SIROCLAVE VS
SIROCLAVE VS

Repair and service instructions


German

https://translate.googleusercontent.com/translate_f 1/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 3
2
Sirona Dental Systems GmbH

General information

About this document SIRONA customer service center


This document describes the Should you get this repair manual
Repair of SiroClave B devices can not help once, then
(Mod. 01), SiroClave B (Mod. 03), our customer service
SiroClave VS (Mod. 01) and SiroClave Center on the phone number
VS (Mod. 03).
(+49) 62 51/16 - 16 16

available at any time.

General information

New from: 09.2006

https://translate.googleusercontent.com/translate_f 2/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Changes compared to the last edition:


Chapter or Abschnit t page
New edition

2nd SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 4
Sirona Dental Systems GmbH

Table of Contents
1 Safety instructions ................................................ ................ 5
2 Technical description ............................................... ........ 6
2.1 Exterior view - Siroclave B ............................................. ................................... 6
2.2 Exterior view - Siroclave VS ............................................. ................................ 7
2.3 Interior view - Siroclave B ............................................. ....................................8th
2.4 Interior view - Siroclave VS ............................................. .................................. 9
2.5 System components - Siroclave B (Mod.01) ........................................ ........... 10
2.6 System components - Siroclave B (Mod.03) ........................................ ........... 11
2.7 System components - Siroclave VS ............................................. ................... 12
2.8 Circuit diagram - Siroclave B (Mod. 01) ........................................ ........................... 13
2.9 Circuit diagram - Siroclave B (Mod. 03) ........................................ ........................... 14
2.10 Circuit diagram - Siroclave VS ............................................. ................................. 15
2.11 Technical data - Siroclave B ............................................ ......................... 16
2.12 Technical data - Siroclave VS ............................................ ...................... 17
3 Technical settings ............................................... ...... 18
3.1 Height compensation ................................................ ........................................ 18
3.1.1 Note on altitude compensation ............................................ .................. 18
3.2 Configuration of the main board .............................................. .......................... 19
3.3 Alignment of the printing board .............................................. .................................. 20
3.3.1 Software adjustment ............................................. ............................. 20
3.3.2 Hardware-side adjustment ............................................. ............................ 21
https://translate.googleusercontent.com/translate_f 3/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

3.4 Adjusting the temperature board .............................................. ......................... 22


3.4.1 Software adjustment ............................................. ............................. 22
3.4.2 Hardware comparison ............................................. ............................ 23
4 Error diagnosis ................................................ ..................... 24
4.1 Manual diagnostic mode ............................................... ............................... 24
4.1.1 Siroclave B ............................................. .................................................. .24
4.1.2 Siroclave VS ............................................. ................................................. 26
4.2 Vacuum test ................................................ .................................................. .... 28
4.3 Checking the vacuum pump .............................................. ............................... 30
4.3.1 Siroclave B ............................................. .................................................. .30
4.3.2 Siroclave VS ............................................. ................................................. 31
4.4 Testing a solenoid valve .............................................. .............................. 32
4.4.1 SIRAI type solenoid valve ............................................ ................................. 32
4.4.2 Solenoid valve type HONEYWELL ............................................ .................... 34
4.5 Checking the water flow meter .............................................. .................. 36
5 Assembly work ................................................ ................... 38
5.1 Annual maintenance ............................................... ............................................. 38
5.2 Disassembly of the housing .............................................. ................................ 40
5.2.1 MOD. 03 ................................................. .................................................. .40
5.2.2 MOD. 01 ................................................. .................................................. .42
5.3 Disassembly of the vacuum pump .............................................. ......................... 43
5.3.1 Siroclave B ............................................. .................................................. .43
5.3.2 Installation of the revision kit - Siroclave B ......................................... ............ 45
5.3.3 Siroclave VS ............................................. ................................................. 47
5.3.4 Installation of the revision kit - Siroclave VS ......................................... ......... 48
5.4 Replacing a temperature sensor .............................................. ...................... 50

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 3rd

Page 5
Sirona Dental Systems GmbH

5.4.1 Upper temperature sensor ............................................. ............................ 50


5.4.2 Lower temperature sensor ............................................. ........................... 52
5.4.3 Steam temperature sensor .............................................. .............................. 54
5.4.4 Pin assignment of temperature board ............................................. ........ 56
5.5 Replacing a solenoid valve .............................................. .............................. 57
5.6 Replacing the main board .............................................. .................................... 59
5.7 Replacing the print board .............................................. .................................... 62
5.8 Replacing the temperature board .............................................. ........................... 64
5.9 Replacing the water flow meter .............................................. ................... 66
5.10 Replacing the condensate pump .............................................. .......................... 67
5.11 Cleaning the cooler .............................................. ..................................... 69
5.12 Removing the display cover .............................................. .................. 70
6 fault trees ................................................ ......................... 72
6.1 Introduction ................................................ .................................................. ....... 72
6.2 BLAC OUT ............................................... .................................................. ...... 73
6.3 MANU STOP ............................................... .................................................. .. 73

https://translate.googleusercontent.com/translate_f 4/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
6.4
6.5 NEED
NEED INST ...............................................
CLEANING ..................................................
............................................... ..... 74
............................................ 74
6.6 NEED SERVICE ............................................... ............................................... 74
6.7 ADD H2O ............................................... .................................................. ........ 75
6.8 FULL H2O ............................................... .................................................. ....... 75
6.9 TEST FAIL ............................................... .................................................. ...... 76
6.10 DRY FAIL ............................................... .................................................. ...... 77
6.11 OPEN DOOR ............................................... .................................................. 78
6.12 NEED TEST CD 1 ............................................. ............................................ 79
6.13 NEED TEST CD 2 ............................................. ............................................ 80
6.14 NEED TEST CD 3 ............................................. ............................................ 81
6.15 NEED TEST CD 4 ............................................. ............................................ 82
6.16 NEED TEST CD 5 ............................................. ............................................ 83
6.17 NEED TEST CD 6 ............................................. ............................................ 83
6.18 NEED TEST CD 7 ............................................. ............................................ 84
6.19 FAIL AL 1 .............................................. .................................................. ....... 85
6.20 FAIL AL 2 .............................................. .................................................. ....... 85
6.21 FAIL AL 3 .............................................. .................................................. ....... 86
6.22 FAIL AL 4 .............................................. .................................................. ....... 86
6.23 FAIL AL 5 .............................................. .................................................. ....... 87
6.24 FAIL AL 6 .............................................. .................................................. ....... 88
6.25 FAIL AL 7 .............................................. .................................................. ....... 89
6.26 FAIL AL 8 .............................................. .................................................. ....... 90
6.27 FAIL AL 9 .............................................. .................................................. ....... 91
6.28 FAIL AL 10 .............................................. .................................................. ..... 92
6.29 FAIL AL 11 .............................................. .................................................. ..... 93
6.30 FAIL AL 12 .............................................. .................................................. ..... 94
6.31 FAIL AL 13 .............................................. .................................................. ..... 95
6.32 FAIL AL 14 .............................................. .................................................. ..... 96
6.33 FAIL AL 15 .............................................. .................................................. ..... 97
6.34 FAIL AL 16 .............................................. .................................................. ..... 98
6.35 FAIL AL 18 .............................................. .................................................. ..... 98
6.36 FAIL AL 31 .............................................. .................................................. ..... 99

4th SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 6
Sirona Dental Systems GmbH

1 Safety instructions

ATTENTION ATTENTION

Repair and maintenance work


Use during repair

https://translate.googleusercontent.com/translate_f 5/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
may only be from SIRONA an ESD bracelet. Connect
or trained by SIRONA
this with the protective conductor.
Service technicians carried out
become.
ATTENTION

ATTENTION When you open the device:


Please note the precautionary
Use for repair and measures when dealing with
Maintenance only original PCBs (EGB). Unload yourself before
SIRONA spare parts. Touch the components by touching them
a grounding point.

ATTENTION
ATTENTION
Disruption of electromedical devices
by radio telephones: Perform after every maintenance or
To ensure the operational Repair a functional
standby electromedical Control by. Turn off the device
Devices, the operation is more mobile at the main switch. Wait that
Radiotelephones in the practice or Self test and run
Clinic area prohibited. then program 1.
Check She, that the
Sterilization process without errors
ATTENTION
will run.

