230Vac 230Vac TB1 (Phase)
MPH A1 MODULE
F18−2
F19−2
O Vac  2167144 O Vac TB1 (Neutral)
I / O PANEL
(AC Distribution Panel)
Ground Plug GND TABLE Power Supply CABLE GND TB1 (Ground)

O Vac
XJ10−7
MIS 27228 O Vac J42−X
XJ10−6
12
 
Bucky Start Bucky Start J42−N
XJ10−4 2145602
MPH A6 A5 MODULE 115/230Vac 220Vac J42−D
XJ10−9
(Table Tomo Board) I / O PANEL
C40E TABLE BUCKY CABLE
XJ10−2 RTN RTN J42−d
XJ10−1 Exp Order Exp Order
J42−b
clamp shield shield connector plate

CAUTION VERIFY THE BUCKY SUPPLY 115/240 SETTING

XJ15−F Right area Right area J40−A


XJ15−D IC output IC output
J40−B
XJ15−A +15VDC MIS 27229 +15VDC J40−C
13
MPH A6 A3 MODULE
XJ15−K Center area
−15VDC.  2145651  Center area
−15VDC.
J40−D
XJ15−C J40−F
(Distribution Board)
XJ15−J Left area Left area J40−H I / O PANEL
C40E Table Ion Chamber CABLE
XJ15−H Start Int. Start Int. J40−J
XJ15−E Relay RTN Relay RTN J40−E
XJ15−B 0VDC 0VDC J40−L
XJ15−P + HV + HV J40−R

clamped shield shield isolated

XT SUSPENSION

J101−9 −10V ref. MANUAL −10V ref.


J43−9
SID pot. SID pot.
J101−10 J43−10
SID pot. rtn
MIS 10705 SID pot. rtn
J43−8
J101−8

14 Control
J101−7 Ref, rtn coll 1
 46−329341G2  Ref, rtn coll 1
Ver. Detent
J43−7
J101−3 Ver. Detent J43−3
Board 3A1
J101−35 Ver. Move Ver. Move J43−35 I / O PANEL
SID CABLE
J101−16 XT ok for Tomo XT ok for Tomo J43−16
J101−12 Ordo Lock Release. Ordo Lock Release. J43−12
J101−11 P21V P21V J43−11

connector support shield shield connector support

37
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

XT SUSPENSION ULTRANET SA/SM COLLIMATOR

CABLE # J1

J1−3,22 0VAC 0VAC J/P2−K


J1−1,20 +24 VAC +24 VAC J/P2−L
P21V P21V
15 J1−5
J1−6 Aux SW Aux SW
J/P2−F
J/P2−B
J1−7 Det Det J/P2−H
Vert SW no det Vert SW no det J/P2 REAR I/O PANEL
J1−9 SW
Vert SW det Vert SW
SW det
J/P2−D
J1−10 J/P2−N
J1−11
Tran SW MIS 10706 Tran SW
J/P2−J
AngSW
Long Long SW
J1−12
J1−8 SW  Ang
Shield
J/P2−E
J/P2−A
46−329291G1 SW J/P2−M clamp on collimator head
shield clamp on collimator head
COLLIMATOR
J1−36 Trans Mtr Hi CABLE Trans Mtr Hi J/P1−X
J1−37 Trans Mtr Lo Trans Mtr Lo J/P1−Z
J1−18 Long Mtr Hi Long Mtr Hi J/P1− −B
Control J1−19 Long Mtr Lo Long Mtr Lo J/P1− −D
Board 3A1 −10VDC −10VDC
J1−32 J/P1−K
J1−33 Trans Pot FdBk Trans Pot FdBk J/P1−H
J1−34 Pot FdBk Rtn Trans FdBk Rtn J/P1−V
J1−14 Long Pot Fbk Long Pot Fbk J/P1−P
J1−15 Pot Fbk Rtn Long FbBk Rtn J/P1−R
J1−16 −10VDC Pot −10VDC Pot Rtn J/P1−M
J/P1 REAR I/O PANEL
J1−27 Undersizing Undersizing J/P1−L
J1−26 PBL PBL J/P1−S
J1−24 Exp Hold Lmp Exp Hold Lmp J/P1− −A
J1−25 Non auto Lmp Non auto Lmp J/P1−F
J1−13 SW ind pwr SW/ind pwr J/P1−B
J1−29 80−100SW −80 to−100 SW J/P1−D
J1−30 −10 − +10 SW −10 to +10 SW J/P1−J
J1−31 +80 − +100 SW +80 to +100 SW J/P1−T
J1−28 Key SW Key SW J/P1− −C

shield

SG60/100 VERTICAL
MPH 50/65/80 CABINET BUCKY STAND

XJ16−F Right area Right area J40−A


XJ16−D IC output IC output
J40−B
+15VDC MIS 27222 +15VDC
16 XJ16−A J40−C

 
XJ16−K Center area Center area
J40−D
2145642
MPH A6 A3 MODULE XJ16−C −15VDC. −15VDC. J40−F
(Distribution Board)
XJ16−J Left area Left area J40−H I / O PANEL
SG100 ION CHAMBER CABLE
XJ16−H Start Int. Start Int. J40−J
XJ16−E Relay RTN Relay RTN J40−E
XJ16−B 0VDC 0VDC J40−L
XJ16−P + HV + HV J40−R

Clamp shield shield isolated

COMPAX 40E TABLE

MIS 27224
17
J44−1 0VDC
SG K7 IN 2145656
 0VDC
SG K7 IN J45−1
J44−2 J45−2
I/O PANEL COMPATIBILITY CABLE Bucky HORIZ I/O PANEL
J44−10 Bucky HORIZ J45−3
shield shield
connector metallic shells connector metallic shells

38
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET SG60/ SG100 AUTO

J/P3−F Bucky VERT Bucky VERT J41−9


J/P3−H Bucky HORIZ Bucky HORIZ J41−10
CABLE # J/P3−P Bucky H N1 MIS 27223 Bucky H N1 J41−13
Bucky H N2 Bucky H N2 J41−12

J/P3−M
J/P3−R
J/P3−N
Bucky ROT
Cassette IN  2145644
Bucky ROT J41−11
Cassette IN J41−6
18 RAFL−S
I/F BOARD J/P3−E +21 VCD
−10VDC
+21 VCD J41−5
−10VDC
I/O PANEL
J/P3−D SG60/100 AUTO TRAY CABLE J41−2
TRANS TRANS J41−4
J/P3−C L/T Cass
J/P3−A L/T Cass J41−3
J/P3−B Long Cass Long Cass J41−7
J/P3−L −10VDC RTN −10VDC RTN J41−1
J/P3−K 0 VDC 0 VDC J41−8
connector metallic plate shield shield connector metallic shells

ROOM INTERLOCK SWITCH


ROOM WARNING LIGHT

 
Door Interlock RTN  OPTIONAL
19 XJ4−1
Door Interlock SUPPLIED IN THE FIELD
ROOM DOOR
INTERLOCK
XJ4−2
MPH A6 A4 MODULE SWICTH
(Wall Bucky Board) System ON (220Vac) WIRING IS FIELD SUP-
XJ4−4 PLIED CONNECT ACCORD-
System ON RTN (0Vac) SUPPLIED IN THE FIELD
XJ4−8 ING TO THE RULES IN
YOUR LOCALLITY

XT SUSPENSION

−10V ref coll1 −10V ref coll1


J201−9 J101−9
20 J201−10
SID Pot SID Pot J101−10
SID Pot Rtn SID Pot Rtn
J201−8 J101−8
Ref Rtn Coll1 Ref Rtn Coll1
J201−7 J101−7
SID A SID A J101−17
J201−17
J201−18 SID B SID B J101−18
J201−19 SID C SID C J101−19
J201−37 SID D SID D J101−37
SID E SID E J101−36
J201−36
J201−11 P21V P21V J101−11
Ordo Lock Rel. Ordo Lock Rel. J101−12
J201−12
Det Valid Det Valid J101−13
J201−13
XT OK for Tomo XT OK for Tomo J101−16
J201−16
Vert Move Vert Move J101−35
J201−35
Vert Det (remote)
MIS 10705 Vert Det (remote)
J201−3 J101−3
J201−14
XT Park Intlk Com.
XT Park Intlk
 46−329341G2  XT Park Intlk Com. J101−14
XT Park Intlk J101−15
J201−15
Trans Mtr Hi Trans Mtr Hi
MPH A6 J201−21 J101−21 Control
I/P Interface Trans Mtr Lo SID&COLLIMATOR
Trans Mtr Lo J101−20 Board 3A1
J201−20 CABLE
Long Mtr Hi Long Mtr Hi J101−2
J201−2
Long Mtr Lo Long Mtr Lo
J201−1 J101−1
−10VDC −10VDC
J201−25 Trans Pot FdBk J101−25
Trans Pot FdBk
J201−24 Pot FdBk Rtn J101−24
Pot FdBk Rtn
J201−23 J101−23
Long Pot Fbk Long Pot Fbk
J201−6 J101−6
Pot Fbk Rtn Pot Fbk Rtn
J201−5 J101−5
−10VDC Pot −10VDC Rtn
J201−4 J101−4
Undersizing Undersizing
J201−30 J101−30
PBL PBL J101−31
J201−31
Exp Hold Lmp Exp Hold Lmp
J201−33 J101−33
Non auto Lmp Non auto Lmp
J201−32 J101−32
SW ind pwr SW ind pwr
J201−34 J101−34
−80 to −100SW −80 to −100SW
J201−28 J101−28
−10−to +10 SW −10 to +10 SW J101−27
J201−27
J201−26 +80 to +100 SW +80 to +100 SW J101−26
Key SW Key SW J101−29
J201−29

connector metallic shell shield shield connector metallic shell

39
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET COMPAX 40E TABLE

CABLE # SID Det Rtn SID Det Rtn


J202−14 J23−14
SID Det SID Det J23−6
J202−6
Det Rtn Det Rtn J23−4
J202−4
21 J202−5 Det MIS 27231 Det J23−5
J202−9 SID out −
SID out +
2145655  SID out −
SID out +
J23−9
J202−1 J23−1
MPH A6 SC2
Diff Com Diff Com I/O PANEL
Interface J202−10 J23−10
FS−Floor Dist SID DETENT FS−Floor Dist
J202−2 CABLE J23−2
−10VDC Ref Rtn −10VDC Ref Rtn
J202−11 J23−11
Vert Move Vert Move
J202−7 J23−7
Vert Move Rtn Vert Move Rtn J23−15
J202−15
P21VDC P21VDC J23−13
J202−13
−10VDC Ref −10VDC Ref J23−3
J202−3
connector metallic shells shield shield connector metallic shells

J/P2−F Table Vert&Cass IN MIS 27230 Table Vert&Cass IN J41−9 not used
Cassette IN Cassette IN
22 J/P2−E J41−10
 
J/P2−P Table −15deg&Cass IN Table −15deg&Cass IN J41−13 not used
J/P2−K Table Horiz&Cass IN 2145618 Table Horiz&Cass IN J41−6
+21VDC isolated +21VDC isolated
J/P2−B J41−5
RAFL−S J/P2−A −10V −10V J41−2 I/O PANEL
TRANS TRANS
I/F BOARD J/P2−C DC
L/T Cass DC
L/T Cass
J41−4
J/P2−L J41−3
J/P2−H Long Cass COMPAX 40E Long Cass J41−7
J/P2−D −10VDC RTN AUTOMATIC TRAY CABLE −10VDC RTN J41−1
J/P2−R 0 VDC isolated 0 VDC isolated J41−8
connector metallic plate shield shield connector metallic shells

MPH 50/65/80 CABINET


INTERNAL
A format A format J/P10−V
XJ5−V
B format B format J/P10−W
XJ5−W
C format C format J/P10−X
XJ5−X
D format D format J/P10−Y
XJ5−Y
XJ5−HH E format  E format J/P10−HH
MIS 27232

common common
XJ5−MM J/P10−MM
MPH A6 A8 MODULE XJ5−J
Undersizing 2145659 Undersizing J/P10−J MPH A6 CL1
Exd Sig Enable Exd Sig Enable J/P10−R
RAFL−S
XJ5−R
Exp Sig
SAS INTERFACE CABLE
Exp Sig I/O PANEL
XJ5−A J/P10−A
Tube Sel Tube Sel J/P10−C
XJ5−C
Tech Dir Sel Tech Dir Sel J/P10−D
XJ5−D
Table Table
J/P10−U
XJ5−U Bucky
common Sel
Bucky
common Sel J/P10−H
XJ5−H Sel Sel
shield shield

clamp on I/F generator plate connector metallic shells

230Vac (isolated) MIS 27233 230Vac (isolated)


P1−3 TB−4
O Vac 2154537 O Vac
MPH A6 CL1 RAFL−S P1−1 TB−5 MPH A6 SC3
I/O PANEL GND POWER SUPPLY CABLE GND
GND TB−6

MIS 27240
230Vac (isolated) 230Vac (isolated)
P F13
MPH A6 SC3 2152750
O Vac O Vac MPH A1
FILTER N F12
GND UPPER CABINET POWER SUPPLY CABLE GND
GND GND Terminal

40
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET MAIN POWER

CABLE #
MIS 27235
1.1
MPH A1 A1 MODULE
2138905 
LINE SAFETY SWITCH
(AC Distribution Panel)
POWER SUPPLY CABLE
TS−GND GROUND

(PU3) MADRID CONSOLE


2.1 (45553701)
MPH A1 MODULE
 45553701 GND (BAR GROUND)
Ground Plug
(AC Distribution Panel)
CONSOLE GROUND CABLE

COMPAX 40E TABLE

MIS 27300
3.1
MPH A1 MODULE  2140362
I / O PANEL
(AC Distribution Panel) Ground Plug TB1 (Ground)
TABLE Ground CABLE

XT SUSPENSION

4.1 MIS 10708


MPH A1 MODULE  (46−316922G2) SUPPLIED with XT SUSPENSION 
(AC Distribution Panel) Ground Plug GND STUD BRIDGE
GROUND CABLE

SG 60/100 VERTICAL BUCKY


STAND
MIS 27301
5.1  2140362
Ground Plug TS2−GND
MPH A1 MODULE I / O PANEL
(AC Distribution Panel) SG 60/100 GROUND CABLE

PRINTER−X

W19
6.1
MPH A1 MODULE
(AC Distribution Panel) Ground Plug
 (2105836) SUPPLIED with PRINTER−X
GND STUD MPH A9
PRINTER−X Ground CABLE MODULE

MAXIRAY100 XRAY TUBE XT SUSPENSION

8.1 (46−216211G10)
GND STUD  46−216211G10 
Ground Stud GND STUD BRIDGE
GROUND CABLE

41
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


1-4-3 Mis Chart

System MPH 50/65/80 and RS85 Tube Stand with Compax 40E / SG 60 or SG100
Note: Some of the cables, or set cables, are factory connected at one end. In the following MIS Charts:
− Numbers on the left presented in grey squares are HLA numbers (CABLES #).
− The  symbol is used to identify those connections which remain to be done during on site
installation.
− The symbol indicates where cable clamping is necessary (refer to Job Card IST006 of this Manual
for EMC rules requirements.

DO NOT FORGET TO CLAMP CABLES ENTERING THE MPH CABINET. IT IS


MANDATORY TO MAINTAIN THE SYSTEM EMC COMPLIANT.

HV TANK X−RAY TUBE

L MIS 107
 
L
CABLE # C C
TUBE−I (+) S 3600208..0/1/2/8 S ANODE
ANODE M M X−RAY TUBE HV
HV
1 TERMINALS TUBE−I (−)
CATHODE
L
C
 3600208..0/1/2/8  C
L
CATHODE
CONNECTIONS
S
S M
M

MPH 50/65/80 CABINET

ST1−1 Black Black (8)−Terminal Strip

2 ST1−3 White MIS 27239 White (7)−Terminal Strip


ST1−5
ST1−7
Green
Brown
  2145653  Green
Brown
(6)−Terminal Strip
(5)−Terminal Strip
MPH A1 A2 MODULE X−RAY TUBE
(Stator Connector Panel) ST1−10 Blue Blue (3)−Terminal Strip (Maxiray−75/100)
STATOR CABLE
ST1−11 Orange Orange (2)−Pressure Switch
ST1−12 Yellow Yellow (1)− Pressure Switch
Clamped Shield Shield GND Screw

PRINTER − X

3 W34
XJ1A/B−1 120/240Vac
XJ1A/B−2 0Vac
 2104204
120/240Vac TSxx−L
0Vac TSxx−N MPH A9 MODULE
MPH A6 A3 MODULE
(Distribution Board) Ground Plug Ground PRINTER−X Power Supply CABLE Ground GND STUD

MPH 50/65/80 CABINET

4 MIS 27238
F15−2 230Vac 230Vac X1−L MPH SCI
MPH A1 MODULE 2165548
F14−2 0Vac 0Vac X1−N RS 85 POWER
(AC Distribution Panel) SUPPLY
Gnd RS85 Power Supply CABLE Gnd PE1−GND
Ground Plug

42
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET


MAIN POWER

CABLE # MIS 27235


(OPTIONAL)
5 P1 Phase 1 (R)

MPH A1 A1 MODULE
P2
P3
 2138905  Phase 2 (S)
Phase 3 (T) LINE SAFETY SWITCH
(AC Distribution Panel)
POWER SUPPLY CABLE
TS−GND GROUND

(PU3) MADRID CONSOLE PRINTER−X


(2125971)
RX RX

XJ4−2 MPH A9−2
2125971 TX
MPH A7 MODULE TX
6 (Control Console)
XJ4−3
Gnd Gnd
MPH A9−3 MPH A9 MODULE
XJ4−5 PRINTER−X Control CABLE MPH A9−7
Shield Shield
Connector Support Connector Support

MPH 50/65/80 CABINET (PU3) MADRID CONSOLE

XJ10−13
XJ10−32 Gnd Gnd
XJ1−4
Gnd Gnd
7 XJ10−9
5Vdc 5Vdc
XJ1−4
XJ10−24 XJ1−24
XJ10−11
TXA W32 TXA
XJ1−11
TXB TXB
XJ10−30
RXA  2126153  RXA
XJ1−30
XJ10−12 XJ1−12
RXB RXB
XJ10−31 XJ1−31
MPH A6 A3 MODULE PROGRAM−X Control CABLE
19V 19V MPH A8 MODULE
(Distribution Board) XJ10−17 XJ1−17
(Program−X Control)
XJ10−18 XJ1−18
XJ10−36 XJ1−36
XJ10−37 XJ1−37
0 / 19V 0 / 19V
XJ10−14 XJ1−14
XJ10−15 XJ1−15
XJ10−33 XJ1−33
XJ10−34 XJ1−34
XJ10−20 XJ1−20
Connectorsupport/clamp Connectorsupport/clamp
Shield Shield

FL switch 2
XJ12−21 FL switch 2
C. switch XJ1−21
XJ12−22 C. switch
OFF XJ1−22
XJ12−24 OFF
8 ON (2126152) XJ1−24

MPH A6 A3 MODULE
XJ12−25
XJ12−26
TXB cons.  2126152  ON
TXB cons.
XJ1−25 MPH A7 MODULE
(Distribution Board) RXB cons. XJ1−26 (Control Console)
XJ12−27 CONSOLE Control CABLE RXB cons.
TXB TAV XJ1−27
XJ12−28 TXB TAV
RXB TAV XJ1−28
XJ12−29 RXB TAV
TXB print XJ1−29
XJ12−30 TXB print
XJ1−30
XJ12−31 RXB print
RXB print XJ1−31
Shield
clamp Shield clamp

SG100 VERTICAL BUCKY STAND


MIS 27225
230Vac 230Vac
TS2−L
TS2−N O Vac 2151729  F11−2
O Vac F10−2
MPH A1 MODULE
9 I / O PANEL
GND STUD GND SG100 Power Supply CABLE GND Ground Plug
(AC Distribution Panel)

43
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SG60/100 VERTICAL
MPH 50/65/80 CABINET BUCKY STAND

CABLE #
O Vac
XJ3−7
MIS 27221 O Vac J42−X
XJ3−6
10  
Bucky Start Bucky Start J42−N
XJ3−4 2145641
MPH A6 A4 MODULE 230Vac 220Vac J42−D
XJ3−9
(Wall Bucky Board) I / O PANEL
SG60/100 BUCKY CABLE
XJ3−2 RTN RTN J42−d
XJ3−1 Exp Order Exp Order
J42−b
clamp shield shield connector plate

