Beruflich Dokumente
Kultur Dokumente
Document History
Table 1 – Document History
Date Revision Description of Revision Author
27/04/2005 1V0 Initial Draft S. King
2/03/2008 2V1 Major upgrade to utility and manual S. King
17/5/2012 2v1 Minor corrections and reformat S. Brede
10/10/2012 2V3 Minor upgrade to utility for Virtual USB and Ethernet B. Fernando
6/6/2014 2V3a Company Details Updated. Major reformatting S. Brede
© Copyright
This document is protected by copyright and the information contained herein is confidential. The
document may not be copied and the information herein may not be disclosed except by written
permission of and in a manner permitted by the proprietors of IPC Global Pty Ltd.
Table of Contents
1 INTRODUCTION ........................................................................................................................................................4
1.1 Scope ............................................................................................................................................................4
1.2 Overview.......................................................................................................................................................4
1.3 Referenced Documents. ..............................................................................................................................4
1.4 Document Location .....................................................................................................................................4
1.5 Terminology. ................................................................................................................................................4
2 MAIN MENU ITEMS...................................................................................................................................................5
2.1 Start-up .........................................................................................................................................................5
2.2 Setup Menu ..................................................................................................................................................5
2.3 View Menu ...................................................................................................................................................8
2.4 Help ..............................................................................................................................................................8
2.5 Screen Pages .................................................................................................................................................8
3 SYSTEM TAB ...........................................................................................................................................................10
3.1 Status Bar ...................................................................................................................................................11
3.2 System Buttons...........................................................................................................................................12
4 CALIBRATION TAB ................................................................................................................................................16
4.1 Calibration page toolbar ............................................................................................................................18
4.2 Mitutoyo control panel ..............................................................................................................................18
4.3 CDAS panel ...............................................................................................................................................19
4.4 Calibration record information panel ......................................................................................................19
5 CDAS DEFAULTS TAB..........................................................................................................................................21
5.1 Servo Control Section ...............................................................................................................................22
5.2 Power Supply Section ................................................................................................................................22
5.3 Setting the HSM High and Low Pressure. ................................................................................................23
6 FIRMWARE UPDATING ON PCBS......................................................................................................................24
7 GUIDE TO LINEARISING A LVDT........................................................................................................................26
7.1 Initial setup ................................................................................................................................................26
7.2 Zeroing the reference device ......................................................................................................................27
7.3 Collecting reference and actual data values ..............................................................................................27
7.4 Generating the Correction Array ..............................................................................................................29
7.5 Transducer Verification .............................................................................................................................29
APPENDIX A TYPICAL DEFAULT VALUES. .....................................................................................................31
A.1 I/O diagnostics page ..................................................................................................................................31
A.2 I/O Address DIP Switch Settings..............................................................................................................31
List of Tables
Table 1 – Document History...................................................................................................................... i
Table 2 Calibration Page typical values................................................................................................... 31
Table 3 I/O Address DIP Switch Settings .............................................................................................. 31
List of Figures
Figure 2-1 System Screen. .......................................................................................................................... 5
Figure 2-2 Linearisation menu options ..................................................................................................... 5
Figure 2-3 Print Setup Screen .................................................................................................................... 6
Figure 2-4 Communications Setup. .......................................................................................................... 6
Figure 2-5 Pendant Setup .......................................................................................................................... 7
Figure 2-6 Virtual Hydraulic Pendant ....................................................................................................... 8
Figure 3-1 System Screen ......................................................................................................................... 10
Figure 3-2 I/O Card ................................................................................................................................ 11
Figure 3-3 System status bar .................................................................................................................... 11
Figure 3-4 CPU Configuration Page ....................................................................................................... 13
Figure 3-5 I/O Module Configuration Display ...................................................................................... 13
Figure 3-6 I/O Calibrate Screen.............................................................................................................. 15
Figure 4-1 Linearisation Options menu list ............................................................................................ 16
Figure 4-2 Typical calibration display page with mitutoyo control panel .............................................. 17
Figure 4-3 Typical calibration display page without mitutoyo control panel ........................................ 17
Figure 4-4 Model information display. ................................................................................................... 20
Figure 5-1 CDAS Defaults Screen ........................................................................................................... 21
Figure 6-1 Firmware Open File Dialogue. .............................................................................................. 24
Figure 6-2 I/O Configuration Page ......................................................................................................... 25
Figure 7-1 IPC Mitutoyo reference device with stepper motor control ................................................. 26
Figure 7-2 IPC Mitutoyo reference device with digital display only. ..................................................... 26
Figure 7-3 Calibration Page with Confirm ............................................................................................. 28
1 INTRODUCTION
1.1 Scope
This document provides the user with sufficient information to use the UTS 000
Maintenance Utility to setup, update and perform transducer calibration checks in an
IMACS controlled UTM system.
