Beruflich Dokumente
Kultur Dokumente
measurements
Introduction the subject in a useful paper. Their the amount of light scattered by the
paper covers both method sample; and the reproducibility and
When undertaking particle size development and method validation, precision of the measurement.
measurements it is important to whereas this note is mainly concerned
validate the analysis method to with the latter.
ascertain both its robustness and Sampling
integrity. This allows the key variables Note that following the procedures One of the first considerations in
associated with variability in the outlined here cannot guarantee that particle size analysis is whether the
results to be determined and then an auditing body will always approve sample selected for analysis is
controlled as part of the measurement a given method. Rather the minimum representative of the bulk material. If
procedure. The implementation of a amount of work expected to be the sample is not representative the
validation study is an absolute performed is outlined. Any additional results obtained may be atypical and
requirement in the pharmaceutical studies will only strengthen the case the experiment is a pointless exercise.
industry and of growing significance for the candidate method. During transit of the sample, settling
throughout the manufacturing sector can occur - in powders large particle
where the need for result consistency tend to settle at the top of the sample
Method Validation container, whereas suspensions show
is important to ensure efficient
production and quality control. The development of a validated the reverse within large particles
method should be carried out using undergoing sedimentation. In each
Validation is described by the US an instrument that has validated case the material must be sampled in
Food and Drug Administration (FDA) software and is regularly tested to such a way as to remove the bias
as “establishing documentary confirm its performance. Validated caused by these processes.
evidence which provides a high software has lifecycle documentation
degree of assurance that a specific detailing its development and Research has shown that use of a
process will consistently produce a maintenance and should be spinning riffler is the most
product meeting its predetermined numerically validated using a peer reproducible method of obtaining a
specification and quality attributes” package such as Microsoft Excel. The representative sample for powder
[1]. The FDA Guidance to Industry instrument installation should conform samples when compared with other
document goes on to state that for to the manufacturers Installation methods [7] (table 1). Riffling works
particle sizing methods (including Qualification (IQ). The recommended best for free-flowing particles but can
laser diffraction) that the intermediate manufacturer’s Operational take a great deal of time if a large
precision and robustness of the Qualification (OQ ) should also be amount of powder is to be handled. If
method should be studied. carried out at least yearly, with the the powder is not free-flowing then
instrument performance being tested particle segregation is minimized and
Users of laser diffraction instruments techniques such as scoop sampling
for particle characterization using secondary standards on a
applications have a wealth of routine basis between each OQ visit.
information on the theory behind the In the current regulatory environment, Table 1: Variability Associated with
technology as well as guidance on it is also essential that the software is different sampling methods [7].
both sampling and dispersion [2,3,4]. compliant to 21 CFR part 11, the
FDA’s rule regarding the use of Method Relative Standard
In 1999 the Pharmaceutical Analytical Deviation (%)
Sciences Group (PASG) [5] laid down electronic records and signatures.
some guidelines as to the process When validating a laser diffraction Cone & Quartering 6.81
which should then be followed during method for particle characterization, Scoop Sampling 5.14
method validation. The work of the the following main variables must be
PASG group is built upon here with Table Sampling 2.09
considered: sampling; sample
practical examples of how validation preparation; instrument range; Chute Riffling 1.01
can be carried out. Lerke and Adams appropriateness of the technique;
[6] have also published their ideas on Spinning Riffling 0.13
robustness of the analytical method;
Deionised Water
Tap Water Sample Dispersion
As part of method development the
0.5 process of sample dispersion should
have been explored and understood.
In the case of dry measurements
0.4 users are required to understand how
Dv50 / Microns
Figure 4: Effect of varying the stirrer rate on the result obtained for a typical
lactose.
4.4
obscuration range. At extremely low
obscurations results with large COVs
may be obtained (owing to the low
signal to noise ratio) whilst at
4.2 extremely high concentrations the
results obtained may be smaller than
expected due to the effects multiple
4.0 scattering. It is suggested that
obscurations of 5, 10, 15, 20 and 25%
are investigated in the same way that
the measurement duration test is
3.8 performed, with the acceptable COVs
0 5 10 15 20 25 30
being similarly specified.
Obscuration (%)
An example of how multiple scattering
Figure 5: Change in Dv50 as a function of obscuration for a pharmaceutical
affects the results obtained is given in
powder.
figure 5 for a pharmaceutical powder.
It should also be noted that the Confirmation of refractive index At low obscurations the result is
application of ultrasound can cause As part of the method development constant. However, beyond 10%
agglomeration in some cases. If this work, the choice of refractive index obscuration multiple scattering causes
occurs then it may be necessary to should be examined. Index matching the reported size to decrease. In this
look at the use of different dispersing fluids can be used to provide case a measurement obscuration of
agents to bring about stability. Full experimental evidence of the real around 7.5% would provide a robust
details of how wet method refractive index. The Fraunhofer method where small changes in the
development may be carried out are approximation should not be used if obscuration from sample to sample
given elsewhere [9]. there are particles smaller than 40 would not overly affect the results.
times the wavelength of light used for
Pump and Stir rates the measurement (25 microns for a Reproducibility
The pump and stir speeds used system using a HeNe red laser light Bell et al defined reproducibility as an
during a wet measurement should be source) present in the distribution1 as indicator of precision between
examined as part of method it may erroneously report the laboratories [5]. It can indeed show
development. The chosen conditions presence of fine material. ISO13320-1 this, but experience at Malvern
should be capable of suspending all provides guidance as to when the Instruments has shown that it is far
the material without causing air optical properties will be critical in more likely to provide a measure of
bubble formation (a particular problem configuring the laser diffraction the effectiveness of the chosen
if surfactants are being used). analysis [2]. sampling method. It can also be used
to flag differences between different
Figure 4 shows how the result instruments (be it of the same or
Linearity and Obscuration
obtained for a lactose sample varies different models). Another
according to the stirrer settings. As Assessing the linearity of a sizing
technique is not considered to be part consideration will be environment
can be seen, the result reaches a conditions experienced in different
plateau above 2000rpm. It is at this of the method validation as sizing
techniques seldom display a linear laboratories, especially where
point that all of the material is saturated solutions are used as
correctly suspended and dispersed. response as a function of particle
size. However, it is important to dispersants, as small temperature
Sample sedimentation causes the differences can cause result changes
results to be smaller than expected at consider how the sample obscuration
affects the measurement. Obscuration due to either particle dissolution or
stirrer rates below 2000rpm. recrystallization.