Service Manual
CONTENTS
1. INTRODUCTION ________________________________________________________________________ 7
2.4 LABELS_____________________________________________________________________________________ 17
6. INSTALLATION ________________________________________________________________________ 24
7.2.7 TIME LIMIT / TUBE LIMIT ALARM / EEROR RESET BUTTON __________________________ 41
9. SPECIFICATION _______________________________________________________________________ 73
REVSON HISTORY
ADVISORY SYMBOLS
The following advisory symbols are used throughout this manual.
Their application and meaning are described below.
WARNING
Warning symbol used to indicate a potential hazard lead operators, service personnel to serious injury, death
or radiation exposure.
CAUTION
Caution symbol used to indicate a potential hazard lead operators, service personnel to injury or damage of
equipment.
NOTE
Note symbol used to indicate important information need to proper use and right operation of equipment.
NOTE
KEEP THIS SERVICE MANUAL WITH THE EQUIPMENT AT ALL TIMES, AND REVIEW THE
IMPORTANT INFORMATION WHENEVER REQUIRED.
THE STATISTICS AND SPECIFICATIONS OF THIS UNIT AND MANUAL CAN BE MODIFIED
WITH OR WITHOUT NOTIFICATION FOR THE IMPROVEMENT OF PERFORMANCE AND
SAFETY.
1. INTRODUCTION
The ANYVIEW-500R Innovation mobile C-arm features excellent x-ray Image, quality, accuracy, reproducibility
and long-term stability. Fast kV rising time minimize unnecessary patient dose and supply superior Image
quality, fast Image response.
The operator control console is designed to be simple and user-friendly and one can select or change x-ray
parameters easily using a large soft membrane switch.
Automatic Calibration function will minimize calibration time and set correct calibration value. Computer
program of ANYVIEW-500R offers graphic waveform of kV, mA and data of x-ray exposure. So, you don’t
have to perform manual calibration and measurement equipment are unnecessary. Naturally, ANYVIEW-500R
supports Automatic, Semi-auto and Manual Calibration functions.
Adaptable calibration function keeps up long-term stability and accuracy by trimming calibration data
whenever x-ray exposed.
Remote Diagnosis Software can make report about information of system, user, service and error by the USB
connection between PC and control console, and transfer report to ECOTRON via internet.
The mobile C-arm Anyview-500R is intended for use to produce X-ray images of patient during medical
operations. The Anyview-500R system consists of a Main power board and Main control board, Control
console, SMPS, X-ray generator, Image intensifier.
X-ray generator contains the HT tanks and control circuits, the filament drivers, FET drivers, Collimator control
circuits and interface connections to the Main control board.
This product, Fluoroscopic x-ray system, is radiation medical equipment only used by professional radiologists.
This product is designed to provide fluoroscopic and spot film images of the patient during diagnostic and
interventional procedures.
This system can be applied in emergency room, operation room, cast room or etc. of hospital.
Indication
Bone fracture, Interventional procedures
NOTE
Side Effects
Operating X-ray on a patient for a lengthy period of time may cause the following side effects.
Dizziness
Nausea
Vomiting
Diplopia
Skin Numbness
Masticatory Atonia
Transient leg paresthesia
Contraindication
In case of taking continuously pulsating X-rays on the parts where artificial cardiac pacemaker or
implantable defibrillator is implanted, such devices may suffer from inappropriate malfunction.
Therefore direct X-ray on these parts should be avoided, and minimum output should be used for a
short period of time.
In case of taking X-rays of infants, genital protective gear should be used on the genitalia.
The subject and the conductor should be at least 20cm away from the equipment when taking the X-
ray.
1.2 COMPOSITION
This product is composed of the X-ray component part, the display image part, accessories part, and options.
Detail information of each component follows below.
(3) Accessories
Operator’s Manual
Hand switch
Foot switch
Laser
(4) Option
DIS(Digital Interface System) with DICOM
Dose Area Meter
Air Kerma
Disinfected covers
NOTE
PLEASE ENSURE THAT YOU READ THE WARNING NOTICES BEFORE USING THE
EQUIPMENT
2.1 SAFETY
X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending over
periods of many months or even years. X-ray operators should avoid any exposure to the primary beam
and take protective measures to safeguard against scatter radiation. Scatter radiation is caused by any object
in the path of the primary beam and may be of equal or less intensity than the primary beam that exposes the
film.
