Beruflich Dokumente
Kultur Dokumente
IDETEC.
ZAC ESPLANADE SUD
650, Av. Saint André de Codols
30 900 Nîmes - France
DECLARATION OF RESPONSIBILITY
The MANUFACTURER is only responsible for the safety of its products when their maintenance, repair
and/or modification has been carried out by the MANUFACTURER or by personnel expressly authorised to
do so by the MANUFACTURER itself.
The manufacturer shall not be held responsible in any way for malfunction, damage and/or danger
due to incorrect use of the system or to disregard of the maintenance Regulations.
The user of the installation where the system is connected is responsible for making sure that the
installation itself is only used by suitably trained and qualified operators.
The system is not used in contact with the patient - only accidental contact with some parts of
the device with the patient is possible and is contact with the operator is foreseen.
Contact with the patient is not of the invasive type.
Contact with the operator is foreseen for reasons strictly linked to the conditions of use
(Operating Manoeuvres).
7
10
6
4
9 1 Pivoting wheel
2 Mechanical support
3 Collimator
4 Handle for positioning monoblock
5 Monoblock
11 6 Key ON
7 User interface display
2 8 Transport handle ''Dead man''
9 Electronic control
10 Handle-lock brake release ragged arm
1 11 Support fortilting
12 12 Lateral wheels
13 N.A.
ON Switch on unit
1.2.3 COLLIMATOR
15
THE MANUFACTURER declines any responsibility for damage to persons and/or things caused by
! improper use of the same!
The user is always responsible for compliance with standards applicable to the installation and
! use of the equipment!
Any shot right after resetting the breaker may indicate bad operation. Immediately contact the
! technical service.
Before carrying out any x-ray exposure, make sure that all the necessary precautions
against radiation have been taken!
During emission of radiation, the personnel present in the x-ray room must follow the instructions in
force on the subject of protection against radiation. For this purpose, bear in mind the following rules:
• where necessary, use protective accessories against radiation in addition to the devices already
provided on the apparatus
• always use the special radio-protective gowns: a radio-protective material equivalent to 0.35 mm of
lead (0.35 mm Pb) attenuates 99.95% of the radiation produced at 50 kV and 94.5% of that
produced at 100 kV
• the best protection against radiation is distance: therefore always keep as far away as possible both
from the radiation source and from the object of the exposure, using the special remote
radiography control with extension cable;
• avoid moving within or staying in the direct beam of radiation
• always use the smallest exposure field possible: the radiation dispersed considerably depends on the
volume of the object radiated
• Use always the smallest field of exposure possible, appropriately closing the
collimator diaphragms. The radiation dispersed does, in fact, depend greatly on the
volume of the object radiated.
In no case must the safety circuits provided against accidental exposures be modified or
! de-activated.
The MOBILE UNIT has been designed and constructed to avoid turning over up to an inclination
of 10° from horizontal in its transport position.
It is therefore advisable:
not to remain, move or place the Apparatus on surfaces with an inclination of more than
10°.
not to try to move the Apparatus without working on the special handle for “releasing the
brakes”;
pay attention, during movement of the UNIT, to any obstacles on the floor (cables, steps
and differences in level of any kind)
For the Residual Risks due to any defect in the protection measures used, remember that:
Protection against electric shocks is made by connecting the metal parts of the apparatus to
earth: it is therefore necessary to periodically check correct operation of the whole earthing circuit
- according to the ORDINARY MAINTENANCE PROGRAMME described in SECTION 4 of this manual.
! X-RAY MONOBLOCK: this overload could result in loss, by the MONOBLOCK itself, of the
insulating medium at very high temperatures.
During movements of the apparatus, take care that the parts do not hit the patient or the
! operator.
Label of the
PHYSIOLOGICAL E FFECTS
unit Control
pane
Pushbutton
RADIOGRAPHY
Radiography
connector
1 BEEP The X-ray pushbutton has been released before completion of the exposure - the
exposure has therefore been interrupted
WARNING!
All types of transmission with mobile radio apparatus must be avoided.
Mobile telephones must be turned off in rooms near the apparatus.
These rules must be applied when the unit is active.
The use of accessories, transducers and cables other than those specified may result in
! increased emission or decreased immunity of the equipment
2.7.2 PERFORMANCE
The essential performances of the SD 30 T with regard to EMC are:
High quality digital X-ray images
Software robustness
Robustness of the communication link with the system
No image lost
Artefacts on the image can occur but without any risk of wrong diagnostic
The mobile unit SD 30 T needs special precautions regarding EMC and needs to be installed
! and put into service according to the EMC information provided below
The model SD 30 T is intended for use in the electromagnetic environment specified below. The customer or the user of
the SD 30 T should assure that is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
The model SD 30 T uses RF energy only for its
RF Emission CISPR 11 Group 1 internal function. Therefore, its RF emission are
very low and are not likely to cause any
interference in nearly electronic equipments.
RF Emission CISPR 11 Class A The SD 30 T is suitable for use in all
Harmonic emissions IEC 61000-3-2 Not applicable establishments other than domestic and those
directly connected to the public low-voltage
Voltage fluctuations/flicker
Not applicable power supply network that supplies buildings
emissions IEC 61000-3-3 used for domestic purposes.
Indicates that the product has the requirements requested by new directives introduced for the
environmental protection (2002/95/EC, 2002/96/EC, 2003/108/EC) and it must be disposed properly once its
life cycle is ended.
When the device has reached the end of its life it must be disposed at the proper centres for the separate
collection of electrical and electrical wastes, or it must be returned to the reseller or to the manufacturer in case
you want to replace the product with another equivalent new one.
The proper separate collection helps to avoid possible negative effect on the environment and on health
and it facilitates the second use and the re-cycle of materials of which the device is composed of. Ask
for further information to the local authorities about the areas dedicated to the wastes disposal.
Who does not dispose the product following what here above mentioned will be responsible for the enforced
rules.
The Mobile Unit is Braked and cannot Move unless the “dead man Brake” is Released (8)
!
3.2 POSITIONING
Do not try to move the system when the brakes are on.
For movements, use the special handles.
To connect the apparatus to the power supply network, only use 16A sockets with earth
terminal.
To turn the Unit on, proceed as indicated in Paragraph 3.3.
