Sie sind auf Seite 1von 7

Cryogenics 41 (2001) 725–731

www.elsevier.com/locate/cryogenics

Description of removable sample mount apparatus for rapid


thermal conductivity measurements
A.L. Pope, B. Zawilski, T.M. Tritt *

Clemson University, 118 Kinard Laboratory, Clemson, SC 29632-0978, USA


Accepted 31 August 2001

Abstract
Described in this paper is an apparatus in which bulk samples can easily be mounted on a removable puck for thermal con-
ductivity measurements and then placed in the described measurement system. This rapid mounting and measurement system uses a
standard steady-state absolute thermal conductivity measurement but allows for rapid measurement and excellent thermal stability
coupled with the use of a closed cycle refrigerator. The distinction of this system is rapid mounting and measurement of thermal
conductivity over a broad temperature range without sacrificing accuracy and precision in data acquisition. In addition, this system
allows for versatility in its use. The design of this apparatus, measurement specifications, and thermal conductivity data on several
standard materials measured in this system are presented. Ó 2002 Elsevier Science Ltd. All rights reserved.

Keywords: Thermal conductivity; Closed cycle refrigerator; Rapid measurement

1. Introduction conducting samples with approximate dimensions of


2  2  8 mm3 . It was determined that the absolute
The determination of the thermal conductance ðKÞ is steady-state technique would be most advantageous to
a solid-state measurement in which a temperature dif- use for these samples. There are power loss terms ðPLOSS Þ
ference ðDT Þ across a sample is measured due to power involved in the measurement of thermal conductivity
input ðPR ¼ I 2 RÞ into a resistive heater, essentially a utilizing the steady-state technique that must be ac-
measure of the heat flow through the sample. Thermal counted for or minimized. Thus, one must determine the
conductivity ðkÞ is the same measurement with the di- true power through the sample ðP ¼ PIN  PLOSS Þ that is
mensions of the sample taken into account and responsible for the temperature gradient.
k ¼ KL=A, where L is the length between thermocouples,
DT is the temperature difference measured, and A is the
cross-sectional area of the sample through which the
heat P flows. In the thermal conductivity measurement 2. Experimental considerations
 
A Previously we developed a custom designed apparatus
P ¼ kTOT DT ;
L for rapid thermopower and electrical resistivity mea-
surements using dismountable integrated circuit chips
where P is the power input into the sample, kTOT is the
incorporated into a commercial cryocooler [3]. Utiliza-
total thermal conductivity.
tion of the absolute steady-state technique (as used in
Many methods exist for measuring thermal conduc-
the dismountable puck system described here) requires
tivity such as the 3x technique [1], thermal diffusivity,
thorough understanding of the errors or losses involved
and the steady-state (absolute or comparative) tech-
in this technique. A disadvantage to using an absolute
nique [2]. Each technique has its own advantages and
technique is the difficulty in determining exactly how
limitations. Our probe is designed to measure electrical
much heat is being lost; consequently, this loss must be
minimized or calculated to reduce this uncertainty. This
*
Corresponding author. Tel.: +1-864-656-5319; fax: +1-864-656-
may be done through the use of small diameter wires
0805. with low thermal conductance, high vacuum to reduce
E-mail address: ttritt@clemson.edu (T.M. Tritt). convection and gas reduction loss and designs to reduce
0011-2275/02/$ - see front matter Ó 2002 Elsevier Science Ltd. All rights reserved.
PII: S 0 0 1 1 - 2 2 7 5 ( 0 1 ) 0 0 1 4 0 - 0
726 A.L. Pope et al. / Cryogenics 41 (2001) 725–731