No practical design can incorporate complete protection for operators or service personnel who do not take
adequate safety precautions. Only authorized and properly trained service and operating personnel
should be allowed to work with this X-ray system equipment. The appropriate personnel must be made
aware of the inherent dangers associated with the servicing of high voltage equipment and the danger of
excessive exposure to X-ray radiation during system operation.
Wear protective clothing. Protective aprons and gloves with an equivalent of a minimum of 1/64”(0.35mm)
of lead are recommended.
To protect the patient against radiation, always use radiation protection accessories in addition to devices
which are fitted to the X-ray system
Keep as large a distance as possible away from the object being exposed and the X-ray tube assembly.
Never operate this X-ray system in areas where there is a risk of explosion. Detergents and disinfectants,
including those used on patients, may create explosive mixtures of gases. Please observe the relevant
regulations.
Do not operate the x-ray system in direct sunlight or near any heat sources.
Do not operate the x-ray system near strong magnetic fields(microwave ovens, speakers, etc), and avoid
routing the x-ray system near these devices.
The x-ray system must be operated in locations that are clean(free of excess dust, dirt, debris, etc),
stable(free of vibration).
Only trained maintenance staff may remove the covers of the x-ray system.
Always select a focal spot to skin distance as large as possible to keep the absorbed dose for the patient
as low as reasonably possible.
WARNING
WARNING
WARNING
THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE
EXPSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.
WARNING
PROPER USE AND SAFE OPERATING PRACTICES WITH RESPECT TO X-RAY SYSTEMS
ARE THE RESPONSIBILITY OF THE USERS OF SUCH X-RAY SYSTEMS.
WARNING
WARNING
CAUTION
CAUTION
DO NOT EXCEED THE TUBE MAXIMUM POERATING LIMITS SHOWN IN THE X-RAY
TUBE DATA SECTION AT THE END OF THE POERATOR’S MANUAL.
INTENDED LIFE AND RELIABILITY WILL NOT BE OBTAINED UNLESS X-RAY SYSTEMS
ARE POERATED WITHIN PUBLISHED SPECIFICATION.
CAUTION
CAUTION
DO NOT CONNECT ANY OTHER ELECTRICAL DEVICE TO THE LAN CONNECTOR AND
THE USB CONNECTOR OF THE EQUIPMENT EXCEPT FOLLOWING ELECTRICAL
DEVICES.
WARNING
The ANYVIEW-500R X-ray system with CE marking complies with the European Council Directive concerning
Medical Devices. One of the harmonized standards of this Directive defines the permitted levels of
electromagnetic emission from this equipment and its required immunity from the electromagnetic emission of
other devise.
It is not possible, however to exclude with absolute certainty the possibility that other high frequency electronic
equipment, which is fully compliant to the EMC regulations, will not adverse effect the operation of this x-ray
system. If the other equipment has a comparatively high level of transmission power and is in close proximity
to the generator, these EMC concerns (the risk of interference) may as mobile telephones, cordless
microphones and other similar mobile radio equipment be restricted from the vicinity of this X-ray system.
ANYVIEW-500R needs special precautions regarding EMC and needs to be installed and use according to
the EMC information provided in this operation manual. For more information, refer to Appendix B.
● Cable List
H-20 5M NO Ecotron
POWER CABLE 6M NO Ecotron
PC POWER CALBE 3M NO Ecotron
HAND S/W CABLE 1.5M NO Ecotron
FOOT S/W CABLE 1.5M NO Ecotron
2.3 SYMBOLS
2.3.1 Symbols on the X-ray system
SYMBOL Description
Consult accompanying documents: This symbol advises the reader to consult the
accompanying documents.
High voltage warning: This symbol means high voltage parts included.
Warning: This symbol indicates not to put your weight on the packing box.
Disposal: For more information, see on Section 10.5 “THE END OF PRODUCT
LIFE”
SYMBOL Description
WARNING
CAUTION
THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE
EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.
