Sie sind auf Seite 1von 5

Federal Register / Vol. 68, No.

9 / Tuesday, January 14, 2003 / Proposed Rules 1925

TABLE 8 OF SUBPART YYYY OF PART 63.—APPLICABILITY OF GENERAL PROVISIONS TO SUBPART YYYY—Continued


Applies to subpart
Citation Subject Explanation
YYYY

§ 63.10(d)(4) ................ Progress reports ............................................. No ............................... Compliance extensions do not apply to new
or reconstructed sources.

§ 63.10(d)(5) ................ Startup, shutdown, and malfunction reports ... No ............................... Subpart YYYY does not require reporting of
startup, shutdowns, or malfunctions.

§ 63.10(e)(1) and (2)(i) Additional CMS reports ................................... Yes.

§ 63.10(e)(2)(ii) ............ COMS-related report ....................................... No ............................... Subpart YYYY does not require COMS.

§ 63.10(e)(3) ................ Excess emissions and parameter Yes.


exceedances reports.

§ 63.10(e)(4) ................ Reporting COMS data .................................... No ............................... Subpart YYYY does not require COMS.

§ 63.10(f) ...................... Waiver for recordkeeping and reporting ......... Yes.

§ 63.11 ......................... Flares .............................................................. No.

§ 63.12 ......................... State authority and delegations ...................... Yes.

§ 63.13 ......................... Addresses ....................................................... Yes.

§ 63.14 ......................... Incorporation by reference .............................. Yes.

§ 63.15 ......................... Availability of information ................................ Yes.

3. Appendix A to Part 63 is proposed uncontrolled and controlled natural gas-fired, 5.0 Safety
to be amended by adding, in numerical stationary combustion sources. 5.1 Prior to applying the method in the
order, Method 323 to read as follows: 1.1.3 Data Quality Objectives. If you
field, a site-specific Health and Safety Plan
adhere to the quality control and quality
should be prepared. General safety
Appendix A to Part 63—Test Methods assurance requirements of this method, then
you and future users of your data will be able precautions include the use of steel-toed
* * * * * to assess the quality of the data you obtain boots, safety glasses, hard hats, and work
and estimate the uncertainty in the gloves. In certain cases, facility policy may
Method 323—Measurement of
Formaldehyde Emissions from Natural Gas- measurements. require the use of fire-resistant clothing while
Fired Stationary Sources—Acetyl Acetone on-site. Since the method involves testing at
2.0 Summary of Method high-temperature sampling locations,
Derivitization Method
An emission sample from the combustion precautions must be taken to limit the
1.0 Introduction exhaust is drawn through a midget impinger potential for exposure to high-temperature
This method describes the sampling and train containing chilled reagent water to gases and surfaces while inserting or
analysis procedures of the acetyl acetone absorb formaldehyde. The formaldehyde removing the sample probe. In warm
colorimetric method for measuring concentration in the impinger is determined locations, precautions must also be taken to
formaldehyde emissions in the exhaust of by reaction with acetyl acetone to form a avoid dehydration.
natural gas-fired, stationary combustion colored derivative which is measured
5.2 Potential chemical hazards associated
sources. This method, which was prepared colorimetrically.
with sampling include formaldehyde,
by the Gas Research Institute (GRI), is based 3.0 Definitions nitrogen oxides (NOX), and carbon monoxide
on the Chilled Impinger Train Method for [Reserved] (CO). Formalin solution, used for field
Methanol, Acetone, Acetaldehyde, Methyl spiking, is an aqueous solution containing
Ethyl Ketone, and Formaldehyde (Technical 4.0 Interferences formaldehyde and methanol. Formaldehyde
Bulletin No. 684) developed and published The presence of acetaldehyde, amines, is a skin, eye, and respiratory irritant and a
by the National Council of the Paper Industry polymers of formaldehyde, periodate, and carcinogen, and should be handled
for Air and Stream Improvement, Inc. sulfites can cause interferences with the accordingly. Eye and skin contact and
(NCASI).1 However, this method has been acetyl acetone procedure which is used to inhalation of formaldehyde vapors should be
prepared specifically for formaldehyde and determine the formaldehyde concentration. avoided.
does not include specifications (e.g., However, based on experience gained from Natural gas-fired combustion sources can
equipment and supplies) and procedures extensive testing of natural gas-fired potentially emit CO at toxic concentrations.
(e.g., sampling and analytical) for methanol, combustion sources using FTIR to measure a Care should be taken to minimize exposure
acetone, acetaldehyde, and methyl ethyl variety of compounds, GRI expects only
to the sample gas while inserting or removing
ketone. To obtain reliable results, persons acetaldehyde to be potentially present when
using this method should have a thorough the sample probe. If the work area is
combusting natural gas. Acetaldehyde has
knowledge of at least Methods 1, 2, 3, and been reported to be a significant interferent enclosed, personal CO monitors should be
4 of 40 CFR part 60, appendix A. only when present at concentrations above used to insure that the concentration of CO
50 ppm.4 However, GRI reports that the in the work area is maintained at safe levels.
1.1 Scope and Application concentration of acetaldehyde from gas-fired 5.3 Potential chemical hazards associated
1.1.1 Analytes. The only analyte sources is very low (typically below the FTIR with the analytical procedures include acetyl
measured by this method is formaldehyde detection limit of around 0.5 ppmv); acetone and glacial acetic acid. Acetyl
(CAS Number 50–00–0). therefore, the potential positive bias due to acetone is an irritant to the skin and
1.1.2 Applicability. This method is for acetaldehyde interference is expected to be respiratory system, as well as being
analyzing formaldehyde emissions from negligible. moderately toxic. Glacial acetic acid is highly

