Data Sheet

POLILYTE Plus ARC


The POLILYTE Plus ARC is an electrochemical pH sensor with integrated electronics, having the full functionality
of a measuring device with self-diagnostics. The sensor is designed for wide range of application with high
demands reproducible and stable measurements over long period of time. The sensor is equipped with a
patented reference system with POLISOLVE Plus electrolyte and double SINGLE PORE diaphragm.

Application Fields:
Chemical Industry, Water applications.

Specifications:
Measuring method: pH sensitive glass electrode
Range: 0 to 14 pH
Zero point (at factory calibration): +/- 20 mV at 25 °C
Sensitivity (at factory calibration): 57 to 59 mV/pH at 25 °C
Measured values can be configured by software according to:
pH: pH, mV
Temperature: °C, K
Sample min conductivity: 2 µS/cm
Operating temperature range:
Analog interface: 0 to 110 °C
Digital interface: 0 to 130 °C
Process pressure range: 0 to 6 bar
Electrolyte: POLISOLVE Plus
Reference system: HAMILTON EVEREF-L
Shaft material: Glass
O-ring material VITON
Membrane glass: HAMILTON type H glass
Diaphragm: SINGLE PORE
Number of diaphragm: 2
Internal solution ground Yes
Shaft diameter: 12 mm
Mounting: PG 13.5 thread
Steam sterilizable: Yes
Autoclavable: Yes
CIP: No
Certificate: Declaration of Quality and materials used, CE conformity
Recommended storage: HAMILTON Storage Solution (p/n: 238931) or 3M KCl
(p/n: 238036 or 238936)
Monitoring of sensor quality: Yes
Automatic standard calibration: Yes
Temperature compensation: Yes
Electrical connector: VP 8.0 socket head
Operating voltage: 7 to 30 VDC max. 150 mW
Analog interface 1 and 2: Galvanically not isolated, pulse width modulation
with 3.5 kHz, recommended PCS settings:
- Use galvanically separated inputs
- Sampling rate < 3 kHz and ≠ n*3.5 kHz
- Average over > 1 s
Analog interface 1: 4–20 mA for pH, programmable
Analog interface 2: 4–20 mA for temperature, programmable
Digital RS485 interface: Modbus RTU protocol, maximal 31 addresses
Baud rate: 4800, 9600, 19200, 38400, 57600,115200 bd
For pricing or any further information, please contact Omni Instruments Ltd.
Ordering Information:
Sensors
Part Number Description a-length
242 111 POLILYTE Plus ARC 120 120 mm
242 112 POLILYTE Plus ARC 225 225 mm
242 113 POLILYTE Plus ARC 325 325 mm
242 114 POLILYTE Plus ARC 425 425 mm

Accessories
Part Number Description
242 411 USB-RS485 Modbus converter
242 180 ARC View Handheld Package
242 170 ARC Wi Sensor Adapter
355 194 Demo Cable (1m, open end, with plug power pack)
355 217 Sensor cable VP8, 1m
355 218 Sensor cable VP8, 3m
355 219 Sensor cable VP8, 5m
355 220 Sensor cable VP8, 10m
355 221 Sensor cable VP8, 15m
355 222 Sensor cable VP8, 20m

Dimensional drawing:

For pricing or any further information, please contact Omni Instruments Ltd.