Sie sind auf Seite 1von 7

Method validation for laser diffraction

measurements
Introduction the subject in a useful paper. Their the amount of light scattered by the
paper covers both method sample; and the reproducibility and
When undertaking particle size development and method validation, precision of the measurement.
measurements it is important to whereas this note is mainly concerned
validate the analysis method to with the latter.
ascertain both its robustness and Sampling
integrity. This allows the key variables Note that following the procedures One of the first considerations in
associated with variability in the outlined here cannot guarantee that particle size analysis is whether the
results to be determined and then an auditing body will always approve sample selected for analysis is
controlled as part of the measurement a given method. Rather the minimum representative of the bulk material. If
procedure. The implementation of a amount of work expected to be the sample is not representative the
validation study is an absolute performed is outlined. Any additional results obtained may be atypical and
requirement in the pharmaceutical studies will only strengthen the case the experiment is a pointless exercise.
industry and of growing significance for the candidate method. During transit of the sample, settling
throughout the manufacturing sector can occur - in powders large particle
where the need for result consistency tend to settle at the top of the sample
Method Validation container, whereas suspensions show
is important to ensure efficient
production and quality control. The development of a validated the reverse within large particles
method should be carried out using undergoing sedimentation. In each
Validation is described by the US an instrument that has validated case the material must be sampled in
Food and Drug Administration (FDA) software and is regularly tested to such a way as to remove the bias
as “establishing documentary confirm its performance. Validated caused by these processes.
evidence which provides a high software has lifecycle documentation
degree of assurance that a specific detailing its development and Research has shown that use of a
process will consistently produce a maintenance and should be spinning riffler is the most
product meeting its predetermined numerically validated using a peer reproducible method of obtaining a
specification and quality attributes” package such as Microsoft Excel. The representative sample for powder
[1]. The FDA Guidance to Industry instrument installation should conform samples when compared with other
document goes on to state that for to the manufacturers Installation methods [7] (table 1). Riffling works
particle sizing methods (including Qualification (IQ). The recommended best for free-flowing particles but can
laser diffraction) that the intermediate manufacturer’s Operational take a great deal of time if a large
precision and robustness of the Qualification (OQ ) should also be amount of powder is to be handled. If
method should be studied. carried out at least yearly, with the the powder is not free-flowing then
instrument performance being tested particle segregation is minimized and
Users of laser diffraction instruments techniques such as scoop sampling
for particle characterization using secondary standards on a
applications have a wealth of routine basis between each OQ visit.
information on the theory behind the In the current regulatory environment, Table 1: Variability Associated with
technology as well as guidance on it is also essential that the software is different sampling methods [7].
both sampling and dispersion [2,3,4]. compliant to 21 CFR part 11, the
FDA’s rule regarding the use of Method Relative Standard
In 1999 the Pharmaceutical Analytical Deviation (%)
Sciences Group (PASG) [5] laid down electronic records and signatures.
some guidelines as to the process When validating a laser diffraction Cone & Quartering 6.81
which should then be followed during method for particle characterization, Scoop Sampling 5.14
method validation. The work of the the following main variables must be
PASG group is built upon here with Table Sampling 2.09
considered: sampling; sample
practical examples of how validation preparation; instrument range; Chute Riffling 1.01
can be carried out. Lerke and Adams appropriateness of the technique;
[6] have also published their ideas on Spinning Riffling 0.13
robustness of the analytical method;

