Service Manual

Ecotron Co., Ltd


Made by Korea
4F&5F, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul 08511, Korea
Obelis S.A
Bd. General Wahis 53, 1030 Brussels, Belgium / E-mail:mail@obelis.net
Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03

D/N: ECO-SM-C500R-001 Rev. C


ANYVIEW-500R SERVICE MANUAL

(This page intentionally left blank)

Rev. C 2/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

CONTENTS

1. INTRODUCTION ________________________________________________________________________ 7

1.1 ANYVIEW-500R MAIN FEATUES _____________________________________________________________ 9

1.2 COMPOSITION _______________________________________________________________________________ 9

2.1 SAFETY _____________________________________________________________________________________ 11

2.3 SYMBOLS __________________________________________________________________________________ 15

2.4 LABELS_____________________________________________________________________________________ 17

2.5 CLASSIFICATION ACCORDING TO IEC 60601-1 __________________________________________ 20

2.6 ESSENTIAL PERFORMANCE ______________________________________________________________ 20

3. ANYVIEW-500R DUTY CYCLE LIMIT _________________________________________________ 21

4. DAILY X-RAY TUBE WARM-UP PROCEDURE ________________________________________ 22

5. APPLICABLE STANDARDS ___________________________________________________________ 23

5.1 SAFETY _____________________________________________________________________________________ 23

5.2 EMC _________________________________________________________________________________________ 23

5.3 INTERVENTIONAL PROCEDURE __________________________________________________________ 23

5.4 OTHERS ____________________________________________________________________________________ 23

5.5 CLINICAL EVALUATION ___________________________________________________________________ 23

6. INSTALLATION ________________________________________________________________________ 24

6.1 UNLOADING AND INSPECTING THE SHIPMENT _________________________________________ 24

6.2 MECHANICAL TRANSPORT POSITION ___________________________________________________ 24

6.3 MECHANICAL INSPECTION _______________________________________________________________ 26

6.4 MONITOR CART INSTALLATION __________________________________________________________ 32

6.5 C-ARM INSTALLATION_____________________________________________________________________ 34

6.6 INSTALLATION FPD and GRID ____________________________________________________________ 35

6.7 SWITCHING THE SYSTEM ON AND OFF __________________________________________________ 36

Rev. C 3/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6.8 EMERGENCY STOP BUTTON _____________________________________________________________ 37

6.9 BOOTING THE SYSTEM ___________________________________________________________________ 38

7. CONSOLE CONTROLS _______________________________________________________________ 39

7.1 Anyview-500R ______________________________________________________________________________ 39

7.1.1 MODE SELECT ___________________________________________________________________________ 39

7.1.2 kV UP/DOWN BUTTON AND DISPLAY __________________________________________________ 40

7.2.3 mA or mAs UP/DOWN BUTTON AND DISPLAY _________________________________________ 40

7.2.4 IMAGE FLIP, ROTATION AND SAVE BUTTON ___________________________________________ 40

7.2.5 COLLIMATOR CONTROL BUTTON ______________________________________________________ 40

7.2.6 I.I SIZE SELECT BUTTON ________________________________________________________________ 41

7.2.7 TIME LIMIT / TUBE LIMIT ALARM / EEROR RESET BUTTON __________________________ 41

7.2.8 X-RAY EXPOSURE CONTROL BUTTON _________________________________________________ 41

7.2.9 LASER POINTER ON / OFF BUTTON ____________________________________________________ 41

7.2.10 WAIT / READY / X-RAY EXPOSURE INDICATOR ______________________________________ 41

7.2.11 VERTICAL MOVEMENT UP/DOWN BUTTON __________________________________________ 42

7.2.12 BEEP SOUND TURN ON/OFF __________________________________________________________ 42

8. SCHEMATICAND TROUBLE SHOOTING _____________________________________________ 44

8.1 BLOCK DIAGRM ___________________________________________________________________________ 44

8.2 PCB BOARD LAYOUT ______________________________________________________________________ 45

8.3 TROUBLE SHOOTING _____________________________________________________________________ 55

9. SPECIFICATION _______________________________________________________________________ 73

9.1 SPECIFICATION TABLE ____________________________________________________________________ 73

9.2 MECHANICAL SPECIFICATION____________________________________________________________ 76

10. ENVIRONMENTAL SPECIFICATIONS _______________________________________________ 77

11. CUSTOMER SUPPORT ______________________________________________________________ 78

Rev. C 4/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

12. MAINTENANCE PROCEDURE ______________________________________________________ 79

12.1 GENERAL CAUTION ______________________________________________________________________ 79

12.2 MAINTENANCE SCHEDULE ______________________________________________________________ 79

12.3 THE EQUIPMENT OPERATION CHECKLIST FOR OPERATOR _________________________ 79

12.4 GLASS IMPLOSION OF I.I ________________________________________________________________ 79

12.5 CLEANING ________________________________________________________________________________ 80

12.6 THE END OF PRODUCT LIFE ____________________________________________________________ 81

13. QUALITY WARRANTY _______________________________________________________________ 82

APPENDIX A EMC COMPATIBILITY _____________________________________________________ 83

APPENDIX B LABEL ____________________________________________________________________ 87

Rev. C 5/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

REVSON HISTORY

Revision Number Date Description

A MAY 12, 2015 First Edition

B OCT 17, 2016 Update Manual

C AUG 30, 2018 Update Manual

ADVISORY SYMBOLS
The following advisory symbols are used throughout this manual.
Their application and meaning are described below.

WARNING

Warning symbol used to indicate a potential hazard lead operators, service personnel to serious injury, death
or radiation exposure.

CAUTION

Caution symbol used to indicate a potential hazard lead operators, service personnel to injury or damage of
equipment.

NOTE

Note symbol used to indicate important information need to proper use and right operation of equipment.

NOTE

KEEP THIS SERVICE MANUAL WITH THE EQUIPMENT AT ALL TIMES, AND REVIEW THE
IMPORTANT INFORMATION WHENEVER REQUIRED.

THE STATISTICS AND SPECIFICATIONS OF THIS UNIT AND MANUAL CAN BE MODIFIED
WITH OR WITHOUT NOTIFICATION FOR THE IMPROVEMENT OF PERFORMANCE AND
SAFETY.

Rev. C 6/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

1. INTRODUCTION

The ANYVIEW-500R Innovation mobile C-arm features excellent x-ray Image, quality, accuracy, reproducibility
and long-term stability. Fast kV rising time minimize unnecessary patient dose and supply superior Image
quality, fast Image response.

The operator control console is designed to be simple and user-friendly and one can select or change x-ray
parameters easily using a large soft membrane switch.

Automatic Calibration function will minimize calibration time and set correct calibration value. Computer
program of ANYVIEW-500R offers graphic waveform of kV, mA and data of x-ray exposure. So, you don’t
have to perform manual calibration and measurement equipment are unnecessary. Naturally, ANYVIEW-500R
supports Automatic, Semi-auto and Manual Calibration functions.

Adaptable calibration function keeps up long-term stability and accuracy by trimming calibration data
whenever x-ray exposed.

Remote Diagnosis Software can make report about information of system, user, service and error by the USB
connection between PC and control console, and transfer report to ECOTRON via internet.

Indications for use / Intended use

The mobile C-arm Anyview-500R is intended for use to produce X-ray images of patient during medical
operations. The Anyview-500R system consists of a Main power board and Main control board, Control
console, SMPS, X-ray generator, Image intensifier.

X-ray generator contains the HT tanks and control circuits, the filament drivers, FET drivers, Collimator control
circuits and interface connections to the Main control board.

This product, Fluoroscopic x-ray system, is radiation medical equipment only used by professional radiologists.
This product is designed to provide fluoroscopic and spot film images of the patient during diagnostic and
interventional procedures.

This system can be applied in emergency room, operation room, cast room or etc. of hospital.

Indication
Bone fracture, Interventional procedures

Rev. C 7/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

Available the patient population for use

 Gender: no gender restrictions


 Age: no age restrictions
 The patient population for use
- Adults
- Pediatric patients

NOTE

FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF


PHYSICIAN OR WITH THE DESCRIPTIVE DESIGNATION OF ANY OTHER PRACTITIONER
LICENSED BY THE LAW OF THE STATE IN WHICH HE PRACTICES TO USE OF ORDER
THE USE OF THE DEVICE.

Side Effects
Operating X-ray on a patient for a lengthy period of time may cause the following side effects.
 Dizziness
 Nausea
 Vomiting
 Diplopia
 Skin Numbness
 Masticatory Atonia
 Transient leg paresthesia

Contraindication
 In case of taking continuously pulsating X-rays on the parts where artificial cardiac pacemaker or
implantable defibrillator is implanted, such devices may suffer from inappropriate malfunction.
Therefore direct X-ray on these parts should be avoided, and minimum output should be used for a
short period of time.
 In case of taking X-rays of infants, genital protective gear should be used on the genitalia.
 The subject and the conductor should be at least 20cm away from the equipment when taking the X-
ray.

