Sie sind auf Seite 1von 3

TECHNICAL INFORMATION PATIENT MONITOR

Type : Philips INTELLIVUE MX700 Standard + X2


Parameter : ECG, NIBP, SpO2, Temp, HR/RR
Brand : PHILIPS
Country of Origin : Germany

Physical
Weight : 11 kg (24.3 lb)
Size : 393 x 321 x 233 mm

Power Supply Specification


Power Consumption : <200 W average
Line Voltage : 100 to 240 V
Current : 1.9 to 0.9 A
Frequency : 50/60 Hz

Monitor Performance
Display : 15” color LCD (TFT)
Resolution : 1280 x 768 (WXGA)
Pixel size : 0.261 mm x 0.261 mm
Waveforms : standard 8, optional 12

Indicators/sounds
Audible feedback for user input
Prompt tone
QRS tone or SpO2 modulation tone
4 different alarm sound
Remote tone for alarms on other beds in network
Tone for Timer expired

Trends
Resolution : MX700 supports up to 48h for up to 24 or 32 numerics

Information : Multiple choices of number of numerics, resolution and duration


depending on trend option and application area.

Events
Information : Trigger condition and time, event classification
and associated detailed view of episode data
Capacity : 25 or 50 events for either 8 or 24 hours

IntelliVue X2 Module (ECG, HR, RR, NIBP, SpO2, T/IBP)


Display : Touchscreen, 3.5 in TFT flat panel display with QVGA
Weight : 1.2 kg (2.6 lb)

Power Specification
Power consumption : <12 W average
<30 W while battery is loading
Operating Voltage : 36 to 60 V DC floating

Battery Specifications
Operating Time : Basic monitor configuration: 2.5 hours (with new, fully charged
battery at 25°C)
Charge Time : When X2 is off : 2 h.
When X2 is in use and connected to Monitor
Without extensions: 12 h approx.
When X2 is in use and connected to the external power: 12 h approx.

Spo2 Performance
Sensor : Wavelength range: 500 – 1000 nm
Emitted Light Energy: ≤15 mW
Pulse rate range : 30 to 300 bpm
Resolution : 1 bpm
Accuracy : ±2% or 1 bpm, whichever is greater

NIBP performance
Measurement range : Systolic--> Adult : 30 to 270 mmhg
Diastolic --> Adult: 10 to 245 mmhg
Mean--> Adult : 20 to 255 mmhg
Pulse rate range --> Adult: 40 to 300 bpm
Accuracy : Max. Std. Deviation : 8 mmHg (1.1 kPa)
Max. Mean Error : ± 5 mmHg (± 0.7 kPa)
Pulse Rate Measurement Accuracy : 40 to 100 bpm : ± 5 bpm
Cuff iflation time : Typical for normal adult cuff:Less than 10 seconds
Typical for neonatal cuff: Less than 2 seconds
Auto Mode Repetition Times : 1,2,2.5,3,5,10,15,20,30,45 minutes or 1, 2, 4, 8, 12, 24 hours

ECG performance
Range : Adult/pedi: 15 to 300 bpm
Neo range: 15 to 350 bpm
Note: for rates equal to or less than 15 bpm,
the displayed heart rate is 0
Accuracy : ±1% of range
Resolution : 1 bpm
Sensitivity : ≥200 µVpeak

Respiration performance
Range : Adult/pedi: 0 to 120 rpm
: Neo: 0 to 170 rpm
Accuracy : at 0 to 120 rpm ±1 rpm
at 120 to 170 rpm ±2 rpm

Resolution : 1 rpm
Accuracy : ±1 rpm to ±2 rpm
Bandwith : 0.3 to 2.5 Hz (-6 dB)
Noise : Less than 25 mΩ (rms) referred to the input

Temperature Performance
Range : - 1 to 45 C ( 30 to 113 F )
Resolution : 0.1 C
Accuracy : 0.1 C
Average time constant : less than 10 second

IBP Performance
Range : -40 to 360 mmHg
Pulse Rate : Range : 25 to 350 bpm
Accuracy : ±1% Full Range
Resolution : 1 bpm
Sensitivity : 5 μV/V/mmHg (37.5 μV/V/kPa)
Frequency Response : dc to 12.5 Hz or 40 Hz
Zero Adjustment : Range : ±200 mmHg (±26 kPa)
Accuracy : ±1 mmHg (±0.1 kPa)
Drift : Less than 0.1 mmHg/°C (0.013 kPa/°C)
Transducer : Load Impedance : 200 to 2000 Ω (resistive)
Output Impedance : ≤3000 Ω (resistive)

Accessories Included
ECGTrunk cable 5 leads : 1 pcs
ECG electrode disposable : 1 pack
NIBP Tubing : 1 pcs
NIBP Cuff Adult : 1 pcs
Reusable Spo2 sensor adult : 1 pcs
Temperature probe : 1 pcs IBP

Das könnte Ihnen auch gefallen