radiation losses. For example, in order to minimize essential when designing this system was the ability to
conductive loss a differential thermocouple is utilized so thermally sink the samples to the cold finger. This was
that there will only be two wires available as a heat leak achieved through the use of custom designed compo-
instead of four wires if standard thermocouples were nents at Clemson, coupled with the use of commercial
used. The absolute technique requires fewer thermo- parts (specifically a removable puck and puck recepta-
couples and provides a smaller quantity of places to cle, provided by Quantum Design) and incorporating
make poor thermal contacts than if a comparative these into a commercial cryocooler system. The com-
technique was used. mercial manufactured parts used here are also used in
Radiative loss is also a major concern when using any the Quantum Design physical properties measurement
solid-state technique to measure thermal conductivity. system (PPMS) to thermally sink experiments to the
There are only a few design features that can be incor- cold finger of the PPMS. The sample puck is made of
porated to overcome these losses and corrections to the oxygen-free high-conductivity copper that couples di-
data are typically necessary. Matching of the tempera- rectly to the puck receptacle that houses the thermom-
ture gradient on a radiation shield surrounding the eter. The sample puck is keyed so that it may only be
sample with the temperature gradient of the sample is inserted into the receptacle in the correct orientation [5].
often used as a solution in larger samples. With smaller Another concern alleviated though the use of commer-
samples, several mm in length, this correction shows cial parts was the effect of thermal cycling on the puck/
little reduction of radiative loss. Another method to receptacle junction since this had been adequately tested
combat radiative loss with small samples is to incorpo- previously.
rate a radiation shield that will reflect the heat back on In order to facilitate rapid throughput of samples, the
the sample. This is accomplished through the use of a puck was designed to accommodate two samples, using
highly reflective surface on the inside of the radiation an absolute steady-state technique. The two mounted
shield. Also essential when making measurements is an samples can be measured simultaneously. Samples are
excellent vacuum which will minimize convection and mounted on the dismountable puck and tested for
heat conduction in the system. Convection will lead to readiness (good electrical contact) in a custom designed
heat transfer along the sample not associated with travel test box before they are plugged into the closed cycle
through the sample. A sweep of power verses DT con- refrigerator. A good ‘‘rule of thumb’’ is that excellent
ducted at a stable temperature should be linear. If it is electrical contact as well as mechanical stability is re-
not a straight line, there are losses due to convection flected into excellent thermal contact. Additionally, the
which need to be corrected for or minimized. dismountable puck allows all sample mountings to be
Another issue that must be taken into consideration performed under a microscope and ensures virtually no
when making thermal conductivity measurements is ‘‘down time’’ in the operation of the closed cycle re-
thermal sinking of the sample. In order to make an ac- frigerator.
curate measurement, the sample must be in excellent Components necessary to thermally couple the puck
thermal contact with the cold finger and heater as well as receptacle to the cold finger were machined (see Fig. 1).
with the thermocouple. Kopp and Slack [4] present an A brass piece is screwed into the cold finger of the closed
excellent discussion of thermal contacting problems with
thermocouples. Uncertainties in measurement of sample
dimension can also lead to a 5–10% uncertainty in the
thermal conductivity measurement since the thermal
conductivity depends on the cross-sectional area of the
sample and the length between the thermocouples. With
many systematic errors that must be considered, then a
5–10% uncertainty in the absolute magnitude of the
thermal conductivity is generally considered to consti-
tute a relatively accurate measurement.

3. Description of removable puck system

With the desire to design for a system in which sam-


ples for thermal conductivity measurements could be
easily introduced into the cold cycle refrigerator, several
requirements were considered. The foremost concern Fig. 1. Pieces necessary for assembling the thermal conductivity re-
when designing this system was the reproducibility and ceptacle for the dismountable puck. These pieces are assembled on the
accuracy of the measurement. One requirement that was cold finger of the closed cycle refrigerator.
A.L. Pope et al. / Cryogenics 41 (2001) 725–731 727

