Beruflich Dokumente
Kultur Dokumente
No. DESCRIPTION
Physical Specifications
External dimensions : 320(W) x 90(H) x 160(D)mm
Weight : Approx. 2,0 kg
Classification : Class II, Type CF, IPX1 (Drip proof)
Fuse : TL 0.2 A (for 100V Area), TL 0.1 A (for 200V Area)
Electrical specifications
Power Supply
AC power source : AC 100-127V±10% 50/60Hz (for 100V Area) and
AC 200-240±10% 50/60Hz (for 200V Area)
External DC power supply : DC 9-15V 5W
Internal battery : Rechargeable, DC 3,6V 1500mAh NiMH Model : BP-55
Continuous operation Battery life is Approx 12 hours (at flow rate of 5mL/h using new battery in fully
recharged state)
Alkaline batteries : Uses four LR6 Type AA size alkaline batteries DC 6V.
Continuous operation Battery life is Approx 24 hours (at flow rate of 5mL/h using new battery)
Performance specifications
Applicable syringe : 10mL, 20mL, 30mL, and 50mL syringes from :
TOP, TERUMO,NIPRO,JMS,B-D,MONOJECT and B.BRAUN syringes
plus one additional syringe registered by user per syringe size
Flow range : 0.1 - 150.0mL/h (10mL syringe)
0.1 - 150.0mL/h (20mL syringe)
0.1 - 150.0mL/h (30mL syringe)
0.1 - 150.0mL/h (50mL syringe)
Settings adjustable in 0.1mL increments
Agent range : 0.1 - 1000.0mg
Solution range : 0.1 - 1000.0mL
Weight range : 0.1 - 300kg
Dosage range : 0.01 - 99.99 ug/kg/min 0.01-99.99mg/kg/h
Total infusion volume range : 0.1 - 1000.0mL
Infusion accuracy : Brand-specified mode
Mechanical accuracy : ±1%
Accuracy including the syringe : ±3%
*for at least one hour of infusion at flow rate of at least 1.0mL/h
Alarm
Low volume
Occlusion
Overlord
Low battery
syringe barrel / syringe barrel flange/
Syringe plunger clamp disengaged
Power supply switch
Reminder
Other device malfunctions
Special functions
Auto power OFF : During battery operation, when pump is left for 3 minutes with infusion stopped
or in an alarm state, the buzzer sound. When buzzer has been left sounding for
3 minutes, the power shuts off automatically.
Alarm reminder : Unless the cause of an alarm is collected within 2 minutes after operator passed
the SILENCE key. The alarm will sound again to remind operator of problem.
Buzzer setting : Audible indication of standby and key operation can be turned on or off.
Power supply switch alarm : An alarm sound when Ac power supply has stopped and the unit has switched
over to battery operation.
History Function : History of pump operation can be chacked when infusion started and stopped.
Key lock function : During infusion, places the system in the key lock state
KVO Function : After infusion completed, the device continues to operate at 0.1mL/h to
prevent thrombus.
Environmental operations
Operting conditions
Ambient temperature : 5 to 40ºC
Relative humidity : 20 to 90% (no condensation)
Atmospheric pressure : 70 to 106kPa
Storage requirements
Storage conditions
Ambient temperature : -20 to 45ºC
Relative humidity : 10 to 95% (no condensation)
Atmospheric pressure : 50 to 106kPa
Compositions :
1 . Syringe pump TOP-5510 : 1 unit
2 . AC Power source cord : 1 pc
3 . Pole Clamp : 1 pc
4 . Operation guide : 1 pc