VerDate Dec<13>2002 19:47 Jan 13, 2003 Jkt 200001 PO 00000 Frm 00039 Fmt 4701 Sfmt 4700 E:\FR\FM\14JAP2.SGM 14JAP2
1926 Federal Register / Vol. 68, No. 9 / Tuesday, January 14, 2003 / Proposed Rules

corrosive and is an irritant to the skin, eyes, 7.2.4 Glacial Acetic Acid. and sampling flow rates and dry gas meter
and respiratory system. Eye and skin contact exhaust temperature every 5–10 minutes
and inhalation of vapors should be avoided. 8.0 Sample Collection, Preservation, during the run, in a signed and dated
Acetyl acetone and glacial acetic acid have Storage, and Transport notebook.
flash points of 41°C (105.8°F) and 43°C 8.1 Pre-test 8.2.5 If possible, monitor and record the
(109.4°F), respectively. Exposure to heat or fuel flow rate to the engine and the exhaust
8.1.1 Collect information about the site
flame should be avoided. oxygen concentration during the sampling
characteristics such as exhaust pipe
period. This data can be used to estimate the
6.0 Equipment and Supplies diameter, gas flow rates, port location, access engine exhaust flow rate based on the
6.1 Sampling Probe. Quartz glass probe to ports, and safety requirements during a Method 19 approach. This approach, if
with stainless steel sheath or stainless steel pre-test site survey. You should then decide accurate fuel flow rates can be determined,
probe. the sample collection period per run and the is preferred for reciprocating IC engine
6.2 Teflon Tubing. Teflon tubing to target sample flow rate based on your best exhaust flow rate estimation due to the
connect the sample probe to the impinger estimate of the formaldehyde concentration pulsating nature of the engine exhaust. The
train. A heated sample line is not needed likely to be present. You want to assure that F-Factor procedures described in Method 19
since the sample transfer system is rinsed to sufficient formaldehyde is captured in the may be used based on measurement of fuel
recover condensed formaldehyde and the impinger solution so that it can be measured flow rate and exhaust oxygen concentration.
rinsate combined with the impinger contents precisely by the spectrophotometer. You may One example equation is Equation 323–2.
prior to sample analysis. use Equation 323–1 to design your test 8.3 Post-test. Perform a sampling system
6.3 Midget Impingers. Three midget program. As a guideline for optimum leak-check (from the probe tip to pump
impingers are required for sample collection. performance, if you can, design your test so outlet). Connect a rotameter to the outlet of
The first impinger serves as a moisture that the liquid concentration (Cl)is the pump. Close off the inlet to the probe and
knockout, the second impinger contains 20 approximately 10 times the assumed observe the leak rate. The leak rate must be
mL of reagent water, and the third impinger spectrophotometer detection limit of 0.2 less than 2 percent of the sampling rate.
contains silica gel to remove residual ppmw. However, since actual detection Weigh and record each impinger
moisture from the sample prior to the dry gas limits are instrument specific, we also immediately after sampling to determine the
meter. suggest that you confirm that the laboratory moisture weight gain. The impinger weights
6.4 Vacuum Pump. Vacuum pump equipment can meet or exceed this detection are measured before transferring the impinger
capable of delivering a controlled extraction limit. contents, and before rinsing the sample probe
flow rate between 0.2 and 0.4 L/min. 8.1.2 Prepare and then weigh the midget and sample line. The moisture content of the
6.5 Flow Measurement Device. A impingers prior to configuring the sampling exhaust gas is determined by measuring the
rotameter or other flow measurement device train. The first impinger is initially dry. The weight gain of the impinger solutions and
to indicate consistent sample flow. second impinger contains 20 mL of reagent volume of gas sampled as described in
6.6 Dry Gas Meter. A dry gas meter is water, and the third impinger contains silica Method 4. Rinse the sample probe and
used to measure the total sample volume gel that is added before weighing the sample line with reagent water. Transfer the
collected. The dry gas meter must be impinger. Each prepared impinger is weighed impinger catch to an amber 40-mL VOA
sufficiently accurate to measure the sample and the pre-sampling weight is recorded to bottle with a Teflon-lined cap. If there is a
