Sie sind auf Seite 1von 6

Australian Institute of Radiography

The Radiographer 2009; 56 (3): 32–37 Literature review

Diagnostic reference levels as a quality assurance tool

KD Edmonds

Medical Physics Section, Medical Radiation Branch, Australian Radiation Protection and Nuclear Safety Agency,
Yallambie, Victoria 3085, Australia.
Correspondence keith.edmonds@arpansa.gov.au

Abstract The objective of diagnostic reference levels (DRL) in radiology is to assist in the optimisation of radiation dose
to patients, while maintaining diagnostic image quality, and to detect unusually high doses that do not contribute signifi-
cantly to the clinical outcome of a medical imaging examination. DRL have been in existence overseas for more than a
decade and its influence has contributed to a steady decline in dose for general radiography and fluoroscopic procedures.
High dose modalities such as CT and interventional procedures are increasing dramatically both locally and internation-
ally resulting in the unwanted outcome of a significant increase in population cumulative effective dose. This calls for
urgent dose reduction and dose constraint measures. Utilising DRL is one method of optimising patient dose. Some local
and international DRL dose levels for some common radiographic, interventional and CT examinations are presented as
a suggestion for the application of this methodology in Australian radiology practice.

Keywords: diagnostic reference levels, guidance levels, radiation dose, reference values, quality assurance.

Introduction of equipment. These levels are expected not to be exceeded for


The objective of establishing diagnostic reference levels (DRL) standard procedures when good and normal practice regarding
in diagnostic imaging (also previously known as Guidance Levels diagnostic and technical performance is applied.”36 It reinforces
or Reference Values1,2) is to provide radiology and other depart- the concept of references doses applying only to “standard” or
ments that use x-ray imaging with a convenient DRL dose com- “representative” patients. DRL therefore are not dose limits but
parison to ensure that radiation doses to patients are kept within a guide of good practice. It is not a dose constraint and the DRL
reasonable limits. values are not used for regulatory or commercial purposes. DRL
The main task of radiation protection is not only to mini- act as an investigation trigger if the numerical values are consis-
mise the stochastic risks but also to avoid deterministic injuries. tently exceeded.
Stochastic refers to effects whose probability increases with
Background
increasing dose and for which there is no threshold dose. Any
dose, no matter how small, has the potential to cause harm and The need for DRL
this becomes apparent years after the exposure. Examples are Patient exposures in diagnostic radiology are increasing at a
leukaemia and hereditary effects. Deterministic effects are those disquieting rate for certain radiographic, fluoroscopic and CT
in which the severity of the effect, rather than the probability, examinations. Regulla and Elder5 pointed out that data obtained
increase with increasing dose and for which there is a threshold from the United Nations Scientific Committee on the Effects of
dose. Examples are epilation, erythema and hematologic damage Atomic Radiation (UNCEAR) show that there are significant
and are known as early effects.3 differences in national radiation exposures and a very uneven
A DRL, as defined by the International Commission on distribution of patient doses among world population for the same
Radiological Protection (ICRP), is “a form of investigation level, or similar procedures. Mean annual x-ray effective dose of the
applied to an easily measured quantity, usually the absorbed dose population can vary by up to a factor of 60. In the United States,
in air, or tissue-equivalent material at the surface of a simple studies such as the Nationwide Exposure X-ray Trends (NEXT)
phantom or a representative patient.”4 surveys also showed that patient doses in radiology vary consider-
The ICRP recommends the establishment of reference levels ably from one facility to the next.6 Gray, et al.1 posed the question,
as a method of optimising the radiation exposure to patients. This “why one radiology facility should use an exposure that is 10,
is accomplished by comparison between the numerical value of 20 or 126 times greater than another facility to produce a radio-
the diagnostic reference level (derived from relevant national, graphic image?” Johnston and Brennan7 and Carroll and Brennan8
regional or local data) and the mean or other appropriate value also reported wide variations in patient doses for the same radio-
observed in practice for a suitable reference group of patients or a graphic examinations among hospitals in the UK and Europe.
suitable reference phantom. These patient doses are attributed to a wide range of factors such
Another definition by the Council of the European Union in its as type of image receptor, exposure factors, fluoroscopic times,
Council Directive 97/43 defines DRL as “dose levels in medical number of images, type of anti-scatter grid and level of quality
radiodiagnostic practices or, in the case of radio-pharmaceuticals, control as reviewed by Seeram,9 Bushong10 and Parry, et al.11
levels of activity, for typical examinations for groups of standard- In Australia, the Australian Radiation Protection and Nuclear
sized patients or standard phantoms for broadly defined types Safety Agency (ARPANSA) Code of Practice, Section 3.1.8
Diagnostic reference levels as a quality assurance tool The Radiographer ­33