1 Mastersizer 2000 Application Note MRK671-01


may yield a representative sample. processes, sample preparation should development rather than as part of the
take this into account in order to avoid method validation process. Details of
For slurries it is important to overcome the break-up of agglomerated the method development process are
sedimentation by re-suspending the particles. In either case, the given elsewhere [8,9].
sample. This can often be achieved dispersion medium – whether air or a
by simple stirring. However, use of liquid – should not cause irreversible
certain stirrers, such as magnetic Detection Limit and Range
changes to the particle size through
fleas, can lead to the large particles Assessment of the detection limit of
processes such as dissolution, milling
being thrown to the outside of the the laser diffraction technique is not
or aggregation.
container (in a similar way to required as part of method validation.
hydrocyclone separation) where they For both dry and wet measurements it However, it is important that the
are not sampled. will be important to understand how dynamic range of the particle-sizing
the dispersion energy affects the instrument covers the size range of
Sample Preparation reported particle size. For wet the sample being tested. This
measurements the addition of normally does not present a problem
The PASG group defined sample for most pharmaceutical samples as
surfactants and additives and the use
preparation as “the pre-treatment and modern laser diffraction
of sonication for dispersion must be
the presentation of the sample to the instrumentation can cover a size
investigated. Comparison of the state
measuring technique in a meaningful range from 20nm to 2000µm in a
of the powder before and after
manner” [5]. Selection of the correct single measurement. Older
dispersion using microscopy can then
method will therefore depend on the instrumentation, however, may have
be used to assess if irreversible
interests of the user. Where, for to use many lenses in order to cover
particle break-up has occurred. In the
example, the primary particle size is the same dynamic range. Here the
case of dry powder measurements
important, correct dispersion of the lens that covers the largest proportion
the effect of dispersion pressure must
sample will be important. This may be of the particle size distribution should
be ascertained, with the correct
the case when control of the solubility be used. Alternatively the result from
pressure being selected by comparing
of the powder is important in defining two lenses can be blended together,
the results against wet dispersion.
its functionality. If the natural although this is not recommended
agglomerated state is of interest, as is Dispersion parameters are normally since the result depends on the
the case in some granulation assessed as part of method mathematical efficacy of the blending
routines.

0.66% 1.02% Specificity


0.98% Specificity, in terms of whether or not
0.58%
22 the technique is appropriate to the
material under analysis, should be
addressed as part of method
development and does not have to be
Dv50 / microns

revisited for method validation. It is, of


21 course, true that different sizing
techniques display different
sensitivities and will therefore provide
0.62%
different results for the same sample.
Selection of an appropriate technique
20 depends on what is of interest – for
instance is the detection of a small
amount of over-sized material
required or does the technique need
19
to differentiate between different
0 2 4 6 8 10 12 14 16
particle types? Laser diffraction
Measurement Time / sec provides a good method for assessing
small changes in the size distribution
Figure 1: Variation in the Dv50 as a function of measurement duration. The in this regard. However, it is difficult to
error bars represent the COV over ten measurements. differentiate between the different

2 Mastersizer 2000 Application Note MRK671-01


80 COV limits are doubled. The
Dv10 increased variability allowed within
Dv50
Dv90
ISO13320 for fine particles reflects
the fact that dispersion is important
60 within this size range. Note that
typically COV values of less than 3%
are achievable across the
measurement range if both dispersion
Size / Microns

40 and sampling are controlled.


Figure 1 shows an example of how
the Dv50 reported for a lactose
20
excipient varies with measurement
duration. In this case measurement
duration of 7 sec was chosen.
Although the COV is low for the 2 sec
0 measurement, the Dv50 is
significantly smaller than for the other
measurements, suggesting that the
0 2 4 6 8 10
large particles were not correctly
Time / min
sampled using such a short
Figure 2: Lactose sample stability measurement summary. measurement time.

components within pharmaceutical short measurements time can be Measurement Stability


