Sie sind auf Seite 1von 4

Measuring power with a

Fluke ScopeMeter®
190 Series
Application Note

Although many electrical test tools are available to measure


voltage, few can measure current, and even fewer are equipped
to measure electrical power directly. Moreover, there is always
the question of how to measure power in electronic systems that
are not operating at mains frequencies. The Fluke ScopeMeter
190 Series has the answer.

Power measurements This elementary power


The prime parameter specified for measurement, using a DC-source for
any electrical system is operating U1 and a resistive load RL requires
voltage. But that voltage in itself has little more than a multimeter, which
little meaning if, once applied, it will usually be a digital multimeter
doesn’t result in power to, for (DMM). The DMM is used first to
example, exert a force, drive a measure the voltage U1 and then to
machine or run a lighting system. measure the current supplied to the
Accurate measurement of electrical load RL, from which power supplied
power is therefore crucial for to the load is calculated from
understanding a system’s behaviour. formula (1).

Suppose we have an electrical Once the switch is closed, all values


circuit consisting of a voltage source in this DC-system are static and
If the load isn’t purely resistive but
U1, a switch Sw and a load resistor therefore the two measurements
also includes inductive or capacitive
RL (see figure 1). By closing the can be made successively using a
elements, a phase difference
switch, the supply voltage U1 is single instrument if so desired.
between the applied voltage and
connected to the load RL and a
If the source is a low-frequency AC the current will result that needs to
current I will flow as a result. Once
source, the same rules basically be taken into account to determine
this current is flowing, the load will
apply. We can measure the RMS the power. The relation now is:
heat as a result of the power
value of the applied voltage and
applied to it according to:
that of the current, and multiply P = U1 * I * cos ϕ (2)
these to get the power handled by
P = U1 * I (1)
the load. As long as we’re working where ϕ is the phase-angle in
at mains frequencies, most DMMs degrees between voltage and
can do the job. current.

Multimeters are generally not


I equipped to make measurements on
A more than one input and therefore
Sw cannot measure phase angle. For
these measurements, a more
U1 RL V specialized instrument is needed,
for instance a phase meter or an
oscilloscope.
If power measurements are made
Figure 1: Basic circuit loop in which voltage more frequently, a power meter like
and current are measured.
the Fluke 43B Power Quality and written on screen, these are
Analyzer would be the most also multiplied and displayed to
appropriate test tool. This create the resultant waveform M on
instrument measures both voltage screen. From this waveform, the
and current simultaneously and (instantaneous) power can be read
automatically takes the phase angle off, at every point in time, for
into account. For the occasional instance using the cursors of the
user, however, this may be an Fluke 190 Series.
expensive option and a more
generic tool like the ScopeMeter Example of a light dimmer
would be preferred. As an example, we’ve applied the
mains voltage to a so-called “light
Use of the Fluke dimmer”. This is a thyristor-based
ScopeMeter power control device in which Figure 3: Menu tree for setting up the
All Fluke ScopeMeters have two current is allowed to flow only waveform multiplication.