Rev. C 8/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

1.1 ANYVIEW-500R MAIN FEATUES

 Real-time monitoring self-diagnosis function.


 Overload & HU protection and error message display.
 Automatic calibration without measurement equipment.
 Adaptable calibration keeps up accuracy through long-term usage.
 Smaller, lighter and convenient modular design and user-friendly system configuration.
 Constant dose output due to kV and mA regulation by high-frequency switching during exposures.
 Remote diagnosis software for system diagnosis via internet.
 The collimator is used to reduce an unnecessary irradiation of X-Ray.
 The user-friendly control console for excellent controllability and visibility.
 Thermal sensor prevents the tube from overheating.

1.2 COMPOSITION

This product is composed of the X-ray component part, the display image part, accessories part, and options.
Detail information of each component follows below.

(1) X-ray component part


 Image Intensifier
 CCD Camera
 Main Power Module with Main Control board
 SMPS
 Mono block X-ray tank
 X-ray control board
 FET drive board
 Collimator with Iris and shutter type
 Collimator control board

(2) Display image part


 Image Monitor Cart
 Dual Monitor
 Interface cable
 PC
 DIS(Digital Interface System) without DICOM

(3) Accessories
 Operator’s Manual
 Hand switch
 Foot switch
 Laser

Rev. C 9/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(4) Option
 DIS(Digital Interface System) with DICOM
 Dose Area Meter
 Air Kerma
 Disinfected covers

NOTE

THIS MANUAL CONTAINS IMPORTANT SAFETY INFORMATION.

AN UNDERSTANDING OF THIS INFORMATION IS CRITICAL TO THE SAFE OPERATION


OF YOUR EQUIPMENT.

PLEASE ENSURE THAT YOU READ THE WARNING NOTICES BEFORE USING THE
EQUIPMENT

Rev. C 10/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2. SAFETY AND SPECIFICATIONS

2.1 SAFETY

X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending over
periods of many months or even years. X-ray operators should avoid any exposure to the primary beam
and take protective measures to safeguard against scatter radiation. Scatter radiation is caused by any object
in the path of the primary beam and may be of equal or less intensity than the primary beam that exposes the
film.

No practical design can incorporate complete protection for operators or service personnel who do not take
adequate safety precautions. Only authorized and properly trained service and operating personnel
should be allowed to work with this X-ray system equipment. The appropriate personnel must be made
aware of the inherent dangers associated with the servicing of high voltage equipment and the danger of
excessive exposure to X-ray radiation during system operation.

 Wear protective clothing. Protective aprons and gloves with an equivalent of a minimum of 1/64”(0.35mm)
of lead are recommended.

 To protect the patient against radiation, always use radiation protection accessories in addition to devices
which are fitted to the X-ray system

 Keep as large a distance as possible away from the object being exposed and the X-ray tube assembly.

 Never operate this X-ray system in areas where there is a risk of explosion. Detergents and disinfectants,
including those used on patients, may create explosive mixtures of gases. Please observe the relevant
regulations.

 ANYVIEW-500R is made available within 1.5m of the patient environment.

 Do not place liquids(coffee, beverages, flowers, etc) on the X-ray system.

 Do not operate the x-ray system in direct sunlight or near any heat sources.

 Do not operate the x-ray system near strong magnetic fields(microwave ovens, speakers, etc), and avoid
routing the x-ray system near these devices.

 The x-ray system must be operated in locations that are clean(free of excess dust, dirt, debris, etc),
stable(free of vibration).

 Only trained maintenance staff may remove the covers of the x-ray system.

 Always select a focal spot to skin distance as large as possible to keep the absorbed dose for the patient
as low as reasonably possible.

Rev. C 11/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

WARNING

TO AVOID RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE CONNECTED


TO A SUPPLY MAINS WITH PROTECTIVE EARTH.

WARNING

DO NOT CONNECT ADDITIONAL EXTENSION CORD TO EQUIPMENT.

WARNING

THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE
EXPSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.

WARNING

PROPER USE AND SAFE OPERATING PRACTICES WITH RESPECT TO X-RAY SYSTEMS
ARE THE RESPONSIBILITY OF THE USERS OF SUCH X-RAY SYSTEMS.

ECOTRON Co., Ltd PROVIEDS INFORMATION ON ITS PRODUCTS AND ASSOCIATED


HAZARDS, BUT ASSUMES NO RESPONSIBILITIES FOR AFTER-SALE POERATING AND
SAFETY PRACTICES.

ECOTRON Co., Ltd ACCEPTS NO RESPONSIBILITY FOR ANY GENERATOR NOT


MAINTAINED OR SERVICED ACCORDING TO THE SERVICE MANUAL OR ANY X-RAY
SYSTEM THAT HAS BEEN MODIFIED IN ANY WAY.

ECOTRON Co., Ltd ALSO ASSUMES NO RESPONSIBILITY FOR X-RAY RADIATION


OVEREXPOSURE OF PATIENTS OR PERSONNEL RESULTING FROM POOR OPERATING
TECHNIQUES OR PROCEDURS.

WARNING

MONITOR CART IS NON-MEDICAL EQUIPMENT. IT MUST USE OUT OF PATIENT


ENVIRONMENT. AND IT MUST COMPLIANCE WITH IEC 60950 STANDARD.

Rev. C 12/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

WARNING

DO NOT ALLOW OPERATION OF THIS APPARATUS BY ANY PERSON OTHER THAN


QUALIFIED PERSONNEL (PHYSICIANS, RADIOTHERAPY ENGINEERS AND CLINICAL X-
RAY ENGINEERS) OR UNDER OBSERVATION BY THEM.

CAUTION

INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN


INJURY OR EQUIPMENT DAMAGE.

CAUTION

DO NOT EXCEED THE TUBE MAXIMUM POERATING LIMITS SHOWN IN THE X-RAY
TUBE DATA SECTION AT THE END OF THE POERATOR’S MANUAL.

INTENDED LIFE AND RELIABILITY WILL NOT BE OBTAINED UNLESS X-RAY SYSTEMS
ARE POERATED WITHIN PUBLISHED SPECIFICATION.

CAUTION

WHEN PATIENTS WITH A PACEMAKER ARE IN RADIOGRAPH/FLUOROSCOPY


SUCCESSIVELY, COULD RESULT IN MALFUNCTION OF THE PACEMAKER.

CAUTION

DO NOT CONNECT ANY OTHER ELECTRICAL DEVICE TO THE LAN CONNECTOR AND
THE USB CONNECTOR OF THE EQUIPMENT EXCEPT FOLLOWING ELECTRICAL
DEVICES.

• MEDICAL ELECTRICAL DEVICES WHICH CONFORM TO IEC60601-1


• NON-MEDICAL ELECTRICAL DEVICES WHICH CONFORM TO RELATED IEC SAFETY
STANDARDS
• NON-MEDICAL ELECTRICAL DEVICES WHICH HAVE SAFETY EQUAL TO DEVICES
CONFORM TO IEC SAFETY STANDARDS

Rev. C 13/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2.2 ELECTROMAGNETIC COMPATIBILITY

WARNING

 OTHER CABLES AND ACCESSORIES MAY NEGATIVELY AFFECT EMC


PERFORMANCE.

 USE OF OTHER ACCESSORIES RESULTS IN NON-COMPLIANCE

 ANYVIEW SERIES SHOULD NOT BE USED ADJACENT TO OR STACKED WITH


OTHER EQUIPMENT AND THAT IF ADJACENT OR STACKED USE IS NECESSARY,
ANYVIEW SERIES SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION IN THE
CONFIGURATION IN WHICH IT WILL BE USED.

The ANYVIEW-500R X-ray system with CE marking complies with the European Council Directive concerning
Medical Devices. One of the harmonized standards of this Directive defines the permitted levels of
electromagnetic emission from this equipment and its required immunity from the electromagnetic emission of
other devise.

It is not possible, however to exclude with absolute certainty the possibility that other high frequency electronic
equipment, which is fully compliant to the EMC regulations, will not adverse effect the operation of this x-ray
system. If the other equipment has a comparatively high level of transmission power and is in close proximity
to the generator, these EMC concerns (the risk of interference) may as mobile telephones, cordless
microphones and other similar mobile radio equipment be restricted from the vicinity of this X-ray system.

ANYVIEW-500R needs special precautions regarding EMC and needs to be installed and use according to
the EMC information provided in this operation manual. For more information, refer to Appendix B.