(Radiation Protection Series No.14) states that “the Responsible 2002 also issued advice on the use of DRL (or guidance levels)
Person must establish a program to ensure that radiation doses in their safety standard series and included guidance levels for
administered to a patient for diagnostic purposes are: typical adult patient doses for general radiography, CT, fluoros-
1 Periodically compared with DRL for diagnostic procedures for copy and mammography.2,14,
which DRL have been established in Australia Many countries worldwide have now incorporated the European
2 If DRL are consistently exceeded, reviewed to determine Community Directive in national legislative documents.1Several
whether radiation has been optimised.”12 organisations currently providing guideline documents for estab-
In addition, the ARPANSA Safety Guide, Section 7.8 lishing DRL include the ICRP,4,20 the Health Protection Agency in
(Radiation Protection Series No.14.1), suggests that “as part of the UK,16,21 the Commission of European Communities22 and the
the QA program, patient dose surveys are undertaken periodically American College of Radiology.23
to establish that the doses are acceptable when compared with
Aim
published DRL.” It also recommends that accrediting bodies such
The overall aim of DRL are to better manage patient dose in
as RANZCR and the Australian Council on Healthcare Standards
diagnostic radiology using the principle of optimisation which is
consider including compliance with DRL for a core set of exami-
defined as exposure to radiation from justified activities should
nations. If the radiology department observes dose values consis-
be kept as low as reasonably achievable, social and economic
tently exceeding the DRL, then this warrants further investigation
factors being taken into account. The European Commission and
however some flexibility should be allowed if higher doses are
the ICRP provide a range of tools to achieve this.15,24
indicated by sound clinical judgement.13,14
Data obtained from patient dose surveys show that typical
However, at this point in time, there are no DRL published in
patient doses for the same type of x-ray examinations can vary
the Code of Practice or the Safety Guides. It would seem logical
considerably from one radiology practice to another. The estab-
therefore to use published values from the literature from exten-
lishment of DRL therefore is to give an indication of unusually
sive surveys in countries with similar healthcare settings e.g.
high values. The DRL are usually set at the third quartile value of
similar levels of education and training for imaging technologists,
the distribution of typical doses derived from dose surveys both
radiologists and similar provision of imaging equipment.15
nationally and internationally. Using the third quartile or 75th
History of DRL percentile is a compromise between being overly stringent and
National surveys of patient doses from x-ray examinations in overly complacent.3
Europe and the USA since the 1950s have demonstrated wide Essentially, if mean doses exceed a reference level dose an inves-
variations in doses between radiology departments and illustrated tigation should take place to establish the cause and take corrective
the need for quantitative guidance on patient exposure. It was only action, unless the dose was clinically justified. Reference doses were
at this stage that dose measurements to patients began in earnest. also used to provide a trigger for practices in need of investigation
National surveys in the USA and UK concentrated on measuring and hopefully lead to dose optimisation. ICRP 734 recommended
entrance surface doses with or without backscatter for common that DRL values be selected by professional bodies, be reviewed at
radiographic projections. The Nationwide Evaluation of X-ray regular intervals and be specific to a country or region.
Trends in the USA in the 1970s measured entrance skin expo- Wide variations in patient doses are to be expected and it
sure free-in-air for average exposure technique factors or used is only sensible to compare mean or median values, which is
a standard phantom. The NRPB national patient dose survey in less influenced by extreme outliers, on representative groups of
the UK in the 1980s measured entrance surface dose directly on patients to monitor trends with time, equipment or technique.
the surface of the patient (including backscatter) using thermolu-
minescence dosemeters. A European trial supporting the Quality Method
Criteria for Diagnostic Radiographic Images in 1991 used the From a practical perspective, the DRL should be expressed
same technique.37 as a readily measurable patient-related quantity for the specified
Dose guidelines began to appear in late 1980s. First was the procedure. For example,
USA, promoted by the Centre for Devices and Radiological Health 1 General radiographic examinations – either entrance skin dose
(CDRH) in conjunction with the Conference of Radiation Control (ESD) or the dose area product (DAP)
Program Directors Inc. Then in the UK it was conducted by the 2 Fluoroscopic examinations – dose area product (DAP)
National Radiation Protection Board in collaboration with relevant 3 CT examinations – computed tomography dose index ( CTDIw
professional bodies. Europe then followed with reference doses or CTDIvol ) and the dose length product (DLP).
incorporated into Working Documents by EC Study Groups.38 New CT scanners in accordance with Australian Standards,
International recommendations then appeared on how to AS/NZS 32002.4,25 should display the volume CTDIvol and/or the
measure and set reference dose levels based on the initiatives led DLP on the operator’s console after the selection of technique
by the USA and the UK. The ICRP Publication 60 first made factors and prior to the initiation of x-rays.13
mention of the concept of “investigation levels” in 1990 followed DRL used for film-screen technology should not necessarily be
by the current definition of DRL in ICRP Publication 73 in 19964 used for new digital radiography without prior adjustment.26
and the EC Medical Exposure Directive in 1997.36 Dosimetry methods
The United Kingdom introduced DRL in 1990 for common International guidance on patient dosimetry techniques for
diagnostic examinations based on a national patient dose survey x-rays used in medical imaging is published by the International
in the mid-1980s conducted by the NRPB, now known as the Commission on Radiation Units and Measurements in ICRU
Health Protection Agency (HPA). They are now based on the Report 74.27 This report contains advice on the relevant dosimet-
five-yearly reviews of the National Patient Dose Database and are ric quantities and how to measure or calculate them in a clinical
currently in their third review.16 setting which is directly applicable to the patient dose surveys
The International Atomic Energy Agency (IAEA) in 1996 and needed to estimate population exposure.
­34 The Radiographer KD Edmonds

