SERVICE MANUAL
V. 1.13 07/2006 ENGLISH
HOSPITEX DIAGNOSTICS
Via Provinciale Lucchese, 145 - 50019
H Sesto Fiorentino
H
FIRENZE - ITALY
Ph +39 055 374083 Fax + 39 055 374084
e-mail : hospitex@hospitex.it
web : http://www.hospitex.com
Hospitex Diagnostics Screen Master Touch – Service manual
Index:
page n°
SERVICE MENU 8
FILTER SEL. 9
FILTER MOT. 10
PUMP
PUMP MOT. 11
READING AD 12
T. CUVETTE 13
T.INCUBATOR 14
WAIT PUSH 15
READING ABS 16
WASHING 18
DOWN LOAD 18
UPLOAD 18
CALIBRATION PUMP 19
CALIBRATION TS (TOUCH SCREEN) 22
HOST 23
Check transformer 24
Check voltage test point 24
Check continuous Voltage 24
Touch Screen Contrast: 24
Check linearity 25
OPTICAL GROUP 25
1
Hospitex Diagnostics Screen Master Touch – Service manual
Introduction
Thank you for choosing Hospitex Diagnostics SCREEN MASTER TOUCH, a new instrument
designed with powerful features in a compact unit for biochemical investigations, enzymes,
turbidimetry and drug tests. With such characteristics as the easy to operate integrated touch screen,
the Screen Master Touch is the perfect solution for you laboratory.
2
Hospitex Diagnostics Screen Master Touch – Service manual
Technical specifications
3
Hospitex Diagnostics Screen Master Touch – Service manual
Located in the rear panel (from left to right as shown below) you will find:
Fuses
Drow
WASTE OUTLET: Before switching on the instrument, remember to connect the plastic outlet
connector (red) to a waste tank by means of a silicone tube.
4
Hospitex Diagnostics Screen Master Touch – Service manual
The instrument should be located in a clean environment, placed on a stable surface, and away
from direct sunlight, which could affect the operating temperature and the quantity of light
measured by the instrument.
- Please check the setting of the power supply switch according to your country’s electrical
network.
- Connect the power plug to a good grounded AC wall outlet, preferably one that is not
shared with other electric appliances and with low fluctuation of line voltage compared to
the standard voltage specified (10-15%).
- Keep the instrument away from other appliances that generate high frequency electrical
noises (e.g. radiological instruments).
- Before connecting the power cord, check that the AC power supply corresponds to the value
that is stated on the instrument’s label.
5
Hospitex Diagnostics Screen Master Touch – Service manual
Caution
Do not connect the instrument to a power supply different from the value indicated on the label.
- Before connecting the power and finishing the installation section, make sure that the
instrument is turned off (check the polysnap located on the rear part of the instrument).
- Make sure that your AC main line has an efficient ground line. A bad ground line
connection may compromise analysis results and damage the instrument.
- After turning on the instrument, pay attention not to spill liquids or micro solid substances
on the surface around the instrument.
If the above procedures are carefully followed, you are now ready to TURN ON the
instrument by using the switch located on the rear panel.
Measurement block
Open the measurement block cover lid, positioned on the top left part of the instrument. Inside you
will see:
- Peristaltic pump that allows the solution to enter the instrument for measurement.
- Incubator well, which keeps the flow-cell or manual cuvette at the temperature of
approximately 37° C.
- Inlet tube.
6
Hospitex Diagnostics Screen Master Touch – Service manual
Tube setting
Make sure that the transparent tube (the one that connects the flow-cell with the internal hydraulic
circuit) is not twisted or squashed.
Place the red tube around the peristaltic pump and make sure that the white plastic connections are
set as shown in the figure below:
INLET TUBE
PUMP
CUVETTE
POSITION
RED TUBE
Under the measurement block there is a lever (the PUSH button). Press the lever for no more than
half a second to pump samples into the flow cell.
MEASUREMENT
PUSH button
INLET PIPE
7
Hospitex Diagnostics Screen Master Touch – Service manual
Incubator
The incubator temperature is brought up to 37 °C by the software and it will remain constant until
the instrument is turned off.
NOTE: It is very important to heat the macro cuvettes to the proper temperature for obtaining the
most accurate analysis results.