COMPAX 40E TABLE

11 MIS 27227
230Vac 230Vac TB1 (Phase)
MPH A1 MODULE
F18−2
F19−2
O Vac  2167144 O Vac TB1 (Neutral)
I / O PANEL
(AC Distribution Panel)
Ground Plug GND TABLE Power Supply CABLE GND TB1 (Ground)

O Vac
XJ10−7
MIS 27228 O Vac J42−X
XJ10−6
12  
Bucky Start Bucky Start J42−N
XJ10−4 2145602
MPH A6 A5 MODULE 230Vac 220Vac J42−D
XJ10−9
(Table Tomo Board) I / O PANEL
C40E TABLE BUCKY CABLE
XJ10−2 RTN RTN J42−d
XJ10−1 Exp Order Exp Order
J42−b
clamp shield shield
connector plate

XJ15−F Right area Right area J40−A


XJ15−D IC output IC output
J40−B
XJ15−A +15VDC MIS 27229 +15VDC J40−C
13 
MPH A6 A3 MODULE
XJ15−K Center area
−15VDC.
 2145651
Center area
−15VDC.
J40−D
XJ15−C J40−F
(Distribution Board)
XJ15−J Left area Left area J40−H I / O PANEL
C40E Table Ion Chamber CABLE
XJ15−H Start Int. Start Int. J40−J
XJ15−E Relay RTN Relay RTN J40−E
XJ15−B 0VDC 0VDC J40−L
XJ15−P + HV + HV J40−R

clamp shield shield isolated

MANUAL
RS 85 TUBESTAND

MIS 27293
24VAC

24VAC
X3−13
14 TOWER X3−14
0VAC 2165551 0VAC J12−1
J12−2
GND GND FUSE BOARD
GND GND

44
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


SM or SA ULTRANET
RS 85 TUBE STAND
COLLIMATOR
MIS 27303
INTERNAL
TOWER
SUPPLIED WITH RS85  REAR I/O PANEL
X3 J/P2
LOCK CABLE

RS 85 TUBE STAND MPH 50/65/80 CABINET


MIS 27304

24VDC 2171350 24VDC 5
X3−1
TOWER 24 VDC CABLE MPH SCI
X3−3,5,7
0VDC 0VDC 4 RS 85 POWER
SUPPLY

RS 85 TUBE STAND
COMPAX 40E TABLE

X3−7 SID Det SID Det J21−4


CABLE # SID Det Rtn MANUAL SID Det Rtn
X3−2 J21−3
X3−8 Vert Move MIS 27292 Vert Move J21−14
15 X3−3 Vert Move Rtn   Vert Move Rtn
J21−15
(2165550)
X3−6 Detent Detent J21−6
TOWER
Detent Rtn Detent Rtn I / O PANEL
X3−5 J21−5
−10VDC Ref RS 85 −10VDC Ref J21−24
X4−5
(SID Pot.&Brakes
X4−2 FS−Floor Dist FS−Floor Dist J21−10
control)
X4−3 Diff Com CABLES Diff Com J21−9
X4−1 −10VDC Ref Rtn −10VDC Ref Rtn J21−23
X4−4 shield shield connector metallic shells
Clamp

SG60/100 VERTICAL
MPH 50/65/80 CABINET BUCKY STAND

XJ16−F Right area Right area J40−A


XJ16−D IC output IC output
J40−B
+15VDC
MIS 27222 +15VDC
16 XJ16−A J40−C

 
XJ16−K Center area Center area
J40−D
2145642
MPH A6 A3 MODULE XJ16−C −15VDC. −15VDC. J40−F
(Distribution Board)
XJ16−J Left area Left area J40−H I / O PANEL
SG100 ION CHAMBER CABLE
XJ16−H Start Int. Start Int. J40−J
XJ16−E Relay RTN Relay RTN J40−E
XJ16−B 0VDC 0VDC J40−L
XJ16−P + HV + HV J40−R

Clamp shield shield isolated

COMPAX 40E TABLE


MIS 27224
17 J44−1 0VDC
SG K7 IN 2145656
 0VDC
SG K7 IN J45−1
J44−2 J45−2
I/O PANEL COMPATIBILITY CABLE Bucky HORIZ I/O PANEL
J44−10 Bucky HORIZ J45−3
shield shield
connector metallic shells connector metallic shells

RAFL−S RACK

J/P3−F Bucky VERT Bucky VERT J41−9


18 J/P3−H Bucky HORIZ
Bucky H N1 MIS 27223 Bucky HORIZ J41−10
Bucky H N1 J41−13
J/P3−P Bucky H N2 Bucky H N2 J41−12

J/P3−M
J/P3−R Bucky ROT  Bucky ROT J41−11
Cassette IN 2145644 Cassette IN J41−6 I/O PANEL
I/F BOARD J/P3−N
J/P3−E +21 VCD +21 VCD J41−5
J/P3−D −10VDC SG60/100 AUTO TRAY CABLE −10VDC J41−2
TRANS TRANS J41−4
J/P3−C L/T Cass
J/P3−A L/T Cass J41−3
J/P3−B Long Cass Long Cass J41−7
J/P3−L −10VDC RTN −10VDC RTN J41−1
J/P3−K 0 VDC 0 VDC J41−8
connector metallic plate shield shield
connector metallic shells

45
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET SG60/ SG100 AUTO

J/P3−F Bucky VERT Bucky VERT J41−9


J/P3−H Bucky HORIZ Bucky HORIZ J41−10
CABLE # J/P3−P Bucky H N1 MIS 27223 Bucky H N1 J41−13
J/P3−M Bucky H N2 Bucky H N2 J41−12
Bucky ROT  Bucky ROT J41−11
J/P3−R Cassette IN  Cassette IN J41−6
J/P3−N 2145644
18 RAFL−S
I/F BOARD J/P3−E +21 VCD
−10VDC
+21 VCD J41−5
−10VDC
I/O PANEL
J/P3−D SG60/100 AUTO TRAY CABLE J41−2
TRANS TRANS J41−4
J/P3−C L/T Cass
J/P3−A L/T Cass J41−3
J/P3−B Long Cass Long Cass J41−7
J/P3−L −10VDC RTN −10VDC RTN J41−1
J/P3−K 0 VDC 0 VDC J41−8
connector metallic plate shield shield connector metallic shells

ROOM INTERLOCK SWITCH


ROOM WARNING LIGHT

 
Door Interlock RTN  OPTIONAL
19 XJ4−1
Door Interlock SUPPLIED IN THE FIELD
ROOM DOOR
INTERLOCK
XJ4−2
MPH A6 A4 MODULE SWICTH
(Wall Bucky Board) System ON (220Vac) WIRING IS FIELD SUP-
XJ4−4 PLIED CONNECT ACCORD-
System ON RTN (0Vac) SUPPLIED IN THE FIELD
XJ4−8 ING TO THE RULES IN
YOUR LOCALLITY

ULTRANET SA COLLIMATOR
J201−15
Trans Mtr Hi
20 J201−21
Trans Mtr Hi
J/P1−X
Trans Mtr Lo Trans Mtr Lo
J201−20 J/P1−Z
Long Mtr Hi Long Mtr Hi J/P1− −B
J201−2
Long Mtr Lo Long Mtr Lo
J201−1 J/P1− −D
−10VDC −10VDC
J201−25 J/P1−K
Trans Pot FdBk Trans Pot FdBk
J201−24 Pot FdBk Rtn J/P1−H
Pot FdBk Rtn
J201−23 J/P1−V
MIS 27236
J201−6
Long Pot Fbk
 Long Pot Fbk
J/P1−P

 
Pot Fbk Rtn Pot Fbk Rtn
MPH A6 J201−5 J/P1−R
2165546
SC2 Interface −10VDC Pot −10VDC Rtn REAR I/O
J201−4 J/P1−M PANEL
Undersizing Undersizing
J201−30 J/P1−L
PBL COLLIMATOR PBL
J201−31 CABLE J/P1−S
Exp Hold Lmp Exp Hold Lmp
J201−33 J/P1− −A
Non auto Lmp Non auto Lmp
J201−32 J/P1−F
SW ind pwr SW ind pwr
J201−34 J/P1−B
−80 to −100SW −80 to −100SW
J201−28 J/P1−D
−10−to +10 SW −10 to +10 SW J/P1−J
J201−27
J201−26 +80 to +100 SW +80 to +100 SW J/P1−T
Key SW Key SW J/P1− −C
J201−29

connector metallic shell shield shield clamp

MPH 50/65/80 CABINET COMPAX 40E TABLE

SID Det Rtn SID Det Rtn


J202−14 J23−14
SID Det SID Det J23−6
J202−6
Det Rtn Det Rtn J23−4
J202−4
21 J202−5 Det MIS 27231 Det J23−5
J202−9 SID out −
SID out +
2145655  SID out −
SID out +
J23−9
J202−1 J23−1
MPH A6 SC2
Diff Com Diff Com I/O PANEL
Interface J202−10 J23−10
FS−Floor Dist SID DETENT FS−Floor Dist
J202−2 CABLE J23−2
−10VDC Ref Rtn −10VDC Ref Rtn
J202−11 J23−11
Vert Move Vert Move
J202−7 J23−7
Vert Move Rtn Vert Move Rtn J23−15
J202−15
P21VDC P21VDC J23−13
J202−13
−10VDC Ref −10VDC Ref J23−3
J202−3
connector metallic shells shield shield connector metallic shells

46
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET COMPAX 40E TABLE

CABLE #
J/P2−F Table Vert&Cass IN MIS 27230 Table Vert&Cass IN J41−9 not used
Cassette IN Cassette IN
22 J/P2−E J41−10
 
J/P2−P Table −15deg&Cass IN Table −15deg&Cass IN J41−13 not used
J/P2−K Table Horiz&Cass IN 2145618 Table Horiz&Cass IN J41−6
+21VDC isolated +21VDC isolated
J/P2−B J41−5
RAFL−S J/P2−A −10V −10V J41−2 I/O PANEL
TRANS TRANS
I/F BOARD J/P2−C DC
L/T Cass DC
L/T Cass
J41−4
J/P2−L J41−3
J/P2−H Long Cass COMPAX 40E Long Cass J41−7
J/P2−D −10VDC RTN AUTOMATIC TRAY CABLE −10VDC RTN J41−1
J/P2−R 0 VDC isolated 0 VDC isolated J41−8
connector metallic plate shield shield connector metallic shells

RS85 TUBE STAND

23 J301−17 SID Det


SID Det Rtn
SID Det
SID Det Rtn
X3−7

J301−16 X3−2
J301−18 Vert Move Vert Move X3−8
Vert Move Rtn Vert Move Rtn X3−3
J301−19
MIS 27237
J301−15 Detent Detent X3−6
J301−14 Detent Rtn  2165547  Detent Rtn X3−5
TOWER − POTENTIOMETER

SID C Intlk SID C Intlk


J301−13 X3−11
MPH A6 SC2 SID D Intlk SID D Intlk
Interface J301−9 X3−12
21VDC isolated RS85 (SID Pot. + Detent) 21VDC isolated
J301−8 X3−15
0VAC CABLES 0VAC
J301−10,11,12,21 X3−14
J301−22,23,24,25 24VAC 24VAC
X3−13
J21−24 −10VDC Ref −10VDC Ref X4−5
J21−10 FS−Floor Dist FS−Floor Dist X4−2
J21−9 Diff Com Diff Com X4−3
J21−23 −10VDC Ref Rtn −10VDC Ref Rtn X4−1

connector metallic shells shield shield X4−4 clamp

MPH 50/65/80 CABINET


INTERNAL
A format A format J/P10−V
XJ5−V
B format B format J/P10−W
XJ5−W
C format C format J/P10−X
XJ5−X
D format D format J/P10−Y
XJ5−Y
XJ5−HH E format  E format J/P10−HH
MIS 27232

common common J/P10−MM
XJ5−MM
MPH A6 A8 MODULE XJ5−J
Undersizing 2145659 Undersizing J/P10−J MPH A6 CL1
XJ5−R
Exd Sig Enable Exd Sig Enable J/P10−R RAFL−S
Exp Sig
SAS INTERFACE CABLE
Exp Sig J/P10−A I/O PANEL
XJ5−A
Tube Sel Tube Sel J/P10−C
XJ5−C
Tech Dir Sel Tech Dir Sel J/P10−D
XJ5−D
Table Table
J/P10−U
XJ5−U Bucky
common Sel
Bucky
common Sel J/P10−H
XJ5−H Sel Sel
shield shield

clamp on I/F generator plate connector metallic shells

230Vac (isolated) MIS 27233 230Vac (isolated)


P1−3 TB−4
O Vac 2154537 O Vac
MPH A6 CL1 RAFL−S P1−1 TB−5 MPH A6 SC3
I/O PANEL GND POWER SUPPLY CABLE GND
GND TB−6

MIS 27240
230Vac (isolated) 230Vac (isolated)
P F13
MPH A6 SC3 2152750
O Vac O Vac MPH A1
FILTER N F12
GND UPPER CABINET POWER SUPPLY CABLE GND
GND GND Terminal

47
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET MAIN POWER

CABLE #
MIS 27235
1.1
MPH A1 A1 MODULE
2138905 
LINE SAFETY SWITCH
(AC Distribution Panel)
POWER SUPPLY CABLE
TS−GND GROUND

(PU3) MADRID CONSOLE


2.1 (45553701)
MPH A1 MODULE
 45553701 GND (BAR GROUND)
Ground Plug
(AC Distribution Panel)
CONSOLE GROUND CABLE

COMPAX 40E TABLE

MIS 27300
3.1
MPH A1 MODULE  2140362
I / O PANEL
(AC Distribution Panel) Ground Plug TB1 (Ground)
TABLE Ground CABLE

RS85 TUBE STAND

4.1 MIS 27302


MPH A1 MODULE  2140362 
(AC Distribution Panel) Ground Plug GND STUD TOWER
GROUND CABLE

SG 60/100 VERTICAL BUCKY


STAND
MIS 27301
5.1  2140362
Ground Plug TS2−GND
MPH A1 MODULE I / O PANEL
(AC Distribution Panel) SG 60/100 GROUND CABLE

PRINTER−X

6.1 W19
MPH A1 MODULE
(AC Distribution Panel) Ground Plug
 (2105836) SUPPLIED with PRINTER−X
GND STUD MPH A9
230Vac MODULE
PRINTER−X Ground CABLE

MPH 50/65/80 CABINET

7.1 MIS 27238


F15−2 X1−L
MPH A1 MODULE 2165548
F14−2 0Vac X1−N Terminals Strip
(AC Distribution Panel)
Ground Plug RS85 Power Supply CABLE Gnd PE1−GND

MAXIRAY100 XRAY TUBE RS85 TUBE STAND

8.1 MIS 27305


GND STUD  2183955 
Ground Stud GND STUD TOWER
GROUND CABLE

48
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


1-4-4 Mis Chart

System MPH 50/65/80 with Compax 40E / SG 60 or SG100 / XT Suspension with Tomo Link
Note: Some of the cables, or set cables, are factory connected at one end. In the following MIS Charts:

− Numbers on the left presented in grey squares are HLA numbers (CABLES #).
− The  symbol is used to identify those connections which remain to be done during on site
installation.
− The symbol indicates where cable clamping is necessary (refer to Job Card IST006 of this Manual
for EMC rules requirements.

DO NOT FORGET TO CLAMP CABLES ENTERING THE MPH CABINET. IT IS


MANDATORY TO MAINTAIN THE SYSTEM EMC COMPLIANT.

HV TANK X−RAY TUBE

L
MIS 107
CABLE #
 
L
C C
TUBE−I (+) S 46−195120G4 S ANODE
ANODE X−RAY TUBE HV
HV M M CONNECTIONS
1 TERMINALS TUBE−I (−)
CATHODE
L
C
 46−195120G4
 L
C CATHODE
S
S M
M
MPH 50/65/80 CABINET

ST1−1 Black Black (8)−Terminal Strip


ST1−3 White MIS 10680 White (6)−Terminal Strip
2 ST1−5
ST1−7
Green
Brown
 2128497  Green
Brown
(7)−Terminal Strip
(5)−Terminal Strip
MPH A1 A2 MODULE X−RAY TUBE
(Stator Connector ST1−10 Blue Blue (4)−Terminal Strip (Maxiray−75/100)
Panel) STATOR
ST1−11 Orange CABLE Orange (2)−Pressure Switch
ST1−12 Yellow Yellow (1)−Pressure Switch
Clamped Shield Shield GND Screw

PRINTER − X

3 W34

MPH A6 A3 MODULE
XJ1A/B−1 120/240Vac
XJ1A/B−2 0Vac
 2104204
120/240Vac TSxx−L
0Vac TSxx−N MPH A9 MODULE
(Distribution Board) Ground Plug Ground PRINTER−X Power Supply Ground GND STUD
CABLE

XT SUSPENSION

MIS10707
4
MPH A1 MODULE
F15−2 230Vac
F14−2 0Vac
 46−329288G2  240Vac J100−10
0Vac J100−12 (Control Board 3A1)
(AC Distribution
Panel) Gnd XT Suspension Power Supply CABLE Gnd J100−6
Ground Plug
J100−5
jumper
J100−1

49
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET


MAIN POWER

CABLE # MIS 27235


(OPTIONAL)
5 P1 Phase 1 (R)

MPH A1 A1 MODULE
P2
P3
 2138905  Phase 2 (S)
Phase 3 (T) LINE SAFETY SWITCH
(AC Distribution Panel)
POWER SUPPLY CABLE
TS−GND GROUND

(PU3) MADRID CONSOLE PRINTER−X


(2125971)
RX RX

XJ4−2 MPH A9−2
2125971 TX
MPH A7 MODULE TX
6 (Control Console)
XJ4−3
Gnd Gnd
MPH A9−3 MPH A9 MODULE
XJ4−5 PRINTER−X Control CABLE MPH A9−7
Shield Shield
Connector Support Connector Support

MPH 50/65/80 CABINET (PU3) MADRID CONSOLE

XJ10−13
XJ10−32 Gnd Gnd
XJ1−4
Gnd Gnd
7 XJ10−9
5Vdc 5Vdc
XJ1−4
XJ10−24 XJ1−24
XJ10−11
TXA W32 TXA
XJ1−11
TXB TXB
XJ10−30
RXA  2126153  RXA
XJ1−30
XJ10−12 XJ1−12
RXB RXB
XJ10−31 XJ1−31
MPH A6 A3 MODULE PROGRAM−X Control CABLE
19V 19V MPH A8 MODULE
(Distribution Board) XJ10−17 XJ1−17
(Program−X Control)
XJ10−18 XJ1−18
XJ10−36 XJ1−36
XJ10−37 XJ1−37
0 / 19V 0 / 19V
XJ10−14 XJ1−14
XJ10−15 XJ1−15
XJ10−33 XJ1−33
XJ10−34 XJ1−34
XJ10−20 XJ1−20
Connectorsupport/clamp Connectorsupport/clamp
Shield Shield

FL switch 2
XJ12−21 FL switch 2
C. switch XJ1−21
XJ12−22 C. switch
OFF XJ1−22
XJ12−24 OFF
8 ON (2126152) XJ1−24

MPH A6 A3 MODULE
XJ12−25
XJ12−26
TXB cons.  2126152  ON
TXB cons.
XJ1−25 MPH A7 MODULE
(Distribution Board) RXB cons. XJ1−26 (Control Console)
XJ12−27 CONSOLE Control CABLE RXB cons.
TXB TAV XJ1−27
XJ12−28 TXB TAV
RXB TAV XJ1−28
XJ12−29 RXB TAV
TXB print XJ1−29
XJ12−30 TXB print
XJ1−30
XJ12−31 RXB print
RXB print XJ1−31
Shield
clamp Shield clamp

SG100 VERTICAL BUCKY STAND


MIS 27225
230Vac 230Vac
TS2−L
TS2−N O Vac 2151729  F11−2
O Vac F10−2
MPH A1 MODULE
9 I / O PANEL
GND STUD GND SG100 Power Supply CABLE GND Ground Plug
(AC Distribution Panel)

50
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SG60/100 VERTICAL
MPH 50/65/80 CABINET BUCKY STAND

CABLE # XJ3−7
O Vac
MIS 27221 O Vac J42−X
XJ3−6
10  
Bucky Start Bucky Start J42−N
XJ3−4 2145641
MPH A6 A4 MODULE 115/230Vac 220Vac J42−D
XJ3−9
(Wall Bucky Board) I / O PANEL
SG60/100 BUCKY CABLE
XJ3−2 RTN RTN J42−d
XJ3−1 Exp Order Exp Order
J42−b
clamp shield shield connector plate

CAUTION VERIFY THE BUCKY SUPPLY 115/240 SETTING


COMPAX 40E TABLE

MIS 27227
11 230Vac 230Vac TB1 (Phase)
MPH A1 MODULE
F18−2
F19−2 O Vac  2167144 O Vac TB1 (Neutral) I / O PANEL
(AC Distribution Panel)
Ground Plug GND TABLE Power Supply CABLE GND TB1 (Ground)

O Vac
XJ10−7
MIS 27228 O Vac J42−X
XJ10−6
12
 
Bucky Start Bucky Start J42−N
XJ10−4 2145602
MPH A6 A5 MODULE 115/230Vac 220Vac J42−D
XJ10−9
(Table Tomo Board) I / O PANEL
C40E TABLE BUCKY CABLE
XJ10−2 RTN RTN J42−d
XJ10−1 Exp Order Exp Order
J42−b
clamp shield shield connector plate

CAUTION VERIFY THE BUCKY SUPPLY 115/240 SETTING

XJ15−F Right area Right area J40−A


XJ15−D IC output IC output
J40−B
XJ15−A +15VDC MIS 27229 +15VDC J40−C
13
MPH A6 A3 MODULE
XJ15−K Center area
−15VDC.  2145651  Center area
−15VDC.
J40−D
XJ15−C J40−F
(Distribution Board)
XJ15−J Left area Left area J40−H I / O PANEL
C40E Table Ion Chamber CABLE
XJ15−H Start Int. Start Int. J40−J
XJ15−E Relay RTN Relay RTN J40−E
XJ15−B 0VDC 0VDC J40−L
XJ15−P + HV + HV J40−R

clamped shield shield isolated

XT SUSPENSION

J101−9 −10V ref. MANUAL −10V ref.


