5123649
Revision 1

HiSpeed Dual, ProSpeed FII Series


Camera Interface Option Installation Manual

Copyright © 2004 by General Electric Company

Operating Documentation
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
WARNING TRANSLATION SERVICES.
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.

• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
AVERTISSEMENT TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

• DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN ENGLISCHER


SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
WARNUNG IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

• ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
AVISO SERVICIO DE TRADUCCIÓN.
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

a
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

• ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL


EM INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
ATENÇÃO GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

• IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN


INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
AVVERTENZA MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

b
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

IMPORTANT! . . . X-RAY PROTECTION


X–ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained
should be thoroughly read and understood by everyone who will use the equipment before you attempt to
place this equipment in operation. The General Electric Company, Medical Systems Group, will be glad to
assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x–radiation other than the useful beam,
no practical design of equipment can provide complete protection. Nor can any practical design compel the opera-
tor to take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others
to radiation.
It is important that everyone having anything to do with x–radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP
Reports available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and
of the International Commission on Radiation Protection, and take adequate steps to protect against injury.

The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its
agents, and representatives have no responsibility for injury or damage which may result from improper use of the
equipment.

Various protective material and devices are available. It is urged that such materials or devices be used.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT

All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment
shall be performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit
shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment,
calibrations, and testing shall be performed by qualified GE Medical personnel. The products involved (and the
accompanying electrical installations) are highly sophisticated, and special engineering competence is required.

In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s
electrical work on these products will comply with the requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third–
party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the
equipment.

c
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a Gen-
eral Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported
to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and con-
tainers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection
is not requested within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 827-3449 / 8*285-3449 immediately after damage is found.
At this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.

OMISSIONS & ERRORS


GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documenta-
tion. Customers, please contact your GE Sales or Service represenatives.

CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.

Devices not to be used near this equipment:

Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmit-
ter, radio-controlled toy, etc.

Keep power to these devices turned off when near this equipment.

Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.

d
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

LIST OF EFFECTIVE PAGES

Tab - Section Page REV 1 (Introduction) 1-1 1

Title page 1 2 (Material List) 2-1 1


-
Title page rear blank
3 (Hardware Installation) 3-1 to 3-9 1
(Warning and other
important information)
a to d 1 4 (Configuration) 4-1 1

(Revision Information) A 1
5 (Final Check) 5-1 to 5-2 1

(Table of Contents) i 1
- Blank/Rear cover -

REVISION HISTORY
REV Date Primary Reason for Change
1 09/30/04 Initial release.

A
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
blank

B
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

SECTION 2 - MATERIAL LIST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

SECTION 3 - HARDWARE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1


3-1 SWITCH SETTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-1-1 LCAM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-1-2 VDB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-2 INSTALLLING THE LCAM DASM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-3 INSTALLLING THE VDB DASM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7

SECTION 4 - CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1


4-1 GENERAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-2 CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

SECTION 5 - FINAL CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1


5-1 ATTACHING THE RATING PLATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5-2 EXPLANATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5-3 SENDING THE PRODUCT LOCATOR CARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2

i
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION

ii
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

SECTION 1 - INTRODUCTION

The option camera interface is required in order to connect CT system and Laser Camera (Imager).
There are two (2) kinds of filming interfaces.
• Analog camera interface (DASM VDB)
• Digital camera interface (DASM LCAM)
The decision which interface is used depends upon laser camera. Consult the camera manufacturer before instal-
lation.
GE provides ONLY the DASM LCAM (or VDB) and its interface that is installed in the CT Operator Console (OC).
GE dose NOT provide the interface between DASM LCAM (or VDB) and Laser Camera.
See Illustration 1-1.

Illustration 1-1 GE and Camera Maker Support Overview

GE Support Camera Manufacturer Support

CT Operator Console
This Option

DASM – LCAM IMAGER LASER CAMERA


OR I/F (IMAGER)
DASM – VDB

1-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION

1-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

SECTION 2 - MATERIAL LIST

DASM TYPE Part No.


LCAM (Digital) 5121408

ITEM PART NO. DESCRIPTION QTY


1 2191524–2 DASM LCAM (without case) 1
2 2401055 SCSI TOWER 1
3 5122642 LCAM CABLE 1
4 2146513 SCREW 4
5 5122644 SCSI CABLE 1
6 2111577 RATING PLATE 1
7 2121680 STUD 1
8 5123700 LCAM PANEL 4
9 N9301BB NUT 4
10 N9306HR SCREW 4
11 5123649 Instruction (This Manual) 1

DASM TYPE Part No.