1.2 Overview
The Maintenance Utility is designed to setup all aspects of a UTM’s controlling IMACS
and associated transducers. With it an operator can perform all configuration, diagnostics,
transducer checking, calibration record creation/modifying. It can also be used to upgrade
firmware in the IMACS modules.
1.5 Terminology.
Note the term CDAS (Control and Data Acquisition System) refers generally to IPC
Global’s system control electronics hardware.
The term IMACS (Integrated Multi-Axis Control System) refers to IPC Global’s second
generation CDAS.
For all practical purposes within this document the terms are interchangeable.
After selecting the appropriate port the user should click the “Enable Tx/Rx display” check
box and then click the “Re-Open CDAS Port” button to ensure communications are
established with the selected parameters. Data transmitted (Tx data) to the IMACS should
be seen in the top display panel while data sent from the IMACS (Rx data) should be
displayed in the lower panel.
Data flow between the IMACS and the PC can also be saved to a log file. A check box
allows the data stream to be saved for off-line analysis. The log file is “Log.Txt” and is
located in the “C:\IPCglobal UTS\000 Maintenance Utility” folder.
The screens are sub divided into several areas and can vary in function depending on the
type of module being addressed at the time (CPU or I/O Module).
3 SYSTEM TAB
Figure 3-1 shows a typical system screen. This shows the connected IMACS chassis is
communicating with the Maintenance software via USB and displays a list of fitted
modules (in this case a CPU and one I/O module). Firmware versions of each fitted
module are displayed as well as other I/O configuration items
Physical
Address Setup
Switch
3.1.2 Jogging
Jogging refers to the movement of a control axis via the virtual pendant or a physical
pendant. Typical jogging control of axis 1 is the movement of the vertical actuator in
position control mode while for a pressure control axis such as axis 2 or 3 it is the
application of applying and dumping air or liquid pressure. Whenever a jogging command
is sent to the IMACS and loading system these arrows will become illuminated and flash.
3.1.3 Machine Status
The Machine Status indicators will appear if
the emergency stop button is pressed,
•
low hydraulic or air pressure is sensed
•
for hydraulic systems warnings will appear for a dirty oil filter, high temperature, frame
•
interlock or
• a general fault occurs.
3.1.4 Test Status
The Test Status panel indicates the current state of the application test software with a
text message and an illuminated background color.
3.1.5 Control Axis Status
The Control Axis Status panel indicates the current control state of all configured control
modules on the IMACS unit. These indicators should be in their default mode with each
controller in its idle state. The above picture shows a unit with 5 axis control.
3.1.6 Machine Limits
A machine limit will appear under a control axis status field whenever a loading axis is
driven to a jog, or machine imposed limit condition on any of its controllable inputs.
3.2 System Buttons
• Edit enters the EDIT screen for the highlighted module.
• Diagnostics enters the DIAGNOSTICS screen for the highlighted I/O module.
• Refresh reloads the table from the connected IMACS system.
By highlighting an I/O PCB and selecting the Diagnostics button the I/O Input calibration
screen is entered. This screen allows recalibration of all the I/O PCB’s input channels.
This screen is a useful diagnostic tool. By clicking Re-Calibrate and looking for any wild
variations in displayed data an errant input channel can be quickly found. Similarly any
values not within the normal range are evident.
Also if required a printed report can be generated by selecting the Print button.
See Table 2 in APPENDIX A for typical values for this page.
4 CALIBRATION TAB
All transducers used by IPC Global testing machines will have a corresponding calibration
file which needs to be correctly allocated and saved into template files for all UTS
application tests that will use that particular transducer. This display page can be used to
create and check calibration files. The calibration files used in the UTS platform all have
the .CAL file extension attached to their file name and generally a calibration file name
will contain some or all of the accompanying manufactures serial number of the
transducer. Also older type .CAR files from previous IPC software platforms may be
imported, converted and saved as .CAL files ready for use in the UTS software platform.
This section allows calibration files to be created for:
• Displacement transducers.
• Load cells.
• Strain gauges.
• Pressure transducers.
• Temperature probes.
• Flow devices.
• Weight measuring devices.
• Torque transducers.
• Voltage reference device.
The calibration program can produce calibration files in either Metric and Imperial
Measurement units as required.
This section of the Maintenance program also allows non-linear displacement transducers
to be linearised by applying a 5th order polynomial curve fitting algorithm to create a 43
point linearisation table which is used by the IMACS. There are a number of options
available for this process and the operator selects the linearization options from the Setup
main menu item as shown.