J43−9
SID pot. SID pot.
J101−10 J43−10
SID pot. rtn
MIS 10705 SID pot. rtn
J43−8
J101−8

14 Control
J101−7 Ref, rtn coll 1
 46−329341G2  Ref, rtn coll 1
Ver. Detent
J43−7
J101−3 Ver. Detent J43−3
Board 3A1
J101−35 Ver. Move Ver. Move J43−35 I / O PANEL
SID CABLE
J101−16 XT ok for Tomo XT ok for Tomo J43−16
J101−12 Ordo Lock Release. Ordo Lock Release. J43−12
J101−11 P21V P21V J43−11

connector support shield shield connector support

51
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

XT SUSPENSION ULTRANET SA/SM COLLIMATOR

CABLE # J1

J1−3,22 0VAC 0VAC J/P2−K


J1−1,20 +24 VAC +24 VAC J/P2−L
P21V P21V
15 J1−5
J1−6 Aux SW Aux SW
J/P2−F
J/P2−B
J1−7 Det Det J/P2−H
Vert SW no det Vert SW no det J/P2 REAR I/O PANEL
J1−9 SW
Vert SW det Vert SW
SW det
J/P2−D
J1−10 J/P2−N
J1−11
Tran SW MIS 10706 Tran SW
J/P2−J
Long SW Long SW
J1−12
J1−8 Ang  Ang
J/P2−E
J/P2−A
SW
46−329291G1 SW
Shield J/P2−M clamp on collimator head
shield clamp on collimator head
COLLIMATOR
J1−36 Trans Mtr Hi CABLE Trans Mtr Hi J/P1−X
J1−37 Trans Mtr Lo Trans Mtr Lo J/P1−Z
J1−18 Long Mtr Hi Long Mtr Hi J/P1− −B
Control J1−19 Long Mtr Lo Long Mtr Lo J/P1− −D
Board 3A1 −10VDC −10VDC
J1−32 J/P1−K
J1−33 Trans Pot FdBk Trans Pot FdBk J/P1−H
J1−34 Pot FdBk Rtn Trans FdBk Rtn J/P1−V
J1−14 Long Pot Fbk Long Pot Fbk J/P1−P
J1−15 Pot Fbk Rtn Long FbBk Rtn J/P1−R
J1−16 −10VDC Pot −10VDC Pot Rtn J/P1−M
J/P1 REAR I/O PANEL
J1−27 Undersizing Undersizing J/P1−L
J1−26 PBL PBL J/P1−S
J1−24 Exp Hold Lmp Exp Hold Lmp J/P1− −A
J1−25 Non auto Lmp Non auto Lmp J/P1−F
J1−13 SW ind pwr SW/ind pwr J/P1−B
J1−29 80−100SW −80 to−100 SW J/P1−D
J1−30 −10 − +10 SW −10 to +10 SW J/P1−J
J1−31 +80 − +100 SW +80 to +100 SW J/P1−T
J1−28 Key SW Key SW J/P1− −C

shield

SG60/100 VERTICAL
MPH 50/65/80 CABINET BUCKY STAND

XJ16−F Right area Right area J40−A


XJ16−D IC output IC output
J40−B
+15VDC MIS 27222 +15VDC
16 XJ16−A J40−C

 
XJ16−K Center area Center area
J40−D
2145642
MPH A6 A3 MODULE XJ16−C −15VDC. −15VDC. J40−F
(Distribution Board)
XJ16−J Left area Left area J40−H I / O PANEL
SG100 ION CHAMBER CABLE
XJ16−H Start Int. Start Int. J40−J
XJ16−E Relay RTN Relay RTN J40−E
XJ16−B 0VDC 0VDC J40−L
XJ16−P + HV + HV J40−R

Clamp shield shield isolated

COMPAX 40E TABLE

MIS 27224
17
J44−1 0VDC
SG K7 IN 2145656
 0VDC
SG K7 IN J45−1
J44−2 J45−2
I/O PANEL COMPATIBILITY CABLE Bucky HORIZ I/O PANEL
J44−10 Bucky HORIZ J45−3
shield shield
connector metallic shells connector metallic shells

52
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET SG60/ SG100 AUTO

J/P3−F Bucky VERT Bucky VERT J41−9


J/P3−H Bucky HORIZ Bucky HORIZ J41−10
CABLE # J/P3−P Bucky H N1 MIS 27223 Bucky H N1 J41−13
Bucky H N2 Bucky H N2 J41−12

J/P3−M
J/P3−R
J/P3−N
Bucky ROT
Cassette IN  2145644
Bucky ROT J41−11
Cassette IN J41−6
18 RAFL−S
I/F BOARD J/P3−E +21 VCD
−10VDC
+21 VCD J41−5
−10VDC
I/O PANEL
J/P3−D SG60/100 AUTO TRAY CABLE J41−2
TRANS TRANS J41−4
J/P3−C L/T Cass
J/P3−A L/T Cass J41−3
J/P3−B Long Cass Long Cass J41−7
J/P3−L −10VDC RTN −10VDC RTN J41−1
J/P3−K 0 VDC 0 VDC J41−8
connector metallic plate shield shield connector metallic shells

ROOM INTERLOCK SWITCH


ROOM WARNING LIGHT

 
Door Interlock RTN  OPTIONAL
19 XJ4−1
Door Interlock SUPPLIED IN THE FIELD
ROOM DOOR
INTERLOCK
XJ4−2
MPH A6 A4 MODULE SWICTH
(Wall Bucky Board) System ON (220Vac) WIRING IS FIELD SUP-
XJ4−4 PLIED CONNECT ACCORD-
System ON RTN (0Vac) SUPPLIED IN THE FIELD
XJ4−8 ING TO THE RULES IN
YOUR LOCALLITY

XT SUSPENSION

−10V ref coll1 −10V ref coll1


J201−9 J101−9
20 J201−10
SID Pot SID Pot J101−10
SID Pot Rtn SID Pot Rtn
J201−8 J101−8
Ref Rtn Coll1 Ref Rtn Coll1
J201−7 J101−7
SID A SID A J101−17
J201−17
J201−18 SID B SID B J101−18
J201−19 SID C SID C J101−19
J201−37 SID D SID D J101−37
SID E SID E J101−36
J201−36
J201−11 P21V P21V J101−11
Ordo Lock Rel. Ordo Lock Rel. J101−12
J201−12
Det Valid Det Valid J101−13
J201−13
XT OK for Tomo XT OK for Tomo
J201−16 J101−16
Vert Move Vert Move J101−35
J201−35
Vert Det (remote)
MIS 10705 Vert Det (remote)
J201−3 J101−3
J201−14
XT Park Intlk Com.
XT Park Intlk
 46−329341G2  XT Park Intlk Com.
XT Park Intlk
J101−14
J201−15 J101−15
Trans Mtr Hi Trans Mtr Hi
MPH A6 J201−21 J101−21 Control
SC2 Interface Trans Mtr Lo SID&COLLIMATOR
Trans Mtr Lo J101−20 Board 3A1
J201−20 CABLE
Long Mtr Hi Long Mtr Hi J101−2
J201−2
Long Mtr Lo Long Mtr Lo
J201−1 J101−1
−10VDC −10VDC
J201−25 J101−25
Trans Pot FdBk Trans Pot FdBk
J201−24 Pot FdBk Rtn J101−24
Pot FdBk Rtn
J201−23 J101−23
Long Pot Fbk Long Pot Fbk
J201−6 J101−6
Pot Fbk Rtn Pot Fbk Rtn
J201−5 J101−5
−10VDC Pot −10VDC Rtn
J201−4 J101−4
Undersizing Undersizing
J201−30 J101−30
PBL PBL J101−31
J201−31
Exp Hold Lmp Exp Hold Lmp
J201−33 J101−33
Non auto Lmp Non auto Lmp
J201−32 J101−32
SW ind pwr SW ind pwr
J201−34 J101−34
−80 to −100SW −80 to −100SW
J201−28 J101−28
−10−to +10 SW −10 to +10 SW J101−27
J201−27
J201−26 +80 to +100 SW +80 to +100 SW J101−26
Key SW Key SW J101−29
J201−29

connector metallic shell shield shield connector metallic shell

53
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET COMPAX 40E TABLE

CABLE # SID Det Rtn SID Det Rtn


J202−14 J23−14
SID Det SID Det J23−6
J202−6
Det Rtn Det Rtn J23−4
J202−4
21 J202−5 Det MIS 27231 Det J23−5
J202−9 SID out −
SID out +
2145655  SID out −
SID out +
J23−9
J202−1 J23−1
MPH A6 SC2
Diff Com Diff Com I/O PANEL
Interface J202−10 J23−10
FS−Floor Dist SID DETENT FS−Floor Dist
J202−2 CABLE J23−2
−10VDC Ref Rtn −10VDC Ref Rtn
J202−11 J23−11
Vert Move Vert Move
J202−7 J23−7
Vert Move Rtn Vert Move Rtn J23−15
J202−15
P21VDC P21VDC J23−13
J202−13
−10VDC Ref −10VDC Ref J23−3
J202−3
connector metallic shells shield shield connector metallic shells

J/P2−F Table Vert&Cass IN MIS 27230 Table Vert&Cass IN J41−9 not used
Cassette IN Cassette IN
22 J/P2−E J41−10
 
J/P2−P Table −15deg&Cass IN Table −15deg&Cass IN J41−13 not used
J/P2−K Table Horiz&Cass IN 2145618 Table Horiz&Cass IN J41−6
+21VDC isolated +21VDC isolated
J/P2−B J41−5
RAFL−S J/P2−A −10V −10V J41−2 I/O PANEL
TRANS TRANS
I/F BOARD J/P2−C DC
L/T Cass DC
L/T Cass
J41−4
J/P2−L J41−3
J/P2−H Long Cass COMPAX 40E Long Cass J41−7
J/P2−D −10VDC RTN AUTOMATIC TRAY CABLE −10VDC RTN J41−1
J/P2−R 0 VDC isolated 0 VDC isolated J41−8
connector metallic plate shield shield connector metallic shells

MPH 50/65/80 CABINET


INTERNAL
A format A format J/P10−V
XJ5−V
B format B format J/P10−W
XJ5−W
C format C format J/P10−X
XJ5−X
D format D format J/P10−Y
XJ5−Y
XJ5−HH E format  E format J/P10−HH
MIS 27232

common common
XJ5−MM J/P10−MM
Undersizing 2145659
MPH A6 A8 MODULE XJ5−J
Undersizing J/P10−J MPH A6 CL1
XJ5−R
Exd Sig Enable Exd Sig Enable J/P10−R RAFL−S
SAS INTERFACE CABLE
XJ5−A
Exp Sig Exp Sig J/P10−A I/O PANEL
Tube Sel Tube Sel J/P10−C
XJ5−C
Tech Dir Sel Tech Dir Sel J/P10−D
XJ5−D
Table Table
J/P10−U
XJ5−U Bucky
common Sel
Bucky
common Sel J/P10−H
XJ5−H Sel Sel
shield shield

clamp on I/F generator plate connector metallic shells

230Vac (isolated) MIS 27233 230Vac (isolated)


P1−3 TB−4
O Vac 2154537 O Vac
MPH A6 CL1 RAFL−S P1−1 TB−5 MPH A6 SC3
I/O PANEL GND POWER SUPPLY CABLE GND
GND TB−6

MIS 27240
230Vac (isolated) 230Vac (isolated)
P F13
MPH A6 SC3 2152750
O Vac O Vac MPH A1
FILTER N F12
GND UPPER CABINET POWER SUPPLY CABLE GND
GND GND Terminal

54
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

TOMOLINK COMPAX 40E TABLE


MANUAL
CABLE # MIS 11212
J106− 12
+21V XT
 2132121  +21V XT
XT OK/TOMO
J22−12
Control Electronics
25 TL1 A1
J106− 13 XT OK/TOMO
Tomo Lock Rel Tomo Lock Rel
J22−13
J22−25
I/O PANEL
J106− 25
XT−TOMO Cable connector support
connector support shield
shield

MPH 50/65/80 CABINET


AUTO
MIS 11212
Control Electronics
J106−12
+21V XT
XT OK/TOMO
 2132121−8  +21V XT
XT OK/TOMO
J200−12
MPH A6 SC2 Interface
25 TL1 A1
J106−13
J106−25 Tomo Lock Rel Tomo Lock Rel
J200−13
J200−25
XT−TOMO Cable
connector support shield connector support
shield

TOMOLINK

+24 CAN +24 CAN


J101−1 J101−1
26 J101−2 +24 CAN +24 CAN J101−2
J101−3 CAN H
MIS 11214 CAN H J101−3
J101−11 CAN L
 
CAN L J101−11
Control Electronics J101−4 Reset*H Reset*H J101−4
2132261 Drive Unit
TL1 A1 J101−12 Rest *L Rest *L J101−12 TL3 A1
Motion EN Motion EN
J101−5 DRIVE UNIT J101−5
+24 CAN RTN +24 CAN RTN
J101−9 Control Cable J101−9
+24 CAN RTN +24 CAN RTN
J101−10 J101−10
Spare Spare
J101−13 J101−13

connector support shield shield connector support

27 MIS 11215

100VAC HI 100VAC HI

Control Electronics
J100−1
100VAC Low
2124056  100VAC Low
J100−1 Drive Unit
TL3 A1
J100−2 J100−2
TL1 A1 Chassis GND Chassis GND
J100−3 DRIVE UNIT Power Supply Cable J100−3

MPH 50/65/80 CABINET

EXP ORD RTN EXP ORD Rtn


J105−11 J12−1
EXP ORD EXP ORD
J105−24 J12−2
28 J105−2 STK STRT RTN STK ST Rtn
J12−4
Tomo STK ST Tomo STK ST
J105−14 J12−6
Prep Sig Rtn Prep Sig Rtn J12−8
J105−6 Prep Sig MIS 11411 Prep Sig J12−9
J105−19
Sel Tomo Rtn Sel Tomo Rtn J14−1
J105−7
Control Electronics J105−20
Sel Tomo   Sel Tomo J14−2
MPH A6 A5 Table
TL1 A1 Tomo Sel 2159597 Tomo Sel J14−3
J105−N.C Tomo Board
Tomo Time1 Tomo Time1
J105−10 TOMO Control Cable J14−4
Tomo Time2 Tomo Time2
J105−23 J14−5
Tomo Time3 Tomo Time3
J105−21 J14−6
Tomo Time4 Tomo Time4
J105−9 J14−7
Tomo Time5
J105−22 Tomo Time5 J14−8
J105−8 Tomo Tomo J14−9
Time/ Time/
connector support Sel
shield Sel
shield clamp on connector plate
CMM CMM

55
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

TOMOLINK TOMOLINK

+24 CAN
J102−1
J102−2 +24 CAN +24 CAN A
J102−3 CAN H MIS Y +24 CAN J
J102−11 CAN L CAN H C
INTERNAL J102−4 Reset*H REA 00074
CAN L D
Control Electronics J102−12 Rest *L Reset*H
E To Tomo−Tower Control TL2
TL1 A1 Motion EN Rest *L via Quick Disconnect connector
J102−5 F
+24 CAN RTN TOMO−TOWER Control Cable Motion EN G
J102−9
+24 CAN RTN +24 CAN RTN
J102−10 B
Spare +24 CAN RTN
J102−13
Spare K
connector support shield H
L
shield

+24 CAN
J103−1
J103−2 +24 CAN +24 CAN J2−A
J103−3 CAN H +24 CAN J2−J
MIS Z
J103−11 CAN L CAN H J2−C
INTERNAL J103−4 Reset*H CAN L J2−D
REA 00072
Control Electronics J103−12 Rest *L Reset*H J2−E
TL1 A1 Motion EN Rest *L J2−F TomoLink Console
J103−5
+24 CAN RTN TOMO−CONSOLE Control Cable Motion EN
J103−9 J2−G
+24 CAN RTN +24 CAN RTN J2−B
J103−10
Spare +24 CAN RTN J2−K
J103−13
Spare J2−H
connector support shield
Clamp
shield

COMPAX 40E TABLE

MIS bb
INTERNAL J107−1 Bucky Brake Release Bucky Rake Release
J8−3
J107−2 Bucky REA 00075
Bucky J8−4
C40E Power Board
Control Electronics Brake Rake
TL1 A1 Release Release
RTN RTN

MIS aa 230VAC
230VAC
J110−L FL1−P
INTERNAL J110−N O VAC
REA 00073 O VAC
C40E Filter
FL1−N
Control Electronics Ground Tomo−Link Power Supply Cable Ground
TL1 A1 GND GND

FIXED CABLE J1−1


+24 CAN

J1−2 +24 CAN +24 CAN A


J1−3 CAN H MIS Y +24 CAN J
J1−11 CAN L CAN H C
J1−4 Reset*H REA 00074
CAN L D
J1−12 Rest *L Reset*H
E
Motion EN Rest *L Quick
J1−5 F
+24 CAN RTN Fixed Tower Cable Motion EN G Disconnect
J1−9
+24 CAN RTN +24 CAN RTN Connector
J1−10 B
Spare +24 CAN RTN
J1−13
Spare K
connector support shield H
L
shield

56
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

MPH 50/65/80 CABINET MAIN POWER

CABLE # MIS 27235


2138905 
1.1 LINE SAFETY SWITCH
POWER SUPPLY CABLE
MPH A1 A1 MODULE
(AC Distribution Panel) TS−GND GROUND

(PU3) MADRID CONSOLE


2.1 (45553701)
MPH A1 MODULE
 45553701 GND (BAR GROUND)
Ground Plug
(AC Distribution Panel)
CONSOLE GROUND CABLE

COMPAX 40E TABLE

MIS 27300
3.1
MPH A1 MODULE  2140362
TB1 (Ground) I / O PANEL
(AC Distribution Panel) Ground Plug
TABLE Ground CABLE

XT SUSPENSION

4.1 MIS 10708


MPH A1 MODULE  (46−316922G2) SUPPLIED with XT SUSPENSION 
(AC Distribution Panel) Ground Plug GND STUD
BRIDGE
GROUND CABLE

SG 60/100 VERTICAL BUCKY


STAND
MIS 27301
5.1  2140362
Ground Plug TS2−GND
MPH A1 MODULE I / O PANEL
(AC Distribution Panel) SG 60/100 GROUND CABLE

PRINTER−X
W19
6.1
MPH A1 MODULE
(AC Distribution Panel) Ground Plug
 (2105836) SUPPLIED with PRINTER−X
GND STUD MPH A9
PRINTER−X Ground CABLE MODULE

COMPAX 40E TABLE COMPAX 40E TABLE

(45433617)
7.1 45433617
TLK Electronics GND TB1 (Ground) I / O PANEL
TOMO LINK Ground CABLE

MAXIRAY100 XRAY TUBE XT SUSPENSION

8.1 (46−216211G10)
GND STUD  46−216211G10 
Ground Stud GND STUD BRIDGE
GROUND CABLE

57
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


THIS PAGE INTENTIONALLY LEFT BLANK.