Printer settings
GREEN LEVER
8
Hospitex Diagnostics Screen Master Touch – Service manual
SERVICE MENU:
1 2
9
Hospitex Diagnostics Screen Master Touch – Service manual
1. FILTER SEL.
2. FILTER MOT.*
3. PUMP*
4. PUMP MOT.*
5. READING AD*
6. T. CUVETTE
7. T.INCUBATOR
8. WAIT PUSH*
9. READING ABS
10. WASHING
11. DOWN LOAD
12. UPLOAD
13. CALIBRATION PUMP
14. CALIBRATION TS (TOUCH SCREEN)
15. HOST
(*) : This menus are dedicated for internal company testing procedure. No intended use for
service.
1. FILTERS SEL.
1)
10
Hospitex Diagnostics Screen Master Touch – Service manual
3. Prepare a white paper sheet of 1x5cm and insert it in the reading well as shown in
picture bellow.
4. Insert paper attach to the preamp side
5. A green circle within a square will be designed by the light of lamp
6. Loose the screws of the black metal lamp support
7. Adjust lamp position operating on lamp support in order to have the square perfectly
centered in the circle (see drawing).
Green circle of
light
Filter wheel
Reading well
Lamp support
11
Hospitex Diagnostics Screen Master Touch – Service manual
In order to rotate the filter wheel and set the selected filter wavelength under the light path,
proceed as follow:
1. Press the “SEL. FILTER” message and the sub-menu will appear.
2. Use the right and left arrows to select a specific filter wavelength, and then press Exec.
3. The filter wheel will rotate in order to set the selected filter under the light path.
4. 340nm> no clour
5. 405> Violet
6. 505>Dark Green
7. 546> light Green
8. 578>Yellow
9. 630>Red
2. T. CUVETTE
Press the “T. CUVETTE” message and the sub-menu will appear.
The 1° row stated as: “Temp. Cuvet” will show the flowcell current temperature.
In case the temperature need adjustment, proceed as follow:
This screen allows you monitoring temperature gradients of incubator and cuvette according to
temperature displayed in user program and digital level acquired by microcontroller (both current
and target).
12
Hospitex Diagnostics Screen Master Touch – Service manual
3. T.INCUBATOR
This screen allows you monitoring temperature gradients of incubator and cuvette according to
temperature displayed in user program and digital level acquired by microcontroller (both current
and target).
The message: ”Temperature” will show the target temperature to be reached into incubator
block.
13
Hospitex Diagnostics Screen Master Touch – Service manual
4. READING ABS
Press the “READING ABS” message and the sub-menu will appear.
In this menu it is possible to read the Photocell amplifier energy, and to perform an Absorbance
reading.
Select Filter: The filter that operator want to set, use the right and left arrows to select:
(340nm 405nm, 505nm, 546nm, 578nm and 630nm)
Actual Filter: The filter currently placed under the light beam
Absorbance: The Abs value of the substance present in the flow cell.
Tension: The mVolt value present on the Photodiode preamp.
1. Select filter with arrows then press wheel icon, the desired filter will be placed under the
light beam
2. In order to obtain the Abs value make and aspiration with water and wait for
measurement.
3. Set the test tube of water (or other substances) in the inlet pipe and press the PUSH
button.
4. After a few seconds, continuous evolution of the ABS value will appear on the display.
5. Press 0 key to have zero on the current ABS value (water or other substances). You will
choose this option when you will set to zero the instrument.
6. Prepare a new sample (reagents, potassium-dichromate or others) and press the PUSH
button to start a new ABS measurement session.
7. At the screen exit, the instrument expels the volume from flow cell via the outlet waste.
14
Hospitex Diagnostics Screen Master Touch – Service manual
In “READING ABS” screen you can monitor (after having filled water) tension coming from
optical group (OPT TENS) and tension coming in A/D converter (A/D TENS) , using a particular
filter (FILTER)
Tension Mode Measurement is very useful to verify lamp calibration, filter / lamp / lens /
photodiode / preamp board quality, position sensor working.
Moreover in this screen you can test correct pump motion.
If you have a correct calibrated optical group, when you use the flow cell, the TENSION
values are between 250 mV (for 340 nm) and about 3.5 V (for 630 nm) for all filters.