The Monoblock / Collimator group can rotate in all directions, as shown in the figures below. To position
it, use the MONOBLOCK POSITIONING HANDLE (4).
The inclination of the X-RAY UNIT is indicated by the GONIOMETER placed in the front (F)
3.3 TURNING ON
To make the system operational:
• Connect the Unit to the power supply mains
• Turn on the unit using the key and the relative pushbutton present on the unit(6)
• The initial automatic TEST of the unit will start immediately
• Wait for the completely start of the software
! Wait for the complete start of the system before carrying out any other operation!
To connect the apparatus to the power supply mains, only use 16A sockets with earth
! terminal
Do not continually turn the unit on
NOTE: Steps below only for reference. Steps may be changed according to your system
configuration.
Steps:
1. Before power up the host system, please powers up the X-ray generator, detector
and all the other x-ray system components needed.
2. Make sure that the peripheral X-ray system components are on and working
properly.
3. Power up the Host PC and login the PC’s operating system
The DROC HOST PC should be powered down once a day to ensure a complete
system reset. IF NOT, system performance deteriorates.
Once the X-ray generator is powered on, perform the following checks:
Remove any patient supports or other objects so that they do not interfere with the
movement of the X-ray tube's suspension arm.
Switch on the collimator’s light to ensure that the area to be exposed to the X-ray
beam is illuminated
Page 27 on 124 MT 9B 50 002 Ø
Check the releases, if applicable, on the X-ray tube's suspension arm, the Bucky, and the
table to ensure that the controls are operating properly.
Power up the HOST PC and Log in to the PC's operating system;; if the monitor is not on,
press the power button on the monitor.
NOTE: Before you power on the DROC HOST PC, make sure the detector have completed their
startup sequence (wait at least several minutes after you power on the UPS).
Start up the DROC application by double clicking the DROC shortcut on desktop. The login
window shows as below:
To sign into the application, choose your level from Level Icon (Technician, Application,
Manager) and your name from the User Name drop-down list, enter your password, and
click button.
NOTE: In case you CAN NOT see this window, and a message box “No Dongle available or
Dongle is invalid! DROC System will exit!” shows up instead. Check your USB
license key to make sure it is well installed or contact our engineers for further support.
The initialization window will show the initialization status of the DROC, shown as below:
There are some initialization tasks of DROC system; the result of each task (successful or
failed) will be displayed one after another.
The general workflow to acquire and send images to output devices involves the following
steps.
Steps:
2. Taking an exposure;
3. Viewing and accepting the image; Optionally, you can reject the image and re-capture
till an acceptable image is acquired;
4. After all images captured, the data will be transmitted to the default node.
Alternatively, users can send the images to other node or print the images;
The following subsections provide specific instructions for each of those steps.
Before capturing images, you must identify the patient with whom they are to be
associated. Proper identification ensures that the image data is associated with the correct
patient information.
Depending on the system configuration at your site, there are three ways to identify the
patient from whom you want to acquire images, as described below.
All systems allow you to add/identify patients manually. Whether you are to acquire
images of a new patient or an existing one, you can add/identify manually by:
If your system supports HIS Worklist queries, you can retrieve patient information from a
Radiology Information System (RIS). Refer to the following topic “Query and Refresh” for more
information.
You can capture images for the patient/study of SCHEDULED or IN PROGRESS status; even can
super-add images to a COMPLETED status patient/study as long as it still existing in the Local
database.
NOTE: As soon as the patient has been accomplished, it will be stored as PPACS data.
Meanwhile, the completed studies will still being stored in Worklist table for a certain
period after exposure, it depends on the Space Reclamation Strategy and individual
configuration. Users can pick up the patient still in the Worklist table and proceed on
capturing images. A Study existing in the PPACS database could not being superadded
images any more.
NOTE: Make sure you enter patient names and IDs correctly. Incorrect entries can cause
duplicate patient records for the same patient. As a result it will be more difficult to find
the patient in future search operations.
To add a new patient without a barcode, fill the information in New Patient page on
Worklist window.
The New Patient page is displayed below. Please note that, for your site, the fields in this
window may be different.
Page 32 on 124 MT 9B 50 002 Ø
Figure 5: New Patient Page
In this page, specify the patient information and determine the examination that patient is going
to take.
Fields labeled with an asterisk (*) are required to be filled in. You can switch to the next
field by pressing Tab key. After birthday entered, age is calculated and showed by system
automatically. The name will be displayed in the format LAST ^MIDDLE ^FIRST. If you
don’t want “^“ character to appear in name, you can input all content of name in Last
Name field.
Click Exam button. The patient information will be loaded on the Exam window for
technicians to capture images.
Click Save To Local button. The patient information save to local worklist database and
system is still in Worklist window.
When specifying the protocol, click Add Protocol button, the Add Protocol window
shows like below.
Select a Body Part from the left list, the corresponding protocol will be listed in the
upper right area. If select “Often Used” item in the left list, only protocols marked
“often used” display in the upper right area (The often used protocol is defined by
DROC configuration).
Click the protocol to load the corresponding views into the lower right area.
To delete the view from the lower right area, click Delete Selected View
2. Click Edit button, patient fundamental information editing page displays with the
original information
From the Query Panel in the Local/Worklist page, you can query a patient/Study by various
searching criteria such as Patient ID, Patient Name, Accession Number and date. The Query
Input the key information associated to the desired patient/Study, and then click Query
button. DROC application will search the local database, and then the result will be displayed
If your site is connecting to RIS, DROC can retrieve the worklist in background and
automatically refresh the Worklist page. You can override this operation manually by clicking
Select the patient/Study that you want to delete from the Local/Worklist page; then click
NOTE: Only the local cached worklist item can be removed by this deleting operation. If
retrieved from RIS, it may appear again.
+ left click mouse; Click the Exam button in Next Step area; Studies relating to the desired
NOTE: The UI in your site may not show the some buttons after its installation due to the
customization. Please refer to your support engineer for more details.
You can select a view by highlighting the view icon, and the Study Description will change as
At the time the system is installed, default exposure settings and capturing sequences are
configured according to every specific view. You can use manual mode or photo timing mode
Technically, you could simply use the default parameters. They should only be changed for
If you need to adjust the exposure parameters or capturing sequence, refer to Section C
For one patient, you may want to open additional procedures, to include all images in the
same Study. For more information, refer to “Working with Multiple Open Procedures” for
more information.