cycle refrigerator. This brass piece is essential to slow


down the cooling rate of the sample to prevent thermally
shocking the sample. It also provides better temperature
control for thermal cycling of the system and stabilizing
at a desired temperature. The brass piece acts as a
thermal decoupler, due to its low thermal conductivity
(k  100 W/m K at T  300 K) and provides thermal
inertia between the sample and the copper cold finger
(k  400 W/m K at T  300 K).
A small copper cup was designed and machined and is
screwed into the brass piece. There is a small hole in the
side of the copper cup through which the electrical lead
wires can be inserted into its interior. The lead wires are
thermally sunk to this Cu cup with thermally conductive
epoxy, called Stycastâ . The wires run from the puck
receptacle to the electrical outlet connectors in the
closed cycle refrigerator. This Cu cup is necessary to
electrically connect to the puck receptacle, which has its
electrical lead pins at the bottom of the receptacle. The
cup also acts as an isothermal environment for the wires
leading to the puck receptacle, minimizing any un-
wanted or extraneous thermal voltages.
In order to attain excellent thermal sinking, a copper
ring was machined to fit snugly around the puck re-
ceptacle and securely in the copper cup. The copper ring
is screwed to the puck receptacle. Twisted pair wires are Fig. 2. Assembly drawing of removable thermal conductivity mount-
thermally sunk with grease to the side of the copper cup ing system.
and then attached to the leads on the puck receptacle. A
Si diode is imbedded into the puck receptacle, providing
excellent thermal coupling as well as placement close to sunk into the copper solder pots and subsequently at-
the sample. The ring/receptacle assembly is placed in the tached to the modified removable puck from Quantum
copper cup using thermal grease to enhance thermal Design. Consequently, the sample is in excellent thermal
contacting. The copper cup and copper ring/receptacle contact with the puck, which in turn is thermally ad-
are fastened together with countersunk screws securing joined to the entire system.
the puck receptacle in place. This system provides a The next challenge in showing the viability of this
hardwired mount to plug the puck into without sacri- dismountable mounting system is in attaining accurate
ficing the excellent thermal contact required for reliable and reproducible data by setting up an absolute thermal
thermal conductivity measurements. In addition, a Cu conductivity measurement on the removable puck
radiation shield (can) which has been polished on the (Fig. 3). The measurements were performed on a set of
inside and sputtered with high reflectivity Au is placed available standards. The sample mounting technique is
around the samples and thermally sunk to the base. See taken from Uher [8]. In order to measure thermal con-
Fig. 2 for an assembly drawing. ductivity using the absolute technique a strain gauge is
employed as a resistive heater (120 X resistor). The
strain gauges are used as heaters due to the fact that
4. Mounting configuration their small size allows for better thermal contact with the
sample as well as less surface area for radiative loss. The
The Quantum Design puck is then modified to ac- strain gauges are attached to the sample with a very thin
commodate samples for the thermal conductivity mea- layer of 5 min epoxy providing very good thermal con-
surement. A small copper plate with four screw holes is tact even though the thermal conductivity of the epoxy
attached with Stycastâ to the thin copper film on the is low. The existing leads of the strain gauge (Cu,
removable puck. Small copper blocks with screw holes k  400 W/m K at 300 K) are removed and replaced
to fasten to the copper block on the removable puck are with phosphor-bronze (k  20 W/m K at 300 K) wires
machined with a hole in the center. This hole acts as a due to their low thermal conductivity and consequently
solder pot with which to thermally contact the sample to minimized heat loss. A two-wire resistance measurement
the cold finger. Stycastâ may also be used to thermally is made from the sample to the puck and a four-wire
adhere the sample to the base. Samples are thermally measurement is made from the puck to the measurement
728 A.L. Pope et al. / Cryogenics 41 (2001) 725–731

establishing a vacuum of 106 Torr in order to minimize


convective losses. The sample chamber is then evacuated
with a roughing pump and then a turbo pump (typical
Ω vacuum  106 torr).

5. Data acquisition

Thermal conductivity measurements are made using


state-of-the-art electronics (measurement equipment
described below) and data acquisition software. Using
LabViewâ , a visual programming language, a custom
designed data acquisition program is developed that
automates the data acquisition process. The cross-sec-
tional area of the samples, which have been measured
using an optical microscope or a set of calipers, is en-
tered into this program. The length between the ther-
mocouples is measured from the inside edge of the two
copper flags and entered into the program. Measure-
ment from inside the wires is necessary because the
copper wires are of much higher thermal conductivity
than the samples that are being measured, generally two
orders of magnitude. The copper wire essentially pro-
Fig. 3. Two samples are mounted on a removable puck. Each sample vides a thermal short across its small diameter.
is mounted as pictured above.
The LabViewâ program controls a Lakeshoreâ 340
temperature controller which allows stabilization of
equipment. This two-wire measurement introduces a temperature to 50 mK. In fact, the program requires
small error (<1%) in the power input and thus the that the base temperature remain within 50 mK before
thermal conductivity, which must be and is corrected for data is acquired. Once the temperature is stable, a small
in the calculation. Two #38 AWG (0.004 in.) insulated power or current (depending on the temperature) is in-
copper wire flags are placed parallel to one another and dependently input into each strain gauge, 120 X resis-
perpendicular to the length of the sample. Once the tors, by two Kepcoâ power supplies. The heat is input
Stycastâ is dry, the coating on the tops of the copper into the sample until the heat flow is uniform and stable,
wires are scraped to provide a surface on which the typically 2–3 min. The power and DT are recorded. The
(0.001 in.)Cn–chromel junction of the differential ther- current through the heater is slightly increased, the
mocouple is soldered. This Cu–thermocouple junction temperature difference allowed to equilibrate, and then
provides excellent thermal contact. The copper flags the power and DT are recorded using a Keithley 2001
help not only with thermal contact, but also ensure that multimeter. This sequence is repeated several times, re-
the differential thermocouple is not shorted to the sulting in a power verses DT sweep at a given tempera-
sample resulting in erroneous voltage measurements. ture. This power sweep is then fit to a straight line, the
This specific type of thermocouple was chosen due to its slope being proportional to the thermal conductivity.
low thermal conductance, which will assist in reducing This measurement is repeated at desired increments of
error due to conductive loss. A single differential ther- temperature from 10 to 300 K. Both samples have in-
mocouple is used in this system instead of two thermo- dependent power supplies attached to their respective
couples in an effort to further minimize conductive heaters. A typical measurement sequence for the entire
losses through the wires (2 wires lead away from the temperature range will take 24–48 h to run two samples
sample instead of 4). on this apparatus.
Once the sample has been examined to ensure all the
wires are installed properly and none of the wires are
electrically shorted, the puck is placed in the puck re- 6. Reproducibility and accuracy of data
ceptacle on the cold finger of the closed cycle refriger-
ator and three shields are placed over the sample. The Data is taken in the thermal conductivity system while
inner shield matches the temperature gradient around the apparatus is warming from 10 to 300 K. It is much
the samples in an effort to minimize the radiative loss. easier to thermally stabilize the experiment in the
The middle shield helps minimized radiation loss from warming mode than the cooling mode of the closed cycle
the cold finger. The outermost shield is a vacuum shield, refrigerator. Representative data can be seen in Fig. 4.
A.L. Pope et al. / Cryogenics 41 (2001) 725–731 729