Regulate trimmer R1 to obtain 3,5 V on the Tension menu (see picture bellow):
15
Hospitex Diagnostics Screen Master Touch – Service manual
5. WASHING
When pressing this message the peristaltic pump will be activated for around 10 seconds.
6. DOWN LOAD
7. UPLOAD
16
Hospitex Diagnostics Screen Master Touch – Service manual
8. CALIBRATION PUMP
Now the pump work following a determinate step numbers 4790 is the default pump step number
for the peristaltic pump, the instrument should aspirate correctly with this number.
In case the aspiration needs to be calibrated, operate as follow:
Search with down scroll arrow the “Pump calibration” when highlighted press it the
following screen will appear:
Pump Calibration
Steps for 10 ml 4795
17
Hospitex Diagnostics Screen Master Touch – Service manual
Exit from service and from main menu enter in setup menu
Look for “Automatic empty” menu and insert 5000 value (air gap delay time, equal
to 5 seconds), then save. See also user manual.
Enter in Run method and run glucose.
Prepare 500 l of water in a cuvette set the aspiration tip in a bottom angle of
manual cuvette and press aspiration button.
If the pump works correctly, the water should be aspired completely from the
manual cuvette, and water should stop at the top of the aspiration tip.
After 5 seconds the air gap will be aspired and liquid front will rise up.
Repeat operation several times
Cuvette hose
Manual
cuvette
Flow cell
After the complete aspiration (sample aspiration + air gap), if water stops after at beginning of
aspiration tip or stops to near the cuvette the red tube has a problem and you have to adjust
aspiration by changing the aspiration pump steps.
18
Hospitex Diagnostics Screen Master Touch – Service manual
19
Hospitex Diagnostics Screen Master Touch – Service manual
Scroll down to the Touch screen Calibration menu and select it:
Inside menu Touch with a pencil and an indicator will appear in correspondence of the pencil tip.
The tip of the pencil must be in the same position of the indicator, try different position of touch,
especially the one where you report the problem on keyboard.
In case the tip and indicator are too far, you need to recalibrate touch screen.
Press Door to exit
20
Hospitex Diagnostics Screen Master Touch – Service manual
10. HOST
A Host program manage (Running under Windows XP and supplied separately) to operate the
Screen Master Touch from an external PC.
In the Host program it is located all the features of the SMT software, plus an extended
database and a large choice of printer will able to be used with SMT.
In order to connect the Screen Master Touch to a Host Computer via RS232 proceed as follow:
21
Hospitex Diagnostics Screen Master Touch – Service manual
1 0 Vac
2 16 Vac
3 –
4 14 Vac
5 0 Vac
6 14 Vac
In order to regulate the contrast for the Touch Screen adjust RV1 placed on the Touch screen board,
see picture bellow:
22
Hospitex Diagnostics Screen Master Touch – Service manual
OPTICAL GROUP
2 5
2 4
1
6
INCUBATOR 7
3
23
Hospitex Diagnostics Screen Master Touch – Service manual
PRINTER
Open case of the instrument, pulling out four screws down the equipment and three ones on the
back side.
Pull out two screws at right and left sides of cuvette site, and move away pump support (the one
with pump rotor).
Move away the little tongue that fixes the lamp, by pulling out the only screw.
Extract old lamp, and set carefully the new one, bringing it up to the bottom of the site (pay
attention to not touch lamp glass with finger skin).
Set up again the little tongue, and check the top of the lamp being inside the little hole.
Set up again pump support and equipment case.
24
Hospitex Diagnostics Screen Master Touch – Service manual
Type A
1. Disconnect the red tube from peristaltic pump by pulling one of the white holders.
2. Free the red tube from holders by pulling the two white pieces apart:
Type A
3. Measure the red tube with a ruler, the length must be 10 cm, if not cut the exceeding of the
tube with sharp cutter and set the tube back. Then check aspiration as appendix 2.
NOTE: Sometimes red tubs go longer than 10cm especially in hot whether
Type B
1. Disconnect the red tube from peristaltic pump by pulling the red holders.
2. Free the red tube from connector by pulling the two white connector out from red tube:
3. Insert a new tube
White Connector
Type B
Holder
s
25
Hospitex Diagnostics Screen Master Touch – Service manual
Main Board:
26
Hospitex Diagnostics Screen Master Touch – Service manual
The Touch Screen board carries a “Managing Firmware” of the instrument. This firmware
can be loaded directly from serial port present in the rear panel. The 99% of upgrades will
be performed on this board.
