NATIONAL METROLOGY DIVISION


TEMPERATURE LABORATORY
CALIBRATION CERTIFICATE No: 111/2020-2021

1.0 Customer’s Details

Name of the customer : T2000 Hotel


Location : Nyarugenge
Tel : 0787791631
E-mail : -
2.0 Details of the Instrument
Name of the instrument : Refrigerator
Manufacturer : Binbomeichu
Model : -
Serial No : -
Identification Number : -
Calibration range/point : < 5°C.
Condition of the Instrument : Good
Sticker Number : JUL20C0271
Calibration Date : 14/07/2020
3.0 Standard(s) used to calibrate the Instrument (s) and traceability

Standard Name Type/ Model Serial number Sticker/Cert. number Calibration Date
Data logger OctTemp2000 Q85484 NC-18-12-19-008 11/2018
Thermocouple TC-J - 9002 04/2018

This calibration certificate documents the traceability to national standards, which realize the units of
measurement according to the international system of units (SI).

4.0 Procedure Used to calibrate the Instrument (s)


The calibration of this instrument was carried out as prescribed in the procedure MEU/3-PRO/04:
Temperature chamber calibration procedure, which was developed according to the guideline,
DKD-R 5-7 (Calibration of climatic chambers).
5.0 Remark
As agreed with the customer, the calibration due date is not later than July, 2021.
KARAMUZI M. Eric
Director of Mechanical and Signature: …………….. Date: ……/………/………..
Electrical Laboratories Unit

ZIMULINDA Philibert
Division Manager Signature: …………….. Date: ……/………/………..
For
Director General
This certificate has been issued without any alteration and may not be reproduced other than in full except with the
approval of the Director General. The results contained herein apply only to that particular equipment and conditions
stated above.

KK 15 Rd, 49 Kicukiro P.O.BOX 7099, Kigali-Rwanda Telephone: (+250)788303492


Hotline: 3250 Email: info@rsb.gov.rw Website: www.rsb.gov.rw
Page 2 of 2

NATIONAL METROLOGY DIVISION


TEMPERATURE LABORATORY
CALIBRATION CERTIFICATE No: 111/2020-2021

6.0 Environmental conditions


Normal ambient conditions
7.0 Calibration results

The operation temperature of the instrument is <5 °C.


The indicator’s temperature no indicator.

7.1 Temperature measurements

The instrument was calibrated by positioning sensors in different locations in useful volume and
measurements were taken at 2 minutes interval for 30 minutes after the instrument stabilized at a
preset temperature as indicated below.

Sensor 1................................................................……. 2.36°C

Sensor 2 .................................................................……2.61°C

7.2 Temperature variations

The maximum temperature variation between the sensors after a period of one hour was ±0.29°C.

7.3 Temperature fluctuations

Sensor 1....................................................................... ± 0.09°C

Sensor 2........................................................................± 0.17C

8.0 Measurement Uncertainty

The Uncertainty of the calibration results for the given range of temperature measurements is ± 0.50C.
The reported expanded uncertainty is stated as expanded uncertainty of measurement multiplied by
coverage factor K= 2, providing a confidence level of approximately 95%.

………………………………………………..END…….……………………………………………….