Empty the water tanks if


You tilt the device.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 5

Page 7
Sirona Dental Systems GmbH

2 Technical description
https://translate.googleusercontent.com/translate_f 6/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

2.1 Exterior view - Siroclave B

6 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

https://translate.googleusercontent.com/translate_f 7/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 8
Sirona Dental Systems GmbH

2.2 Exterior view - Siroclave VS

https://translate.googleusercontent.com/translate_f 8/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 7

Page 9
Sirona Dental Systems GmbH

2.3 Interior view - Siroclave B

https://translate.googleusercontent.com/translate_f 9/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

8th SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 10
Sirona Dental Systems GmbH

2.4 Interior view - Siroclave VS

https://translate.googleusercontent.com/translate_f 10/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 9

Page 11
Sirona Dental Systems GmbH

2.5 System components - Siroclave B (Mod.01)

https://translate.googleusercontent.com/translate_f 11/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

10th SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 12
Sirona Dental Systems GmbH

2.6 System components - Siroclave B (Mod.03)

https://translate.googleusercontent.com/translate_f 12/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 11

Page 13
Sirona Dental Systems GmbH

2.7 System components - Siroclave VS

https://translate.googleusercontent.com/translate_f 13/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

12th SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 14
Sirona Dental Systems GmbH

2.8 Wiring diagram - Siroclave B (Mod. 01)

https://translate.googleusercontent.com/translate_f 14/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 13

https://translate.googleusercontent.com/translate_f 15/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 15

Sirona Dental Systems GmbH

2.9 Circuit diagram - Siroclave B (Mod. 03)

https://translate.googleusercontent.com/translate_f 16/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

14 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 16
Sirona Dental Systems GmbH

2.10 Circuit diagram - Siroclave VS

https://translate.googleusercontent.com/translate_f 17/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 15

Page 17
Sirona Dental Systems GmbH

2.11 Technical data - Siroclave B

Inner dimensions of the


Ø 240 mm, depth 384 mm
chamber
Chamber capacity 18 l
4 kg (unpacked instruments)
Cargo capacity
1.5 kg (porous material)
20 minutes at room temperature
Heating times
10 minutes with the preheated chamber
Sterilization holding times 3-90 min. depending on the program
dry season 3-14 min. depending on the program
443 x 590 x 428 mm
Outer dimensions
(W x D x H)
Net weight 55 kg
Input voltage 200-250 VAC
frequency 50 Hz
Maximum power consumption 2160 W.
Average
1000 W
Power consumption
"Standby" consumption 1w

Fuses 2 x 10 AT (model 6.3 x 32 CT) - IEC 127

30 min. after the cycle ends


Auto shutdown
the device
Double water tank 4 l each
Vacuum pump 20 l / min.
0.3 µm at 99.97% specification 21 see 820
Bacteriological filter
FDA autoclavable
Differentiated heating system - SDR
https://translate.googleusercontent.com/translate_f 18/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

B class - prEN13060
3 LCD displays and LED displays
Sensor buttons
10 programs

- 2 test programs
- 6 sterilization programs
- 1 disinfection program
- 1 individual program

Double insulation door with opening protection


Control of water quality

16 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 18
Sirona Dental Systems GmbH

2.12 Technical data - Siroclave VS

Inner dimensions of the


Ø 240 mm, depth 384 mm
chamber
Chamber capacity 18 l
4 kg (unpacked instruments)
Cargo capacity
1.5 kg (porous material)
20 minutes at room temperature
Heating times
10 minutes with the preheated chamber
Sterilization holding times 3-90 min. depending on the program
dry season 3-14 min. depending on the program
443 x 570 x 428 mm
Outer dimensions
(W x D x H)
Net weight 55 kg
Input voltage 200-250 VAC
frequency 50 Hz
Maximum power consumption 1600 W.
Average
600 W.
Power consumption
"Standby" consumption 1w

Fuses 2 x 10 AT (model 6.3 x 32 CT) - IEC 127

https://translate.googleusercontent.com/translate_f 19/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
30 min. after the cycle ends
Auto shutdown
the device
Double water tank 4 l each
Vacuum pump 20 l / min.
0.3 µm at 99.97% specification 21 see 820
Bacteriological filter
FDA autoclavable
Differentiated heating system - SDR
S class - prEN13060
3 LCD displays and LED displays
Sensor buttons
6 programs

- 2 test programs
- 3 sterilization programs
- 1 individual program

Double insulation door with opening protection


Control of water quality

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 17th

Page 19
Sirona Dental Systems GmbH

3 Technical settings

3.1 Height compensation


1. Switch on the device using the main switch.

2. Hold down the ➊ button and press the POWER button .


3. <SET ALT 100 MT> is displayed (100 is the default setting). Give

with the buttons ➊ and Tasten the current sea level.


4. Press the SET button to program the altitude and start the
automatic filling of the waterways.

https://translate.googleusercontent.com/translate_f 20/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

3.1.1 Note on height compensation

In order to ensure the proper functioning of the pressure sensor, in


the autoclave integrates a function that allows the device to different
Air pressure ratios can be adjusted.

When installing the autoclave, you must take into account the height above the
The installation site of the device. This adjustment must
be made every time the autoclave is at a different altitude
is set up.

The factory setting specifies a height of 100 meters. For heights between 0 and
The setting does not have to be changed 200 meters, as height differences of
+100 meters have no effect on the correct operation of the device.
For safe sterilization it is important that the selected value is no more
than 200 meters from the actual value. An incorrect setting means one
extended vacuum cycle and can cause incorrect or premature AL8 or AL5
Trigger error messages.

18th SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 20
Sirona Dental Systems GmbH

3.2 Configuration of the main board

After replacing the main board, it must be set to the SIRONA version.
The correct configuration ensures that the
Float switch.

1. Switch on the device using the main switch.

https://translate.googleusercontent.com/translate_f 21/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
2. Press the POWER button on the control panel and wait for the self-test.

3. Enter the following code on the control panel.

SET Year 66
SET Month 05
SET Day 05
SET Hour 05

After another SET , the display shows <L on> or


<L off> and a number between <0> and <3> in the bottom display.

4. Use the ➊ and ➋ buttons to set the correct values.


<L on> and <0> = SIRONA version

<L off> and <0> = SIRONA version with deactivated float switch
of the waste water tank. The connection of one
Hose from the overflow (rear of the device) into the
Drain is necessary.

5. Re-enter the date and time.

NOTE

When connected to the drain, angle of the overflow inside the


Remove the waste water tank.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 19th

Page 21
Sirona Dental Systems GmbH

3.3 Alignment of the print board

After replacing the print board, it may be necessary to adjust the print board
https://translate.googleusercontent.com/translate_f 22/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

his. An unbalanced print board can lead to incorrect or premature AL 5


and AL 8 cause error messages.

3.3.1 Software adjustment

1. Switch on the device using the main switch.

2. Press the POWER button on the control panel and wait for the self-test.

3. Set the offset pressure deviation with the key combination SET and
TEST firmly with the chamber door open.

4. Press the SET button to return.

5. Enter the following code on the control panel.

SET Year 88
SET Month 05
SET Day 05
SET Hour 05

After another SET , the display shows <bar> and


the display shows <AdJ> in the bottom display .

6.
TheEnter the determined pressure deviation using the ➊ and ➋ keys .
adjustable range is between +/- 0.04 bar with a step accuracy of
0.0025 bar.

7. Use the SET key to return to normal operation.

8. Re-enter the date and time.

NOTE

If a pressure deviation is found, first check the setting of the


Height compensation.

20th SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

https://translate.googleusercontent.com/translate_f 23/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 22
Sirona Dental Systems GmbH

3.3.2 Hardware adjustment

1. Switch on the device using the main switch.

2. Press the POWER button on the control panel and wait for the self-test.

3. Set the offset pressure deviation with the key combination SET and
TEST firmly with the chamber door open.

4. Adjust the offset printing with the "Offset" potentiometer on the printing board
0.00 bar. Turn the potentiometer screw counterclockwise, so
the pressure value increases. Turn the potentiometer screw clockwise
meaning, the pressure value drops.

5. Use the SET key to return to normal operation.

NOTE

If a pressure deviation is found, first check the setting of the


Height compensation.

NOTE

The pressure display reacts sluggishly to changes in the potentiometer screw.

ATTENTION

Never change the setting of the "Gain" potentiometer.

https://translate.googleusercontent.com/translate_f 24/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 21

Page 23
Sirona Dental Systems GmbH

3.4 Adjustment of the temperature board


After replacing the temperature board, the pressure board can be adjusted
to be required. An unbalanced temperature board can lead to incorrect or
cause premature AL 5 and AL 8 error messages. For correct implementation
For a comparison, we recommend that the upper and lower temperature sensors
have the same version status (see section 5.4.4).

3.4.1 Software adjustment

1. Switch on the device using the main switch.

2. Press the POWER button on the control panel and wait for the self-test.

3. Siroclave B: Start program 1. Set after the 3rd vacuum phase


at 0.00 bar pressure the temperature deviation compared to 100 ° C
firmly. Cancel the cycle.

Siroclave VS: Start program 1. Adjust the pressure deviation


2.15 bar at 134 ° C. Cancel the cycle.

NOTE

If the pressure is above 2.15 bar, the one shown is


Temperature value too low. If the pressure is below 2.15
bar, the displayed temperature value is too high.

4. Enter the following code on the control panel.

SET Year 77
SET Month 05
SET Day 05
SET Hour 05

After another SET , the display shows <° C> and in

https://translate.googleusercontent.com/translate_f 25/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
the bottom display shows <AdJ>.

5. Enter the determined temperature deviation using the ➊ and ➋ buttons


on. The adjustable range is between +/- 4 ° C with a step accuracy
of 0.25 ° C.