volume to within 2 percent, calibrated at the the nearest 0.5 gm. small amount of liquid in the dropout
selected flow rate and conditions actually 8.1.3 Assemble the sampling train (see impinger (<10 mL), the impinger catches can
encountered during sampling, and equipped Figure 1). Ice is packed around the impingers be combined in one 40 mL VOA bottle. If
with a temperature sensor (dial thermometer, in order to keep them cold during sample there is a larger amount of liquid in the
or equivalent) capable of measuring collection. A small amount of water may be dropout impinger, use a larger VOA bottle to
temperature accurately to within 3 °C added to the ice to improve thermal transfer. combine the impinger catches. Rinse the
(5.4 °F). 8.1.4 Perform a sampling system leak- impingers and combine the rinsate from the
check (from the probe tip to the pump outlet) sample probe, sample line, and impingers
6.7 Spectrophotometer. A
as follows: Connect a rotameter to the outlet with the impinger catch. In general,
spectrophotometer is required for
of the pump. Close off the inlet to the probe combined rinse volumes should not exceed
formaldehyde analysis, and must be capable
and observe the leak rate. The leak rate must 10 mL. The volume of the rinses during
of measuring absorbance at 412 nm.
be less than 2 percent of the planned sample recovery should not be excessive as
7.0 Reagents and Standards sampling rate of 0.2 or 0.4 L/min. this may result in your having to use a larger
8.1.5 Source gas temperature and static VOA bottle. This in turn would raise the
7.1 Sampling Reagents pressure should also be considered prior to detection limit of the method since after
7.1.1 Reagent water. Deionized, distilled, field sampling to ensure adequate safety combining the rinses with the impinger
organic-free water. This water is used as the precautions during sampling. catches in the VOA bottle, the bottle should
capture solution, for rinsing the sample be filled with reagent water to eliminate the
probe, sample line, and impingers at the 8.2 Sample Collection
headspace in the sample vial. Keep the
completion of the sampling run, in reagent 8.2.1 Set the sample flow rate between sample bottles over ice until analyzed on-site
dilutions, and in blanks. 0.2—0.4 L/min, depending upon the or received at the laboratory. Samples should
7.1.2 Ice. Ice is necessary to pack around anticipated concentration of formaldehyde in be analyzed as soon as possible to minimize
the impingers during sampling in order to the engine exhaust. (You may have to refer possible sample degradation. Based on a
keep the impingers cold. Ice is also needed to published data 5 6 for anticipated limited number of previous analyses,
for sample transport and storage. concentration levels.) If no information is samples held in refrigerated conditions
available for the anticipated levels of showed some sample degradation over time.
7.2 Analysis formaldehyde, use the higher sampling rate
7.2.1 Acetyl acetone Reagent. Prepare the of 0.4 L/min. 8.4 Quality Control Samples
acetyl acetone reagent by dissolving 15.4 g of 8.2.2 Record the sampling flow rate every 8.4.1 Field Duplicates. During at least one
ammonium acetate in 50 mL of reagent water 5–10 minutes during the sample collection run, a pair of samples should be collected
in a 100-mL volumetric flask. To this period. concurrently and analyzed as separate
solution, add 0.20 mL of acetyl acetone and 8.2.3 Monitor the amount of ice samples. Results of the field duplicate
0.30 mL of glacial acetic acid. Mix the surrounding the impingers and add ice as samples should be identified and reported
solution thoroughly, then dilute to 100 mL necessary to maintain the proper impinger with the sample results. The percent
with reagent water. The solution can be temperature. Remove excess water as needed difference in exhaust (stack) concentration
stored in a brown glass bottle in the to maintain an adequate amount of ice. indicated by field duplicates should be
refrigerator, and is stable for at least two 8.2.4 Record measured leak rate, within 20 percent of their mean
weeks. beginning and ending times and dry gas concentration. Data are to be flagged as
7.2.2 Formaldehyde. Reagent grade. meter readings for each sampling run, suspect if the duplicates do not meet the
7.2.3 Ammonium Acetate. impinger weights before and after sampling, acceptance criteria.