58
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SYSTEM COMPAX 40E Job Card IST 001 1 of 4

Purpose: X−RAY TUBE & COLLIMATOR Version No.: 2


INSTALLATION ON XT SUSPENSION Date: November 10, 2006

Time: Personnel: 2

SECTION 1
SPARE PARTS REQUIRED

SECTION 2
EQUIPMENT REQUIRED

SECTION 3
SAFETY PRECAUTIONS

SECTION 4
PRELIMINARY STEPS

Always use extreme caution when removing XT Column shipping rod. The
counterpoise should be wound for a maxiray 100 when you receive it.. If
this is not the case, a quick check can be made by pulling on the cable
from the counterpoise. If it is taut, it is probably wound.
Also note that the vertical lock is on when the power is off. The vertical
lock will support approximately 30 lbs. This does not mean that the
counterpoise is set correctly.
See Chapter 3 or 4 of the SM XT RADIOGRAPHIC SUSPENSION, Dir.
2118733−100.

SECTION 5
INSTALLATION PROCEDURES

Note:
− During HLA the X−Ray tube unit is aligned with the Collimator Interface Plate.
In the field it is only necessary to mount the X−ray Tube assembly on the XT Tube Support arm
and install the collimator following instructions in Job Card IST001.
− The Collimator Interface Plate is adjusted parallel to the X−Ray tube axis so that the middle of the
two focal spots is centered on Collimator Alignment Tool crosshair.

59
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 2 of 4

ILLUSTRATION 1
Fixing Screws

Interface Plate of Collimator


X−Ray Tube (ULTRANET SA & SM)

Note: The Collimator Interface Plate is adjusted in factory (HLA)

60
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 3 of 4

5.1 X−Ray Tube Installation

Note: Spacers are only required for MX75 X−ray tubes.


1. Mount the MX100 X−ray tube with cable mounting bracket on the XT Suspension support and
position it to align with the holes located on the front side of XT Suspension support. Insert
screws to hold the X−ray tube.
2. Level the tube so that the output window is horizontal.
3. Tighten the four screws (M6x35) to attach the X−ray tube and adapter to the XT Suspension
support (screws are furnished in the XT box).

5.2 Cable Connections (MIS 10680) and Conduit Routing

THE COMMON LEAD ON THE MPH STATOR CABLE IS NOT CONNECTED


TO GROUND CABLE.
CONNECT GROUND CABLE PROVIDED ON XT TO X−RAY TUBE HOUSING.

1. Secure stator cable shield to fan mounting bracket using the supplied clamp, bolt and nut.
Connect stator leads to terminal strip as indicated in the MIS MAP and MIS CHART of this
Manual.
To route the stator cable correctly, see illustration 13−8 of XT Suspension SM, dir.
2118910−100, Chapter 3, Section 9−6 or Chapter 4, Section 13−4.
(Position 1 of terminal strip on X−ray tube is on anode side).
5.3 Collimator Installation

1. Rotate the X−Ray Tube support so bottom side of tube unit faces up.
2. Fasten the Collimator on the X−Ray tube unit with four 1/4 x 1 1/16 in screws.
For ULTRANET Collimator use supplied spanner wrench to tighten the Collimator coupling
ring.

LOCTITE THREADLOCKER MUST BE APPLIED DURING INSTALLATION TO


PREVENT THE SET SCREWS FROM BACKING OUT OVER TIME.

INSTALL AND TIGHTEN THE THREE SET SCREWS (DOWELS) INTO


COLLIMATOR LOCKING RING. YOU MUST FOLLOW THE INSTALLATION
PROCEDURE OF COLLIMATOR SERVICE MANUAL CHAPTER 3 TO
ENSURE THE COLLIMATOR IS SAFELY SECURED ON THE INTERFACE
PLATE.

61
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 4 of 4

3. Rotate the Tube support arm so the collimator is facing down, then remove the Collimator
rear cover and connect to J/P2 and J/P1 plugs, Collimator cable Refer to MIS MAP and MIS
CHART of this Manual).
Note: The J/P1 plug is used only with Ultranet Sa Collimator (Auto) when JP2 plug is used with both
SA and SM Collimator.
4. Re−install the Collimator rear cover and reconnect small ground wire if removed in Step 3.

62
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SYSTEM COMPAX 40E Job Card IST 001 1 of 4

Purpose: X−RAY TUBE & COLLIMATOR Version No.: 2


INSTALLATION ON XT SUSPENSION Date: August 7, 1997

Time: Personnel: 1

SECTION 1
SPARES REQUIRED

SECTION 2
EQUIPMENT REQUIRED

SECTION 3
SAFETY PRECAUTIONS

SECTION 4
PRELIMINARY STEPS

WITCH THE POWER OFF ON XT SUSPENSION BEFORE CHANGING TAP


SELECTION ON XT CONTROL BOARD.

SECTION 5
INSTALLATION PROCEDURES

5.1 Configuration During HLA

For XT Suspension installation in Europe the Collimator lamp voltage is configured by connecting
an electrical jumper to J10 (24VAC) for the Philips lamp or J11 (26VAC) for the GE type lamp, when
a 230VAC input line is on XT transformer (refer to System Inspection Report and Illustration 1).
Note: An XT Suspension integrated in the US is furnished with jumper connected to J9.

63
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 2 of 4

ILLUSTRATION 2
24VAC Test Point (J10)
26VAC Test Point (J11)
Common Test Point

XT Control Board
(2146642 B)

64
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 3 of 4

5.2 Collimator Lamp voltage Selection Procedure

Before starting this procedure ensure that the XT control board and the Collimator lamp are
accessible (refer to Collimator SM for instructions).
1. Select on digital multimeter:
− VAC function
− Auto range (or manual range > 240 VAC).
2. Measure the actual voltage at Collimator lamp bulb base with the lamp switched ON.
3. With The GE type lamp, ensure that voltage is greater than 23V when the electrical jumper
J11 is in place.
4. With The Philips type lamp, if the voltage is within 21.5−VAC 24VAC range be sure that one
of the electrical jumper position next to J9 or J10 cannot be used to set the voltage closer to
22.5VAC (always above 21. VAC), this depends on the actual room line voltage available on
your system.
If so, then move jumper from J10 to J9. (Refer to table below for tap selection voltages).

Terminal blade Terminal blade


Field Lamp TAP select
straight straight

J7 J8 20VAC TAP

J7 J9 22VAC TAP

J7 J10 24VAC TAP

Selected in HLA
J7 J11 26VAC TAP

5. If the voltage is out of the range 21.5 − 24 VAC, check the input line on the XT control board
(between J100−1 & J100−5). If it is not 230 VAC +/− 12.5%, redo IST 02 of SM MPH 50/65/80
Dir. 2137354−100 and correct line voltage setting on generator.
If the input line is correct, move the electrical jumper to a more appropriate position in order to
keep the voltage within 21.5 and 24 VAC and idealy close to 22.5 VAC for the Philiips lamp and
greater than 23 VAC for the GE lamp.
6. Record actual final voltage reading at Collimator lamp bulb base only if you have to do HHS
measurement on your system.

Actual Voltage VAC

65
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 4 of 4

THIS PAGE INTENTIONALLY LEFT BLANK.

66
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SYSTEM COMPAX 40E Job Card IST 001 1 of 8

Purpose: X−RAY TUBE & COLLIMATOR Version No.: 3


INSTALLATION ON XT SUSPENSION Date: November 10, 2006

Time: Personnel: 1

SECTION 1
SPARES REQUIRED

SECTION 2
EQUIPMENT REQUIRED

SECTION 3
SAFETY PRECAUTIONS

Do not remove the shipping anchor rod from the Vertical


carriage until instructed to do so See step A8 in the Steer-
ing Guide. (fig. 1)

SECTION 4
PRELIMINARY STEPS

Slightly loosen the Tube Stand Arm anchor bolts to insert


the Tube Stand Arm into the brackets.

SECTION 5
INSTALLATION PROCEDURES

WARNING
Note: During the HLA procedure in factory:
− The Tubestand arm is installed, levelled and DO NOT REMOVE THE ROD
adjusted.
− The X−Ray tube is aligned with the Collimator shown without
Interface Plate left which is installed. cabling
− In the field it is only necessary to mount the
Tube Assy on the Tube Support Arm and
install the collimator with the following
instructions.

Note: The Collimator Interface Plate is adjusted


parallel to the X−Ray tube axis so that the
middle of the two focal spots is centered on
the Collimator Alignment Tool crosshair. Fig.1

67
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION Job Card IST 001 2 of 8

ILLUSTRATION 3

FIXING SCREWS

COLLIMATOR INTERFACE PLATE


X−RAY TUBE (ULTRANET SA & SM)

REMARK: THE COLLIMATOR INTERFACE PLATE IS ADJUSTED IN FACTORY (HLA)

68
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION Job Card IST 001 3 of 8

5.1 X−Ray Tube Installation

The compensating support spacer (item 160, illustration 4−3 in RS 85 Mounting instructions,
drawing 0327 7223) must be mounted between the X−ray tube housing and the RS 85 tube arm.
1.5 mm thick plates (item 161, illustration 4−3 in RS 85 Mounting instructions) are supplied to
allow precise alignment between the X−ray tube transversal axis and the table.
One or more plates can be mounted between the spacer and the tube arm plate (Illustration 4),
between the spacer and the tube housing (Illustration 5), or both.
The tube housing is secured on the spacer with four 1/4”−20 screws, a flat washer and a lock
washer. These items are contained in bag 2192415. Screws of two lengths (25 and 30 mm) are
provided depending on the quantity of 1.5 mm plates used between the spacer and the tube
housing.
Screws must be chosen according to Table 1:

TABLE 1
CHOICE OF SCREWS TO MOUNT TUBE HOUSING ON SPACER
Number of plates 1/4” x 20 1/4” x 20
between spacer and tube 25 mm screws 30 mm screws
housing
0 x
1 ( thick. = 1.5mm) x
2 ( thick. = 3.0mm) x
3 ( thick. = 4.5mm) x

Verify that:
 when the screws are fully tightened, they do not reach the end of the tube housing hole (no
play),
 the length of the screws engaged in the tube housing is at least 8mm.

69
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 4 of 8

ILLUSTRATION 4
TUBE ASSEMBLY OF RS85 ARM WITH SPACER (1.5 MM PLATES ON ARM SIDE)

ILLUSTRATION 5
TUBE ASSEMBLY OF RS85 ARM WITH SPACER (1.5 MM PLATES ON TUBE HOUSING SIDE)

70
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 5 of 8

5.2 Cable Connections (MIS 27239) and Conduit Routing

THE COMMON LEAD ON THE MPH STATOR CABLE IS NOT GROUNDED.


CONNECT GROUND CABLE PROVIDED ON RS 85 TO X−RAY TUBE
HOUSING.

1. Secure stator cable shield to fan mounting bracket using the supplied clamp, bolt and nut.
Connect stator leads to the terminal strip as indicated in the MIS map and MIS chart MIS
27239 of this Manual.
To route the stator cable correctly, refer to RS 85 Mounting Instructions, Dir 03277223,
Chapter 1, section 1−4−3.
(Position 1 of terminal strip on X−ray Tube is on anode side).
5.3 Collimator Installation

1. Rotate the X−ray Tube Support so the bottom side of the tube unit is facing up.

2. Fasten the Collimator over to the X−Ray tube.

LOTITE THREADLOCKER MUST BE APPLIED DURING INSTALLATION TO


PREVENT THE SET SCREWS FROM BACKING OUT OVER TIME.
WARNINGS ARE ALL BOLD CAPS.

INSTALL AND TIGHTEN THE THREE SET SCREWS (DOWELS) INTO


COLLIMATOR LOCKING RING.

71
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 6 of 8

3. Rotate the Tube Support Arm so the collimator is facing down, then remove the collimator
rear cover and connect to J/P1 and J/P2 plugs and collimator cables (27303 and 27236) (See
MIS MAP / MIS CHART). Route the cables according to the RS−85 Mounting Instructions, dir.
0327 7223 (Chapter 1, section 1.7). Secure collimator cables shield with strain relief plate.
Note: The J/P1 plug is used only on the SM Collimator with Ultranet SA collimator (Auto).
The J/P2 is used for both types of Collimator (SA and SM).
4. Re−install the Collimator rear cover and reconnect small ground wire if removed in Step 3.

5.4 Collimator Lamp Voltage Checks

5.4.1 For Automatic Systems with GE Type Lamp:

1. Check that lead J/P6 H is on terminal strip 29V of RAFL−S transformer TR1 (see Ultranet and
RAFL−S Service Manual Y00W20L.04, Back Panel RAFL−S Schematic 45296256).
2. Measure the voltage at the bulb base. This voltage must be greater than 23 VAC.

5.4.2 For Automatic Systems with Philips Type Lamp:

1. Check that lead J/P6 H is on terminal strip 28V of RAFL−S transformer TR1 (see Ultranet and
RAFL−S Service Manual Y00W20L.04, Back Panel RAFL−S Schematic 45296256).
2. Measure the voltage at the bulb base. This voltage must be between 21.5 and 23 VAC.

5.4.3 For Manual Systems:

1. Measure voltage at the bulb base; this voltage must be between 21.5 V and 23.0 V for Philips
type lamp and greater than 23V for GE type lamp.
2. If not, check voltage configuration on Table secondary fuse panel (2144512).
When the Compax 40E Table is fed with 230 VAC, the lead J11−2 (secondary fuse panel)
must be connected to terminal strip 17d for GE lamp or to 17b for Philips type lamp, on
transformer T1 (see Compax 40E Schematics SM 2160704−100, schematic 2149008).

72
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 7 of 8

5.5 RS 85 Cable Drape Mount

Two cable drapes are provided with the RS 85 tube stand to cover the high voltage and system
cables going to the tube housing and collimator.
One drape is 2.2 meters long and is mounted between the top of the column and the tube arm.
The other one is initially 4 meters long and is mounted between the column top and the wall
fixation.
The cable drapes must be installed after all cables are connected to the tube and column, and the
harnesses are made.
1. Remove the column and wall box top covers.
2. Install the 2.2 meter cable drape between the column top and the tube arm:
Wrap the cables with the appropriate drape.
At the top of the column, secure the drape on the cable with a ty−rap as close as possible to the channel
entrance as shown in Illustration 6.
On the tube housing end, wrap the drape as far as possible along the cables. Allow the cable going to the
connection point in the arm to exit the drape through the Velcro.
Secure the cables and the drape on the tube arm with two ty−raps as shown in Illustration 6.
3. Install the 4 meter cable drape between the column top and the wall fixation point:
Note: The initial length of the drape for this section is enough to accommodate an installation with
the rail installed far from the wall.
Allowing about 10 cm inside the cable channel, cut the drape to the appropriate length.
Clamp the cables with the drape inside the wall box.
At the top of the column, secure the drape on the cable with a ty−rap as as close as possible to the
channel entrance as shown in Illustration 6.
4. Reinstall the column cover. Make sure the ty−raps are hidden inside the cover (Illustration 6).
5. Reinstall the wall box cover.

73
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 8 of 8

ILLUSTRATION 6
RS85 CABLE DRAPE MOUNTING

74
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SYSTEM COMPAX 40E Job Card IST 001 1 of 4

Purpose: X−RAY TUBE & COLLIMATOR Version No.: 1


INSTALLATION ON XT SUSPENSION Date: February 10, 1997

Time: Personnel: 1

SECTION 1
SPARES REQUIRED

SECTION 2
EQUIPMENT REQUIRED

SECTION 3
SAFETY PRECAUTIONS

SECTION 4
PRELIMINARY STEPS

SECTION 5
INSTALLATION PROCEDURES

Room light, X−Ray light and Wall Bucky are generally powered with:
. 220 VAC in EUROPE
. 110 VAC in USA
Refer to this Jobcard section 5.2 to select the proper setting.

For DOOR OPEN SWITCH and EXPOSURE ON SIGNAL LAMP connection, refer to MPH A6 A4
Wall Bucky Board schematic, Illustration 7 of this Job Card.

75
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 2 of 4

ILLUSTRATION 7
MPH A6 A4: WALL BUCKY BOARD

INSTALL JUMPER IF
NO DOOR INTERLOCK

RAD

CAUTION

See MPH
Service
Manual

76
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 3 of 4

ILLUSTRATION 8
MPH CABINET

MPH A6 A3 MPH A6 A4
Distribution Board Wall Bucky Board

5.1 Connection of Door Open safety switches (see schematic).

With this feature, X−Ray generation is inhibited while the door of the X−Ray room remains open.
Connect the safety switch to J4 pin 1 & 2 (or a jumper if there is no safety switch) if Wall Bucky
Board.
5.2 Connection of Room Lamp, X−Ray Exposure on Signal Lamp and Bucky.

The generator is configured in the factory to supply Buckies with 220VAC. If the Buckies must be
supplied in 115VAC, change the fuse F9 of the Room Distribution board MPH A6 A3 from position
F9A to position F9B. All the Buckies are now supplied on 110VAC ( it is not possible to mix the
voltages). The room lamps are also supplied in 110VAC.

77
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 4 of 4

THIS PAGE INTENTIONALLY LEFT BLANK.

78
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SYSTEM COMPAX 40E Job Card IST 001 1 of 4

Purpose: X−RAY TUBE & COLLIMATOR Version No.: 2


INSTALLATION ON XT SUSPENSION Date: July 6, 1997

Time: Personnel: 1

SECTION 1
SPARES REQUIRED

SECTION 2
EQUIPMENT REQUIRED

SECTION 3
SAFETY PRECAUTIONS

SECTION 4
PRELIMINARY STEPS

SECTION 5
INSTALLATION PROCEDURES

 Recommendation for EMC During Installation, any disassembly/reassembly or


Preventive (Electronic Magnetic Compliance) Maintenance, ensure that each System
Cable shield is properly secure in it’s clamp.

− Refer to Illustration 9 of this Job Card for clamps location in MPH Cabinet. Use EMC
clamps provided in bag 2206408.

− For other sub−systems, cable shield must be clamped according to MIS CHART marks
(see MIS CHART Chapter 2 of this manual).

− Ensure that each metal plug is properly secured by its mounting screws (EMC, ground
continuity).
 RAFL−S Rack Accessibility To better access the back of the RAFL rack or read the
keyboard display, it is possible to slide out the complete rack after the fixing screws are
removed. Be sure you tilt down the Extension Rack MPHA6 before and that the RAFL Power
Board is well secured. Refer also to Job Card IST002 of MPH Service Manual.

79
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 2 of 4

 System Cables Storing STANDARD CONFIGURATION (MPH)

During Installation, coil cables extra length onto the cable hanger in the upper left MPH
Cabinet (Refer to Illustration 9).
ILLUSTRATION 9
CABLE STORAGE

Separator panel

Cable Hanger

Cable Shield
Clamps area

Reserved space
to coil cable
extra length

80
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

CHAPTER 3 − DISASSEMBLY/REASSEMBLY

SECTION 1
SPARE PARTS REQUIRED

None

SECTION 2
EQUIPMENT REQUIRED

None

SECTION 3
SAFETY PRECAUTIONS

None

SECTION 4
PRELIMINARY STEPS

None

SECTION 5
PROCEDURES

Follow DR Job Card instructions carefully and read the procedure through
before starting work.

The following Chapter defines the disassembly and reassembly procedures for the Compax 40E System.

81
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


THIS PAGE INTENTIONALLY LEFT BLANK.

82
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

SYSTEM COMPAX 40E Job Card IST 001 1 of 4

Purpose: X−RAY TUBE & COLLIMATOR Version No.: 1


INSTALLATION ON XT SUSPENSION Date: February 10, 1997

Time: Personnel: 1

SECTION 1
SPARES REQUIRED

MX100 Tube

SECTION 2
EQUIPMENT REQUIRED

Protective pad, metal rod.

SECTION 3
SAFETY PRECAUTIONS

None

SECTION 4
PRELIMINARY STEPS

None

SECTION 5
INSTALLATION PROCEDURES

The following procedure describes the steps for replacing an MX100 Tube.

NEVER REMOVE THE ROD DURING X−RAY TUBE (OR COLLIMATOR)


REPLACEMENT UNTIL STEP 8. OF THIS PROCEDURE

Follow these instructions carefully and read the procedure through before
starting work.

83
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

X−RAY TUBE & COLLIMATOR


INSTALLATION ON XT SUSPENSION Job Card IST 001 2 of 4

1. Place a protective pad on the foot end of the Table top.


2. Insert the metal rod (refer to Illustration 10) onto the Column to immobilize the Tube Arm
vertically.
3. Immobilize the RS 85 Column carriage so it cannot move longitudinally.
4. Rotate the Tube Assy 180° to ”Tube Output Up” to remove the Collimator.
5. Remove the Collimator and set it aside on the pad. Cable disconnect should not be
necessary.
6. Remove MX100 Tube and replace it with the new tube.
7. Re−install the Collimator onto the X−ray Tube assy.
8. Remove the rod from the RS 85 Column to free the Tube Arm.