Figure 4-2 Typical calibration display page with mitutoyo control panel
There are 2 manual modes where the user can enter up to 43 data reference points and
then apply the curve fitting algorithm. With linearization option 3 selected IMACS
readings can be automatically recorded at the selected reference points or they can be
manually entered if option 4 is selected from the main menu list. When one of these
options is selected then the Mitutoyo control panel (shown above) is hidden. A typical
display page is shown below.
Figure 4-3 Typical calibration display page without mitutoyo control panel
The black background display box at the end of the panel will display the current Mitutoyo
reference devices position reading. When active the displayed values will be highlighted in
a yellow colour.
4.3 CDAS panel
The transducer which is to be calibrated can be plugged into any available channel on a
IMACS unit and then by selecting the appropriate axis or group number and channel the
graphical display can be activated to display the transducers position on a bar graph along
with digital displays showing the converted engineering value and its raw data count. The
decimal digits selection box allows the user to format the number of decimal digits to be
displayed in the converted engineering units display box.
The transducer reading from the IMACS can also be filtered by selecting x2, x4, x8 or x16
averaging.
The Gain section box has a number of selection items however for calibration purposes the
setting should be set auto or x1 mode.
4.4 Calibration record information panel
The Calibration record information panel contains the .CAL file description, transducer
description and type, span, offset, date of calibration and whether the transducer has curve
correction enabled.
If the transducer requires linearisation (curve correction is enabled) then the Correction
array, Calibration and Verification array and linearization chart will be used to enter and
display additional data.
The correction array contains 43 points that are used by the IMACS unit to compensate
for any non-linearity in the transducer.
The calibration and verification array has 2 sections, the before (calibration data) and after
(verification data) columns. The first column in each section contains the reference values,
the second column contains measured readings and the third column displays the
percentage deviation from linear (the difference of the reference and measured values).
The linearisation chart can plot the fitted curve, before and after readings with the y axis
being the percentage deviation from linear and the x axis the reference points.
The four buttons between the correction array and the calibration and verification array
are:
• The Interpret button can be used to reverse engineer the before section of the
calibration and verification array so that a new curve fit can be applied or the
transducer and the calibration record can be verified.
• The Fit button can be used when enough data has been presented in the before
section of the calibration and verification array to apply a 5th order polynomial curve
fit algorithm. The curve fit is also plotted onto the linearization chart.
• The Info button can be used to display the coefficients of the 5th order polynomial. A
typical example is shown below.
connected to the IMACS. If the box is unchecked then when that axis is selected on the
Virtual Pendant the HSM controls will be greyed out.
Figure 7-1 IPC Mitutoyo reference device with stepper motor control
• If Manual modes are selected then ensure all available hardware is setup correctly.
Figure 7-2 IPC Mitutoyo reference device with digital display only.
• Attach the LVDT to the reference device and connect it to the IMACS unit, preferably
in the channel where it will be mainly used.
• Select the Setup-Communications main menu item and check all settings are correct.
• Select the Calibration tab to display the calibration screen. If required adjust the size
of the window by maximizing it or selecting the View-Set Default Window size.
• Configure the IMACS panel on the Calibration display screen so that it matches the
axis/group and channel number where the LVDT has been plugged into. Set the
Averaging pull down selection box to x16, the Gain selection box to x1 and depending
on the span of the transducer set the Decimal digits selection box to the desired
number.
• Select the Levels toolbar button to activate the communications between the IMACS
and the PC. The graphic display bar and digital displays should now become active.
• For the Mitutoyo Controlled modes check that the digital display box is working on
the Mitutoyo Control Panel.
• Click the New toolbar button to start a fresh calibration record or the Open button if
an existing record has been saved for this transducer.
• For a new calibration record enter the transducer type, span, offset and description
into the Calibration Record Information panel. Note: The date field is automatically
loaded with the current date however this can be overwritten.
• Select the Save toolbar button to save the current values. Generally the filename is
made up of the transducers serial number making it easy to find when allocating it to a
UTS software test.
7.2 Zeroing the reference device
• Ensure that the LVDT display on the IMACS panel has the LVDT positioned within
it operating range. If not then manually adjust the transducer until it is.
• For the Mitutoyo Controlled (auto) mode select the Zero button on the Mitutoyo
Control Panel. The stepper motor controller will now attempt to position the LVDT to
its central point.
• For other calibration modes adjust the LVDT and or reference device so that the
LVDT is positioned at its central point.
• Once this is done the IMACS panel levels display should be reading zero.