6. Use the SET key to return to normal operation.

7. Re-enter the date and time.

22 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 24
Sirona Dental Systems GmbH

3.4.2 Hardware adjustment

1. Switch on the device using the main switch.

2. Press the POWER button on the control panel and wait for the self-test.

3. Siroclave B: Start program S5 with 10 min. Holding time and 134 ° C.


After 5 min. Holding time the pressure deviation to 2.05-
2.08 bar at 134 ° C.

Siroclave VS: Start program S5 with 10 min. Holding time and 134 ° C.
After 5 min. Hold time the pressure deviation to 2.14-
2.16 bar at 134 ° C.

NOTE

If the pressure at 134 ° C is above the specified pressure range, the is


displayed temperature value too low. If the pressure is below 134 ° C
specified pressure range, the displayed temperature value is too high.

4. Adjust the temperature display with the “CAM” offset potentiometer


Temperature board. Turn the potentiometer screw against the
Clockwise, the temperature value increases. Turn the potentiometer
screw clockwise, the temperature value drops.

https://translate.googleusercontent.com/translate_f 26/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

NOTE

A quarter turn corresponds to approx. 1 ° C. The temperature display reacts slowly


Changes to the potentiometer screw. The adjustment is complete when the below
Point 3 specified parameters can be achieved stably. Finally, check the
Alignment with program 1.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 23

Page 25
Sirona Dental Systems GmbH

4 Fault diagnosis
4.1 Manual diagnostic mode

4.1.1 Siroclave B

1. Switch on the device at the main switch.

2. Press the POWER button on the control panel and wait for the self-test.

PHASE 1

Action Ads on the display

Keys in sequence The display shows <TEST> , offset


Press SET and TEST temperature and offset pressure of the chamber.

It is registered on the top of the chamber



to letkey pressed Temperature.

It is registered on the bottom of the chamber



to letkey pressed Temperature.

The display <CICL> and the total number of executed


➌ key pressed Program runs appear on the display.
https://translate.googleusercontent.com/translate_f 27/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
to let
The display <ABOR> and the number of failed

to letkey pressed Program runs appear on the display.

Indicates the display <TEST IN> and the status of the


Button TEST Press
Level indicators and the door switch.

Indicates the display <ALARM> and the encrypted number of


Button POWER Press
last 3 alarms.

The key SET press for a return to normal operation

NOTE

While the device is in manual diagnostic mode, it is impossible


switch off the autoclave with the POWER button .

24th SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 26
Sirona Dental Systems GmbH

PHASE 2

Action Ads on the display

SET keys in succession


<TEST OUT> appears on the display .
and press POWER

https://translate.googleusercontent.com/translate_f 28/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS


to letkey pressed
Solenoid valve 1 opens.


to letkey pressed
Solenoid valve 2 opens.

The solenoid valve 3 opens, the solenoid valve 5 closes. The



to letkey pressed The vacuum pump and the condensate pump start up.


to letkey pressed
The solenoid valve 4 opens.

Button POWER pressed


The solenoid valve 5 closes.
to let

TEST key pressed


The fan must run.
to let

The key SET press for a return to normal operation

NOTE

The condensate pump is only available in model 03.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 25th

Page 27
Sirona Dental Systems GmbH

4.1.2 Siroclave VS

1. Switch on the device at the main switch.

2. Press the POWER button on the control panel and wait for the self-test.

https://translate.googleusercontent.com/translate_f 29/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

PHASE 1

Action Ads on the display

Keys in sequence The display shows <TEST> , temperature


Press SET and TEST and chamber pressure

It is registered on the top of the chamber



to letkey pressed Temperature

It is registered on the bottom of the chamber



to letkey pressed Temperature

The display <CICL> and the total number of executed



to letkey pressed Program runs appear on the display

The display <ABOR> and the number of failed



to letkey pressed Program runs appear on the display

Indicates the display <TEST IN> and the status of the


Button TEST Press
Level indicators and the door switch.

Indicates the display <ALARM> and the encrypted number of


Button POWER Press
last 3 alarms

The key SET press for a return to normal operation

NOTE

While the device is in manual diagnostic mode, it is impossible


switch off the autoclave with the POWER button .

26 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 28
Sirona Dental Systems GmbH

https://translate.googleusercontent.com/translate_f 30/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

PHASE 2

Action Ads on the display

SET keys in succession


<TEST OUT> appears on the display
and press POWER


to letkey pressed
Solenoid valve 1 opens


to letkey pressed
Solenoid valve 2 opens

The solenoid valve 3 opens. The vacuum pump starts up



to letkey pressed Business.


to letkey pressed
The solenoid valve 4 opens

TEST key pressed


The fan must run
to let

The key SET press for a return to normal operation

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 27

Page 29
https://translate.googleusercontent.com/translate_f 31/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Sirona Dental Systems GmbH

4.2 Vacuum test

1. Switch on the device at the main switch.

2. Press the TEST button . The test starts automatically and takes some time
Minutes.

Phase t0 : The vacuum pump pulls a vacuum up to


reach from –0.82 bar (Siroclave B) or
-0.8 bar (Siroclave VS).

Phase t1 : Holding time of 5 min.

Phase t2 : Holding time of 10 min.

In the event of an error, a leak is indicated with <TEST FAIL> .

3. Press the POWER button to switch on the autoclave.

NOTE

The test should be taken at the beginning of each working day with the device cold (below 35 ° C)
Check the system from the "Stand By" mode.

NOTE

If the vacuum test has failed despite the tightness of the chamber, it can
there is a defect in the vacuum pump. To test the vacuum pump, a
Vacuum measuring instrument (Ref 58 93 198) required. (see section 4.3)

https://translate.googleusercontent.com/translate_f 32/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

28 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 30
Sirona Dental Systems GmbH

https://translate.googleusercontent.com/translate_f 33/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 29

Page 31
Sirona Dental Systems GmbH

4.3 Checking the


Vacuum pump

4.3.1 Siroclave B

5. Activate the vacuum pump

by pressing the ➌ button

1. Disassemble the housing


(see section 5.2).

6. The vacuum gauge shows


2. Remove the right hose
now apply a vacuum. The
the vacuum pump.
displayed pressure must be lower than
Be -0.88 bar. Check that
3. Close the vacuum measuring
the vacuum pump the negative pressure
instrument (Ref 58 93 198) to the
after the shutdown stops.
Vacuum pump on.

NOTE

https://translate.googleusercontent.com/translate_f 34/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
If the required vacuum is not
reaches or holds the vacuum pump
not the vacuum, so swap
the vacuum pump or run
If possible, a revision of the
Vacuum pump (see section
5.3).

4. Switch to manual
Diagnostic mode - phase 2
(see section 4.1).

30th SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 32
Sirona Dental Systems GmbH

4.3.2 Siroclave VS Diagnostic mode - phase 2


(see section 4.1).

5. Activate the vacuum pump

by pressing the ➌ button

1. Disassemble the housing 6. The vacuum gauge shows


(see section 5.2). now apply a vacuum. The
displayed pressure must be lower than
2. Remove the front one Be -0.82 bar. Check that
Vacuum pump hose. the vacuum pump the negative pressure
https://translate.googleusercontent.com/translate_f 35/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

after the shutdown stops.


3. Close the vacuum measuring
instrument (Ref 58 93 198) to the
Vacuum pump on. NOTE

If the required vacuum is not


reaches or holds the vacuum pump
not the vacuum, so swap
the vacuum pump or run
If possible, a revision of the
Vacuum pump (see section
5.3).

4. Switch to manual

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 31

Page 33
Sirona Dental Systems GmbH

4.4 Checking a
Solenoid valve

4.4.1 SIRAI solenoid valve

5. Measure with a multimeter


the electrical resistance to the
Terminals. The opponent
level value must be approx. 17 Ω .

6. Build the magnet to be tested


valve (see section 5.5).

https://translate.googleusercontent.com/translate_f 36/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
1. Switch on the device
Mainswitch on.

2. Press the POWER button on


Control panel and service the
Self test from. Will not
Drive displayed error message 7. Remove the clamping ring and

continue with step 3. the washers of the


Solenoid valve.

NOTE 8. Pull the solenoid


upwards.
Becomes one after the self test
Error message FAIL AL 1 to 4
is displayed, so is the corresponding
Solenoid valve electrically defective. To deceive
she then the appropriate
Solenoid valve off
(see section 5.5).

3. Disassemble the housing


(see section 5.2)

4. Pull out the electrical 9. Loosen the housing screws


Connections from the solenoid valve. exercises of the solenoid valve.

32 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 34
Sirona Dental Systems GmbH

10. Open the housing and


remove the pestle and that
Valve.

https://translate.googleusercontent.com/translate_f 37/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

11. Check the valve and the


Valve seat for damage and
Pollution.

12. Check the freedom of movement


of the plunger in its guide.

13. Check the permeability


of the inlet and outlet channel.

14. Install the solenoid valve in


reverse order again
together.