● Cable List

Cable Name Cable Length Shield Note (Manufacturer)

H-20 5M NO Ecotron
POWER CABLE 6M NO Ecotron
PC POWER CALBE 3M NO Ecotron
HAND S/W CABLE 1.5M NO Ecotron
FOOT S/W CABLE 1.5M NO Ecotron

Rev. C 14/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2.3 SYMBOLS
2.3.1 Symbols on the X-ray system

SYMBOL Description

Consult accompanying documents: This symbol advises the reader to consult the
accompanying documents.

Protection Class I, Type B

Radiation Warning: This symbol indicates radiation warning.

High voltage warning: This symbol means high voltage parts included.

Warning: This symbol indicates hazard.

Warning: This symbol indicates not to sit on the out box.

Warning: This symbol indicates not to put your weight on the packing box.

Heavy weight warning: This symbol indicates heavy weight warning.

Model name: This symbol means model name of system

Serial Number: This symbol is followed by the system serial number.

Manufacture Date: This symbol is followed by the system manufacture date.

Manufacturer information: This symbol is followed by the name and address of


the system manufacturer.

Rev. C 15/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

Disposal: For more information, see on Section 10.5 “THE END OF PRODUCT
LIFE”

Protective earth: This symbol means protective earth.

2.3.2 Symbols on the Packing of X-ray system

SYMBOL Description

KEEP DRY: This symbol means keep away from rain

Fragile: This symbol means do not drop of product

UP: This symbol means “This way up”

WARNING

INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN


INJURY OR EQUIPMENT DAMAGE.

CAUTION

THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE
EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.

Rev. C 16/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2.4 LABELS
2.4.1 COMPOSITION LABEL
IDENTIFICATION LABEL- DIGITAL IMAGING SYSTEM
This label is attached to Digital Imaging System. (Example)

IDENTIFICATION HV MONO TANK


This is label is attached to side of the HV MONO Tank inside of the X-ray generator. (Example)

Rev. C 17/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM


This label is attached to rear cover of the x-ray system. (Example)

IDENTIFICATION LABEL- COLLIMATOR


This label is attached to Collimator. (Example)

Rev. C 18/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2.4.2 SAFETY AND WARNING LABEL

RADIATION WARNING LABEL

This label is attached to the outside of the X-ray system, and


indicates warning for radiation.

HEAVY WEIGHT WARNING LABEL

This label is attached to the outside of the x-ray system, and


states the approximate weight of the main cabinet with the
HT oil tank. Do not attempt to lift this unit without proper
assistance.

HV 3 MINUTES WARNING LABEL


This label is attached to the outside of the main power
module. The DC link capacitor (approximately 325 VDC with
line voltage at 230VAC) will remain charged for up to
3 minutes after the AC mains is disconnected or the AC
switch is switched off.

DANGER HIGH VOLTAGE LABEL


This label attached to the main fuse cover and FET driver
board. Main voltage is present inside the x-ray system
whenever the main disconnect is switched on. Additionally,
the DC link capacitors will remain charged for up to 3
minutes after the AC mains is disconnected or the AC switch
is switched off.

SAFE USE WARNING LABEL

This label attached to a control panel. This X-ray unit may


be dangerous to patient and operator unless safe exposure
factors, operating instructions and maintenance schedules
are observed.

Rev. C 19/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2.5 CLASSIFICATION ACCORDING TO IEC 60601-1


 Protection Method against Electric Shock
- Class I equipment
 Degree of Protection against Electric Shock
- No applied part
 Operation Mode
- Continual operation
 Degree of Protection against Liquid Ingress
- Ordinary equipment (Foot switch IP68)
 Use in Flammable Atmosphere
- Not to be used in presence of flammable anesthetics gas
 Allowable weight for Monitor Cart
- Up to 20kg
 ME system could use within the patient environment.

NOTE

IF NEEDS, ECOTRON CO., LTD. COULD PROVIDE THE TECHNICAL DESCRIPTION AS


FOLLOWING LISTS.

- CIRCUIT DIAGRAMS
- COMPONENT PART LISTS
- CALIBRATION INSTRUCTIONS AND ETC.

2.6 ESSENTIAL PERFORMANCE

Essential Performance requirements according to IEC 60601-2-54:2009

Accuracy of Loading Factors

Reproducibility of the Radiation Output

Automatic Control System

Imaging Performance

Rev. C 20/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

3. ANYVIEW-500R DUTY CYCLE LIMIT

NOTE

THE FOLLOWING SECTION CONTAINS IMPORTANT INFORMATION. PLEASE READ AND


UNDERSTAND THIS MATERIAL BEFORE CONTINUING.

Internal X-ray system components will heat up during normal use of the x-ray system. This is similar to X-ray
tube heating during normal x-ray system operation. The amount of heat produced is proportional to the
product of kV, mA and time.

CAUTION

THIS X-RAY SYSTEM HAS TEMPERATURE MONITORING OF FET-DRIVE BOARD TO


PROTECT THE EXCESSIVE HEAT BUILD-UP.

IF THE X-RAY SYSTEM DETECTS THE OVER-HEAT OF FET-DRIVE, AN ERROR CODE


“E04” OR “E05” WILL BE DISPLAYED. EXPOSURE WILL BE INHIBITED WHEN THIS
MESSAGE IS DISPLAYED, AND IT SHOULD BE UNDERSTOOD THAT CONTINUING TO
MAKE EXPOSURES MIGHT CAUSE X-RAY SYSTEM DAMAGE DUE TO OVERHEATING.
THE GENERATOR SHOULD BE ALLOWED TO COOL SUFFICIENTLY SUCH THAT THIS
MESSAGE IS NO LONGER DISPLAYED.

Rev. C 21/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

4. DAILY X-RAY TUBE WARM-UP PROCEDURE

NOTE

THE FOLLOWING PROCEDURE PRODUCES X-RAYS. OBSERVE ALL SAFETY


PRECAUTIONS TO PROTECT PERSONNEL.

Use the procedure when the generator has not been used for several days. This procedure provides for
exposures at medium power before the tube is used at maximum mA or kV values. This will reduce the
possibility of damaging the X-ray tube and high voltage components. No test setup is required.

For maximum stability and reliability, use the following techniques at start up:

Select the following:


 Small focal spot.
 80kV / 2mA.
 Use approximately 10sec per exposure.

Make three to five exposures at 30 second intervals.

Rev. C 22/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

5. APPLICABLE STANDARDS

The ANYVIEW-500R of X-ray system complies with the regulatory requirements and design standards in this
section as follows:

This product complies with DHHS requirements of 21 CFR Subchapter J.

5.1 SAFETY
● IEC 60601-1:2015+A1:2012
● IEC 60601-1-6:2010+A1:2013
● IEC 62366:2007
● IEC 60601-2-28:2010
● IEC 60601-2-54:2009

5.2 EMC
● IEC60601-1-2:2007
● EN55011:2007+A2:2007GROUP 1 CLASS A
● EN61000-3-2:2000+A2:2005
● EN61000-3-3:1995+A1:2001

5.3 INTERVENTIONAL PROCEDURE


● IEC60601-2-43:2010

5.4 OTHERS
● ISO15223-1:2012 Medical devices -Symbols to be used with medical device labels, labeling and
information to be supplied
● IECTR60878:2003 Graphical symbols for electrical equipment n medical practice
● IEC60417-1:2012 Graphical symbols for use on equipment-part1:overview and application
● ENISO14971:2007 Medical devices - Application of risk management to medical devices
(ISO 14971:2007)
● ENISO13485:2003 Medical devices - Quality management systems-Requirements for regulatory
purposes
● ISO9001:2008 Quality management systems-Requirements
● 93/42/EEC:1993 Council Directive concerning medical devices
● EN1041:2008 Information supplied by the manufacturer with medical devices
● EN62304:2006 Medical device software-Software lifecycle processes

5.5 CLINICAL EVALUATION


● EN ISO 14155-1:2003 Clinical investigation of medical devices for human subjects - Part1: General
requirements (ISO14155-1:2003)
● MEDDEV.2.7.1 Evaluation of Clinical Data: A Guide for Manufacturers and Notified Bodies

Rev. C 23/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6. INSTALLATION

6.1 UNLOADING AND INSPECTING THE SHIPMENT


Before unpacking, the customer should inspect the shipping containers and exposed system components for
any apparent damage. Any shipping container that appears damaged should be unpacked with the carrier’s
agent present. The customer should also ensure that all equipment specified in the purchase order has been
delivered.
If any part of the shipment is damaged or missing, immediately notify the carrier and the service office.
Retain the shipping container and packing material for the carrier’s inspection.

6.2 MECHANICAL TRANSPORT POSITION


(1) Transport position
For safety reasons, you must return both the C-arm system and the monitor cart to their respective transport
position before transporting them.

(2) C-arm system transport position.