VerDate Dec<13>2002 19:47 Jan 13, 2003 Jkt 200001 PO 00000 Frm 00040 Fmt 4701 Sfmt 4700 E:\FR\FM\14JAP2.SGM 14JAP2
Federal Register / Vol. 68, No. 9 / Tuesday, January 14, 2003 / Proposed Rules 1927

8.4.2 Spiked Samples. An aliquot of one acceptable if the measured spike recovery is same manner as the other samples, should be
sample from each source sample set should 80 to 120 percent. Use Equation 323–4. Data collected with each set of source samples.
be spiked at 2 to 3 times the formaldehyde are to be flagged as suspect if the spike The field blank results should be less than 50
level found in the unspiked sample. It is also recovery do not meet the acceptance criteria. percent of the lowest calibration standard
recommended that a second aliquot of the 8.4.3 Field Blank. A field blank used in the sample analysis. If this criteria is
same sample be spiked at around half the consisting of reagent water placed in a clean not met, the data should be flagged as
suspect.
level of the first spike; however, the second impinger train, taken to the test site but not
spike is not mandatory. The results are sampled, then recovered and analyzed in the 9.0 Quality Control

QA/QC Specification Acceptance criteria Frequency Corrective action

Leak-check—Sections 8.1.4, 8.3 ... <2% of Sampling rate ................... Pre- and Post-sampling ................ Pre-sampling: Repair leak and re-
check Post-sampling: Flag data
and repeat run if for regulatory
compliance.
Sample flow rate ............................ Between 0.2 and 0.4 L/min .......... Throughout sampling .................... Adjust.
VOA vial headspace ...................... No headspace .............................. After sample recovery .................. Flag data.
Sample preservation ...................... Maintain on ice ............................. After sample recovery .................. Flag data.
Sample hold time ........................... 14 day maximum .......................... After sample recovery .................. Flag data.
Field Duplicates—Section 8.4.1 ..... Within 20% of mean of original One duplicate per source sample Flag data.
and duplicate sample. set.
Spiked Sample—Section 8.4.2 ...... Recovery between 80 and 120% One spike per source sample set Flag data.
Field Blank—Section 8.4.3 ............ <50% of the lowest calibration One blank per source sample set Flag data.
standard.
Calibration Linearity—Section 10.1 Correlation coefficient of 0.99 or Per source sample set ................. Repeat calibration procedures.
higher.
Calibration Check Standard—Sec- Within 10% of theoretical value .... One calibration check per source Repeat check, remake standard
tion 10.3. sample set. and repeat, repeat calibration.
Lab Duplicates—Section 11.2.1 .... Within 10% of mean of original One duplicate per 10 samples ..... Flag data.
and duplicate sample analysis.
Analytical Blanks—Section 11.2.2 <50% of the lowest calibration One blank per source sample set Clean glassware/analytical equip-
standard. ment and repeat.

