Sie sind auf Seite 1von 18

Error analysis tree for Sphinx surgical lasers

Sphinx

Version Author Date Description


01 NA / UA 07.10.2009 Start Version
02 NA / UA 18.11.2009
03 NA / UA 19.03.2010
04 NA / UA 07.07.2010 Texte überarbeitet, Konsolenfehler eingepflegt.
05 NA 19.07.2010 Von HT abgenommene Texte eingepflegt.
06 NA 05.08.2010 FPB-ErrorCodes ‚Startup’ jeweils in der Priorisierung an das Ende der jeweiligen FPB-ErrorCodes gestellt.
07 UA 14.02.2011 For Software OSCF04v06

© LISA laser products OHG, D-37191 Katlenburg, Max-Planck-Straße 1, Tel. (49) 05556-9938-0, Fax (49) 05556-9938-10
RelationOfErrorcodeToFailnumber- gespeichert am 20.04.12
KI: Error code vs. fail number of MPL / Sphinx

Explanatory texts
German

Text Text
number
0 Wärmeüberlast.
1 Fehler.
20 Interlock offen.
21 Fußschalterfehler.
22 Interlock-Schaltkreis fehlerhaft.
23 Fehler an der externen Warnlampe.
24 Faser entfernt oder überhitzt.
25 Laser Stop ausgelöst.
26 Fehler in der Netzversorgung.

English

Text Text
number
0 Heat overload.
1 Laser error.
20 Interlock open.
21 Foot switch error.
22 Interlock circuit faulty.
23 External warning light error.
24 Fiber disconnected or fiber connector hot.
25 Laser Stop activated.
26 Error in mains supply.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 2 von 18
KI: Error code vs. fail number of MPL / Sphinx

Handling advices
German

Advice Text
number
0 Laser neustarten. Bei erneutem Fehler Service informieren.
1 Bitte warten.
2 Bei erneutem Fehler Service informieren.
20 Interlock prüfen.
21 Fußschalter und -anschluss prüfen.
22 Interlock-Schaltkreis prüfen.
23 Externe Warnlampe prüfen.
24 Faser prüfen und anschließen.
25 Laser Stop prüfen.
26 Netzleitung oder Anschluss prüfen.

English

Advice Text
number
0 Restart laser. Inform laser service if error recurs.
1 Please wait.
2 Inform laser service if error recurs.
20 Please check interlock and interlock connector.
21 Please check foot switch and foot switch connector.
22 Please check the interlock circuit.
23 Please check external warning light circuitry.
24 Please check and attach fiber.
25 Please check mains supply and power supply cord.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 3 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
1 40 0/1 Overtemperature of the F436 The system became too hot Keep the system switched-on and wait
chiller fluid. during normal operation. until the cooling fluid has reached its
setpoint. This may take time depending
on the current temperature.
Setpoint temperature: [Tset /
Defect of hot gas bypass valve. Change the hot gas bypass valve.
°C]
Current temperature: [Tcur /
°C] Defect of compressor. Change compressor.

2 41 1/0 Undertemperature of the F437 Cooling fluid is too cold. Keep the system switched-on and wait
chiller fluid. until the cooling fluid has reached its
setpoint. This may take time depending
on storage temperature.
Setpoint temperature: [Tset /
Defect of hot gas bypass valve. Change the hot gas bypass valve.
°C]
Defect of compressor. Change compressor.
Current temperature: [Tcur /
°C]
3 28 1/0 Chiller fluid level is too low. F433 Fluid level too low. Fill up fluid.
4 35 0/1 Pressure in cooling system F431, System became too hot during Keep the system switched-on and wait
is too high. F432 normal operation. until the pressure has reached its
setpoint.
Maximum pressure: [Pset /
bar] Defect of airfan. Replace airfan.
Current pressure: [Pcur /
bar]
5 36 1/0 Pressure in cooling system F430 Too less refrigerant. Fill up refrigerant.
is too low.
Defect of hot gas bypass valve. Change the hot gas bypass valve.
Minimum pressure: [Pset /
bar]
Current pressure: [Pcur /
bar]
RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 4 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
6 27 1/0 Airfan speed is out of F439 Defect of airfan. Replace airfan.
tolerance.
Setpoint speed: [SpeedSet /
rpm]
Current speed: [SpeedCur /
rpm]
7 37 1/0 Cooling fluid flow is too low. F440 Leaking cooling system. Search leak and seal properly. Refill
cooling fluid.
Setpoint flow: [FlowSet /
l/min] Defect of fluid pump. Replace fluid pump.
Current flow: [FlowCur /
l/min]
8 25 1/0 No valid signal from the fluid F435 Fluid temperature sensor defect. Replace sensor.
temperature sensor.
9 26 1/0 No valid signal from the air F434 Air temperature sensor defect. Replace sensor.
temperature sensor.
10 23 1/0 Failure of the service switch. F004 Service switch defect. Replace SCU board or service switch.