Page 44 on 124 MT 9B 50 002 Ø
Adding a View to a Procedure (Optional)
If you need to capture extra view image, you could add views to current procedure. Refer to
Taking an Exposure
While taking an exposure behind the leaded wall, you must maintain
verbal communication with the patient and keep the patient within your
sight at all times.
If the X-ray tube heat index exceeds the maximum allowable limit, the X-
ray tube and generator will be disabled until the heat index drop to the
safe level. If this condition occurs, a message will be displayed and an
audible beep will sound.
Steps:
1. Check the detector state in Detector Status area. It shows three colors corresponding to
three detector states: green indicates “ready to exposure”, yellow indicates “preparing
for exposure”, red indicates “on exposure”. If the indicator turns into yellow, please
wait until it turns to green.
2. Optionally, to get higher quality image, you have to give the patient instructions, such as
to hold breath or do not move.
When the exposure is completed, the Preview window pops up and the raw captured image
displays in it. It will close automatically after several seconds. For more details of the other
NOTES: Releasing the Prep/Expose button on the hand-switch before exposure beep will
cease the exposure process without any problem.
If there is any error occurs during the exposure, the Message Field under the
Exposure Status button will display the warning or error message.
If it is just a warning, the message disappears automatically after several
seconds.
If it is not an auto recoverable error, the message will remain. Click the Reset
button before performing subsequent exposure.
NOTE: The UI in at your site may not show the Reset button after its installation. Please refer to
your support engineer on how to enable this button.
Choose one post-process algorithm from ESA and Symphony (if available).
printer.
Add markers.
From the ESA/IT panel, you can change the current ESA/Symphony TM setting by:
Clicking on the drop-down list in the ESA panel and choosing a different setting.
Pressing and holding the right button, and moving the mouse on image preview area,
the LUT is changed. Vertical movement changes window width of LUT, and horizontal
Adjusting the image’s LUT settings by placing the pointer on the three rectangle handles
of the curve, pressing and holding the left mouse button, and dragging the handle.
ESA setting changes or LUT adjustments are typically not needed. For more information
about making LUT adjustments, refer to the topic “Changing the ESA Setting or Adjusting
Cropping an Image
You can crop an image in order to cut the background and only save the area with useful
information. The cropping applies only to the image when it is sent to a DICOM printer or a
storage device. In DROC Configure page, the default crop size is set for each procedure.
When previewing image, the initial crop rectangular is positioned according to this size. If
DROC is capable of controlling the collimator’s movement, the collimator will receive the
position parameters from DROC and sets its field to the specified size before image taking;
position of collimator adjustment manually, the size of crop rectangle will be updated
You can add text comments to an image at the time of acquisition; the comment will be sent to
the DICOM devices which are chosen as output destination. There are already three
For more information, refer to the sub-topic “Adding Comments to an Image” of topic
For more information, refer to the sub-topic “Selecting ESA or SymphonyTM” of topic “Working
You can add markers to an image which allows further information about the image (for
example, supine, upright, and so on), and the markers will be sent to the DICOM devices
which are chosen as an output destination. In DROC Configure page, you can configure the
default text and its position in the raw image for each view. For more information, refer to the
Section C.
You can rotate or mirror the image. For more information, refer to the sub-topic “Rotating or
Full Size” under the topic “Working with Acquired Images” in Section C.
To accept the image, click the Accept button. The Process window closes and a thumbnail
(minimized view) of the image will be displayed in Exam window, as shown below. The
applied exposure parameters will also be found under its thumbnail image.
Once accept the image, that image is stored in DROC and is queued to be sent to default
output devices (Automatic sending function depends on your configuration). Now the image
By default, the next view to be exposed will be selected in image area automatically. If
needed, you can select a different view and adjust the exposure settings before acquiring
Steps:
The Image Rejection Reason list displays, as shown below. In this list, you can select a
reason of the rejection.
2. Check the corresponding checkboxes to choose one or multiple reasons from the
list of default comments. If you check the Other Reason checkbox, you need to type
the reason in field under the checkbox.
3. Click the Reject button to save the rejection reason and reject the image.
Then the Rejection Reason Window and Process window will close then you can
recapture the image.
To dismiss the window and return to Process window without rejecting the image,
click the Cancel button.
Rejected images are stored in a designated reject bin storage area in DROC, and can be
discarded or resent to an output device.
NOTE: Rejected images can be manually deleted by the system administrator. And the system
will delete rejected images when it reclaims disk space.
After captured, the images will be sent to default destinations or printers automatically. But
you can still re-send the images to different destinations or resend images from Output
window.
There are two types of output destinations: archive server for storage and printer for
printing. These two destinations (may contain zero or multiple nodes) can be configured for
all images belong to the same procedure. Furthermore, the film orientation, size and format
can be configured for each different procedure. When an exam is opened, these
configurations are loaded automatically. You can also change them manually for current
Study.
For more information, please refer ''Managing the Images": Working with Worklist Studies.
You can quit the Exam window to perform other operations, no matter whether all the
To quit the Exam window, click the button in Next Step area.
If all of the images in the current study has being acquired, the status of this Study will
You can pick up the incomplete Study from the Worklist window to proceed with the
To log off the DROC application, click the Exit button from the Menu Bar. The DROC Close
dialog displays.
Exit:Close DROC
Typically, you should leave the system components on. The DROC HOST PC, monitor, and
Detector are able to enter an energy-saving "sleep" mode1 when they are not used for a
period of time.
To power off the system, it is recommended that you perform the power off tasks in the
following sequence,
Steps:
Firstly close all the applications running in PC; then from the Start Menu, choose
“Turn off the computer”→“Turn off”.
Powering off the X-ray generator also typically DOES power off the other X-ray
system components, such as X-ray tube, tube suspension arm, collimator, Bucky,
and table (if applicable).
1
The “Sleep” mode is available for Trixell Detector.
When the study(ies) opens in the EXAM interface, you can perform following procedures.
When the View is added successfully, it will be indicated in the Image Area of Exam window.
