a Fluke ScopeMeter®
190 Series
Application Note
Basics of MRI
Once the nuclei are aligned by this
NMR is based on the fact that the
static magnetic field, an alternating
atoms of many chemical elements
magnetic field in a direction
and their isotopes have some
perpendicular to the main field will
degree of magnetic orientation.
make the protons precess around
Under normal conditions, the
the axis of their static orientation.
individual magnetic fields of
When this occurs, the atoms turn
individual atoms show a random
orientation, and the net field is zero
(Figure 1).
When placed in a very strong static
Given this resonance, the nuclei can
magnetic field, the magnetic fields
be thought of as resonators that
of the individual atoms line up,
absorb energy from the applied
resulting in a net magnetization
RF-field.
(Figure 2). The magnetic fields used
Nuclei of different elements and
to accomplish this are around 0.1 to
different isotopes all have different
4 tesla (1 T = 10000 gauss) with a
resonance frequencies.
tendency to the ever stronger fields.
For as long as the RF-field is
Figure 2: When placed in a strong magnetic
field, the magnetic orientation of nuclei align
applied, all nuclei will resonate in
with the external field. phase, but once this stimulus ends
the individual nuclei will lose
out to have a resonance frequency synchronism, or de-phase, and
in the RF-range. For instance the return to their original position
hydrogen atom (1H) has a resonance aligned with the static magnetic
frequency of approximately field. This process of returning to
42.58 MHz when placed in a static the original orientation takes a
magnetic field of 1 T. length of time that is again
characteristic of each different type
Figure 1: Normal randomly arranged magnetic of nucleus.
orientation of nuclei.
By measuring the resonance display electrical signals of various
frequencies and the decay times, frequencies and that are not
the individual elements present in influenced by magnetic fields (in
the body can be determined. contrast to a traditional CRT-based
oscilloscope) since the display used
is an LCD.
Oscilloscopes used in MRI
For the proper functioning of the
ScopeMeters used in MRI
MRI-equipment, it is essential that
Figure 5: A short duration burst is captured,
the RF stimuli are switched on and The Fluke ScopeMeters are but the envelope is more difficult to identify
off in a well controlled manner, and available in different bandwidth (Fluke 199C)
with good reproducibility. During and with different sample rates (see
development and adjustment of the table 1). Given the RF- frequencies frequencies, the effects of aliasing
equipment, a proper representation commonly used in MRI of 42.58, may also influence the quality of
of these RF-bursts is most 63.8 and 127 MHz, the Fluke 190 the displayed bursts.
important. For this purpose an Series ScopeMeters are the most
oscilloscope is a most valuable appropriate instruments for test and Here, some of the additional
piece of test equipment, if it can alignment purposes. functions found in the ScopeMeters
display the RF carrier frequency and
is capable of showing the Fluke type number Bandwidth Max. sample rate (single shot)
envelopes of the bursts properly. Fluke 123 20 MHz 25 MS/s
These envelopes show you Fluke 124 40 MHz 25 MS/s
Smart averaging
Smart averaging shows the
averaged curve of repeated signal
parts, as well as incidental
deviations.
Figure 7: The Fluke 199C is used here with
‘Digital Persistence’ to capture a natural enve- Extended offset
lope of a short duration burst of an RF-carrier.
Allows vertical zoom-in for study of
small details of the larger amplitude
traditional high-frequency signal or when a larger DC-
oscilloscope with its phosphor component is included.
ensuring the slow trace fade-out.
An additional benefit of the Digital Medical models available are the
Persistence mode is that it Fluke 199CM, Fluke 199BM,
Fluke 196CM and Fluke 196BM.
See techical data sheet for full
details.