11 14 20 / 20 Interlock is disconnected. F006 - Connect interlock.

12 13 21 / 21 Failure of the foot switch. F007, Short circuit in foot switch Replace foot switch.
F008, channel 1 and/or 2.
F017
Foot switch channels 1 and 2 are No action required. In the event of
not symmetric. repeated occurrence, replace foot switch.

13 15 1/0 Failure of the shutter. F009 Defect of shutter. Replace shutter.


14 12 22 / 22 Fuse on interlock connector F005 Current on interlock connector > Limit current on interlock connector to 1A.
has triggered. 1A.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 5 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
15 99 1/0 Error in the user console. F003 Error in BPC. Refer to error message on BPC.
16 20 23 / 23 Failure of the external F014 Error in SCU board. Change SCU board.
warnlamp control.
17 0 1/0 FPGA on SCU board has F029 FPGA on SCU board not Program FPGA on SCU board.
not been programmed. programmed correctly.

18 16 24 / 24 Release was set without a F020, Fiber removed while release was No action required. Do not remove the
connected fiber. F101 set. fiber while release is set.

Fiber port became hot while Check coupling on fiber connector.


release was set.
Defect of fiber detection. Replace fiber detection.
19 96 1/0 Failure of the SCU system F030 The system test of the SCU did Refer to the in parallel occurred failures
test. not succeed. and/or error codes to examine the reason
of this failure.
Failed test: [Number of
failed test]
More Details in SI002

20 95 1/0 Failure of the PEC system F444 The system test of the PEC did Refer to the in parallel occurred failures
test. not succeed. and/or error codes to examine the reason
of this failure.
Failed test: [Number of
failed test]
21 4 1/0 Failure of the SCU low F000, Defect of low voltage power Replace the power supply unit in the
voltage supply. F011, supply unit in LCU. LCU.
F012,
F028
Defect in one of the attached Replace the defect part.
units of the power supply unit
causing overcurrent.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 6 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
22 81 1/2 Laser pulse energy has not F021 Defect of laser resonator or laser Replace resonator or lower the requested
been reached. resonator at its limit. energy and/or frequency and/or change
the pulse width.
Defect of pump power supply. Change defect part of pump power
supply.

23 38 1/2 Deviation between both F022 Defect of detector. Replace the detector (sensor and/or LHD
laser pulse detector values depending on defect part).
is too large.
Detector not calibrated correctly. Perform a calibration of the detector.

24 82 1/2 Detected energy at least F023 Error in regulation. No action required. In the event of
100% higher than setpoint. repeated occurrence, contact LISA.
Resonator characteristics Restart system. In the event of repeated
changed since system test. occurrence, replace resonator.

Pump energy deviation from Replace pump energy unit.


setpoint too large.

25 83 1/2 Detected energy at least F024 Error in regulation. No action required. In the event of
20% higher than setpoint. repeated occurrence, contact LISA.
Resonator characteristics Restart system. In the event of repeated
changed since system test. occurrence, replace resonator.
Pump energy deviation from Replace pump energy unit.
setpoint too large.
26 85 1/2 Detected energy at least F025 Error in regulation. No action required. In the event of
20% less than setpoint. repeated occurrence, contact LISA.
Resonator characteristics Restart system. In the event of repeated
changed since system test. occurrence, replace resonator.
Pump energy deviation from Replace pump energy unit.
setpoint too large.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 7 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
27 84 1/2 Detected energy at least F026 Error in regulation. No action required. In the event of
50% less than setpoint. repeated occurrence, contact LISA.
Resonator characteristics Restart system. In the event of repeated
changed since system test. occurrence, replace resonator.
Pump energy deviation from Replace pump energy unit.
setpoint too large.
28 21 1/0 Communication error F038 The data received by the SCU is Check that the software MplPmaScu and
between SCU and BPC. invalid. of the BPC are compatible. In the event
of repeated occurrence, contact LISA.
Defect of the communication Check the cable between SCU and BPC.
hardware.