2.4 LABELS
2.4.1 COMPOSITION LABEL
IDENTIFICATION LABEL- DIGITAL IMAGING SYSTEM
This label is attached to Digital Imaging System. (Example)
NOTE
- CIRCUIT DIAGRAMS
- COMPONENT PART LISTS
- CALIBRATION INSTRUCTIONS AND ETC.
Imaging Performance
NOTE
Internal X-ray system components will heat up during normal use of the x-ray system. This is similar to X-ray
tube heating during normal x-ray system operation. The amount of heat produced is proportional to the
product of kV, mA and time.
CAUTION
NOTE
Use the procedure when the generator has not been used for several days. This procedure provides for
exposures at medium power before the tube is used at maximum mA or kV values. This will reduce the
possibility of damaging the X-ray tube and high voltage components. No test setup is required.
For maximum stability and reliability, use the following techniques at start up:
5. APPLICABLE STANDARDS
The ANYVIEW-500R of X-ray system complies with the regulatory requirements and design standards in this
section as follows:
5.1 SAFETY
● IEC 60601-1:2015+A1:2012
● IEC 60601-1-6:2010+A1:2013
● IEC 62366:2007
● IEC 60601-2-28:2010
● IEC 60601-2-54:2009
5.2 EMC
● IEC60601-1-2:2007
● EN55011:2007+A2:2007GROUP 1 CLASS A
● EN61000-3-2:2000+A2:2005
● EN61000-3-3:1995+A1:2001
5.4 OTHERS
● ISO15223-1:2012 Medical devices -Symbols to be used with medical device labels, labeling and
information to be supplied
● IECTR60878:2003 Graphical symbols for electrical equipment n medical practice
● IEC60417-1:2012 Graphical symbols for use on equipment-part1:overview and application
● ENISO14971:2007 Medical devices - Application of risk management to medical devices
(ISO 14971:2007)
● ENISO13485:2003 Medical devices - Quality management systems-Requirements for regulatory
purposes
● ISO9001:2008 Quality management systems-Requirements
● 93/42/EEC:1993 Council Directive concerning medical devices
● EN1041:2008 Information supplied by the manufacturer with medical devices
● EN62304:2006 Medical device software-Software lifecycle processes
6. INSTALLATION
CAUTION
KEEP IN MIND THAT THE C-ARM SYSTEM MAY ROLL AWAY ON UNEVEN FLOORS.
LEAVE THE C-ARM STAND BEHIND ONLY ON EVEN FLOORS AND LOCK ITS BRAKES.
RELEASE THE PARKING BRAKE ONLY TO MOVE OR POSITION THE C-ARM SYSTEM.
Levers in upper position: you can move the monitor cart to the left and right side.
Angulation Brake
Panning Brake
Horizontal
Movement
Brake
Orbital
Steering
Rotation Brake
Handle
Vertical Movement
Parking Brake
Up/Down Switch
CAUTION
DO NOT MOVE THE SYSTEM OVER FLOORS WITH AN INCLINATION OF MORE THAN 10°
FROM LEVEL DURING TRANSPORT. THIS APPLIES TO BOTH THE C-ARM STAND AND
THE MONITOR CART.
EXERCISE EXTREME CAUTION WHEN MOVING THE C-ARM STAND AND THE MONITOR
CART OVER ROUGH SURFACES SUCH AS TILE FLOORING, PAVEMENT, ASPHALT OR
CARPET. TAKE CARE THAT THE CABLE GUARDS DO NOT DRAG AND THE WHEELS DO
NOT CATCH OR TILT CAUSING DAMAGE TO THE EQUIPMENT.
CAUTION
BEFORE ROTATING THE C-ARM, MAKE SURE THAT IT DOES NOT COLLIDE WITH ANY
PERSONS OR OBJECTS.
CAUTION
BEFORE MOVING THE C-ARM UP OR DOWN, MAKE SURE THAT IT DOES NOT COLLIDE
WITH ANY PERSONS OR OBJECTS.
CAUTION
BEFORE MOVING THE C-ARM BACK AND FORTH, MAKE SURE THAT IT DOES NOT
COLLIDE WITH ANY PERSONS OR OBJECTS.
Monitor Arm
Plug the foot switch cable connector into the foot switch socket located at the front of the C-arm system
and lock the connector.
Make sure that a suitable supply voltage is available and that the socket-outlet is properly grounded and
fused.