There are various methods of recording dose levels. review of DRL for general radiography for adults (Table 1) and
Thermoluminescent dosimeters (TLD) are often used for plain fluoroscopy for adults (Table 2), Paediatric procedures (Table
film examinations and include dose contribution from backscat- 3), the UK 2003 CT survey (Table 4), and mammography (Table
ter if placed on the patient or phantom surface. The small TLD 5), be adopted and /or adapted in the Australian context as there
sachets are usually placed in the centre of the irradiated field on are currently no established national DRL.29 Some state regula-
the entrance surface of the patient or phantom. The TLD can be tors though have provided local DRL guidance on radiography,
stuck directly and unobtrusively to the patient’s skin with very fluoroscopy and CT.34 Dose values should be reviewed as com-
little interference in patient mobility or comfort. They do not puted/digital radiography becomes more widespread in order to
interfere with the examination or obscure important diagnostic minimise the detrimental influence of ‘exposure creep’.30 This
information on the radiographic image. They need to be calibrated phenomenon occurs after the change over from film-screen
with respect to radiation qualities used in diagnostic radiology. radiography to digital radiography where exposure factors may
TLD are also prone to some inaccuracy due to signal fade, non- actually increase in order to reduce image noise. Uncoupling of
linear response and dependency on beam energy.3 Ionisation display from acquisition in digital radiography introduces the
chambers are bulky and more difficult to attach to patients. The potential for systematic overexposure without necessarily com-
parallel plate ionisation chambers measure back scatter but the promising image quality.31 The wide exposure latitude and linear
lead backed solid state detectors do not. They are not recom- response to x-ray energy provides an image appearance that
mended for direct measurement of entrance surface dose on the remains consistent throughout the exposure range and this in turn
skin of the patient. They can, however, be used to make measure- provides little feedback to the technologist. Underexposed images
ments of the absorbed dose to air, in free air, without a patient typically have a grainy, mottled appearance that causes radiolo-
or phantom present. The measurements can then be corrected gists to reject images. Over-exposed images, on the other hand,
using appropriate backscatter factors and the inverse square law have a crisp, sharp appearance. In order to prevent repeating the
to estimate the entrance surface dose. Newer technology such as image, the technologist may increase exposure factors especially
optically stimulated luminescence dosimeters (OSLs) and radio- for manual and mobile radiography. Exposure indices or expo-
chromic film may replace TLD. Radiochromic film is currently sure indicators provided by the various CR/DR manufacturers
being evaluated by ARPANSA. also have a wide range of acceptable values and are currently not
Alternatively, the entrance skin dose may also be calculated standardised throughout the industry.
from x-ray tube output measurements (mGy/mAs) and the expo- In the case of CT examinations, care should be taken when
sure parameters, kVp, filtration and mAs. The incident air kerma following overseas DRL because of the wide variety of CT scan-
is calculated from the tube output using the inverse square law and ners and local examination protocols employed. In addition, the