15. Reassemble the solenoid valve


on.

16. The solenoid valve now works


still not perfect, so
exchange it.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 33

Page 35
Sirona Dental Systems GmbH

4.4.2 Solenoid valve type


HONEYWELL

4. Pull out the electrical


https://translate.googleusercontent.com/translate_f 38/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Connections from the solenoid valve.

5. Measure with a multimeter


the electrical resistance to the
Terminals. The opponent
level value must be approx. 17 Ω .

6. Build the magnet to be tested


valve (see section 5.5).

1. Switch on the device


Mainswitch on.

2. Press the POWER button on


Control panel and service the
Self test from. Will not
Drive displayed error message
continue with step 3.

NOTE
7. Loosen the nut of the coil
Becomes one after the self test frame.
Error message FAIL AL 1 to 4
is displayed, so is the corresponding
Solenoid valve electrically defective. To deceive
NOTE
she then the appropriate
To loosen the mother of the
Solenoid valve off
Coil frame becomes an 11/16 "mouth
(see section 5.5).
or ring spanner needed.

3. Disassemble the housing


8. Pull the solenoid together
(see section 5.2)
the coil frame upwards.

34 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 36

https://translate.googleusercontent.com/translate_f 39/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
Sirona Dental Systems GmbH

9. Loosen the housing screws


Practice the valve block.

NOTE

11. Check the valve and the


Valve seat for damage and
Pollution.

To the To solve the Casing-


the valve block is screwed 12. Check the freedom of movement

that, the "Bellows Rep Kit" (Ref 59 06 of the plunger in its guide.

891) enclosed, tool and a


7 / 8´´ combination wrench or ring spanner 13. Check the permeability

needed .. of the inlet and outlet channel.

14. Install the solenoid valve in


10. Open the housing and
reverse order again
remove the plunger and the
together.
Feather.

15. Reassemble the solenoid valve


on.

16. The solenoid valve now works


still not perfect, so
exchange it.

https://translate.googleusercontent.com/translate_f 40/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 35

Page 37
Sirona Dental Systems GmbH

4.5 Checking the


Water flow meter

1. Switch to manual
4. Loosen the housing screws
Diagnostic mode - phase 2
exercises of the water meter.
(see section 4.1).

2. Simultaneously press the buttons


➌ and ➍ for fresh water in the
Pump chamber. If the
Vacuum LED on the control panel
the water volume is flashing moderately
counter okay. Should the
Vacuum LEDs flash irregularly
or don't light up at all, like this
go to step 3 . 5. Remove the housing cover
and the rotor.

NOTE 6. Check the cleanliness of the


Rotor axis and the rotor.
The water flow meter measures that
Amount of water inductive. As soon as
The rotor turns and water flows
the sensor electronics counts that
Revolutions. The turns are

https://translate.googleusercontent.com/translate_f 41/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
a measure of the flow rate.

3. Disassemble the water


quantity counter
(see section 5.9).

36 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 38
Sirona Dental Systems GmbH

7. Check the permeability


of the inlet and outlet channel.

8. Put the rotor back on


its axis and check the
Free movement of the rotor.

9. Mount the housing again


of the water flow meter.

10. Build the amount of water


counter again.
(see section 5.9)

11. Repeat steps 1 and


2 . Should the vacuum LED always
not flashing regularly yet, so
swap the amount of water
counter off.

https://translate.googleusercontent.com/translate_f 42/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 37

Page 39
Sirona Dental Systems GmbH

5 assembly work
5.1 Annual maintenance

https://translate.googleusercontent.com/translate_f 43/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Installation instructions service kit


(Ref 59 70 681)

1. Switch on the device Final work


Main switch off.
7. Remove the cooler from the
2. Exchange the bacteriological Back of the autoclave. Clean
Filter. The fans and the cooler
thoroughly. (see section 5.10)
3. Replace the door seal.
8. Test the fans.
4. Empty both water tanks. (see section 4.1)

5. Replace the fresh water 9. Perform a vacuum test.


filter. (see section 4.2)

6. Exchange the two filters in the


Pressure chamber.

38 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 40
Sirona Dental Systems GmbH

NOTE

Unscrew the closing cover of the


Fresh water filter holder with one
Coin from.

NOTE

After assembling the new one


Door seal is for closing the door
more effort is required.

NOTE

For unscrewing the filter holder

https://translate.googleusercontent.com/translate_f 44/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
you need a "filter key"
(REF 58 36 361).

ATTENTION

Screw the filter holder just like that


firmly believe that the filter screens are not
be squeezed.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 39

Page 41
Sirona Dental Systems GmbH

5.2 Disassembly of the


Housing

5.2.1 MOD. 03

https://translate.googleusercontent.com/translate_f 45/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

1. Loosen the side housing


screw connections.
3. Pull the protective conductor from
Upper part of the housing.

4. Remove the upper case


part.

5. Press the side parts of the


Housing down and
2. Lift the top of the
then backwards.
Housing slightly at the back and
pull the top back.

40 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 42
Sirona Dental Systems GmbH

6. Pull the protective conductor from the


Side parts of the housing.

7. Remove the side parts.

https://translate.googleusercontent.com/translate_f 46/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

NOTE

Installation takes place in reverse


Sequence.

ATTENTION

Close all protective conductors at the


Reassembly.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 41

Page 43
Sirona Dental Systems GmbH
https://translate.googleusercontent.com/translate_f 47/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

5.2.2 MOD. 01

3. Pull the protective conductor from


Housing.

4. Spread the side parts of the


Housing easily and take
the housing completely.

NOTE

Installation takes place in reverse


Sequence.
1. Loosen the housing
screws on the bottom
of the device.
ATTENTION

Close all protective conductors at the


Reassembly.

2. Lift the case back


lightly and pull that
Housing a little to the rear.

https://translate.googleusercontent.com/translate_f 48/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

42 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 44
Sirona Dental Systems GmbH

5.3 Disassembly of the


Vacuum pump

5.3.1 Siroclave B

4. Loosen the two nuts


1. Disassemble the housing
which the vacuum pump with the
(see section 5.2.1).
Connect the carrier plate.

2. Loosen the screw connections of the


5. Remove the front right
lower housing strut and
the vacuum pump connected
remove them.
his hose.

NOTE

The lower housing strut only has to


be removed from Mod. 01. At the
Mod. 03 the strut is shortened and
can remain installed.
https://translate.googleusercontent.com/translate_f 49/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

3. Disassemble the cooler without


the hose connections too
disconnect (see section 5.11).

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 43

Page 45
Sirona Dental Systems GmbH

6. Remove the rear hose


on the silencer. NOTE

Installation takes place in reverse


7. Disconnect the electrical
Sequence.
Vacuum pump connections.

8. Carefully pull out the pump ATTENTION


the housing. Turn here
the pump counterclockwise In exchange only one is allowed
sense. Vacuum pump with a speed of
2200 rpm can be installed if there is
a Siroclave B (Mod.03) with
a condensate pump (see section
5.10).
All other Siroclave B models
may only be used with a vacuum pump
at a speed of 2700 rpm
become.

https://translate.googleusercontent.com/translate_f 50/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

44 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 46
Sirona Dental Systems GmbH

5.3.2 Installation of the revision kit - Siroclave B

ATTENTION

The revision kit is only suitable for devices with a serial no. from 1973 or for
Devices with a nickel-coated lid (A). This is the case if the cover (A)
is darker than the part (H).

NOTE

You need the following revision kits (see type plate):

Siroclave B (Mod.01): Engine speed: 2700 rpm ➡ 2 x 59 31 592


Siroclave B (Mod.03): Engine speed: 2700 rpm ➡ 2 x 59 31 592
Siroclave B (Mod.03): Engine speed: 2200 rpm ➡ 2 x 60 31 236

After two revisions, the complete pump should be replaced as the


Engine needs to be overhauled.

https://translate.googleusercontent.com/translate_f 51/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

1. Disassemble the vacuum pump 2. Mark the position of the parts


(see section 5.3). (A) and (B) before disassembly. This
is required to get the correct one
To ensure assembly.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 45

Page 47
Sirona Dental Systems GmbH

3. Remove the hose (S), the


connects both pistons together.
ATTENTION

4. Remove the screws (C)


and clean the parts (A) and
(B).

5. Take the circlip (D)


from.

The flat side of the valves (V) must


6. Loosen the screws (E).
on the conical side of the lid
(A).
7. Pull the old piston (F) with you
a puller for ball bearings from
13. Replace the parts (A) and (B)
the motor shaft (G).
considering the attached
Mark and drag the
8. Place the new piston (F)
Slightly tighten screws (C).
in the correct position from the same side
https://translate.googleusercontent.com/translate_f 52/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

on.
14. Turn the motor shaft (G)
a few turns by hand to
9. Tighten the screws (E).
to be sure that the membrane
sits correctly. Pull the
10. Make sure again
Tighten screws (C).
that all parts are installed correctly
are. The ball bearing must be on the outside
15. Close the hose (S),
sit!
the two pistons together
reconnects.
ATTENTION
16. Reinstall the pump
The ball bearing may only be pressed
(see section 5.3).
be pressed onto the inner ring

17. Measure the maximum


11. Put the locking ring (D) Vacuum that the pump
on. can achieve with the vacuum
Measuring instrument (Ref 58 93 198)
12. Install the new O-rings (O)
(see section 4.3).
and the valves (V).