Connect the C-arm system and Monitor cart power cable plug to the power supply.
Grid cover
Grid
Plastic Bolt (4ea)
CAUTION
YOU MUST USE STANDARD BOLT WHEN INSTALL X-RAY GRID AND GIRD COVER.
YOU MUST INSTALL GRID TO MATCH ANODE DIRCTION OF X-RAY TUBE AS SHOWN
ABOVE.
Both the C-arm system and the monitor cart have power switches for witching the system on.
You must turn on first C-arm system power switch before turn on monitor cart power switch.
You must wait 20 seconds to execute DIS icon ( ) after turn on the C-arm system.
NOTE
DO NOT PLUG ANY USB STICK INTO THE USB PORT UNTIL THE SYSTEM HAS FULLY
COMPLETED ITS POWER-UP SEQUENCE AND THE START SCREEN IS DISPLAYED.
NOTE
DUE TO BACKGROUND RADIATION, THE AIR KERMA, AIR KERMA RATE OF DOSE
AREA PRODUCT DISPLAY MAY INDICATE SOME SMALL VALUE AFTER POWER-UP OF
THE SYSTEM.
NOTE
DURING POWER-UP, DO NOT TOUCH EITHER OF THE CONTROL PANELS. THE SYSTEM
CAN STILL BE CONTROLLED VIA THE CONTROL PANELS.
NOTE
PRESS THE EMERGENCY STOP BUTTON ONLY IF THIS CANNOT BE AVOIDED, SINCE
UNSAVED PATIENT DATA AND IMAGES WILL BE LOST.
The EMERGENCY STOP button is locked in this position. All electrical functions of the system are disabled.
Any unsaved patient data and/or images are lost.
You must turn on first C-arm system power switch and turn on monitor cart power switch.
Initial status
When powered on, the startup screen (figure 6.8) is displayed on the SEGMENT. (During about 5sec and not
allowed input with key)
After 5sec, Initial display will be changed like as figure 6.8 and displayed normal fluoroscopy offset condition
“80KV, 1mA”
7. CONSOLE CONTROLS
7.1 Anyview-500R
X-RAY IMAGE STORE BUTTON (when you want to store x-ray image)
press
counterclockwise, press
To reset time limit, press ( ). This key is used to get out of the 'Time Limit'
status without 1 minute standby when a period of cumulative exposure time
exceeds 5 minutes but you cannot irradiate X rays.
This key is used to get out of the 'Tube Limit' status for emergent re-
radiography when the 'Tube Limit' is in action to protect the tube and therefore
you cannot radiograph.
To turn off beep sound, keep press laser button during 3sec.
To turn on beep sound, keep press laser button during 3sec.
CAUTION
CAUTION
THE HAND SWITCH BUTTON MUST REMAIN PRESSED DOWN TO COMPLETE X-RAYS
IRRADIATION.
CAUTION
THE HAND SWITCH BUTTON MUST REMAIN PRESSED DOWN TO COMPLETE X-RAYS
IRRADIATION.
CAUTION
CAUTION
Voltage Current
Fuse Rating
250VAC 25A
LED
I.I Size
D21 D22 D23
LED DESCRIPTION
LED DESCRIPTION
A FET BOARD
B VR COLLIMATOR BOARD
D XCU_MCU BOARD
LED DESCRIPTION
D18 X-Tube Filament Current Over Error Indicator
Warning
Display Description
Code
The heat sink is overheated on FET driver board.
0x00004000 29
Please wait for the heat sink to cool down.