determined, being the difference between kT and


kE ðkL ¼ kT  kE ). The lattice contribution is usually
observed to have a very characteristic shape, in this case
the curve goes as 1=T from 80 to 150 K. Above 150 K,
the curve deviates from this fit. When performing ther-
mal conductivity measurements the sample temperature
is given by ðT þ DT Þ and upon doing a Taylor expansion
4 4
of ½ðTSAM Þ  ðTSURR Þ the radiation loss to first order is
found to be proportional to T 3 DT . Consequently, the
radiation loss is proportional to T 3 as a function of
overall temperature. If a curve of the measured lattice
thermal conductivity kL is plotted in conjunction with
the extrapolated lattice ðkLC ¼ Að1=T Þ, where A ¼ con-
stant) and the difference between the two is called D
(where D ¼ kL  kLC ), it is observed that near room
temperature the difference in the calculated and assumed
(a) lattice thermal conductivity is on the order of 1 W/m K
(Fig. 4(b)). The difference, D, is assumed to be due to
radiation loss proportional to T 3 . If D is plotted verses
T 3 , since there is a temperature dependence of the
thermal conductivity, it is seen (see inset in Fig. 4(b))
that the relationship is linear, indicating that the differ-
ence is indeed due to radiative losses. The total thermal
conductivity corrected for radiative losses can be de-
termined by adding the calculated electronic thermal
conductivity and the corrected lattice thermal conduc-
tivity together. Using this iterative process, corrections
for radiative losses can be achieved.
The true test of any measurement technique’s viability
comes through the accuracy and reproducibility of the
measured data. In Fig. 5, thermal conductivity from
NIST 1461 standard stainless steel is shown with the
corresponding data taken in the system described above.

(b)

Fig. 4. (a) Total measured thermal conductivity with electronic and


lattice contributions shown. The electronic thermal conductivity ðkE )
can be extracted using the Wiedemann–Franz Law. (b) Lattice thermal
conductivity ðkL ) is extracted and subtracted from the total thermal
conductivity ðkTOT Þ. A radiation term ðDÞ can then be determined as
described in the text.

One of the drawbacks to using a standard steady-state


method is that above 150 K radiation loss can become
a serious problem. For larger samples (1–2 sq in.), the
radiation effects are typically negligible below 150 K.
These large samples are usually very difficult to obtain in
research grade samples. Thus, in order to correct for
radiation effects in our smaller samples the following
procedure is employed. Once the thermal conductivity
(kT ) is measured, the Wiedemann–Franz law
ðkE ¼ L0 rT ; where L0 is the Lorentz number, r is the
electrical conductivity and T is the temperature) is used Fig. 5. Thermal conductivity of stainless steel as measured in the
to extract the electronic contribution (kE ) to the thermal closed cycle refrigerator with no corrections compared to NIST 1461
conductivity. From this the lattice contribution (kL ) is Stainless Steel standard data for stainless steel.
730 A.L. Pope et al. / Cryogenics 41 (2001) 725–731