To Delete a View/Image
If you like to delete the reduplicate/unsuitable view, click the “Delete View” button which
is located in the upper right corner of Exam window.
window
When there are many view(s) or/and captured image(s) which cannot be displayed in one
page, click the “Page Up” or “Page Down” button to see the view or image on other page.
Image area displays the views to be acquired and thumbnails (minimized views) of images
already acquired, as shown below.
By default, the views to be acquired are arranged sequentially from left to right as images
Before capturing image, you can adjust the exposure technique parameters (Detail is
introduced in next topic). The modified settings will be saved automatically.
You can quit Exam window by clicking other available items (Worklist or Output) on the
Menu bar. For study with all the images being captured, the quit operation will change the
status of this study into COMPLETED, and return to Worklist window. The completed study
will exit in the worklist page for a certain period, and you can add view to the study by
retrieving the study from worklist page.For study with not all the images being captured, the
Studies’ status confirmation dialog like below figure will pop up.
Choose Suspend button, the closing operation will change the status of unfinished
study into IN PROGRESS, and return to Work window.
Page 61 on 124 MT 9B 50 002 Ø
Choose Finish button, the closing operation will change the status of unfinished
study into COMPLETED, and return to Work window.
Choose Cancel button, stay in Exam window.
Whatever the study’s status is, you can proceed to taking exposures by retrieving the
study from Worklist page.
Adjust exposure settings using AEC (Automatic Exposure Control); this includes selecting
one or more AEC fields, and adjusting default density settings.
To select a body type, click on the icon for the required type—Large, Medium, Small, and
Pediatric. The selected button will be pushed down
Adjusting kV
To adjust the kV value, click on the Increase KV button and Decrease KV button located to
the right of the kV value field. The kV value increases and decreases in steps of 1 kV.
Pressing and holding the button causes displayed kV value to increase or decrease
automatically until the button is released. If the kV value reaches maximum or minimum,
the value will remain unchanged.
The Automatic Exposure Control (AEC) system monitors X-radiation dosage which is passing
through ion chambers and automatically terminates the exposure after the X-radiation
dosage reaches a predetermined amount of radiation.
NOTE: The default density (0) is calibrated for normal contrast and density levels. Larger,
heavily muscled patients will require an increase in density. Smaller, lightly muscled
patients will require a decrease in density. However, only use this density adjustment if
the image is grainy. If an image is too dark or light, adjust the ESA settings (refer to the
Typically you use the AEC and detector field default settings. You can override them under
special circumstances (for example, the patient has only one lung).
NOTE: When using the AEC function, the patient must be positioned over the enabled AEC
field and the radiation field must be larger than the AEC field. In cases the preferred
radiation field is smaller than the AEC field, manual techniques are recommended.
You can select AEC detector fields in each ion chamber individually or in any combination.
You can vary the density setting between +4 and -4 (0 is neutral).
To change the density setting, click on the Increase Density button or Decrease Density button
located to the right of Density value field, as shown above. For the increases or decreases
amplitude of each time, please refer to the Service Manual of Generator.
With the mAs button selected, the TIME controls are dimmed. Since the TIME settings and
MAS setting are dependent to each other.
You can manually specify exposure timing by setting the mA and the ms value. When using the
TIME settings (shown below), the exposure is terminated when the specified time value is
reached. This mode is used when a specific time must be achieved, such as in a
breathing study or for stop motion. Regarding the reciprocal grid’s requirement on exposure
window, the ms value should be guaranteed above 50 when the system employs reciprocal
grid.
To specify the exposure by Time setting, make sure the Time button in the Exposure Control
panel is selected (shown below), then click on the Increase mA/ms button or Decrease
mA/ms button to set the desired values. The value increases and decreases in steps of one unit.
Pressing and holding the button causes the value to increase or decrease automatically until the
button is released. If the value reaches maximum or minimum, the value will remain unchanged.
With the Time button selected, the mAs controls are dimmed. Since the TIME settings and
mAs setting are dependent to each other. Changing mA/ms values causes the mAs value
changes.
To select the focal spot size, click on the corresponding size button. The selected button will be
pushed down.
Page 66 on 124 MT 9B 50 002 Ø
Figure 17: Focal Setting
NOTE: If the selected Focal Spot Size is not compatible with the tube and focal spot, the
system inhibits exposure and displays an overload condition. An overload condition
may exist if the technique selected exceeds the maximum or is below the minimum
allowable X-ray tube focal spot rating. The tube parameters are consistent with the
X-ray tube manufacturer's rating charts.
Hardware Status
The status of hardware is displayed in the Common Message area, as shown below.
Reset Hardware
If there has an error and it cannot auto recover, the error message will remain in message field. Click
the Reset button before performing subsequent exposure.
After taking an exposure, the acquired image could be loaded into the Process Window for
processing.
In Process window, you can see the patient/study information on the top left of the image,
and the real exposure dose information on the top right of the image.
These markers can be used to designate left and right, the position of the patient.
NOTE: It is up to four markers and/or comments that could be added on the image.
By clicking Define Mark button, you can add a customized label which will be displayed on
the image.
Clicking Clear All Marks button, all of added marks and comments are removed from the
image.
Click the Original Size button; you can view the image in full size by holding the left mouse key
pressed and moving mouse to desired direction. In Original Size mode you can only view the
image, cannot permitted to clip, rotate and do other operations. To enable such operations,
you must return to Fit Window mode by click Fit Window button
The image can be rotated 90° (clockwise) or 90° (anti-clockwise) clockwise as required. It is
recommended that, if the image is rotated, please use markers to designate either left or
right rotation.
Clip an Image
When an image has been captured and is displayed in the Process window, you can clip the
image for output to a DICOM print or storage device.
When you accept the image, the full image along with any clipping information, is stored in
the local disk and can be recalled for clipping again within a period of time. Meanwhile, the
clipped image is sent to the DICOM output device.
To remove the clip line and not clip the image: click the Remove Clip button.
To Rotate the image by Rotate Any Angle button, make sure the Rotate Any Angle button is
clicked (shown below), then click the Increase button or Decrease button to set the
desired values. The value increases and decreases in steps of one unit. Pressing and holding
the button causes the value to increase or decrease automatically until the button is released.