The main board controls all the peripherals (device), frequency of re- programming for the
“Peripherals Firmware” is rare. In order to upgrade the it is necessary to open the
instrument.
In order to upgrade the instrument firmware, please follow step by step the bellow stated
instruction:
1. Connect one side of the re-program cable to one of the available PC COM sockets and the
other side to the SMT RS232 sockets
2. Close the contact present on the re-program cable PC/instrument.
3. Switch on the instrument.
4. Start on PC the program FlashSta and perform the following operations:
Select the COM to which the re-program cable has been connected, then press OK
27
Hospitex Diagnostics Screen Master Touch – Service manual
Press OK
Press Setting…
28
Hospitex Diagnostics Screen Master Touch – Service manual
Select Baud rate 57600 and Program intervals (ms) 40 then press OK
The program will perform an automatic parameters settings (the operation will least few seconds)
Press E.P.R…
29
Hospitex Diagnostics Screen Master Touch – Service manual
Press OK
Now the programming mode start, it will last some seconds and on the PC screen will appear the
following:
The programming time is displayed by the blue bar; at the end of programming the blue bar will be
complete.
30
Hospitex Diagnostics Screen Master Touch – Service manual
1 g 1
Grenn
White
Blue
Cable
Switch
9 9
31
Hospitex Diagnostics Screen Master Touch – Service manual
In order to install the Peripherals Firmware on Screen Master Touch, follow the procedure stated
below.
Find on your PC, with “Open” buttom, the file “SM_002_13_Ott_2005.mhx” with which you
want re-program the instrument.
SM_002_13_Ott_2005.mhx
32
Hospitex Diagnostics Screen Master Touch – Service manual
4. Connect serial cable to PC and instrument (COM1), select from: Set Environment button
5. Open cover of SMT and connect special adapter to J28 (see board).
9. Push Reset button on the main board press OK in message then wait for the end of the
automatic programming session, then press OK.
In case of fail of the automatic software loading try the manual loading:
FEMALE MALE
1 1
6 6
9 9
5 5
33
Hospitex Diagnostics Screen Master Touch – Service manual
In order to upgrade the instrument Assay parameters and make a backup of all the assay parameters
present inside the SMT memory, please follow step by step the bellow stated instruction.
In order to operate these procedure the relevant program named SetupSmpDownload.exe must be
installed in the PC (the PC must be RS232 connection equipped)
In order to install SetupSmpDownload.exe, copy it on desktop then bobble click and follow
installation indications
After the installation is complete an icon will appear on desktop
Click on icon and the following program will appear:
Select Com by pressing Com button in case the PC has more than 1 com connection
Select File, open the file and select the Met_V002_16_Feb_2006.met (this is Hospitex
Diagnostics assay parameters adaptation) or other.
In case no file is present in Select file option, go in C\SMP\Method and copy the file
DOWNLOAD PROCEDURE: This program allows to load in the SMT automatically methods
parameters from a PC via RS232.
34
Hospitex Diagnostics Screen Master Touch – Service manual
1) Connect one side of the re-program cable to one of the available PC COM sockets and the
other side to the SMT RS232 sockets
2) Open the contact present on the re-program cable PC/instrument.
3) Switch on the instrument.
4) Start on PC the program SmpDownload.exe
5) Select File, open the file and select the Met_V002_16_Feb_2006.met, than press OK
button
6) In the SMT enter in service menu and select Download
7) In the SmpDownload program press Download button
8) The methods will be loaded into the SMT automatically
UPLOAD PROCEDURE: The same program allows to record the methods parameter stored
inside a SMT in the PC via RS232.
1) Connect one side of the re-program cable to one of the available PC COM sockets and the
other side to the SMT RS232 sockets
2) Open the contact present on the re-program cable PC/instrument.
3) Switch on the instrument.
4) Start on PC the program SmpDownload.exe
5) Press File name button and name a file where all the assays parameters present on SMT
will be stored in PC, then press Ok
6) In the SMT enter in service menu and select Upload
7) In the SmpDownload program press Upload button
8) The methods of SMT will be loaded into the PC automatically and stored in
C\SMP\Method
35