ILLUSTRATION 10
METAL ROD ON VERTICAL CARRIAGE
WARNING

METAL ROD

Fig.1

84
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

CHAPTER 4 − FUNCTIONAL CHECKS

SECTION 1
PRELIMINARIES

The functional tests are divided into two parts:


1. The operational tests are intended to ensure that all of the interconnected system components are
operating properly (for different techniques).
2. The performance tests are intended to check that the system’s components conform to their
operating specifications.

READ SAFETY INSTRUCTIONS ATTENTIVELY PRIOR TO STARTING ANY


FUNCTIONAL TESTS.

SECTION 2
OPERATIONAL TESTS

Switch system ON from the MPH Console.

2-1 Table Operational Tests

Test all movements, functionalities and limit positions of the Table by referring to the Operator Manual
2150277, Chapter 4:
Emergency Stop Section 4.1
Patient Support Sections 4.2 & 4.3 (including anticollision device)
SID Display Section 4.4
Positioning the Bucky Section 4.5
Cassette Tray Section 4.7
Tomography Section 4.6
Grid Oscillation −

Note: Operate cafefully by testing safety and limitation devices simultaneously.

2-2 Technique Tests

From the MPH Console, select the various techniques in turn and check for the correct functionning of
each of them by referring to :
− Operator Manual, Chapter 3 for operating instructions,
− MPH Operator Manual for setting parameters or any other selection of data as required (as for the
Automatic Exposure device).

85
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

2-3 SG 60 Operational Tests

Test all movements, functionalities and limit positions. Refer to Operator Manual 2150272, Chapter 3:
Vertical Positioning page 3−1
Cassette Tray page 3−3
Grid Oscillation page 3−10 (during preparation)

2-4 SG 100 Operational Tests

Test all movements, functionalities and limit positions. Refer to Operator Manual 2150275, Chapter 3:
Vertical Positioning page 3−1
Patient Support Legs page 3−2
Bucky Rotation page 3−3
Bucky Angulation page 3−4
Cassette Tray page 3−7
Grid Oscillation page 3−5 (during Preparation)

2-5 XT Suspension Operational Tests

Test all movements, functionalities and limit positions. Refer to Operator Manual 2116790−100:
Overhead Rail System Section 2
Telescopic Column and Carriage Section 2
Longitudinal and Transverse Detents Section 2
Lock Operation Section 2

2-6 RS 85 Tubestand

Test all movements, functionalities and limit positions. Refer to Operator Manuals 03277324, 03277341,
03277361, 03277303, 03277351 and 03277371.

2-7 Ultranet SM Collimator

Test all movements, functionalities and limit positions. Refer to Operator Manual 2144901:
Continuous Rotation of the Collimator Chapter 3 page 3−6
Rotation of the Tube Unit/Collimator Assy Chapter 3 page 3−6
Knobs, Buttons, Sensitive Handle Chapter 4 page 4−13
Light Beam Centering Device Chapter 4 page 4−16
Tape Measure

2-8 Ultranet SA Collimator

Test all movements, functionalities and limit positions. Refer to Operator Manual 2144901:
Continuous Rotation of the Collimator Chapter 3 page 3−6

86
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


Rotation of the Tube Unit/Collimator Assy Chapter 3 page 3−6
Knobs, Buttons, Sensitive Handle Chapter 4 page 4−13
Light Beam Centering Device Chapter 4 page 4−16
Tape Measure

SECTION 3
PERFORMANCE TESTS

MPH See SM 2137354−100, Job Card VF002


Tomo−Link See Dir 2133821−100, Chapter 18−10
IQ (including AEC) See Tomo−Link, Chapter 18−10
See SG 60 SM 2150293−100; VF01
See SG 100 SM 2150297−100; VF01

87
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


THIS PAGE INTENTIONALLY LEFT BLANK.

88
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

CHAPTER 5 − SYSTEM CABINET SPARE PARTS

89
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


ILLUSTRATION
SYSTEM CABINET UPPER PART (2146714)

A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5

− 2175225 2  SYSTEM CABINET FRONT DOOR ASSY

− 2153822 2   255 X 900 X 10 MM ALU AIR FILTER 1

− 2153823 2   DOOR LOCK 1

− N  RAD OPTION

1 2152751 2   MAINS EMI FILTER 2P A6 250 V 1

2 2150568 2   INTERMEDIATE PANEL FOR RAFL−S OPTION 1−2

3 2150565 1   RAFL−S RACK CE MARKED (OPTIONAL) 1−2

4 2179445 1   LOW VOLTAGE POWER SUPPLY FOR RS 85 TUBE STAND OPTION 1

2206408 2   EMC TIE−WRAPS (BAG) 1

−  SYSTEM CABLES

− 2145602−2 2 N   18M 40E SYSTEM BUCKY CABLE 1

− 2167144−2 2 N   18M 40E SYSTEM POWER CABLE 1

− 2140362−2 2 N   18M 40E SYSTEM GROUND CABLE 1

− 2145651−2 2 N   18M 40E SYSTEM AEC−MPH CABLE 1

− 2145618−2 2 N   18M 40E SYSTEM CASSETTE TRAY CABLE 1

− 2145655−2 2 N   18M 40E SYSTEM SID CABLE 1

− 2128497−2 2 N   18M STATOR CABLE 1

− 2145602−3 2 N   24M 40E SYSTEM BUCKY CABLE 2

− 2167144−3 2 N   24M 40E SYSTEM POWER CABLE 2

− 2140362−3 2 N   24M 40E SYSTEM GROUND CABLE 2

− 2145651−3 2 N   24M 40E SYSTEM AEC−MPH CABLE 2

− 2145618−3 2 N   24M 40E SYSTEM CASSETTE TRAY CABLE 2

− 2145655−3 2 N   24M 40E SYSTEM SID CABLE 2

− 2128497−3 2 N   24M STATOR CABLE 2

Note: Other spare parts are sub−systems. Spare parts are found in
sub−system service manual (Chapter Spare Parts).

90
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


ILLUSTRATION
SYSTEM CABINET UPPER PART (2143779)

RAFL−S space

91
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


ILLUSTRATION
SYSTEM CABINET UPPER PART (2146714)

A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5

− 2145641−2 2 N   18M SG SYSTEM BUCKY CABLE 1

− 2140362−2 2 N   18M SG SYSTEM GROUND CABLE 1

− 2145644−2 2 N   18M SG SYSTEM CASSETTE TRAY CABLE 1

− 2145656−2 2 N   18M SG SYSTEM COMPATIBILITY CABLE 1

− 2145642−2 2 N   18M SG SYSTEM AEC−MPH CABLE 1

− 2151729−2 2 N   18M SG SYSTEM POWER CABLE 1

− 2145641−3 2 N   24M SG SYSTEM BUCKY CABLE 2

− 2140362−3 2 N   24M SG SYSTEM GROUND CABLE 2

− 2145644−3 2 N   24M SG SYSTEM CASSETTE TRAY CABLE 2

− 2145656−3 2 N   24M SG SYSTEM COMPATIBILITY CABLE 2

− 2145642−3 2 N   24M SG SYSTEM AEC−MPH CABLE 2

− 2151729−3 2 N   24M SG SYSTEM POWER CABLE 2

Note: Other spare parts are sub−systems. Spare parts are found in
sub−system service manual (Chapter Spare Parts).

92
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


ILLUSTRATION
SYSTEM CABINET UPPER PART (2143779)

RAFL−S space

93
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Blank page.

94
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

CHAPTER 6 − PREVENTIVE MAINTENANCE

SECTION 1
PERIODICITY

1-1 Table

1 visit per year

1-2 MPH Generator

1 visit per year

1-3 XT

visit per year

1-4 RS 85

1 visit per year

1-5 SG60

1 visit per year

1-6 SG100

1 visit per year

1-7 Tomo−Link

Periodically

1-8 Collimator

1 visit per year


Total Time PM at System Level: 8 Hours

95
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


SECTION 2
OPERATIONS

Note: At the end of each operation of periodic maintenance, the complete ”functional tests” procedure must be ap-
plied (refer to Chapter 4 of this manual).

2-1 Table

See SM ref.: 2150299−100, Chapter 6

2-2 MPH Generator

See SM MPH ref.: 2137354−100, Chapter 4.

2-3 XT

See SM 2118733−100, Dir. 2116793−100

2-4 RS 85

See SM 03277223, Chapter 4

2-5 SG60

See SM 2150293−100, Chapter 5

2-6 SG100

See SM 2150297−100, Chapter 5

2-7 Tomo−Link

See SM 2133821−100, Chapter 19

2-8 Collimator

The following PM procedure applies to Ultranet “SM” Collimators. You can also reference the following
directives for this procedure.
− SM Y00W21L.04, Chapter 10
− SM Y00W20L.04, Chapter 10
2-8-1 Equipment Required for Collimator PM

− Standard tools
− LUBRIPLATE 630−AA Grease (or equivalent)
− Set of wrenches (furnished with the collimator and left on site)
> Hook wrench for collimator mounting ring (Part No. 45298477)
> Hex adapter (Part No. 2223142)
> Two wrenches for 10/32” and 9/16” screws (Part No. 45298477)
> Two hook wrenches for 25,, castle nuts (Part No. 45298476)
> One knob position tool (Part No. 2188963)
− Set of wrenches (not furnished)
> Torque wrench (for torquing the mounting ring): 3/8” square drive torque of 20 to 30 Nm (180 to
270 in lbs.)

96
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


> Torque screwdriver (for torquing the set screws): torque of 1.4 Nm (12 in lbs.)
2-8-2 Removing the Covers from the Collimator

The following identifies the collimator covers and the steps required for disassembling.
− Main cover
a. Remove the rear lid
b. Disconnect the ground cable
c. Remove the six securing screws and remove the main cover
− Internal cover
a. Remove the five screws and remove the internal cover
− Front panel
a. Remove the knobs
b. Loosen the two screws and open the front panel
c. Disconnect the ground cable
d. Disconnect connectors PR1A and PR2
e. Remove the front panel

2-8-3 Lubrication

Apply a light coating of Lubriplate 630−AA or equivalent to the following parts:


− Worm screws, which are coupled to motor M2 and M1
− Clutch − Do not loosen the clutch; apply grease to its outer surface and to the clutch for motor M1
and M2.
− Blade pivot points − Lubricate all pivot points (14 points total)
2-8-4 Collimator General Inspection

Note: Take extra care when cleaning the window and mirror, as the surfaces can easily be scratched.

1. Light field quality is affected by dust build−up or smudges on the window or mirror. Check for grime
on the inner and outer surfaces of the window. Clean with X−ray screen cleaner.
2. Carefully clean the mirror with a soft cloth moistened in a mild soap and water solution. Wipe dry
using another soft, dry cloth. (A commercial glass cleaner can also be used to clean the mirror.)
2-8-5 Check Tightness of Collimator Mounting Ring (every 12 months)

If the collimator has play, you must remount it. Use the following steps to remount the collimator.
1. Rotate tube/collimator assembly to bring the collimator bottom up
2. Disconnect the cables at the back of the collimator
3. Remove the blocking screws (dowels) and unscrew the mounting ring
4. Remove the collimator from the interface plate.
5. Put it on the table top, taking care to not scratch the top surface.
6. Reassemble all parts following the installation procedure in “Section 3” of this chapter.
Pay particular attention to foreign particles or chips that could prevent the ring or dowels being correctly
tightened.

97
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


2-8-6 If the collimator has no play:

1. Locate the wrench or extension tools provided with all collimators.


2. Check that it is not possible to slide the queue of one of these tools into the space between the ring
top surface and the bottom of the interface plate.
3. If the tool can be inserted, it means that the ring is not fully tighten. In this case, perform the steps in
2−8−5.

ANY TIME YOU SCREW OR UNSCREW THE BLOCKING SCREWS


(DOWELS) YOU MUST APPLY LOCTITE THREADLOCKER TO ENSURE THE
GLUE EFFECTIVENESS INSIDE THE DOWELS THREADED HOLES.
FAILURE TO APPLY LOCTITE MAY RESULT IN THE SCREWS BACKING
OUT OVER TIME.

2-8-7 Checking Electrical Cables and Connections, and Cable Connection Points (every 12 months)

Verify the following:


− electrical cables are not scratched
− connectors are secure on their sockets
− cable fasteners are secure
2-8-8 Reassembling the Parts

Note: Before mounting the covers, ensure all the ground cables are connected.

Reassemble the covers in the following order:


1. internal cover
2. main cover
3. front panel

Note: At the end of each operation of periodic maintenance, the complete ”functional tests” procedure must be ap-
plied (refer to Chapter 4 of this manual).

98
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


SECTION 3
COLLIMATOR INSTALLATION VERIFICATION

1. Apply LOTO to system power.


2. Remove the locking set screws (item 4) from the collimator locking ring. Refer to Illustration 6−1.
3. Using the service wrench supplied, fully loosen the collimator locking ring (item 2) to detach the
collimator from the X−ray tube mounting plate (PN. CGY0750381). To loosen the ring, turn the
wrench counterclockwise. Refer to Illustration 6−1.
ILLUSTRATION 6−1
LOCKING RING AND SET SCREWS

4
4

CCW

4. Inspect the threads of the collimator mounting ring (item 1) on the x−ray tube and the locking ring
(item 2) on the collimator for evidence of a previous cross threaded condition. If any visible thread
damage is present due to cross threading, order a new plate and and/or new collimator as
necessary.

Note: If replacement of the mounting plate is necessary, alignment of the new mounting plate to the X−ray tube focal
spot is required. Refer to the Ultranet SA or Ultranet SM service manual for this procedure.

THE COLLIMATOR IS SLIGHTLY FRONT HEAVY. APPLY A SLIGHT UPWARD


FORCE TO THE COLLIMATOR HANDLE TO AVOID CROSS THREADING
THE LOCKING RING.

5. If no damage is visible, install the collimator and tighten the locking ring. Be careful not to cross
thread the locking ring on the mounting plate. Make sure the pin (item 3) lines up with the slot in the
collimator mounting ring. Refer to Illustration 6−1.

99
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


6. Using the service wrench supplied, tighten the collimator locking ring until. Do not overtighten. The
locking ring will be torqued at a later step.
7. Using GEMS torque wrench, 46−315668P1, Sturtevant Richmond 3SDR−200I, 40−200 in−lbs, set
the torque wrench to 200 in−lbs (22.5 Nm) and connect it to the collimator service wrench using the
square in the wrench.

Note: To apply proper torque to the locking ring ensure that the long axis of the torque wrench is aligned with the long
axis of the wrench handle when applying torque. Refer to Illustration 6−2. Also, make sure that you hold the
torque wrench at the handle ONLY. Holding it in other places will cause the applied torque to be wrong.

8. Tighten the collimator locking ring until the torque wrench clicks. Be careful not to jam the torque
wrench against the screws which hold the mounting plate.

Stop tightening when the torque wrench clicks.


DO NOT overtighten. If you overtighten, you may strip the set screw
threads or break the torque wrench.

ILLUSTRATION 6−2
LONG AXIS OF TORQUE WRENCH
ALIGNED WITH LONG AXIS OF HANDLE

9. Remove the locking ring service wrench and torque wrench.

BE CAREFUL NOT TO CROSS THREAD THE SET SCREWS. PROPER


TORQUE CANNOT BE ACHIEVED IF THE SET SCREWS ARE CROSS
THREADED.

LOCTITE THREADLOCKER MUST BE APPLIED DURING INSTALLATION TO


PREVENT THE SET SCREWS FROM BACKING OUT OVER TIME.

10. Apply Loctite Threadlocker to the supplied set screws and install at least 2 set screws in the 3 holes
in the collimator locking ring using the supplied hex socket wrench. Apply either the medium strength
Loctite Threadlocker 242 (PN. 46−170686P2) or the oil resistant/medium strength Loctite
Threadlocker 243 (PN. 2257866) to the set screws.

100
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


Note: It is required to install only 2 set screws. Three holes are provided in the locking ring to avoid clearance/acces-
sibility problems.

Note: Use enough loctite to coat the threads when the set screws are installed. DO NOT use an excessive amount of
Loctite that could seep into the interior of the collimator and cause damage to the collimator.

11. Using the GEMS torque wrench 46−282782P1, Sturtevant Richmond CAL−36/4, 2−36 in−lbs, and
the supplied hex socket wrench and adapter, set the torque wrench to 12 in−lbs (1.4 Nm) and tighten
the set screws until the torque wrench clicks. Refer to Illustration 6−3.

Stop tightening when the torque wrench clicks.


DO NOT overtighten. If you overtighten, you may strip the set screw
threads or break the torque wrench.

12. Store the spanner wrench on site for future service of the collimator.
13. Remove LOTO and apply system power.
14. Return XT suspension to operation.
15. Store the torque wrench, ensuring it is set to zero before storing.
ILLUSTRATION 6−3
ADJUSTING THE TORQUE WRENCH

LINE UP WIRE IN BUBBLE


WITH 12 IN−LBS.

USE PULL OUT LEVER ON END OF


HANDLE TO ADJUST SETTING OF
IN−LBS OF TORQUE.

101
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


SECTION 4
ULTRANET COLLIMATOR INSTALLATION CHECK LIST

Complete the following checklist in Table 6−1 to ensure proper installation. If all of the answers to the
checklist questions are ‘yes’, then the collimator is properly installed.

TABLE 6−1
ULTRANET COLLIMATOR INSTALLATION CHECKLIST

CHECKLIST YES NO

Try to insert the collimator spanner wrench between the lock ring and the collimator mounting plate. Refer
1 to Illustrations 6−4. Does the space between the ring and the mounting plate prevent the wrench from being
inserted? (indicates full thread engagement)

Is the collimator roughly parallel to the X−ray tube? (If not, it could indicate a
2
potential cross threaded condition)

Are the threads of the mounting plate on the X−ray tube covered up by the locking ring or not visible? (indi-
3
cates full thread engagement)

4 Has the proper torque been applied to the collimator locking ring?

5 Have the 2 set screws been installed with Loctite?

6 Has the proper torque been applied to the 2 set screws?

ILLUSTRATION 6−4
CHECKING THE LOCKING RING AND COLLIMATOR MOUNTING PLATE SPACING

102
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

CHAPTER 7 − SCHEMATICS

SECTION 1

103
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


THIS PAGE INTENTIONALLY LEFT BLANK.

