Beruflich Dokumente
Kultur Dokumente
PLAN 2
Optimization Structures
The optimization structures created for Plan 1 were used for Plan 2. I created two more control
structures called ‘NS_Avoid_02a and NS_Avoid_02b’(Fig 7a and Fig 7b) By using the Region
of Interest (ROI) to define the extent of these structures, I cropped the body structure (or
external), 2mm from PTV_4140 and PTV_5040 respectively. I put an upper constraint on these
structures at level 3 of my optimization to prevent stray high dose outside of the PTV_4140 and
PTV_5040. I am able to reduce stray high doses for static field IMRT plans with these control
structures (NS_Avoid_02a and NS_Avoid_2b).
Fig7a: Showing body cropped 2mm from PTV_4140 Fig7b: Showing body cropped 2mm from PTV_5040
1
Plan Parameters
I used the same plan parameters for both Phase 1 and Phase 2
Energy: 6MV
Gantry Angles: I used the suggested gantry angles for the SupaFirefly technique: 60°, 80°, 120°,
140°, 160°, 180°, 200° (Fig 8). I set my isocenter at the geometric center of the PTV. I used the
same gantry and collimator arrangements for the phase 1 and phase 2 plans.
2
Calculation Algorithm: Anisotropic Analytical Algorithm (AAA) version 15.6.06
I started this plan with an objective to have 100% of dose to the PTV with a priority of 150. For
the OARs, I started with a priority of 85%. I put in the minimum dose constraint requirement for
the OARs but made sure I started off with good target coverage. During optimization, I changed
the priorities and objectives for the targets and OARs as needed by increasing priority for hard to
meet constraints and minimizing less to meet constraints. For both plans, I set my NTO priority
3
to 100, distance from target border to 0.3cm, Start-dose to 100%, end dose to 60% and fall off t0
0.15. I ran this plan a couple of times to get what I wanted.
After my plan optimization and I renormalized each plan (Phase 1 and Phase 2) to have 95 % of
each PTV receiving 100 % of the prescription dose. I then converted the 110% isodose line into
structure and dose painted it to reduce the hotspot. I then formed a composite of Phase 1 and
Phase 2 plans to a total of 5040 cGy.
4
5
DVH of Plan 2
6
Dose Constraint Comparison
Discussion
The dose constraint outcome of the “SupaFirefly” Esophagus technique is slightly superior than
the two-arc technique plan. However, I do not think the difference in the achieved dose
constraints are significant. The “Supafirefly” technique resulted in less integral dose than in the
two-arc technique, hence, producing a better dose sparing to parallel organs like the lungs and
kidneys.
Also, the “Supafirefly” technique produced a total of 2249 MUs compared to a total of 916 MUs
from the two-arc plan. Patients treated with the “Supafirefly” technique will spend twice as much
time on-beam than with the two-arc plan. A longer on-beam time is likely to result in patient
motion during treatment. The actual received dose to organs at risk from treatment could be
7
significantly different from the achieved dose from the plan when there is a significant patient
motion during treatment.
The ‘Supafirefly” technique was helpful and is definitely a choice technique in situations where
VMAT treatment is not available. The “Supafirefly” technique could also produce very
significant difference in dose to the lungs for mid-esophageal cancers.