then multiplied by the backscatter factor to obtain the entrance rapid advances in CT technology have also resulted in constantly
skin dose. changing scanning protocols. Nevertheless, the DRL provided in
A dose area product (DAP) meter consists of an ionisation Tables 1–5, serve as a rough guide until new DRL emerge from
meter that is usually attached to the x-ray tube collimator and current surveys in Australia.
measures the dose in Gy. square centimetre (Gy cm2) which is In future, a web based interactive dose survey software pro-
proportional to the beam area and incident air kerma. The unit gram will be provided by ARPANSA where each radiology
unfortunately does not measure backscatter which is important department can access it to calculate their dose levels and com-
in higher dose examinations such as cardiac and vascular inter- pare them with DRL.
ventional procedures. However DAP meters can be used for Discussion
radiographic and fluoroscopic procedures such as barium meals, The development of DRL practice in diagnostic radiology within
angiography and on mobile image intensifiers.2,3,28 Australia is still at an early stage as no national surveys have been
For CT machines, the CTDIw and /or CTDIvol (mGy) and the carried out for any radiological examinations for the express pur-
DLP values(mGy cm) are conveniently provided at the opera- pose of establishing national DRL. At a local level, various organi-
tor console before or after the examination. The CTDIw is the sations, regulatory authorities and individual practices have carried
weighted sum of the CT dose (or air kerma ) index measured out limited general radiography, fluoroscopy and CT surveys.34
in the centre and periphery (1 cm under the surface) of a 16 cm There is a clear need to manage (optimise) the radiation doses
diameter (head) or a 32 cm diameter (trunk) standard polymeth- from diagnostic radiology in order to minimise the risks from
ylmethacrylate (PMMA) CT dosimetry phantom. The CT dose radiation induced cancers. The establishment and use of DRL is
index is measured with a 100mm long pencil ionization chamber recommended by international radiation protection organisations
inside a standard PMMA CT dosimetry phantom. as an important component of the management of these doses
1 2 and many countries have incorporated them into their radiation
CTDIw = 3 CTDIc + 3 CTDIp where c is the centre position protection regulations12,36
and p is the peripheral position of the phantom. Units: mGy Data from European countries shows a wide variation in com-
mon DRL which may be due to differences in socio-economic
CTDIw corrected for pitch is the CTDIvol. conditions, regulatory regime, activeness of professional bod-
CTDIw ies and health care implementation (private/public mix etc).7,32
CTDIvol = Units: mGy International radiation protection bodies such as the IAEA and
pitch
ICRP therefore recommend that each country carry out its own
The DLP is the product of CTDIvol and the scan length of the national wide scale DRL survey. It is for this reason that Australia
examination. must develop its own set of common national DRL.
Thus DLP = CTDIvol x length irradiated. Units: mGy cm15 The introduction of computed and digital radiography in recent
As a starting point it is suggested that the UK 2000 survey years has had a significant impact on the potential for higher dose
Diagnostic reference levels as a quality assurance tool The Radiographer ­35