If the value reaches maximum or minimum, the value will remain unchanged.
If you like to see only part of the image, you can add the mask on the image. Click the
Add/Delete Mask button, click several points around your interested region and double
click the end point. The system will add mask on the image except the setting region, like
following figure.
If you like to see the inverse image, click the Invert button.
The DROC supports both LUT algorithms to adjust the image contrast,and Symphony
algorithms to enhance the image based on the anatomy selected. Click the corresponding
button to use the algorithms.
Click the LUT, Symphony button or Custom Symphony, system would use the default parameters
to process the image.
When previewing an image, you can modify the LUT settings of the image. The LUT
controls the minimum and maximum densities used in the printed or viewed image.
Page 75 on 124 MT 9B 50 002 Ø
When the system is installed, customized default LUT settings are configured for each
procedure type. The image’s original LUT is created by applying the procedure’s specified
LUT settings to the image data.
Typically, you need to change the LUT only for patients with health conditions that may
affect the image quality, such as emphysema for chest views.
To modify an image’s LUT: drag one of the three square handles of the graph line until the
image shows the way you want. If the top or the bottom one is moved out of display are
and is invisible, one the middle one can be dragged. You can also press and hold the right
mouse button, and move the mouse on image preview area to change the LUT. Vertical
movement changes window width of LUT, horizontal movement changes window level of
LUT.
Besides viewing image at the time of taking exposures, you can retrieve any captured
images from the Work window for sending, reprinting, reviewing or doing other operations.
You can resend images to the output destinations (for example, film printers or display
workstations) that are originally selected at the time of initialization, or to other output
destinations. For more information, refer to the following topic “Resending Images to
Output Devices”
NOTE: Images are not stored infinitely. To ensure that there is always sufficient room to store
newly acquired images, the system deletes the oldest images as needed to reclaim
storage space. However, you can protect specific patient images so that they will not be
deleted. Refer to the following topic “Protecting a Study along with the Associated
Images”.
Patient images are stored in the DROC PC temporarily, until storage space begins to get low.
The system deletes the oldest patient records and images to reclaim storage space for new
one.
When the system begins to do space reclamation depends on the configuration of PPACS. You
can set a minimum threshold for local hard disk space. When the space of the directory
for image storage is lower than this threshold value, the system begins reclamation. The
threshold value is recommended to be set as the maximum image volume of all patients
in one day at least. For example, the maximum patient flow in one day is 100.For each patient
creates 2 images on average, one image’s size is 16M, so the threshold should be 16*100 *2 =
3200M or higher.
You can protect patient related images from being deleted during this reclamation process. You
might need to do this if you may want to resend images to output devices in future.
In Work window, the items in the list can be arranged by clicking the criteria which shown in
the status bar, such as Patient Name, Patient ID, Sex, Birth Date, Age, Study Desp, Study
Date, and Accession Num etc.
You can edit a patient’s fundamental information even after capturing images. The status of the
patient could be SCHEDULED, IN PROGRESS or COMPLETED and also listed in Local/Worklist
page.
Double-click an item from the Local/Worklist page to load the study on View/Output window. Or
Ctrl+click to select multiple studies, which belong to same patient, then click View menu on
menu bar. All associated images would display on the View/Output window.
There are four groups of operation in the View page. We can perform the following functions
as:
Select image layouts.
Move or magnify the image.
Rotate or mirror the image.
Change the image to original size or fit window.
Invert the image.
Process the image.
View the image information.
Restore the image.
Output the image via send.
Page 81 on 124 MT 9B 50 002 Ø
Now we will introduce these four parts respectively:
View operations
When you zoom in image, it may become larger than the image area. Panning (moving the
image within the image area) then becomes important. Click the Pan button, then click and
drag the left mouse button within the image area to move the image. Release the mouse button to
drop the image in its new position.
Click the Zoom button, then press and drag the left mouse button up or down to zoom in
or zoom out for the image. Dragging up increases the image zooms and dragging down
decreases it.
Press the Magnifier button and then click on the interested region which you want to magnify,the
magnifying window will appear on the interested region.
The image can be rotated 90° (clockwise) or 90° (anti-clockwise) clockwise as required. It is
recommended that, if the image is rotated, please use markers to designate either left or
right rotation.
The image can be mirrored by clicking Flip Horizontally button or Flip Verticallybutton.
If you like to see the inverse image, click the Invert button.
If you would like to adjust the brightness & contrast of the image, you can do the operation
by clicking the adjust button as shown below.
You can browse the DICOM information stored with the image file by clicking DICOM
Information button.
Output operations
Sending image
You can transmit the images to one or more specified Send Image nodes. In addition to the
images, the patient fundamental information, comments, and text annotations will be sent as
well.
There are 3 ways to transmit image, Send All Image(s), Send One Specified Image and Send
Multi Specified Images.
Page 85 on 124 MT 9B 50 002 Ø
Send All Image(s):
Load one or more images into the film sheet of View page
Click Send All Image(s) button to send all images in the film sheet
Send One Image:
Load one image into the film sheet of View page
Click Reverse Select Hot Image button to select the image or highlight the image
Click Send Selected Image(s) button to send the only & specified image to the
selected Send Image node
Repeat this step to click Reverse Select Hot Image button to select multi-
image or highlight multi-image
Click Send Selected Image(s) button to send the specified images to the selected
Send Image node
To process an image
On View/Output window, you can switch to Process window by clicking Process menu item
on Menu Bar. On Process window, you can perform image operating functions such as
cropping, rotating, mirroring, adding markers, just as the same as what you do when you
viewing an image at the time of taking exposures. Please refer to the topic “Additional
Operations In Process window” in "Additional Operations": to perform the operations.
When the available storage space drops below a predefined level, local record will be
deleted on a first-in-first-out strategy. When images of a patient are protected, the images
will not be deleted during reclamation.
NOTE: If any image of a patient is in the send queue, all images of this patient will not be reclaimed.
Images are deleted on a study basis: once acquired images in a study have been sent to
their output devices and the study is closed, the images in that study are eligible for deletion
(unless the study is protected. Please refer to the following topic “Protecting a Study along
with the Associated Images”). This is true even if the study is in incomplete status.