104
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

1-1 Schematics

ILLUSTRATION 7−1
MIS MAP (SYSTEM MPH 50/65/80 GENERATOR WITH COMPAX 40E−MANUAL / SG100 / XT SUSPENSION)

2145642 16 27222
2145641
10
27221

2151729
9 27225

ÈÈÈÈ
ÈÈÈÈ
46−329288G2 10707
XT SUSPENSION
4 J100

TS2
ROOM1 A1

J42
J101

ÈÈÈÈ

J40
Door Safety
46−195120G4 107 J1
1
ROOM1 A2 SUPPLIED BY THE FIELD
Room Indicators 19
10680 2128497
2

10706
MPH 50/65/80 CABINET
MAIN POWER

9 15
MPH A6 A8 Plugs Panel

46−329291G1
J4
XJ3
MPH A6 A6
MPH A6 A4 FAN COVER
II Sensor Board ANODE
Wall Bucky & HOUSING CATHODE
XJ12
XJ16
LINE XJ10 XJ10 SG 100 WALL STAND
SAFETY
XJ1A/B XJ15
SWITCH
OPTIONAL

MPH A6 A3 MPH A6 A5
MPH A1 A2 Distribution Table Tomo
Stator Connector Board Collimator J/P2
Panel 4
27235 MPH A4
Front Electronic Panel
MPH A3 A2 10705
X−RAY TUBE+
Main Power
2 +COLLIMATOR
5
ST1 MPH A3 A1
DC Power MPH A5
Real Electronic Panel

46−329341G2
2138905

MPH A3 MPH A3 A3 1
FL1 Inverter Board

MPH A1
F11 AC Distribution
F10 I+ I−

P1/P2 F18 F15 MPH A2


/P3 F19 F14 HV TANK

W34 14

8 2126152
2126152

12
2126153

3 MPH A8 MPH A7
XJ1
Program−X

W32 MPH A7 13
Control Console
MPH A8
XJ1
2104204

COMPAX 40E TABLE


11
27228
MPH A7
XJ4

MPH A9
MPH A9 6 27229
Printer−X J43

MPH A9
2125971 serial
2145602
link 12 J42

2145651
13 J40

2167144 TB1(Phase)
27227 TB1(Neutral)
11
TB1(GND)

105
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ILLUSTRATION 7−2
GROUNDING INTERCONNECTING MAP (SYSTEM MPH 50/65/80 GENERATOR WITH COMPAX 40E−MANUAL / SG100 / XT SUSPENSION)

Note: All grounding cables are yellow−green. XT SUSPENSION


CARRIAGE
GND STUD

46−316922G2
MPH 50/65/80 CABINET XT Ground Cable ( Supplied with XT ) I/O PANEL GND
4.1

MPH A6 A8 Plugs Panel

46−216211G10 8.1
MPH A6 A6 MPH A6 A4
XT Ground Cable
II Sensor Board Wall Bucky ( Supplied with XT )

MAIN GROUND
10708
MPH A6 A3 MPH A6 A5
MPH A1 A2 Distribution Table Tomo
Stator Connector Board Collimator
Panel MPH A4
Front Electronic Panel
MPH A3 A2
Main Power

SG 100 WALL STAND


MPH A3
A1 MPH A5
DC Power Real Electronic Panel

MPH A3 MPH A3 A3
FL1 Inverter Board
2138905 ( Power Supply Cable ) MPH
A1 Ground Plug I+ I−
27235 1.1 MPH A2
HV TANK
OPTIONAL 5.1
REFER TO JOBCARD IST002
IN APPROPRIATE VERSION 27301
MPH SERVICE MANUALS 2140362 ( SG Ground Cable )
FOR GROUND PLUG LAYOUT
MPH V1 SM: 2137354−100
MPH V2/V3 SM: 2165156−100
6.1
2105836
Printer Ground Cable 2.1
(Supplied with Printer) 2140362 ( TABLE Ground Cable )
45553701
(Console Ground Cable)
3.1
27300
LEGENDE MPH A8
MPH A7
S SIGNAL POINT GND Program−X
Control Console
CHASSIS
COMPAX 40E TABLE
PROTECTIVE
EARTH TERMINAL

REFERENCE TERMINAL
GND STUD
EQUIPOTENTIAL TERMINAL

MIS RUN − FIELD GND STUD


000 SUPPLIED WIRE
A SAFETY GROUND WIRE
MPH A9
B SIGNAL GROUND WIRE Printer−X
SAFETY & SIGNAL
C GROUND WIRE TB1−GND
000 MIS RUN − CABLE
REFER TO OPTION
MAPS FOR MIS NUMBER
FACTORY INSTALLED WIRE

106
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ILLUSTRATION 7−3
MIS MAP (SYSTEM MPH 50/65/80 GENERATOR WITH COMPAX 40E−AUTO / SG100 / XT SUSPENSION)
2145644
18 27223
2145642 16 27222
2145641
10 27221

ÈÈÈ
2151729
9
27225

ÈÈÈ
ROOM1 A1
Door Safety
46−329288G2 XT SUSPENSION

ÈÈÈ
4 10707 J100
ROOM1 A2 SUPPLIED BY THE FIELD
19

TS2
Room Indicators

J41
J101

J42

J40
J45
46−195120G4 J1
107 1
20
2128497 2 10680
21
46−329341G2 10705 10706
20

MAIN POWER
MPH 50/65/80 CABINET
SC2 RAFL−S Electronic P3
Interface 18 15
9 CL1 P2
J202

J201
MPH A6 A8 Plugs Panel

46−329291G1
J4

XJ3

2145656
MPH A6 A6
MPH A6 A4 FAN COVER
II Sensor Board ANODE
Wall Bucky & HOUSING CATHODE
XJ12
XJ16
LINE XJ10 XJ10 SG 100 WALL STAND
SAFETY
SWITCH XJ1A/B XJ15
OPTIONAL

MPH A6 A3 MPH A6 A5 J / P1
MPH A1 A2 Distribution Table Tomo
Stator Connector Board Collimator J / P2 17
Panel 4
27235 MPH A4
Front Electronic Panel
MPH A3 A2 X−RAY TUBE+
Main Power
2 +COLLIMATOR
5
ST1
MPH A3 A1
DC Power MPH A5
Real Electronic Panel
2138905

MPH A3 MPH A3 A3 1
FL1 Inverter Board

MPH A1 27224
F11
F10
AC Distribution I+ I−

P1/P2 F18 F15 MPH A2


/P3 F19 F14 HV TANK
21

7 22

W34
2126152
8 2126152
17
2126153

12 27231
MPH A7
XJ1
3 MPH A8
Program−X
MPH A7 13 27230
W32 Control Console
MPH A8
2104204

XJ1
COMPAX 40E TABLE
27228
11

MPH A7
XJ4

J44

27229
2145655
MPH A9 21 J23
MPH A9 6
Printer−X 2145618 J41
22
MPH A9 2125971 serial 2145602
link 12 J42

2145651
13 J40

2167144 TB1(Phase)
27227 11 TB1(Neutral)
TB1(GND)

107
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ILLUSTRATION 7−4
MIS MAP (SYSTEM MPH 50/65/80 GENERATOR AND RS85 TUBE STAND WITH COMPAX 40E−MANUAL AND SG100)

2145642 16 27222
2145641
10
27221

2151729
9 27225

27304

ÈÈÈÈ
2171350
2165548 4 27238

ÈÈÈÈ

TS2
ROOM1 A1

J42

J40
Door Safety
36002080 107 1

ÈÈÈÈ
ROOM1 A2 SUPPLIED BY THE FIELD
Room Indicators 2145653
27239 2

19

MPH 50/65/80 CABINET


MAIN POWER

9
MPH A6 A8 Plugs Panel
L/N L+ / L−
J4
X3
(RS−85)24VDC XJ3 TOWER
Tube−Stand MPH A6 A6
MPH A6 A4 X3
Power Supply II Sensor Board
Wall Bucky
XJ12
XJ16 X4
LINE XJ10 XJ10 SG 100 WALL STAND
SAFETY
XJ1A/B XJ15
OPTIONAL

SWITCH
MPH A6 A3 MPH A6 A5

2165551
MPH A1 A2 Distribution Table Tomo 27303
Stator Connector Board Collimator
Panel 4
27235 MPH A4
Front Electronic Panel
MPH A3 A2
Main Power
2 J1 ANODE
5
FAN COVER CATHODE
ST1 MPH A3 A1 & HOUSING
DC Power MPH A5
Real Electronic Panel
27293
2138905

MPH A3 MPH A3 A3 1 J/P2


FL1 Inverter Board
X−RAY TUBE
MPH A1 27292
F11 AC Distribution +
I+ I−
F10 ULTRANET SM
P1/P2 F18 F15 MPH A2 COLLIMATOR
/P3 F19 F14 HV TANK

14 15
7

W34 RS85 TUBE STAND


2126152
8

2165550
12
2126153

3 MPH A8 MPH A7
XJ1
Program−X

W32 MPH A7 13
Control Console
MPH A8
XJ1
2104204

COMPAX 40E TABLE


11
27228
MPH A7
XJ4

MPH A9
MPH A9 6 27229
Printer−X J12

MPH A9 2125971 serial


link 2145602 12 J42

2145651
13 J40

2167144 TB1(Phase)
27227 11 TB1(Neutral)
TB1(GND)

J21

108
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ILLUSTRATION 7−5
GROUNDING INTERCONNECTING MAP (SYSTEM MPH 50/65/80 GENERATOR AND RS85 TUBE STAND WITH COMPAX 40E−MANUAL AND SG100)

Note: All grounding cables are yellow−green.

MPH 50/65/80 CABINET


I/O PANEL GND

2165548
POWER SUPPLY CABLE
MPH A6 A8 Plugs Panel
GND
POWER SUPPLY
MPH A6 A6 MPH A6 A4
II Sensor Board Wall Bucky
7.1
RS 85
MAIN GROUND
24VDC
TUBESTA
MPH A1 A2 ND MPH A6 A3 MPH A6 A5
Distribution Table Tomo 27238
Stator Connector POWER Board Collimator 2183955
Panel
SUPPLY MPH A4
Front Electronic Panel
2140362 ( SG Ground Cable ) GND
MPH A3 A2
RS85 TOWER
7.1 Main Power 5.1
8.1
MPH A3
A1 MPH A5
DC Power Real Electronic Panel
SG 100 WALL STAND
MPH A3 MPH A3 A3 2140362
FL1 Inverter Board 27302
2138905 ( Power Supply Cable ) 27305
MPH 27301
A1 Ground Plug I+ I−
4.1
27235 1.1 MPH A2
HV TANK
OPTIONAL
RS85 Ground Cable
REFER TO JOBCARD IST002
IN APPROPRIATE VERSION
MPH SERVICE MANUALS X−RAY TUBE
FOR GROUND PLUG LAYOUT GND X−RAY TUBE
+
MPH V1 SM: 2137354−100 SA/SM Ultranet
MPH V2/V3 SM: 2165156−100 COLLIMATOR
2165548
POWER SUPPLY CABLE
2105836
Printer Ground Cable 2.1
(Supplied with Printer) 2140362 ( TABLE Ground Cable )
45553701 RS85 TUBE STAND
6.1
(Console Ground Cable) 3.1
27300
LEGENDE MPH A8
MPH A7
S SIGNAL POINT GND Program−X
Control Console
CHASSIS
COMPAX 40E TABLE
PROTECTIVE
EARTH TERMINAL

REFERENCE TERMINAL
GND STUD
EQUIPOTENTIAL TERMINAL

MIS RUN − FIELD GND STUD


000 SUPPLIED WIRE
A SAFETY GROUND WIRE
MPH A9
B SIGNAL GROUND WIRE Printer−X
SAFETY & SIGNAL
C GROUND WIRE TB1−GND
000 MIS RUN − CABLE
REFER TO OPTION
MAPS FOR MIS NUMBER
FACTORY INSTALLED WIRE

109
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ILLUSTRATION 7−6
MIS MAP (SYSTEM MPH 50/65/80 GENERATOR AND RS85 TUBE STAND WITH COMPAX 40E−AUTO AND SG100)
2145644
18 27223
2145642 16 27222
2145641
10 27221

ÈÈÈ
2151729
9
27225

ÈÈÈ
ROOM1 A1
Door Safety
27304
4 27238

ÈÈÈ
2165548

2171350
ROOM1 A2 SUPPLIED BY THE FIELD
19

TS2
Room Indicators

J41
J42
J45

J40
36002080 107 1
20
2145653 2 27239
21
2165546
20 27236

MAIN POWER
MPH 50/65/80 CABINET
SC2 RAFL−S Electronic P3
Interface 18 X3
9 CL1 P2
J202 TOWER
MPH A6 A8 Plugs Panel X3
J201
J4
J301 X4
XJ3

2145656
MPH A6 A6 2165547
MPH A6 A4 27237 23
II Sensor Board
Wall Bucky
XJ12
XJ16
27303
LINE XJ10 XJ10 SG 100 WALL STAND
SAFETY
SWITCH XJ1A/B XJ15
L+ / L− L / N
(RS−85)24VDC MPH A6 A3 MPH A6 A5
MPH A1 A2 Tube−Stand Distribution Table Tomo
Stator Connector Power Supply Board Collimator 17
Panel 4 ANODE
27235 MPH A4 FAN COVER CATHODE
Front Electronic Panel & HOUSING
MPH A3 A2
OPTIONAL

Main Power
2
5 J/P2
ST1 MPH A3 A1 X−RAY TUBE
DC Power MPH A5
Real Electronic Panel + J /P1
ULTRANET SA
2138905

MPH A3 MPH A3 A3 1 COLLIMATOR


FL1 Inverter Board

MPH A1
F11
F10
AC Distribution I+ I−

P1/P2 F18 F15 MPH A2


/P3 F19 F14 HV TANK
21 RS85 TUBE STAND

7
22

W34
2126152
8 2126152
2126153

12 27231
MPH A7
XJ1
3 MPH A8 27224
Program−X
MPH A7 27230
13
W32 Control Console
17
2104204

MPH A8
XJ1
COMPAX 40E TABLE
27228
11

MPH A7
XJ4

J44

27229
2145655
MPH A9 21 J23
MPH A9 6
Printer−X 2145618 J41
22
MPH A9 2125971 serial 2145602
link 12 J42

2145651
13 J40

2167144 TB1(Phase)
27227 11 TB1(Neutral)
TB1(GND)

110
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ILLUSTRATION 7−7
MIS MAP (SYSTEM MPH 50/65/80 GENERATOR WITH COMPAX 40E−MANUAL / SG100 / XT SUSPENSION WITH TOMO LINK)

2145642 16 27222
2145641
10
27221

2151729
9 27225

ÈÈÈÈ ÂÂÂÂ
46−329288G2 10707
XT SUSPENSION
4 J100

ÈÈÈÈ ÂÂÂÂ

TS2
ROOM1 A1

J42
J101

J40
Door Safety TOMO LINK
46−195120G4 107 J1

ÈÈÈÈ
1 DRIVE UNIT
ROOM1 A2 SUPPLIED BY THE FIELD
Room Indicators 19
10680 2128497 J100 J101
2

TL3

TL3
10706
MPH 50/65/80 CABINET
MAIN POWER

9 15
MPH A6 A8 Plugs Panel

26

46−329291G1
J4
XJ3
MPH A6 A6
MPH A6 A4 FAN COVER
II Sensor Board ANODE
Wall Bucky & HOUSING CATHODE
OPTIONAL

XJ12 27
XJ16
XJ10
LINE XJ10 XJ12 SG 100 WALL STAND
SAFETY
XJ1A/B XJ14
SWITCH
XJ15
MPH A1 A2 MPH A6 A3 MPH A6 A5
Stator Connector Distribution Table Tomo J/P2
Panel Board Collimator 4
27235 MPH A4
Front Electronic Panel 11214
MPH A3 A2 10705
X−RAY TUBE+
Main Power
2 +COLLIMATOR
5
ST1 MPH A3 A1 11215
DC Power MPH A5
Real Electronic Panel

46−329341G2
2138905

MPH A3 MPH A3 A3 1
FL1 Inverter Board

MPH A1
F11 AC Distribution
F10 I+ I− 2132261
P1/P2 F18 F15 MPH A2
/P3 F19 F14 HV TANK

2124056

2159597
7 28
11411

ÂÂÂÂ
W34
Quick Disconnect
8 2126152 Connector

ÂÂÂÂ
2126152
12
2126153

ÂÂÂÂ
3 MPH A8 MPH A7
XJ1
Program−X
14 TL2
W32 MPH A7 13 Y
Control Console TOWER

MPH A8
XJ1 COMPAX 40E TABLE
2104204

11
27228 CONSOLE
MPH A7
XJ4

TL4
J2

J43
MPH A9 27229
MPH A9 6 Z
Printer−X
T J43
2125971 serial J103 TL1
MPH A9 2145602
link 12 J42 J102 TL1

2145651 J107
13 J40 bb
J8 (PCB POWER C40E) J110
TB1(Phase)
TB1(Neutral) J100 TL1
2167144 TB1(GND) FL1 FILTER aa
27227 11
2132121 TL1 J106 J101 TL1
J22 25 11212
J105 TL1

111
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ILLUSTRATION 7−8
GROUNDING INTERCONNECTING MAP (SYSTEM MPH 50/65/80 GENERATOR WITH COMPAX 40E−MANUAL / SG100 / XT SUSPENSION WITH TOMO LINK)

Note: All grounding cables are yellow−green. XT SUSPENSION

ÂÂÂÂ
CARRIAGE
GND STUD

ÂÂÂÂ
TOMO LINK
46−316922G2 DRIVE UNIT
MPH 50/65/80 CABINET XT Ground Cable ( Supplied with XT ) I/O PANEL GND
4.1

MPH A6 A8 Plugs Panel


46−216211G10 8.1
MPH A6 A6 XT Ground Cable
MPH A6 A4
II Sensor Board Wall Bucky
( Supplied with XT )

MAIN GROUND
10708
MPH A6 A3 MPH A6 A5
MPH A1 A2 Distribution Table Tomo
Stator Connector Board Collimator
Panel MPH A4
Front Electronic Panel
MPH A3 A2
Main Power

SG 100 WALL STAND


MPH A3
A1 MPH A5
DC Power Real Electronic Panel

MPH A3 MPH A3 A3
FL1 Inverter Board
2138905 ( Power Supply Cable ) MPH
A1 Ground Plug I+ I−

27235 1.1 MPH A2


HV TANK
OPTIONAL
REFER TO JOBCARD IST002 5.1
IN APPROPRIATE VERSION
MPH SERVICE MANUALS 27301 2140362 ( SG Ground Cable )
FOR GROUND PLUG LAYOUT
MPH V1 SM: 2137354−100
MPH V2/V3 SM: 2165156−100
6.1
2105836 2.1
Printer Ground Cable
(Supplied with Printer) 2140362 ( TABLE Ground Cable )
45553701
(Console Ground Cable)

ÂÂÂÂ
Â
3.1
27300

ÂÂÂ
ÂÂ
LEGENDE MPH A8

ÂÂÂ
ÂÂ
MPH A7
S SIGNAL POINT GND Program−X
Control Console
CHASSIS
COMPAX 40E TABLE
PROTECTIVE
EARTH TERMINAL TOWER

REFERENCE TERMINAL
GND STUD
EQUIPOTENTIAL TERMINAL
CONSOLE
MIS RUN − FIELD GND STUD
000 SUPPLIED WIRE
A SAFETY GROUND WIRE
MPH A9
B SIGNAL GROUND WIRE Printer−X TB1−GND
SAFETY & SIGNAL
C GROUND WIRE
000 MIS RUN − CABLE
REFER TO OPTION
MAPS FOR MIS NUMBER
FACTORY INSTALLED WIRE

7.1
(Internal) GND (TLK ELECTRONICS)
Tomo Link Ground Cable )
45433617

112
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

ILLUSTRATION 7−9
MIS MAP (SYSTEM MPH 50/65/80 GENERATOR WITH COMPAX 40E−AUTO / SG100 / XT SUSPENSION WITH TOMO LINK)
2145644
18 27223
2145642 16 27222
2145641
10 27221

ÈÈÈ
2151729
9 9
27225

ÈÈÈ ÂÂÂÂ
ROOM1 A1
Door Safety
46−329288G2 XT SUSPENSION

ÈÈÈ ÂÂÂÂ
4 10707 J100
ROOM1 A2 SUPPLIED BY THE FIELD
19

TS2
Room Indicators

J41
J101

J42

J40
J45
ÂÂÂÂ
46−195120G4 J1 TOMO LINK
107 1
DRIVE UNIT
20
2128497 2 10680 J100 J101

TL3

TL3
21
46−329341G2 10705 10706
20

MAIN POWER
MPH 50/65/80 CABINET
SC2 RAFL−S Electronic P3
Interface 18 15
9 CL1 P2
J202
MPH A6 A8 Plugs Panel
J201

46−329291G1
J200 J4 26
XJ3

2145656
MPH A6 A6
II Sensor Board MPH A6 A4 FAN COVER ANODE
Wall Bucky & HOUSING CATHODE
XJ12
XJ16
27
XJ10
LINE XJ10
XJ12 SG 100 WALL STAND
SAFETY
SWITCH XJ1A/B XJ14
XJ15
J / P1
MPH A1 A2 MPH A6 A3 MPH A6 A5
Stator Connector Distribution Table Tomo J / P2 17
Panel Board Collimator 4
27235 MPH A4
Front Electronic Panel
11214
MPH A3 A2 X−RAY TUBE+
OPTIONAL

Main Power
2 +COLLIMATOR
5
ST1
MPH A3 A1 11215
DC Power MPH A5
Real Electronic Panel
2138905

MPH A3 MPH A3 A3 1
FL1 Inverter Board

MPH A1 27224
F11
F10
AC Distribution I+ I−
2132261
P1/P2 F18 F15 MPH A2
/P3 F19 F14 HV TANK
21
2124056
2159597
7 28
22 11411

ÂÂÂÂ
W34
2126152 Quick Disconnect
8 2126152 Connector

ÂÂÂÂ
17
2126153

12 27231

ÂÂÂÂ
MPH A7
MPH A8 XJ1
3
Program−X TOWER
MPH A7 TL2
Control Console 13 27230
11212 W32 Y

MPH A8 COMPAX 40E TABLE
2104204

XJ1

27228
11
CONSOLE
MPH A7
XJ4

J44 TL4
J2
27229
2145655
MPH A9 21 J23
MPH A9 6 Z
Printer−X 2145618 J41
22
2125971 serial J103 TL1
MPH A9 2145602
link 12 J42
J102 TL1
TL1
25 J107
2145651 bb
13 J40
TL1 J110
J8 (PCB POWER C40E)
2167144 TB1(Phase)
27227 11 J100 TL1
TB1(Neutral) aa
TB1(GND) FL1 FILTER
2132121 J101 TL1
11212 TL1 J106
25
J105 TL1