dosage forms using the technique. used. If polydisperse or coarse
particle size distributions are analyzed In order to determine whether
then longer measurement times may samples are stable over the period of
Robustness analysis and not subject to
be required in order to ensure that a
The robustness of an analytical agglomeration, de-agglomeration or
representative volume of larger
method is an indication of its ability to dissolution, it is necessary to monitor
particles has been sampled.
remain unaffected by small variations the particle size distribution at known
in the test parameters and so The suggested procedure for time points.
provides assurance of its reliability assessing the correct measurement
during routine use. The method duration is to carry out ten repeat A sample should be prepared in
robustness should be considered measurements using measurement accordance with the method under
before repeatability, reproducibility times of 2, 5, 7, 10 and 15 seconds. investigation. It is recommended that
and intermediate precision are The individual and mean readings for at least five repeat measurements at
assessed. each measurement time can be over- the previously established duration be
plotted and any shift in particle size taken at varying times after sample
Measurement duration and dispersion (1, 3, 5, 7 and 10 minutes).
distribution noted. The appropriate
measurement stability are the two The mean and COV for the Dv10,
duration period can be selected by
main variables to be considered as Dv50 and Dv90 values should then be
looking at the coefficient of variation
part of a robustness study. Others, determined and should be shown to
(COV) for the median particle size
such as air pressure (dry be within the limits of ISO13320 [2].
(Dv50), Dv10 and Dv90.
measurements) and pump/stir rates
(wet measurement) are normally The COV should be within limits of Typical measurement stability results
considered as part of method acceptability laid down in ISO13320 obtained for a lactose sample are
development, but are briefly described [2]. This states that for samples with a show figure 2. From these results it is
here. median size of greater than 10 µm, clear that the lactose sample is stable
the COV should be less than 3% for whilst in suspension. It was decided
cut-off values close to the centre of from these results that measurements
Measurement Duration
the distribution (e.g. Dv50) and less should be taken after 1 minute in
The duration of each laser diffraction order to give the sample time to
measurement should be set so as to than 5% for cut-off values towards the
limits of the distribution (e.g. Dv10 equilibrate.
ensure that representative sampling
has been achieved. For narrow and Dv90). For samples with a An example of how the dispersion
distributions or fine particles relatively median size of less than 10 µm, the conditions can yield poor

3 Mastersizer 2000 Application Note MRK671-01


measurement stability is shown in (and therefore time). The results water be observed. Obviously the
figure 3. Here, the particle size of an obtained for single measurements are precision of measurements carried
emulsion product dispersed in tap similar. Only by assessing the out in tap water will be poor in this
water and deionised water is shown measurement stability over time can case and deionised water should be
as a function of measurement number the instability of the dispersion in tap used.

Deionised Water
Tap Water Sample Dispersion
As part of method development the
0.5 process of sample dispersion should
have been explored and understood.
In the case of dry measurements
0.4 users are required to understand how
Dv50 / Microns

the measured particle size varies with


the air pressured selected for
dispersion [2]. A suitable pressure is
0.3 one at which dispersion is achieved
without milling of the particles. In the
case of pharmaceutical materials, the
sample under test can be friable and
0.2 may be milled to a finer particle size if
the dispersion air pressure is set too
high. Often both dispersion and
milling occur simultaneously (which
0.1 leads to a broadening of the
0 2 4 6 8 10 distribution) [10]. The best way of
Measurement Number
proving that no attrition is occurring is
Figure 3: Emulsions sample stability measurements. to achieve near–identical results for
both wet and dry dispersion [2]. Full
details of how dry method
development may be carried out are
Dv10
Dv50
given elsewhere [8].
60 Dv90
For measurements made using wet
dispersion the role of ultrasound in
assisting dispersion must be
understood [2]. Using high-energy
Particle Size / Microns

ultrasound can fracture some


40
crystalline materials, although this is
extremely unusual. As part of method
development the affect of varying the
duration and power of ultrasound on
the particle size distribution should be
20
examined. Ideally measurements
should be taken before, during and
after ultrasound to examine what
effect sonication has on the
robustness of the measurement.
0
Micrographs should also be obtained
to ensure that particle break-up has
0 500 1000 1500 2000 2500 3000 not occurred during dispersion.
Pump Rate / rpm

Figure 4: Effect of varying the stirrer rate on the result obtained for a typical
lactose.

4 Mastersizer 2000 Application Note MRK671-01


is a measure of the amount of light
scattered by the sample and
4.8
correlates with the concentration of
material present within the
measurement zone. For most particle
4.6 size distributions the reported particle
size should be independent of the
measurement obscuration for a wide
Dv50 / Microns