Table 1: Recommended diagnostic reference doses for individual radiographs on adult patients.
Radiograph ESD per radiograph (mGy) DAP per radiograph (Gy cm2)
Skull AP/PA 3 -
Skull LAT 1.5 -
Chest PA 0.2 0.12
Chest LAT 1 -
Thoracic spine AP 3.5 -
Thoracic spine LAT 10 -
Lumbar spine AP 6 1.6
Lumbar spine LAT 14 3
Lumbar spine LSJ 26 3
Abdomen AP 6 3
Pelvis AP 4 3
Adopted from the UK 2000 DRL survey review. 29
Note: Adult is defined as a person of average size (70–80 kg) ESD = Entrance Skin Dose, DAP = Dose Area Product.

Table 2: Recommended diagnostic reference doses for fluoroscopic/interventional examinations on adult patients.
Examination DAP per exam (Gy cm2) Fluoroscopy time per exam (mins)
Barium (or water soluble) swallow 11 2.3
Barium meal 13 2.3
Barium follow through 14 2.2
Barium (or water soluble) enema 31 2.7
Small bowel enema 50 10.7
Biliary drainage/intervention 54 17
Femoral angiogram 33 5
Hickman line 4 2.2
Hysterosalpingogram 4 1
IVU 16 -
MCU 17 2.7
Nephrostogram 13 4.6
Nephrostomy 19 8.8
Retrograde pyelogram 13 3
Sialogram 1.6 1.6
T-tube cholangiogram 10 2
Venogram (leg) 5 2.3
Coronary angiogram 36 5.6
Oesophageal dilation 16 5.5
Pacemaker implant 27 10.7
Adopted from the UK 2000 survey review.29 DAP = Dose Area Product.

delivery.30,31 In addition, the exponential increase in CT examina- required across as many radiology departments as possible. Complex
tions has lead to the unwanted outcome of a significant increase calculations will only discourage participation. Furthermore, setting
in population cumulative effective dose.35 Other causal agents that DRL is a resource intensive activity and requires a national response.
are linked to high doses include type of image receptor, exposure Priority should be given to procedures with greatest dose implica-
factors, fluoroscopic time, number of images, type of antiscatter tions, i.e. CT and Interventional procedures. DRL should be owned
grid and level of quality control.9,10,11 by the professions such as the Australian Institute of Radiography and
DRL depend significantly on local practice and equipment. the Royal Australian New Zealand College of Radiology. ARPANSA
They may also change with time as optimisation strategies become will assist in facilitating their development.
successful. The UK experience over the past 20 years has shown
that the implementation of DRL together with a dose optimisation Conclusion
program has resulted in a gradual reduction of doses.29,32 Overseas experience has shown that the use of DRL have prov-
For DRL to succeed, acceptance and application of the concept is en be a useful quality assurance tool in optimising patient dose in
­36 The Radiographer KD Edmonds

Table 3: Recommended diagnostic reference doses for complete examinations on paediatric patients.
Examination Standard age (y) DAP per exam (Gy cm 2)
MCU 0 0.4
  1 1.0
  5 1.0
  10 2.1
  15 4.7
Barium meal 0 0.7
  1 2.0
  5 2.0
  10 4.5
  15 7.2

Barium swallow 0 0.8


  1 1.5
  5 1.5
  10 2.7
  15 4.6
Adopted from the UK 2000 survey review. 29

Table 4: Recommended diagnostic reference levels for CT examinations (CTDIvol and DLP).
Patient group Scan region CTDIvol (mGy) DLP (mGy cm)
single slice/multi slice Single slice/multi slice

Adults Post fossa 65/100


Cerebrum 55/65
Whole exam 760/930

Abdomen (liver metastases ) 13/14


Whole exam 460/470

Abdomen and pelvis 13/14


(abscess) Whole exam 510/560

Chest, abdomen and pelvis 22/26


(lymphoma staging or follow up).
Whole exam 760/940

Chest (lung cancer) 10/13 430/580

Chest Hi-res 3/7


Whole exam 80/170
Head (post fossa) 35 270 (whole exam)
Children
Head (cerebrum) 30
0–1 year-old
Thorax 12 200
Head (post fossa) 50 470 (whole exam)
5-year-old Head (cerebrum) 45
Thorax 13 230
Head (post fossa) 65 620 (whole exam)
10-year-old Head (cerebrum) 50
Thorax 20 370
Adopted from UK 2003 CT dose survey. 33