VDB (Analog) 5121409

ITEM PART NO. DESCRIPTION QTY


1 2191523–3 DASM VDB (without case) 1
2 2401055 SCSI TOWER 1
3 2146513 SCREW 4
4 5122644 SCSI CABLE 1
5 2111577 RATING PLATE 1
6 46–296980G1 ADAPTER CABLE 1
7 5123649 Instruction (This Manual) 1

2-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION

2-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

SECTION 3 - HARDWARE INSTALLATION

3-1 SWITCH SETTING

3-1-1 LCAM
1. Set the switches as shown below:

Illustration 3-1 LCAM Switch Setting


BOTTOM VIEW

RS232/RS422 Selection Select “OFF” as Active


SCSI termination.

3 4 5
2 6 Select “3” as SCSI ID.
1 0 7
SCSI Termination
RS232
SCSI ID
RS422

Select “RS232”.

Printed Circuit Board VIEW

Pin Header : Not in use

Closed
SP
DIP Switch
4

0
OPEN
3

The bit swtiches corre-


1 sponding to SP ~ 2
printed on the board
2

2 MUST be set as
1

shown.

Attention

3-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

3-1-2 VDB
1. Set the switches as shown below:

Illustration 3-2 VDB Switch Setting

BOTTOM VIEW

Select “OFF” as Active


SCSI termination.
SCSI Termination
SCSI ID 3 4 5
2 6 Select “3” as SCSI ID.
1 0 7

Printed Circuit Board VIEW


Pin Header : Not in use
60Hz

Attention
1

2
0

OPEN

Closed
4 3 2 1

DIP Switch
Must be set to Open The bit swtiches corre-
(60Hz). sponding to 0 ~ 2
printed on the board
MUST be set to
“Closed”.

3-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

3-2 INSTALLLING THE LCAM DASM


1. Remove the two cosmetic covers of the DASM Assy.
2. Install the LCAM DASM (with LCAM Panel) into the DASM Assy by screwing its eight (8) screws on the two
sides.
3. Connect SCSI Power Cable and SCSI Chain Cable to the LCAM DASM.

Illustration 3-3 Installing the LCAM DASM Panel

LCAM Panel SCSI Power Cable SCSI Chain Cable

DASM Assy
4. Assemble the LCAM Cable with the holding cover.

Illustration 3-4 Assembly LCAM Cable Assy


LCAM Cable

Holding Panel

3-3
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

3-2 INSTALLLING THE LCAM DASM (CONTINUED)


5. Install the LCAM Cable Assy into the LCAM Assy by screwing its eight (8) screws (see illustration 3-5 and con-
nect it to LCAM DASM (see illustration 3-6).

Illustration 3-5 installing LCAM Cable Assy

Push

Illustration 3-6 connect LCAM Cable to LCAM DASM

3-4
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

3-2 INSTALLLING THE LCAM DASM (CONTINUED)


6. Connect the SCSI Cable to the SCSI Tower Assy of the OC (see illustration 3-7).

Illustration 3-7 Connect Data Cable

DASM ASSY

SCSI Cable

SCSI TOWER

Power Cable
Connectors

3-5
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

3-2 INSTALLLING THE LCAM DASM (CONTINUED)


7. Connect the following cables to camera (see illustration 3-8).
Note
GE does NOT provide these two interface cables. They are components of laser camera. Please
refer to the camera specification for their usage.

Illustration 3-8 Interface cables between LCAM DASM and Laser Camera
Interface Cables

3-6
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

3-3 INSTALLLING THE VDB DASM


1. Remove the lower cosmetic cover of the DASM Assy.
2. Install the VDB DASM (with VDB Panel) into the DASM Assy by screwing its eight (8) screws on the two sides.
3. Connect SCSI Power Cable and SCSI Chain Cable to the VDB DASM (illustration 3-9).

Illustration 3-9 Installing the LCAM DASM Panel


SCSI Power Cable SCSI Chain Cable
Cosmetic Cover VDB Panel

DASM Assy

3-7
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

3-3 INSTALLLING THE VDB DASM (CONTINUED)


4. Connect the SCSI Cable to the SCSI Tower Assy of the OC (see illustration 3-10).

Illustration 3-10 Connect Data Cable

DASM ASSY

SCSI Cable

SCSI TOWER

Power Cable
Connectors

3-8
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

3-3 INSTALLLING THE VDB DASM (CONTINUED)


5. Connect the following cables to camera (see illustration 3-11).
Note
GE does NOT provide these two interface cables. They are components of laser camera. Please
refer to the camera specification for their usage.