1. From the Local/Worklist page on Work window, assign a study you want to protect;
2. Click Protect button located on the Patient panel, then the status of “unlocked” status
study is changed to “locked”, otherwise, the status of “locked” status study is changed
to “unlocked”.
Locked status
Unlocked status
This function is used to manually clear a study along with the associated images.
2. Click Delete button located in the Patient panel, the confirmation window displays as
showed below:
From System Menu, you can perform advanced operation concerning account
management, emergency setting, statistic and device control etc. Although these functions
are out of the examining workflows, they do have relatively great effect upon the imaging
quality and the flexibility, accessibility of the entire system.
From System menu (located in the Menu Bar), click User Management menu to open the
User Management window, shown below. When you login with the manager privilege account,
all users are listed. Otherwise, only your own account is listed.
In this window, you can create a new account, edit the existing user or delete an existing user.
1. Click New button located in the bottom of the window. The New User window is
showed as below;
2. Input the new user’s information and choose a user level you want to create
3. Click OK button to accept the new account, and the New User window closes. Then the
new account information displays in the User list on the User Management window.
Click Clear button to empty all of new user information, and reenter the information in
New User window.
In the User Management window, choose the user you want to delete and click Delete
button.
NOTE: Passwords are case sensitive. That asterisk (*) characters displays for each entered
character to protect the privacy of the password. That requires verifying that you are
typing the new password correctly, since you cannot see what characters you are
typing.
Click Reset button to empty the fields and then reentered information.Click
From the System menu, you can obtain the access to Emergency Setting. The
Emergency Setting window is showed below:
To choose/create a prefix
For emergency registration, the patient’s default name is prefix + date + sequence
number. This prefix can indicate it is a patient of emergency or the patient’s corporation
for physical examination. You can choose, add or delete the prefix.
To add a prefix: input text in the field under prefix label, and then click New button, this
prefix is added to the Patient ID Prefix list.
To delete a prefix: select one item from Patient ID Prefix list and then click the Delete
button, you will see this prefix is deleted.
To set the default procedure Choose the default procedure on the lower half of Emergency
Setting window.
On current page, the users’ name and the respective amount of accepted images and rejected
images are listed.
In the Rejected Detail page on the Statistic window, set the query criteria including date range
and Body Part/Reject Reason, then click Query button. The information about the rejected
images matching the criteria will be listed.
In the X-Ray Book page on the Statistic window, set the query criteria, then click
Query button. The x-ray information of matching images will be listed.
Click Export button to save the current statistic information page as a excel file to PC.
NOTE: Calibration must be performed at least half an hour after system startup or power up
from the sleep mode.
The panel displays instructions for how to set the technique settings and take
exposures for the calibration. Click the Begin Calibration button, to start calibration
sequence.
4. Change the technique settings, follow the recommended value in your calibration
panel.
NOTES: You should reject an image if it exhibits any edge cut-off due to collimation or
misalignment or if there is any artifact from debris or obstructions. Due to the
diverse characteristic of detector, it is necessary for your system to acquire
calibration images in different doses: high, medium, low. And for each dose,
accumulate several images to gain an approximately normal value. You should
follow the instruction strictly during the entire calibration procedure.
5. Take an exposure. The Calibration window with the flat field image displays.
6. Compare the average pixel value of the image with the recommended one, click
the Accept button if they are close, and otherwise click the Reject button.
If you accept the image, calibration runs. The calibration panel displays the instruction
for next exposure, Go to step 4.
If you reject the image, the calibration panel displays. Correct the problem and return to
step 5.
8. Once you finish the calibration sequence, a dialogue box will prompt you whether to
accept this calibration sequence, Click Save Calibration button to save this
calibration result.
If you want to discard the accumulated calibrations, click Abort Calibration button.
Click Close button on the right bottom of the window to quit the calibration.
Station Information
You can see the manufacture name and software version number in Station Information
message window. You can choose the System Menu located in the Menu Bar.
DROC offers a flexible and versatile system for printing image on film or paper, includes
printing capabilities that go well beyond those of most other Windows applications. This
chapter covers how you can print series and studies in Print/Output window using a wide
range of layouts, how you can copy and paste images in the same film or between films, as
well as how to create your own layouts for printing.
Printing is done in the background and will no tie up your system, so you can continue
viewing and reporting while your print jobs are running.
DROC lets you combine various images on “virtual film sheets” which you can then print out.
1. Select study on Work window, click Print button in Next Step area.
A standard film layout was defined during configuration of your system. This layout contains
all the settings required for filming.
All the necessary settings have therefore already been made for each new job. If you are
happy with these settings, you can start filming images without any further intervention.
If these presetting are not suitable, you can change the following settings:
Some of these settings apply to all sheets of current job; others may apply to current
sheet only. For the next job, the system returns to the standard layout.
Image index on the left of Print/Output window, shows all image’s thumbnail views. You
could drag the image from the image index into the viewpoint of the current film sheet.
After entering Print tab, the first film sheet of the current print job is displayed.
With the tab in the top of sheets, you can page through all the film sheets of the current print
job.
If the last film sheet of your job is already full and you want to copy or move images to the end
of your film job, you must first append a new film sheet.
Click the Add Film button . A new (empty) film sheet is appended to the end of the job.
Click the Delete Film button to delete current film sheet. It would delete all images
on current film sheet as well. But this operation will not delete the image thumbnail index.
You can switch current film sheet’s orientation between portrait and landscape.
If you do not like to expose or print the current job with the default film size, you could select
film size from list which can be triggered by the Define Film Size Button.
By setting the layout of the film sheet, you can define how many images per film sheet. For
example, if you like to see diagnostic details on the images, switch to a larger format with
fewer images per film sheet.
The default layout of the film sheet is 1 image per film sheet. If needs you can click the
related layout button to adjust the layout of the film sheet, e.g. Define Film Layout button
means 1-16 images per film sheet with portrait/ landscape mode, Define Child
Layout button means divide one cell to 1-16 parts, and Cancel Child Layout
button means cancel the enabled child layout . Define Special Layout button
means special image layout for print can be enabled according to the custom
needs.
If you want to print an image, you have to add the image into the film by clicking the Add
You could also drag the image from the image index into the viewpoint of the current film
sheet to add the image into the film.