29 6 1/0 Connection error between F010, SCU and PEC not or not Check the optical fibers between SCU
SCU and PEC. F013, correctly connected by optical and PEC and that the correct ports are
F031, fibers. connected together.
F415,
Either the SCU, the PEC or both Restart system. In the event of repeated
F416
are not working correctly. occurrence, replace SCU and/or PEC.
The FPGA software of the SCU Check that both the FPGA of the SCU
and the PEC are not compatible. and of the PEC are programmed with a
compatible software.
30 18 1/0 Communication timeout F019 PEC is not connected to SCU. Check connection between SCU and
from PEC to SCU. PEC.
Software error in PEC. Restart system. In the event of repeated
occurrence, replace PEC.
31 79 1/2 Unexpected laser emission F016, Heat pump energy too high. Contact LISA.
has been detected. F108,
F109 Defect of SCU. Replace SCU.
Defect of PPS. Replace defect part of PPS.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 8 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
Detector not calibrated correctly. Perform a calibration of the detector.
32 39 1/2 Pulse configuration has not F015 The parameters of this laser Use other parameters and contact LISA.
been accepted by FPGA on pulse configuration are not valid
SCU. for the FPGA.

33 2 26 / 26 The voltage of at least one F035, Line supply below 207V or above Measure line voltage under full load.
supply phase is out of F036, 253V Ensure line voltage is in specification
tolerance. F037 (230V +/- 10%).

34 98 1/2 An error has been detected F002 System behaviour not consistent Refer to the in parallel occurred failures
by the supervisory circuit. to setpoints. and/or error codes to examine the reason
of this failure.
35 5 1/0 A software error occured on F001 Software error. Contact LISA and restart system.
the SCU.
Software class code:
[Software class code]
Class internal error: [Class
internal error]
36 1 26 / 26 At least one supply phase is F032, The line voltage of at least one Measure line voltage under full load.
defective. F033, phase is missing. Ensure line voltage is in specification
F034 (230V +/- 10%).

37 97 1/2 An error occured in the PPS. F018 An error occurred in the PPS. Refer to the in parallel occurred failures
and/or error codes to examine the reason
of this failure.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 9 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
38 3 1/0 LHD is not connected. F100 LHD board not connected to Connect the LHD board.
SCU.

39 29 1/2 The shutter was open F102 Defect of shutter. Replace shutter.
without release.

Malfunction in shutter control on Restart system. In the event of repeated


SCU board. occurrence, replace SCU.

40 30 1/0 The shutter was open during F103 Defect of shutter. Replace shutter.
the system test.
Malfunction in shutter control on Restart system. In the event of repeated
SCU board. occurrence, replace SCU.
41 31 1/2 The active signal was set F104 Defect of SCU board. Replace SCU.
without release.
42 32 1/2 Release was set while a fail F105 Defect of SCU. Replace SCU.
signal was active.
43 33 1/2 The shutter was open while F106 Defect of shutter. Replace shutter.
a fail signal was active.
Malfunction in shutter control on Restart system. In the event of repeated
SCU board. occurrence, replace SCU.
44 34 1/2 The sync signal to the PEC F107 Defect of SCU board. Replace SCU.
was high while a fail signal
was active.
45 80 1/2 The pulse frequency differs F027 Malfunction of SCU board. Restart system. In the event of repeated
from its setpoint. occurrence, replace SCU.
46 10 25 / 25 The laser stop button is not F421 Laser stop button not attached. Attach laser stop button.
attached to the system.