To prepare the C-arm system for a transport, do the following:
 Wind the foot switch cable onto the foot switch support and hang up the foot switch there.
 Rotate the C-arm orbitally until the Image Intensifier is positioned directly above the generator and secure
it with relevant orbital rotation brake.
 Lower the lifting column completely using the move up/down buttons.
 Panning the horizontal carriage into a central position and secure it with the panning brake.
 Retract the horizontal carriage completely and secure it with the horizontal movement brake.
 Switch off the system and disconnect the power cable from the power supply.
 Release the parking brake.

(3) Braking and steering the C-arm system


The C-arm system has a combined steering handle & Parking brake. It is located at the rear end and the
bottom side of the C-arm system.

The C-arm system parking brake operates as follows:


 Press parking brake: Wheels are locked.
 Release parking brake: Wheels can move freely.

To steer the C-arm system, do the following:


 Rotate the steering handle until reaching the desired position and Release parking brake.

Rev. C 24/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

CAUTION

KEEP IN MIND THAT THE C-ARM SYSTEM MAY ROLL AWAY ON UNEVEN FLOORS.
LEAVE THE C-ARM STAND BEHIND ONLY ON EVEN FLOORS AND LOCK ITS BRAKES.
RELEASE THE PARKING BRAKE ONLY TO MOVE OR POSITION THE C-ARM SYSTEM.

(4) Monitor cart transport position


To prepare the monitor cart for a transport, do the following:
 Wind the power cable onto the cable support on the back of the monitor cart
 Wind the video cable onto the cable support on the back of the monitor cart
 Unlock the parking brake by lifting the foot levers on the front wheels.

(5) Braking and steering the monitor cart.


On the wheels of the monitor cart there are foot levers which serve for locking and releasing the brakes.
 Levers in lower position: the wheels are locked, the brake is locked as well.

 Levers in upper position: you can move the monitor cart to the left and right side.

Figure 6.1 Braking and Steering wheels

Rev. C 25/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6.3 MECHANICAL INSPECTION


Inspect each part of C-arm system.

(1) Mechanical brake inspection

Angulation Brake

Panning Brake
Horizontal
Movement
Brake

Orbital
Steering
Rotation Brake
Handle
Vertical Movement
Parking Brake
Up/Down Switch

Figure 6.2 Mechanical Brake

CAUTION

DO NOT MOVE THE SYSTEM OVER FLOORS WITH AN INCLINATION OF MORE THAN 10°
FROM LEVEL DURING TRANSPORT. THIS APPLIES TO BOTH THE C-ARM STAND AND
THE MONITOR CART.

EXERCISE EXTREME CAUTION WHEN MOVING THE C-ARM STAND AND THE MONITOR
CART OVER ROUGH SURFACES SUCH AS TILE FLOORING, PAVEMENT, ASPHALT OR
CARPET. TAKE CARE THAT THE CABLE GUARDS DO NOT DRAG AND THE WHEELS DO
NOT CATCH OR TILT CAUSING DAMAGE TO THE EQUIPMENT.

Rev. C 26/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(2) Orbital rotation inspection


 Unlock the orbital rotation brake.
 Rotate the C-arm in the desired position.
 Lock the orbital rotation brake.

Figure 6.3 Orbital Rotation

CAUTION

BEFORE ROTATING THE C-ARM, MAKE SURE THAT IT DOES NOT COLLIDE WITH ANY
PERSONS OR OBJECTS.

Rev. C 27/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(3) Vertical movement inspection


 Press the UP/DOWN Switch.
 Move the vertical column in the desired position.

Figure 6.4 Vertical Movement

CAUTION

BEFORE MOVING THE C-ARM UP OR DOWN, MAKE SURE THAT IT DOES NOT COLLIDE
WITH ANY PERSONS OR OBJECTS.

Rev. C 28/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(4) Angulation movement inspection


 Unlock the angulation brake.
 Rotate the C-arm in the desired position.
 Lock the angulation brake.

Figure 6.5 Angulation Movement

Rev. C 29/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(5) Panning movement inspection


 Unlock the panning brake.
 Move the C-arm in the desired position.
 Lock the panning brake.

Figure 6.6 Panning Movement

Rev. C 30/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(6) Horizontal movement inspection


 Unlock the horizontal movement brake.
 Push or pull the cross-arm in the desired position.
 Lock the horizontal movement brake.

Figure 6.7 Horizontal Movement

CAUTION

BEFORE MOVING THE C-ARM BACK AND FORTH, MAKE SURE THAT IT DOES NOT
COLLIDE WITH ANY PERSONS OR OBJECTS.

Rev. C 31/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6.4 MONITOR CART INSTALLATION

(1) Install Main Live Monitor and Sub Monitor as below

Rev. C 32/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(2) To Install Monitor on Monitor arm

Monitor Arm

Install Monitor on the Monitor Arm

Main Live Monitor & Sub Monitor

Rev. C 33/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6.5 C-ARM INSTALLATION


Before switching on the system, after a transport, you must connect the two system components to each other
and also connect the entire system to the power supply.

To prepare the system operation, do the following:


 Plug the video input cable connector into the video output socket located at the front of the C-arm system
and lock the connector.

 Plug the foot switch cable connector into the foot switch socket located at the front of the C-arm system
and lock the connector.

 Make sure that a suitable supply voltage is available and that the socket-outlet is properly grounded and
fused.

Rev. C 34/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

 Connect the C-arm system and Monitor cart power cable plug to the power supply.

6.6 INSTALLATION FPD and GRID

Grid cover
Grid
Plastic Bolt (4ea)

CAUTION

YOU MUST USE STANDARD BOLT WHEN INSTALL X-RAY GRID AND GIRD COVER.

YOU MUST INSTALL GRID TO MATCH ANODE DIRCTION OF X-RAY TUBE AS SHOWN
ABOVE.

Rev. C 35/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6.7 SWITCHING THE SYSTEM ON AND OFF

Both the C-arm system and the monitor cart have power switches for witching the system on.
You must turn on first C-arm system power switch before turn on monitor cart power switch.

You must wait 20 seconds to execute DIS icon ( ) after turn on the C-arm system.

NOTE

DO NOT PLUG ANY USB STICK INTO THE USB PORT UNTIL THE SYSTEM HAS FULLY
COMPLETED ITS POWER-UP SEQUENCE AND THE START SCREEN IS DISPLAYED.

NOTE

DUE TO BACKGROUND RADIATION, THE AIR KERMA, AIR KERMA RATE OF DOSE
AREA PRODUCT DISPLAY MAY INDICATE SOME SMALL VALUE AFTER POWER-UP OF
THE SYSTEM.

NOTE

DURING POWER-UP, DO NOT TOUCH EITHER OF THE CONTROL PANELS. THE SYSTEM
CAN STILL BE CONTROLLED VIA THE CONTROL PANELS.

Rev. C 36/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6.8 EMERGENCY STOP BUTTON


There is an EMERGENCY STOP button on the C-arm system, enabling you to switch off all electrical
functions of the unit in case of emergencies.

NOTE

PRESS THE EMERGENCY STOP BUTTON ONLY IF THIS CANNOT BE AVOIDED, SINCE
UNSAVED PATIENT DATA AND IMAGES WILL BE LOST.

 Fully press down the EMERGENCY STOP button.

The EMERGENCY STOP button is locked in this position. All electrical functions of the system are disabled.
Any unsaved patient data and/or images are lost.

Rev. C 37/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6.9 BOOTING THE SYSTEM


Turn off the power switch and insert the power plug into an outlet. Turn the power switch on, and check the
messages on the SEGMENT.

You must turn on first C-arm system power switch and turn on monitor cart power switch.

 Initial status
When powered on, the startup screen (figure 6.8) is displayed on the SEGMENT. (During about 5sec and not
allowed input with key)

After 5sec, Initial display will be changed like as figure 6.8 and displayed normal fluoroscopy offset condition
“80KV, 1mA”

Figure 6.8 Initial State & Initial Mode State

Rev. C 38/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

7. CONSOLE CONTROLS

7.1 Anyview-500R

7.1.1 MODE SELECT

RADIOGRAPHY MODE SELECT BUTTON

FLUOROSCOPY MODE SELECT BUTTON

THE PULSED FLUOROSCOPY MODE SELECT BUTTON


(YOU CAN SELECT 8, 15, 30 PULSED MODE)

BOOST MODE SECECT BUTTON

ABS MODE SELECT BUTTON (only Use F-mode)

IDR (Intelligent Dose Reduction) MODE SELECT BUTTON

MAR (Motion Artifact Reduction) MODE SELECT BUTTON

DS (Digital Shot) MODE SELECT BUTTON

DISPLAYS THE SELECTED MODE INFORMATION


- rA: R-Mode
- FL: F-Mode
- PL: P-Mode
- b0: Boost
- dr: IDR
- Ar: MAR
- dS: DS-Mode

Rev. C 39/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

7.1.2 kV UP/DOWN BUTTON AND DISPLAY

To increase kV, press / To decrease kV, press


Displays demanded kVp of the exposure.