Page 105 on 124 MT 9B 50 002 Ø
Remove the image
If you don’t like to print an image anymore, click on the image and then click the Delete
Image button . The image will disappear from the film. The DEL key on the keyboard is
the same as The Delete Image button.
Move images
You can move images within a film sheet.
First define mouse action by clicking the Drag drop button . Then drag & drop within the
same sheet.
Film repack
Before you send the film sheets to a camera/printer, if there are some empty viewports on the
film, you should reorganize the film sheets to make better use of the film material.
Click the Film Repack button to repack all film sheets. The empty sheets are
removed if necessary.
Click the Same Zoom Ratio button . All images in the current sheet will be magnified to
same factor as the current focus.
NOTE: The mark is always printed out on film. Only the annotation can be hidden.
If you do not like to expose or print your current job with the default camera/printer, you could select
another camera/printer for current job. Select a camera/printer from Print Node list. This selection list
contains all the cameras and printers connected to your system.
DICOM print
Click the DICOM Print button , the film(s) with image(s) will be transmitted to the selected film
printer node.
Windows print
Click the Windows Print button , the film(s) with image(s) will be transmitted to the selected
18
17
16
14
15
l Following a long period of inactivity (3 or more months), it is very important to proceed with
! FORMATION OF THE X-RAY TUBE. This is to prevent high voltage discharges which might
damage the X-ray tube. The formation procedure is described in the Service Manual.
Present on the “EMISSION STATUS” area(see § 3.4.11.1) At this point You can also
switch on the collimator light pushing on the relative icon:
It will appear the control keyboard for the Kv and mAs selection
Set the desired values for these parameters by means of the KV SELECTION and mAS SELECTION
pushbuttons.
Position the panel
Remove the X-RAY CONTROL pushbutton from the X-RAY CONTROL SUPPORT. This device is a
pushbutton with double click (CLICK A and CLICK B - see figures) and must be used as shown
below.
Use the X-RAY CONTROL pushbutton and carry out the first click (A). Keep the pushbutton in
this position for 3 seconds. When You are ready, make the second click (B) with the X-RAY
CONTROL pushbutton and keep it in that position. At this moment x-ray radiation emission
starts and the Emission in progress LED lights up.
At the end of X-RAY RADIATION emission, the EMISSION IN PROGRESS LED turns off and the acoustic
signal ends. At this point it is possible to release both the clicks of the X-RAY CONTROL
pushbutton. After some seconds the TOUCH-SCREEN will display the image acquired.
TReleasing the first click (A) early does not have any
Hconsequences.
EEarly release of the second click (B) interrupts x-ray radiation
! emission. The Display will show the wording “Man X-ray Stop”
Ufor about 10 seconds and the Unit will emit an intermittent
Nsound.
I
T
! Do not remove the connector from the power supply socket unless the unit has been turned off.
Should it be necessary to replace parts which could jeopardise safety of the machine, only use
! original spare parts
The weekly, six-monthly and annual checks are reserved for qualified and authorised
! personnel of the technical service.
DAILY CHECKS
Check operation of the signals, the touch-screen monitor and the luminous LEDs. Check
operation of the parking brake.
Check integrity of the warning and danger labels.
WEEKLY CHECKS
Check that there are no oil leaks from the Monoblock.
Check there are no unusual noises in the Monoblock during X-ray emission.
SIX-MONTHLY CHECKS
Check correct operation and the relative value of the whole earthing circuit. Check the
power supply voltage value.
Check the value of the direct voltages generated inside the system. Check fixing
and general state (dust and corrosion) of the cards.
Check centring of the monoblock-collimator group.
ANNUAL CHECKS
Check kV and mAs (*)
Check leakage current
Safety Test Check electric
strength Check earth
resistance
(*) Call the Service to carry out the safety and reproducibility tests, as indicated in the IEC 1223-2 and IEC 1223-11 standards,theother
functional checksof the apparatus, as described in the programmedmaintenanceplan (Service manual)
Turn the system off and disconnect the mains power supply cable .
Make sure that no liquid enters the apparatus, in order to prevent short-circuits and corrosion
of the electrical and electro-mechanical components.
The apparatus is not suitable for use in the presence of anesthetics and/or flammable disinfection and
! cleaning products. Should products which form explosive gaseous mixtures be used, make sure that
the gases are dispersed before turning the unit on again.
For any replacements of the unit components (Monoblock, collimator), please refer to the Service
! Manual– Section 3 – Maintenance
Description Data
Voltage 230 Vac
Frequency 50 Hz
Continuousoperation<1A
Max absorbed current
Intermitted operation 12A
Line compensation Automatic
Line resistance < 2.5 Ω
Class I
Classification EN60601-1
With type B part applied
Conditionsofuse EN60601-1 Continuous operation with intermitted load
Directive classification 2007/47/CE Class II b(annexIX norm 10)
Description Data
Rear-lit Touch Screen for all the operating parameters and messages for any
Operator interface
possible anomalous conditions. Managed by Microcontroller.
Selectable languages Italian, English
Radiography
Remote control with double click and extension cable or WIFI
exposure control
Radiography with
Technique with 2 points (kV, mAs).
free technique
– mAmin and mAmax Safety device
– Maximum X-ray tube load safety device
– Maximum exposure time safety device
– Temperature and Monoblock
Safety devices – Max kV, min kV, max I
– Capacitor Faulty
– Starter Anode Faulty
– Microprocessor auto-test with display of the diagnostic messages..
TABLE6.3
TRANSPORT AND STORAGE OPERATION
Description Data Data
Maximum Temperature From –10°Cto55°C From 10 °Cto 40°C
Recommended temperature From 0 °Cto 40°C From 10 °Cto 40°C
Relative humidity From 20%to 90% From 30%to 75%
Pressure From 500 to 1060 hPa From 700 to 1060 hPa
TABLE6.4
Maximum power 32 kW
Large Focus
mA Table 6.5A
s22 32 / / / / / / / / / / / / / /
0 0
N.A.