113
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

1-2 Cables Identification According to the Mis Map and Mis Chart

TABLE 7−1
SYSTEM COMPAX 40E AND XT SUSPENSION WITH MPH 50/65/80 AND SG−60/100 (ILLUSTRATIONS 3.1, 3.2 AND 3.3)

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

1 46−195120G4 VHV CABLE MPH 50/65/80 CABINET XT SUSPENSION  X−Ray TUBE 18m
UL E4413 − 75kV (MPH A2 − HV TANK)  
MIS 107−A 75kV  TUBE I ( + ) TERMINAL ANODE TERMINAL
19.94 ± 0.38mm

1 46−195120G4 VHV CABLE MPH 50/65/80 CABINET XT SUSPENSION  X−Ray TUBE 18m
UL E4413 − 75kV (MPH A2 − HV TANK)  
MIS 107−C 75kV  TUBE I ( − ) TERMINAL CATHODE TERMINAL
19.94 ± 0.38mm

2 2128497 STATOR MPH 50/65/80 CABINET XT SUSPENSION  X−Ray TUBE  18m


CABLE (MPH A1 A2−Stator Connector Panel)  (Terminals Strip) 
MIS 10680 UL 2516 − 600V ST1 Connector AMP TERMINALS TERMINAL
240VAC
12.19 ± 0.51mm

3 2104204 PRINTER−X MPH 50/65/80 CABINET PRINTER−X  18m


Power Supply (MPH A6 A3−Distribution Board) (MPH A9 Module) 
(W34) UL 1011 − 600V CABLE XJ1 Tinned Wire TS TERMINAL
230VAC
9.7" 0.4 mm

4 46−329288G2 XT Suspension MPH 50/65/80 CABINET XT SUSPENSION  18m


Power Supply (MPH A1−AC Distribution Panel) (A6 A1 module) 
MIS10707 UL/CSA − 300V CABLE F15−2 Tinned Wire J100 Connector AMP
240VAC F14−2
9.65" 0.51 mm Ground Plug

5 2138905 MPH POWER MPH 50/65/80 CABINET LINE SAFETY SWITCH 9m


X SUPPLY (MPH A1 A1−AC Distribution Panel) 
MIS27235 UL 1019 − 600V CABLE P1 Ring Terminal Phase 1 (R) Ring Terminal
480VAC P2 Phase 2 (S)
33.6" 1.6 mm P3 Phase 3 (T)
GND GROUND

6 2125971 PRINTER−X (PU3) MADRID CONSOLE  PRINTER−X  18m


UL E125349 Control CABLE (MPH A7 − I/O Console) SUD−D (37p/F) (MPH A9 Module)  SUD−D (37p/F)
(2125971) 24VAC XJ4 Connector MPH A9 Connector
2.7" 0.15 mm

7 2126153 Program−X (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET 18m


UL E125349 − 250V Control CABLE (MPH A8 − Program−X)  SUD−D (37p/F) (MPH A6 A3−Distribution Board) SUD−D (37p/M)
(W32) 24VAC XJ1 Connector XJ10 Connector
9.1" 0.15 mm

114
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

8 2126152 Console (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET 18m


UL E125349 − 250V Control CABLE (MPH A7 − I/O Console)  SUD−D (37p/F) (MPH A6 A3−Distribution Board) SUD−D (37p/F)
(2126152) 24VAC XJ1 Connector XJ12 Connector
9.1" 0.15 mm

9 2151729 SG100 MPH 50/65/80 CABINET  SG100 BUCKY STAND  18m


UL 2464 − 300V Power Supply (MPH A1−AC Distribution Panel)  (I/O PANEL) 
MIS 27225 240VAC CABLE F11 − 2 Tinned Wire TS2 − L Tinned Wire
8" 0.2 mm F10 − 2 TS2 − N
Ground Plug GND STUD

10 2145641 SG 60/100 MPH 50/65/80 CABINET  Female Connector SG 60 or SG100 BUCKY STAND  18m
UL 2464 − 300V BUCKY CABLE (MPH A6 A4 Wall Bucky Board)  (9p) (I/O PANEL) 
MIS 27221 240/110 (USA)VAC XJ3 J42 AMP_(M26p/F)
8" 0.2 mm Connector

11 2167144 TABLE MPH 50/65/80 CABINET  COMPAX 40E TABLE  18m


UL 2463 − 600V Power Supply (MPH A1−AC Distribution Panel)  (I/O PANEL) 
MIS 27227 480VAC CABLE F18 − 2 Tinned Wire TB1 (PHASE) Tinned Wire
9.1" 0.15 mm F19 − 2 TB1 (NEUTRAL)
Ground Plug TB1 (GROUND)

12 2145602 C40E TABLE MPH 50/65/80 CABINET  COMPAX 40E TABLE  18m
BUCKY CABLE (MPH A6 A5 Table Tomo Board)  (I/O PANEL) 
MIS 27228 UL 2463 − 300V XJ10 Tinned Wire J42 AMP_(M26p/F)
240/110 (USA)VAC Connector
8" 0.2 mm

13 2145651 TABLE MPH 50/65/80 CABINET COMPAX 40E TABLE  18m


UL 2464 − 300V ION CHAMBER (MPH A6 A3−Distribution Board) AMP_(M14p/M) (I/O PANEL)  AMP_(M14p/F)
MIS 27229 200/15 (USA)VDC CABLE XJ15 Connector J40 Connector
8" 0.2 mm

14 46−329341G2 SID CABLE XT SUSPENSION  COMPAX 40E TABLE  18m


UL 2464 − 300V (3A1 module)  SUD−D (37p/F) (I/O PANEL)  SUD−D (37p/F)
MIS 10705 200/15 (USA)VDC J101 Connector J43 Connector
8" 0.2 mm

15 46−329291G1 (Manual System) XT SUSPENSION  XT SUSPENSION  4m


UL 2464 − 300V COLLIMATORCA- (3A1 module)  SUD−D (37p/F) (Ultranet SM Collimator) AMP_ (26p/F)
MIS 10706 24VDC BLE J1 Connector J/P2 Connector
12.70" 0.51 mm

16 2145642 SG 60/100 MPH 50/65/80 CABINET SG100 BUCKY STAND  18m


UL 2464 − 300V ION CHAMBER (MPH A6 A3−Distribution Board) AMP_(M14p/M) (I/O PANEL)  AMP_(M14p/F)
MIS 27222 200/15 (USA)VDC CABLE XJ16 Connector J40 Connector
8" 0.2 mm
Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

1.1 2138905 Power Supply MPH 50/65/80 CABINET LINE SAFETY SWITCH 9m
X UL 1019 − 600V CABLE (MPH A1 A1−AC Distribution Panel) Ring Terminal  Ring Terminal
MIS 27235 480VAC TS − GND GROUND
33.6" 1.6 mm

115
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


2.1 45553701 Console (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET  18m
UL 1019 − 600V Ground CABLE (MPH A7 − I/O Console)  (MPH A1−AC Distribution Panel) 
(45553701) 0VDC GND Terminal Ground Plug Tinned Wire
7.2" 0.3 mm

3.1 2140362 TABLE COMPAX 40E TABLE  MPH 50/65/80 CABINET  18m
UL 1015 − 600V Ground CABLE  (MPH A1−AC Distribution Panel) 
MIS 27300 0VDC TB1−GND Tinned Wire Ground Plug Tinned Wire
4.5" 0.2 mm

4.1 46−316922G2 GROUND XT SUSPENSION  MPH 50/65/80 CABINET  100 FT


SUPPLIED with XT CABLE (CHASSIS)  Terminal (MPH A1−AC Distribution Panel)  Terminal
MIS 10708 SUSPENSION GND STUD Ground Plug
UL 8AWG − 600V
0VDC
0.260 ± 0.005in

5.1 2140362 SG 60/100 SG 100 BUCKY STAND MPH 50/65/80 CABINET  18m
UL 1015 − 600V Ground CABLE (I/0 PANEL) (MPH A1−AC Distribution Panel) 
MIS 27301 0VAC GND STUD Terminal Ground Plug Tinned Wire
4.5" 2.2 mm

6.1 2105836 PRINTER−X PRINTER−X  MPH 50/65/80 CABINET  18m


SUPPLIED with Ground CABLE (MPH A9 Module)  (MPH A1−AC Distribution Panel) 
(W19) PRINTER−X GND (Ground) Terminal Ground Plug Tinned Wire
UL 1011 − 600V
0VDC
4.8" 0.2 mm

8.1 46−216211G10 XT XT SUSPENSION  X−RAY TUBE  4m


SUPPLIED with XT X−Ray Tube (CHASSIS)  
(46−216211 SUSPENSION Ground CABLE GND STUD Terminal Ground Plug Terminal
G10) UL 8AWG − 600V
0VDC
0.260 ± 0.005in

15 46−329291G1 (Auto System) XT SUSPENSION  AMP 37 XT SUSPENSION  AMP_ (26pos/F)/ 4m


UL 2464 − 300V COLLIMATORCA- (3A1 module)  CONT RECP (Ultranet SA Collimator) (14pos/F) Connec-
MIS 10706 24VDC BLE J1 Connector J/P1−J/P2 tors
12.70" 0.51 mm

17 2145656 COMPATIBILITY SG 60 or SG100 BUCKY STAND  COMPAX 40E TABLE  18m


UL 2464 − 300V CABLE (I/O PANEL)  SUD−D (9p/F) (I/O PANEL) 
MIS 27224 24VDC J45 Connector J44
4.5" 0.2 mm

116
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

18 2145644 SG 60/100 MPH 50/65/80 CABINET SG60/100 BUCKY STAND  18m


UL 2464 − 300V Automat. Tray (MPH A6 CL1 − RAFL−S, I/O Panel) (I/O PANEL)  SUD−D (15p/M)
MIS 27223 10VDC CABLE P3 J41 Connector
8" 0.2 mm

20 46−329341G2 SID CABLE XT SUSPENSION  AMP 37 MPH 50/65/80 CABINET AMP 37 24m
UL 2464 − 300V (3A1 module)  CONT RECP (MPH A6 SC2−INTERFACE) CONT RECP
MIS 10705 24VDC J101 Connector J201 Connector
12.70" 0.51 mm

21 2145655 SID DETENT COMPAX 40E TABLE  MPH 50/65/80 CABINET 18m
UL 2464 − 300V CABLE (I/O PANEL)  (MPH A6 SC2−INTERFACE)
MIS 27231 24VDC J23 J202
8" 0.2 mm

22 2145618 C40E COMPAX 40E TABLE  MPH 50/65/80 CABINET 18m


UL 2464 − 300V Automat. Tray (I/O PANEL)  (MPH A6 CL1 − RAFL−S, I/O Panel)
MIS 27230 10VDC CABLE J41 P2
8" 0.2 mm

INTERNAL 2145659 SAS INTERFACE MPH 50/65/80 CABINET MPH 50/65/80 CABINET  0.7m
UL 2464 − 300V CABLE (MPH A6 CL1 − RAFL−S, I/O Panel) (MPH A6 A8 Extension) 
MIS 27232 24VDC P10 XJ5
8" 0.2 mm

INTERNAL 2154537 Power Supply MPH 50/65/80 CABINET MPH 50/65/80 CABINET  1.5m
UL 2464 − 300V CABLE (MPH A6 CL1 − RAFL−S, I/O Panel) (MPH A6 SC3) 
MIS 27233 240VAC P1 Tinned Wire TB4−5/6 & GND Tinned Wire
8" 0.2 mm

INTERNAL 2152750 Power Supply MPH 50/65/80 CABINET  MPH 50/65/80 CABINET  1.5m
UL 2464 − 300V CABLE (MPH A6 SC3)  (MPH A1−AC Distribution Panel) 
MIS 27240 240VAC EMI FILTER IEC Connector F13 − 2 Tinned Wire
8" 0.2 mm F12 − 2
Ground Plug

117
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

TABLE 7−2
SYSTEM COMPAX 40E AND RS85 TUBE STAND WITH MPH 50/65/80 AND SG−60/100 (ILLUSTRATIONS 3.4, 3.5 AND 3.6)

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

1 36002080 VHV CABLE MPH 50/65/80 CABINET RS85 TUBE STAND  ARM  X−Ray TUBE 18m
(MPH A2 − HV TANK)  
MIS 107−A  TUBE I ( + ) TERMINAL ANODE TERMINAL

1 36002080 VHV CABLE MPH 50/65/80 CABINET RS85 TUBE STAND  ARM  X−Ray TUBE 18m
(MPH A2 − HV TANK)  
MIS 107−C  TUBE I ( − ) TERMINAL CATHODE TERMINAL

2 2145653 STATOR MPH 50/65/80 CABINET RS85 TUBE STAND  ARM  X−Ray TUBE 18m
CABLE (MPH A1 A2−Stator Connector Panel)  (Terminals Strip) 
MIS 27239 UL 2516 − 600V ST1 Connector AMP TERMINAL TERMINAL
240VAC
12.19 ± 0.51mm

3 2104204 PRINTER−X MPH 50/65/80 CABINET PRINTER−X  18m


Power Supply (MPH A6 A3−Distribution Board) (MPH A9 Module) 
(W34) UL 1011 − 600V CABLE XJ1 Tinned Wire TS TERMINAL
230VAC
9.7" 0.4 mm

4 2165548 RS85 MPH 50/65/80 CABINET RS85 POWER SUPPLY  3m


UL 2464 − 300V Power Supply (MPH A1−AC Distribution Panel) (Terminals strip) 
MIS 27238 240VAC CABLE F15−2 Tinned Wire X1−N Connector AMP
7.1 mm F14−2 X1−L
Ground Plug PE1 (GND Screw)

5 2138905 Power Supply MPH 50/65/80 CABINET LINE SAFETY SWITCH 9m


X CABLE (MPH A1 A1−AC Distribution Panel) 
MIS 27235 UL 1019 − 600V P1 Ring Terminal Phase 1 (R) Ring Terminal
480VAC P2 Phase 2 (S)
33.6" 1.6 mm P3 Phase 3 (T)
GND GROUND

6 2125971 PRINTER−X (PU3) MADRID CONSOLE  PRINTER−X  18m


UL E125349 Control CABLE (MPH A7 − I/O Console) SUD−D (37p/F) (MPH A9 Module)  SUD−D (37p/F)
(2125971) 24VAC XJ4 Connector MPH A9 Connector
2.7" 0.15 mm

7 2126153 Program−X (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET 18m


UL E125349 − 250V Control CABLE (MPH A8 − Program−X)  SUD−D (37p/F) (MPH A6 A3−Distribution Board) SUD−D (37p/M)
(W32) 24VAC XJ1 Connector XJ10 Connector
9.1" 0.15 mm

118
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

8 2126152 Console (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET 18m


UL E125349 − 250V Control CABLE (MPH A7 − I/O Console)  SUD−D (37p/F) (MPH A6 A3−Distribution Board) SUD−D (37p/F)
(2126152) 24VAC XJ1 Connector XJ12 Connector
9.1" 0.15 mm

9 2151729 SG100 MPH 50/65/80 CABINET  SG100 BUCKY STAND  18m


UL 2464 − 300V Power Supply (MPH A1−AC Distribution Panel)  (I/O PANEL) 
MIS 27225 240VAC CABLE F11 − 2 Tinned Wire TS2 − L Tinned Wire
8" 0.2 mm F10 − 2 TS2 − N
Ground Plug GND STUD

10 2145641 SG 60/100 MPH 50/65/80 CABINET  Female Connector SG 60 or SG100 BUCKY STAND  18m
BUCKY CABLE (MPH A6 A4 Wall Bucky Board)  (9p) (I/O PANEL) 
MIS 27221 UL 2464 − 300V XJ3 J42 AMP_(M26p/F)
240/110 (USA)VAC Connector
8" 0.2 mm

11 2167144 TABLE MPH 50/65/80 CABINET  COMPAX 40E TABLE  18m


UL 2463 − 600V Power Supply (MPH A1−AC Distribution Panel)  (I/O PANEL) 
MIS 27227 480VAC CABLE F18 − 2 Tinned Wire TB1 (PHASE) Tinned Wire
9.1" 0.15 mm F19 − 2 TB1 (NEUTRAL)
Ground Plug TB1 (GROUND)

12 2145602 C40E TABLE MPH 50/65/80 CABINET  COMPAX 40E TABLE  18m
BUCKY CABLE (MPH A6 A5 Table Tomo Board)  (I/O PANEL) 
MIS 27228 UL 2463 − 300V XJ10 Tinned Wire J42 AMP_(M26p/F)
240/110 (USA)VAC Connector
8" 0.2 mm

13 2145651 TABLE MPH 50/65/80 CABINET COMPAX 40E TABLE  18m


UL 2464 − 300V ION CHAMBER (MPH A6 A3−Distribution Board) AMP_(M14p/M) (I/O PANEL)  AMP_(M14p/F)
MIS 27229 200/15 (USA)VDC CABLE XJ15 Connector J40 Connector
8" 0.2 mm

14 2165551 Collimator Lamp RS 85 TUBE STAND  TOWER  COMPAX 40E TABLE  18m
UL 2464 − 300V CABLE (Terminals Strip)  (FUSE PANEL) 
MIS 27293 24VAC X3 Tinned Wire J12 Molex Connector
7.1 mm

15 2165550 SID Pot. & RS 85 RS 85 TUBESTAND TOWER COMPAX 40E TABLE  18m
MIS 27292 UL 2464 − 300V Brakes Control (Termnal Strip and Amp Connector) Tinned Wire (I/O PANEL)  SUD−D (25p/F)
24VDC CABLE X3 & X4 & J21 Connector
9.1 mm AMP Connector

INTERNAL Supplied by Pausch LOCK CABLE RS85 TUBE STAND  TOWER  RS85 TUBE STAND  ARM 
UL 1007 − 300V (Terminals Strip)  ULTRANET SM / SA Collimator SUD−D (37p/F)
MIS 27303 24VAC X3 Tinned Wire (rear PANEL)  Connector
12.6" 0.4 mm J / P2

119
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

16 2145642 SG 60/100 MPH 50/65/80 CABINET SG60/100 BUCKY STAND  18m


UL 2464 − 300V ION CHAMBER (MPH A6 A3−Distribution Board) AMP_(M14p/M) (I/O PANEL)  AMP_(M14p/F)
MIS 27222 200/15 (USA)VDC CABLE XJ16 Connector J40 Connector
8" 0.2 mm

17 2145656 COMPATIBILITY SG 60 or SG100 BUCKY STAND  COMPAX 40E TABLE  18m


UL 2464 − 300V CABLE (I/O PANEL)  SUD−D (9p/F) (I/O PANEL)  SUD−D (15p/F)
MIS 27224 24VDC J45 Connector J44 Connector
4.5" 0.2 mm

18 2145644 SG 60/100 MPH 50/65/80 CABINET SG60/100 BUCKY STAND  18m


UL 2464 − 300V Automat. Tray (MPH A6 CL1 − RAFL−S, I/O Panel) AMP (I/O PANEL)  SUD−D (15p/M)
MIS 27223 10VDC CABLE P3 Connector J41 Connector
8" 0.2 mm

20 2165546 COLLIMATOR RS85 TUBE STAND  ARM  MPH 50/65/80 CABINET 18m
UL 2464 − 300V CONTROL ULTRANET SA Collimator  (MPH A6 SC2−INTERFACE) AMP 37
MIS 27236 24VDC CABLE (rear PANEL)  AMP  CONT RECP
10.8mm J / P1 Connector J201 Connector

21 2145655 SID DETENT COMPAX 40E TABLE  MPH 50/65/80 CABINET 18m
UL 2464 − 300V CABLE (I/O PANEL)  SUD−D (15p/F) (MPH A6 SC2−INTERFACE) SUD−D (15p/F)
MIS 27231 24VDC J23 Connector J202 Connector
8" 0.2 mm

22 2145618 C40E COMPAX 40E TABLE  MPH 50/65/80 CABINET 18m


UL 2464 − 300V Automat. Tray (I/O PANEL)  SUD−D (15p/M) (MPH A6 CL1 − RAFL−S, I/O Panel)  AMP
MIS 27230 10VDC CABLE J41 Connector P2 Connector
8" 0.2 mm

23 2165547 RS 85 DETENT RS85 TUBE STAND  TOWER  MPH 50/65/80 CABINET 18m 22m
UL 2464 − 300V SID CABLE (Terminals Strip / Connector)  Tinned Wire and (MPH A6 SC2−INTERFACE) SUD−D (25p/F)
MIS 27237 24VDC X3 and X4 AMP Connector  Connector
10.8mm J301