7.2.3 mA or mAs UP/DOWN BUTTON AND DISPLAY

To increase mA or mAs, press / To decrease mA or mAs, press


Displays demanded mA or mAs of the exposure.

7.2.4 IMAGE FLIP, ROTATION AND SAVE BUTTON

To rotate a stored Image clockwise, press / To rotate a stored Image


counterclockwise, press

To reverse a stored Image horizontally, press / To reverse a stored Image


vertically, press

X-RAY IMAGE STORE BUTTON (when you want to store x-ray image)

7.2.5 COLLIMATOR CONTROL BUTTON

To enlarge iris collimator, press / to reduce iris collimator, press


Displays open range of Iris collimator.

To enlarge shutter collimator, press / to reduce shutter collimator,

press

To rotate shutter collimator clockwise, press / to rotate shutter collimator

counterclockwise, press

Rev. C 40/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

7.2.6 I.I SIZE SELECT BUTTON

To select 4.5", 6", 9" I,I size, press


Displays I.I size

7.2.7 TIME LIMIT / TUBE LIMIT ALARM / EEROR RESET BUTTON

To reset time limit, press ( ). This key is used to get out of the 'Time Limit'
status without 1 minute standby when a period of cumulative exposure time
exceeds 5 minutes but you cannot irradiate X rays.

This key is used to get out of the 'Tube Limit' status for emergent re-
radiography when the 'Tube Limit' is in action to protect the tube and therefore
you cannot radiograph.

7.2.8 X-RAY EXPOSURE CONTROL BUTTON

To make x-ray exposure for radiography or fluoroscopy, press ( ). Shooting


operates according to seleted shooting mode.

7.2.9 LASER POINTER ON / OFF BUTTON

The laser positioning device generates a laser-beam crosshair, the central


point of which marks the position of the central X-ray beam on the patient.

7.2.10 WAIT / READY / X-RAY EXPOSURE INDICATOR

The ready indicator ( , white) will light when ready to make an


exposure.
The x-ray exposure indicator ( , Green) will light when an X-ray
exposure is being taken.
The wait indicator ( , Yellow) will light when x-ray exposure is ended
during 3 seconds.

Rev. C 41/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

7.2.11 VERTICAL MOVEMENT UP/DOWN BUTTON

Press up/down button. C-arm system is moving vertical position.

7.2.12 BEEP SOUND TURN ON/OFF

To turn off beep sound, keep press laser button during 3sec.
To turn on beep sound, keep press laser button during 3sec.

Rev. C 42/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

CAUTION

THE HAND SWITCH HAS A 2-STEP OPERATIONAL STRUCTURE. IF YOU GO DIRECTLY


TO THE SECOND STEP WITHOUT GOING THROUGH THE FIRST STEP, X-RAYS WILL
NEVERTHELESS START TO BE IRRADIATED AFTER 3 SECONDS.

CAUTION

THE HAND SWITCH BUTTON MUST REMAIN PRESSED DOWN TO COMPLETE X-RAYS
IRRADIATION.

CAUTION

THE HAND SWITCH BUTTON MUST REMAIN PRESSED DOWN TO COMPLETE X-RAYS
IRRADIATION.

CAUTION

UNDER ANY CIRCUMSTANCES, RELEASING THE HAND SWITCH BUTTON DURING X-


RAYS IRRADIATION WILL IMMEDIATELY PREVENT IRRADIATION (DEAD MAN'S
SWITCH). IN THIS CASE, THE MACHINE HAS NO TROUBLES, AND YOU CAN START
IRRADIATION AGAIN.

CAUTION

TO PROTECT THE TUBE, SUFFICIENT WAITING TIME MUST BE GIVEN. AFTER


IRRADIATION ONE TIME, A WAITING TIME OF 60 SECONDS IS GIVEN TO THIS MACHINE.

Rev. C 43/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8. SCHEMATICAND TROUBLE SHOOTING

8.1 BLOCK DIAGRM

Rev. C 44/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.2 PCB BOARD LAYOUT


8.2.1 SMPS1, SMPS2

MODEL Voltage Current Max Watt

SMPS1: SNP-Z109-M 24V 4.5A 100W

SMPS2: SNP-Z08F 5V/24V/12V 6A/2A/0.5A 85W

8.2.2 MAIN FUSE BOARD

Voltage Current
Fuse Rating
250VAC 25A

Rev. C 45/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.2.3 MAIN BORAD AND MAIN CONTORL BOARD

Rev. C 46/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

1) Main board LED Display

LED
I.I Size
D21 D22 D23

OFF OFF OFF 9 inch

ON OFF OFF 6 inch

OFF ON OFF 4.5 inch

LED DESCRIPTION

D19 CCD Camera Trigger Out ON Indicator

Rev. C 47/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2) Main Control board LED Display

LED DESCRIPTION

D7 Flashing operate periodically by CLK

D8 Flashing operate periodically by CLK

D9 Flashing operate periodically by CLK

D10 X-ray Exposure Indicator

D11 NOT USED

D12 X-ray Ready Indicator

D13 X-ray Wait Indicator


Fault state: Hold the current state when system is malfunction.
D14
Normal state: Flashing operate periodically

Rev. C 48/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

3) Main Control board DIP Switch Position

DIP S/W 1 Description


OFF: Normal mode
S/W1
ON: External program mode

S/W2 NOT USED

S/W3 NOT USED

S/W4 NOT USED

S/W5 NOT USED

S/W6 NOT USED

S/W7 NOT USED


OFF: Normal mode
S/W8
ON: Initialize internal memory

DIP S/W 2 Description


OFF: Normal mode
S/W1
ON: Reset program
OFF: Normal mode
S/W2
ON: Program download mode

Rev. C 49/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.2.4 HFG TANK BOARD LAYOUT

No. BOARD NAME

A FET BOARD

B VR COLLIMATOR BOARD

C XCU DRIVER BOARD

D XCU_MCU BOARD

Rev. C 50/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.2.5 X-RAY CONTROL BOARD

1) X-ray Control Board LED Display

LED DESCRIPTION
D18 X-Tube Filament Current Over Error Indicator

D20 X-Tube Current Over Error Indicator

D22 X-Tube Voltage Over Error Indicator

D24 FET Pulse by Pulse

D4 X-ray Wait Indicator

D5 X-ray Exposure Indicator

D6 X-ray Ready Indicator

D7 System Working Indicator(Flashing every 1 sec)


Detecting Collimator Origin Indicator
D32 - LED ON : Origin Position
- LED OFF: Another Position

Rev. C 51/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2) X-ray Control Board DIP Switch Position

DIP S/W 1 Description


OFF: HV tank Internal thermal sensor enable mode
S/W1
ON: HV tank Internal thermal sensor disable mode
OFF: Error enable
S/W2
ON: Error disable
S/W3 NOT USED
OFF: Preheat data reference normal mode
S/W4
ON: Preheat data reference half mode
S/W5 NOT USED
OFF: X-ray control board normal mode
S/W6
ON: X-ray control board reset mode

Rev. C 52/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.2.6 FILAMENT AND FET DRIVER BOARD

8.2.7 FET BOARD

Rev. C 53/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.2.8 Display and Key Board

Display Main Board

Display Sub Board

Rev. C 54/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.3 TROUBLE SHOOTING


8.3.1 Error code
If an error occurs, do the following
- Contact your after-sales service center.

List of error messages on the C-arm

Error Code Display Description

Tube rotor is not spinning (Detecting by H/W).


0x00080000 26
Please contact to service engineer.

IP (Feedback Current) is over the limit (Detecting by H/W).


0x00040000 32
Please contact to service engineer.

EP (Feedback Voltage) is over the limit(Detecting by H/W).


0x00020000 30
Please contact to service engineer.

Filament Current is over the limit (Detecting by H/W).


0x00010000 31
Please contact to service engineer.

PBP (Pulse by Pulse) is over the limit (Detecting by H/W).


0x00008000 28
Please contact to service engineer.

Filament feedback is over the limit in standby state.


0x00001000 10
Please contact to service engineer.

EP feedback is detected in standby state.


0x00000800 11
Please contact to service engineer.

IP feedback is detected in standby state.


0x00000400 12
Please contact to service engineer.

Filament feedback is abnormal while exposing x-ray.


0x00000200 14
Please contact to service engineer.

IP feedback is lower than reference level while exposing x-ray.


0x00000100 18
Please contact to service engineer.

IP feedback is higher than reference level while exposing x-ray.


0x00000080 16
Please contact to service engineer.

EP feedback is lower than reference level while exposing x-ray.


