EyeSuite Perimetry
Setup Guide
© Haag-Streit AG #7220060 (Rev 3, 12/2007)
HAAG-STREIT AG, Switzerland, Phone: (+41-31) 978 0111, Fax: (+41-31) 978 0282, info@haag-streit.ch
HAAG-STREIT AUSTRIA GmbH, Austria, Phone (+43-1) 895 0364, Fax (+431) 895 0368, info@haag-streit.at
HAAG-STREIT DEUTSCHLAND GmbH, Germany, Phone: (+49-4103) 709 02, Fax: (+49-4103) 709 370, info@haag-streit.de
HAAG-STREIT FRANCE, France, Phone (+33-4) 5009 0033, Fax (+33-4) 5009 7190, info@haag-streit.fr
HAAG-STREIT UK, United Kingdom, Phone (+44-1279) 414969, Fax (+44-1279) 635232, info@haag-streit-uk.com
HAAG-STREIT USA, INC., USA, Phone: (+1-513) 336 6858, Fax: (+1-513) 336 7828, octopus@haag-streit-usa.com
OCTOPUS 900 & EyeSuite Perimetry Setup
CONTENTS
1 General Issues 3
1.1 Requirements 3
1.2 Included materials 3
1.3 Disclaimer 3
1.4 Safety measures 3
2 Software Installation 4
1.1 Requirements Haag-Streit has taken all required efforts to protect and
safely handle any test result. The reliability of a PC
Minimum PC requirements for EyeSuite Perimetry are environment is depending on many factors including
hardware components, operating systems or even
• Windows XP SP2 (Home or Professional) computer viruses that may cause loss of data under
• P4 or compatible processor certain conditions.
• 1GB RAM
Haag-Streit thus strongly recommends to perform a
• 40GB free hard disk space
regular backup of all examination data to external media –
• CD-ROM or DVD drive
for example CD-R or Flash memories.
• Screen resolution 1280 x 1024 pixels
(17" recommended) Haag-Streit AG does not warrant for any consequences,
complications or data loss caused by the usage of and/or
For the operation of an Octopus 900: decisions based on this software. By applying this
• T100 Ethernet interface software you agree with this disclaimer.
All HAAG-STREIT AG software including
If the instrument runs in a network environments EyeSuite™Perimetry is copyright protected by
• Network Switch HAAG-STREIT AG.
2 SOFTWARE INSTALLATION
[mysqld]
basedir=<installation folder>\
datadir=<installation folder>\data\
Run "Setup.bat"
If you would like to access perimetry data from multiple In every viewing station perform the following steps
PCs you need to make a network installation. In the
8. Follow the procedure in EyeSuite Perimetry CD-
addendum 6 you find the templates for various
ROM
configurations to document your environment.
9. Open the folder C:\program files\EyeSuite in
Windows explorer explorer
2.2.1 Small network: Control PC acts as Server 10. Rename the file EyeSuiteOctopus.jar to
It is required that a local PC controls the Octopus 900. EyeSuiteOctopus.aaa
Local means, it needs to be on the same network switch. 11. Do a right click on "Start", "EyeSuite", "EyeSuite" (or
If you have a small network you might want to use the OC on the shortcut on the desktop) and click on
controlling the Octopus 900 also as a server for your "Properties"
perimetry data. The according configuration is
documented in Figure 5.
3 CONNECTING AND UPDATING THE OCTOPUS 900 To configure the Octopus 900 you might need to
temporarily change the IP address of your PC to
172.16.nnn.mmm. The change of the IP address requires
The Octopus 900 has the default IP address
Administrator rights and can be done under "Start",
172.16.42.32. Commands are sent from the PC to the
"Control Panel", "Network Connections". Please contact
Octopus using the TCP protocol. The video stream from
the responsible person for your network to define the final
the Octopus is transmitted from the Octopus 900 to the
static IP address of your Octopus 900.
PC using the UDP protocol.
To change the IP address you need to configure your Please refer to your computer's operating system manual
computer to be in the same address range for further details on setting the computers IP address.
172.16.nnn.mmm with the subnet mask 255.255.0.0.
nnn and mmm can be any number between 1 and 254. To change the IP address of the Octopus 900
Attention: To separate the UDP video stream which takes a 1. Make sure your computer's IP address is in the
considerable amount of the network bandwidth from the same range as the Octopus 900's IP address
rest of your network you have to use a network switch
between the Octopus 900, the controlling PC and the 2. Connect the Octopus 900 and your PC either
Ethernet wallplug (the connection to the network).
directly via a crossed Ethernet cable or via a
If the Octopus 900 and the Control PC are directly network Switch using straigt Ethernet cables.
connected to the network, the network may become
slow or even hang. 3. Run EyeSuite Perimetry, go to "Perimetry",
"Service / Diangostic", "Connection"
4. For the Flash update select the "*.bin" file, for the
7. Wait about 30 seconds until you heard the FPGA update select the "*.pom" file and click
instrument gave one high and 3 low beep sounds "Open"
8. Change the IP address under "OCTOPUS
connection", "IP address" and click on "Connect
to OCTOPUS" to reconnect again. It may be
required that you change the IP address of your
own computer to become part of the same
network address range again
3. When you click "Print", the PC-ID is shown on 6. Enter the dongle code and click "Save". You
the right side and a form is printed out. will then need to restart EyeSuite Perimetry.
EyeSuite Perimetry directly imports PeriTrend static and EyeSuite Perimetry does support sorted exportation of
GKP kinetic examinations. results for statistical analysis as well as interface functions
to third party software. These functions will be extended in
the future.
To import existing data from Peritrend go to "Import",
"Import examination data" and click the button "…" 5.4.1 PeriData export
Peridata is a perimetry analysis software by Dr. Jörg
Weber, Germany.
In order to export visual fields to PeriData go to "Export",
"Setup", "Perimetry".
6 ADDENDUM
Figure 4: Always set a network switch between the Octopus 900 and the PC. This ensures correct data flow.
Figure 5 Never attach the Octopus 900or its control PC to a hub or to a wall plug directly. Always use a local network switch for connection.
Figure 6 Never attach the Octopus 900or its control PC to a hub or to a wall plug directly. Always use a local network switch for connection.