INTERNAL 2145659 SAS Interface MPH 50/65/80 CABINET MPH 50/65/80 CABINET  0.7m
UL 2464 − 300V CABLE (MPH A6 CL1 − RAFL−S, I/O Panel) AMP (MPH A6 A8 Extension)  AMP
MIS 27232 24VDC P10 Connector XJ5 Connector
8" 0.2 mm

INTERNAL 2154537 Power Supply MPH 50/65/80 CABINET MPH 50/65/80 CABINET  1.5m
UL 2464 − 300V CABLE (MPH A6 CL1 − RAFL−S, I/O Panel) (MPH A6 SC3) 
MIS 27233 240VAC P1 Tinned Wire TB4−5/6 & GND Tinned Wire
8" 0.2 mm

INTERNAL 2152750 Power Supply MPH 50/65/80 CABINET  MPH 50/65/80 CABINET  1.5m
UL 2464 − 300V CABLE (MPH A6 SC3)  (MPH A1−AC Distribution Panel) 
MIS 27240 240VAC EMI FILTER IEC Connector F13 − 2 Tinned Wire
8" 0.2 mm F12 − 2
Ground Plug

120
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

2171350 RS85 TUBE STAND  TOWER  RS85 POWER SUPPLY 


MIS 27304 Supplied by Pausch 24 VDC (Terminals Strip)  (Terminals Strip) 
UL 2464 − 300V CABLE X3 Tinned Wire L + and L − Terminals Tinned Wire
24VDC
7.1 mm

1.1 2138905 POWER MPH 50/65/80 CABINET LINE SAFETY SWITCH 9m


X UL 1019 − 600V SUPPLY (MPH A1 A1−AC Distribution Panel) Ring Terminal  Ring Terminal
MIS 27235 480VAC CABLE TS − GND GROUND
33.6" 1.6 mm

2.1 45553701 Console (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET  18m
UL 1019 − 600V Ground CABLE (MPH A7 − I/O Console)  (MPH A1−AC Distribution Panel) 
(45553701) 0VDC GND Terminal Ground Plug Tinned Wire
7.2" 0.3 mm

3.1 2140362 TABLE COMPAX 40E TABLE  MPH 50/65/80 CABINET  18m
UL 1015 − 600V Ground CABLE  (MPH A1−AC Distribution Panel) 
MIS 27300 0VDC TB1−GND Tinned Wire Ground Plug Tinned Wire
4.5" 0.2 mm

4.1 2140362 GROUND RS85 TUBE STAND  TOWER  MPH 50/65/80 CABINET  18m
UL 1015 − 600V CABLE  Terminal (MPH A1−AC Distribution Panel) 
MIS 27302 0VDC GND STUD Ground Plug Tinned Wire
4.5" 0.2 mm

5.1 2140362 SG 60/100 SG 100 BUCKY STAND MPH 50/65/80 CABINET  18m
UL 1015 − 600V Ground CABLE (I/0 PANEL) (MPH A1−AC Distribution Panel) 
MIS 27301 0VAC GND STUD Terminal Ground Plug Tinned Wire
4.5" 2.2 mm

6.1 2105836 PRINTER−X PRINTER−X  MPH 50/65/80 CABINET  18m


SUPPLIED with Ground CABLE (MPH A9 Module)  (MPH A1−AC Distribution Panel) 
(W19) PRINTER−X GND (Ground) Terminal Ground Plug Tinned Wire
UL 1011 − 600V
0VDC
4.8" 0.2 mm

7.1 2165548 RS85 MPH 50/65/80 CABINET RS85 POWER SUPPLY  18m
UL 2464 − 300V Power Supply (MPH A1−AC Distribution Panel) (Terminals strip) 
MIS 27238 240VAC CABLE Ground Plug Tinned Wire PE1 (GND Screw) Terminal
7.1 mm

8.1 2183955 RS85 RS 85 TUBESTAND  TOWER  X−RAY TUBE  3m


UL 1015 − 600V X−Ray Tube  
MIS 27305 0VDC Ground CABLE Ground Plug Terminal Ground Plug Terminal
4.5 " 0.2 mm

121
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

TABLE 7−3
SYSTEM COMPAX 40E AND XT SUSPENSION−TOMO LINK, WITH MPH 50/65/80 AND SG−60/100 (ILLUSTRATIONS 3.7, 3.8 AND 3.9)

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

1 B2055BE VHV CABLE MPH 50/65/80 CABINET XT SUSPENSION  X−Ray TUBE 18m
UL E4413 − 75kV (MPH A2 − HV TANK)  
MIS 107−A 75kV  TUBE I ( + ) TERMINAL ANODE TERMINAL
19.94 ± 0.38mm

1 B2055BE VHV CABLE MPH 50/65/80 CABINET XT SUSPENSION  X−Ray TUBE 18m
UL E4413 − 75kV (MPH A2 − HV TANK)  
MIS 107−C 75kV  TUBE I ( − ) TERMINAL CATHODE TERMINAL
19.94 ± 0.38mm

2 2128497 STATOR MPH 50/65/80 CABINET XT SUSPENSION  X−Ray TUBE  18m


CABLE (MPH A1 A2−Stator Connector Panel)  (Terminals Strip) 
UL 2516 − 600V ST1 Connector AMP TERMINAL TERMINAL
MIS 10680 240VAC
12.19 ± 0.51mm

3 2104204 PRINTER−X MPH 50/65/80 CABINET PRINTER−X  18m


Power Supply (MPH A6 A3−Distribution Board) (MPH A9 Module) 
(W34) UL 1011 − 600V CABLE XJ1 Tinned Wire TS TERMINAL
230VAC
9.7" 0.4 mm

4 46−329288G2 XT Suspension MPH 50/65/80 CABINET XT SUSPENSION  18m


Power Supply (MPH A1−AC Distribution Panel) (A6 A1 module) 
MIS10707 UL/CSA − 300V CABLE F15−2 Tinned Wire J100 Connector AMP
240VAC F14−2
9.65" 0.51 mm Ground Plug

5 2138905 MPH POWER MPH 50/65/80 CABINET LINE SAFETY SWITCH 9m


X SUPPLY (MPH A1 A1−AC Distribution Panel) 
MIS27235 UL 1019 − 600V CABLE P1 Ring Terminal Phase 1 (R) Ring Terminal
480VAC P2 Phase 2 (S)
33.6" 1.6 mm P3 Phase 3 (T)
GND GROUND

6 2125971 PRINTER−X (PU3) MADRID CONSOLE  PRINTER−X  18m


UL E125349 Control CABLE (MPH A7 − I/O Console) SUD−D (37p/F) (MPH A9 Module)  SUD−D (37p/F)
(2125971) 24VAC XJ4 Connector MPH A9 Connector
2.7" 0.15 mm

7 2126153 Program−X (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET 18m


UL E125349 − 250V Control CABLE (MPH A8 − Program−X)  SUD−D (37p/F) (MPH A6 A3−Distribution Board) SUD−D (37p/M)
(W32) 24VAC XJ1 Connector XJ10 Connector
9.1" 0.15 mm

122
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


Option Reference CABLE CODE DESIGNATION CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

8 2126152 Console (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET 18m


UL E125349 − 250V Control CABLE (MPH A7 − I/O Console)  SUD−D (37p/F) (MPH A6 A3−Distribution Board) SUD−D (37p/F)
(2126152) 24VAC XJ1 Connector XJ12 Connector
9.1" 0.15 mm

9 2151729 SG100 MPH 50/65/80 CABINET  SG100 BUCKY STAND  18m


UL 2464 − 300V Power Supply (MPH A1−AC Distribution Panel)  (I/O PANEL) 
MIS 27225 240VAC CABLE F11 − 2 Tinned Wire TS2 − L Tinned Wire
8" 0.2 mm F10 − 2 TS2 − N
Ground Plug GND STUD

10 2145641 SG 60/100 MPH 50/65/80 CABINET  Female Connector SG 60 or SG100 BUCKY STAND  18m
UL 2464 − 300V BUCKY CABLE (MPH A6 A4 Wall Bucky Board)  (9p) (I/O PANEL) 
MIS 27221 240/110 (USA)VAC XJ3 J42 AMP_(M26p/F)
8" 0.2 mm Connector

11 2167144 TABLE MPH 50/65/80 CABINET  COMPAX 40E TABLE  18m


UL 2463 − 600V Power Supply (MPH A1−AC Distribution Panel)  (I/O PANEL) 
MIS 27227 480VAC CABLE F18 − 2 Tinned Wire TB1 (PHASE) Tinned Wire
9.1" 0.15 mm F19 − 2 TB1 (NEUTRAL)
Ground Plug TB1 (GROUND)

12 2145602 C40E TABLE MPH 50/65/80 CABINET  COMPAX 40E TABLE  18m
BUCKY CABLE (MPH A6 A5 Table Tomo Board)  (I/O PANEL) 
MIS 27228 UL 2463 − 300V XJ10 Tinned Wire J42 AMP_(M26p/F)
240/110 (USA)VAC Connector
8" 0.2 mm

13 2145651 TABLE MPH 50/65/80 CABINET COMPAX 40E TABLE  18m


UL 2464 − 300V ION CHAMBER (MPH A6 A3−Distribution Board) AMP_(M14p/M) (I/O PANEL)  AMP_(M14p/F)
MIS 27229 200/15 (USA)VDC CABLE XJ15 Connector J40 Connector
8" 0.2 mm

14 46−329341G2 SID CABLE XT SUSPENSION  COMPAX 40E TABLE  18m


UL 2464 − 300V (3A1 module)  SUD−D (37p/F) (I/O PANEL)  SUD−D (37p/F)
MIS 10705 200/15 (USA)VDC J101 Connector J43 Connector
8" 0.2 mm

15 46−329291G1 (Manual System) XT SUSPENSION  XT SUSPENSION  4m


UL 2464 − 300V COLLIMATORCA- (3A1 module)  SUD−D (37p/F) (Ultranet SM Collimator) AMP_ (26p/F)
MIS 10706 24VDC BLE J1 Connector J/P2 Connector
12.70" 0.51 mm

16 2145642 SG 60/100 MPH 50/65/80 CABINET SG100 BUCKY STAND  18m


UL 2464 − 300V ION CHAMBER (MPH A6 A3−Distribution Board) AMP_(M14p/M) (I/O PANEL)  AMP_(M14p/F)
MIS 27222 200/15 (USA)VDC CABLE XJ16 Connector J40 Connector
8" 0.2 mm

123
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100


Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

15 46−329291G1 (Auto System) XT SUSPENSION  AMP 37 XT SUSPENSION  AMP_ (26pos/F)/ 4m


UL 2464 − 300V COLLIMATORCA- (3A1 module)  CONT RECP (Ultranet SA Collimator) (14pos/F) Connec- −
MIS 10706 24VDC BLE J1 Connector J/P1−J/P2 tors
12.70" 0.51 mm

17 2145656 COMPATIBILITY SG 60 or SG100 BUCKY STAND  COMPAX 40E TABLE  18m


UL 2464 − 300V CABLE (I/O PANEL)  SUD−D (9p/F) (I/O PANEL)  −
MIS 27224 24VDC J45 Connector J44
4.5" 0.2 mm

18 2145644 SG 60/100 MPH 50/65/80 CABINET SG60/100 BUCKY STAND  18m


UL 2464 − 300V AUTOMAT. TRAY (MPH A6 CL1 − RAFL−S, I/O Panel) (I/O PANEL)  SUD−D (15p/M) −
MIS 27223 10VDC CABLE P3 J41 Connector
8" 0.2 mm

20 46−329341G2 SID CABLE XT SUSPENSION  AMP 37 MPH 50/65/80 CABINET AMP 37 24m
UL 2464 − 300V (3A1 module)  CONT RECP (MPH A6 SC2−INTERFACE) CONT RECP −
MIS 10705 24VDC J101 Connector J201 Connector
12.70" 0.51 mm

21 2145655 SID DETENT COMPAX 40E TABLE  MPH 50/65/80 CABINET 18m
UL 2464 − 300V CABLE (I/O PANEL)  (MPH A6 SC2−INTERFACE) −
MIS 27231 24VDC J23 J202
8" 0.2 mm

22 2145618 C40E COMPAX 40E TABLE  MPH 50/65/80 CABINET 18m


UL 2464 − 300V Automat. Tray (I/O PANEL)  (MPH A6 CL1 − RAFL−S, I/O Panel) −
MIS 27230 10VDC CABLE J41 P2
8" 0.2 mm

INTERNAL 2145659 SAS INTERFACE MPH 50/65/80 CABINET MPH 50/65/80 CABINET 
UL 2464 − 300V CABLE (MPH A6 CL1 − RAFL−S, I/O Panel) (MPH A6 A8 Extension)  −
MIS 27232 24VDC P10 XJ5 0.7m
8" 0.2 mm

INTERNAL 2154537 Power Supply MPH 50/65/80 CABINET MPH 50/65/80 CABINET  1.5m
UL 2464 − 300V CABLE (MPH A6 CL1 − RAFL−S, I/O Panel) (MPH A6 SC3) 
MIS 27233 240VAC P1 Tinned Wire TB4−5/6 & GND Tinned Wire
8" 0.2 mm

INTERNAL 2152750 Power Supply MPH 50/65/80 CABINET  MPH 50/65/80 CABINET  1.5m
UL 2464 − 300V CABLE (MPH A6 SC3)  (MPH A1−AC Distribution Panel) 
MIS 27240 240VAC EMI FILTER IEC Connector F13 − 2 Tinned Wire
8" 0.2 mm F12 − 2
Ground Plug

124
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Option Reference CABLE CODE DESIGNATION CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

25 2132121 (Manual System) COMPAX 40E TABLE  SUB−D (25 p/f) COMPAX 40E TABLE  SUB−D (25 p/f) 24m (−8) 18m (−6)
X UL CL3 − 300V XT−TOMO (Control Electronics TL1/A1)  connector (I/O PANEL)  connector 30m (−10)
MIS 11212 24VDC CABLE J106 J22
11.7 +/− 0.2mm

25 2132121−8 (Auto System) COMPAX 40E TABLE  MPH 50/65/80 CABINET 


XT−TOMO (Control Electronics TL1)  (MPH A6 SC2−INTERFACE)  −
MIS 11212 CABLE J106 J200 24m

26 2132261 DRIVE UNIT Con- XT SUSPENSION  SUB−D (15 p/f) COMPAX 40E TABLE  SUB−D (15 p/f) 18m (−6)
UL CL3 − 300V trol CABLE (Drive Unit TL3/A1)  connector (Control Electronics TL1/A1)  connector 24m (−8) 30m (−10)
X MIS 11214 24VDC J101 J101
9.1 +/− 0.2mm

27 2124056 DRIVE UNIT XT SUSPENSION  3 Pin Mate & Lock COMPAX 40E TABLE  3 Pin Mate & Lock
X UL SJT − 300V Power Supply (Drive Unit TL3/A1)  Hsg − (Control Electronics TL1/A1)  Hsg − 24m (−2) 60 ft (−1)
MIS 11215 120VAC CABLE J100 Female Pins con- J100 Female Pins con- (80ft) 100ft (−3)
9.7 +/− 0.2mm nector nector

28 2159597−3 TOMO Control MPH 50/65/80 CABINET LMI 9 pins COMPAX 40E TABLE  SUB−D (25 p/f) 24m (−4)
X UL CL3 − 300V CABLE (MPH A6 A5 Table Tomo Board)  connectors (Control Electronics TL1/A1)  connector 18 30m (−5)
MIS 11411 24VDC XJ12 & XJ14 J105
9.1 +/− 0.2mm

INTERNAL REA 00074 TOMO−TOWER COMPAX 40E TABLE  SUB−D (15 p/f) COMPAX 40E TABLE  BURNDY
X (INTERNAL) Control CABLE (Control Electronics TL1/A1)  connector (Quick Disconnect connector)  connector 2.6m −
MIS Y J102 Jx

INTERNAL REA 00072 TOMO−CON- COMPAX 40E TABLE  SUB−D (15 p/f) COMPAX 40E TABLE  SUB−D (15 p/f)
X UL CL3 − 300V SOLE (Control Electronics TL1/A1)  connector (Tomo−Console Control TL4/A1)  90° −
MIS Z 24VDC Control CABLE J103 J2 connector 3.6m
9.1 +/− 0.2mm

X INTERNAL REA 0073 TOMO Power COMPAX 40E TABLE  IEC Power con- COMPAX 40E TABLE  1/4in push on Ter- 1.5m
(INTERNAL) Supply CABLE (Control Electronics TL1) nector (EMI Filter)  minals −
MIS aa J110 Female FL1

X INTERNAL REA 0075 TOMO BUCKY COMPAX 40E TABLE  3 Mate & Lock COMPAX 40E TABLE  6 Mate & Lock 1.8m
(INTERNAL) BLADE RELEASE ( CONTROL ELECTRONICS TL1/A1)  HSG − Female (Power Board)  HSG − Female −
MIS bb CABLE J107 pins connector J8 pins connector

X Cable at- REA 0054 TOMO TOWER TOMO TOWER  SUB−D (15 p/f) COMPAX 40E TABLE  BRUNDY connec- 1.0m
tached to UL CL3 − 300V FIXED CABLE (Control Board TL2/A1)  90° (Frame of Table Top)  tor −
Tower 24VDC J1 connector Quick Disconnect Connector
9.1 +/− 0.2mm

125
System Service Manual − SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE with MPH 50/65/80 (including XT, RS 85 and Tomo−Link)

REV 9 DIRECTION 2154260−100

Option Reference CABLE CODE CONNECTION CONNECTOR CONNECTION CONNECTOR TOTAL AVAILABLE
DESIGNATION
=X MIS MAP NUMBER FROM TYPE TO TYPE LENGTH LENGTH

1.1 2138905 Power Supply MPH 50/65/80 CABINET LINE SAFETY SWITCH 9m
X UL 1019 − 600V CABLE (MPH A1 A1−AC Distribution Panel) Ring Terminal  Ring Terminal −
MIS 27235 480VAC TS − GND GROUND
33.6" 1.6 mm

2.1 45553701 Console (PU3) MADRID CONSOLE  MPH 50/65/80 CABINET  18m
UL 1019 − 600V Ground CABLE (MPH A7 − I/O Console)  (MPH A1−AC Distribution Panel)  −
(45553701) 0VDC GND Terminal Ground Plug Tinned Wire
7.2" 0.3 mm

3.1 2140362 TABLE COMPAX 40E TABLE  MPH 50/65/80 CABINET  18m
UL 1015 − 600V Ground CABLE  (MPH A1−AC Distribution Panel)  −
MIS 27300 0VDC TB1−GND Tinned Wire Ground Plug Tinned Wire
4.5" 0.2 mm

4.1 46−316922G2 GROUND XT SUSPENSION  MPH 50/65/80 CABINET  100 FT


SUPPLIED with XT CABLE (CHASSIS)  Terminal (MPH A1−AC Distribution Panel)  Terminal −
MIS 10708 SUSPENSION GND STUD Ground Plug
UL 8AWG − 600V
0VDC
0.260 ± 0.005in

5.1 2140362 SG 60/100 SG 100 BUCKY STAND MPH 50/65/80 CABINET  18m
UL 1015 − 600V Ground CABLE (I/0 PANEL) (MPH A1−AC Distribution Panel)  −
MIS 27301 0VAC GND STUD Terminal Ground Plug Tinned Wire
4.5" 2.2 mm

6.1 2105836 PRINTER−X PRINTER−X  MPH 50/65/80 CABINET  18m


SUPPLIED with Ground CABLE (MPH A9 Module)  (MPH A1−AC Distribution Panel)  −
(W19) PRINTER−X GND (Ground) Terminal Ground Plug Tinned Wire
UL 1011 − 600V
0VDC
4.8" 0.2 mm

X 7.1 45433617 TOMO−LINK COMPAX 40E TABLE  COMPAX 40E TABLE  1.5m
UL1007 −300V Ground CABLE  Eye−let (TLK Electronics)  Eye−let −
(45433617) 0VDC TB1−GND GND
3.2 +/− 2mm

8.1 46−216211G10 XT XT SUSPENSION  X−RAY TUBE  4m


SUPPLIED with XT X−Ray Tube (CHASSIS)  
(46−216211 SUSPENSION Ground CABLE GND STUD Terminal Ground Plug Terminal
G10) UL 8AWG − 600V
0VDC
0.260 ± 0.005in

126
GE Healthcare

GE Medical Systems: Telex 3797371


P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)
GE Medical Systems  Europe: Telex 698626
283, rue de la Miniére, B.P. 34, 78533 Buc Cedex
France

Das könnte Ihnen auch gefallen