47 45 1/2 The motor in the laser head F400 Motor defect / blocked. Replace or free motor.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 10 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
is blocked. Defect in motor control MST. Replace motor control MST.
48 43 1/0 No motor is attached to the F401 Motor not attached. Attach motor.
MST.
Defect in motor control MST. Replace motor control MST.
49 44 0/1 Overtemperature of the F402 System over heated. Wait for system cool down.
motor conrol.
50 42 1/0 No MST is attached. F403 Motor control MST not attached. Attach MST.
51 78 1/2 No resonator is selected for F404 No resonator selected by the Select at least one resonator depending
the next pulse. switches on the PEC. on the current system.
No resonator selected by the Restart system. In the event of repeated
SCU. occurrence, contact LISA.
52 8 1/0 A software error occured on F405 Software error. Contact LISA and restart system.
the PEC.
Software class code:
[Software class code]
Class internal error: [Class
internal error]
53 7 1/0 Failure of the PPS low F424, Defect of low voltage power Replace the low voltage power supply
voltage supply. F426, supply unit in PPS. unit in the PPS.
F427,
Defect in one of the attached Replace the defect part.
F428,
units of the low voltage power
F429
supply unit causing overcurrent.
54 46 1/0 Error in the high voltage F407 Ready input doesn’t exist or the Check the wire and connection of
supply 1. heatsink temperature is above connector X7 at HVS. Wait until the
85°C. heatsink temperature is less than 79°C.
55 47 1/0 Error in the high voltage F408 Ready input doesn’t exist or the Check the wire and connection of
supply 2. heatsink temperature is above connector X7 at HVS. Wait until the
85°C. heatsink temperature is less than 79°C.
56 48 1/0 Failure of simmer 1. F409, No ignition of the flash lamp. Check the simmer or the flash lamp.
RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 11 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
F445
57 49 1/0 Failure of simmer 2. F410, No ignition of the flash lamp. Check the simmer or the flash lamp.
F446
58 91 1/0 An error occured in FPU1. F411 Error on FPU1. Refer to the in parallel occurred failures
and/or error codes to examine the reason
of this failure.
59 92 1/0 An error occured in FPU2. F412 Error on FPU2. Refer to the in parallel occurred failures
and/or error codes to examine the reason
of this failure.
60 11 25 / 25 The laser stop button is F420 Laser stop button pressed. Release laser stop button.
pressed.
61 100 1/2 An error occured on the F414 Error on SCU. Refer to the in parallel occurred failures
SCU. and/or error codes to examine the reason
of this failure.
62 101 1/2 An error occured on the F417 Error on SCU. Refer to the in parallel occurred failures
SCU or the connection and/or error codes to examine the reason
between SCU and PEC of this failure.
timed out.
SCU is not connected to PEC. Check connection between PEC and
SCU.
Software error in SCU. Restart system. In the event of repeated
occurrence, replace SCU.
63 9 1/0 FPGA on PEC board has F418 FPGA on PEC board not Program FPGA on PEC board.
not been programmed. programmed correctly.
Note: error may not be
transmitted to BPC because of
missing program in FPGA.
64 51 1/2 Minimum frequency of PEC F419 SCU transmitted a frequency Use other parameters and contact LISA.
has been underrun. value too low for the PEC.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 12 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
65 50 0/1 Temperature of the heat F438 Temperature of heat sink too Wait for cool down.
sink is too high. high.
Current temperature: [Tcur /
°C]
66 89 1/2 Voltage on the capacitors is F413 Capacitors not discharged by Remove other failure.
too high after the laser flash lamp due to other failure.
emission.
Defect of thyristor bank. Replace thyristor bank.
67 88 1/2 Faulty control of the pulse F422 Software error in PEC. Contact LISA.
generation in the FPGA on
the PEC.
68 90 1/2 The voltage on the thyristor F423 Calibration of thyristor bank Perform calibration of thyristor bank.
bank exceeded its setpoint. wrong.
Capacitor not discharged on No action required.
previous pulse, because the pulse
generation has been stopped
unexpectedly (e.g. due to a
previously occurred error).
69 64 1/0 FPB1 is starting. F501 Start of FPB1. No action required.