10.0 Calibration and Standardization 11.0 Analytical Procedure equipment should be cleaned and the
10.1 Spectrophotometer Calibration. 11.1 Sample Analysis. A 2.0-mL aliquot analytical blank repeated.
Prepare a stock solution of 10 ppm of the impinger catch/rinsate is transferred to 12.0 Calculations and Data Analysis
formaldehyde. Prepare a series of calibration a screw-capped vial. Two mL of the acetyl
standards from the stock solution by adding acetone reagent are added and the solution is 12.1 Nomenclature
0, 0.1, 0.3, 0.7, 1.0, and 1.5 mL of stock thoroughly mixed. Once mixed, the vial is A = measured absorbance of 2 mL aliquot
solution (corresponding to 0, 1.0, 3.0, 7.0, placed in a water bath (or heating block) at B = estimated sampling rate, lpm
10.0, and 15.0 µg formaldehyde, respectively) 60 °C for 10 minutes. Remove the vial and Cl = target concentration in liquid, ppmw
to screw-capped vials. Adjust each vial’s allow to cool to room temperature. Transfer D = estimated stack formaldehyde
volume to 2.0 mL with reagent water. Add the solution to a cuvette and measure the concentration (ppmv)
2.0 mL of acetyl acetone reagent, thoroughly absorbance using the spectrophotometer at E = estimated liquid volume, normally 40,
mix the solution, and place the vials in a 412 nm. The quantity of formaldehyde mL (the size of the VOA used)
water bath (or heating block) at 60 °C for 10 present is determined by comparing the cform = formaldehyde concentration in gas
minutes. Remove the vials and allow to cool sample response to the calibration curve. Use stream, ppmvd
to room temperature. Transfer each solution Equation 323–5. If the sample response is out cform @15%02 = formaldehyde concentration in
to a cuvette and measure the absorbance at of the calibration range, the sample must be gas stream corrected to 15% oxygen,
412 nm using the spectrophotometer. diluted and reanalyzed. Such dilutions must ppmvd
Develop a calibration curve from the be performed on another aliquot of the Csm = measured concentration of
analytical results of these standards. The original sample before the addition of the formaldehyde in the spiked aliquot
acceptance criteria for the spectrophotometer acetyl acetone reagent. The full procedure is Cu = measured concentration of
calibration is a correlation coefficient of 0.99 repeated with the diluted sample. formaldehyde in the unspiked aliquot of
or higher. If this criteria is not met, the the same sample
11.2 Analytical Quality Control
calibration procedures should be repeated. Cs = calculated concentration of
10.2 Spectrophotometer Zero. The 11.2.1 Laboratory Duplicates. Two formaldehyde spiking solution added to
spectrophotometer should be zeroed with aliquots of one sample from each source the spiked aliquot
reagent water when analyzing each set of sample set should be prepared and analyzed df = dilution factor, 1 unless dilution of the
samples. (with a minimum of one pair of aliquots for sample was needed to reduce the
10.3—Calibration Checks. Calibration every 10 samples). The percent difference absorbance into the calibration range
checks consisting of analyzing a standard between aliquot analysis should be within 10 Fd = dry basis F-factor from Method 19, dscf
separate from the calibration standards must percent of their mean. Use Equation 323–3. per million btu
be performed with each set of samples. The Data are flagged if the laboratory duplicates GCVg = Gross calorific value (or higher
calibration check standard should not be do not meet this criteria. heating value), btu per scf
prepared from the calibration stock solution. 11.2.2 Analytical blanks. Blank samples Kc = spectrophotometer calibration factor,
The result of the check standard must be (reagent water) should be incorporated into slope of the least square regression line
within 10 percent of the theoretical value to each sample set to evaluate the possible (Note: Most spreadsheets are capable of
be acceptable. If the acceptance criteria are presence of any cross-contamination. The calculating a least squares line.)
not met, the standard must be reanalyzed. If acceptance criteria for the analytical blank is K1 = 0.3855 °K/mm Hg for metric units,
still unacceptable, a new calibration curve less than 50 percent of the lowest calibration (17.65 °R/in.Hg for English units.)
must be prepared using freshly prepared standard. If the analytical blank does not MW = molecular weight, 30 g/g-mole, for
standards. meet this criteria, the glassware/analytical formaldehyde