0x00000040 15
Please contact to service engineer.

EP feedback is higher than reference level while exposing x-ray.


0x00000020 17
Please contact to service engineer.

Rev. C 55/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

DC Voltage is not detected.


0x00000010 18
Please contact to service engineer.

Foot switch is keeping ‘Ready’ state.


0x00000008 5
Please check the foot switch.

Foot switch is keeping ‘Exposure’ state.


0x00000004 6
Please check the foot switch.

Hand switch is keeping ‘Ready’ state.


0x00000002 7
Please check the hand switch.

Hand switch is keeping ‘Exposure’ state.


0x00000001 8
Please check the hand switch.

List of warning messages on the C-arm

Warning
Display Description
Code
The heat sink is overheated on FET driver board.
0x00004000 29
Please wait for the heat sink to cool down.

The HV-Tank is overheated in Monoblock.


0x00002000 2
Please wait for the HV-Tank to cool down.

< Example of error or warning message display>

Rev. C 56/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.3.2 Display Error Code

Error Code Example:


 Err code: 001001(Err 001000 + Err 000001)

: ERROR RESET

Rev. C 57/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

8.3.3 Trouble shooting check list


Please fill in following the state check list.
1) Main board LED Display State

LED STATE
I.I Size
D21 D22 D23
(ON or OFF) (ON or OFF) (ON or OFF) 9 inch

(ON or OFF) (ON or OFF) (ON or OFF) 6 inch

(ON or OFF) (ON or OFF) (ON or OFF) 4.5 inch

LED STATE
D19 (ON or OFF)

Rev. C 58/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2) Main Control board LED Display

LED STATE
D7 (Flashing ON or OFF)

D8 (Flashing ON or OFF)

D9 (Flashing ON or OFF)

D10 (ON or OFF)

D11 (ON or OFF)

D12 (ON or OFF)

D13 (ON or OFF)

D14 (Flashing ON or OFF)

Rev. C 59/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

3) Main Control board DIP Switch Position

DIP S/W 1 STATE

S/W1 (ON or OFF)

S/W2 (ON or OFF)

S/W3 (ON or OFF)

S/W4 (ON or OFF)

S/W5 (ON or OFF)

S/W6 (ON or OFF)

S/W7 (ON or OFF)

S/W8 (ON or OFF)

DIP S/W 2 STATE

S/W1 (ON or OFF)

S/W2 (ON or OFF)

Rev. C 60/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

4) X-ray Control Board LED Display

LED STATE
D18 (ON or OFF)

D20 (ON or OFF)

D22 (ON or OFF)

D24 (ON or OFF)

D4 (ON or OFF)

D5 (ON or OFF)

D6 (ON or OFF)

D7 (Flashing ON or OFF)

D32 (ON or OFF)

Rev. C 61/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

5) X-ray Control Board DIP Switch Position

DIP S/W 1 Description

S/W 1 (ON or OFF)

S/W 2 (ON or OFF)

S/W 3 (ON or OFF)

S/W 4 (ON or OFF)

S/W 5 (ON or OFF)

S/W 6 (ON or OFF)

Rev. C 62/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6) Driver Board and LED

LED STATE
D29 (ON or OFF)

D30 (ON or OFF)

D31 (ON or OFF)

D32 (ON or OFF)

Rev. C 63/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(7) kV, mA Reference Waveform Check


1) Turn off AC switch
2) Please connect probe to EP_REF(kV feedback_TP3) ,IP_REF (mA feedback_TP14) test point on
operation board below picture.
3) Turn on AC switch.
4) Exposure X-ray

5) Please disconnect DC Line during kV, mA reference voltage test as below

Rev. C 64/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

6) kV / mA reference table

kV / V REF Voltage mA / V REF Voltage

R-mode F-mode R-mode F-mode

1 40kV / 2V 40kV / 2V 1 20mA / 2V 0.5mA / 0.2V

2 50kV / 2.5V 50kV / 2.5V 2 40mA / 4V 1mA / 0.4V

3 60kV / 3V 60kV / 3V 3 60mA / 6V 2mA / 0.8V

4 70kV / 3.5V 70kV / 3.5V 4 80mA / 8V 3mA / 1.2V

5 80kV / 4V 80kV / 4V 5 100mA / 10V 4mA / 1.6V

6 90kV / 4.5V 90kV / 4.5V 6 5mA / 2.0V

7 100kV / 5V 100kV / 5V 7 10mA / 4.0V

8 110kV / 5.5V 110kV / 5.5V

9 120kV / 6V 120kV / 6V

10 125kV / 6.25V 125kV / 6.25V

Rev. C 65/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

mAs TABLE

Anyview-500R (5kw Rotor) 125KV, 100mA


KV 40KV ~ 50KV
mA 100 80 60
mAs 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100 125 150 200
Sec 0.008 0.010 0.013 0.016 0.020 0.025 0.032 0.040 0.050 0.064 0.080 0.100 0.163 0.200 0.250 0.313 0.400 0.667 0.833 1.067 1.333 1.667 2.083 2.500 3.333
51KV ~ 60KV
80 60 40
0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100 125 150 200
0.010 0.013 0.016 0.020 0.025 0.031 0.040 0.050 0.063 0.080 0.100 0.125 0.217 0.267 0.333 0.417 0.533 1.000 1.250 1.067 2.000 2.500 3.125 3.750 5.000
61KV ~ 70KV
70 50 30
0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100 125 150
0.011 0.014 0.019 0.023 0.029 0.036 0.046 0.057 0.071 0.091 0.114 0.143 0.260 0.320 0.400 0.500 0.640 1.333 1.667 2.133 2.667 3.333 4.167 5.000
71KV ~ 80KV
60 40 30
0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
0.013 0.017 0.022 0.027 0.033 0.042 0.053 0.067 0.083 0.107 0.133 0.167 0.325 0.400 0.500 0.625 0.800 1.333 1.667 2.133 2.667 3.333
81KV ~ 90KV
50 40 30
0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
0.016 0.020 0.026 0.032 0.040 0.050 0.064 0.080 0.100 0.128 0.160 0.200 0.325 0.400 0.500 0.625 0.800 1.33 1.67 2.13 2.67 3.33
91KV ~ 100KV
40 30 20
0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
0.020 0.025 0.033 0.040 0.050 0.063 0.080 0.100 0.125 0.160 0.200 0.250 0.433 0.533 0.667 0.833 1.067 2.000 2.500 3.200 4.000 5.000
101KV ~ 110KV
30 20
0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
0.027 0.033 0.043 0.053 0.067 0.083 0.107 0.133 0.167 0.213 0.267 0.333 0.433 0.533 0.667 0.833 1.067 2.000 2.500 3.200 4.000 5.000
111KV ~ 125KV
25 20
0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
0.032 0.040 0.052 0.064 0.080 0.100 0.128 0.160 0.200 0.256 0.320 0.400 0.520 0.640 0.800 1.000 1.280 2.000 2.500 3.200 4.000 5.000

Rev. C 66/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(8) kV, mAs feedback waveform check


1) Turn off AC switch
2) Please connect probe to EP_FB (kV feedback_TP1), IP_FB (mA feedback_TP15) test point on
operation board below picture.
3) Turn on AC switch.
4) Exposure X-ray

5) kV, mA feedback voltage table

kV/kV feedback Voltage mA/mA feedback Voltage


R-mode F-mode R-mode F-mode
1 40kV / 2V 40kV / 2V 1 20mA / 2V 0.5mA / 0.2V

2 50kV / 2.5V 50kV / 2.5V 2 40mA / 4V 1mA / 0.4V

3 60kV / 3V 60kV / 3V 3 60mA / 6V 2mA / 0.8V

4 70kV / 3.5V 70kV / 3.5V 4 80mA / 8V 3mA / 1.2V

5 80kV / 4V 80kV / 4V 5 100mA / 10V 4mA / 1.6V

6 90kV / 4.5V 90kV / 4.5V 6 5mA / 2.0V

7 100kV / 5V 100kV / 5V 7 10mA / 4.0V

8 110kV / 5.5V 110kV / 5.5V

9 120kV / 6V 120kV / 6V

10 125kV / 6.25V 125kV / 6.25V

Rev. C 67/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

A. F-mode

1) 40kV/0.5mAs waveform

Please set kV, mAs below table and take exposure

Exposure condition : 40kV/0.5mAs


CH 1: kV 2V/div
CH 2: mA 2V/div

Setting value EP_FB IP_FB

Ref voltage 40kV 2V 0.2V

Pass/Fail Pass Fail Pass Fail

Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. C 68/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2) 80kV/0.5mAs waveform

Please set kV, mAs below table and take exposure

Exposure condition : 80kV/0.5mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div

Setting value EP_FB IP_FB

Ref voltage 80kV 4V 0.2V

Pass/Fail Pass Fail Pass Fail

Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. C 69/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