70 52 1/0 DC overvoltage of FPU1. F502 The DC voltage in the capacitors Wait until the overvoltage discharged.
is higher than the maximum The FPU will startup automatically.
voltage allowed (> 407V).
71 53 1/0 The FPU1 supply voltage is F503 The AC input voltage is higher Wait until the AC input voltage is less
above the maximum value. than the maximum voltage allowed than 260V. The FPU will startup
(> 270V). automatically.
72 54 1/0 The FPU1 supply voltage is F504 The AC input voltage is less than Wait until the AC input voltage is above
below the minimum value. the minimum voltage allowed. 170V. The FPU will startup automatically.
73 55 1/0 No DC voltage link at FPU1. F505 No DC voltage exists after power Check the fuse F2. Check for a short
up. circuit between UZ+ and UZ-. Check the
wire from UZ+ and UZ-.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 13 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
74 56 1/0 The DC output voltage of F506 The DC voltage is below the Restart the laser system.
FPU1 is below its minimum minimum voltage allowed after
value at startup. power up.
75 57 1/0 Overcurrent in FPU1 (PFC). F507 The AC current of the FPU Restart the laser system.
exceeded the maximum limit.
76 58 1/0 Overcurrent in FPU1 (buck F508 The DC current exceeded the Check the high voltage supply for a short
converter). maximum limit at power up. circuit.
77 59 1/0 Supply voltage of FPU1 is F509 The AC input voltage is switched If the error occurs during normal
missing. off. operation of the laser, check the fuse F1
or the wire connection.
78 60 1/0 Error in the analog interrupt F510 An error exists at the FPB Change the FPB microcontroller unit.
of FPU1. microcontroller unit.
79 61 0/1 The temperature of the F511 The temperature of the heatsink Wait until the temperature is less than
FPU1 heat sink is out of exceeded 65°C. 60°C.
tolerance.
80 62 1/0 Overcurrent in FPU1 (load). F512 The DC current exceeded the Check the High voltage supply for a short
maximum limit. The current was circuit.
too high for a time of 4s.
81 63 1/0 Error in the FPU1 hardware. F513, There are hardware errors at the Change the FPB passive components.
F514, FPB passive components.
F515,
F516
82 77 1/0 FPB2 is starting. F601 Start of FPB2. No action required.

83 65 1/0 DC overvoltage of FPU2. F602 The DC voltage in the capacitors Wait until the overvoltage discharged.
is higher than the maximum The FPU will startup automatically.
voltage allowed (> 407V).
84 66 1/0 The FPU2 supply voltage is F603 The AC input voltage is higher Wait until the AC input voltage is less
above the maximum value. than the maximum voltage allowed than 260V. The FPU will startup
(> 270V). automatically.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 14 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
85 67 1/0 The FPU2 supply voltage is F604 The AC input voltage is less than Wait until the AC input voltage is above
below the minimum value. the minimum voltage allowed. 170V. The FPU will startup automatically.
86 68 1/0 No DC voltage link at FPU2. F605 No DC voltage exists after power Check the fuse F2. Check for a short
up. circuit between UZ+ and UZ-. Check the
wire from UZ+ and UZ-.
87 69 1/0 The DC output voltage of F606 The DC voltage is below the Restart the laser system.
FPU2 is below its minimum minimum voltage allowed after
value at startup. power up.
88 70 1/0 Overcurrent in FPU2 (PFC). F607 The AC current of the FPU Restart the laser system.
exceeded the maximum limit.
89 71 1/0 Overcurrent in FPU2 (buck F608 The DC current exceeded the Check the high voltage supply for a short
converter). maximum limit at power up. circuit.
90 72 1/0 Supply voltage of FPU2 is F609 The AC input voltage is switched If the error occurs during normal
missing. off. operation of the laser, check the fuse F1
or the wire connection.
91 73 1/0 Error in the analog interrupt F610 An error exists at the FPB Change the FPB microcontroller unit.
of FPU2. microcontroller unit.
92 74 0/1 The temperature of the F611 The temperature of the heatsink Wait until the temperature is less than
FPU2 heat sink is out of exceeded 65°C. 60°C.
tolerance.
93 75 1/0 Overcurrent in FPU2 (load). F612 The DC current exceeded the Check the High voltage supply for a short
maximum limit. The current was circuit.
too high for a time of 4s.
94 76 1/0 Error in the FPU2 hardware. F613, There are hardware errors at the Change the FPB passive components.
F614, FPB passive components.
F615,
F616

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 15 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
95 19 1/0 Comparison of both detector F110 Connection between main Check connection of the PMA on the
channels is impossible due regulation (PMA) and supervisory SCU board.
to at least one missing circuit defective.
Replace SCU board.
value.