4.4
obscuration range. At extremely low
obscurations results with large COVs
may be obtained (owing to the low
signal to noise ratio) whilst at
4.2 extremely high concentrations the
results obtained may be smaller than
expected due to the effects multiple
4.0 scattering. It is suggested that
obscurations of 5, 10, 15, 20 and 25%
are investigated in the same way that
the measurement duration test is
3.8 performed, with the acceptable COVs
0 5 10 15 20 25 30
being similarly specified.
Obscuration (%)
An example of how multiple scattering
Figure 5: Change in Dv50 as a function of obscuration for a pharmaceutical
affects the results obtained is given in
powder.
figure 5 for a pharmaceutical powder.
It should also be noted that the Confirmation of refractive index At low obscurations the result is
application of ultrasound can cause As part of the method development constant. However, beyond 10%
agglomeration in some cases. If this work, the choice of refractive index obscuration multiple scattering causes
occurs then it may be necessary to should be examined. Index matching the reported size to decrease. In this
look at the use of different dispersing fluids can be used to provide case a measurement obscuration of
agents to bring about stability. Full experimental evidence of the real around 7.5% would provide a robust
details of how wet method refractive index. The Fraunhofer method where small changes in the
development may be carried out are approximation should not be used if obscuration from sample to sample
given elsewhere [9]. there are particles smaller than 40 would not overly affect the results.
times the wavelength of light used for
Pump and Stir rates the measurement (25 microns for a Reproducibility
The pump and stir speeds used system using a HeNe red laser light Bell et al defined reproducibility as an
during a wet measurement should be source) present in the distribution1 as indicator of precision between
examined as part of method it may erroneously report the laboratories [5]. It can indeed show
development. The chosen conditions presence of fine material. ISO13320-1 this, but experience at Malvern
should be capable of suspending all provides guidance as to when the Instruments has shown that it is far
the material without causing air optical properties will be critical in more likely to provide a measure of
bubble formation (a particular problem configuring the laser diffraction the effectiveness of the chosen
if surfactants are being used). analysis [2]. sampling method. It can also be used
to flag differences between different
Figure 4 shows how the result instruments (be it of the same or
Linearity and Obscuration
obtained for a lactose sample varies different models). Another
according to the stirrer settings. As Assessing the linearity of a sizing
technique is not considered to be part consideration will be environment
can be seen, the result reaches a conditions experienced in different
plateau above 2000rpm. It is at this of the method validation as sizing
techniques seldom display a linear laboratories, especially where
point that all of the material is saturated solutions are used as
correctly suspended and dispersed. response as a function of particle
size. However, it is important to dispersants, as small temperature
Sample sedimentation causes the differences can cause result changes
results to be smaller than expected at consider how the sample obscuration
affects the measurement. Obscuration due to either particle dissolution or
stirrer rates below 2000rpm. recrystallization.

5 Mastersizer 2000 Application Note MRK671-01


give a pooled mean and a pooled Table 3: Lactose results obtained for a
Table 2: Variation in the results RSD (which should be <3%). second analyst.
obtained for seven separate
The reproducibility data obtained for a Sample D(v,0.1) D(v,0.5) D(v,0.9)
scooped-sampled lactose samples.
second analyst analyzing the lactose Number µm µm µm
Sample D(v,0.1) D(v,0.5) D(v,0.9) sample mentioned above is shown in
Number µm µm µm Table 3. From this a pooled mean and
COV can be determined across both 1 1.06 22.92 61.01
1 1.22 23.68 63.23 operators and is shown in Table 4. As
2 1.17 23.77 60.02 can be seen, the variation in the Dv50 2 1.08 22.08 56.54
is only just within the ISO13320 limits.
3 1.09 22.79 56.59 This is related to the method of 3 1.04 21.66 62.17
4 1.16 23.63 62.55 sampling used within this study. If the
powder had been sampled using a 4 0.97 22.55 60.23
5 1.11 22.26 59.68 spinning riffler this would have
6 1.18 22.78 65.36 improved the overall precision. The 5 1.04 22.74 57.98
broader acceptance limits employed
7 1.12 23.41 61.47 in USP<429> could also be applied in 6 0.99 23.58 59.86
Mean 1.15 23.19 61.27 this case if scoop sampling is the only
viable method for obtaining a sample. 7 0.95 22.11 62.78
%RSD 3.95 2.50 4.63