: ERROR RESET
LED STATE
I.I Size
D21 D22 D23
(ON or OFF) (ON or OFF) (ON or OFF) 9 inch
LED STATE
D19 (ON or OFF)
LED STATE
D7 (Flashing ON or OFF)
D8 (Flashing ON or OFF)
D9 (Flashing ON or OFF)
LED STATE
D18 (ON or OFF)
D4 (ON or OFF)
D5 (ON or OFF)
D6 (ON or OFF)
D7 (Flashing ON or OFF)
LED STATE
D29 (ON or OFF)
6) kV / mA reference table
9 120kV / 6V 120kV / 6V
mAs TABLE
9 120kV / 6V 120kV / 6V
A. F-mode
1) 40kV/0.5mAs waveform
2) 80kV/0.5mAs waveform
3) 110kV/0.5mAs waveform
CH 1: kV 2V/div
CH 2 : mA 2V/div
B. R-mode
1) 50kV/1mAs/40mA waveform
2) 80kV/1mAs waveform
9. SPECIFICATION
9.1 SPECIFICATION TABLE
Model Name Anyview-500R
1) System power requirement
(1) Line voltage 220-230V~, 50/60Hz
2)X-ray Generator
(1)Type HFG INVERTER
(3)Fluoroscopy
- DS mode 10-30mA
(4)Radiography mode
3)X-ray tube
(1) X-ray tube type Rotating
(2)Max kV 125kV
(3)Focal spot
-Small 0.3mm
-Large 0.6mm
4)Collimator
(1)Type Motorized
(2)Opening/Closing 2 way
(3)Rotation 360°
(4)Iris Motorized
5)Image Intensifier
(1)Input field view size 9inch
(3)Central resolution(lp/cm) 52 / 58 / 68
6)CCD camera
(1)Type CCD
(2)Size 19inch
(3)q'ty 2
9)Reproducibility :
10)Accuracy
kV<(±1%+1kV), mA<(±4%+1mA), mAs<(±4%+0.1mAs)
11)Linearity
Coefficient of Linearity <0.01:CL=(X1-X2)/(X1+X2),where X is mR/mAs
12)Weight
(1)G-weight
(2)Net-weight
● RADIATION FIELD
- 25cm*25cm(Circle) at 1m (SID)
Momentary Apparent
Mains Voltage
Line Current Mains Resistance
220V 50A 0.08Ω
230V 48A 0.0833Ω
(1)C-ARM movement
PANNING 12.5
Width 800mm
(2)C-ARM rotation
Locking Manual
SID
Fluoro 1m
RAD 1m
(3)COLLIMATOR
Speed Rotation
(4) Accessories
OPERATING
Ambient temperature range 10°C to 40°C (50°F to 104°F)
Relative humidity range 30% to 75%, non-condensing
Atmospheric pressure range 800hPa to 1060hPa
This information is provided to help you establish safe operating conditions for both you and your X-
ray system.
WARNING
ECOTRON Co.,Ltd
4F & 5F, Hanshin IT Tower II, 47, Digital-ro 9-gil,
Geumcheon-Gu, Seoul 08511, Republic of Korea
TEL:+82-2-2025-3760, FAX:+82-2-2025-3764
E-mail: ecotron@ecotron.co.kr
Web-site: http://www.ecotron.co.kr
In USA
Contact DRGEM USA Inc.
TH
7018 NW 50 Terrace, Gainsville, Florida, 32653, USA
TEL:201-370-6672, FAX: 352-337-1271
E-mail: drgemusa@gmail.com
Obelis S.A
Bd. General Wahis 53, 1030 Brussels, Belgium / E-mail:mail@obelis.net
Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03
RAD mA :
1 1 Year
Should be within accuracy boundary of ±3%, with the voltage set by 40 kV and 110 kV.
Fluoro mA :
2 1 Year
Should have accurate 1mA step for 1~5mA, with the voltage range of 40 ~ 110 kV
RAD mAs :
3 Should be within accuracy boundary of ±5%, when the mAs value is between 0.4mAs and 1 Year
100 mAs.
RAD KV :
4 Should be within accuracy boundary of ±10%, when the kV value is between 40kV and 1 Year
110kV.
Fluoro KV:
5 1 Year
Should be within accuracy boundary of ±3%, with the voltage range of 40 ~ 110 kV
Optical Maintenance :
6 When dirt appears on the image, remove dirt on the camera lens with soft brush or ‘Kodak 1 Year
Lens Cleaner’.
DAP (Option):
7 5 Year
Should be calibrated for accuracy of DAP information.
12.5 CLEANING
Keep the machine clean when it gets dirt, especially after a surgery. To clean the main-body, turn off the
power and pull the plug out from the outlet. Scrub surfaces smoothly with soft fabric or gauze. Keep the
machine clean when it gets dirt, especially after a surgery.
To clean the main-body, turn off the power and pull the plug out from the outlet as pictures above. Do not allow
liquids and water enter inside the unit, to prevent from short circuits and corrosion of inner parts.