• The reference device also needs to be zeroed. If the Mitutoyo Control Panel is active
then the digital display box should also be displaying zero millimetres.
7.3 Collecting reference and actual data values
• Un-check the Enable Curve Correction check box.
• For the Mitutoyo Controlled (auto) mode select the number of points to capture and
then press the Start button on the Mitutoyo Control Panel. The controller’s stepper
motor will commence from the zero point in the positive direction, then return to zero
and then run through the negative range of the LVDT before returning back to the
zero point. As a reference point is positioned that value and the actual value from the
IMACS will be saved to the Calibration and Verification Array in the Before sections.
• For the Mitutoyo Controlled (manual) mode position the reference device to the
required position then move the mouse pointer to a free location on the Calibration
and Verification Array and then right click the mouse to bring up a popup menu. From
the menu select the Take Calibration Reading option. A confirm prompt will appear for
the operator to check that the data looks ok. Selecting the Yes button will then take
the reference and actual reading and save it to the array. This is repeated for all
reference points and Note that there should be equal points taken in both the positive
and negative ranges of the LVDT.
• For the Manual (User enters Reference value only) mode left click the mouse pointer
on the mm Ref. column in the Before section of the Calibration and Verification Array
and enter all the reference point values including the zero point. At least 11 points (5
positive, 5 negative and the zero point) should be entered but up to 43 points can be
used. Then adjust the reference device to the required positions, move the mouse
pointer to the corresponding reference value location in the Calibration and Verification
Array and right click the mouse to bring up a popup menu. From the menu select the
Take Calibration Reading option. A confirm prompt will appear for the operator to
check that the data looks ok. Selecting the Yes button will then take the actual reading
from the IMACS and save it to the array. This is repeated for all reference points.
• For the Manual (User enters Reference and IMACS values) mode left click the mouse
pointer on the mm Ref. column in the Before section of the Calibration and Verification
Array and enter all the reference and IMACS values, including the zero reference
point.
• The calibration chart will also be updated as each reference and IMACS value is added
to the calibration and verification array.
7.4 Generating the Correction Array
• Once enough reference and IMACS values are captured and saved then the next step
is to apply a 5th order polynomial curve fitting algorithm to produce a correction curve
on the chart. This is done by pressing the Fit button which is located between the
Correction Array and the Calibration and Verification array.
• Press the Apply button to generate the 43 point correction array that the IMACS uses
for linearization.
• Select the Save toolbar button to save the current values. Generally the filename is
made up of the transducers serial number making it easy to find when allocating it to a
UTS software test.
7.5 Transducer Verification
• Tick the Enable Curve Correction check box to enable curve correction when the
calibration record is loaded into the IMACS.
• Check that the LDVT central point (0mms) and reference device are zeroed correctly.
If not then follow the instructions in section 3.3.2 Zeroing the reference device.
• For the Mitutoyo Controlled (auto) mode select the Verify button on the Mitutoyo
Control Panel. The controller’s stepper motor will commence from the zero point in the
positive direction, then return to zero and then run through the negative range of the
LVDT before returning back to the zero point. As a reference point is positioned that
value and the actual value from the IMACS will be saved to the Calibration and
Verification Array in the After sections.
• For the Mitutoyo Controlled (manual) mode position the reference device to the
reference position then move the mouse pointer to the corresponding reference value
location in the Calibration and Verification Array and right click the mouse to bring up a
popup menu. From the menu select the Take Verification Reading option. A confirm
prompt will appear for the operator to check that the data looks ok. Selecting the Yes
button will then save the verification data to the array. This should be repeated for all
reference points.
• For the Manual (User enters Reference value only) mode adjust the reference device to
the reference position then move the mouse pointer to the corresponding reference
value location in the Calibration and Verification Array and right click the mouse to
bring up a popup menu. From the menu select the Take Verification Reading option. A
confirm prompt will appear for the operator to check that the data looks ok. Selecting
the Yes button will then save the verification data to the array. This should be repeated
for all reference points.
• The calibration chart will also be updated as each reference and IMACS value is added
to the calibration and verification array.
• Select the Save toolbar button to save the verification data. Generally the filename is
made up of the transducers serial number making it easy to find when allocating it to a
UTS software test.
Parameter X1 X4 X16
Zero error <± 25 <± 100 <±400
Zero error noise <± 10 <± 10 <± 10
Span Reference 16384 8192 26214
Span measured 14900 ± 100 6400 ± 50 24600 ± 150
Span error 1450 ± 40 1800 ± 25 1650 ± 100
Span noise <15 <15 <15
Table 2 Calibration Page typical values.
- END -