Dose values for adults relate to the 16 cm diameter CT dosimetry phantom for examinations of the head and the 32 cm diameter CT dosimetry phantom for
examinations of the trunk.All dose values for children relate to the 16 cm diameter CT dosimetry phantom.
CTDI vol = Computed Tomography Dose Index Volume, DLP = Dose Length Product.
Diagnostic reference levels as a quality assurance tool The Radiographer ­37

Table 5: Recommended diagnostic reference level for mammography for a 13 ARPANSA. Safety Guide for Radiation Protection in Diagnostic and
Interventional Radiology Radiation Protection Series 2008 RPS 14.1.
typical adult patient.
14 IAEA 1996. International Atomic Energy Agency, International Basic safety
• For film screen examinations using a grid, the mean glandular dose (MGD) standards for protection against ionizing radiation and for the safety of radia-
is 2 mGy based on the 4.2 cm acrylic American College of Radiologists phan- tion sources. Safety Series No. 115, IAEA, Vienna.
tom.17,18,19 15 European Commission. European guidance on estimating population doses
• For a 50% adipose, 50% glandular 5 cm thick phantom the MGD is 3 mGy17 from medical x-ray procedures.2008. Radiation Protection Publication No 154.
Note: For digital mammography, the values quoted above represent an upper TREN/04/NUCL/S07.39241.
limit 16 National Radiological Protection Board. Guidelines on patient dose to promote
the optimisation for diagnostic medical exposures. Documents of the NRPB
1999; 10 (1).
diagnostic radiology. It is recommended that local dose surveys be
performed annually while national surveys every five years.32,33 17 Craig AR, Heggie JCP, McLean ID, Coakley KS, Nicoll JJ. Recommendations
for a mammography quality assurance program ACPSEM Position paper.
The imaging technologist is the main person who decides on Australas Phys Eng Sci Med 2001; 24 (107): 131.
the exposure factors and the visual image quality for the radi-
18 Mammographic Quality Control Manual. Royal Australian and New Zealand
ologist to make a diagnosis. The technologist should therefore be College of Radiologists 2002.
aware of the exposure options that minimise radiation doses while 19 National Accreditation Standards. Breastscreen Australia 2002; pp 157.
still maintaining good image quality and monitoring dose levels.
20 International Commission on Radiological Protection. 1990 recommendations
DRL would therefore serve as an important means of minimising of the International Commission on Radiological Protection. ICRP Publication
radiation doses as well as dose variations at minimal cost to radiology No. 60. Oxford: Pergamon; 1991.
departments. They also increase staff awareness and imaging tech- 21 HPA 2008. National protocol for patient dose measurements in diagnostic
nologists will be better equipped to deal with patient enquiries.3 radiology. Dosimetry Working Party of the Institute of Physical Sciences in
ARPANSA is responsible for carrying out national DRL Medicine. Health Protection Agency.Chilton.Didcot.
surveys in consultation with relevant stakeholders such as the 22 Commission of European Communities. European Guidelines on Quality Criteria
Royal Australian & New Zealand College of Radiology, The For Diagnostic Radiographic Images. The European Commission 1996.
Australian Institute of Radiography, Australasian College of 23 ACR Technical Standard for Diagnostic Medical Physics Performance
Monitoring of Radiographic and Fluoroscopic Equipment. Revised. Reston:
Physical Scientists & Engineers in Medicine, Australian & New
American College of Radiology; 2001: 729–32.
Zealand Society of Nuclear Medicine, Department of Health and
24 International Commission on Radiological Protection. Diagnostic reference
Aging and the various State/Territory Regulators. levels in medical imaging: review and additional advice. Available online at:
www.icrp.org/docs/DRL_for_web.pdt [verified 21st March].
Acknowledgements
25 AS/NZS 2005. Standards Australia and Standards New Zealand. Medical