3) 110kV/0.5mAs waveform

Please set kV, mAs below table and take exposure

Exposure condition : 110kV/0.5mAs

CH 1: kV 2V/div

CH 2 : mA 2V/div

Setting value EP_FB IP_FB

Ref voltage 110kV 5.5V 0.2V

Pass/Fail Pass Fail Pass Fail

Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. C 70/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

B. R-mode

1) 50kV/1mAs/40mA waveform

Please set kV, mAs below table and take exposure

Exposure condition : 50kV/1mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div

Setting value EP_FB IP_FB

Ref voltage 50kV 2.5V 4V

Pass/Fail Pass Fail Pass Fail

Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. C 71/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

2) 80kV/1mAs waveform

Please set kV, mAs below table and take exposure

Exposure condition : 80kV/1mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div

Setting value EP_FB IP_FB

Ref voltage 80kV 4V 4V

Pass/Fail Pass Fail Pass Fail

Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

Rev. C 72/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

9. SPECIFICATION
9.1 SPECIFICATION TABLE
Model Name Anyview-500R
1) System power requirement
(1) Line voltage 220-230V~, 50/60Hz

(2) Line phase Single

(3) Input power (at 230VAC) Standby: 50VA, Momentary: 11.5kVA

2)X-ray Generator
(1)Type HFG INVERTER

(2)Output power 5.0kW

(3)Fluoroscopy

- Continuous mode 0.5-10mA

- Pulsed mode 0.5-20mA

- Boost mode 30mA

- DS mode 10-30mA

- ABS, IDR, MAR mode YES

(4)Radiography mode

-kV range 40-125kV

-mA range 20-100

-mAs range 0.8~200mAs

(5)X-ray tube XRR-2251

3)X-ray tube
(1) X-ray tube type Rotating

(2)Max kV 125kV

(3)Focal spot

-Small 0.3mm

-Large 0.6mm

(4)Target angle 10°

(5)Anode heat storage capacity 210kHU

4)Collimator
(1)Type Motorized

(2)Opening/Closing 2 way

(3)Rotation 360°

(4)Iris Motorized

5)Image Intensifier
(1)Input field view size 9inch

(2)Nominal entrance field size 9 / 6 / 4.5 inch

(3)Central resolution(lp/cm) 52 / 58 / 68

(4)Contrast ration(10%) 30:1

6)CCD camera
(1)Type CCD

Rev. C 73/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

(2)Pixel Size 1K*1K

7)Monitor (Included by Monitor stand)


(1)Type TFT-LCD DUAL

(2)Size 19inch

(3)q'ty 2

8) Digital Imaging System(DIS)


(1)Type AIS

Intel Core i3-2130(3.4GHz) or higher


Windows 7 or higher
(2) PC Spec
4GB DDR3 SDRAM,1600MHz-1X4GB or higher
500GByte Hard drive or higher

9)Reproducibility :

(1) Coefficient of Variation kV<0.005, Time<0.005, <mAs<0.001

10)Accuracy
kV<(±1%+1kV), mA<(±4%+1mA), mAs<(±4%+0.1mAs)

11)Linearity
Coefficient of Linearity <0.01:CL=(X1-X2)/(X1+X2),where X is mR/mAs

12)Weight
(1)G-weight
(2)Net-weight

● If needs the monitor and the pc manual, we will provide separately.


● Anyview-500R mAs range: 0.8 - 200mAs, 25step

0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0


6.4 8.0 10 13 16 20 25 32 40
50 64 80 100 125 150 200

● RADIATION FIELD
- 25cm*25cm(Circle) at 1m (SID)

● Significant Zone of Occupancy


- To refer below appendix C.2

● Apparent Mains Resistance

Momentary Apparent
Mains Voltage
Line Current Mains Resistance
220V 50A 0.08Ω
230V 48A 0.0833Ω

Rev. C 74/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

● Circuit Breaker Specification


- Voltage Rating: 250VAC
- Limit Current: 20A

● Emission & Filament Characteristics of X-ray Tube (XRR-2251)


- Construction: Rhenium-Tungsten faced Molybdenum
- Inherent Filtration: At least 0.7 mmAl at 75kV

Rev. C 75/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

9.2 MECHANICAL SPECIFICATION

(1)C-ARM movement

C-ARM Inner circle space 790mm

C-ARM depth 695mm

C-ARM Orbital movement 135


Horizontal movement 200mm

Vertical movement 500mm

PANNING  12.5
Width 800mm

(2)C-ARM rotation

rotation range right 180°, left 180°

Locking Manual

SID

Fluoro 1m

RAD 1m

(3)COLLIMATOR

Operating type REMOTE MOTOR DRIVE

Structure Lead Shutters

Speed Rotation

Open –Close Maximum less than 15 sec

(4) Accessories

CASSETTE HOLDER CASSETTE SIZE 8  10 Inch


GRID spec 8: 1, 60 L/cm

Disinfected covers two sets of three kinds of cover (Four of clips)

Rev. C 76/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

10. ENVIRONMENTAL SPECIFICATIONS

OPERATING
Ambient temperature range 10°C to 40°C (50°F to 104°F)
Relative humidity range 30% to 75%, non-condensing
Atmospheric pressure range 800hPa to 1060hPa

TRANSPORT AND STORAGE


Ambient temperature range -10°C to 70°C (14°F to 158°F)
Relative humidity range 5% to 95%, non-condensing
Atmospheric pressure range 500hPa to 1100hPa

This information is provided to help you establish safe operating conditions for both you and your X-
ray system.

WARNING

DO NOT OPERATE THIS X-RAY SYSTEM EXCEPT IN ACCORDANCE WITH INFORMATION


INCLUDED IN THIS SECTION, AND ANY ADDITIONAL INFORMATION PROVIDED BY THE
X-RAY SYSTEM MANUFACTURER AND/ OR COMPETENT SAFETY AUTHORITIES.

Rev. C 77/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

11. CUSTOMER SUPPORT


Address any questions regarding X-ray system operation to:

ECOTRON Co.,Ltd
4F & 5F, Hanshin IT Tower II, 47, Digital-ro 9-gil,
Geumcheon-Gu, Seoul 08511, Republic of Korea
TEL:+82-2-2025-3760, FAX:+82-2-2025-3764
E-mail: ecotron@ecotron.co.kr
Web-site: http://www.ecotron.co.kr

In USA
Contact DRGEM USA Inc.
TH
7018 NW 50 Terrace, Gainsville, Florida, 32653, USA
TEL:201-370-6672, FAX: 352-337-1271
E-mail: drgemusa@gmail.com

Obelis S.A
Bd. General Wahis 53, 1030 Brussels, Belgium / E-mail:mail@obelis.net
Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03

Rev. C 78/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

12. MAINTENANCE PROCEDURE

12.1 GENERAL CAUTION


It is recommended to follow the maintenance procedure described below, for the reliable operation. The
routine inspection should be committed by a trained expert

12.2 MAINTENANCE SCHEDULE

Maintenance/control item Period

RAD mA :
1 1 Year
Should be within accuracy boundary of ±3%, with the voltage set by 40 kV and 110 kV.

Fluoro mA :
2 1 Year
Should have accurate 1mA step for 1~5mA, with the voltage range of 40 ~ 110 kV
RAD mAs :
3 Should be within accuracy boundary of ±5%, when the mAs value is between 0.4mAs and 1 Year
100 mAs.
RAD KV :
4 Should be within accuracy boundary of ±10%, when the kV value is between 40kV and 1 Year
110kV.
Fluoro KV:
5 1 Year
Should be within accuracy boundary of ±3%, with the voltage range of 40 ~ 110 kV
Optical Maintenance :
6 When dirt appears on the image, remove dirt on the camera lens with soft brush or ‘Kodak 1 Year
Lens Cleaner’.
DAP (Option):
7 5 Year
Should be calibrated for accuracy of DAP information.

12.3 THE EQUIPMENT OPERATION CHECKLIST FOR OPERATOR


Operator needs to conduct the maintenance once a month, by the following procedures.

Check Item Period


ABS:
1 After selecting ABS, use Phantom to take images of particular parts (eg: hand, foot, 1 Month
pelvis, vertebrae etc), to check whether kV and mA changes automatically.

12.4 GLASS IMPLOSION OF I.I


The X-ray image intensifiers (I.I.) are a high-vacuumed electron tube containing glass. Do not scratch or strain
the glass of the I.I. for the prevention of the implosion.
(1) Because of the risk of implosion, a protection mask shall be put on and extreme care shall be taken when
manipulating the bare tube.
(2) The aluminum plate (for light-shielding) on the entrance plane of housing also plays as a protection against
the injury due to implosion. When removing the plate, the input side shall not face a person. The protection
plate such as anti-scattering grid is recommended when removing the plate.