VerDate Dec<13>2002 19:47 Jan 13, 2003 Jkt 200001 PO 00000 Frm 00041 Fmt 4701 Sfmt 4700 E:\FR\FM\14JAP2.SGM 14JAP2
1928 Federal Register / Vol. 68, No. 9 / Tuesday, January 14, 2003 / Proposed Rules

24.05 = mole specific volume constant, liters Zu = volume fraction of unspiked (native) 12.5 Percent Recovery of Spike
per g-mole sample contained in the final spiked
m = mass of formaldehyde in liquid sample,
mg
aliquot [e.g., Vu/(Vu + Vs), where Vu +
Vs should = 2.0 mL ] %R =
(Csm − Z u C u ) ∗ 100 Eq. 323- 4
Pstd = Standard pressure, 760 mm Hg (29.92 Zs = volume fraction of spike solution Z sCs
in.Hg) contained in the final spiked aliquot
Pbar = Barometric pressure, mm Hg (in.Hg) [e.g., Vs/(Vu + Vs)] 12.6 Mass of Formaldehyde in Liquid
PD = Percent Difference R = 0.02405 dscm per g-mole, for metric units
Sample
Y = Dry Gas Meter calibration factor
Qe = exhaust flow rate, dscf per minute
Qg = natural gas fuel flow rate, scf per minute 12.2 Pretest Design
V 
Tm = Average DGM absolute temperature, °K m = K c AF  t  Eq. 323-5
(°R). B∗ t∗ D∗30  Va 
Tstd = Standard absolute temperature, 293 °K Cl = Eq. 323-1
(528 °R). 24.05∗ E 12.7 Dry Sample Gas Volume, Corrected to
t = sample time (minutes) 12.3 Exhaust Flow Rate Standard Conditions
Vm = Dry gas volume as measured by the
DGM, dcm (dcf).
Vm(std) = Dry gas volume measured by the Fd Q g GCVg  20.9 
Vm (std ) =
(Vm YTstd Pbar )
Q=   Eq. 323-2 Eq. 323-6
DGM, corrected to standard conditions,
dscm (dscf).
10 6  20.9 − O 2 d  (Tm Pstd )
Vt = actual total volume of impinger catch/ 12.4 Percent Difference.—(Applicable to K i YVm Pbar
rinsate, mL =
Field and Lab Duplicates) Tm
Va = volume (2.0) of aliquot analyzed, mL

(X1 − X 2 )
X1 = first value
12.8 Formaldehyde Concentration in Gas
X2 = second value
PD = ∗ 100 Eq. 323-3 Stream
O2d = oxygen concentration measured,  X1 + X 2 
percent by volume, dry basis
%R = percent recovery of spike
 2 

R  m  1g 
c form = 
MW  Vm (std )   1000 mg 
(
 1 × 10 ppm
6
) Eq. 323-7

12.9 Formaldehyde Concentration,

EP14JA03.009</MATH>
Corrected to 15% Oxygen

(20.9 − 15)
c form @15%O2 = c form ∗ Eq. 323-8
(20.9 − O2 d )

EP14JA03.008</MATH>
13.0 Method Performance with the method validation, sample Data Analysis Protocols,’’ Technical Bulletin
13.1 Precision. Based on a Method 301 degradation for 7 and 14-day hold times does No. 684, December 1994.

