2218610
Revision 0

Gantry
Positioning Light Visor
Kit Installation

e        

    


GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

D THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING D IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
D DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
D FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.

D CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


AVERTISSEMENT D SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
D NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
D LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

D DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN


ENGLISCHER SPRACHE.
WARNUNG
D FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
D VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
D WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

D ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


AVISO D SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
D NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
D LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

a
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

D ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA


DISPONÍVEL EM INGLÊS.
ATENÇÃO
D SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
D NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
D O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

D IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE


SOLTANTO IN INGLESE.
AVVERTENZA
D SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
D SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
D NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

b
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should
be thoroughly read and understood by everyone who will use the equipment before you attempt to place this
equipment in operation. The General Electric Company, Medical Systems Group, will be glad to assist and
cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no
practical design of equipment can provide complete protection. Nor can any practical design compel the operator to
take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents,
and representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective material and devices are available. It is urged that such materials or devices be used.

    



 
All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required.

In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

c
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.

OMISSIONS & ERRORS

GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.

d
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

CAUTION

Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.

Devices not to be used near this equipment:

Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.

Keep power to these devices turned off when near this equipment.

Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.

e
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
blank 2218610

f
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

LIST OF EFFECTIVE PAGES

REVISION DATE PRIMARY REASON FOR CHANGE


0 . . . . June 10, 1998 . . . Initial Release

PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . . . 0 A ..................0 Tab 1 Tab 2 TBb 3
Title page rear . . blank B . . . . . . . . . . . . . . blank 1–1 . . . . . . . . . . . . . . . . 0 2–1 . . . . . . . . . . . . . . . . 0 3–1 to 3–5 . . . . . . . . . . 0
a to e . . . . . . . . . . . . . . 0 i...................0 1–2 . . . . . . . . . . . . blank 2–2 . . . . . . . . . . . . blank 3–6 . . . . . . . . . . . . blank
f . . . . . . . . . . . . . . blank ii . . . . . . . . . . . . . . blank

A
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
blank 2218610

B
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 – INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1


1-1 SPECIAL TOOLS (TO BE PREPARED BY FE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1-2 MANPOWER REQUIRED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1

SECTION 2 – MATERIAL LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1

SECTION 3 – INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1


3-1 LIGHT LEAKS CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3-2 POSITIONING LIGHT MODIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3-3 POSITIONING LIGHT ADJUSTMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–4
3-4 FUNCTIONAL CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–5

i
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
blank 2218610

ii
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

SECTION 1 – INTRODUCTION

The Gantry has miner light leaks from the mylor ring attached to the Gantry bore.

The option installs the light visor to the two outer positioning lights to prevent light leaks.

GANTRY REAR VIEW

LIGHT LEAKS

1-1 SPECIAL TOOLS (TO BE PREPARED BY FE)


D None

1-2 MANPOWER REQUIRED


One service engineer for 1 hour plus travel.

1–1
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
blank 2218610

1–2
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

SECTION 2 – MATERIAL LIST

Positioning Light Visor Kit Option Part No.


2218833

ITEM PART NO. DESCRIPTION QTY


1 2216838 Plate 2
2 2216839 Visor 2
3 N9306JS Screw M3 x 6 2
4 2218610 Instruction (This Manual) 1

2–1
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
blank 2218610

2–2
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

SECTION 3 – INSTALLATION

3-1 LIGHT LEAKS CHECK


1. Power ON the main switch of the Operator Console.

2. Turn ON the positioning light of the Gantry.

3. Verify that the Gantry has light leaks from the mylor ring.

4. Power OFF the main switch of the Operator Console.

3-2 POSITIONING LIGHT MODIFICATION


1. Turn OFF ALL switches and breakers at the Gantry rear base.

2. Remove the left side, right side, and front covers from the Gantry.

3. Locate the two outer positioning lights on the Gantry.

4. Remove the two positioning light attaching screws, then remove the positioning light from its support plate.

OUTER POSIITONING LIGHT (LEFT SIDE)

SCREWS

3–1
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

3-2 Positioning Light Modification (Continued)


5. Remove the stop screws for lenses, then remove the lenses from the plate.
Discard the plate.

STOP SCREWS

PLATE LENSES

6. First install the lens B onto the new plate, shipped with this kit, and push the lens B as far as it will go.
Align the adjusting knob of the lens B with holes for stop screws of the plate.

7. Secure the lens B stop screw.

8. Next install the lens A into the new plate as far as it will go.

9. Secure the lens A stop screw.

STOP SCREWS

ADJUSTING KNOB

ALIGN

LENSE B

PLATE LENSE A

3–2
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

3-2 Positioning Light Modification (Continued)


10. Install the positioning visor, shipped with this kit, onto the plate.

Note
Install the visor so that it is flushed with the lens B side of the plate.

LENS B SIDE OF PLATE FLUSHED

VISOR

SCREWS

11. Install the positioning light to the support plate on the Gantry.

POSITIONING LIGHT
SCREWS

3–3
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

3-3 POSITIONING LIGHT ADJUSTMENT


1. Power ON the main switch of the Operator Console.

2. Turn ON ALL switches and breakers at the Gantry rear base.

3. Turn ON the positioning light of the Gantry.

4. Adjust the light angle using the ‘D’ lever.

LIGHT
ANGLE
LIGHT D
ANGLE
d (A hex screw is installed inside this hole)
LIGHT
Before adjusting D, loosen d; after adjusting light
SCREEN angle using D, tighten d.

3–4
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
REV 0 2218610

3-4 FUNCTIONAL CHECK


1. Power OFF the main switch of the Operator Console.

2. Install the Gantry covers.

3. Power ON the main switch of the Operator Console.

4. Turn ON the positioning light of the Gantry.

5. Verify that the Gantry has NO light leaks from the mylor ring.

3–5
GE MEDICAL SYSTEMS GANTRY POSITIONING LIGHT VISOR KIT INSTALLATION
blank 2218610

3–6