NOTE

A value must be lower than - 0.88 bar


can be achieved.

46 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 48
Sirona Dental Systems GmbH

5.3.3 Siroclave VS

https://translate.googleusercontent.com/translate_f 53/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

4. Remove the two from the


Vacuum pump connected
Hoses.

5. Disconnect the electrical


1. Disassemble the housing.
Vacuum pump connections.
(see section 5.2.1)

6. Carefully pull out the pump


2. Disassemble the fan
the housing.
(see section 5.12).

7. Unscrew the silencer


3. Loosen the two nuts
on the pump head.
which the vacuum pump with the
Connect the carrier plate.

NOTE

Installation takes place in reverse


Sequence.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 47

Page 49
Sirona Dental Systems GmbH

5.3.4 Installation of the revision kit - Siroclave VS

ATTENTION

https://translate.googleusercontent.com/translate_f 54/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
The revision kit is only suitable for devices with a serial no. from 1070 or for
Devices with a nickel-coated lid (A). This is the case if the cover (A)
is darker than the part (H).

NOTE

After two revisions, the complete pump should be replaced as the


Engine needs to be overhauled.

1. Disassemble the vacuum pump 3. Remove the screws (C)


(see section 5.3). and clean the parts (A) and
(B).
2. Mark the position of the parts
(A) and (B) before disassembly. This 4. Take the locking ring (D)
is required to get the correct one from.
To ensure assembly.
5. Loosen the screws (E).

48 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 50
Sirona Dental Systems GmbH

https://translate.googleusercontent.com/translate_f 55/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

6. Pull the old piston (F) with you 12. Put the parts (A) and (B)
a puller for ball bearings from considering the attached
the motor shaft (G). Mark and drag the
Slightly tighten screws (C).
7. Install the new piston (F)
in the correct position from the same side 13. Turn the motor shaft (G)
on. a few turns by hand to
to be sure that the menbrane
8. Tighten the screws (E). sits correctly. Pull the
Tighten screws (C).
9. Make sure again
that all parts are installed correctly 14. Reinstall the pump
are. The ball bearing must be on the outside (see section 5.3)
sit!
15. Measure the maximum
Vacuum that the pump
ATTENTION can achieve with the vacuum
Measuring instrument (Ref 58 93 198).
The ball bearing may only be pressed
(see section 4.3)
be pressed onto the inner ring

10. Put the retaining ring (D)


NOTE
on.
A value must be lower than - 0.82 bar
11. Install the new O-rings (O) can be achieved.
and the valves (V).

ATTENTION

The flat side of the valves (V) must


on the conical side of the lid
(A).

https://translate.googleusercontent.com/translate_f 56/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 49

Page 51
Sirona Dental Systems GmbH

5.4 Exchange one


Temperature sensor

5.4.1 Upper temperature sensor

5. Disconnect the sensor cable from the


Temperature board.
1. Disassemble the housing.
(see section 5.2)
NOTE
2. Remove the thermal insulation
of the temperature sensor. To the better To reach the
You can use the connecting terminals
3. Loosen the screw connection of the Remove the temperature board.
Sensors. (see section 5.8)

6. Carefully pull the sensor


out of the case.

https://translate.googleusercontent.com/translate_f 57/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

4. Remove all cable ties


along the sensor line.

50 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 52
Sirona Dental Systems GmbH

ATTENTION

If the tightening torque is too high


the thread becomes irreparable
damaged.

9. Route the sensor cable in


same location as the old and
connect the sensor to the
Temperature board.

7. Prepare the new sensor like this


before that the metallic part the NOTE
has the same curvature as the
old. The temperature sensors must be with
the correct polarity
become. (see section 5.4.4)
ATTENTION

To break the sensor 10. Check the correct and

to prevent, bend the sensor tension-free laying of the


Sensor cable. Bring too
in a generous radius and not
new cable ties on the cables
at its ends.
instructions.

8. Screw the new sensor in


11. Mount the housing and
same location as the old firm
do a function test
(see image). Bring the heat
by.
insulation again.

NOTE

https://translate.googleusercontent.com/translate_f 58/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

It is recommended temperature
sensors of the same type for the
upper and lower chamber sensor too
use. (see section 5.4.4)

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 51

Page 53
Sirona Dental Systems GmbH

5.4.2 Lower temperature


sensor

5. Loosen the screw connection of the


Sensors.
1. Empty all water tanks.
6. Remove all cable ties
2. Disassemble the housing. along the sensor line.
(see section 5.2)

3. Place the device on the left


Side and remove the
Plastic plug on the bottom of the
Device.

https://translate.googleusercontent.com/translate_f 59/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

7. Disconnect the sensor cable from the


Temperature board.

NOTE

To the better To reach the


4. Remove the heat You can use the connecting terminals
insulation of the temperature sensor. Remove the temperature board
(see section 5.8).

8. Carefully pull the sensor


out of the case.

52 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 54
Sirona Dental Systems GmbH

NOTE

The temperature sensors must be with


the correct polarity
(see section 5.4.4)

12. Check the correct and


tension-free laying of the
Sensor cable. Bring too
new cable ties to the
Lines on.
9. Prepare the new sensor like this
before that the metallic part the
13. Install the plastic stopper
has the same curvature as the
old. the bottom of the device again
on.

ATTENTION 14. Assemble the housing,

https://translate.googleusercontent.com/translate_f 60/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
fill the fresh water tank
To break the sensor and run a function test
to prevent, bend the sensor by.
in a generous radius and not
at its ends.
NOTE
10. Screw the new sensor
It is recommended temperature
in the same position as the old one
sensors of the same type for the
firmly. Bring the heat
upper and lower chamber sensor too
insulation again.
use (see section 5.4.4)

ATTENTION

If the tightening torque is too high


the thread becomes irreparable
damaged.

11. Route the sensor cable in


same location as the old and
connect the sensor to the
Temperature board.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 53

Page 55
Sirona Dental Systems GmbH

5.4.3 Steam temperature sensor

https://translate.googleusercontent.com/translate_f 61/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

1. Disassemble the housing.


4. Remove the plastic plug
(see section 5.2)
the back of the device.

2. Remove all cable ties


along the sensor line.

3. Disconnect the sensor cable from the


Temperature board.

NOTE

To the better To reach the


You can use the connecting terminals
Remove the temperature board 5. Bend the sensor straight,
(see section 5.8). pull the sensor cable through
the round hole from the case.
Unscrew the sensor from the
Pressure chamber out.

6. Clean the thread of the


Pressure chamber of residual gasket.

7. Apply the sealant


"RTV 159" (Ref 18 74 192)
the thread of the new sensor.

54 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 56
Sirona Dental Systems GmbH

8. Screw the new sensor in


the pressure chamber.

9. Bend the sensor so that it


has the same curvature as
the old.
https://translate.googleusercontent.com/translate_f 62/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

ATTENTION

To break the sensor


to prevent, bend the sensor
in a generous radius and not
at its ends.

10. Route the sensor cable in


same location as the old and
connect the sensor to the
Temperature board.

NOTE

The temperature sensors must be with


the correct polarity
become. (see section 5.4.4)

11. Check the correct and


tension-free laying of the
Sensor cable. Bring too
new cable ties to the
Lines on.

12. Mount the housing and


do a function test
by.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 55

Page 57
Sirona Dental Systems GmbH

https://translate.googleusercontent.com/translate_f 63/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

5.4.4 Pin assignment of temperature board

https://translate.googleusercontent.com/translate_f 64/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

56 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 58
Sirona Dental Systems GmbH

5.5 Exchange one


Solenoid valve

1. Disassemble the housing.


(see section 5.2)

2. Find the defective solenoid valve.


(see section 2.3 or 2.4)

3. Pull out the electrical


Connections from the solenoid valve.

4. Loosen the screw connections of the


Solenoid valve.

NOTE

https://translate.googleusercontent.com/translate_f 65/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

With Mod. 01, when exchanging


of solenoid valve 2, 3 or 4 previously
Carrier plate can be solved.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 57

Page 59
Sirona Dental Systems GmbH

5. Screw the hose


connections.
ATTENTION

Make sure that the hoses


6. Remove the solenoid valve.
not crushed or kinked
become.
NOTE

ATTENTION
Installation takes place in reverse
Sequence.
The solenoid valves are depending on the type
opened or closed when de-energized.

NOTE Pay attention when replacing the


The right type.

If you have an old type solenoid valve Use the

replace with a new type, so Spare part list.