EP14JA03.007</MATH>
2 National Council of the Paper Industry for
validation using quad train arrangement with not exceed 2.3 and 4.6 percent, respectively,
post sampling spiking study of the method at based on a 95 percent level of confidence. Air and Stream Improvement, Inc., ‘‘Field
a natural gas-fired IC engine, the relative Therefore, the recommended maximum Validation of a Source Sampling Method for
standard deviation of six pairs of unspiked sample holding time for the underivatized Formaldehyde, Methanol, and Phenol at
samples was 11.2 percent at a mean stack gas impinger catch/rinsate is 14 days, where Wood Products Mills,’’ 1997 TAPPI
concentration of 16.7 ppmvd. International Environmental Conference.

EP14JA03.005</MATH>
projected sample degradation is below 5 3 Roy F. Weston, Inc., ‘‘Formaldehyde
13.2 Bias. No bias correction is allowed. percent.
The single Method 301 validation study of Sampling Method Field Evaluation and
the method at a natural gas-fired IC engine, 14.0 Pollution Prevention Emission Test Report for Georgia-Pacific
indicated a bias correction factor of 0.91 for Resins, Inc., Russellville, South Carolina,’’
Sample gas from the combustion source
that set of data. An earlier spiking study got August 1996.
exhaust is vented to the atmosphere after 4 Hoechst Celanese Method CL 8–4, EP14JA03.004</MATH>
similar average percent spike recovery when passing through the chilled impinger
spiking into a blank sample. This data set is ‘‘Standard Test Method for Free
sampling train. Reagent solutions and Formaldehyde in Air Using Acetyl acetone,’’
too limited to justify using a bias correction
factor for future tests at other sources. samples should be collected for disposal as Revision 0, September 1986.
13.3 Range. The range of this method for aqueous waste. 5 Shareef, G.S., et al. ‘‘Measurement of Air

Toxic Emissions from Natural Gas-Fired


EP14JA03.003</MATH>

formaldehyde is 0.2 to 7.5 ppmw in the 15.0 Waste Management


liquid phase. (This corresponds to a range of Internal Combustion Engines at Natural Gas
0.27 to 10 ppmv in the engine exhaust if Standards of formaldehyde and the Transmission and Storage Facilities.’’ Report
sampling at a rate of 0.4 Lpm for 60 minutes analytical reagents should be handled No. GRI–96/0009.1, Gas Research Institute,
and using a 40 mL VOA bottle.) If the liquid according to the Material Safety Data Sheets. Chicago, Illinois, February 1996.
6 Gundappa, M., et al. ‘‘Characteristics of
sample concentration is above this range,
EP14JA03.002</MATH>

16.0 References
perform the appropriate dilution for accurate Formaldehyde Emissions from Natural Gas-
1 NationalCouncil of the Paper Industry for
measurement. Any dilutions must be taken Fired Reciprocating Internal Combustion
from new aliquots of the original sample Air and Stream Improvement, Inc., ‘‘Volatile Engines in Gas Transmission. Volume I:
before reanalysis. Organic Emissions from Pulp and Paper Mill Phase I Predictive Model for Estimating
13.4 Sample Stability. Based on a sample Sources, Part X—Test Methods, Quality Formaldehyde Emissions from 2–Stroke
EP14JA03.001</MATH>

stability study conducted in conjunction Assurance/Quality Control Procedures, and Engines.’’ Report No. GRI–97/0376.1, Gas

VerDate Dec<13>2002 19:47 Jan 13, 2003 Jkt 200001 PO 00000 Frm 00042 Fmt 4701 Sfmt 4700 E:\FR\FM\14JAP2.SGM 14JAP2
Federal Register / Vol. 68, No. 9 / Tuesday, January 14, 2003 / Proposed Rules 1929

Research Institute, Chicago, Illinois, 17.0 Tables, Diagrams, Flowcharts, and


September 1997. Validation Data
BILLING CODE 6560–50–P

[FR Doc. 03–86 Filed 1–13–03; 8:45 am]


BILLING CODE 6560–50–C
EP14JA03.010</GPH>

VerDate Dec<13>2002 19:47 Jan 13, 2003 Jkt 200001 PO 00000 Frm 00043 Fmt 4701 Sfmt 4700 E:\FR\FM\14JAP2.SGM 14JAP2

Das könnte Ihnen auch gefallen