Illustration 3-11 Interface cables between VDB DASM and Laser Camera

Interface Cables

3-9
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION

3-10
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

SECTION 4 - CONFIGURATION

4-1 GENERAL
The CAMERA INTERFACE software has been installed in the system standard software, but there are parameter
settings required before sending data to the camera device. In addition to that, the communication test between the
console and DASM (LCAM or VDB) should be run to verify the camera interface functions.

4-2 CONFIGURATION
Refer to the following sections:
• To enter and exit from the configuration screen, refer to Functional Check / Adjustment, Introduction,
Performing the System Configuration.
• To setup the camera configuration, refer to Functional Check / Adjustment, System, Software Loading,
Changing System Configuration (Reconfig).

4-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION

4-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

SECTION 5 - FINAL CHECK

5-1 ATTACHING THE RATING PLATE


1. Attach the provided rating plate to the following location (in the label area left to the rating plate of the OC).

Illustration 5-1 Rating Plate Location

OC Rear Side

Label area Rating plate

5-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

5-2 EXPLANATION
Provide an explanation of the use of a laser camera.

5-3 SENDING THE PRODUCT LOCATOR CARD


1. Fill in the blanks on the Product Locator Card provided in this kit.
2. Send the Product Locator Card to the appropriate address provided below.

Address for GEMS-AM sites:


Send to the address printed on the Product Locator Card.

Address for GEMS-E sites:


Send to the address printed on the Product Locator Card.

Address for GEMS-A sites:


Send to the local PLS or Service Administrator.

5-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
blank

5-3
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION

5-4
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION
REV 1 5123649

APPENDIX A - LASER CAMERA SUPPORT LIST

The following list is what GE recommends to connect with Laser Camera.

Maker Type Connection Test Formatting On Film


1 FUJI FL–IM D VDB T.B.D. 1 2 4 6 12 20
LCAM T.B.D. 1 2 4 6 12 20
2 FUJI FL–IM3543II VDB T.B.D. 1 2 4 6 12 20
LCAM T.B.D. 1 2 4 6 12 20
3 KONICA Li–10A VDB OK 1 2 4 6 9 12 15 16 20
LCAM T.B.D. 1 2 4 6 9 12 15 16 20
4 KONICA Li–7 VDB OK 1 2 4 6 9 12 15 16 20
LCAM OK 1 2 4 6 9 12 15 16 20
5 KONICA Li–21 VDB OK 1 2 4 6 9 12 15 16 20
LCAM T.B.D. 1 2 4 6 9 12 15 16 20
6 KODAK KELP–2180 VDB T.B.D. 1 2 4 6 8 9 12 15 16 20
LCAM T.B.D. 1 2 4 6 8 9 12 15 16 20
7 KODAK KELP–1120 VDB OK 1 2 4 6 8 9 12 15 16 20
LCAM OK 1 2 4 6 8 9 12 15 16 20
8 KODAK 100XLP VDB T.B.D. 1 2 4 6 8 9 12 15 16 20
LCAM T.B.D. 1 2 4 6 8 9 12 15 16 20
9 AGFA LR3000 VDB T.B.D. 1 2 4 6 8 9 12 15 16 20
LCAM T.B.D. 1 2 4 6 8 9 12 15 16 20
10 DUPONT LINX VDB T.B.D. 1 2 4 6 8 9 12 15 16 20
LCAM T.B.D. 1 2 4 6 8 9 12 15 16 20
11 3M 969HQ VDB OK 1 2 4 6 8 9 12 15 16 20
LCAM OK 1 2 4 6 8 9 12 15 16 20
12 3M 959XL VDB OK 1 2 4 6 8 9 12 15 16 20
LCAM OK 1 2 4 6 8 9 12 15 16 20
13 3M 952PLUS VDB OK 1 2 4 6 8 9 12 15 16 20
LCAM OK 1 2 4 6 8 9 12 15 16 20

A-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS CAMERA INTERFACE OPTION INSTALLATION

A-2
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)

GE Medical Systems - Europe: Telex 698626


283, rue de la Miniére, B.P. 34, 78533 Buc Cedex, France