96 22 1/2 BPC mode is unknown to F111 Software versions are not Update the software. In the event of
supervisory circuit. compatible. repeated occurrence, contact LISA.
97 86 1/2 More dose pulses have F112 Software error in PMA on SCU Contact LISA.
been emitted than set. board.
98 87 1/2 More burst pulses have F113 Software error in PMA on SCU Contact LISA.
been emitted than set. board.
99 17 1/0 A watchdog reset occured F406 Software error in PEC. Restart system. In the event of repeated
for the PMA on the PEC occurrence, replace PEC.
board.
100 24 1/0 The connection between the F441 Cable between PEC and CDB Connect or replace cable.
PEC and the chiller is disconnected or defective.
defective.
Software versions are not Update the software. In the event of
compatible. repeated occurrence, contact LISA.
101 93 1/0 The connection between the F442 Cable between PEC and CDB Connect or replace cable.
PEC and the FPB is disconnected or defective.
defective.
Software versions are not Update the software. In the event of
compatible. repeated occurrence, contact LISA.
102 94 1/0 The HVS test of the PEC did F443 HVS defect. Restart system. In the event of repeated
not succeed. occurrence, replace HVS.
HVDU defect. Restart system. In the event of repeated
occurrence, replace HVDU.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 16 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
FPB defect. Restart system. In the event of repeated
occurrence, replace FPB.
200 Highest 1/0 The connection between the F200, Cable between Console and SCU Connect or replace cable.
Priority user panel and the SCU is F201, disconnected or defective.
defective. F202,
F203, Software error in Console. Restart system. In the event of repeated
F204, occurrence, replace Console.
F205, Software error in SCU. Restart system. In the event of repeated
F206, occurrence, replace SCU.
F207
201 Highest 1/1 User memo data defective. F226 CF card with the user date is Replace CF card.
Priority The system is trying to missing or defective
repair the file system.
Software error in the User panel. Restart system. In the event of repeated
Please wait at least 3
occurrence, replace Console.
Minutes
202 Highest 1/2 User memo data defective. F220, CF card with the user date is Replace CF card.
Priority F221, missing or defective
F222,
F223, Software error in the User panel. Restart system. In the event of repeated
F224, occurrence, replace the User panel.
F225,
F227,
F228
203 Highest 1/0 A watchdog reset occurred F300 Software error in the User panel. Restart system. In the event of repeated
Priority in the user panel. occurrence, replace the User panel.
204 Highest 1/2 Required laser power not F301 Resonator hot or may be fogged. Wait a few minutes and then restart the
Priority reached. System is working system. If this did not work try again later.
with reduced power.
Resonator defective. Restart system. In the event of repeated
occurrence, replace the Resonator.

Software error in the SCU. Restart system. In the event of repeated


occurrence, replace the SCU.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 17 von 18
KI: Error code vs. fail number of MPL / Sphinx

Error Priority Explanatory Detailed message Fails Cause of defect Recommended action
code text
[ ] = Parameter 1/2
/ handling
advice
Software error in the User panel. Restart system. In the event of repeated
occurrence, replace the User panel.

205 Highest 1/0 Unknown laser power option F302 Software error in the User panel. Restart system. In the event of repeated
Priority received. occurrence, replace the User panel.

Software error in the SCU. Restart system. In the event of repeated


occurrence, replace the SCU board.

206 Highest 1/0 FRAM memory error in user F303 Software or hardware error in the Restart system. In the event of repeated
Priority panel. User panel. occurrence, replace the User panel.

207 Highest 1/0 User panel cannot establish F304 Cable between Console and SCU Connect or replace cable.
Priority connection to SCU. disconnected or defective.
Software error in Console. Restart system. In the event of repeated
occurrence, replace Console.
Software error in SCU. Restart system. In the event of repeated
occurrence, replace SCU.

RelationOfErrorcodeToFailnumber
gespeichert am 20.04.12
Seite 18 von 18

Das könnte Ihnen auch gefallen