The author thanks Anthony Wallace for his advice in preparing electrical equipment –Particular requirements for safety-x-ray equipment for
this article. computed tomography, AS/NZS 3200.2.44 Ed 2.1
26 ICRP. Managing patient dose in digital radiography. ICRP Publication 93
The author
Annals of the ICRP 2004; 34 (1).
KD Edmonds DCR BHA Grad Dip Pub Health
27 ICRU. Patient Dosimetry for x-rays used in medical imaging. ICRU Report 74.
References Journal of the ICRU 2005; 5 (2).
1 Gray JE, Archer BR, Butler PF, Hobbs BB, Mettler FA, Pizzutiello RJ, et al. 28 IPSM (1992). Dosimetry Working Party of the Institute of Physical Sciences
Reference values for diagnostic radiology: application and impact. Radiology in Medicine. National protocol for patient dose measurements in diagnostic
2005; 235: 354–8. radiology. Chilton, UK: NRPB.
2 IAEA 2002 .Radiological Protection for Medical Exposure to Ionizing 29 Hart D, Hillier MC, Wall BF. Doses to patients from medical x-ray examina-
Radiation. IAEA Safety Standards Series. Safety Guide No RS-G-1.5. Vienna, tions in the UK-2000 review. Chilton: NRPB 2002; NRPB-W14.
Austria: International Atomic Energy Agency. 30 Warren-Forward H, Arthur L, Hobson L, et al. An assessment of exposure
3 Seeram,E, Brennan P C, Diagnostic reference levels in radiology. Radiologic indices in computed radiography for the posterior-anterior chest and the lateral
Technology 2006: 77: 373–84. lumbar spine. Br J Radiol 2007; 80: 26–31.
4 ICRP. Radiological Protection and Safety in Medicine. ICRP Publication 73, 31 Willis C. Strategies for dose reduction in ordinary radiographic examinations
Annals of the ICRP 1996; 26 (2). using CR and DR. Pediatr Radiol 2004; 34 (Suppl 3): S196–S200.
5 Regulla DF, Eder H. Patient exposures in medical x-ray imaging in Europe. 32 Hart D, Hillier MC,Wall BF. National reference doses for common radio-
Radiat Prot Dosimet 2005; 14: 11–25. graphic, fluoroscopic and dental x-ray examinations in the UK. Br J Radiol
6 Conference of Radiation Control Program Directors. Nationwide Evaluation of 2009: 82: 1–12.
X-ray Trends (NEXT): Tabulation and Graphical Summary of 1995 Abdomen 33 Shrimpton PC,Hillier MC,Lewis MA,Dunn M. National survey of doses from
and Lumbosacral Spine Survey 2000. CRCPD Publication 00-2. CT in the UK: 2003. Br J Radiol 2006: 79: 968–80.
7 Johnston DA, Brennan PC. Reference dose levels for patients undergoing com- 34 Boal T, Hedt J, Einsiedel P. A survey of patient dose and image quality factors
mon diagnostic x-ray examinations in Irish hospitals. Br J Radiol 2000: 73: for CT scanners in Victoria. Aust Phys Engineer Sci Med 1999; 22 (3): 103–12.
396–402. 35 Brenner DJ, Hall EJ. Computed tomography – an increasing source of radia-
8 Carroll EM, Brennan PC. Radiation doses for barium enema and barium meal tion exposure. New Engl J Med 2007; 357 (22): 2277–84.
examinations in Ireland: potential diagnostic reference levels. Br J Radiol 36 Office for Official Publications of European Communities, Luxemburg, 1997.
2003; 76: 393–7. European Commission, Council Directive 97/43/Euroatom of June 1997 on
9 Seeram E. Rad Tech’s Guide to Radiographic Protection. Boston: Blackwell the health protection of individuals against the dangers of ionizing radiation
Science; 2001. in relation to medical exposure. Official Journal of the European Communities
L180/22 1997.
10 Bushong S. Radiologic Science for Technologists 8th ed. St Louis: Mosby; 2004.
37 Wall BF, Shrimpton PC. The Historical development of reference doses in
11 Parry RA, Glaze SA, Archer BJ. Typical patient radiation doses in diagnostic diagnostic radiology. Radiat Prot Dosimet 1998; 80 (1–3): 15–9.
radiology. Radiographics 1999; 19: 1289–1302.
38 EC Study Group. Quality Criteria for Diagnostic Radiographic Images.
12 ARPANSA. Code of Practice for Radiation Protection in the Medical Working Document, 2nd ed, CEC XII/173/90 (Brussels: EC) 1990.
Applications of Ionizing Radiation, Radiation Protection Series 2008 RPS 14.

Das könnte Ihnen auch gefallen