Rev. C 79/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

12.5 CLEANING
Keep the machine clean when it gets dirt, especially after a surgery. To clean the main-body, turn off the
power and pull the plug out from the outlet. Scrub surfaces smoothly with soft fabric or gauze. Keep the
machine clean when it gets dirt, especially after a surgery.

System power On/off

To clean the main-body, turn off the power and pull the plug out from the outlet as pictures above. Do not allow
liquids and water enter inside the unit, to prevent from short circuits and corrosion of inner parts.

WARNING

Make sure the disconnection from the mains before cleaning the unit!

Clean the painted surfaces using only a damp cloth


and neutral detergents. Take care to dry the parts
with a dry cloth. Do not use corrosive solvents
(Alcohol, Gasoline)

Clean the extra covers, carrying handles, foot switch, hand switch, breakers, monitors and the
accessories using a damp cloth.
Periodically and when necessary decontaminate the op panel, the carrying handles, the breakers, the
surfaces of monitors using a proper decontaminator.

WARNING

Do not keep near the magnetic stuffs and water pails etc. They might bring a serious
harm to the system.

Rev. C 80/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

12.6 THE END OF PRODUCT LIFE

If the X-ray system has completed its useful service life, local environmental
regulations must be complied with in regard to disposal of possible hazardous
materials used in the construction of the generator.
In order to assist with this determination, the noteworthy materials used in the
construction of this generator are itemized below:

ITEM

• Electrical insulating oil in HT tank. This is a mineral oil with trace additives (11 Liter)

• Solder (lead/tin).

• Epoxy fiberglass circuit board materials, tracks are soldering on copper.

• Wire, tinned copper. Insulated with PVC or silicone.

• Steel and / or aluminum (x-ray system body and console chassis).

• Plastic (console membrane).

• Electrical and electronic components: IC’s, transistors, diodes, resistors, capacitors, etc.

Rev. C 81/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

13. QUALITY WARRANTY


 Scope & Duration of Warranty
ANYVIEW-500R manufactured by ECOTRON Co., Ltd. are warranted to be free from defects for a period of
one year after purchase date. If during the warranty period the product you purchased is found to be defective,
it will be repaired free of charge.

In the case of one of the following, however, a certain amount of service fees will be charged.
• Defect or damage found after the warranty period.
• Defect or damage in appearance which is not related to main function of the system.
• Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes.
• Damage resulting from either improper movement or inattention to the precautions.
• Damage resulting from repair or modification by someone other than ECOTRON or the authenticated by
ECOTORN Co., Ltd.
• Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes.
• Incidental or indirect loss caused during system manipulation.

Any or all defect or damages in appearance, which do not affect the main functions of the product are
not covered by this free of charge warranty.

 Prerequisites for Repair Request


- When a defect is found, stop the using immediately. It is strongly recommended to refer to related
material on the Service Manual.
- Before a service request, must power off the entire system and check the model number, serial
number, and the purchase date. Then contact an authorized service office.
 Any product with a defect in appearance only shall not be returned to nor replaced by ECOTRON Co., Ltd.
ECOTRON Co., Ltd. shall not be liable any incidental or consequential damages arising out of or relating
to the use of the product.
 ECOTRON Co., Ltd. shall not be liable for any damages or losses occurring after the warranty period.
 This Quality Warranty prevails over the detailed Warranty for fitness or all other warranties in relation to
the product.

Rev. C 82/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

APPENDIX A EMC COMPATIBILITY

Electromagnetic emission

The ANYVIEW-500R is intended for use in electromagnetic environment specified below. The customer or the
user of the ANYVIEW-500R should assure that it is used in such an environment.

Emission test Compliance Electromagnetic environment - guidance

The ANYVIEW-500R uses RF energy only for its


internal function. Therefore, its RF emissions are very
RF emissions – CISPR11 Group 1
low and are not likely to cause any interference in
nearby electronic equipment.

The ANYVIEW-500R is suitable for use in all


establishments (i.e. hospitals, doctors practice etc.)
RF emissions- CISPR11 Class A
other than domestic. ANYVIEW-500R is intended for
professional use only.

Harmonic emissions IEC


Class A
61000-3-2

Voltage fluctuations/flicker
emissions IEC6100-3-3

Rev. C 83/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

Electromagnetic immunity

The ANYVIEW-500R is intended for use in the electromagnetic environment specified below. The customer or
the user of the ANYVIEW-500R should assure that it is used in such an environment.

Electromagnetic environment –
IMMUNITY test IEC 60601 test level Compliance level
guidance
Floors should be wood, concrete or
Electrostatic
± 6 kV contact ceramic tile. If floors are covered
discharge (ESD)
± 8 kV air with synthetic material, the relative
IEC 61000-4-2
humidity should be at least 30 %.

Electrical fast ± 2 kV for power Mains power quality should be that


supply lines
transient/burst of a typical commercial or hospital
± 1 kV for input/output
IEC 61000-4-4 lines environment.

± 1 kV line(s) to Mains power quality should be that


Surge
line(s) of a typical commercial or hospital
IEC 61000-4-5
± 2 kV line(s) to earth environment.

<5 % UT
(>95 % dip in UT)
for 0,5 cycle Mains power quality should be that
of a typical commercial or hospital
Voltage dips, short 40 % UT environment. If the user of the [ME
interruptions and (60 % dip in UT) EQUIPMENT or ME SYSTEM]
voltage variations for 5 cycles requires continued operation during
on power supply power mains interruptions, it is
70 % UT
input lines (30 % dip in UT) recommended that the [ME
IEC 61000-4-11 for 25 cycles EQUIPMENT or ME SYSTEM] be
powered from an uninterruptible
<5 % UT power supply or a battery.
(>95 % dip in UT)
for 5 s

Power frequency Power frequency magnetic fields


(50/60 Hz) should be at levels characteristic of
3 A/m
magnetic field a typical location in a typical
IEC 61000-4-8 commercial or hospital environment.

NOTE UT is the a.c. mains voltage prior to application of the test level.

Rev. C 84/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

Electromagnetic immunity
The ANYVIEW-500R is intended for use in the electromagnetic environment specified below. The customer or
the user of the ANYVIEW-500R should assure that it is used in such an environment.
IMMUNITY test IEC 60601 Compliance
Electromagnetic environment – guidance
TEST LEVEL level
Portable and mobile RF communications
equipment should be used no closer to any part of
the ANYVIEW-500R, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.

Recommended separation distance


Conducted RF 3 Vrms [V1] V
IEC 61000-4-6 150 kHz to 80 MHz

80MHz to 800 MHz


Radiated RF 3 V/m [E1] V/m
IEC 61000-4-3 80 MHz to 2,5 GHz 800MHz to 2,5 GHz

where P is the maximum output power rating of


the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as


determined by an electromagnetic site survey,a
should be less than the compliance level in each
b
frequency range .

Interference may occur in the vicinity of equipment


marked with the following symbol:

NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
ANYVIEW-500R is used exceeds the applicable RF compliance level above, the ANYVIEW-500R should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the ANYVIEW-500R.

b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.

Rev. C 85/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

Recommended separation distances between portable and mobile RF communications


equipment and the ANYVIEW-500R.
The ANYVIEW-500R is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the ANYVIEW-500R can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the ANYVIEW-500R as recommended below, according to the maximum output power of
the communications equipment.

Rated maximum output Separation distance according to frequency of transmitter (m)


power of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
W

0.01

0.1

10

100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

Rev. C 86/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

APPENDIX B LABEL

LABEL

12

1 11
11
5

14, 15

13

6
7

9
8

Rev. C 87/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

3
10, 11

Rev. C 88/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

The labels shown in the following table are merely reproductions. Their sizes and colors may differ
from reality and serve illustrative purposes only.

NO LABEL COMMENTS

MOBILE C-ARM
/ MOBILE C-ARM X-RAY
EQUIPMENT
1
ANYVIEW-500R

X-RAY GENERATOR

ANYVIEW-500R

2 MAX KVP
FILTRATION
FOCAL SPOT
TUBE TYPE
TUBE SERIAL NO.

For systems with DIS


3
ANYVIEW-500R

For systems with Collimator


4
ANYVIEW-500R

5 X-ray Radiation

Rev. C 89/90 ECOTRON Co., Ltd.


ANYVIEW-500R SERVICE MANUAL

System with laser positioning


6 device only

Systems with laser positioning


7 device only

Systems with laser positioning


8
device only

Systems with laser positioning


9 device only

10 HEAVY OBJECT

Read the User’s manual or


11
Operator manual

12 CE0123 (EU countries)

13 Radiation Warning

14 HV 3 Minutes Warning

15 Danger High Voltage

16 Safe Use Warning Label

Rev. C 90/90 ECOTRON Co., Ltd.