input channels and can measure during a selected part of the mains screen; this may also be changed
voltage and current at the same cycle. The effective output voltage to later. The two input waveforms and
time, whilst also measuring and the load (the lights) can be the resultant (M) will now be visible.
displaying the phase angle between controlled by changing the phase If the resultant is too small or too
them. See also the section “practical angle. The measured voltage (see large, the scaling can be modified.
set-up hints” near the end of this figure 2, trace A) shows that the After pressing softkey F3 we can
publication for further details. output is active during approxi- also change the vertical position of
mately two-thirds of the time, or the resultant trace M.
If the voltage is non-sinusoidal or if
about 120 degrees during each half-
the load isn’t purely resistive,
cycle. During one-third of the time,
power measurements become too
the dimmer is ‘off’ and no output
complex to be performed on a
voltage is supplied to the lights. If
DMM. The best way to determine
we change this phase angle, the
power under these conditions is to
lights will brighten or dim.
take a large number of current and
voltage measurements over each
cycle of the supply voltage. The
measurements need to be made
Figure 4: Voltage, Current and Power curves
simultaneously on both signals. on the Fluke 190 Series ScopeMeter.
Each set of simultaneous
Figure 4 shows the voltage
measurements can then be
(waveform A, in red), the current (B,
multiplied to produce a
in blue) and the multiplied curve
corresponding set of data points
(M, in green), representing the
from which a curve can be
power supplied to the lights.
constructed of the power handling Figure 2: Output voltage and current of a
dimmer driving a string of lights.
at successive moments in time. On the Fluke ScopeMeter 190
Series, the cursors can be used to
The Fluke 190 Series ScopeMeter is In figure 2 we can also see the
measure the power at any point in
capable of performing this specific resulting current through the light-
time, as shown in figure 5.
function for you! bulbs (channel B, blue curve). We
can now set up the ScopeMeter to
Included in the functionality of the
calculate the power applied to the
Fluke ScopeMeter 190 Series is the
lights (see figures 3 and 4).
ability to multiply individual curves
To do this select:
(waveforms) to create a resultant
SCOPE,
curve. With this function, sets of
then F4 (= Waveform Options),
samples from channels A and B are
next Mathematics, and Enter,
multiplied to create a resultant
then A * B, and Enter.
curve labeled M. In other words,
every time a sample for channel A Next we select a scaling that we
Figure 5: The ScopeMeter’s cursor is used for
and a sample for channel B is taken expect will keep the power curve on power measurement at a specific point in time.
Here, the cursor is set at about the mains’ safety ground). In contrast,
maximum peak of the power curve, the ScopeMeter 190 Series has
and the reading tells us that the independently floating inputs. This
lights are handling a peak of allows for voltage measurements at
approximately 1.7 kW at that voltage test-points as well as across
instant in time. current-sensing resistors, which
may be at different voltage levels.
Power measurement in a See the ScopeMeter’s technical
switched-mode power specification for full details.
supply (SMPS) Figure 7: Voltage and current waveforms are When used with current clamps or
used to create the power-curve.
In electronic systems, the with current sensing resistors in the
frequencies of the signals are often From the curves in figure 6, we network, the ScopeMeter inputs can
much higher and the waveforms calculate the power that the be set-up to read amplitudes in
much more diverse than those in transistor is handling by multiplying amps directly.
above example. the two graphs, as in figure 7. Here
To do so, select the input channel
also the timebase-setting has been
As an example, consider some key (e.g.: “B”) and then softkey F3
changed to see the part of the
waveforms from a switched-mode (“Probe channel B”). A selection can
waveform that is of particular
power supply (figure 6). In this then be made if the input signal
interest in more detail.
system, the mains voltage is represents a voltage or a current, or
On these curves, we can use the
rectified and filtered, resulting in a even a temperature. What’s more,
cursors of the Fluke 190 Series to
DC voltage of about 350 V. This is the sensitivity of the current clamp
measure the peak in power
then applied to a switching or the circuit’s sensing resistor can
transistor that drives a step-down be selected from a selection table,
transformer. indicated in mV/mA (equivalent to
V/A or simply Ω).

