Beruflich Dokumente
Kultur Dokumente
Fluke ScopeMeter®
190 Series
Application Note
input channels and can measure during a selected part of the mains screen; this may also be changed
voltage and current at the same cycle. The effective output voltage to later. The two input waveforms and
time, whilst also measuring and the load (the lights) can be the resultant (M) will now be visible.
displaying the phase angle between controlled by changing the phase If the resultant is too small or too
them. See also the section “practical angle. The measured voltage (see large, the scaling can be modified.
set-up hints” near the end of this figure 2, trace A) shows that the After pressing softkey F3 we can
publication for further details. output is active during approxi- also change the vertical position of
mately two-thirds of the time, or the resultant trace M.
If the voltage is non-sinusoidal or if
about 120 degrees during each half-
the load isn’t purely resistive,
cycle. During one-third of the time,
power measurements become too
the dimmer is ‘off’ and no output
complex to be performed on a
voltage is supplied to the lights. If
DMM. The best way to determine
we change this phase angle, the
power under these conditions is to
lights will brighten or dim.
take a large number of current and
voltage measurements over each
cycle of the supply voltage. The
measurements need to be made
Figure 4: Voltage, Current and Power curves
simultaneously on both signals. on the Fluke 190 Series ScopeMeter.
Each set of simultaneous
Figure 4 shows the voltage
measurements can then be
(waveform A, in red), the current (B,
multiplied to produce a
in blue) and the multiplied curve
corresponding set of data points
(M, in green), representing the
from which a curve can be
power supplied to the lights.
constructed of the power handling Figure 2: Output voltage and current of a
dimmer driving a string of lights.
at successive moments in time. On the Fluke ScopeMeter 190
Series, the cursors can be used to
The Fluke 190 Series ScopeMeter is In figure 2 we can also see the
measure the power at any point in
capable of performing this specific resulting current through the light-
time, as shown in figure 5.
function for you! bulbs (channel B, blue curve). We
can now set up the ScopeMeter to
Included in the functionality of the
calculate the power applied to the
Fluke ScopeMeter 190 Series is the
lights (see figures 3 and 4).
ability to multiply individual curves
To do this select:
(waveforms) to create a resultant
SCOPE,
curve. With this function, sets of
then F4 (= Waveform Options),
samples from channels A and B are
next Mathematics, and Enter,
multiplied to create a resultant
then A * B, and Enter.
curve labeled M. In other words,
every time a sample for channel A Next we select a scaling that we
Figure 5: The ScopeMeter’s cursor is used for
and a sample for channel B is taken expect will keep the power curve on power measurement at a specific point in time.
Here, the cursor is set at about the mains’ safety ground). In contrast,
maximum peak of the power curve, the ScopeMeter 190 Series has
and the reading tells us that the independently floating inputs. This
lights are handling a peak of allows for voltage measurements at
approximately 1.7 kW at that voltage test-points as well as across
instant in time. current-sensing resistors, which
may be at different voltage levels.
Power measurement in a See the ScopeMeter’s technical
switched-mode power specification for full details.
supply (SMPS) Figure 7: Voltage and current waveforms are When used with current clamps or
used to create the power-curve.
In electronic systems, the with current sensing resistors in the
frequencies of the signals are often From the curves in figure 6, we network, the ScopeMeter inputs can
much higher and the waveforms calculate the power that the be set-up to read amplitudes in
much more diverse than those in transistor is handling by multiplying amps directly.
above example. the two graphs, as in figure 7. Here
To do so, select the input channel
also the timebase-setting has been
As an example, consider some key (e.g.: “B”) and then softkey F3
changed to see the part of the
waveforms from a switched-mode (“Probe channel B”). A selection can
waveform that is of particular
power supply (figure 6). In this then be made if the input signal
interest in more detail.
system, the mains voltage is represents a voltage or a current, or
On these curves, we can use the
rectified and filtered, resulting in a even a temperature. What’s more,
cursors of the Fluke 190 Series to
DC voltage of about 350 V. This is the sensitivity of the current clamp
measure the peak in power
then applied to a switching or the circuit’s sensing resistor can
transistor that drives a step-down be selected from a selection table,
transformer. indicated in mV/mA (equivalent to
V/A or simply Ω).
clamp. These are commercially Figure 10: Adding a current sensing resistor to For more information call:
the circuitry to allow current measurement. In the U.S.A. (800) 443-5853 or
available for AC and for DC+AC Fax (425) 446-5116
measurements, and for various In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
current ranges. The Fluke 80i-100s, If no switch is provided, or if the In Canada (800)-36-FLUKE or
Fax (905) 890-6866
for instance, measures DC and AC DMM has no current-measuring From other countries +1 (425) 446-5500 or
currents from 0.1 to 100A; the Fluke capability, we may also add a Fax +1 (425) 446-5116
Web access: http://www.fluke.com
i1010 can even be used up to current sensing resistor of a known
1000A AC and/or DC. value Rs to the circuit, which needs
to be small in value compared with ©2003 Fluke Corporation. All rights reserved.
Printed in the Netherlands.
10/2003 Pub-ID 10654-eng Rev. 01