Right the Hole spacing the


Screw connections of the solenoid valve
ATTENTION
no longer with the hole spacing of the
Housing or the carrier plate
Empty the water tanks if
match. In this case, it is
You tilt the device.
Solenoid valve with just one screw
fasten.

https://translate.googleusercontent.com/translate_f 66/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

58 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 60
Sirona Dental Systems GmbH

5.6 Exchange of
Motherboard

4. Loosen the screw connections of the


upper housing strut and
remove them.

5. Disassemble the display


cover
(see section 5.12).

6. Loosen the two screws on the


ATTENTION
Motherboard carrier plate.

Use during repair


an ESD bracelet. Connect
this with the protective conductor. You're welcome
observe the precautionary

https://translate.googleusercontent.com/translate_f 67/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
took when handling circuit boards
(ETUC). Discharge yourself before touching
the components by touching one
Grounding point

1. Disassemble the housing.


(see section 5.2)

2. Disassemble the print board


without removing the hose.
(see section 5.7)

3. Disassemble the temperature


circuit board without the sensor cables
remove. (see section 5.8)

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 59

Page 61
Sirona Dental Systems GmbH

7. Loosen the screw connection of the


Carrier plate on the lower strut.

8. Remove all electrical


Lines from the motherboard.

ATTENTION
10. Press with flat-nose pliers

https://translate.googleusercontent.com/translate_f 68/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
the retention pins of the individual
Hold pins together and pull
The board carefully a bit
over the retaining pegs.
Pay attention when pulling the
Lines to the position of the two
marked connector. A mix up
the plug leads to AL3 and AL1
Error messages.

9. Now pull the carrier plate


together with the motherboard carefully
up and lay it flat
on the water tanks.

60 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 62
Sirona Dental Systems GmbH

NOTE

If the board does not overlap


Let the holding pins pull, snap
the retention pins with pliers
from. Renew damaged ones
Retaining pins before reinstalling the
Circuit board.

11. Once you get the board from everyone


Have loosened retaining pins
You remove the circuit board.
https://translate.googleusercontent.com/translate_f 69/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

NOTE

Installation takes place in reverse


Sequence.

ATTENTION

When changing the motherboard


need the data from the old one
Board to be transferred to the new one.
You need the technician
software.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 61

Page 63
Sirona Dental Systems GmbH

5.7 Exchange of
Printing board

https://translate.googleusercontent.com/translate_f 70/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

ATTENTION NOTE

If the board does not overlap


Use during repair
Let the holding pins pull, snap
an ESD bracelet. Connect
the retention pins with pliers
this with the protective conductor. You're welcome
from. Renew damaged ones
observe the precautionary
took when handling circuit boards Retaining pins before reinstalling the

(ETUC). Discharge yourself before touching Circuit board.

the components by touching one


Grounding point. 3. Once you get the board from both
Have loosened retaining pins
The board carefully out of the
1. Disassemble the housing.
(see section 5.2) Plug socket on the main board
pull.

2. Press with flat-nose pliers


the retention pins of the individual
Hold pins together and pull
The board carefully a bit
over the retaining pegs.

62 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 64
Sirona Dental Systems GmbH

https://translate.googleusercontent.com/translate_f 71/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

4. Remove the cable tie on the


Pressure sensor and pull the
Hose off.

NOTE

Installation takes place in reverse


Sequence.

NOTE

Tip: Moisten the connector


prune before postponing the
Hose. Then the hose slides
better. No lubricant use

ATTENTION

Always push the hose


completely up to the stop on the
Sensor on. Make sure that
the hose is not squeezed or
is kinked.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 63


https://translate.googleusercontent.com/translate_f 72/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 65
Sirona Dental Systems GmbH

5.8 Exchange of
Temperature board

NOTE

If the board does not overlap


Let the holding pins pull, snap
the retention pins with pliers
from. Renew damaged ones
ATTENTION
Retaining pins before reinstalling the
Circuit board.
Use during repair
an ESD bracelet. Connect
this with the protective conductor. You're welcome 3. Once you get the board from both

observe the precautionary Have loosened retaining pins


took when handling circuit boards The board carefully out of the
(ETUC). Discharge yourself before touching Plug socket on the main board

the components by touching one pull.

Grounding point.
4. Loosen the compression fittings

1. Disassemble the housing. exercises of the sensor lines and

(see section 5.2) pull the lines out of the


Base out.

2. Press with flat-nose pliers


the retention pins of the individual
Hold pins together and pull
The board carefully a bit
over the retaining pegs.

https://translate.googleusercontent.com/translate_f 73/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

64 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 66
Sirona Dental Systems GmbH

NOTE

Note the assignment of the


Sensor cables for reinstallation.
(see also section 5.4.4)

NOTE

Installation takes place in reverse


Sequence.

https://translate.googleusercontent.com/translate_f 74/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 65

Page 67
Sirona Dental Systems GmbH

5.9 Exchange of
Water flow meter

4. Pull the water quantity


counter a bit, pull
the electrical lines from
Water flow meter from and
screw the hose
connections on the meter.

1. Empty the fresh water


tank.

1. Disassemble the housing.


(see section 5.2)

3. Loosen the screw connection of the


Water meter on the front
NOTE
of the device. You have to do this
first the closing cover of the
Note the assignment of the
Filter holder with a coin
electrical connections for the
screw off.
https://translate.googleusercontent.com/translate_f 75/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
Replacement.

NOTE

Installation takes place in reverse


Sequence.

66 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 68
Sirona Dental Systems GmbH

5.10 Exchange of
Condensate pump

1. Disassemble the housing.


(see section 5.2)
4. Keep the rubber buffer with you
2. Loosen the screw connection of the a pair of pliers and loosen the
Condensate pump on the back Mother. Take the pump out
of the device. the frame.

https://translate.googleusercontent.com/translate_f 76/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

3. Pull out the electrical


Connections of the pump.
5. Remove the cable ties
both hoses. Now pull
the two hoses carefully from
the condensate pump.

NOTE

Note the flow direction of the


Condensate pump for
installation.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 67

Page 69
Sirona Dental Systems GmbH

NOTE

Installation takes place in reverse


Sequence.

NOTE

Tip: Moisten the connector


prune before postponing the
Hose. Then the hose slides
better. No lubricant use

https://translate.googleusercontent.com/translate_f 77/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

ATTENTION

Always push the hose


completely up to the stop on the
Sensor on. Make sure that
the hose is not squeezed or
is kinked.

68 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 70
Sirona Dental Systems GmbH

5.11 Cleaning the


Cooler

https://translate.googleusercontent.com/translate_f 78/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

3. Now pull the cooler


1. Loosen the four screws on the a little bit carefully. Clean
Cover grid. The cooling fins and the fans
thoroughly.

4. Test the fans.


2. Loosen the two screws of the (see section 4.1.1)
Cooler.

NOTE

Installation takes place in reverse


Sequence.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 69

Page 71
Sirona Dental Systems GmbH

https://translate.googleusercontent.com/translate_f 79/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

5.12 Dismantling the


Display cover
4. Pull the display cover
upwards from the housing.

5. Put the display cover on


a flat surface and loosen
the screw connections of the display
circuit board.

1. Disassemble the housing.


(see section 5.2)

2. Loosen the screw connection


at the bottom of the housing front.

6. Remove the display board.

NOTE

Keep the spacers of the


Display board safely until
Refit on.

7. Press with a slotted


screwdriver on the
Snap connection of the PC
3. Disconnect the ribbon cable from the
Connection socket and pull the
Displays from the main board.
Socket from the display cover
from.

70 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

https://translate.googleusercontent.com/translate_f 80/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 72
Sirona Dental Systems GmbH

NOTE

Installation takes place in reverse


Sequence.

ATTENTION

Use during repair


an ESD bracelet. Connect
this with the protective conductor. You're welcome
observe the precautionary
took when handling circuit boards
(ETUC). Discharge yourself before touching
the components by touching one
Grounding point.

https://translate.googleusercontent.com/translate_f 81/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 71

Page 73
Sirona Dental Systems GmbH

6 fault trees
6.1 Introduction
The electronic control program of the autoclave checks the various
Stages of the sterilization process continuously and controls the correct operation
every component. In the event of a malfunction, an error message appears on the display
spent. You can use this error message in this chapter
systematically determine the cause of the error. Go the appropriate one
Error tree step by step. Actions described in normal script
can be carried out with the device closed. In italics on yellow
The activities described above can only be carried out when the device is open
become.

Example:

" Error message " importance

Can at Can when closed


1 closed device Yes Device performed
being checked. become.

No

You must first You have to do that beforehand


https://translate.googleusercontent.com/translate_f 82/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
2nd open the device No Open the device (see
(see section 5.2) Section 5.2)

Yes

You have to do that beforehand


Open the device (see
Section 5.2)

72 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 74
Sirona Dental Systems GmbH

6.2 BLAC OUT

Importance:
"BLAC OUT" Temporary
Power cut

Check that the power cord is secure.


Check glass butt fuses. Moisture in
let the chamber dry, close the door
and restart the program.

6.3 MANU STOP

Importance:
Program was done by hand
"MANU STOP"
stopped. Sterilization is not
successful.