If we have connected a 1 Ω current


sensing resistor in our network and
want to measure the voltage across
it, the current sensitivity would be
set at 1 V/A. If we include a 0.1 Ω
resistor in our network to sense the
Figure 8: Cursor used to measure the peak of current, then the sensitivity would
the power, handled by the transistor
be set at 100 mV/A. If the voltage
across that resistor is measured
Figure 6: Voltage and current handled by the handling, as is done in figure 8.
switcher component in a SMPS. using the standard (10:1) voltage
From this we see that the switching
probe, the overall sensitivity would
In figure 6 we can see the voltage transistor is handling a peak power
be 10 mV/A. By selecting this
across the switching transistor of 123 W. During design of the
sensitivity from the menu, the
(curve A, in red) and the current SMPS, one has to be aware that
current that we read on the
through the transistor (curve B, in such power peaks may occur and
ScopeMeter screen directly indicates
blue). The voltage reaches peak the components need to be selected
the true value.
values of over 400 V (see the 4 with this in mind.
divisions of amplitude at 100 V/div)
Total Energy
and the current has a peak-value of Instrument set-up
The total energy handled over time
over 200 mA. A single cycle of this The ScopeMeter is unique in that its
can be calculated by multiplying the
converter’s signal takes inputs are fully insulated from one
continuous power and the time the
approximately 26 µs, which means another, allowing direct connection
system is active:
the operating frequency is around to electrical wiring, even if this is at
36 kHz. Note, however, that for a mains potential. Most standard
given SMPS, the frequency may oscilloscopes have a common W=P*t (3)
vary with changes in the line ‘ground’ connection on the inputs
voltage and loading. which has to be used as a common The result is expressed in watt-
reference point for all input signals seconds (Ws), also known as joules
(and which is connected to the (J). The value in joules can be
rescaled into kWh, whereby These clamps eliminate the need to
1 kWh = 1000 Wh = open up the circuit loop when the load-resistance RL (see figure
1000 * 3600 Ws = 3.6 * 106 Ws making measurements and provide 10). We can now measure the
good isolation between any ‘live’ voltage across this sensing resistor
Some ScopeMeter models also
wiring and the test instrument. and calculate the current from
contain a function for calculating
When working on power circuits, Ohms‘ law. Adding the resistor is a
the total power accumulated over a
this is definitely the safest way to one-time modification, which is
period of time, which is then
measure the current. more convenient than repeatedly
selected using the cursors. Power
If a current clamp is used for opening up the circuit loop.
over time equals energy, and we
measuring small currents and the
can read this on the ScopeMeter as
sensitivity of the clamp in If Rs is more than 10 times smaller in
watt-seconds directly.
insufficient, the effective sensitivity value than the load RL, less than a
can be increased by feeding percent of the energy will be handled
multiple turns of the wire through by the series resistance, and thus the
the clamp. The actual current is error in the power measurement that
now the measured current divided results from adding the resistor will
by the number of turns. be less than 1%.

Sometimes, however, it’s not so


simple to cut the wiring of an Conclusion
existing circuit loop in order to Power measurements on low-
include a current meter, for example frequency linear systems can be
Figure 9: Power curve and energy
measurement. when all wiring is part of a printed performed using a DMM. The
circuit board. A possible way to measurement of power in electronic
Looking at figure 9, we can see that bring the current meter into the systems where waveforms are more
between the two cursors, i.e. within loop of a low-power circuit would complex and frequencies often much
a single cycle of the mains voltage, then be to set up the DMM for higher than the mains frequency
an energy of 10.09 Ws is delivered current measurement and select the requires more sophisticated tools.
to the lights. A cycle of the mains in highest current-range provided. The Fluke 190 Series of ScopeMeters
this case takes 20 ms. Now connect the meter over the are well equipped to make these
contacts of the on/off-switch. If the measurements, and can even make
Per second this leads to a power
switch is left open, the meter will fast peak-power measurements to
consumption of:
close the loop and read the current, determine the power handling of fast
10.09 Ws / 20ms = while no wiring needs to be electronic components, for instance
10.09 * 50 W*s/s = 505 W interrupted or modified. in switched-mode power supplies.

And over a time-span of one hour,


this equals a total energy I
consumption of: A
Sw Fluke. Keeping your world
up and running.
505 W * 1 h = 505 Wh =
0.505 kWh = 1.8 MWs U1 RL V
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Practical set-up hints
Fluke Europe B.V.
The best way to measure a current PO Box 1186, 5602 BD
in a circuit loop is with a current RS Eindhoven, The Netherlands

clamp. These are commercially Figure 10: Adding a current sensing resistor to For more information call:
the circuitry to allow current measurement. In the U.S.A. (800) 443-5853 or
available for AC and for DC+AC Fax (425) 446-5116
measurements, and for various In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
current ranges. The Fluke 80i-100s, If no switch is provided, or if the In Canada (800)-36-FLUKE or
Fax (905) 890-6866
for instance, measures DC and AC DMM has no current-measuring From other countries +1 (425) 446-5500 or
currents from 0.1 to 100A; the Fluke capability, we may also add a Fax +1 (425) 446-5116
Web access: http://www.fluke.com
i1010 can even be used up to current sensing resistor of a known
1000A AC and/or DC. value Rs to the circuit, which needs
to be small in value compared with ©2003 Fluke Corporation. All rights reserved.
Printed in the Netherlands.
10/2003 Pub-ID 10654-eng Rev. 01

Das könnte Ihnen auch gefallen