WARNING
Make sure the disconnection from the mains before cleaning the unit!
Clean the extra covers, carrying handles, foot switch, hand switch, breakers, monitors and the
accessories using a damp cloth.
Periodically and when necessary decontaminate the op panel, the carrying handles, the breakers, the
surfaces of monitors using a proper decontaminator.
WARNING
Do not keep near the magnetic stuffs and water pails etc. They might bring a serious
harm to the system.
If the X-ray system has completed its useful service life, local environmental
regulations must be complied with in regard to disposal of possible hazardous
materials used in the construction of the generator.
In order to assist with this determination, the noteworthy materials used in the
construction of this generator are itemized below:
ITEM
• Electrical insulating oil in HT tank. This is a mineral oil with trace additives (11 Liter)
• Solder (lead/tin).
• Electrical and electronic components: IC’s, transistors, diodes, resistors, capacitors, etc.
In the case of one of the following, however, a certain amount of service fees will be charged.
• Defect or damage found after the warranty period.
• Defect or damage in appearance which is not related to main function of the system.
• Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes.
• Damage resulting from either improper movement or inattention to the precautions.
• Damage resulting from repair or modification by someone other than ECOTRON or the authenticated by
ECOTORN Co., Ltd.
• Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes.
• Incidental or indirect loss caused during system manipulation.
Any or all defect or damages in appearance, which do not affect the main functions of the product are
not covered by this free of charge warranty.
Electromagnetic emission
The ANYVIEW-500R is intended for use in electromagnetic environment specified below. The customer or the
user of the ANYVIEW-500R should assure that it is used in such an environment.
Voltage fluctuations/flicker
emissions IEC6100-3-3
Electromagnetic immunity
The ANYVIEW-500R is intended for use in the electromagnetic environment specified below. The customer or
the user of the ANYVIEW-500R should assure that it is used in such an environment.
Electromagnetic environment –
IMMUNITY test IEC 60601 test level Compliance level
guidance
Floors should be wood, concrete or
Electrostatic
± 6 kV contact ceramic tile. If floors are covered
discharge (ESD)
± 8 kV air with synthetic material, the relative
IEC 61000-4-2
humidity should be at least 30 %.
<5 % UT
(>95 % dip in UT)
for 0,5 cycle Mains power quality should be that
of a typical commercial or hospital
Voltage dips, short 40 % UT environment. If the user of the [ME
interruptions and (60 % dip in UT) EQUIPMENT or ME SYSTEM]
voltage variations for 5 cycles requires continued operation during
on power supply power mains interruptions, it is
70 % UT
input lines (30 % dip in UT) recommended that the [ME
IEC 61000-4-11 for 25 cycles EQUIPMENT or ME SYSTEM] be
powered from an uninterruptible
<5 % UT power supply or a battery.
(>95 % dip in UT)
for 5 s
NOTE UT is the a.c. mains voltage prior to application of the test level.
Electromagnetic immunity
The ANYVIEW-500R is intended for use in the electromagnetic environment specified below. The customer or
the user of the ANYVIEW-500R should assure that it is used in such an environment.
IMMUNITY test IEC 60601 Compliance
Electromagnetic environment – guidance
TEST LEVEL level
Portable and mobile RF communications
equipment should be used no closer to any part of
the ANYVIEW-500R, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
ANYVIEW-500R is used exceeds the applicable RF compliance level above, the ANYVIEW-500R should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the ANYVIEW-500R.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
0.01
0.1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
APPENDIX B LABEL
LABEL
12
1 11
11
5
14, 15
13
6
7
9
8
3
10, 11
The labels shown in the following table are merely reproductions. Their sizes and colors may differ
from reality and serve illustrative purposes only.
NO LABEL COMMENTS
MOBILE C-ARM
/ MOBILE C-ARM X-RAY
EQUIPMENT
1
ANYVIEW-500R
X-RAY GENERATOR
ANYVIEW-500R
2 MAX KVP
FILTRATION
FOCAL SPOT
TUBE TYPE
TUBE SERIAL NO.
5 X-ray Radiation
10 HEAVY OBJECT
13 Radiation Warning
14 HV 3 Minutes Warning