Moisture in the
Dry the chamber
leave, close the door and
Restart the program.
https://translate.googleusercontent.com/translate_f 83/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 73

Page 75
Sirona Dental Systems GmbH

6.4 NEED INST

Importance:
"NEED INST" The installation must be done
become.

To install. See instructions for use.

6.5 NEED CLEANING

Importance:
"NEED CLEANING" There were 60 cycles without
Go through intermediate cleaning.

https://translate.googleusercontent.com/translate_f 84/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Carry out cleaning cycle. Please refer


Instructions for use.

6.6 NEED SERVICE


Importance:
After a year or 1500
"NEED SERVICE"
Cycles is maintenance
perform.

Perform maintenance. See section


5.1.

74 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 76
Sirona Dental Systems GmbH

6.7 ADD H2O

Importance:
"ADD H2O"
The fresh water tank is empty.

Is the Top up with fresh water


1 Fresh water tank Yes and program new
almost empty? start.

https://translate.googleusercontent.com/translate_f 85/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
No

Float switch
and sensor cables
check.

6.8 FULL H2O

Importance:
"FULL H2O"
The process water tank is full.

Is the Float switch


1 Domestic water tank Yes and sensor cables
been emptied? check.

No

Domestic water tank


empty and
Restart the program.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 75

Page 77
Sirona Dental Systems GmbH

6.9 TEST FAIL

Importance:
"TEST FAIL" The vacuum test is
failed.

https://translate.googleusercontent.com/translate_f 86/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Is the door seal


Clean the door seal or
1 clean and No
replace.
undamaged?

Yes

Is the chamber filter Clean the chamber filter or


2nd No
clean? replace.

Yes

Is the
Vacuum pump
Functionality
overtake or
3rd the vacuum pump No
replace.
given?
(see section 5.3)
(see section 4.3)

Yes

Is a leak in the Check the


4th Hose feed lines No Solenoid valves 1 to 5.
the chamber? (see section 4.4)

Yes

Hose feed
replace.

76 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 78
Sirona Dental Systems GmbH

6.10 DRY FAIL


https://translate.googleusercontent.com/translate_f 87/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Importance:
"DRY FAIL" The drying was not
successfully completed.

Was the material before


end of
1 Yes Press the STOP key.
Drying program
removed?

No

Check the
Door lock and
the door switch.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 77

https://translate.googleusercontent.com/translate_f 88/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 79
Sirona Dental Systems GmbH

6.11 OPEN DOOR

Importance:
Door was not getting right
"OPEN DOOR"
locked or during that
Sterilization process opened.

Became the program


Close door and
1 with the door open Yes
Restart the program.
started?

No

The door was open during Moisture in the


of Dry the chamber
2nd Yes
Sterilization process leave, close the door and
open? Restart the program.

No

Check the
Door lock and
the door switch.

https://translate.googleusercontent.com/translate_f 89/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

78 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 80
Sirona Dental Systems GmbH

6.12 NEED TEST CD 1

Importance:
The water drain has
"Need Test CD 1"
longer than 4 minutes
lasted.

Was the
Clean the chamber filter or
1 Chamber filter cleaned No
replace.
or replaced?

Yes

Are the
Hose feed lines Hose feed
2nd No
from solenoid valve 2 clean or replace.
continuously?

Yes

Check that out


Solenoid valves 2.
(see section 4.4)

https://translate.googleusercontent.com/translate_f 90/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 79

Page 81
Sirona Dental Systems GmbH

6.13 NEED TEST CD 2


Importance:
The heating-up phase is longer than 40
"Need Test CD 2"
Minutes. Top
Heating element

Program run with


Became the chamber
1 Yes lower load
overloaded?
carry out.

No

Is that
Check the mains supply
2nd Mains voltage at No
to let.
230V +/- 10%?

Yes

Are the lines


to the heating element
and the
3rd No Replace cables
Security thermo-
did electric

https://translate.googleusercontent.com/translate_f 91/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
continuously?

Yes

Is the heating element


4th electric No Replace heating element
continuously?

Yes

Safety thermostat
replace.

80 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 82
Sirona Dental Systems GmbH

6.14 NEED TEST CD 3


Importance:
The heating-up phase is longer than 40
"Need Test CD 3"
Minutes. Lower
Heating element

Program run with


Became the chamber
1 Yes lower load
overloaded
carry out.

No

Is that
Check the mains supply
2nd Mains voltage at No
to let.
230V +/- 10%?

https://translate.googleusercontent.com/translate_f 92/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Yes

Are the lines


to the heating element
and the
3rd No Replace cables
Security thermo-
did electric
continuously?

Yes

Is the heating element


4th electric No Replace heating element
continuously?

Yes

Safety thermostat
replace.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 81

Page 83
Sirona Dental Systems GmbH

6.15 NEED TEST CD 4

Importance:
"Need Test CD 4" Has the water filling phase
lasted longer than 50 seconds.

Was the
Replace fresh water filter
1 Fresh water filter No
or clean.
cleaned or replaced?

Yes

https://translate.googleusercontent.com/translate_f 93/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Is the
Amount of water
Water flow meter
2nd counter No
replace.
working?
(see section 4.5)

Yes

Are the
Hose feed lines Hose feed
3rd No
from solenoid valve 4 clean or replace.
continuously?

Yes

Solenoid valve 4
check.
(see section 4.4)

82 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 84
Sirona Dental Systems GmbH

6.16 NEED TEST CD 5

Importance:
Solenoid valve 4 does not close
"Need Test CD 5"
Completely. The reason is usually
dirt

https://translate.googleusercontent.com/translate_f 94/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

If the error occurs for the 3rd


Not an acute one
1 Times in a row or No
Need for action
always on?

Yes

Solenoid valve 4
check.
(see section 4.4)

6.17 NEED TEST CD 6

Importance:
"Need Test CD 6"
The bacterial filter is clogged.

Is the bacterial filter


1 Yes Replace bacterial filter.
dirty?

No

Solenoid valve 1
check.
(see section 4.4)

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 83

Page 85
Sirona Dental Systems GmbH

6.18 NEED TEST CD 7

https://translate.googleusercontent.com/translate_f 95/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
Importance:
The required vacuum level
"Need Test CD 7"
was within 8 minutes
not reached.

Was the height and


Correct skew
Inclination of the
1 No and adjust height.
Device correctly
(see section 3.1.)
set?

Yes

Was the annual Perform maintenance.


2nd No
Performed maintenance? (see section 5.1)

Yes

Is the
Vacuum pump
Functionality
overtake or
3rd the vacuum pump No
replace.
given?
(see section 5.3)
(see section 4.3)

Yes

Is the
Functionality
Repair solenoid valve
4th the solenoid valves 1 No
or replace.
given to 5?
(see section 4.4)

Yes

Check the
Is a leak in the Condensate pump and
5 Hose feed lines No the neighboring
the chamber ? Check valve.
(only mod. 03)

Yes

Hose feed
replace.

84 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

https://translate.googleusercontent.com/translate_f 96/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 86

Sirona Dental Systems GmbH

6.19 FAIL AL 1

Importance:
"FAIL AL 1"
Solenoid valve 1 is defective.

Are the electrical


Lines of
1 No Replace cables.
Solenoid valve 1
continuously?

Yes

Solenoid valve 1
check.
(see section 4.4)

6.20 FAIL AL 2

Importance:
"FAIL AL 2"
Solenoid valve 2 is defective.

Are the electrical


Lines of
1 No Replace cables.
Solenoid valve 2
continuously?

Yes

Solenoid valve 2
check.
(see section 4.4)
https://translate.googleusercontent.com/translate_f 97/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 85

Page 87
Sirona Dental Systems GmbH

6.21 FAIL AL 3

Importance:
"FAIL AL 3"
Solenoid valve 3 is defective.

Are the electrical


Lines of
1 No Replace cables.
Solenoid valve 3
continuously?

Yes

Solenoid valve 3
check.
(see section 4.4)

6.22 FAIL AL 4

Importance:
"FAIL AL 4"
Solenoid valve 4 is defective.

Are the electrical


Lines of
1 No Replace cables.
Solenoid valve 4
continuously?

https://translate.googleusercontent.com/translate_f 98/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Yes

Solenoid valve 4
check.
(see section 4.4)

86 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 88
Sirona Dental Systems GmbH

6.23 FAIL AL 5
Importance:
"FAIL AL 5"
The pressure build-up is incomplete.

Program run with


Became the chamber
1 Yes lower load
overloaded?
carry out.

No

Was the annual Perform maintenance.


2nd No
Performed maintenance? (see section 5.1)

Yes

The height was right Adjust height.


3rd No
set? (see section 3.1)

Yes

Is the
Float switch
4th Fresh water tank No
check.
filled?

Yes

https://translate.googleusercontent.com/translate_f 99/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Will the
Hose feed and
Vacuum test
solenoid valves 1 to 4
5 successful No
check.
passed? (see section 4.4)
(see section 4.2)

Yes

Is the
Amount of water
Water flow meter
6 counter No
replace.
working?
(see section 4.5)

Yes

Check the safety thermostat and that


Pressure relief valve of the chamber.
Do a comparison of the temperature and the
Pressure board (see section 3.3 and 3.4).