20 32 32 / / / / / / / / / / / / /
0
16 0
32 0
32 / / / / / / / / / / / / /
0
13 0
32 0
32 32 32 / / / / / / / / / / /
0
10 0
32 0
32 0
32 0
32 32 32 / / / / / / / / /
080 0
32 0
32 0
32 0
40 0
40 0
32 32 32 25 / / / / / /
63 0
32 0
32 0
32 0
40 0
40 0
32 0
32 0
32 0
32 32 25 20 / / /
50 0
32 0
32 0
32 0
40 0
40 0
32 0
32 0
32 0
32 0
32 0
25 0
20 20 20 16
40 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
25 0
20 0
20 0
20 0
20
32 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
32 0
25 0
25 0
20 0
20
25 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
32 0
25 0
25 0
20 0
20
20 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
25 0
25 0
20 0
20
16 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
25 0
25 0
25 0
20
13 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
25 0
25 0
25 0
20
10 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
25 0
25 0
25 0
20
8 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
25 0
25 0
25 0
20
6. 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 0
32 0
25 0
25 0
25 0
20
53 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 0
30 0
25 0
25 0
25 0
20
4 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 2
30 0
25 0
25 0
25 0
20
3. 0
32 0
32 0
32 0
40 0
40 0
40 0
40 0
40 0
32 0
32 2
32 0
25 0
25 0
25 0
20
2
2. 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
25 0
25 0
25 0
25
25 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
25 0
25 0
25 0
25
1. 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
32 0
25 0
25 0
25 0
25
6
1. 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25
13 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25 0
25
0. 0
20 0
20 0
20 0
20 0
20 0
20 0
20 0
20 0
20 0
20 0
20 0
20 0
20 0
20 0
20
8
0. 0
16 0
16 0
16 0
16 0
16 0
16 0
16 0
16 0
16 0
16 0
16 0
16 0
16 0
16 0
16
5
0. 06 06 06 06 06 06 06 06 06 06 06 063 063 063 063
2 3
40 43 43 3 53 63 63 73 3 83 93 10 10 11 12
kV 5 8
-4 3-
4 6-
4 1-
6 1-
6 6-
7 1-
8 2-
9 1-
10 1-
10 4-
11 1-
12 1-
12
0 1
2 5 9 0 5 0 0 0 0 3 0 0 5
mA Table 6.5B
s22 12 / / / / / / / / / / / / / /
0
20 5
N.A.
12 12 / / / / / / / / / / / / /
0
16 5
12 5
12 / / / / / / / / / / / / /
0
13 5
16 5
16 16 16 / / / / / / / / / / /
0
10 0
16 0
16 0
16 0
16 12 10 / / / / / / / / /
08 0
16 0
16 0
16 0
16 5
12 0
10 10 8 8 / / / / / /
0
6 0
16 0
16 0
16 0
16 5
12 0
10 0
10 100 0
8 8 8 63 / / /
3
5 0
16 0
16 0
16 0
16 5
12 0
12 0
12 0
10 0
8 0
8 0
8 63 63 50 50
0
4 0
16 0
16 0
16 0
16 5
12 5
12 5
12 0
10 0
8 0
8 0
8 63 63 63 50
0
3 0
16 0
16 0
16 0
16 5
16 5
12 5
12 0
10 0
10 100 0
8 80 80 63 50
2 0
16 0
16 0
16 0
20 0
16 5
12 5
12 0
10 0
10 0
10 0
8 80 80 63 50
5
2 0
16 0
16 0
16 0
20 0
16 5
12 5
12 0
10 0
10 0
10 0
8 80 80 63 50
0
1 0
16 0
16 0
16 0
20 0
16 5
12 5
12 0
10 0
10 0
10 0
8 80 80 80 63
6
1 0
16 0
16 0
16 0
20 0
16 5
12 5
12 0
12 0
10 0
10 0
8 80 80 80 63
3
1 0
16 0
16 0
16 0
20 0
16 5
12 5
12 5
12 0
10 0
10 0
10 80 80 80 63
80 0
16 0
16 0
16 0
20 0
16 5
16 5
16 5
12 0
10 0
10 0
10 80 80 80 80
6. 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 0
10 0
10 0
10 80 80 80 80
53 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 0
10 0
10 0
10 10 10 80 80
4 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 0
10 0
10 0
10 0
10 0
10 80 80
3. 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 0
10 0
10 0
10 0
10 0
10 80 80
2
2. 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 0
10 0
10 0
10 0
10 0
10 80 80
25 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 0
10 0
10 0
10 0
10 0
10 80 80
1. 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 0
10 0
10 0
10 0
10 0
10 80 80
6
1. 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 0
12 0
10 0
10 0
10 0
10 80 80
13 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 5
12 0
10 0
10 0
10 0
10 80 80
0. 0
16 0
16 0
16 0
20 0
16 0
16 0
16 5
12 5
12 0
10 0
10 0
10 0
10 80 80
8
0. 0
16 0
16 0
16 0
16 0
16 0
16 0
16 5
12 5
12 0
10 0
10 0
10 0
10 80 80
5
0. 06 06 06 06 06 06 06 56 56 06 06 063 063 36 63
2 3
40 3
43 3
46 3 53 3
61 3
66 73 3 3
82 3
91 10 10 11 12
k 5 8
-4 -4 -4 1-
6 -6 -7 1-
8 -9 -10 1-
10 4-
11 1-
12 1-
12
V 0 1
2 5 9 0 5 0 0 0 0 3 0 0 5
400,00
360,00
320,00 49
81
mA Grap hy
280,00
240,00
101
200,00
51 121
160,00
120,00
111
80,00 81 125
40,00
35 45 55 65 75 85 95 105 115 120 125
kV
TABLE6.6
Description Data
Width 694 mm
Length 1152 mm
Height 1525 mm
Focus’s Heightmax, with outreach 1867 mm
Weight < 350 kg
Handling Dead man brake
TABELLA 6.7
Description Data
Multiple level blades parallel and perpendicular
Blade handling Manual
Dimensions of radiated field withfocus-film distanceofdi 1m:43x43cm
Focus-film distance indication with retractable tape measure(2m)
Light brightness (lux) >160 lux a 1m
Inherent collimator filtration 2 mm Al
All the other information regarding the collimator are present in its relative manual.
System Composition
Rack PC industrial
CPU Intel >2GHZ, memorie 2GB
System operative WINDOWS 7 or superior
Hard Disk archive 200 GB or superior with possibility to memorize until 10.000 imaging