Other Considerations Mean 1.02 22.52 60.08

Validation terms such as


To assess reproducibility, a number of % RSD 4.79 2.83 3.69
quantification limit do not apply to
samples (at least five) should be
laser diffraction methods and are
taken from the same batch and tested
therefore outside the scope of this
in accordance with the method under
application note. Assessment of the
investigation. For each sample at
measurement accuracy is also not a Table 4: Pooled mean and COV
least five repeat measurements
requirement as, when measuring the values.
should be taken and the individual
particle size of non-spherical particles,
and average results should be
it is difficult to define exactly what
obtained. The sample-to-sample COV Mean Dv50 / 22.85
accuracy actually means. This is
for should then be determined and microns
because all particle sizing techniques
should be within the limits stated in
use approximations to derive a Standard
ISO13320 [2], or in certain cases 0.68
characteristic particle diameter. In the Deviation
within the limits of USP<429> [4].
case of laser diffraction there is also COV (%) 2.98
An example of the results obtained no calibration procedure involved in
using scoop sampling for a lactose deriving a particle size distribution that
excipient is shown in Table 2. In this would need to be assessed as part of
case the COV obtained was within the an accuracy study. However, it is such as laser diffraction. Instead,
limits expected on the basis of the important that the system’s analysts are only required to assess
sampling statistics shown in Table 1. performance is verified following the the intermediate precision and
manufacturers recommended OQ robustness in order to show that the
Intermediate Precision procedure. The requirements for selected method provides a
verification are also outlined in reproducible method of controlling
Assessing the Intermediate Precision ISO13320 [2]. product quality [1]. This application
requires users to determine the
note has outlined some of the
method variability when it is followed
Conclusions important variables that should be
by a second analyst or using a
assessed when considering the
second instrument (or both). This is The FDA’s guidance regarding the precision of laser diffraction
essentially a repeat of the validation of particle size analysis measurements. However, the
reproducibility test, thus similar COV methods states that the validation validation protocol applied to
limits should be applied. Both sets of concepts associated with other measurements carried out on a given
results should then be combined to analytical methodologies, such as material will ultimately depend of an
HPLC, do not transfer to techniques

6 Mastersizer 2000 Application Note MRK671-01


assessment of the risk associated laser diffraction" Paper 0208 14th
with any measurement errors and is International Congress of Chemical
therefore the responsibility of the and Process Engineering
analyst involved. "CHISA'2000", 27-31 August, Praha,
Czech Republic
References
[1] FDA Draft CMC Guidance for
Industry; Analytical Procedures and
Methods Validation; Section XI; Part F
(Aug 2000).
[2] ISO 13320-1 Particle Size Analysis
– Laser Diffraction Methods Part 1:
General Principles (1999)
[3] Jillavenkatesa, A, Dapkunas, S. J.
and Lum, L. S., (2001) Particle Size
Characterization Practice Guide,
N.I.S.T 960-1
[4] USP General Chapter <429>
“Light Diffraction Measurement of
Particle Size”, Pharmacopoeial
Forum, 28(4), 1293-1298 (2002).
[5] Bell, R., Dennis, A., Hendriksen,
B., North, N. and Sherwood, J.,
(1999) “Position paper on Particle
Sizing: Sample Preparation, Method
Validation and Data Presentation,
Pharmaceutical Technology Europe,
November 1999
[6] Lerke, S.A. and Adams, S.A,
(2002) “Development and Validation
of a Particle Size Distribution Method
for Analysis of Drug Substance”,
American Pharmaceutical Review,
Fall 2002
[7] Allen, T. Particle Size
Measurement, (1999) 5th edition,
Volume 1,p 38, Chapman and Hall,
London
[8] Developing a Method for
Malvern Instruments Ltd
Dry Powder Analysis, Malvern
Application Note MRK524. Enigma Business Park • Grovewood Road • Malvern • Worcestershire • UK • WR14 1XZ
Tel: +44 (0)1684 892456 • Fax: +44 (0)1684 892789
[9] Wet Method Development
for Laser Diffraction Malvern Instruments Worldwide
Measurements, Malvern
Application Note MRK561. Sales and service centers in over 50 countries for details visit www.malvern.co.uk/contact

[10] Rawle, A. F., (2000)


"Attrition, dispersion and
sampling effects in dry and wet
particle size analysis using more information at www.malvern.co.uk

7 Mastersizer 2000 Application Note MRK671-01

Das könnte Ihnen auch gefallen