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 87

Page 89
Sirona Dental Systems GmbH

6.24 FAIL AL 6

Importance:
The required vacuum level
"FAIL AL 6"
is not reached (outside the
Tolerance).

Was the annual Perform maintenance.


1 No
Performed maintenance? (see section 5.1)

Yes

Was the height and


Correct skew
Inclination of the
2nd No and adjust height.
Device correctly
(see section 3.1.)
set?

Yes

Is the
Vacuum pump
Functionality
https://translate.googleusercontent.com/translate_f 100/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
3rd the vacuum pump No overtake or
replace.
given?
(see section 5.3)
(see section 4.3)

Yes

Is the
Functionality
Repair solenoid valve
4th the solenoid valves 1 No
or replace.
given to 5?
(see section 4.4)

Yes

Check the
Is a leak in the Condensate pump and
5 Hose feed lines No the neighboring
the chamber ? Check valve.
(only mod. 03)

Yes

Hose feed
replace.

88 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 90
Sirona Dental Systems GmbH

6.25 FAIL AL 7

Importance:
"FAIL AL 7" The door was opened during the
Sterilization process open

The door was open during Moisture in the


of Dry the chamber
1 Yes
Sterilization process leave, close the door and
open? Restart the program.

No

https://translate.googleusercontent.com/translate_f 101/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Check the
Door lock and
the door switch.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 89

Page 91
Sirona Dental Systems GmbH

6.26 FAIL AL 8

Importance:
"FAIL AL 8" After three vacuum phases
there is too much air in the chamber

Has been maintenance Perform maintenance.


https://translate.googleusercontent.com/translate_f 102/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
1 carried out? No (see section 5.1)

Yes

The height was right Adjust height.


2nd No
set? (see section 3.1)

Yes

Is the
Functionality
Repair solenoid valve
3rd the solenoid valves 1 No
or replace.
given to 5?
(see section 4.4)

Yes

Is a leak in the
Hose feed
4th Hose feed lines Yes
replace.
the chamber ?

No

Is the
Amount of water
Water flow meter
5 counter No
replace.
working?
(see section 4.5)

Yes

Do a comparison of the
Temperature and the
Printing board through.
(see section 3.3 and 3.4)

90 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 92
Sirona Dental Systems GmbH

6.27 FAIL AL 9

https://translate.googleusercontent.com/translate_f 103/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
Importance:
Countdown break longer
"FAIL AL 9"
than 60 seconds during the
Sterilization phase

Has been maintenance Perform maintenance.


1 No
carried out? (see section 5.1)

Yes

Is a leak in the
Hose feed
2ndHose feed lines Yes
replace.
the chamber ?

No

Is the
Functionality
Repair solenoid valve
3rd the solenoid valves 1 No
or replace.
given to 4?
(see section 4.4)

Yes

Safety thermostat
check.

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 91

https://translate.googleusercontent.com/translate_f 104/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 93
Sirona Dental Systems GmbH

6.28 FAIL AL 10

Importance:
During sterilization, the
"FAIL AL 10"
Pressure above the reference value
gone up.

Turn off the


Heating elements after Temperature sensors
1 No
Reaching the check.
Target temperature from?

Yes

Is the
Plug connection of the Secure fit
2nd No
Pressure measuring board guarantee.
securely engaged?

Yes

Motherboard
check and
possibly
replace.

https://translate.googleusercontent.com/translate_f 105/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

92 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 94
Sirona Dental Systems GmbH

6.29 FAIL AL 11

Importance:
During sterilization, the
"FAIL AL 11"
Pressure below the reference value
like.

Has been maintenance Perform maintenance.


1 No
carried out? (see section 5.1)

Yes

Is a leak in the
Hose feed
2ndHose feed lines Yes
replace.
the chamber ?

No

Is the
Functionality
Repair solenoid valve
3rd the solenoid valves 1 No
or replace.
given to 4?
(see section 4.4)

Yes

Safety thermostat
check.

https://translate.googleusercontent.com/translate_f 106/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 93

Page 95
Sirona Dental Systems GmbH

6.30 FAIL AL 12

Importance:
During sterilization it is
"FAIL AL 12"
Temperature above
Reference value increased.

Turn off the


Heating elements after Temperature sensors
1 No
Reaching the check.
Target temperature from?

Yes

Is the
Plug connection of the Secure fit
2nd No
Pressure measuring board guarantee.
securely engaged?

Yes

Does that work


3rd Motherboard No Replace the main board
flawless?

https://translate.googleusercontent.com/translate_f 107/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Yes

Do a comparison of the
Temperature and the
Printing board through.
(see section 3.3 and 3.4)

94 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 96
Sirona Dental Systems GmbH

6.31 FAIL AL 13

Importance:
"FAIL AL 13" The steam temperature sensor is
malfunction.

Is that The
1 Ambient temperature No Ambient temperature
temperature above 4 ° C must be above 4 ° C.

Yes

Are the lines


the steam
Lines correct
2nd temperature sensor No
connect.
correctly
connected?

Yes
https://translate.googleusercontent.com/translate_f 108/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Steam temperature
replace sensor.
(see section 5.4)

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 95

Page 97
Sirona Dental Systems GmbH

6.32 FAIL AL 14

Importance:
"FAIL AL 14" The top temperature sensor is
malfunction.

Is that The
1 Ambient temperature No Ambient temperature
temperature above 4 ° C must be above 4 ° C.

Yes

https://translate.googleusercontent.com/translate_f 109/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Are the lines


of the top
Lines correct
2nd Temperature sensor No
connect.
correctly
connected?

Yes

Upper temperature
replace sensor.
(see section 5.4)

96 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

Page 98
Sirona Dental Systems GmbH

6.33 FAIL AL 15

Importance:
"FAIL AL 15" The lower temperature sensor is
malfunction.

https://translate.googleusercontent.com/translate_f 110/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
Is temperature
Ambient that The
Ambient temperature
1 No
temperature above 4 ° C must be above 4 ° C.

Yes

Are the lines


of the bottom
Lines correct
2nd Temperature sensor No
connect.
correctly
connected?

Yes

Lower temperature
replace sensor.
(see section 5.4)

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 97

Page 99
Sirona Dental Systems GmbH

6.34 FAIL AL 16

https://translate.googleusercontent.com/translate_f 111/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS
"FAIL AL 16" Importance:
The pressure sensor is defective.

Is the
The secure fit of the
Plug connection of the
1 No Pressure measuring board
Pressure measuring board
guarantee.
securely engaged?

Yes

Pressure measuring board


replace.
(see section 5.7)

6.35 FAIL AL 18

Importance:
"FAIL AL 18" The door was opened during the
Drying process open.

The door was open during Moisture in the


of Dry the chamber
1 Yes
Drying process leave, close the door and
open? Restart the program.

No

Check the
Door lock and
the door switch.

98 SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03)

https://translate.googleusercontent.com/translate_f 112/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

Page 100
Sirona Dental Systems GmbH

6.36 FAIL AL 31

Importance:
During the drying process
"FAIL AL 31"
was not the vacuum level
sufficient.

Was the annual Perform maintenance.


1 No
Performed maintenance? (see section 5.1)

Yes

Was the height and


Correct skew
Inclination of the
2nd No and adjust height.
Device correctly
(see section 3.1.)
set?

Yes

Is the
Vacuum pump
Functionality
overtake or
3rd the vacuum pump No
replace.
given?
(see section 5.3)
(see section 4.3)

Yes

Is the
Functionality
Repair solenoid valve
4th the solenoid valves 1 No
or replace.
given to 5?
(see section 4.4)

Yes

Check the
Is a leak in the Condensate pump and
5 Hose feed lines No the neighboring
the chamber ? Check valve.
(only mod. 03)

Yes

Hose feed
replace.

https://translate.googleusercontent.com/translate_f 113/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

SIROCLAVE B / SIROCLAVE VS (Mod. 01 / Mod.03) 99

Page 102
101

Subject to changes in the course of technical development.

https://translate.googleusercontent.com/translate_f 114/115
3/8/2020 SIROCLAVE VS SIROCLAVE VS

«Sirona Dental Systems GmbH 2000 Language: German Printed in Germany


D 3377.076.01.01.01 03.2007 û.JNr .: 000 000 Imprimé en Allemagne

Sirona Dental Systems dmbe


in the rSA: in Canada:
Fabrikstrasse 31 Sirona Dental Systems LLC Sirona Canada Order no. 61 47 842 D 3377
64625 Bensheim 4835 Sirona Drive, Suite 100 3250 Ridgeway Drive J Unit 5
Germany Charlotte, NC 28273 Mississauga, Ontario L5L 5Y6
www.sirona.com United States Canada

https://translate.googleusercontent.com/translate_f 115/115

Das könnte Ihnen auch gefallen