With corrections made for radiative loss and conduction cut from the same boule. These samples were all
along the wires we conclude that our data is accurate to mounted for thermal conductivity measurements, di-
within 6% of the NIST data. However, the temperature mensions were ascertained, and thermal conductivity
dependence is identical. Much of the difference in mag- recorded. In Fig. 6(a), thermal conductivity of the four
nitude is due to uncertainty in the measurement of the samples is shown. The data are all within 10% of one
sample dimensions as well as conductive and convective another. This variability is probably due to dimensional
losses, which may be estimated and corrected for. Low uncertainties as well as possible slight variation in
thermal conductivity samples will give off a smaller sample composition between the samples measured. If
signal, so care must be taken to get a very low noise level the data are normalized (Fig. 6(b)), it is seen that the
from the thermocouples. curves lie on top of one another. The slight deviation at
The thermal conductivity measurement is found to be higher temperatures is most likely due to differences in
very accurate, with essentially the same temperature sample size leading to different amounts of radiation
dependence and magnitude of the stainless steel stan- contributions. This data shows that our system has very
dard. However, precision and reproducibility of the good precision and reproducibility. In addition, our
measurement are also very important. In order to check quasicrystalline data show an excellent agreement with
these issues, four AlCuFe quasicrystalline samples were the values in the literature [6].

7. Conclusions
(a)
In conclusion, a custom designed measurement sys-
tem in which thermal conductivity can be measured
accurately and precisely using a dismountable mounting
system has been developed. This system utilizes an ab-
solute thermal conductivity technique, but has the ad-
vantage that two samples are mounted on a puck that is
easily introduced into a closed cycle refrigerator. This
system offers enhanced throughput of samples and ease
of mounting. One added advantage to this system is its
versatility in the type of measurement to be executed on
the dismountable puck. For example, a system has also
been developed to measure thermal conductivity of very
thin fibers (A  0:0001 to 0:01 mm2 ) using a modifica-
tion of this design and incorporating with an extensive
(b) data acquisition program [7]. This parallel thermal
conductivity (PTC) technique measures a background
thermal conductance of the wire heaters and platform
and then the sample is placed in parallel with the
background and measured again. By subtracting the
background the thermal conductance of the sample may
be determined.

Acknowledgements

We acknowledge Matt Marone for his work on the


first generation of this thermal conductivity system.
Michael Kaeser measured thermal conductivity data for
Fig. 4 in this paper. Thanks are due to Roy Littleton IV
for his help in modifying the experimental setup for
better temperature control. We would like to thank Ian
Fig. 6. Thermal conductivity of four AlCuFe samples cut from the Fisher, Paul Canfield, and Cynthia Jenks at Ames Na-
same boule and measured in the dismountable puck system. (a) As
tional Lab for providing us with the AlCuFe samples
measured data and are the same to within 10% or roughly the error in
measuring the dimensions of the sample. (b) Normalized data are presented in this paper. A special thanks to Ctirad Uher
shown to behave the same with a slight variation at higher temperature for useful ideas and discussion on mounting samples for
due to radiation from the sample. thermal conductivity measurements. We would also like
A.L. Pope et al. / Cryogenics 41 (2001) 725–731 731

to thank Quantum Design for providing the parts and [3] Pope AL, Littleton IV RT, Tritt TM. Apparatus for the rapid
support necessary to design this system. We acknowledge measurement of electrical transport properties for both ‘‘needle-
like’’ and bulk materials. Rev Sci Instrum 2001;72:3129.
support from ONR, ARO, DARPA (ONR No. N00014- [4] Kopp J, Slack GA. Thermal contact problems in low temperature
98-0271, ONR/DARPA No. N00014-98-0444, and thermocouple thermometry. Cryogenics February, 1971.
ARO/DARPA No. DAAG55-97-0-0267) and from re- [5] Physical Properties Measurement System: Hardware and Opera-
search funds provided (TMT) from Clemson University. tion Manual. Quantum Design Corporate Headquarters, 11578
Sorrento Valley Road, San Diego, CA 92121, USA.
[6] Perrot A, Duboise JM, Cassart M, Issi JP. In: Proceeding of the 6th
International Conference on Quasicrystals, 1997. p. 588.
References [7] Zawilski BM, Littleton IV RT, Tritt TM. Rev Sci Instrum
2001;72:1770.
[1] Cahill DG. Rev Sci Instrum 1990;61:802. [8] Uher C. private communication.
[2] Parrott JE, Stuckes AD. Thermal conductivity of solids. London:
Pion; 1975.

Das könnte Ihnen auch gefallen