Sie sind auf Seite 1von 2

The Comedy of Errors of Light here is no maximum value is

Booths and its Calibration mentioned, high lux values can


also cause problems.
Today I want to write about the
errors committed in calibration of So the Light Box makers bought
Light boxes by almost all ISO low cost Lux meters , that is mostly
17025 certified labs in India , and used in the film industry to
its acceptance by the auditors. measure lux at the Heroines face
and started measuring the light
ISO 17025 calls for very strict inside the cabinet to be 600 Lux
calibration , verification and and above and started selling
maintenance of instruments used certified Light boxes.
for testing.
The major problem is no one know
I strongly beleive that more by what lamps were generating the
design , users of light boxes have 600+ Lux , and it could be
been made to believe that Lux anything D65, F02 F11 etc.
value alone need to be calibrated
and the Spectral Power Distribution The auditors who were accredited
and Color Temperature ( the last 2 by NABL ( National accreditation
terms are seldom mentioned by Board for Testing and calibration
anyone leave alone understand it ). labs ) also went only by Lux values
and did not bother to check the
The reason for this is largely lack of SPD or the CCT.
knowledge of the need to to
measure these values not only by One very much acclaimed Auditor
the Indian companies that made even dismissed me saying that
low cost light boxes. Not only were SPD and CCT are not required to be
these people ignorant , due to their measured.
propaganda and the lack of
knowledge even among dyers , Lux So several Indian made light boxed
measurements were considered as had D55, CWF TL84 or any lamp
sufficient measurement. with 600 lux and got away as
calibrated light boxes.
The choice of Lux as a parameter
was in my view due to 2 reasons , This is the the biggest cause for
the measuring instrument is cheap debates in visual matches as well
and easy to achieve with as Grey scale values.
commercial Fl Lamps.
Comments and Recommendations.
Further ISO 105 A01also does not
say anything about the need to • Please note Light Boxes
measure SPD and CCT , it simply need to be calibrated for
says 600 Lux and above is Lux, SPD and CCT.
considered as good for Grey scale • Refer to AATCC EP 9 for
assessments , another mistake more details on what
must be CCT of the lamps
and the desirable Lux
values.
• CIE has full details of the
SPD of all the lamps
• Calibration has to be
carried out every 1000
hours of operation or 1
year which ever is earlier.
• Fluorescent Lamps cannot
be adjusted and need to
replaced with new ones if
the required values are
not achieved.
• Do not be taken in by
price, compliance to
standards and
specification are more
important, as Color is the
biggest property based on
which the Apparel
business is done. Right
from design to purchase
of a product by a
consumer.
• Good Light boxes show
the color as close as it is
in its original state and
there by reduces debates.
• Clean the Tubes daily, or
fiber keeps accumulating
on the tubes and over a
period of time changes
the Lux , CCT and SPD .
• Paint the Light Box
Munsell N5 or N7 as per
the manufacturers
instructions.